Abstract
An electron beam of very high energy (50–250 MeV) can potentially produce a more favourable radiotherapy dose distribution compared to a state-of-the-art photon based radiotherapy technique. To produce an electron beam of sufficiently high energy to allow for a long penetration depth (several cm), very large accelerating structures are needed when using conventional radio-frequency technology, which may not be possible due to economical or spatial constraints. In this paper, we show transport and focusing of laser wakefield accelerated electron beams with a maximum energy of 160 MeV using electromagnetic quadrupole magnets in a point-to-point imaging configuration, yielding a spatial uncertainty of less than 0.1 mm, a total charge variation below \(1 \%\) and a focal spot of \(2.3 \times 2.6\;{\text {mm}}^2\). The electron beam was focused to control the depth dose distribution and to improve the dose conformality inside a phantom of cast acrylic slabs and radiochromic film. The phantom was irradiated from 36 different angles to obtain a dose distribution mimicking a stereotactic radiotherapy treatment, with a peak fractional dose of 2.72 Gy and a total maximum dose of 65 Gy. This was achieved with realistic constraints, including 23 cm of propagation through air before any dose deposition in the phantom.
Similar content being viewed by others
Introduction
Over the past decades, electron beams have mainly been used for treatment of superficial lesions due to the limited range of available electron energies, approximately 4–25 MeV1,2. However, studies have shown a potential for very high energy electrons (VHEE)3,4, ranging in energies from 50 to 250 MeV, in the treatment of lung cancers5,6, prostate cancers6,7,8, pediatric brain tumors and head and neck cancers6. Compared to photons, VHEE have some advantages for more deeply seated tumors. VHEE exhibits a more reliable dose deposition when there are inhomogeneities in the beam path and a sharper cut off in the dose depth profile, sparing healthy tissue more effectively6. Another more practical advantage is the possibility for beam steering that comes with VHEE. Due to the relatively low scattering and high penetration this makes VHEE suitable for a pencil beam electromagnetic scanning technique to irradiate the tumor. This does not require patient-specific apertures, saving time and reducing cost, and can have greater precision compared to multileaf collimators9. The possibility for other irradiation schemes have recently been demonstrated, such as multi-field and intensity modulation10. Furthermore, the low scattering may also allow for a higher dose conformality compared to photons11.
Achieving a VHEE beam with a traditional radio frequency-based linear accelerator requires large and costly structures as the accelerating electric fields are limited due to electrical breakdown (\(<100\) MV/m). This has led to an interest in using laser-plasma based accelerators with stronger accelerating fields on the order of 100 GV/m, reducing the accelerating distance with several orders of magnitude for a given particle energy, which could result in a more compact and cost efficient solution.
Laser-plasma based accelerators can, with some modifications, accelerate different particle species, such as electrons12, positrons13, protons14 and ions15. At the current stage, only electrons are relevant for radiotherapy, as the acceleration of other particle species are limited in beam quality, having too low energy to be clinically usable because of their limited tissue penetration. One of the laser-plasma based accelerator types is the laser wakefield accelerator12,16,17,18,19 (LWFA). At the Lund High-Power Laser Facility, LWFA is used to accelerate electrons and is accomplished by focusing a high-power laser pulse onto a gas target. Having a high intensity at the focus, on the order of \(10^{18}\;{\text {W/cm}}^2\), the leading edge of the pulse creates a plasma, a mixture of ions and free electrons, that the main part of the pulse interacts with. Due to the ponderomotive force, electrons are pushed out from high intensity regions which creates a region void of electrons inside the plasma, trailing the laser pulse like a wake. This is referred to as “the blowout regime”, in which a very strong electrostatic potential gradient is formed. Some electrons may be trapped within the wake and be accelerated by the electrostatic field over a few mm to reach hundreds of MeV.
Electron beams of different characteristics may be produced depending on the injection mechanism, i.e. the process of trapping electrons inside the plasma wake20,21,22,23,24,25. Common to all injection mechanisms is that the electron beam charge and energy depends on the laser pulse characteristics and, in general, a higher laser power can produce electron beams of higher charge while a higher laser energy usually results in higher electron energy. In this experiment, ionization injection26 was used which gives a large amount of charge (several tens to hundreds of pC), a very broad energy spectrum and an ultra short pulse duration of a few fs. Other injection schemes, such as shockfront injection, gives only a few tens of pC in charge for a 50 TW laser but a more narrow-band beam (\(\Delta E \approx 5\) MeV)23. Ionization injection is easy to implement and one of the more robust and reproducible injection schemes which made it suitable for this experiment.
In this work, we present our advancements towards making a LWFA source suitable for fractionated stereotactic VHEE radiotherapy, i.e. the total dose is delivered in fractions where each fraction is delivered by irradiating the tumor over several different angles. Typically only one fraction per day is delivered as it allows healthy tissue to self-repair. Specifically, we report a large decrease in charge fluctuation and spatial uncertainty by using electromagnetic quadrupoles (EMQs) in a point-to-point imaging configuration. By focusing the electron beam inside a phantom using the EMQs, a more deeply penetrating beam is achieved which is used to create a 3D dose distribution similar to a typical stereotactic radiotherapy treatment.
Method
This experiment was conducted at the Lund High-Power Laser Facility which is a titanium-doped sapphire system, having an energy on target of 600 mJ, pulse duration of 37 fs, and focused to a focal spot size of 12 \(\upmu\)m full width half maximum (FWHM) using an off-axis parabolic mirror with an effective focal length of 775 mm, where it reaches a peak intensity of \(10^{18}\;{\text {W/cm}}^2\).
The experimental setup is shown in Fig. 1. Using a supersonic gas jet with a 1.5 mm orifice, a gas mixture of helium containing \(1\%\) nitrogen is ejected at a backing pressure of 4 bar which corresponds to a peak electron density of approximately \(4\times 10^{18}\;{\text {cm}}^{-3}\) which, in our case, allowed for stable and reproducible ionization injected electron beams. For characterizing the electron energy spectrum before the EMQs, there is a translatable 200 mm long dipole magnet with a peak magnetic field of 0.8 T, which is used to disperse the electron beams onto a scintillating screen. The screen is imaged using a 16-bit scientific CMOS camera (Andor Zyla 4.2 Plus, Oxford Instruments, Oxfordshire, UK).
An extending vacuum pipe passes through the 3 EMQs, keeping the electron beam in vacuum until the pipe is terminated with a Kapton window, 29 cm after the last EMQ. Another scintillating screen is used to image the electron beam focus, along with a translatable 0.68 T dipole magnet to characterize the transported and focused electron beam energy spectrum. Both dipoles along with the scintillating screen were removed for radiochromic film measurements.
Gafchromic EBT3 films (Ashland Inc., Wilmington, United States) were cut into 8 cm diameter circles and 5 cm strips using scissors with care to minimize splitting of the film-edges as the films were irradiated from the side. Seven circular films were each positioned between a pair of 2 mm thick cast acrylic slabs (Perspex) in a stacked cylindrical phantom, which was then clamped to remove any air gaps which could lead to an over-response27. The water equivalence of cast acrylic is 1.045, i.e. 1 mm of acrylic is dosimetrically equivalent to 1.045 mm of water for an electron beam28.
The films were scanned 24 h after irradiation using an Epson Perfection V800 Photo scanner (Epson, Suwa, Japan) in transmission mode with no color correction, 1200 dpi and 48 bit color. The Epson V800 scanner has previously been compared to the more commonly used Epson 10000XL29 and shown to have 0.3% lower dose uncertainty but a stronger lateral response function, 0.2% over 8 cm and 2.7% less sensitivity on the relevant red channel30. The films were analyzed using the polynomial netOD method31 taking the red channel, calculating the dose as \({\mathcal {P}}({\text {log}}_{10}(A_{pre}/A_{post}))\) where \({\mathcal {P}}\) is the polynomial fit to the dose calibration curve and \(A_{pre}\), \(A_{post}\) are the red channel values of the EBT3 film pre- and post-irradiation. The calibration films (\(5 \times 5\) cm) were irradiated by a clinical 10 MeV electron beam, delivered by an ELEKTA Precise linear accelerator (Elekta AB, Stockholm, Sweden), calibrated with an ion chamber according to the IAEA standard32. During the irradiation, the actual output was \(-0.1\%\) compared to the expected output.
As previously shown, the Epson Perfection V800 Photo scanner has a more non-uniform response (i.e. uneven illumination across the scanner) due to its smaller size compared to the Epson 10000XL30. This was characterized and corrected by scanning large unirradiated EBT3 films, fitting a 2D cubic function which was used to correct all the scanned films.
Results
By using EMQs, the VHEE electron beam pointing (standard deviation of the electron beam offset to the laser axis, measured 169 cm from the source) improved by more than an order of magnitude (from 1.95 to 0.09 mm), and the electron beam spatial profile a factor 3 horizontally and 2 vertically (from \(7.8 \times 5.1\) to \(2.3 \times 2.6\) mm, respectively). The standard deviation in photon counts from the scintillating screen, which is proportional to the total amount of charge (found by integrating the electron beam energy spectrum only considering energies above 30 MeV in both the pre- and post-focused electron beam energy spectra), was reduced by a factor 5 (from 5.16 to \(0.94\%\)).
The ionization injection results in a very broad electron energy spectrum, see Fig. 2a,b. For our beam, that resulted in a maximum energy of 160 MeV with the peak in charge at the higher end of the spectrum located at 90 MeV, which was the energy the EMQs were set to focus which can be seen (in orange) in Fig. 2d. The energy spectrum from an ionization injected beam is fairly stable in charge with a low standard deviation (SD) and largely populated by electrons of 40 MeV and below. This part was however lost in the EMQs during the electron beam transport, and the resulting energy spectrum is Gaussian-like with a 40 MeV cut-off and a focus at 95 MeV, as shown in Fig. 2c,d. This energy is close to the 100 MeV that has previously been shown to be superior to state-of-the-art volumetric modulated arc therapy treatment (VMAT) plans6 and the pre-focused electron beam spectra peaks slightly at this energy, allowing for a focused electron beam with higher charge. The SD of the averaged focused electron beam spectra (average of seven pulses) was larger compared to the SD of the averaged electron beam before being focused (eight pulses), the SD is indicated by a blue area in Fig. 2b,d. Although the total charge remains more stable (SD of \(0.94\%\)), suggesting that only the shape of the spectrum is changing.
By varying the focal position at the phantom (a single sheath of EBT3 clamped between a pair of 2 mm thick cast acrylic slabs, radiated from the side), the depth dose profile could be tuned to some extent. By focusing at the exit of the phantom, a higher dose may be reached in the central region, see Fig. 3a–c, which shows the average dose profiles for a phantom with varying focal position of the electron beam. For each focal length, 5 VHEE pulses were delivered. The position where \(50\%\) of the max dose was delivered, \(R_{50}\), was moved 14 mm deeper in the phantom by focusing at the back of the phantom, see Fig. 3d. The beam penumbra was smallest when focusing at the entrance and slightly asymmetric, 0.5 on one side of the beam axis and 0.67 mm on the other. When focusing at the exit the beam penumbra was slightly larger and more asymmetric, 1.6 mm and 0.76. When focusing at the exit, \(R_{50}\) was positioned at 34.9 mm, \(R_{80}\) at 4.5 and \(R_{90}\) at 2.2 mm. The maximum dose was located at the phantom entrance except when focusing 3 cm from the phantom entrance, here \(R_{100}\) was reached at 7.2 mm. The highest dose per pulse was reached when focusing at the phantom exit with 0.4 Gy at \(R_{100}\) for a single pulse and the dose in the target area (40 mm from entrance) was 0.13 Gy per pulse. Focusing at the exit resulted in a \(50\%\) higher maximum dose at \(R_{50}\), with a wider lateral profile with a FWHM of 2.97 mm, compared to focusing at the entrance with a FWHM of 1.04 mm, see Fig. 3e.
A cylindrical phantom consisting of Perspex with seven layers of EBT3 disks (8 cm in diameter) located at different heights (0, 2.1, 2.3, 2.4, 2.6, 2.7 and 4.8 mm) was irradiated to measure the delivered dose. To imitate a stereotactical case, a bean-shaped target volume of \(0.07\;{\text {cm}}^3\) and an organ at risk (OAR) volume of \(0.33\;{\text {cm}}^3\) were assumed, see Fig. 4a. The size of the phantom was chosen as a mean of relevant anatomical sites, comparable in depth and size for a head and neck case, smaller than a brain case and larger than a lung case33. For a typical brain metastasis stereotactical treatment34, the target volume is 1–\(20\;{\text {cm}}^3\), 100 times larger compared the target volume in this paper, showing that our beam has sufficient precision for this treatment. Radiating a larger target volume can be achieved with the added (minor) complexity of scanning the electron beam over the target volume. The OAR was chosen to be of a size comparable to that of the optic chiasm and optical nerves for a brain case (typically located at a surface depth of less than 4 cm), the spinal cord for a head and neck case and the bronchial tree for lung cases.
Comparing the target volume in the circular phantom, see Fig. 4a, to a typical fractionated stereotactic radiotherapy treatment for meningiomas, vestibular schwannomas, pituitary adenomas or craniopharyngioma, the target volume and OAR are comparably small . Having a larger target volume would require the electron beam to be scanned over a larger area or be less focused, which is no major complication.
For a single fraction, the maximum dose was confined within the target area and reached 2.72 Gy, using 36 entrance angles with \(5^{\circ }\) increments rotated around an offset of 5.2 mm from the center and three pulses from each angle. The electron beam was focused at the distal edge of the phantom to increase the dose deposition at the centrally located target. The result was a very localized dose deposition where the entrance path contained only 10–20\(\%\) of the maximum dose, see Fig. 4b. The isodose curve for \(70\%\) of the maximum dose was approximately confined within a circle with a diameter of 11.7 mm (not shown). For an unfocused electron beam, the circle was more than double in size, approximately 26 mm in diameter (not shown). By performing 25 fractions of this order, a total dose to the target area would reach 65 Gy, see Fig. 4c, while keeping the peak dose to the organ at risk below 8 Gy.
Discussion
By focusing our electron beam at the end of the rectangular phantom, an increase in dose of \(50\%\) at the position of \(R_{50}\) was achieved while the dose for the first 10 mm remained very similar. Since the beam is polychromatic and the focal point is determined for a single electron energy (90 MeV in our experiment), all energies below 90 MeV will be focused before the focal point for 90 MeV (energies below 50 MeV will however hit the walls of the vacuum tube and be lost). This explains the higher dose seen in Fig. 3c and also why the shape of the depth dose curve for the different cases varies. The electron beam will enter the phantom converging but with lower energies converging more (at shallower depths) and higher energies less (at deeper depths). This works to offset the electron scattering inside the phantom, the lower the electron energy is, the stronger the scattering is but at the same time the focusing is stronger. Focusing at the entrance of the phantom means that all energies below 90 MeV will be diverging when entering the phantom and results in a shallower depth dose curve. As seen in Fig. 2d, the SD of the electron beam charge in the energy spectrum is increased, which is most likely caused by a slight spatial misalignment of the EMQ axis in relation the the electron beam axis and imperfect imaging conditions, resulting in increased angular fluctuations. Another, less likely, possibility would be fluctuations in the electron beam source position before the EMQs. A shift horizontally of the electron beam before the EMQs will be an offset in the horizontal direction afterwards as well, making it appear as a different energy in the spectrometer. To some extent, this would be analogous to coma in the case of an optical beam. There could also be some hysteresis present as the EMQs were not cycled.
The SD of the total amount of charge in the beams decreased from \(5.16\%\) to a much more clinically acceptable level of \(0.94\%\) when the beam is focused with the EMQs. As previously mentioned, this indicates that, while the shape of the energy spectrum changes (likely due to spatial misalignments) the total charge in each pulse is more stable with EMQs. This decrease is most likely due to the fact that the EMQs will only allow electrons within a certain energy band to be transmitted, the rest will be lost in the walls of the vacuum pipe. This makes for a more robust solution compared to using a dipole and an aperture to cut the electron energy spectrum, as any pointing fluctuation will translate into a charge and energy fluctuation. For a monochromatic beam, there would be less losses in the EMQs and all electrons would be focused to the same distance, resulting in an improved control in the position of \(R_{50}\).
There will always be some inherent charge and spectral fluctuation, as seen in the pre-focused electron energy spectrum in Fig. 2b. However, this is random in nature so the SD will decrease as \(1/\sqrt{N}\), where N is the total number of pulses, and thus will benefit from the larger number of pulses required for a single fraction, 108 in this case. A main concern with standard stereotactic treatment is that it takes a long time, making it difficult to keep the patient correctly positioned throughout the treatment. In this case however, the pulse repetition rate could be very high as it mainly depends on the laser system. A kHz system35 could deliver this fraction in 100 ms, “freezing” any potential movement of the target volume assuming it’s possible to irradiate from all angles simultaneously. Irradiating at such ultra-high dose rates would, as preclinical data has shown, have the additional benefit of a decrease in toxicity to normal tissue while maintaining the effectiveness in treating the tumor. This beneficial radiobiological effect is generally known as FLASH36 and has recently received a lot of attention in radiation research. To monitor the delivered dose from each pulse an integrating current transformer could be used, as the ion collection process in an ion chamber is too slow for fs electron pulses and will show an inadequate response37.
Further studies are necessary to investigate the long term stability38 and achieve better statistics of a system such as this and to further improve the VHEE beam. Another injection mechanism could be used, such as shockfront injection, which results in a more monochromatic energy spectrum. The downside of shockfront injection is that it generates less charge per TW of laser power, approximately 5.5 pC/TW of laser power above a 15 TW threshold39 compared to 9.2 pC/TW above 6.8 TW threshold for ionization injection40. Since all electrons below approximately 50 MeV are lost in the EMQs during transport anyway, this disadvantage may not be an issue as shockfront injection has a much smaller energy spread and the electron beam charge will be better preserved during beam transport and focusing. Shockfront injection produced beams do however suffer from some fluctuation in mean energy, which could be an issue during beam transport and focusing. Another possibility is self-truncated ionization injection41 which produces electron beams of high charge and an energy spread of only a few percent, although these beams also suffer from fluctuations in mean energy. Another possibility for lasers of moderate power, such as the one used in this study, would be to cut the desired energy from the broad ionization injection energy spectrum using a slit and focus this part. To mitigate the pointing fluctuation of the electron beam one could use a plasma lens42,43 to image the beam on to a slit positioned in the middle of a chicane, i.e. two dipoles to disperse the beam, followed by two more to recombine it. By imaging the beam onto the slit, the energy selection is insensitive to electron beam pointing from the source. Alternatively, solving the beam pointing at the source44 would make the plasma lens redundant. Once the beam is recombined at the end of the chicane it can be imaged onto a phantom using EMQs in a point-to-point imaging configuration. Having a motorized slit would allow for some additional flexibility in central energy selection and transmitted bandwidth. Although this takes an additional set of beam optics and is more complicated, it is possibly the most stable solution.
The number of patients requiring radiotherapy, in Europe alone, is expected to reach 2 million over the next 4 years45. As mentioned previously, VHEE plans have been shown to be similar or superior to state-of-the-art VMAT plans due to the reduced scattering, sharper penumbra and insensitive to tissue inhomogeneities, offering a significant sparing to OARs. In this comparison a VHEE beam of 100 MeV, 3 mm FWHM and emittance of 0.3\(\deg\) was used6. A focused electron beam such as ours will further improve the dose conformality and OAR sparing due to the reduced surface and exit dose, increased dose depth and smaller electron beam size at the target.
Conclusion
In this paper, we have shown that combining ionization injection LWFA and EMQs in a point-to-point imaging configuration can result in a clinically acceptable focused VHEE electron beam, with less than \(1\%\) charge uncertainty, 0.1 mm spatial uncertainty and a spatial size of \(2.3 \times 2.6\;{\text {mm}}^2\) FWHM after propagating through 23 cm of air. In addition, we have shown that the EMQs can be used to control the position of \(R_{50}\) in the phantom by focusing at different longitudinal positions, moving \(R_{50}\) up to 14 mm deeper.
Finally, we demonstrated the potential of our electron beam by using a cylindrical phantom to mimic a fractionated stereotactic radiotherapy treatment of a small intracranial tumor. The resulting peak fractional dose delivered to the target was 2.72 Gy which, with a total of 25 fractions, corresponds to a total maximum dose of 65 Gy to the target area while keeping the maximum dose to the OAR below 8 Gy.
We believe that this study represents a step towards clinical trials with VHEE beams generated by LWFA which can increase the availability and efficiency along with reducing the cost of radiotherapy treatment in the future.
References
DesRosiers, C., Moskvin, V., Bielajew, A. & Papiez, L. 150–250 mev electron beams in radiation therapy. Phys. Med. Biol. 45, 1781–1805. https://doi.org/10.1088/0031-9155/45/7/306 (2000).
Hogstrom, K. R. & Almond, P. R. Review of electron beam therapy physics. Phys. Med. Biol. 51, R455–R489. https://doi.org/10.1088/0031-9155/51/13/r25 (2006).
Glinec, Y. et al. Radiotherapy with laser-plasma accelerators: Monte carlo simulation of dose deposited by an experimental quasimonoenergetic electron beam. Med. Phys. 33, 155–162 (2006).
Lundh, O. et al. Comparison of measured with calculated dose distribution from a 120-mev electron beam from a laser-plasma accelerator. Med. Phys. 39, 3501–3508 (2012).
DesRosiers, C., Moskvin, V., Cao, M., Joshi, C. & Langer, M. Lung tumor treatment with very high energy electron beams of 150–250 mev as compared to conventional megavoltage photon beams. Int. J. Radiat. Oncol. Biol. Phys 72, S612. https://doi.org/10.1016/j.ijrobp.2008.06.243 (2008).
Schüler, E. et al. Very high-energy electron (vhee) beams in radiation therapy; treatment plan comparison between vhee, vmat, and ppbs. Med. Phys. 44, 2544–2555. https://doi.org/10.1002/mp.12233.Epub2017May4 (2017).
Desrosiers, C., Moskvin, V., Cao, M., Joshi, C. J. & Langer, M. Laser-plasma generated very high energy electrons in radiation therapy of the prostate. In Commercial and Biomedical Applications of Ultrafast Lasers VIII Vol. 6881 (eds Neev, J. et al.) 49–62 (International Society for Optics and Photonics SPIE, 2008). https://doi.org/10.1117/12.761663.
Fuchs, T. et al. Treatment planning for laser-accelerated very-high energy electrons. Phys. Med. Biol. 54, 3315–3328. https://doi.org/10.1088/0031-9155/54/11/003 (2009).
Bazalova-Carter, M. et al. Treatment planning for radiotherapy with very high-energy electron beams and comparison of vhee and vmat plans. Med. Phys. 42, 2615–2625. https://doi.org/10.1118/1.4918923 (2015).
Labate, L. et al. Toward an effective use of laser-driven very high energy electrons for radiotherapy: Feasibility assessment of multi-field and intensity modulation irradiation schemes. Sci. Rep. 10, 17307. https://doi.org/10.1038/s41598-020-74256-w (2020).
Kokurewicz, K. et al. Focused very high-energy electron beams as a novel radiotherapy modality for producing high-dose volumetric elements. Sci. Rep. 9, 10837. https://doi.org/10.1038/s41598-019-46630-w (2019).
Tajima, T. & Dawson, J. M. Laser electron accelerator. Phys. Rev. Lett. 43, 267–270. https://doi.org/10.1103/PhysRevLett.43.267 (1979).
Sarri, G. et al. Generation of neutral and high-density electron-positron pair plasmas in the laboratory. Nat. Commun. 6, 6747. https://doi.org/10.1038/ncomms7747 (2015).
Clark, E. L. et al. Measurements of energetic proton transport through magnetized plasma from intense laser interactions with solids. Phys. Rev. Lett. 84, 670–673. https://doi.org/10.1103/PhysRevLett.84.670 (2000).
Clark, E. L. et al. Energetic heavy-ion and proton generation from ultraintense laser-plasma interactions with solids. Phys. Rev. Lett. 85, 1654–1657. https://doi.org/10.1103/PhysRevLett.85.1654 (2000).
Tajima, T. & Dawson, J. Ultrahigh gradient particle acceleration by intense laser-driven plasma density waves. Nature 311, 525–529. https://doi.org/10.1063/1.1148194 (1984).
Mangles, S. P. D. et al. Monoenergetic beams of relativistic electrons from intense laser–plasma interactions. Nature 431, 535–538. https://doi.org/10.1038/nature02939 (2004).
Faure, J. et al. A laser-plasma accelerator producing monoenergetic electron beams. Nature 431, 541–544. https://doi.org/10.1038/nature02963 (2004).
Geddes, C. G. R. et al. High-quality electron beams from a laser wakefield accelerator using plasma-channel guiding. Nature 431, 538–541. https://doi.org/10.1038/nature02900 (2004).
Yu, L.-L. et al. Two-color laser-ionization injection. Phys. Rev. Lett. 112, 125001. https://doi.org/10.1103/PhysRevLett.112.125001 (2014).
Bourgeois, N., Cowley, J. & Hooker, S. Two-pulse ionization injection into quasilinear laser wakefields. Phys. Rev. Lett. 111, 155004. https://doi.org/10.1103/PhysRevLett.111.155004 (2013).
Lehe, R., Lifschitz, A., Davoine, X., Thaury, C. & Malka, V. Optical transverse injection in laser-plasma acceleration. Phys. Rev. Lett. 111, 085005. https://doi.org/10.1103/PhysRevLett.111.085005 (2013).
Buck, A. et al. Shock-front injector for high-quality laser-plasma acceleration. Phys. Rev. Lett. 110, 185006. https://doi.org/10.1103/PhysRevLett.110.185006 (2013).
Schmid, K. et al. Density-transition based electron injector for laser driven wakefield accelerators. Phys. Rev. Accel. Beams https://doi.org/10.1103/PhysRevSTAB.13.091301 (2010).
Faure, J. et al. Controlled injection and acceleration of electrons in plasma wakefields by colliding laser pulses. Nature 444, 737–739. https://doi.org/10.1038/nature05393 (2006).
Oz, E. et al. Ionization-induced electron trapping in ultrarelativistic plasma wakes. Phys. Rev. Lett. 98, 084801. https://doi.org/10.1103/PhysRevLett.98.084801 (2007).
Suchowerska, N., Hoban, P., Butson, M., Davison, A. & Metcalfe, P. Directional dependence in film dosimetry: Radiographic and radiochromic film. Phys. Med. Biol. 46, 1391–1397. https://doi.org/10.1088/0031-9155/46/5/305 (2001).
Mihailescu, D. & Borcia, C. Water equivalency of some plastic materials used in electron dosimetry: A monte carlo investigation. Roman. Rep. Phys. 58, 415–425 (2006).
Spelleken, E., Crowe, S. B., Sutherland, B., Challens, C. & Kairn, T. Accuracy and efficiency of published film dosimetry techniques using a flat-bed scanner and ebt3 film. Aust. Phys. Eng. Sci. Med. 41, 117–128. https://doi.org/10.1007/s13246-018-0620-4 (2018).
Lárraga-Gutiérrez, J. M., García-Garduño, O. A., Treviño-Palacios, C. & Herrera-González, J. A. Evaluation of a ledbased flatbed document scanner for radiochromic film dosimetry in transmission mode. Phys. Medica 47, 86–91. https://doi.org/10.1016/j.ejmp.2018.02.010 (2018).
Aland, T., Kairn, T. & Kenny, J. Evaluation of a gafchromic ebt2 film dosimetry system for radiotherapy quality assurance. Aust. Phys. Eng. Sci. Med. 34, 251–260. https://doi.org/10.1007/s13246-011-0072-6 (2011).
Accuracy Requirements and Uncertainties in Radiotherapy. No. 31 in Human Health Series (International Atomic Energy Agency, Vienna, 2016).
Ammirati, M. et al. A retrospective case series on the usefulness of fractionated stereotactic radiotherapy for benign intracranial tumors. Interdiscip. Neurosurg. 16, 123–128. https://doi.org/10.1016/j.inat.2019.01.013 (2019).
Lippitz, B. et al. Stereotactic radiosurgery in the treatment of brain metastases: The current evidence. Cancer Treat. Rev. 40, 48–59. https://doi.org/10.1016/j.ctrv.2013.05.002 (2014).
Guénot, D. et al. Relativistic electron beams driven by khz single-cycle light pulses. Nat. Photonics 11, 293–296. https://doi.org/10.1038/nphoton.2017.46 (2017).
Wilson, J. D., Hammond, E. M., Higgins, G. S. & Petersson, K. Ultra-high dose rate (flash) radiotherapy: Silver bullet or fool’s gold?. Front. Oncol. 9, 1563. https://doi.org/10.3389/fonc.2019.01563 (2020).
Subiel, A. et al. Challenges of dosimetry of ultra-short pulsed very high energy electron beams. Phys. Medica 42, 327–331. https://doi.org/10.1016/j.ejmp.2017.04.029 (2017).
Maier, A. R. et al. Decoding sources of energy variability in a laser-plasma accelerator. Phys. Rev. X 10, 031039. https://doi.org/10.1103/PhysRevX.10.031039 (2020).
Götzfried, J. et al. Physics of high-charge electron beams in laser-plasma wakefields. Phys. Rev. X 10, 041015. https://doi.org/10.1103/PhysRevX.10.041015 (2020).
Guillaume, E. et al. Physics of fully-loaded laser-plasma accelerators. Phys. Rev. ST Accel. Beams 18, 061301. https://doi.org/10.1103/PhysRevSTAB.18.061301 (2015).
Mirzaie, M. et al. Demonstration of self-truncated ionization injection for gev electron beams. Sci. Rep. 5, 14659. https://doi.org/10.1038/srep14659 (2015).
Thaury, C. et al. Demonstration of relativistic electron beam focusing by a laser-plasma lens. Nat. Commun. 6, 6860. https://doi.org/10.1038/ncomms7860 (2015).
van Tilborg, J. et al. Active plasma lensing for relativistic laser-plasma-accelerated electron beams. Phys. Rev. Lett. 115, 184802. https://doi.org/10.1103/PhysRevLett.115.184802 (2015).
Hansson, M. et al. Enhanced stability of laser wakefield acceleration using dielectric capillary tubes. Phys. Rev. ST Accel. Beams 17, 031303. https://doi.org/10.1103/PhysRevSTAB.17.031303 (2014).
Schüller, A. et al. The european joint research project uhdpulse—Metrology for advanced radiotherapy using particle beams with ultra-high pulse dose rates. Phys. Medica 80, 134–150. https://doi.org/10.1016/j.ejmp.2020.09.020 (2020).
Acknowledgements
We acknowledge the support of the Knut and Alice Wallenberg Foundation (KAW 2019.0318), the Swedish Research Council (2019-04784), the Crafoord Foundation (20170845), Laserlab-Europe (EU-H2020 871124) and ARIES (EU-H2020 730871). The authors would like to thank Emma Spelleken for providing a reference calibration curve for the V800 Epson scanner and Elise Konradsson for performing the irradiation of calibration films.
Funding
Open access funding provided by Lund University.
Author information
Authors and Affiliations
Contributions
K.S., D.G., J.B.S. and O.L. conceived the experiment, K.S., D.G. and J.B.S. built the experimental setup, K.S., D.G. and A.P. conducted the experiment, K.S. analysed the results. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Svendsen, K., Guénot, D., Svensson, J.B. et al. A focused very high energy electron beam for fractionated stereotactic radiotherapy. Sci Rep 11, 5844 (2021). https://doi.org/10.1038/s41598-021-85451-8
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-85451-8
This article is cited by
-
Extending deterministic transport capabilities for very-high and ultra-high energy electron beams
Scientific Reports (2024)
-
First theoretical determination of relative biological effectiveness of very high energy electrons
Scientific Reports (2021)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.