Introduction

Neurological disorders are among the world’s leading causes of disability; among these, Parkinson’s disease (PD) is the fastest-growing neurodegenerative disease with a prevalence of more than 6 million afflicted individuals worldwide, projected to more than double in coming decades1,2. PD is characterized by loss of dopamine transporter (DaT) sites in striatum, with cardinal motor symptoms of bradykinesia, tremor, rigidity, and postural instability. Besides these signs, however, a wide spectrum of manifestations include non-motor symptoms (e.g., constipation, dysphagia, sleeping disorder, depression). The sole treatment consists of symptomatic interventions intended to raise dopamine availability. The most common medication is the amino acid L-dihydroxyphenylalanine (L-DOPA) that transports across the blood–brain barrier and undergoes conversion to dopamine in neurons equipped with the appropriate enzyme3,4.

PD primarily is a clinical diagnosis, but so-called functional single-photon emission computed tomography (fSPECT) can aid the diagnosis with the 123I-labeled cocaine analogue ioflupane (N-3-fluoropropryl-2β-carbomethoxy-3β-(4-iodophenyl)nortropane ([123I]FP-CIT) that binds with high affinity to presynaptic DaT proteins5. The imaging reveals neurodegeneration by loss of dopaminergic neurons in the nigrostriatal pathways as indication of PD3,5. The EANM/SNMMI guidelines for dopaminergic imaging of parkinsonian syndromes state that DaT SPECT uses a gamma camera equipped with parallel-hole collimators at 3–6 h post-injection (p.i.)5. The evaluation is based mainly on visual interpretation although semiquantitative measures often are used to aid the assessment.

With the recent introduction of multi-pinhole collimators into the clinical setting, large gains of sensitivity and resolution significantly may improve overall image quality and visual interpretation6. Initial tests with a three-headed SPECT/CT camera (AnyScan Trio SC, Mediso Medical Imaging System, Hungary) with multi-pinhole collimators, designed specifically for brain imaging, reveal a 3.9-fold increase of peak spatial sensitivity (700 cps/MBq) and a spatial resolution of approximately 5 mm half that of conventional two-headed SPECT with parallel-hole collimators7,8. The improved resolution lowers partial volume effects (PVE) and improves delineation of striatum while the higher sensitivity enables dynamic scanning and time framing with the potential to more accurately reveal tracer dynamics and hence the pharmacokinetics of [123I]FP-CIT.

We designed the present study to explore the clinical feasibility of dynamic multi-pinhole fSPECT to investigate pharmacokinetic modelling of [123I]FP-CIT in a series of patients suspected of PD, compared to conventional clinical imaging. We completed the study also as an exploration of the technical abilities of the multi-pinhole system. A primary objective was thus to examine the feasibility of dynamic DaT SPECT (fSPECT). Additional objectives included the acquisition of dynamic and static measurements that would aid the diagnostic process, and the study of tracer dynamics with the goal of optimizing the conventional imaging protocol.

Materials and methods

Subjects

Thirteen patients (three women) referred to the department for imaging by DaT SPECT underwent acquisitions of dynamic brain scans of [123I]FP-CIT uptake by means of fSPECT. The patients gave informed consent prior to participation.

Procedures and image acquisition

We imaged patients with the three-headed Mediso AnyScan Trio SC SPECT/CT device equipped with three multi-pinhole collimators. Following low-dose CT (120 kV, 20 mAs, 1 s rotation, 20 mm collimation, 1.5 pitch), we intravenously injected the patients with 120 MBq [123I]FP-CIT at the time of initiation of the dynamic fSPECT protocol. The protocol included the following steps: At 0–20 min, we completed four 5-min SPECT acquisitions (45 projections, 20 s/projection), immediately followed by repeated 10-min SPECT acquisitions (60 projections, 30 s/projection) as frequently as accepted by the patient. After a break, we planned to achieve at least two 10-min SPECT acquisitions beginning at about two hours post-injection (p.i.) to avoid large time gaps in the data sets. After the break, we redid the low-dose CT scan before continuing the SPECT protocol. At three hours p.i., we initiated the imaging of the patient according to standard clinical SPECT procedures (15 min, 90 projections, 30 s/projection). After the first five patients, we increased the number of projections. The dynamic protocol then consisted of 4 × 6-min fSPECT (60 projections, 18 s/projection) followed by repeated 12-min fSPECT acquisitions (72 projections, 30 s/projection) for as long as tolerated by the patient, or till the end of the planned protocol.

Post-processing

All image acquisitions underwent reconstruction utilizing Interview XP 3 Workstation (Mediso Medical Imaging System) with the Tera-Tomo Q reconstruction framework developed by Mediso. This framework utilizes a regularized ordered subset expectation maximization (OSEM) reconstruction method, which integrates a comprehensive model of the entire system geometry and accounts for the physical interactions, including scatter and attenuation, of gamma photons through Monte Carlo simulation. Transaxial images were reconstructed with regularization filter configured to adaptive bilateral high9, with 54 iterations with 3 subsets, while transaxial image dimension were defined to 108 × 108 pixels, Monte Carlo quality was set to medium. The resulting transaxial images had an isotropic voxel size of 2.13 mm. The acquired field of view measured 19 cm in the axial direction and 23 cm in the transaxial direction. All reconstructions underwent quantitative analysis, providing decay-corrected values in Becquerels per milliliter (Bq/ml), facilitating comparison of tracer uptake across various image acquisitions within and among patients.

Image analysis

Senior nuclear physicians (PG and ZA) interpreted the 3-h p.i. image, independently, with any differences resolved by discussion. Interpretation rested on the specific binding ratio calculated by the commercially available DaTQUANT™ program (GE Healthcare, Brondby, Denmark), SBRDaTQUANT, as mainly based on the visual grading system of Kahraman et al.10 with grade 5 indicating normal tracer uptake and lower grades indicating lower tracer uptake in striatum. For the analyses, we divided the patients into two groups, based on the 3-h p.i. interpretations as low-score (LS, grades 1–3) and high-score (HS, grades 4–5) groups.

Using Inveon Research Workplace 4.2 (Siemens Molecular Solutions, Knoxville, Tennessee, USA), we adjusted for movement between frames by manual co-registration of the different SPECT images. We assumed a standard left and right striatal volume of 8.7 ml on either side, based on healthy controls reported previously11,12. As illustrated in Fig. 1, we created several volumes (V) of interest, beginning with a right-sided standard volume of striatum, VS. The same volumes were applied to all patients. We manually aligned the standard striatal volume VS to encompass the maximum activity concentration of striatum. We created a new volume around VS by adding a 10 mm wide margin around VS in all directions (Vmargin), intended to include spill-over from PVE and possible minor movement artefacts and would lower the impact of inter-individual differences in striatal volume. Vmargin did not include VS. To correct the Vmargin for activity from spill-in from the surrounding tissue, we created a rectangular-shaped background volume (BG) around Vmargin. We divided the VS into anteromedial and posterolateral volumes to delineate the caudate nucleus and putamen, respectively. We drew all right-sided volumes on images from the first patient, mirrored for the contralateral hemisphere and used as template for all patients going forward, and adjusted the VOIs placement and orientation individually, without changing the size or shape. In addition, we drew reference volumes in cerebellum and occipital cortex, taking care not to include venous sinuses or skull. Using an modified version of the partial correction technique of Tissici-Bolt13 the mean activity concentration of striatum (CS) was adjusted for spill-over effects by means of the relationship,

$$C_{{S\left( {adj.} \right) }} = C_{S} + \frac{{V_{margin} }}{{V_{S} }}\left( {C_{margin} - C_{BG} } \right)$$
(1)
Figure 1
figure 1

Volumes of interest as shown on a high-score patient scan. Pink: Vcaudate nucleus. Red: Vputamen. VS = Vcaudate nucleus + Vputamen. White: Vmargin. Blue (rectangularly shaped): VBG. Green: Occipital cortex. Orange: Cerebellum.

We adjusted the activity in putamen and caudate nucleus by using Eq. (1) with only half of volume Vmargin, as the volumes of putamen and caudate nucleus each constituted approx. 50% of the striatal volume.

We calculated the specific binding ratio (SBRcalc) with mean activity concentrations from Eq. (1) and occipital cortex as the reference,

$$SBR_{calc} = \frac{{C_{{S\left( {adj.} \right)}} - C_{occipital} }}{{C_{occipital} }}$$
(2)

We also extracted SBRDaTQUANT values from 3-h images as done in the clinical setting.

The method was initially validated utilizing an anthropomorphic basal ganglia phantom provided by RSD (Radiology Support Devices Inc., Long Beach, CA, USA). The discrete volumes corresponding to the striatal and brain regions of the phantom were filled and imaged twice with varying 123I activity concentrations, resulting in true specific binding ratio (SBR) values of 1.0, 2.5, 3.9, and 9.2. Employing Eqs. (1) and (2), the observed SBR values were determined as 0.9, 2.3, 3.8, and 8.0. The accuracy of the measured brain volume activity concentration (Bq/ml) was confirmed to be within 5% in both imaging sessions.

Dynamic parameters

The binding potential of bound ligand relative to non-displaceable reference activity (BPND) is the ratio, at equilibrium, between the concentrations of receptor-bound and non-displaceable (non-specifically bound and free radioligand)14 in a specific region. We calculated the value of BPND by means of the simplified reference tissue model (SRTM)15 with the occipital cortex as the specified reference region. We also used the SRTM to calculate the relative index of tissue perfusion (R1) as the ratio between the tracer clearance from plasma to tissue (K1) for the striatum and the corresponding rate constant for the occipital cortex (K1). The third SRTM parameter, k2, represents the rate constant of tracer removal from tissue to plasma. We estimated the SRTM parameters (R1, k2, and BPND) from the measured data with least-square method lmfit implemented in Python with weighted data points, based on the duration of image acquisitions.

Using the model, we obtained estimates of the variables BPND, R1, and k2 for striatum, as well as for putamen and caudate nucleus separately, of left and right hemispheres, respectively.

Statistical analyses

We reported all continuous variables as median (min;max) unless otherwise stated. We reported categorical variables by frequencies (percentages). We employed spaghetti plots for visualization purposes per patient, and we applied exploratory linear regression when deemed appropriate. For the regressions of the SRTM model and analyses of pharmacokinetic parameters, χ2 values, and standard errors, we used the lmfit module in Python. We analyzed SBRcalc and completed all graphical presentations with Microsoft EXCEL 2016. We calculated ratios between striatum of both hemispheres by dividing the lowest left or right value by the highest value of the left and right striatum.

Ethics approval and consent to participate

The study protocol received approval from the institutional review board (Approval Code: 21/39019), but as the study did not entail intervention or storing of tissue, approval from the Danish ethics committee was waived. Approval from local ethics committee was not necessary as the study qualifies as a quality assurance study16. All procedures and methods were conducted in strict adherence to applicable guidelines and regulations. Overall, the study was carried out in full compliance with the principles outlined in the Helsinki Declaration. Informed consent was obtained from all individual participants included in the study.

Results

Subjects

We tested all patients for the presence of PD. Patient median age was 73 years ranging from 55 to 83 years. We categorized four patients as LS; one patient as grade 2 and three patients as grade 3. Nine patients were catergorized as HS; two patients as grade 4 and seven patients as grade 5. Ten patients (77%) had tremor-dominant symptoms with an equal distribution of left and right lateralization, all LS categorized patients had tremor-dominant symptoms. Three patients (23%) had other primary symptoms, including sleep behavior disorder and motor symptoms besides tremors, or a familial disposition of first degree. Based on referral information, at least two patients had verified vascular comorbidity, including one patient with MRI-verified lacunar infarcts in the caudate nucleus, and one patient suffered from sequelae from a prior hemorrhage outside striatum.

Dynamic acquisitions

We present original mean activity concentrations of striatum obtained at hourly intervals in Table 1, per patient as well as for caudate nucleus and putamen. The activities appeared to differ between LS and HS patients at 2 h p.i. and tended to do so as early as at 1 h p.i. At all times, the difference between LS and HS patients was more marked in the putamen than in the caudate nucleus (Table 1).

Table 1 Adjusted mean activity concentration in hourly intervals, shown by median (min;max).

We used data from the dynamic SPECT images to construct the time-activity curves (TAC) presented in Fig. 2. Figure 2a shows uptake in Vs, while Fig. 2b–d show corrected uptake values according to Eq. (1). Uncorrected uptake values were approximately 30% lower than adjusted values; however, the degree of underestimation was smaller for patients with lower uptake values (e.g., patient #1). Figure 2a–d show two TAC per patient representing both left and right striatum. The shapes of TACs in Fig. 2a–d were similar for patients with an initial increase prior to 100 min, followed by a plateau. From Fig. 2a–d, TACs from HS and LS patients diverted after about an hour and onwards, although patient #5 appeared to maintain low uptake consistent with the LS patients, most clearly seen in the case of putamen (Fig. 2d). We note that one patient (#13) appeared to have higher uptake values mostly in striatum (Fig. 2a,b), albeit also in reference tissues (Fig. 2e,f). Regarding reference volumes, the occipital cortex (Fig. 2e) presented with the lower mean activity concentration initially, compared to cerebellum (Fig. 2f).

Figure 2
figure 2

Relation between activity and time. (a) Uncorrected mean activity concentration for striatum. (b) Striatal mean activity concentration corrected for partial volume and minor movement artefacts. (c) Corrected mean activity concentration in the caudate nucleus. (d) Corrected mean activity concentration in the putamen. (e) Mean activity concentration in the occipital cortex. (f) Mean activity concentration in the cerebellum. Each different marker represents a patient with values from both the left and right striatum shown in (ad). Patients #1–4 had Karahman scores of < 4 at 3-h images and are illustrated in orange.

The striatal SBRcalc values (Fig. 3) had a more linear course than TACs. The majority of patients presented with ratios of 4–8 at 3 h p.i. As expected, we note a correlation between SBRcalc values (Fig. 3) and mean activity concentrations in the striatum (Fig. 2a–d), because the majority of patients with low SBRcalc had low striatal activity concentrations, while the patients with the higher SBRcalc values had higher striatal activity concentrations at all times of the imaging. We present the values of 3 h p.i. in Table 2. In agreement with the difference between uncorrected and adjusted uptake values, SBRDaTQUANT values reached roughly 2/3 of the values of SBRcalc. Table 3 presents averages of the 3-h SBRcalc values for each patient group. HS patients presented with putamen-caudate ratios of close to unity, while the LS group members presented with distinctly lower values of SBRcalc in putamen than in caudate nucleus (Table 3).

Figure 3
figure 3

Calculated specific binding ratio (SBRcalc) for all patients throughout the scan duration.Each patient is represented by a different mark and illustrated with a dataset from both the left and the right striatum. Patients with #1–4 had Karahman scores of < 4 at 3 h p.i. images and are illustrated in orange.

Table 2 Static and dynamic parameters related to the left and right striatum for each patient.
Table 3 Ratios between striatum as well as caudate nucleus and putamen, shown by median (min;max).

SRTM parameters

The regression results of SRTM were consistent with the measured time activity curves. The dynamic parameters extracted from the model had standard errors of less than 2% of the median, except for the BPND of the caudate nuclei of two patients with standard errors of more than 100%. One patient (#5) had lacunar infarcts in the left caudate nucleus with a standard error for the left striatum of 5%. Another patient (#2) had no reported comorbidity. We excluded both patients’ data from the BPND, R1, and k2 analyses of the left caudate nucleus.

Figure 4 shows the BPND values of each striatum among the patients. LS patients had more putamen-caudate nucleus asymmetry of BPND estimates than HS patients, as shown in Table 3.

Figure 4
figure 4

Non-displaceable binding potential (BPND) for each patient’s striata. Each patient is illustrated with a left column representing values for the left striata (N = 11) and a right column representing data for the right striata (N = 13). Patients #1–4 were categorized as low score, and #5–13 were categorized as high score. Right side of #2 and #5 is excluded.

We compared estimates of SBRcalc with estimates of BPND for both left and right striatum, as shown in Fig. 5. The linear correlation had a coefficient of determination (R2) of 0.75.

Figure 5
figure 5

Scatterplot for correlation between measured non-displaceable binding potential (BPND) and calculated specific binding ratio (SBRcalc). Each data mark is the value of either a left or right striatum from one of the 13 patients with patients #2 and #5 excluded. Orange triangles indicate low-score patients, and blue circles indicate high-score patients.

We show the R1 values of all patients in Table 2. Patient #5 with the highest values of R1 for both left and right striatum had vascular comorbidity in the form of lacunar infarcts. Patient #3 with vascular comorbidity had values close to 1, but with asymmetry of left and right striatum as listed in Table 2. The variability was high among patients for values of k2, with no clear difference between LS and HS.

Correlation with clinical presentation

Compared to tremor lateralization, three of the LS group of patients (30% of the total) had lower mean activity concentration in contralateral striatum, while six patients of the HS group (60% of the total) had slightly lower mean activity concentration in ipsilateral hemisphere as listed in Table 2. An LS patient with lateralized tremor had equal activities in left and right striatum, with similar results for SBRcalc. For BPND and R1, results varied (Table 2).

Discussion

The main objective was an examination of the feasibility of functional dynamic brain SPECT (fSPECT) with the three-headed multi-pinhole gamma camera. We created time-activity curves with quantitative uptake values and corrected for PVE. We further showed the feasibility of examining patients in terms of quantitative pharmacokinetic parameters.

We applied standardized volumes of striatum to all patients. The PVE correction by means of Eq. (1) affected all recordings. In the study, we did find correlation between the 3 h p.i. visual assessment by Kahraman et al.10 and the striatal mean activity uptake values, SBRcalc, and the BPND values obtained by SRTM, strongest of which appeared to be the uptake value. In addition, the findings indicate that uptake values distinguished LS and HS patients as early as 2 h p.i. We found reasonable correlation between the dynamic parameter BPND and the SBRcalc value. The resulting estimates of the parameters R1 and k2 varied among patients but without apparent correlation to the visual assessment.

A correlation between tremor-dominant hemisphere and contralateral declines of both uptake values and SBRcalc estimates appeared to exist among LS patients, but none of the kinetic parameters confirmed the correlation.

We examined the dynamic uptakes in both occipital cortex and cerebellum, the latter higher initially, but otherwise similar. As usual with DaT SPECT, we used the occipital cortex as reference. Indeed, in a recent PET study, the occipital cortex emerged as better suited as reference compared to cerebellum, because of slightly higher temporal stability and lower bias17.

The imaging protocol outlined previously was chosen as a compromise between image quality and acquisition speed. We recognize that the frame length may remove pharmacokinetic information, however we did not want to sacrifice image quality any further. We did change the imaging protocol halfway through the study, in order to raise the visual presentation of the scan. While the visual presentations did improve slightly, we saw no clear differences among any of the uptake values in striatum, depending on the imaging duration.

Conventional imaging of PD is limited to a simple visual interpretation of tracer uptake at a certain time, often aided by the ratio between activities in striatum and a reference region. The study of quantitative dynamic pharmacokinetic parameters extracted from brains of patients with PD is likely to lead to more nuanced diagnostic tools that improve understanding and differentiation. The binding potential is an index of the numbers of receptors present in a given region and their availability, as a value related to the product of receptor density and affinity of the ligand. Thus, if PD is afflicted by loss of neurons or loss of receptor affinity, the estimate of BPND is affected as well. R1 describes relative perfusion and could thus enlight perfusion differences.

In a prior study, Mazère et al. completed dynamic fSPECT imaging with 123I-iodobenzovesamicol, a vesicular acetylcholine transporter, in healthy volunteers and patients with Lewy body dementia18. The group also managed to create dynamic SPECT image series and to determine BPND. In the present study, however, we used half as much injected radioactivity, and the imaging lasted less than half as long. To our knowledge, there are no previous examples of dynamic dopaminergic brain SPECT imaging with pharmacokinetic modelling in the literature. Nonetheless, we can draw parallels to PET imaging. PET studies with the FE-PE2I DaT tracer show significantly lower values of BPND in PD patients compared to healthy control subjects19,20. Similarly, we found that values of BPND tended to be lower in patients with lower visual inspection scores. In the present study, we noted an underestimation of activity in striatum by 30% without correction for PVE. A PET study with both 11C-FLB457 and 18F-fallypride tracers revealed increases of estimates of BPND of 50% by correction for PVE21. Likewise, in a prior study of 123I-labeled tracer in a brain phantom, the authors reported underestimation of activity in striatum of close to 50%, and further underestimation of BP without correction for PVE22. The present evidence shows that dynamic brain scans are feasible by fSPECT, thereby increasing the possibilities of performing dynamic brain scans independently of PET-CT.

While a dynamic PET brain scan requires only 60 min of acquisition5, the dynamic fSPECT by contemporary protocols requires longer durations, and parkinsonian patients may have difficulty resting on the back for longer periods, also because the patients often have stiff necks and tense shoulders. We did experience several of the patients colliding with the collimators due to tense and raised shoulders, and the movements of the patients’ increased the uncertainty of measurements, also of the effect of small movements on the uptake values. Of the two patients for whom the method and SRTM provided unrealistic and imprecise values of BPND, at least one patient had vascular pathology in striatum. Therefore, we need more dynamic studies in case of vascular comorbidity.

Increasing age provides lower DaT availability with as much as 11% decline per decade23,24. While both gender and BMI influence DaT availability24,25, we now need further studies of these factors to evaluate the effects on dynamic uptake. The correlations between visual assessment and uptake values and BPND warrant consideration of shorter imaging durations, but further studies are needed to assess the imaging durations affecting these parameter estimates, and to determine the imaging durations that limit the determination of values of BPND. Previously we thought that estimates of R1 reflect perfusion or extraction effects16, but we were unable here to detect a correlation that may be evident in larger studies.

A neurologist referred all the participants in the study, all as suspected of having PD, although none displayed clear Parkinson-like symptoms that would allow diagnosis on the basis of clinical presentation alone.

Strengths and weaknesses

The present study is a feasibility study that included only a small number of participants, with no normal control subjects. Nonetheless, as far as we know, the study is the first to confirm the feasibility of dynamic fSPECT.

We gathered no follow-up data to determine whether or not PD was actually diagnosed in any of the participants. We did not collect the participants’ medical journals, and many of the participants were referred by private neurology practitioner that ruled out clinical correlation such as Hoehn-Yahr staging. We obtained no height or weight of the participants, and we had no estimates of blood distribution volume that may have had an impact on the availability of the tracer.

Conclusion

We tested the claim of using a three-headed SPECT device to obtain fSPECT images with multi-pinhole collimators. The findings show that dynamic brain SPECT with [123I]FP-CIT is feasible and that it is possible to obtain values of quantitative uptake and pharmacokinetic parameters from the images. The findings show the feasibility of examining patients suspected of PD in new ways by means of fSPECT. Further studies are warranted to examine the optimal future protocol of fSPECT and to examine the diagnostic value of dynamic parameters.