Introduction

The main objective of radiotherapy is to kill cancer cells while sparing normal tissue from damage1 and minimising the creation of secondary cancers. This is usually achieved through modalities such as volumetric modulated arc therapy (VMAT)2, stereotactic radiosurgery (SRS)3, intra-operative radiation therapy (IORT)4,5 or brachytherapy.6 To reduce irradiation of healthy tissue and ensue complete irradiation of tumours7, dose must be delivered precisely in such a way that it conforms to the tumour shape. A common approach is to physically shield healthy tissue and apply multiple beams to irradiate tumours from different directions8. Megavoltage (MV) X-rays (4–25 MV) are the commonly used radiation for treating cancers. As illustrated in Fig. 1, the depth–dose distribution produced by a single X-ray beam is high at the entrance and reduces exponentially with depth. Proton and ion beams can improve the contrast between the doses in the tumour with that in healthy tissue, especially in the distal region, by taking advantage of the Bragg peak9,10. Better conformity is achieved using a spread out Bragg peak (SOBP) that is produced by varying the beam energy (e.g. from 60 to 110 MeV and 135 to 147 MeV). However, this results in a build-up of an anterior dose11, as shown in Fig. 1. Moreover, the dose distribution is sensitive to tissue density inhomogeneity, which results in range uncertainties12 and potentially under-irradiation of the tumour, or over-irradiation of healthy tissue13.

Fig. 1: Simulations of axial dose distributions in a water phantom for different beam geometries and energies.
figure 1

(i) Collimated 6 MV photons, (ii) 60–110 MeV and (iii) 135–147 MeV protons (SOBP), 158 MeV electron beams focused in the horizontal plane with (iv) f/11.2 and (v) f/12.3 (vi) 201 MeV focused in the horizontal plane with f/18.2 and (vii) 200 MeV symmetrically focused with f/1.2. The f-number is a measure of focusing strength defined as f/D, where D is the beam diameter at the focusing optics and f is the focal length of the lens, as in optics. The dose is normalised to the peak dose of each individual depth–dose profile.

Over the last 20 years, very high-energy electrons (VHEEs), in the range of 60–250 MeV14, have been considered as a promising alternative to photons for the treatment of deep-seated tumours15,16,17,18,19. Low-energy electrons (<MeV), from current clinical linacs20, have shallow penetration depths and thus are only suitable for treatment of superficial tumours21 or for intra-operative radiotherapy22. VHEEs, in contrast, are very penetrating due to their high inertia, which enables them to reach deep-seated tumours19,23,24,25,26,27. Their dose is characterised by a sharp transverse penumbra and low scattering at tissue interfaces14,23,28, compared with current low-energy clinical electrons29,30. VHEE beams are also relatively easy to form into pencil beams or scan using electromagnet deflectors15. However, VHEE accelerators are currently relatively large and expensive, compared with 4–22 MeV clinical accelerators, but compact VHEE sources should become available with the development of new technologies such as X-band radio-frequency (RF) accelerators31 and laser-plasma wakefield accelerators32,33,34,35,36.

The depth–dose distribution of collimated VHEE beams is nearly uniform19,26 and several methods have been investigated to improve it. Theoretical and experimental studies have shown that lateral scattering of low-energy (<70 MeV) electron beams can be reduced by imposing a longitudinal or transverse magnetic field on the medium37,38,39, which results in a depth–dose profile that resembles a degraded Bragg peak. Recent Monte Carlo (MC) studies have shown that for VHEEs this dose enhancement at depth can be obtained simply by externally focusing the electron beam using a magnetic lens19. As in optics, we define the f-number as f/D, where D is the beam diameter on the focusing optics and f is the focal length of the lens. Here we present experimental measurements of the depth–dose distribution of VHEE beams focused in a water phantom for several f-numbers and electron energies, and compare these results with theoretical predictions.

Results and discussion

The experiment has been performed at the CERN Linear Electron Accelerator for Research (CLEAR) beamline40, which is based on an S-band RF accelerator that delivers quasi-mono-energetic electron beams with energies between 60 and 220 MeV. The experimental layout is shown in Fig. 2 and more details given in Methods. The electron beam is first expanded by making it diverge over a distance of about 3 m using three quadrupole magnets placed after the last RF deflecting cavity (Fig. 2). The beam is then focused by a second quadrupole triplet into a water phantom, which is placed 21.5 cm downstream of the last quadrupole magnet. The magnetic field strength and aperture size of the quadrupoles only allows focusing with a short focal length in one plane, which is chosen to be horizontal. The beam is collimated or diverges slightly in the vertical plane, resulting in a line focus. Symmetric focusing could have been achieved by upgrading the power supplies of the first quadrupole triplet and employing larger aperture quadrupoles in the second triplet. However, this was not possible in the allocated experimental time. After the final quadrupole magnet, the beam passes through a 0.1 mm thick aluminium window into ambient air, and then through an integrating current transformer (ICT - Bergoz, ICT-055-070-5.0, SN 3350), and finally into the phantom. The ICT records the total charge of the beam entering the phantom. The transverse dose profile is measured at different depths in the phantom using a stack of 16 Gafchromic® films (EBT3, Ashland ISP Advanced Materials) spaced by 1 cm, as discussed in Methods.

Fig. 2: The CLEAR accelerator, beamline and schematic of the experimental setup.
figure 2

The electron beam (travelling from right to left) is produced by a radio-frequency (RF) gun and accelerated up to 220 MeV by three 4.5 m long S-band accelerating structures driven by 2 modulators/klystrons (MKS11 and MKS15). A third klystron (MKS31) can power a transverse deflector. The first structure (Buncher) can also be used to longitudinally compress the electron bunches. Beam diagnostics include a beam position monitor (BPM) and integrating current transformer (ICTs) to measure the charge at different stages. The electron beam energy is measured using a dipole electromagnet that also directs the beam to the VESPER test bench. Two triplets of quadrupole magnets focus the beam inside the water phantom, which is placed at the end of the beamline and is shown in detail in Fig. 5.

Measured dose profiles

Figure 3 shows beam for selected irradiated films spaced by 3 cm (every third film apart from the last film, which is spaced by 2 cm). A 3D simulated reconstruction of the dose distribution is also included, which is based on fits of the sizes measured in the horizontal and vertical planes across the whole film stack. The quadrupole magnetic fields are set to produce a focus at a distance of 14 cm for 158 MeV, and f/11.2 and 10 cm for 158 MeV focused with f/12.3 and 201 MeV focused with f/18.2, from the phantom entrance in air, as measured using an optical transition radiation (OTR) screen.

Fig. 3: 3D reconstruction of the dose profiles based on measured radiation beam profiles using EBT3 films.
figure 3

a 158 MeV with f/11.2, b 158 MeV with f/12.3 and c 201 MeV with f/18.2 focused beams with selected EBT3 film samples placed at depths of 3.1, 6.1, 9.1, 12.1, 14.1 cm in water.

The beam size at the final quadrupole has been varied to obtain different focusing strengths. Here we present results for f/11.2 and f/12.3 (158 MeV), and f/18.2 (201 MeV). Since the CLEAR beamline is not designed for short focal lengths (because of the small diameter of the vacuum pipe), the beam is clipped by the walls of the metal pipe and rectangular drift-chamber in the magnetic spectrometer, resulting in a rectangular beam profile in the vertical plane on the EBT films. In the horizontal plane, the beam profile is approximately Gaussian for f/12.3 and f/18.2, but rectangular for f/11.2. The full-width at half-maximum (FWHM) and depth–dose profiles obtained from the films and FLUKA MC simulations are shown in Fig. 4. The simulated curves are obtained using the beam size measured at the phantom entrance, the focus position in air and the total charge recorded on the ICT monitor, as discussed in Methods. Measurements and simulations show that the maximum of the dose distribution is located about 5–6 cm inside the phantom at a depth corresponding to the position of the beam focus. The smallest spot size measured in the plane where the beam is focused is 3 mm FWHM for 201 MeV energy (Table 1).

Fig. 4: Measured dose distributions.
figure 4

Beam envelope (a–c) and depth dose in the water phantom (d–f) for experimental focused beams. (a, d) 158 MeV focused with f/11.2, (b, e) 158 MeV focused with f/12.3, and (c, f) 201 MeV focused with f/18.2. Solid red curves are simulated depth–dose profiles for beams focused only in the horizontal plane. Dashed green lines are simulated dose profiles obtained for symmetric focusing, while the dashed blue lines represent collimated beams. The depth–dose curves for collimated and symmetrically focused beams are normalised to the peak dose of the line focus. The vertical lines in depth–dose profile for experimental beam indicate measurement uncertainties.

Table 1 Peak dose and full-width at half-maximum (FWHM) obtained from x and y projections of the transverse beam profile at the focus depth, for different beam geometries obtained with the FLUKA MC code.

Our previous theoretical study showed that for symmetrically focused beams the spot size at the focus decreases with f-number19 along the beam direction and increases in the transverse direction, resulting in an asymmetric shape of the high-dose volumetric element. However, for a line focus, as shown in this study, this effect is obscured. As the beam propagates further into the phantom the transverse shape becomes symmetric due to scattering, which also shifts the focus towards the phantom entrance, with respect to its position in air, by 7.6, 2.3 and 4.0 cm, for f/11.2, f/12.3 and f/18.2, respectively.

Simulations have also been performed for a collimated beam, indicated by the blue curves in Fig. 4, with the beam size corresponding to the FWHM in the plane where the beam is focused. Focusing enhances the peak dose, compared with collimated beams, by 6, 4.2 and 8.2 times for f/11.2, f/12.3 (158 MeV) and f/18.2 (201 MeV), respectively (Table 1). For higher energies, symmetric focus shifts the maximum further into the phantom, while the peak dose decreases due to scattering and therefore becomes lower than for a beam focused in one plane for the same incident beam charge. However, for symmetrically focused beams (the green curves in Fig. 4), the peak dose is more pronounced than for the experimental curves and provides a more favourable dose distribution by reducing the proximal and distal dose.

Conclusion

We measured the depth–dose profile of 158 and 201 MeV electron beams focused into a water phantom, demonstrating on-axis dose enhancement at a depth of 5–6 cm, which validates theoretical predictions that focused VHEE beams concentrate dose into a well-defined volume deep in tissue. The dose delivered to surrounding tissue is distributed over a larger volume, which reduces surface and exit doses compared with collimated beams. At fixed electron energy, the increase of the peak dose is more pronounced for smaller f-numbers (f/11.2 for the results presented here). Symmetrically focused beams would require higher charge, but the peak dose can be delivered deeper into the body. VHEE beams are a promising modality in radiotherapy of deep-seated tumours where other modalities may not be appropriate due to strong scattering at tissue interfaces, high skin dose or weak penetration.

The focusing method can also be used to deliver therapeutic doses at high rates (FLASH41 radiotherapy), a regime accessible with the accelerator at the CLEAR user facility. Treatment can be administered using a single beam, potentially reducing the detrimental effects of irradiating healthy tissue, and therefore lowering the risk of radiation-induced secondary cancers, compared with conventional treatment methods combined with FLASH therapy. Deposition of high dose into a small volume by a single beam can also increase precision. Furthermore, similar geometric effects can be observed with other modalities, such as photons and protons, as shown in Supplementary Fig. 1a, b and c, d, respectively. It has been shown previously that a focused photon beam can be produced simply by passing a focused VHEE beam through a bremsstrahlung converter42. The resulting dose distribution of the focused gamma rays is similar to that of a focused VHEE beam42.

Millimetre long laser-plasma wakefield accelerators43 are currently capable of producing quasi-mono-energetic electron beams44 with a few tens of picocoulomb bunch charge45, femtoseconds bunch durations46,47,48 and in the 100 MeV to GeV energy range49. They have been already used for dosimetry studies using VHEE fields, which show excellent agreement with EBT3 film responses using clinical accelerators26. Most medical accelerators are based on 3 GHz, S-band cavities with accelerating gradients less than 100 M/V. However, X-band linacs, operating at 12 GHz, can have higher accelerating gradients, and are potentially more compact. The experiments presented here have been carried out in a laboratory environment, which differs from a clinical setting. However, production of secondary radiation is largely absent because beam-shaping components, as used in conventional clinical accelerators, are not required. Focusing of the electron beam only requires quadrupole magnets, which do not intercept the beam.

Methods

Accelerator and beam focusing system

The experiment has been conducted at the CERN CLEAR beamline. An electron beam with an energy up to 220 MeV is passed through two sets of triplet electromagnetic quadrupoles (EMQs), with specifications given in Table 2, and focused inside a phantom (30 × 30 × 30 cm3 PMMA tank filled with water) mounted on a movable stage to allow precise translation in the direction perpendicular to the beam propagation axis (Fig. 5).

Table 2 Quadrupole magnet specification.
Fig. 5: Experimental setup.
figure 5

Experimental setup (a) and EBT3 stack in the custom made holder (b). (1) focusing magnet, (2) vacuum window, (3) laser for beam alignment, (4) lead collimator, which was not used during irradiation, (5) ICT for charge measurement, (6) PMMA tank, (7) OTR screen, (8) motorised translation stage for transverse displacements of the PMMA tank, (9) CCD camera, (10) beam dump.

When the phantom is moved out, the electron beam passes through an OTR screen50 mounted on a stage that can also be translated along the beamline axis. This is imaged onto a CCD camera, which enables measurement of the beam envelope in air. The phantom is placed at the end of the beamline and the electron beam profile is measured at different depths using a stack of 16 EBT3 films immersed in water and spaced by 1 cm. These are cut into 21.5 × 6.75 cm2 strips, which enables several shots to be recorded on the same strip simply by moving the phantom transversally. The water phantom entrance is placed 32.8 cm from the centre of the last quadrupole. The electron beam passes through a 0.1 mm thick aluminium window and propagates in air for 21.5 cm, passing through a charge monitor unit prior to the phantom, as shown in Fig. 5a. Measurements were carried out for 158 and 201 MeV electron energies. In each case, the electron beam Twiss parameters51,52 (Table 3) were measured using the quadrupole scan technique53 and the beam transport is optimised using the software package MAD-X54. Reconstructions of the beam envelopes based on beam Twiss parameters and optics for all beam geometries are shown in Supplementary Fig. 2.

Table 3 Optimised Twiss parameters and normalised emittance of the electron beams.

As the CLEAR beamline is not designed for short focal lengths, it was decided to focus the beam only in one plane, which produces a line focus at a chosen reference depth (in vacuum) and inside the water phantom. The EMQ settings for all f-numbers are shown in Table 4. The focal length used to calculate the f-number is defined as the distance from the centre of the final quadrupole to the focal spot position.

Table 4 Optimised gradients of the quadrupole magnets Q350-Q520.

Dosimetry

EBT3 films are irradiated with a charge of 0.92 nC (8 shots) for f/11.2, 0.58 nC (10 shots) for f/12.3 and 0.62 nC (10 shots) for f/18.2, measured with the uncertainty of 2%. After exposure, the films are read using an Epson Expression 10000XL flatbed scanner. The digitised images are processed using a 2D Wiener filter to remove scanner noise. The background is defined as the average pixel value of four regions with dimensions of 5 × 5 pixels at the corners of each spot. The dose is calculated as the peak value using the red colour channel, which provides the most sensitive response up to 8 Gy, using the formula

$$D = a \cdot {\mathrm{netOD}} + b \cdot {\mathrm{netOD}}^n,$$
(1)

where netOD is the change in optical density before and after irradiation, given by netOD = log10(PVbefore/PVafter), where PV is a pixel value. The other coefficients are a = 9.1, b = 58 and n = 3.2, obtained by calibrating films from the same batch (lot 06141702) at the Physikalisch-Technische Bundesanstalt (PTB) research accelerator that produces electrons with energies up to 50 MeV55.

Monte Carlo simulations

MC simulations with the code FLUKA56 have been performed to model the depth dose distribution in water for the same configurations as used in the experiments. The beam size at the front of the water tank is taken as the FWHM of the dose profile measured on the corresponding EBT3 film, whereas the focus position is determined from the measurements performed in air using the OTR screen. Simulations show that scattering in air has negligible effect on the position of the focal spot. The OTR diagnostic system was not used to estimate the beam size because it is not configured to store the full 2D beam profile, but only recorded the results of Gaussian fits in the horizontal and vertical plane, respectively. This data was not reliable because the electron beam profile was not Gaussian. The dose deposition is scored using a USRBIN card56 with resolution of 0.025 and 0.079 cm in the transverse and longitudinal directions, respectively. The dose is averaged within the volume of 0.025 × 30 cm2 on the central beam axis. Simulations are performed for 107 histories and 5 cycles. The electron beam is mono-energetic and has a rectangular profile in the vertical plane for all f-numbers. The profile in the horizontal plane is Gaussian for f/12.3 and f/18.2 and rectangular for f/11.2. The PRECISIOn physics setting56 is chosen to provide interaction models for all electromagnetic and nuclear processes that are relevant to the transport of VHEEs, including photonuclear interactions. The threshold for transport and production of delta particles and photons (ECUT, PCUT) is set to 10 keV with the EMCUT card56. Survival probability of a single photon produced via hadronic interactions is set to 0.002. The energy transfer to electrons lower than the threshold is estimated according to the continuous slowing down approximation57.