Introduction

Lung cancer continues to be one of the most common malignancies worldwide and is the leading cause of cancer-related deaths1. Diagnosing and characterizing solitary pulmonary lesions (SPLs), ranging from lung cancer to infections and other benign tumors, are still a major medical challenge2. Magnetic resonance imaging (MRI) is emerging as an alternative and complementary to lung imaging3. Lung MRI is more advantageous than chest radiography and computed tomography (CT) in terms of avoiding ionizing radiation. Moreover, diffusion weighted imaging (DWI) has the potential to gather diffusion and perfusion data, aiding in the characterization of SPLs. However, addressing geometric distortion resulting from magnetic field inhomogeneities in lung MRI is crucial. Therefore, effective methods should be developed to obtain high-quality images while simultaneously ensuring the precision of the derived parameters.

DWI is a noninvasive modality used to assess the diffusivity of water molecules within living tissues. Intravoxel incoherent motion DWI (IVIM-DWI), which uses a biexponential model, enables the separate determination of diffusion and perfusion parameters, unlike mono-exponential DWI4. Echo-planar imaging (EPI) is the preferred DWI technique because of its rapid acquisition time5. Also, many studies6,7 including meta-analysis studies8,9 have demonstrated the considerable benefit of EPI-DWI for the diagnosis and characterization of pulmonary nodules. However, the presence of magnetic field inhomogeneities in the lungs can lead to signal loss or geometric distortion in pulmonary nodule EPI-DWI, thereby compromising the accuracy of pulmonary nodule measurements of derived parameters.

Turbo spin echo–based DWI may be a useful alternative to EPI-DWI in the lung, which is particularly susceptible to magnetic inhomogeneity. Turbo spin echo (TSE) is the most frequently used DWI technique less sensitive to magnetic inhomogeneity. Previously published studies10,11,12 have highlighted the ability of TSE-DWI to improve image quality in conditions such as head and neck cancers, orofacial lesions, and breast cancer. In addition, a recent study13 already presented that TSE-DWI provides significantly improved image quality and test–retest robustness of IVIM parameters in patients with SPLs. However, its role in the assessment of pulmonary lesions remains to be elucidated. As part of further validating lung TSE-DWI feasibility, it should be valuable to explore and compare the role of TSE-DWI in the diagnosis and characterization of SPLs.

The objective of this study was to explore and compare the diagnostic efficacy of the lesion-to-spinal cord signal intensity ratio (LSR) method, diffusion, and perfusion parameters derived from multi-b-value TSE-DWI in patients with SPLs. Furthermore, to the best of our knowledge, this study represents the initial utilization of multi-b-value TSE-DWI and its demonstration of diagnostic efficacy in lung cancer. We believe that the proposed method is a highly promising alternative to lung diffusion-weighted magnetic resonance imaging currently used in clinical practice.

Materials and methods

Patient population

This study was approved by the institutional review board of Guangdong Second Provincial General Hospital (protocol number 2023-DW-KZ-060–02), and written informed consent was obtained from all enrollees. Patients with solitary pulmonary lesions detected by chest CT at our institution from March 2018 to August 2023. The inclusion criteria were as follows: (a) a maximum tumor diameter of more than 10 mm as measured on chest CT images, and with ground glass, necrosis, calcification, and cavity components comprising less than one-third of the lesion; (b) no contraindication for MRI examinations; and (c) patients who had not undergone any treatments. A total of 105 patients (66 males and 39 females; age range, 19–78 years; mean age, 54 years) diagnosed with SPLs through chest CT examination, including 77 malignant tumors (47 invasive adenocarcinomas, 15 squamous carcinomas, six small-cell carcinomas, five lymphoid epithelioid carcinomas, two mucosa-associated lymphoid tissue lymphomas, one carcinoid, and one mucoepidermoid carcinoma) and 28 benign lesions (two hamartomas, five tuberculous granuloma, five cryptococcal granuloma, and 16 chronic organizing inflammation) were enrolled in our study after excluding 32 patients according to exclusion criteria (Fig. 1). The pathological information was obtained or planned by surgical resection or transbronchial or transthoracic biopsy, and the interval between the lung DWI and biopsy was less than 30 days.

Figure 1
figure 1

Flow diagram of the study. Note- DWI = diffusion weighted imaging; TSE = turbo spin echo.

Image acquisition and postprocessing

With the use of a 3.0 T MRI scanner equipped with a 16-channel body phased array coil (Achieva; Philips Healthcare, Best, the Netherlands), axial T1-weighted imaging, T2-weighted imaging, and DWI were performed for all examinations. DWI was acquired during free breathing using a single-shot TSE–based sequence (Table 1).

Table 1 Scanning parameters of magnetic resonance imaging.

DWI data were postprocessed using PRIDE software (Philips Healthcare) to generate apparent diffusion coefficient (ADC) maps with b values of 0 and 800 s/mm2. In IMAge/enGINE MRI Diffusion Toolbox (Beta V2.0.3; Vusion Tech, Hefei, China) with monoexponential and biexponential fit models, ADCtotal and IVIM parameters were calculated, respectively, for all b values from 0 to 800 s/mm2.

The ADC value was calculated based on the following equation:

$$ S\left( b \right)/S_{0} \, = \,{\text{exp}}( - \,b{\text{ADC}}) $$

According to the biexponential IVIM model, the correlation between DWI signal intensity (SI) and the b-factor is as follows4:

$$ S\left( b \right)/{\text{S}}_{0} \, = \,{\text{f}}.{\text{exp}}( - \,b{\text{D}}_{{{\text{star}}}} )\, + \,\left( {{\text{1}}\, - \,{\text{f}}} \right).{\text{exp}}( - \,b{\text{D}}) $$

where S0 and S(b) are the diffusion-weighted signal intensities for b = 0 s/mm2 and a preset b value, respectively.

All regions of interest (ROIs) were manually drawn in the solid lesion on the slice with maximum transverse diameter, avoiding necrosis and hemorrhage, as well as the spinal cord (12–16 mm2) for mean SI measurements, under the guidance of an experienced radiologist. LSR was calculated using the following equation14:

$$ {\text{LSR}}\, = \,{\text{SI}}_{{{\text{lesion}}}} /{\text{SI}}_{{{\text{spinal cord}}}} $$

Statistical analysis

Data that followed a normal distribution were reported as mean (standard deviation), and data that followed a non-normal distribution were reported as median (interquartile range). With the use of SPSS (version 22.0, IBM Corporation, USA) and MedCalc software (version 18.2.1, Mariakerke, Belgium), the Mann–Whitney U test and receiver operating characteristic analysis were performed. A p value less than 0.05 was considered statistically significant.

Ethics approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Informed consent

This retrospective study was approved by the medical ethical committee of Guangdong Second Provincial General Hospital, and informed written consent was obtained from all enrollees.

Results

MRI appearance

The maximum diameters of the SPLs at the largest level ranged from 1.000 to 9.800 cm, with a mean of 1.700 cm. Figures 2 and 3 depict the representative multi-b-value TSE-IVIM-DWI appearance of two SPLs pathologically confirmed to be benign and malignant.

Figure 2
figure 2

Pulmonary invasive adenocarcinoma in the right middle lobe. (a) Diffusion-weighted imaging (DWI) (b = 800 s/mm2); the signal intensity (SI) of the lesion was increased, and the lesion-spinal signal intensity ratio (LSR) was 1.230. (b) ADC800 map. (c) ADCtotal map. (d) Pure diffusion coefficient (D) map. (e) Perfusion fraction (f) map. (f) Perfusion-related diffusion coefficient (Dstar) map.

Figure 3
figure 3

Pulmonary hamartoma in the right upper lobe. (a) Diffusion-weighted imaging (DWI) (b = 800 s/mm2); the signal intensity (SI) of the lesion was decreased, and the lesion-spinal signal intensity ratio (LSR) was 0.530. (b) ADC800 map. (c) ADCtotal map. (d) Pure diffusion coefficient (D) map. (e) Perfusion fraction (f) map. (f) Perfusion-related diffusion coefficient (Dstar) map.

Diagnostic performance of multiple parameters

Table 2 presents the LSR, ADC800, ADCtotal, D, f, and Dstar values for both lung cancer and benign groups.

Table 2 Diffusion parameters of malignant and benign solitary pulmonary lesions.

Figure 4 shows the receiver operating characteristic curves used to predict lung cancer. The AUC values for LSR and ADC800 were determined to be 0.789 and 0.824, respectively; the corresponding cutoff value, sensitivity, and specificity were found to be 0.954, 67.5%, and 82.1%, respectively, for LSR and 1.268, 68.8%, and 89.3%, respectively, for ADC800. In addition, the AUC values for ADCtotal and D were 0.844 and 0.887, respectively; the corresponding cutoff value, sensitivity, and specificity were 1.131, 75.3%, and 82.1%, respectively, for ADCtotal and 1.182, 67.5%, and 96.4%, respectively, for D (Table 3). However, there were no significant differences in diagnostic efficacy between LSR, ADC800, ADCtotal, and D, as presented earlier.

Figure 4
figure 4

Results of receiver operating characteristic analysis for lesion-spinal signal intensity ratio (LSR), ADC800, ADCtotal, pure diffusion coefficient (D), perfusion fraction (f), and perfusion-related diffusion coefficient (Dstar) derived from TSE-DWI in differentiating solitary pulmonary lesions.

Table 3 Sensitivity and specificity of diffusion parameters at optimal cutoff values in differentiating malignant from benign solitary pulmonary lesions.

Discussion

Lung cancer remains a major threat to human health. The results showed that the TSE technique is considerable clinical promise as an alternative to diffusion-weighted magnetic resonance imaging of the lung. Furthermore, the findings of this study demonstrated that multi-b-value TSE-DWI is both clinically feasible and valuable in distinguishing between malignant and benign SPLs. The results also showed that lung cancer had notably lower values of diffusion-related parameters D and ADC, as well as a higher LSR value, than did benign lesions. This difference can be attributed to the high cell density of malignant tumor tissues, which results in increased brightness on DWI due to the restricted diffusion.

This study was performed using TSE-based acquisition techniques, in contrast to most existing lung DWI studies. EPI is still the clinical mainstream technique for DWI because of its extensive usage3. Although echo-planar acquisition has the advantage of reducing the acquisition time for DWI, it still presents significant limitations, especially for some specific site applications, such as lung imaging. These limitations include susceptibility to motion artifacts and geometric distortion caused by magnetic field inhomogeneity, which are more prominent in precise lung DWI. This is primarily attributed to the accumulation of phase errors in echo planar–based acquisition, which can result in signal loss or image distortion. There are some ways to reduce geometric distortion, such as reverse phase encoding to correct EPI-related distortion15 and replacing EPI with TSE16. Diffusion-weighted TSE is less sensitive to magnetic inhomogeneity existing at the air–lung interface and exhibits better geometrical accuracy than EPI-DWI17.

Our study found that diffusion-related parameters derived from TSE-IVIM-DWI, D, and ADC values could distinguish between malignant and benign SPLs and possess higher diagnostic accuracy than Dstar and f, consistent with most lung DWI studies using echo planar–based acquisition techniques18,19. Furthermore, the AUC of D was slightly higher than that of ADCtotal, followed by ADC800. Also, D had the highest specificity, and ADC achieved the highest sensitivity, in line with some studies6, but not with others18. These inconsistent results may be caused by the bias of the included cases and lesion size and variance in acquisition techniques. Nevertheless, these findings suggest that combining D and ADC in clinical practice could be more accurate than using either index alone.

IVIM perfusion-related Dstar and f values did not demonstrate a good diagnostic efficacy for differentiating malignant SPLs from benign SPLs, consistent with most previous studies6,18,19, but not the same as the studies conducted by Deng et al.20 and Wang et al.21. The inconsistencies between these studies may be associated with the diverse pathological types included in each study and the poor measurement reproducibility of Dstar and f. Ultimately, Dstar and f values could be used to obtain tissue perfusion information without the injection of contrast agents, and this possibility should be confirmed in future studies.

In our study, we found that the LSR method can be valuable in distinguishing malignant SPLs from benign SPLs. The LSR values exhibit a notable increase in lung cancer cases compared with the benign group, aligning with previous research findings14,22. The LSR method demonstrates satisfactory diagnostic precision and presents itself as a practical and efficacious alternative for regular clinical implementation. However, the reproducibility of the LSR method may be a challenge, as it does rely on subjective evaluation.

This study has several limitations. First, echo planar–based acquisition, the most common DWI technique, was not performed as a control group, which may have resulted in excessive scan times. A small preliminary pilot study by our group indicated no statistically significant difference in the diagnostic efficacy of EPI-DWI and TSE-DWI16. Second, only nodules larger than 10 mm in diameter and pathologically confirmed were included; including all nodules encountered in clinical practice may yield different results. In addition, the analysis solely encompassed a comparison between lung cancer and benign SPL groups and did not consider the classification of benign lesions as benign tumors or inflammatory granulomas. Furthermore, additional investigation is necessary to explore various pathological types and degrees of differentiation in malignant lesions.

In conclusion, lung DWI using TSE-based acquisition techniques is feasible for differentiating malignant SPLs from benign SPLs. Both the LSR method and conventional DWI perform similarly to IVIM in the assessment of SPLs.