Introduction

Dental implants have been used to rehabilitate edentulous and partially edentulous patients with fixed prostheses for over 30 years.1 Systematic reviews have reported relatively high survival rates for dental implants supporting crowns2 and fixed bridgework.3,4,5,6,7 However, these reviews have also identified a variety of complications that may be associated with these prostheses.2,3,4,5,6,7 Although these complications rarely result in the complete failure of treatment, their management can be frustrating, expensive and time consuming for the patient, clinician and technician. This review discusses the incidence, aetiology, prevention and management of common prosthodontic complications, which may occur following the delivery of implant retained crowns and bridgework.

Complications

Prosthodontic complications can be broadly categorised as biological, mechanical and aesthetic.5 Biological complications affect the tissues supporting the implant and prosthesis.4 Commonly reported biological complications include peri-implant mucositis, peri-implantitis, fistulae at the level of the implant-abutment interface, soft tissue hyperplasia and implant loss.3,4 Mechanical complications affect the structural integrity of the implant, abutment or superstructure.6 Commonly reported mechanical complications include fracture of the veneering material or framework, loss of retention, screw loosening, screw fracture and implant fracture.2,3,4,5,6,7 Aesthetic complications affect the appearance of the restoration and may be associated with the prosthesis itself or the surrounding soft tissues.

Multiple studies reporting prosthodontic complications associated with implant retained crowns and bridgework have been published.2,3,4,5,6,7 These studies frequently vary in the definition of what constitutes a complication and rarely include all of the complications outlined in this article. Furthermore, these studies are of variable design and quality and use a variety of implant systems and components, many of which are not currently available.2,4,7 Unsurprisingly, reported complication rates vary widely.8,9 This heterogeneity has prevented the calculation of precise complication rates and as a result, published systematic reviews have reported trends3,4,5 or complication rates with wide confidence intervals.2,4,6,7 Therefore, published complication rates should be interpreted with caution.

Biological complications

Biological complications refer to disturbances in the function of the implant characterised by biological processes that affect the tissues supporting the implant.4 Common biological complications include the following.

Peri-implant mucositis

Peri-implant mucositis is a reversible inflammatory lesion, analogous to gingivitis, affecting the marginal soft tissues surrounding osseointegrated dental implants and is not associated with peri-implant bone loss (Fig. 1).10 Peri-implant mucositis may affect up to 50% of implants and is characterised by bleeding on probing.11 Peri-implant mucositis usually resolves with improved homecare, antiseptic mouthwash and careful mechanical cleaning.11,12 A lack of keratinised mucosa may impair oral hygiene procedures and increase the incidence of peri-implant mucositis.13

Figure 1
figure 1

Peri-implant mucositis associated with implant retained crown UR1

Peri-implantitis

Peri-implantitis is an inflammatory lesion, analogous to periodontal disease, affecting the supporting bone in addition to the soft tissues around an implant.10,14 Peri-implantitis is characterised by marginal bone loss and may be associated with bleeding on probing, suppuration, pocket formation and mucosal recession.11 Marginal bone loss during the first year of function is attributed to remodelling and peri-implantitis cannot be diagnosed until after this time.10 It has been suggested that peri-implantitis may be diagnosed when the bone adjacent to ≥3 threads of an implant is lost after one year in function.10 However, as implant designs and distances between threads vary, this definition would appear unsatisfactory. Putative risk factors for peri-implantitis include smoking, diabetes mellitus, poor oral hygiene, a history of periodontal disease, interleukin-1 genotype, alcohol, rough surfaced implants, retained cement, (Figs 2 and 3) and poorly fitting abutments or superstructures.3,14,15 A lack of keratinised mucosa does not appear to be associated with an increased incidence of peri-implantitis.13 Jung et al.2 reported that 6.3% of implant supported single crowns had bone loss >2 mm over a 5 year observation period. Pjetursson et al.7 reported that less than 8.6% (95% CI; 5.1%-14.1%) of implant supported fixed partial dentures were affected by peri-implantitis over 5 years. However, much higher incidences have been reported in other studies.11

Figure 2
figure 2

Retained cement associated with implant crowns UL1 and UL2

Figure 3
figure 3

Peri-implant bone loss UL1 and UL2 visible following mucoperiosteal flap elevation

Following the provision of an implant retained crown or bridge baseline peri-implant pocket depths and radiographic bone levels should be established.11 Gentle probing (≤0.25 N) is unlikely to damage the implant surface or the epithelial attachment and should be repeated annually.10,11,14 Studies suggest that future attachment loss is unlikely to occur in the absence of bleeding on probing.14 Probing depths are generally greater around implants than teeth and a probing depth of 6 mm is not uncommon in health.16 Periapical radiographs should not be routinely prescribed during follow-up and should only be taken when changes in clinical parameters indicate disease.17 Radiographs should be interpreted cautiously because they only show interproximal bone levels and have poor reproducibility.14,18

There is little evidence to favour either the non-surgical or surgical management of peri-implantitis.19 However, a surgical approach is often preferred as it offers improved access.14 Implant surfaces may be cleaned mechanically, chemically, by laser or air abrasion.20 No single method of surface debridement has been shown to be superior.11,19,20 Where mechanical debridement is undertaken, plastic or titanium curettes have been recommended to avoid damaging the implant surface.10 However, Romeo et al.21 have reported promising results following the deliberate removal of the implant surface. Guided bone regeneration may be used in conjunction with surgical debridement and this may result in greater bony infill although re-osseointegration is unlikely to occur.20 Studies suggest that surgical debridement may be successful in 60-100% of cases.20,21

Fistulae

Fistulae have been reported at the level of the implant-abutment interface in 1% of cases (study duration not reported)5 (Fig. 4) and they are usually associated with loose abutments screws or ill-fitting frameworks.3 These fistulae usually resolve following removal of the prosthesis, irrigation with 0.2% chlorhexidine gluconate (Corsodyl, GlaxoSmithKline) and replacement of the prosthesis at the correct torque. When a prosthesis is found to be ill-fitting it should be remade. When fistulae occur in conjunction with temporary metal cylinders, corrosion products may cause permanent discolouration of the soft tissues (Fig. 5).

Figure 4
figure 4

Fistula associated with a loose abutment screw UL2

Figure 5
figure 5

Healed fistula with permanent mucosal discolouration associated with temporary implant retained crown at UR2 site

Soft tissue hyperplasia

Hyperplasia of the peri-implant tissues has been reported to affect up to 20% of fixed prostheses over a 9 year observation period.3 This hyperplasia represents an over-exaggerated response to plaque accumulation and may be associated with poor oral hygiene, framework misfit, dead space beneath frameworks and a lack of attached gingivae.3 Hyperplasia usually resolves following correction of predisposing factors, improved oral hygiene and mechanical debridement. However, it may occasionally be necessary to surgically resect the hyperplastic tissue (Figs 6,7,8,9). Figures 6,7,8,9 show a case which required surgical resection following lack of resolution after initial debridement and improvement of oral hygiene. The use of electrosurgery is discouraged in these situations as there may be accidental heating of the implant-bone interface.

Figure 6
figure 6

Hyperplasia of the peri-implant soft tissues associated with the distal aspect of an ill-fitting lower implant retained bridge

Figure 7
figure 7

Radiograph of case shown in Fig. 6. Superstructure is ill fitting at LR3, LL3 and LL4 sites

Figure 8
figure 8

The full extent of soft tissue hyperplasia is revealed following removal of the superstructure shown in Fig. 6

Figure 9
figure 9

Resolution of the soft tissue hyperplasia shown in Fig. 8

Implant loss

Implant losses are categorised as early or late.3 Early implant loss occurs before functional loading and results from a failure to establish osseointegration.18 Early implant loss has been associated with overheating during implant placement, infection, poor bone quantity and poor bone quality.18 Late implant loss occurs following functional loading because of a breakdown of an established osseointegration.18 Late implant loss has been associated with occlusal overload, peri-implantitis, poor bone quantity and poor bone quality.18,22 The reported survival rates for implants supporting crowns and fixed partial dentures exceed 95% over 5 years2,6,7 and 93% over 10 years.6,7 However, survival does not equate to success and these rates must be interpreted cautiously because a surviving implant may be failing, unrestorable or may have had extensive remedial treatment.16 When implant loss occurs, it occurs before functional loading in approximately 55% of cases.4,22 When an implant fails an alternative treatment may be considered, the site may be allowed to heal before placing a new implant (Figs 10 and 11) or a new, wider implant may be placed immediately.23 There is limited evidence to suggest that implants placed at the site of previously failed implants have lower survival rates.23 However, this may simply reflect poor initial case selection.

Figure 10
figure 10

Failing implant UL2

Figure 11
figure 11

New implant placed following a 6 month period of bony healing. Midline diastema has been closed with direct composite

Mechanical complications

Commonly reported mechanical complications associated with implant supported fixed prostheses include fracture of the veneering material or framework, loss of retention, screw loosening, screw fracture and implant fracture.2,6,7 The incidence of these complications associated with implant retained single crowns,2 and implant retained fixed partial dentures7 with or without cantilever extensions,6 after five years, is presented in Table 1.

Table 1 Reported incidence of mechanical complications associated with implant retained single crowns and implant retained fixed partial dentures with, or without, cantilever extensions

Mechanical complications occur when the capacity of the prosthesis to withstand applied forces is exceeded. The precise complication will depend upon the magnitude and the direction of the applied force. However, the weakest parts of the system, that is, the screws and veneering material, are usually affected (Table 1). Implant supported prostheses are believed to be particularly susceptible to mechanical complications because implants have limited proprioception and adaptive capacity.24 Excessive loading has been associated with non axial forces, cantilevers, parafunction and increased implant-abutment ratios.25,26,27 Mechanical complications have also been associated with the inappropriate selection or use of materials, manufacturing imperfections, ill-fitting frameworks and trauma.3,28,29

In partially edentulous patients implant supported prostheses should be designed so that they allow light occlusal contact in the intercuspal position. This allows for compression of the periodontal ligament of adjacent teeth during function and reduces excessive occlusal loading on the implants.24 Furthermore, wherever possible, occlusal forces should be directed axially, occlusal tables should be narrow with shallow cuspal inclines, cantilevers should be minimised and eccentric excursions should be guided by natural teeth.24,27 Following treatment, hard nocturnal splints have been advocated to reduce the incidence of mechanical complications.25 However, these are only likely to be effective if patient compliance is good. Frameworks should be constructed from appropriate materials, be of adequate thickness and designed appropriately to support the veneering material which may be porcelain, acrylic or composite resin.3,28 Frameworks milled using CAD-CAM (computer aided design-computer aided manufacture) technology are preferable as they are not susceptible to distortion or casting porosity and have improved marginal fit when compared to cast gold frameworks.30 However, the accuracy of a CAD-CAM framework is still dependent upon the accuracy of the cast model, the scanning technique and the computer based designing capabilities of the dental technician.

Fracture of the veneering material or framework

Fracture of the veneering material is one of the most common mechanical complications reported in the literature (Figs 12 and 13). Early studies reported incidences as high as 30% over 3 years,3 although this incidence has reduced (Table 1). Superstructures may be veneered with porcelain, acrylic or composite resin. Porcelain maintains appearance, is least likely to fracture and is often considered the veneering material of choice.3,5,7 Fracture of the underlying framework is less common (Table 1, Fig. 14).5

Figure 12
figure 12

Fractured porcelain UL1

Figure 13
figure 13

Screw retained implant bridge with fractured composite veneer UR1

Figure 14
figure 14

Midline fracture of implant retained bridge replacing LR3-LL3. Metalwork is too thin in cross-section

When a superstructure fractures it is important to address the cause of fracture before any remedial treatment (Fig. 14). Small fractures in non aesthetic areas can be smoothed while larger fractures usually require repair. Minor fractures can be repaired intra-orally using metal-ceramic repair systems, such as Cojet (3Mâ„¢ESPEâ„¢). However, the clinical performance of these repairs is uncertain.43 It is usually preferable to remove fractured superstructures and repair them in the laboratory. This process is facilitated when restorations are screw-retained.15 If provisional restorations have been retained, they can be replaced to maintain aesthetics and function while repairs are undertaken. Porcelain absorbs water intra-orally and may crack when re-introduced into a ceramming furnace. Therefore, superstructures veneered with porcelain must be thoroughly dried in a low temperature desiccator before attempting repair.32 Complex restorations with individual crowns cemented on to implant retained superstructures have been described in the literature.33 These restorations are designed so that individual crowns can easily be replaced in the event of fracture, however, they are expensive and complex to manufacture. When recurrent veneer fracture occurs occlusal and palatal surfaces may be constructed in metal.

Loss of retention

Loss of retention is a complication specific to cement retained prosthesis. Loss of retention is commonly associated with abutments that provide inadequate retention or resistance form (Fig. 15). This usually reflects incorrect abutment selection, poor abutment design or a lack of restorative space. The amount of restorative space should be assessed before implant placement. When restorative space is limited, provided bone volume is adequate, consideration should be given to deeper placement of the implant.

Figure 15
figure 15

Decemented crown. Custom abutment has excessive axial taper

Screw loosening and screw fracture

Screw loosening is a commonly encountered complication, as indicated in Table 1. Most studies do not differentiate between prosthesis and abutment screws, therefore they are considered together in this article. The shanks of abutment and prosthetic screws are designed to stretch upon insertion. The tension generated by stretching is termed preload and maintains the integrity of the joint.34 Screw loosening usually occurs as a result of micromovement at the joint interface. This micromovement may be associated with an inadequate initial torque, ill-fitting frameworks and occlusal overload (Fig. 16).3 Early studies reported very high incidences of screw loosening,3 however, alterations in screw design, materials and surface coatings in conjunction with a move towards implants with internal connections appear to have considerably reduced the incidence.28 Screws should be inserted at the correct torque and reused with care because their mechanical properties may have altered.34 Fractured screws are relatively unusual (Fig. 17) and may be associated with over-torquing, occlusal overload and ill-fitting frameworks.3 Fractured screws can often be unwound with a probe or an ultrasonic scaler. However, it is occasionally necessary to use a proprietary screw removal kit or cut a slot in the coronal aspect of the fractured screw to allow the engagement of a screwdriver. Care should be taken to avoid damaging the internal threads of the implant when the slot is prepared and magnification is advised.

Figure 16
figure 16

Loose abutment screw UL3 implant retained crown. UL3 provides sole guidance in left lateral excursion

Figure 17
figure 17

Fractured abutment screw

Implant fracture

Implant fracture is a rare occurrence, which usually affects hollow or narrow diameter implants.4,29 Implant fracture is usually catastrophic and necessitates removal or submergence of the fractured fragment. Occasionally the fractured implant may be recontoured and salvaged.29

Aesthetic complications

Aesthetic complications detract from the overall appearance of the implant supported prosthesis. Missing interdental papillae (Fig. 18), mucosal recession, poor restoration contour and shade mismatch may all cause aesthetic failure.35 Implant supported restorations have been considered successful when they mimic adjacent natural teeth in a well balanced soft tissue framework.36 However, aesthetic outcomes are highly subjective and care should be taken to avoid imposing our own aesthetic parameters on patients. Studies have revealed that there is often poor agreement between patients and professionals regarding aesthetic outcome37 and that patients are usually highly satisfied with the aesthetic outcome.38,39 Several reviews have reported an overall aesthetic complication rate approaching 10%.2,5 However, neither of these reviews reported the criteria used to assess appearance.

Figure 18
figure 18

Absent interdental papilla between implants at UR1 and UR2 sites

The final aesthetic outcome is largely dependent upon the pre-existing state of the hard and soft tissues,35,40,41 the care with which surgical procedures are performed42,43 and lip position (Fig. 19). Operator experience and patient selection are important factors and it is vital that the implants are placed in the correct three dimensional position. Careful preoperative assessment can usually inform us as to the likely cosmetic result. The patient's existing prosthesis, photographs and rehearsals can be used to communicate this to the patient. Where deficiencies are identified with the hard and soft tissues preoperatively, augmentation procedures may be considered, or the patient may elect to accept a compromised result.42,43

Figure 19
figure 19

Smile line of patient in Fig. 18

Patients complaining of poor aesthetics following the completion of treatment can be challenging to manage. Mucogingival grafting44 and the use of pink porcelain may be useful in these cases. However, inadequate treatment planning and poor implant positioning can often not be disguised (Fig. 20).

Figure 20
figure 20

Implant retained crown UL1: a disastrous result in a young girl with a high lip line

Conclusions

Although implant retained crowns and bridgework are highly successful they may be associated with a variety of biological, mechanical and aesthetic complications. The precise incidence of these complications is unclear and future studies should report complications according to standardised criteria. Clinicians providing implant treatment should be aware of these potential complications and the strategies by which they can be prevented and managed. Potential complications should be communicated to the patient before commencing treatment as part of the informed consent process. Clinicians are encouraged to audit their own work and determine their personal complication rates.