Abstract
Single-molecule techniques provide opportunities for molecularly precise imaging, manipulation, assembly and biophysical studies. Owing to the kinetics of bond rupture processes, rapid single-molecule measurements can reveal novel bond rupture mechanisms, probe single-molecule events with short lifetimes and enhance the interaction forces supplied by single molecules. Rapid measurements will also increase throughput necessary for technological use of single-molecule techniques. Here we report a nanomechanical sensor that allows single-molecule force spectroscopy on the previously unexplored microsecond timescale. We probed bond lifetimes around 5 μs and observed significant enhancements in molecular interaction forces. Our loading-rate-dependent measurements provide experimental evidence for an additional energy barrier in the biotin–streptavidin complex. We also demonstrate quantitative mapping of rapid single-molecule interactions with high spatial resolution. This nanomechanical interface may allow studies of molecular processes with short lifetimes and development of novel biological imaging, single-molecule manipulation and assembly technologies.
Similar content being viewed by others
Introduction
Precise recognition mechanisms of biological molecules offer a tremendous potential for the assembly1,2,3, imaging4,5 and analysis of complex biological materials6. Single-molecule force spectroscopy (SMFS) experiments with the atomic force microscope (AFM)7 have demonstrated the potential of intermolecular forces to be used for chemically specific nanoscale imaging8, manipulation9, directed assembly10, and biophysical studies of deformation and failure of biomolecules11. Forces measured during the rupture of molecular bonds are affected by thermally activated dissociation, which results in loading-rate-dependent characteristics. Consequently, the slow mechanical interface to single-molecule processes makes it difficult to access short-lived single-molecule events12. For example, a comparison of experiments and molecular dynamics (MD) simulations of the well-studied biotin–avidin complex shows that these molecules have a potential to supply forces much larger than that encountered in experiments13,14,15. The larger forces will emerge only if bound molecules are pulled apart on a shorter timescale. Furthermore, experimental access to short timescales can provide insights into failure of protein-based materials in which failure mechanisms depend on the timescales of external forces16.
In principle, SMFS can be integrated into the tapping-mode imaging platform, where the cantilever oscillates back and forth against the surface on resonance and at a high frequency. Evidence for this is provided by the molecular recognition images obtained in the topography and recognition imaging mode (TREC)5. In this mode, molecular binding events produce detectable patterns in the oscillation amplitude. This suggests that single molecules can efficiently bind and unbind at the short timescales offered by the resonating microcantilever. If a rapid nanomechanical interface to these interactions is built, high-speed SMFS can be obtained at the frequency of the vibrating cantilever.
In this study, we have designed an AFM cantilever that establishes a rapid nanomechanical interface to single-molecular interactions, thereby increasing the rate of SMFS by nearly four orders of magnitude. The new design is using the torsional harmonic cantilever concept that was developed for high-resolution nanomechanical analysis of samples17,18. These T-shaped cantilevers generate high-speed force-distance curves during the tapping-mode AFM imaging process. Measurements carried out with this nanomechanical interface allowed explorations of single-molecule processes on the microsecond timescale. Analysis of the rupture forces of the well-studied biotin–streptavidin complex at the short timescales has provided experimental evidence for an additional energy barrier. The high speed of measurements further allowed generating high-resolution maps of the quantitative rupture forces.
Results
SMFS on the microsecond timescale
We adapted the torsional harmonic cantilever concept for SMFS on the microsecond timescale using the scheme illustrated in Figure 1. First, we attach the ligand to the tip of the T-shaped cantilever via a short flexible polymer strand and immobilize its binding partner on a substrate (Fig. 1a). Originally used for antibody recognition images, attaching the ligand to the tip of the AFM via a short flexible spacer has several benefits enabling high-speed SMFS19. It enhances binding probability by providing conformational flexibility for the ligand to reach the receptor and orient itself for binding. The flexible linker also provides a temporal delay between the tip–substrate interaction and the unbinding event5,20, which encodes the unbinding information onto the tip-sample force waveform. This information can be recovered by a computer that monitors the twisting motion of the T-shaped cantilever.
Different from the previous design of the torsional harmonic cantilever, the scheme described in Figure 1b requires an ability to monitor forces at every single oscillation cycle with sufficient signal-to-noise ratio. We have designed the cantilever seen in Figure 2a that has force sensitivity sufficient to observe single-molecule interactions with adequate detail. With this cantilever, we have investigated the interaction of biotin with streptavidin adsorbed onto a mica substrate. A 7-nm-long polyethylene glycol (PEG) is used as the flexible linker. Two curves plotted in Figure 2b show the effect of unbinding on the force waveforms. The forces originating from the stretching of the spacer are observed between 80 and 95 μs. The lowest point in the force waveform approximately corresponds to the unbinding force; in the two cases seen in Figure 2b, these points are located near 95 μs. The rupture forces emerging from each pulling event are different because of the stochastic nature of unbinding.
Statistical distribution of the lowest forces obtained from repeatedly measured force waveforms is given in Figure 2c (examples of entire force waveforms and force-distance curves are given in Supplementary Movie 1). A large portion of measured forces does not exhibit binding and reflect measurement noise in the range of −40 to −100 pN. However, points with force values below −150 pN are mostly localized between 70 and 90 μs. These points also exhibit a trend in which larger pulling forces occur later. This trend is due to the stretching of the polymer spacer between the tip and the bound biotin–streptavidin complex that provides rapidly increasing forces as end to end distance approaches the contour length. As seen in the inset of Figure 2c, the number of data points in this region of the graph reduced significantly after adding free biotin to the imaging buffer to block binding sites of streptavidin molecules, providing evidence to the specificity of the measurements. A linear fit to data in the region below −150 pN estimates that the forces increase at a rate of 102 pN μs−1. Considering that the pulling forces reach as much as 750 pN, we estimate the lifetime of the bound molecular complex to be around 4–8 μs.
Discrimination of multiple unbinding events
We note that forces as high as the ones measured here can be encountered in conventional SMFS experiments if multiple ligand–receptor interactions are formed and disrupted in one measurement cycle7. Multiple interactions are less likely to occur in the present measurements owing to the short period of oscillations. For biotin to bind streptavidin on the surface, the tip-sample distance has to be less than the length of the PEG spacer (∼7 nm). As the amplitude of the oscillations is 9 nm, binding has to occur in less than half the oscillation period (∼55 μs). Using the association constant determined from surface plasmon resonance measurements21 (Ka≈5×106 M−1 s−1) and assuming an effective volume Veff∼718 nm3 for the tethered biotin (volume of a half sphere with 7-nm radius), we estimate the time that would lead to 50% binding probability to be ∼86 μs (t0.5≈NAVeff/Ka)19. The binding probability is likely to be even lower than this estimate, because of the additional energy barrier imposed by the PEG spacer used in the experiment. Furthermore, owing to the high bandwidth (135 kHz) and effective spring constant (0.98 N m−1) of the torsional mode, multiple unbinding events within the same oscillation cycle can be potentially discriminated as multiple force minima. We have analysed a set of 2,300 force waveforms (all exhibiting a pulling force of more than 200 pN) and found only six such curves. Two of these curves are plotted in Figure 2d.
Quantitative mapping of rapid single-molecule interactions
The ability to perform SMFS at each cycle of the oscillating cantilever permits mapping of molecular interaction forces quantitatively across surfaces. Figure 3a demonstrates one such image generated by mapping the maximum pulling forces across a 400-nm-wide region of the sample. We observe that the unbinding forces are well above the noise floor so that they appear as clear spikes. This is a direct consequence of the specificity of the information provided by SMFS. The spikes generated by the distinct interaction forces are locating the receptor molecules within ∼7 nm, because the ligand can bind to a receptor within that radius. A comparison of the force map to the simultaneously recorded topography image (Fig. 3b,c) shows that the locations of the force spikes correlate well with topographical features. A force map obtained with another tip-sample combination also exhibited this characteristic (Supplementary Fig. S1). We note that, a number of force spikes do not have a corresponding topographical feature. These may represent unspecific interactions or detachment of streptavidin molecules from the surface. We have also recorded the conventional phase image, simultaneously (Fig. 3d). The phase image does not show as many features as the force map or topography. Phase signal is sensitive to mechanical properties of the surface. Therefore, the features in the phase signal are likely originating from aggregated streptavidin, which can create larger mechanical contrast. Note that the topographical features corresponding to those locations are wider than other topographical features. We also observed that the locations of the force spikes recorded in trace and retrace directions correlate strongly (Fig. 3e). In addition, these force spikes disappeared after introducing biotin to the imaging buffer to block binding sites (Fig. 3f,g). These observations show that microsecond timescale single-molecule interactions can be quantitatively measured and spatially mapped with high resolution.
Statistics of rupture forces
Histograms generated from the pulling forces recorded while the cantilever scans the surface are given in Figure 4a. Each histogram exhibits two peaks. The first peak at the lower force level corresponds to the background determined by unspecific interactions and measurement noise (first row in Fig. 4a). The second peak is seen only after adjusting the scale of the frequency axis (second row in Fig. 4a). This is because most oscillation cycles do not result in biotin–streptavidin binding. Note that the histograms in the first and second rows correspond to the measurements in Figure 3a. According these measurements, the majority of rupture forces are around 625 pN. Forces as much as 750 pN are measured as well. These values are much higher than the rupture forces previously reported for this molecular complex7,15,22,23, which is a consequence of the significantly higher loading rate achieved.
Varying the loading rate
We have varied the loading rate by reducing the vibration amplitude at a constant oscillation frequency and observed the resulting changes in the values of most likely rupture forces (third and fourth rows in Fig. 4a). The histograms in the second, third and fourth rows are recorded at 9, 6 and 4.5 nm, respectively. We estimated the loading rates by linearly fitting the force versus time plots as in Figure 2c. Figure 4b shows the most likely rupture forces plotted against the loading rate. The loading rate dependency of most likely rupture forces contains information about the energy landscape of receptor–ligand interactions. When measured over a wide dynamic range and plotted on a logarithmic scale, different energy barriers can be resolved from the changes in the slope15,24,25. The position of the barrier xβ is related to the slope fβ by fβ=kBT/xβ (here kB is Boltzmann constant and T is temperature)24,26. According to this formulation, the slope of the forces measured here corresponds to a barrier position xβ ≅0.27 Å.
Discussion
The barrier position derived from the loading-rate-dependant measurements are considerably smaller than previous estimates of the innermost barrier position15 of 1.2 Å, suggesting the possibility of an additional barrier. This is also supported by the results of Zhou et al.27, combined MD and continuum mechanics simulations of which identified barriers at 0.3, 1.0 and 2.6 Å. According to their simulations, the innermost barrier is maintained by two hydrogen bonds to biotin by Thr90 and Asn49 of streptavidin. Note that steered MD simulations of other protein materials also revealed barrier positions comparable to the one measured here16.
Recent theoretical and experimental works showed that the stiffness of the force transducer can influence the measured rupture forces28,29. Using analytical models generalized to include stiff transducers, Maitra and Arya29 have found that the comparison of rupture force spectra obtained with different devices is permissible if spring constants K and barrier positions xβ satisfy the relationship Kx2β/(2kBT)≪1. In the present experiment, forces are measured by the torsional deflections, of which the effective spring constant at the position of the offset tip is ∼0.98 N m−1. Using xβ≅0.27 Å, we obtain Kx2β/(2kBT)≅0.08. Note that the effective spring constant can be even lower because of the presence of the linker attaching biotin to the tip. Therefore, we argue that the measured xβ corresponds to an additional barrier in the energy landscape of biotin–streptavidin.
In addition to the evidence provided for an additional energy barrier, the data also provide experimental evidence that the forces can be as high as that predicted by MD simulations14. Those calculations identified several unbinding paths for avidin–biotin complex, with rupture forces between 650 and 800 pN. The MD simulations were carried out on a timescale on the order of 1 ns; however, recent MD simulations now reach hundreds of nanoseconds16,30. As a result, the timescales of experiments and simulations are now within about an order of magnitude. This gap can be bridged with the use of small AFM cantilevers, as Ando et al.31 demonstrated a small cantilever AFM system that can detect torsional vibrations while the cantilever was driven at 1 MHz.
The rapid nanomechanical interface presented here can have various applications beyond probing the energy landscapes of receptor–ligand interactions. For example, cut-and-paste-based molecular assembly schemes rely on selective formation and rupture of a series of chemical bonds between an ink molecule with the carrier tip and the surfaces of depot or target areas10. Theoretical calculations have shown that when chemical bonds in series are subject to external force, loading rate can selectively determine which bond will break first32. This rate-dependent behaviour would allow picking up molecular inks at one loading rate and delivering them at another loading rate using a universal chemistry at the depot and target surfaces, which would also permit reversible rearrangements of assembled structures. This can potentially be realized experimentally by switching between conventional SMFS and the high-speed approach presented here.
Variability in bond rupture mechanisms at different loading rates or timescales has biological implications as well. MD simulations of protein building blocks demonstrated variations in failure mechanisms in response to loading rates16. Experiments with the rapid nanomechanical interface presented here can provide new insights to deformation and failure mechanisms of protein-based materials like collagen and silk, strengths of which are critical for their biological functions33. Experiments also afford greater control of the environment so that effects of solvent pH and ionic strength can be investigated in depth.
Methods
Torsional harmonic cantilevers
T-shaped cantilevers are fabricated by using conventional manufacturing protocols (Applied Nanostructures) with custom geometry. The body of the cantilever is made of silicon nitride and the wider region at the free end including the tip is made of silicon. The probes are 130-μm long and 35-μm wide. The width at the free end is 55 μm. The tip offset distance is 22 μm (measured by electron microscopy). The thickness of the silicon nitride is 600 nm. It has 30-nm-thick gold coating. The thickness of the silicon part is ∼1 μm. The resonance frequencies of these cantilevers in solution are around 9.5 kHz, determined from the thermal noise spectrum. Their torsional resonances and quality factors are ∼115 kHz and 3.5, respectively. The force sensitivity of the torsional mode is calibrated against the thermal noise and found to be 8.0 nN V−1. This corresponds to an effective spring constant of 0.98 N m−1 (defined by ratio of the tip displacement in the torsional mode to the force acting on the tip).
Experimental setup
A commercial AFM system (Multimode series with a Nanoscope 5 controller, and a signal breakout box) has been used. Measurements and imaging are done in fluid tapping mode. The T-shaped cantilever is driven at 9 kHz. Vibration amplitude was set to 4.5, 6 or 9 nm. Cantilever vibration signals are collected with a personal computer equipped with two processors (3.2 GHz, 2 GB shared memory) and a data acquisition card (NI-DAQ S-series 6115). Calculations of tip-sample force waveforms from the torsional vibration signals are carried out in Labview (National Instruments). We followed the same procedures reported previously17 to invert the transfer function of the torsional mode and eliminate cross talk from large vertical deflection signals; however, every oscillation cycle is analysed individually. This approach is based on a framework used by Stark et al.34 for the analysis of higher harmonic vibrations. In a previous work35, we have tested tip-sample force waveforms generated from the torsional vibrations by comparing the measured forces with the previously developed theories36,37 of cantilevers vibrating in the tapping mode. See also Solares and Hoelscher38 for numerical analysis of the measurements based on torsional harmonic cantilever. In the calculations made here, torsional vibration signals up to 135 kHz are included. At a typical scan speed of 1 Hz, approximately nine oscillation cycles are analysed per pixel of the AFM image. The lowest force value measured in those nine cycles is recorded at that pixel to generate unbinding force maps as in Figure 3a.
Tip modification
T-shaped cantilevers are cleaned in an O2 plasma for 5 min and then transferred to a piranha solution (sulphuric acid/H2O2, 8:2) for 30 min. After extensive rinsing with nanopure H2O, the tips are transferred into a fresh ethanol solution, which contains 5% APTES (3-aminopropy-triethoxysilane; Sigma Aldrich), 5% de-ionized water and 90% absolute ethanol, for 30 minutes at room temperature. The amino-functionalized AFM tips are rinsed with de-ionized water five times and pure ethanol five times, followed by drying with a gentle stream of nitrogen. Next, carboxyl-group polyethylene oxide PEG polymers with biotin end groups (Biotin PEG propionic acid, C37H69N3O16S, Polypure AS) were covalently linked to the silanized AFM tip surfaces using Polylink protein coupling kit (Bangs Laboratories). The Biotin-PEG-COOH was diluted in coupling buffer (50 mM MES, pH 5.2, 0.05% Proclin300). The EDAC (carbodiimide) solution was prepared by dissolving EDAC in coupling buffer. AFM tips were incubated in the mixture solution of Biotin-PEG-COOH and EDAC for 45 min at room temperature. Right after incubation, Biotin-functionalized AFM tips were rinsed with washing buffer (10 mM Tris, pH 8.0 bovine serum albumin, 0.05% Proclin300) for three times.
Sample preparation
A stock solution of streptavidin (Sigma Aldrich) in 1 mg ml−1 PBS was diluted with 100 mM NaCl to a final streptavidin concentration of 0.8 μg ml−1. Fresh-cleaved mica was incubated with streptavidin solution for 20 min and subsequently rinsed with PBS buffer solution for two times. The mica sheet was mounted in a fluid cell for AFM experiments. Measurements are carried out in PBS.
Testing the specificity of interaction forces
Specificity of the measured rupture forces is tested by blocking the binding sites of immobilized streptavidin molecules with free biotin introduced into the imaging buffer. The results are given in Figure 3.
Additional information
How to cite this article: Dong, M. & Sahin, O. A nanomechanical interface to rapid single-molecule interactions. Nat. Commun. 2:247 doi: 10.1038/ncomms1246 (2011).
References
Zhang, S. G. Fabrication of novel biomaterials through molecular self-assembly. Nat. Biotechnol. 21, 1171–1178 (2003).
Ke, Y. G., Lindsay, S., Chang, Y., Liu, Y. & Yan, H. Self-assembled water-soluble nucleic acid probe tiles for label-free RNA hybridization assays. Science 319, 180–183 (2008).
Rothemund, P. W. K. Folding DNA to create nanoscale shapes and patterns. Nature 440, 297–302 (2006).
Hinterdorfer, P. & Dufrene, Y. F. Detection and localization of single molecular recognition events using atomic force microscopy. Nat. Methods 3, 347–355 (2006).
Stroh, C. et al. Single-molecule recognition imaging-microscopy. Proc. Natl Acad. Sci. USA 101, 12503–12507 (2004).
Neuman, K. C. & Nagy, A. Single-molecule force spectroscopy: optical tweezers, magnetic tweezers and atomic force microscopy. Nat. Methods 5, 491–505 (2008).
Florin, E. L., Moy, V. T. & Gaub, H. E. Adhesion forces between individual ligand-receptor pairs. Science 264, 415–417 (1994).
Dupres, V. et al. Nanoscale mapping and functional analysis of individual adhesins on living bacteria. Nat. Methods 2, 515–520 (2005).
del Rio, A. et al. Stretching single talin rod molecules activates vinculin binding. Science 323, 638–641 (2009).
Kufer, S. K., Puchner, E. M., Gumpp, H., Liedl, T. & Gaub, H. E. Single-molecule cut-and-paste surface assembly. Science 319, 594–596 (2008).
Buehler, M. J. & Yung, Y. C. Deformation and failure of protein materials in physiologically extreme conditions and disease. Nat. Mater. 8, 175–188 (2009).
Liu, R. C., Hu, D. H., Tan, X. & Lu, H. P. Revealing two-state protein-protein interactions of calmodulin by single-molecule spectroscopy. J. Am. Chem. Soc. 128, 10034–10042 (2006).
Grubmuller, H., Heymann, B. & Tavan, P. Ligand binding: molecular mechanics calculation of the streptavidin biotin rupture force. Science 271, 997–999 (1996).
Izrailev, S., Stepaniants, S., Balsera, M., Oono, Y. & Schulten, K. Molecular dynamics study of unbinding of the avidin-biotin complex. Biophys. J. 72, 1568–1581 (1997).
Merkel, R., Nassoy, P., Leung, A., Ritchie, K. & Evans, E. Energy landscapes of receptor-ligand bonds explored with dynamic force spectroscopy. Nature 397, 50–53 (1999).
Ackbarow, T., Chen, X., Keten, S. & Buehler, M. J. Hierarchies, multiple energy barriers, and robustness govern the fracture mechanics of alpha-helical and beta-sheet protein domains. Proc. Natl Acad. Sci. USA 104, 16410–16415 (2007).
Sahin, O., Magonov, S., Su, C., Quate, C. F. & Solgaard, O. An atomic force microscope tip designed to measure time-varying nanomechanical forces. Nat. Nanotechnol. 2, 507–514 (2007).
Dong, M. D., Husale, S. & Sahin, O. Determination of protein structural flexibility by microsecond force spectroscopy. Nat. Nanotechnol. 4, 514–517 (2009).
Hinterdorfer, P., Baumgartner, W., Gruber, H. J., Schilcher, K. & Schindler, H. Detection and localization of individual antibody-antigen recognition events by atomic force microscopy. Proc. Natl Acad. Sci. USA 93, 3477–3481 (1996).
Gabai, R., Segev, L. & Joselevich, E. Single polymer chains as specific transducers of molecular recognition in scanning probe microscopy. J. Am. Chem. Soc. 127, 11390–11398 (2005).
Qureshi, M. H., Yeung, J. C., Wu, S. C. & Wong, S. L. Development and characterization of a series of soluble tetrameric and monomeric streptavidin muteins with differential biotin binding affinities. J. Biol. Chem. 276, 46422–46428 (2001).
Lee, G. U., Kidwell, D. A. & Colton, R. J. Sensing discrete streptavidin biotin interactions with atomic-force microscopy. Langmuir 10, 354–357 (1994).
Wong, J., Chilkoti, A. & Moy, V. T. Direct force measurements of the streptavidin-biotin interaction. Biomol. Eng. 16, 45–55 (1999).
Evans, E. & Ritchie, K. Dynamic strength of molecular adhesion bonds. Biophys. J. 72, 1541–1555 (1997).
Lo, Y. S., Zhu, Y. J. & Beebe, T. P. Loading-rate dependence of individual ligand-receptor bond-rupture forces studied by atomic force microscopy. Langmuir 17, 3741–3748 (2001).
Husson, J. & Pincet, F. Analyzing single-bond experiments: influence of the shape of the energy landscape and universal law between the width, depth, and force spectrum of the bond. Phys. Rev. E 77, 026108 (2008).
Zhou, J. et al. Unbinding of the streptavidin-biotin complex by atomic force microscopy: a hybrid simulation study. Journal of Chemical Physics 125, 104905 (2006).
Walton, E. B., Lee, S. & Van Vliet, K. J. Extending Bell's model: how force transducer stiffness alters measured unbinding forces and kinetics of molecular complexes. Biophys. J. 94, 2621–2630 (2008).
Maitra, A. & Arya, G. Model accounting for the effects of pulling-device stiffness in the analyses of single-molecule force measurements. Phys. Rev. Lett. 104, 108301 (2010).
Scheraga, H. A., Khalili, M. & Liwo, A. Protein-folding dynamics: overview of molecular simulation techniques. Annu. Rev. Phys. Chem. 58, 57–83 (2007).
Ando, T., Uchihashi, T. & Fukuma, T. High-speed atomic force microscopy for nano-visualization of dynamic biomolecular processes. Prog. Surf. Sci. 83, 337–437 (2008).
Neuert, G., Albrecht, C. H. & Gaub, H. E. Predicting the rupture probabilities of molecular bonds in series. Biophys. J. 93, 1215–1223 (2007).
Keten, S., Xu, Z. P., Ihle, B. & Buehler, M. J. Nanoconfinement controls stiffness, strength and mechanical toughness of beta-sheet crystals in silk. Nat. Mater. 9, 359–367 (2010).
Stark, M., Stark, R. W., Heckl, W. M. & Guckenberger, R. Inverting dynamic force microscopy: from signals to time-resolved interaction forces. Proc. Natl Acad. Sci. USA 99, 8473–8478 (2002).
Sahin, O. Time-varying tip-sample force measurements and steady-state dynamics in tapping-mode atomic force microscopy. Phys. Rev. B 77, 115405 (2008).
Garcia, R. & San Paulo, A. Attractive and repulsive tip-sample interaction regimes in tapping-mode atomic force microscopy. Phys. Rev. B 60, 4961–4967 (1999).
Holscher, H., Ebeling, D. & Schwarz, U. D. Theory of Q-controlled dynamic force microscopy in air. J. Appl. Phys. 99, 084311 (2006).
Solares, S. D. & Holscher, H. Numerical analysis of dynamic force spectroscopy using the torsional harmonic cantilever. Nanotechnology 21, 075702 (2010).
Acknowledgements
The study is supported by the Rowland Junior Fellows Program and the Wyss Institute for Biologically Inspired Engineering at Harvard. We thank Claudia Friedsam and Wesley P. Wong for helpful discussions.
Author information
Authors and Affiliations
Contributions
M.D. and O.S. performed experiments; M.D. prepared functionalized tips, O.S. designed the cantilevers; M.D. and O.S. analysed the data, O.S. directed the research and wrote the paper.
Corresponding author
Ethics declarations
Competing interests
The authors declare competing financial interests. A patent application has been filed by Harvard University. U.S. Patent Application No. 12/652,925 ‘Measurement and Mapping of Molecular Stretching and Rupture Forces’.
Supplementary information
Supplementary Figure S1
Supplementary Figure S1 (PDF 233 kb)
Supplementary Movie 1
Tip-sample interactions recorded during the experiments are plotted against time and tip-sample distance. The curves are plotted 300 times slower than the actual recording rate. A pause equal to 300 ms is added right after curves that exhibit pulling forces more than 200 pN. (MOV 2181 kb)
Rights and permissions
About this article
Cite this article
Dong, M., Sahin, O. A nanomechanical interface to rapid single-molecule interactions. Nat Commun 2, 247 (2011). https://doi.org/10.1038/ncomms1246
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/ncomms1246
This article is cited by
-
Analysis of the vibrational characteristics of diamane nanosheet based on the Kirchhoff plate model and atomistic simulations
Discover Nano (2023)
-
Atomic force microscopy-based characterization and design of biointerfaces
Nature Reviews Materials (2017)
-
High-speed multiple-mode mass-sensing resolves dynamic nanoscale mass distributions
Nature Communications (2015)
-
Imaging and three-dimensional reconstruction of chemical groups inside a protein complex using atomic force microscopy
Nature Nanotechnology (2015)
-
Quantitative measurement of the mechanical properties of human antibodies with sub-10-nm resolution in a liquid environment
Nano Research (2015)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.