Abstract
Blood pressure (BP) measurement plays a critical role in cardiac stress testing and is most commonly assessed manually. The emphasis of social distancing during the COVID-19 pandemic has renewed the interest in and the need for an automated BP device for incremental exercise stress testing. We assessed the accuracy of a new automated blood pressure device specifically manufactured for cardiac stress testing. Thirty-five adults aged 35 ± 16 years were studied during an incremental stress test on the cycle ergometer. Three observers measured BP simultaneously, two listening to Korotkoff sounds using a dual-headed stethoscope and one using headphones to listen to sounds generated by an automated BP device. With increasing workload, systolic BP increased progressively without significant differences in BP readings between any observer compared with the automated monitor at any stage during exercise. Systolic BP obtained with the BP machine was strongly correlated with those obtained by the stethoscope observers (r = 0.96) and the observer with headphones (r = 0.95). Diastolic BP obtained with the BP machine was moderately and significantly associated with those obtained by the stethoscope observers (r = 0.75) and the observer with headphones (r = 0.75). The automated BP monitor specifically made for cardiac stress testing accurately measured both systolic and diastolic blood pressure during exercise.
Similar content being viewed by others
Introduction
Blood pressure (BP) is an essential measurement during exercise testing as it is a relative contraindication to terminate exercise [1]. Typically, manual BP measurements by a single technician are used during exercise testing in clinical settings. However, using a manual sphygmomanometer to measure BP during exercise has limitations as noise and motion artifact from the exerciser make these measurements difficult to take [2, 3]. This reduces the accuracy of BP measurements during exercise and could alter their clinical interpretation [4,5,6,7,8,9]. As many world organizations have expressed the importance of proper social distancing during the COVID-19 pandemic, the interest in an accurate and reliable automated BP monitor that can be used during exercise has been renewed.
A new automated BP monitor has recently been developed for the purpose of exercise stress testing. This monitor uses R-wave gating from an electrocardiogram paired with acoustic transduction via a microphone placed over the brachial artery within the BP cuff in order to measure BP. The purpose of the present study is to evaluate the accuracy of this new automated blood pressure device at rest, during exercise, and recovery. These automated measurements were compared to manual blood pressure measurements which served as the reference standard. We hypothesized that the measurements taken by the automated device would not be different from the reference measurements.
Methods
Participants
A total of 35 apparently healthy adults (20 men and 15 women) varying widely in age (19–66 years) were studied (Table 1). Due to limitations placed during the COVID-19 pandemic, we were instructed by the Institutional Review Board to avoid recruiting hypertensive participants as they fell into the “high-risk” category for COVID-19. As such, the International Organization for Standardization (ISO) requirement that 10% of participants must have a resting BP ≥ 140 mmHg was not met. The health status of each participant was determined using a questionnaire. The 2020 PAR-Q + questionnaire was used to assess the safety of exercise. Informed consent was obtained from all participants before involvement in the study. This study was conducted in accordance with the Declaration of Helsinki, and all procedures were approved by the Institutional Review Board of The University of Texas at Austin.
Procedures
Participants came into the laboratory on two separate occasions separated by at least 2 days and at the same time of day. All participants refrained from caffeine on the day of the test, food intake >4 h prior to testing, and alcohol intake and strenuous exercise for >24 h prior to testing. Height and body weight were measured using an electronic balance scale (Seca #769, Vogel & Halke, Hamburg, Germany).
Upon entry into the laboratory, each participant was equipped with a 5-lead ECG, and the circumference of the upper arm was measured to ensure that a properly-sized blood pressure cuff was used. A new automated BP monitor (FBX-1000, Fukuda Denshi, Tokyo, Japan) specifically developed for exercise testing was evaluated during the study. Prior to testing, an aneroid sphygmomanometer was calibrated using a pressure gauge and was found to be within ±2 mmHg. The automated BP monitor was calibrated against the aneroid sphygmomanometer and was within ±2 mmHg.
Each participant was then seated in an armchair with their back, elbows, and forearms supported. All participants were instructed to rest their feet flat on the ground. A BP cuff from the automated BP monitor was placed over the brachial artery of the designated arm, chosen at random. The opposite arm was used during the second day of testing. Following 10 min of seated rest, resting BP was simultaneously measured using three different methods: (1) by two experienced observers using a stethoscope with dual earpieces (one set for each observer) (reference), (2) by the Fukuda automated BP monitor, and (3) by an observer using headphones to listen to sounds generated by the microphone of the Fukuda BP monitor. The armrest of the chair was utilized so the measurement arm could remain relaxed and supported throughout the BP measurements. In addition, the height of the armrest allowed the BP cuff to be positioned at heart level. Participants were asked to keep their arm relaxed and not to tense their shoulder throughout the duration of the BP measurement. Each observer measured BP using the same method throughout the duration of the study and for each participant. The automated BP monitor was connected to a standard aneroid manometer using a Y-tube connector allowing each observer to be blinded to the BP readings of the automated monitor. The appearance of Korotkoff sounds (phase I) defined systolic BP and the disappearance of Korotkoff sounds (phase V) defined diastolic BP. Each observer was blinded to the BP measurements taken by each other. Heart rate was also obtained from the automated BP monitor.
An incremental exercise test was performed on a cycle ergometer (Excalibur, Lode B.V., Groningen, The Netherlands). Exercise began at 25 W and increased by 25 W every 2 min until a rating of perceived exertion (RPE) of 16 (Borg’s original scale) or 80% of age-predicted maximal heart rate [10] was reached. BP was measured one minute into each exercise stage. The measurement arm of the participant was extended to the side so the BP cuff was positioned at heart level and supported by an observer to avoid excessive movement. All participants were asked to keep their arm relaxed during the BP measurement. Heart rate and RPE were measured within the final 15 s of each exercise stage. Immediately upon completion of exercise, the participant was quickly moved from the cycle ergometer and seated into an armchair with their back, elbow, and forearms supported. Heart rate and BP were measured every 2 min until BP returned to resting values ±5 mmHg.
Data analyses
Data analyses were performed using SPSS version 22 (IBM Corp., Armonk, NY). Shapiro–Wilk tests indicated a normal (Gaussian) distribution for all continuous variables. Differences between BP values obtained with the different methods were evaluated using two-factor analysis of variance (ANOVA) with repeated measures (BP methodologies × exercise stage). The LSD post-hoc analysis was used to determine specific differences between the groups. Pearson correlation analyses were conducted to observe relationships between the automated BP devices and the reference manual BP measurements. The Bland-Altman analyses of agreement were then conducted to observe the relative differences between two different methods. P < 0.05 was considered significant for all analyses. Sample size calculation was conducted using a power calculation software (G*Power 3.0.10, Dusseldorf, Germany) with power = 0.8, alpha level = 0.05, and effect size = 0.33.
Results
Selected participant characteristics are shown in Table 1. The participants were relatively diverse in age, sex/gender, and racial distributions. Pearson correlation coefficients for the two observers using a stethoscope were 0.98 for systolic BP and 0.91 for diastolic BP. BP measurements between the two observers using a stethoscope were not significantly different at any stage. Due to the similarity in BP values, the measurements of the two stethoscope observers were averaged to provide one set of reference data in accordance with ISO guidelines. Systolic and diastolic BP at rest were not different between the reference, the automated BP monitor, and the observer using headphones (Fig. 1 and Table 2). Systolic BP showed gradual and significant increases throughout the incremental exercise protocol (p < 0.05, Fig. 1 and Table 2). Diastolic BP increased slightly and significantly during the incremental exercise protocol. There were no significant differences in systolic or diastolic BP between the methods at any individual stage during the incremental exercise protocol.
Systolic BP decreased progressively for the first 4 min of recovery (Fig. 1 and Table 3). Immediately after exercise, the automated BP monitor recorded a higher systolic and diastolic BP than the other two methods (p < 0.05, Fig. 1 and Table 2).
As shown in Fig. 2, systolic BP obtained with the reference was strongly associated with the Fukuda automated BP monitor (p < 0.001) as evidenced by a strong Pearson correlation coefficient of 0.96. Similar significant associations were obtained with diastolic BP although the strength of the association was slightly weaker at 0.75. These agreements are consistent with the Bland-Altman analysis. Comparable results were observed between the observer using headphones and the automated BP monitor (Supplemental Figs. 1 and 2). Systolic and diastolic BP with the reference showed significant associations with the observer listening via headphones as shown by strong Pearson correlation coefficients of 0.97 and 0.90 and Bland-Altman analysis.
During exercise, all participants obtained a heart rate at least 30% greater than their heart rate at rest. Heart rate increased progressively and significantly throughout the incremental exercise protocol (p < 0.05, Table 2). Heart rate significantly decreased throughout recovery from exercise (p < 0.05).
Discussion
In the present study, the accuracy of a newly-developed automated BP device was compared with standard BP measurements at rest, during exercise, and recovery. The automated BP monitor accurately measured both systolic and diastolic BP at rest and during exercise. Both correlational analyses and Bland-Altman plots indicated that the BP obtained with the automated BP device were strongly associated with those obtained with the reference methods. The Fukuda automated BP device may be a suitable alternative to manual BP measurement especially when the proper social distance is needed.
Previous studies have assessed the accuracy of automated BP devices during exercise [2, 4, 11,12,13,14]. These studies produced highly mixed findings. In one of the earlier studies that evaluated four different BP monitors during cycle exercise at 100 W [14], the two BP devices (Colin 630 and Accutracker II BP monitors) were found to accurately measure BP at 100 W with minimal arm movement. However, it was stated that these devices would be unlikely to produce accurate and reliable measurement at higher intensities [14]. The other two devices (Del Mar P-IV and SpaceLabs 90202 BP devices) could not accurately measure BP until the participant had completely stopped movement [14]. Colin later developed a new automated BP monitor (Colin 680) specifically for the purpose of exercise stress testing which was evaluated during high-intensity cycle and treadmill exercise [11]. The device produced significantly higher systolic BP during cycle ergometer and treadmill exercise compared with the reference (stethoscope) measurements. In contrast, in a follow-up study conducted by a different group [2], the monitor accurately measured systolic BP during low, moderate, and high-intensity cycle exercise but did not accurately measure diastolic BP at moderate and high-intensity exercise. The Colin BP monitor is no longer manufactured or sold in the market. The Tango M2 stress test monitor was also specifically developed for exercise stress testing [15]. Compared with manual BP measurements during incremental cycle exercise, the device produced reliable BP values with absolute BP differences within an acceptable clinical range [15]. Similar findings have been reported in individuals with diabetes [16] and other populations [17,18,19,20,21]. The present study found the newly-developed Fukuda automated BP monitor accurately measured both systolic and diastolic BP during exercise indicating that this is a suitable device which can be used during exercise stress testing.
It should be noted that the precise ISO guidelines were not followed strictly due to the design of our full experiment. We evaluated the Fukuda automated BP monitor as well as a new wristwatch that measures BP during incremental exercise. It was not feasible to measure brachial BP in both arms simultaneously as the arm with the wristwatch was required to be in a flexed position in order for the wristwatch to be positioned at heart level. The wristwatch was not able to accurately measure BP during exercise due to excessive movement artifact and was therefore removed from the paper. In addition, the use of the “same arm sequential” method of measuring BP as described in the ISO guidelines was not feasible during an incremental exercise protocol as BP tends to change even at the same exercise stage as exercise intensity progressed. Although ISO standard guidelines were constructed for resting BP and were not meant to be applied to exercise BP, the automated BP monitor has passed the validation criteria for non-invasive sphygmomanometers intended for use during exercise stress testing.
Overall, the automated brachial BP monitor accurately measured both systolic and diastolic BP at rest and during incremental cycle ergometer exercise. The accuracy and simplicity of this BP monitor make it a suitable option for use in laboratory and clinical settings and can aid proper social distancing in situations such as the COVID-19 pandemic.
Summary
What is known about the topic
-
Exercise stress testing is conducted to predict the risk of developing hypertension as well as diagnose and evaluate the treatment of cardiovascular disease. Manual blood pressure measurements during exercise are difficult to perform as motion and sound artifact from exercising participants often impede the ability of a technician to record accurate blood pressure.
-
Blood pressure (BP) is a vital measurement during exercise stress testing as it is an indicator to terminate exercise.
-
The onset of the COVID-19 pandemic has led international organizations to recommend social distancing to reduce the spread of the coronavirus.
What this study adds
-
The validation of a new automated blood pressure device which was specifically developed for exercise stress testing and can be utilized in environments where social distancing is recommended.
Data availability
The datasets generated and analyzed for the present study are available from the corresponding author upon reasonable request.
References
American College of Sports Medicine, Riebe D, Ehrman JK, Liguori G, Magal M. ACSM’s Guidelines for Exercise Testing and Prescription. 10th ed. Philadelphia: Wolters Kluwer; 2018.
Bond V, Bassett DR, Howley ET, Lewis J, Walker AJ, Swan PD, et al. Evaluation of the Colin STBP-680 at rest and during exercise: an automated blood pressure monitor using R-wave gating. Br J Sports Med. 1993;27:107–9.
Lightfoot JT, Tuller B, Williams DF. Ambient noise interferes with auscultatory blood pressure measurement during exercise. Med Sci Sports Exerc. 1996;28:502–8.
Gould BA, Hornung RS, Altman DG, Cashman PM, Raftery EB. Indirect measurement of blood pressure during exercise testing can be misleading. Heart 1985;53:611–5.
Mora S, Redberg RF, Cui Y, Whiteman MK, Flaws JA, Sharrett AR, et al. Ability of Exercise Testing to Predict Cardiovascular and All-Cause Death in Asymptomatic WomenA 20-Year Follow-up of the Lipid Research Clinics Prevalence Study. JAMA 2003;290:1600–7.
Sharman JE, LaGerche A. Exercise blood pressure: clinical relevance and correct measurement. J Hum Hypertens. 2015;29:351–8.
Morris SN, Phillips JF, Jordan JW, McHenry PL. Incidence and significance of decreases in systolic blood pressure during graded treadmill exercise testing. Am J Cardiol. 1978;41:221–6.
Tanaka H, Bassettjr D, Turner M. Exaggerated Blood Pressure Response to Maximal Exercise in Endurance-Trained Individuals*. Am J Hypertens. 1996;9:1099–103.
Filipovský J, Ducimetière P, Safar ME. Prognostic significance of exercise blood pressure and heart rate in middle-aged men. Hypertension 1992;20:333–9.
Tanaka H, Monahan KD, Seals DR. Age-predicted maximal heart rate revisited. J Am Coll Cardiol. 2001;37:153–6.
Lightfoot JT, Tankersley C, Rowe SA, Freed AN, Fortney SM. Automated blood pressure measurements during exercise. Med Sci Sports Exerc. 1989;21:i.
Modesti PA, Carrabba N, Gensini GF, Bonechi F, Taddei T, Malfanti PL. Automated Blood Pressure Determination During Exercise Test. Clinical Evaluation of a New Automated Device. Angiology 1992;43:980–7.
Shinohara T, Tsuchida N, Seki K, Otani T, Yamane T, Ishihara Y, et al. Can blood pressure be measured during exercise with an automated sphygmomanometer based on an oscillometric method? J Phys Ther Sci. 2017;29:1006–9.
White WB, Lund-Johansen P, Omvik P. Assessment of four ambulatory blood pressure monitors and measurements by clinicians versus intraarterial blood pressure at rest and during exercise. Am J Cardiol. 1990;65:60–6.
Cameron JD, Stevenson I, Reed E, McGrath BP, Dart AM, Kingwell BA. Accuracy of automated auscultatory blood pressure measurement during supine exercise and treadmill stress electrocardiogram-testing. Blood Press Monit. 2004;9:269–75.
Moore MN, Picone DS, Callisaya ML, Srikanth V, Sharman JE, Schultz MG. Comparison of manual and automated auscultatory blood pressure during graded exercise among people with type 2 diabetes. J Clin Hypertens. 2019;21:1872–8.
Bassett DR, Duey WJ, Walker AJ, Torok DJ, Howley ET, Tanaka H. Exaggerated blood pressure response to exercise: importance of resting blood pressure. Clin Physiol. 1998;18:457–62.
O’Donnell CJ, Ridker PM, Glynn RJ, Berger K, Ajani U, Manson JE, et al. Hypertension and Borderline Isolated Systolic Hypertension Increase Risks of Cardiovascular Disease and Mortality in Male Physicians. Circulation 1997;95:1132–7.
Pickering TG, Davidson K, Gerin W, Schwartz JE. Masked Hypertension. Hypertension 2002;40:795–6.
Kuwabara M, Harada K, Hishiki Y, Kario K. Validation of two watch-type wearable blood pressure monitors according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-6410T-ZM and HEM-6410T-ZL. J Clin Hypertens. 2019;21:853–8.
Kario K, Shimbo D, Tomitani N, Kanegae H, Schwartz JE, Williams B. The first study comparing a wearable watch-type blood pressure monitor with a conventional ambulatory blood pressure monitor on in-office and out-of-office settings. J Clin Hypertens. 2020;22:135–41.
Acknowledgements
The authors would like to thank the volunteers that donated their time for the success of this study.
Funding
This study was supported by The University of Texas at Austin.
Author information
Authors and Affiliations
Contributions
NM, and HT designed the experiments, analyzed, and interpreted the data. NM, TW, and LC performed the data collection. NM wrote the initial draft of the paper and prepared figures. All authors edited, revised, and approved the final version of the paper.
Corresponding author
Ethics declarations
Competing interests
Fukuda Denshi provided the automated BP device evaluated in the present study.
Ethics approval
This study was approved by Institutional Review Board at The University of Texas at Austin.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
McLaurin, N.N., Wang, T., Chen, LS. et al. A new and specific automated blood pressure device for exercise stress testing. J Hum Hypertens 37, 150–154 (2023). https://doi.org/10.1038/s41371-022-00784-9