Main

The study of respiratory mechanics and control of breathing ideally requires an accurate and noninvasive measurement of lung volume changes (ΔVL) not interfering with patients' spontaneous activity. Unfortunately, the currently available methods for the measurement of ΔVL in infants present several intrinsic and unresolved limitations.

Most of the techniques require either connections to the airway opening or critical patient- and posture-dependent calibration procedures. Connecting to the airway opening may introduce leaks, annoy the infant, and affect the breathing pattern (1,2), while the use of patient-dependent calibration procedures can lead to inaccurate measurements when a patient's condition or posture changes with respect to the baseline state (3). Moreover, none of the existing techniques have proved to be reliable in tracking long-term changes in end-expiratory lung volume (EELV) due to integration drift or problems in the measurements stability (4,5).

Optoelectronic plethysmography (OEP) is a noninvasive technique, which estimates chest wall volume (VCW) by measuring the three-dimensional position of several reflective markers placed on the patient's thorax (68). This technique has proved to be reliable in adults in different postures and conditions (7,8).

OEP does not require a connection to the airway opening or subject-specific calibration procedures. It provides an accurate measurement of the VCW and it allows, through the use of subsets of the markers placed on the thorax, the measurement of separate compartments, such as the rib cage, abdomen, or smaller components (9). Being based on the direct measurement of chest wall volumes, OEP is not affected by integration drift and can be used to track breath-by-breath changes of EELV (10).

The aim of this study was to develop an appropriate experimental set-up and a geometrical model to apply OEP to term and preterm newborns and to validate the system by comparison with pneumotachography.

METHODS

All measurements were performed in a tertiary level NICU. The study was approved by the institutional ethical committee of the San Gerardo Hospital (Monza, Italy), and informed consent was obtained from parents.

Study population.

Any infant in a stable condition at the time of the measurement and with no lung dysfunction was eligible for the study.

Experimental protocol.

Infants were studied during quiet breathing laying in supine position on an infant warmer (Air-Shields Vicker, Hatboro, PA). Airflow was measured at the mouth using a facemask connected to a pneumotachograph (PNT). To avoid leaks, the mask was held firmly in place by an operator during data recording for 2 min at maximum.

Measurements.

Detailed descriptions of the OEP for the analysis of the VCW in standing, seated, and supine adults have been published previously (8,11,12).

To apply OEP in supine newborns, 24 hemispheric reflective markers (6 mm diameter) were positioned on the anterior thoracicabdominal surface from the clavicles to the anterior superior iliac spines at the intersection between predefined vertical and horizontal lines (Fig. 1).

Figure 1
figure 1

Markers positioning and geometrical model for OEP measurement in newborns. Left panel: Markers' positioning and chest wall compartments for optical analysis in newborns in supine position. Transverse lines: clavicular line; manubriosternal joint, lower costal margin, and anterior superior iliac crest. Craniocaudal lines: midsternal line (extended to the abdomen), the midpoints of the interval between the midsternal and the midaxillary lines; and two midaxillary lines. Right panel: Three-dimensional model of the chest wall from one representative model. Black thick lines: triangles connecting real markers (solid circles); gray lines: triangles connecting to at least one “virtual” marker (open circles). The two compartments are drawn separately by introducing an offset on the craniocaudal direction for clarity.

Six horizontal lines were designated for placing the markers: 1) the clavicular line; 2) the nipples; 3) the xiphoid process; 4) the lower costal margin; 5) umbilicus; 6) superior iliac spine; as well as five vertical lines: 1) the midline (along the sternum and continuing caudally below the xiphoid through the umbilicus); 2–3) two parasternal lines; 4–5) two midaxillary lines. To better outline the costal margin, two markers were added between the parasternal and the midaxillary lines. Similarly, to increase marker density in the abdomen, one more marker was added between the xiphoid process and the umbilicus.

The markers were placed on the skin using bi-adhesive hypoallergenic tape. The three-dimensional position of each marker was measured by an automatic optoelectronic motion analyzer (OEP System, BTS, Milan, Italy) at a frequency rate of 60 Hz using six specially designed infrared video cameras provided with infrared flashing light-emitting diodes. Cameras were positioned so that each marker was simultaneously seen by at least two cameras (Fig. 2) to reconstruct their three-dimensional position and displacement during respiration by stereo-photogrammetric methods. Three cameras were placed on one lateral side about 1 m from the bed. Three additional ones were placed symmetrically on the other side of the bed. The four cameras on the corners were placed about 70 cm above the subject while the two in the middle were placed 1.1 m above the subject; all of them were slanted downward. The operative calibrated volume was large enough to include the whole trunk of the infant lying on the bed.

Figure 2
figure 2

Experimental setup for the application of optoelectronic plethysmography in newborns. See methods section for details.

Total VCW was finally determined by approximating chest wall surface by 34 triangles connecting the markers and computing the volume enclosed by all these triangles. The posterior surface of the chest wall (hidden in supine position) was defined by nine “virtual” markers assumed to lie on the supporting plane. The supporting plane was assumed to be parallel to the plane of the bed and 5 mm below the most posterior lateral marker (lateral markers being: 7, 8, 12, 15, 20, and 24, see Fig. 1). The positions of the virtual markers were determined by projecting onto the reference plane the initial frontal coordinates (from the first image frame) of the following points: 1) the average between marker 4 and 7; 2) the average between marker 6 and 8; 3) marker 11; 4) marker 14; 5) the average between marker 20 and 21; 6) marker 21; 7) marker 22; 8) marker 23; and 9) the average between marker 23 and 24.

To compute VCW and its division in compartments, such as rib cage (VRC) and abdomen (VAB), all the considered triangles must define a closed surface surrounding the compartment. This condition is satisfied only if the position of the markers does not differ significantly from the indications of Figure 1. If a marker were placed incorrectly, the previous mathematical procedure (8) would provide erroneous results with no warnings to the user. Here, we developed a new algorithm, which is capable of automatically identifying improperly positioned markers (see supplementary material http://links.lww.com/PDR/A54). When markers are properly positioned, the two algorithms provide exactly the same volumes.

Measurements of lung volume changes.

The subjects breathed through a facemask connected to a heated mesh-type (PNT) (Hans-Rudolph 3700A) coupled with a differential pressure transducer (Pascaline PXLA02X5DN) for the measurement of airway opening flow (·VAO). The flow signal was sampled at 960 Hz. Corrections for humidity, pressure, and temperature changes of the gas between inspiration and expiration were not applied. Lung volume changes were obtained by numerical integration of the flow signal. The resulting volume drift was corrected by estimating the linear trend of the volume trace over time by linear regression analysis and by subtracting it from the integrated signal, providing a drift-corrected ΔVL trace.

Data analysis.

For each infant data were recorded during 1 to 2 min of quiet breathing. All the recorded breaths not corrupted by clearly identifiable artifacts (such as disconnection of the mask, presence of leaks, or major patient movements) were manually selected. For each breath, end-expiratory and end-inspiratory points were identified in the ΔVL trace. These points were used to compute the tidal volume of each breath from both ΔVL and VCW tracings (VT,L and VT,CW, respectively).

The Bland-Altman analysis, which is a plot of the difference between data measured by two methods against their mean, was performed on raw data to test the agreement of the two ways of measuring ΔVL (13). The same data were also used to perform a linear regression analysis between the techniques.

OEP is based on the approximation of the chest wall shape and dimensions by a group of triangles. If this approximation were not accurate, the accuracy of the measurements could be affected and biased by the chest wall shape, size, or motion pattern. To evaluate this issue, body weight and GA were also added as independent variables in the regression analysis.

The percentage difference (%diff) between OEP and PNT was computed for each single breath as follows:

Equation

The percentage differences were then averaged for each patient.

There are no available gold standards for the measurement of compartmental volumes in newborns. Even if the absolute compartmental accuracy cannot be evaluated, it is still possible to detect differences in the accuracy of the rib cage or the abdominal compartment resulting from inappropriate distribution of markers on the chest wall by studying the relationship between %diff and the percentage contribution of each compartment. If the rib cage and the abdominal volumes were measured with different accuracies, the most inaccurate would negatively affect the overall accuracy depending on how much it contributes to total VT,CW, introducing a trend in this relationship.

To quantify thoracoabdominal asynchronies, the following parameters were calculated from the total and compartmental VCW tracings: rib cage and abdominal contribution to VT,CW (RC% and AB%) and phase angle between abdominal and thoracic motions (θ). The phase angle was calculated, as previously suggested (14), through the analysis of Lissajous figures obtained by plotting rib cage versus abdominal volume tracings (Konno-Mead plots). When rib cage and abdomen are synchronous, a narrow or closed loop is produced, which widens with increasing asynchrony. The width of the loop is a measure of asynchrony and can be quantified by the phase angle (θ), which is θ = sin−1 (m/s) for θ < 90°, and θ = 180° − sin−1 (m/s) for θ > 90°, where m is the width of Lissajous figure at mid-rib cage excursion, and s is the width of Lissajous figure at its largest abdominal excursion. Phase angle can range from 0° (synchronous) to 180° (paradoxical breathing). Whether outward motion of the abdomen preceded or lagged the motion of the rib cage was determined by considering the direction of Lissajous figures (anticlockwise or clockwise). Negative angles were used for anticlockwise direction, indicating an outward motion of the abdomen preceding the rib cage.

The overall accuracy of OEP in measuring tidal volume was evaluated by Bland-Altman and linear regression analysis pooling all breaths from all patients. Multiple linear regression analysis between VT,L and VT,CW on pooled data including body weight and GA as independent variable was performed to examine the extent to which they alter the regression coefficient. Data from each patient, expressed as mean ± SD, were also reported.

RESULTS

A total of 20 newborns with body weight ranging between 1300 and 3790 g at the time of the study were recruited. Relevant data on the patient population are reported in Table 1.

Table 1 Patients' characteristics

A representative experimental tracing with total and compartmental chest wall volumes, VAO, and ΔVL is shown in Figure 3. Rib cage volume fluctuates around 350 mL, abdominal volume around 510 mL, and VCW around 860 mL. These values represent the whole volume of the rib cage, abdomen, and chest wall, and they are not related to lung volume. However, changes in VCW can be associated to lung volume changes and compared with data obtained by PNT. Altogether, a total of 771 breaths were selected, providing 1542 volume changes. The Bland-Altman analysis (Fig. 4) showed a good agreement between VT,L and VT,CW. The mean difference was −0.08 mL indicating that, when considering the whole patients population, there is no significant bias in the measurement. The limit of agreement ranged from −2.98 to 2.83 mL, and we did not find any relationship between the difference and the mean of the measurements, suggesting that OEP does not introduce any systematic error related to the magnitude of volume changes. The correlation analysis on pooled data (Fig. 4) showed a good linear relationship (m = 0.96, r2 = 0.95, p < 0.001) between VT,L and VT,CW.

Figure 3
figure 3

Experimental tracings. VRC, rib cage volume changes; VAB, abdomen volume changes; VCW, total VCW changes; Flow, airflow measured at the inlet of the full-face mask.

Figure 4
figure 4

Comparison between pneumotachography and OEP. Left panel: Bland-Altman analysis; the solid horizontal line is the mean and the dashed lines are mean ± 2 SD. Right panel: linear regression between VT,L (tidal volume measured by integration of the flow signal) and VT,CW (chest wall tidal volume measured by OEP); the slope of the regression line is 0.96, the intercept is 1 mL, and the coefficient of determination is r2 = 0.95.

The inclusion of body weight and GA as independent variables in the regression analysis did not affect the coefficient of determination, indicating that these two parameters do not introduce bias into the extent of agreement between VT,L and VT,CW.

Table 2 reports VT,L, VT,CW, and the percentage difference between the two for each infant. The average tidal volumes ranged between 12.8 and 31.0 mL, allowing the evaluation of OEP on a wide range of tidal volumes, whereas the average percentage differences ranged between −10.1 and 6.1%.

Table 2 Comparison between pneumotachography and OEP

Compartmental VCW changes.

The comparison between %diff and AB% showed the absence of a trend between the two, suggesting that the geometrical models used to describe the rib cage and of the abdomen are equally accurate.

Table 3 shows VT,CW measured by OEP and its subdivision into rib cage and abdominal compartments (VT,RC and VT,AB, respectively), expressed both in absolute values and as a percentage of total VCW variations. The abdomen contributed 95.3 ± 13.7% and the rib cage 4.7 ± 13.7% to VT,CW. These values were compared with compartmental VCW changes reported in the literature for healthy adults in supine position (7). As expected, the main contribution to tidal volume is provided by the abdomen and this contribution is notably higher in infants than in adults.

Table 3 Total and compartmental chest wall volume changes

An example of the Konno-Mead plots obtained in three different representative subjects is shown in Figure 5. In five infants, the rib cage contribution was negative, showing the presence of paradoxical breathing. The phase angle was on average −61.1 ± 45.9, and it was negative for all subjects except one (Table 3), suggesting that the outward motion of the abdomen on average tends to precede ribcage one.

Figure 5
figure 5

Konno-Mead loops. Konno-Mead loops for three representative subjects presenting synchronous (A), asynchronous (B), and paradoxical (C) chest wall motion.

DISCUSSION

The accurate measurement of ΔVL is particularly difficult in newborns because they cannot cooperate, and connecting to the airway opening without leaks is extremely difficult. Moreover, the measurement technique has to be noninvasive and should not interfere with the breathing pattern by increasing dead space or breathing load.

Spirometry and PNT are currently used for the measurement of ΔVL, but they require a connection to the airway opening that may introduce leaks, annoy the infant, and affect the breathing pattern (1,2). Moreover, they cannot measure breath-by-breath changes in absolute lung volumes because of the unavoidable presence of drift in the volume trace, which hinders the assessment of changes in EELV over long time-scales.

Infant whole body plethysmography allows the measurement of tidal volumes and EELV, but it requires the use of a cumbersome rigid box, which blocks the access to the infant. Furthermore, similarly to pneumotachography, it requires a connection to the airway opening and is affected by long-term volume drift.

Respiratory inductive plethysmography (RIP) is a simple and noninvasive respiratory monitoring technique that quantifies breathing pattern, including thoracoabdominal asynchrony, through the changes in the cross-sectional area of the rib cage and the abdominal compartments, without requiring a connection to the airway opening (1). Unfortunately, the calibration procedure is critical and patient-dependent. In addition, measurements are not reliable if the position and condition of the patient differ from those assumed during calibration (3). Moreover, the use of RIP for tracking long-term changes in EELV is still questioned (4,5). Finally, RIP is based on the assumption that the chest wall has only two degrees of freedom (rib cage and abdominal cross-sectional areas). However, the validity of this assumption is questionable, especially with high degrees of distortion and an extremely compliant chest wall, as frequently occurs in neonates (1517).

OEP has potentially the advantage of overcoming these limitations but, to be applied on newborns, it requires special measurement protocols (i.e., number of markers and their positioning on the chest wall and geometrical models), which have to be developed purposely for this application. In this study, a new experimental protocol designed to apply OEP in newborns was introduced. The accuracy provided by the protocol was evaluated by comparison with pneumotachography during short periods of spontaneous breathing.

Comparison between volume measurements by PNT and OEP.

While PNT measures lung gas volume changes, OEP measures the variations of the volume enclosed within the chest wall. For this reason, there are several possible sources of discrepancy between the two measurements.

The measurement of VCW changes provides an estimation of ΔVL in body temperature and pressure, saturated with water vapor (BTPS) conditions. The PNT measures the volume of gas in BTPS during expiration. Conversely, during inspiration, the gas is in ambient temperature and pressure, saturated with water vapor (ATPS) conditions. These differences would require the application of a correction factor of about 8% (1.091–1.069 for 22–26°C ambient temperature in the NICU) to the inspired volume measured by pneumotachography. Because the expected difference is small (or even smaller as the PNT was heated) and is impossible to estimate the real conditions of the gas passing through the PNT mesh (18), it was decided not to correct the data for this source of discrepancy.

Similar considerations can be referred to the different composition of the inhaled and exhaled gas, as the absorption of oxygen and the release of carbon dioxide in the alveolar gas affect the viscosity, leading to possible differences in the pressure drop across the PNT mesh.

Another source of discrepancy between PNT and OEP is represented by possible blood shifts between the trunk and the extremities during breathing, which might produce small but measurable differences between ΔVL and changes in VCW, particularly during deep breaths or positive pressure ventilation (6). As in this study, only spontaneous breathing was tested, and it was assumed that changes of gas volume in the lung were equalled to changes in VCW.

Finally, the average %diff value close to zero suggests that, despite the potential sources of discrepancy described earlier, in our experimental conditions, OEP provides an unbiased, reliable, and accurate estimation of ΔVL without requiring any correction factor.

Division of VCW changes in rib cage and abdominal compartments.

Several studies focused on the clinical interest and usefulness of the measurement of thoracoabdominal asynchronies in the diagnosis of sleep disordered breathing or as an indicator of airway obstruction during REM sleep (19,20). Furthermore, it was shown that chest wall motion is strictly correlated with lung mechanics (21) in spontaneously breathing infants with BPD.

OEP enables the measurement of volume variations of different compartments, which, in this study, were limited to the rib cage and abdomen. The accuracy of each single compartment was tested by comparing the total accuracy of OEP with the relative contribution of the abdominal compartment, taking advantage of the wide distribution of this parameter in newborns (it ranges from 71.5 to 129% in the examined patients). In case of a systematically greater error in the volume estimation of one compartment compared with the other, the total accuracy would be biased by the percentage contribution of that compartment. The absence of any correlation between these variables indicates that the volume estimation of the two compartments is equally accurate.

Table 3 showed high intersubject variability in the degree of asynchrony between abdominal and rib cage compartments, with paradoxical breathing in five subjects, as expected in healthy infants during spontaneous breathing. Previous studies suggested that the time lag between the movement of rib cage and abdomen differs between term and preterm infants (16), is correlated with changes in the mechanical properties of the lung and chest wall (2123) and that the percentage rib cage contribution to tidal volume can be used as a measure of rib cage maturation (24). The capability of OEP to provide anatomically based compartmental volumes could help in advancing the research in these applications.

Critiques of the technique.

Even if suitable for research, at the actual stage of development, the use of OEP for clinical application in NICU environment is affected by some technical issues, such as the need to attach a high number of markers on the skin of the infant and the need to keep the infant in supine position with the thorax naked. Moreover, movements of the back with respect to the bed may introduce artifacts because the model adopted to close the chest wall surface uses the projection of the initial position of some markers onto the bed plane. Finally, the set-up is cumbersome, the system cannot easily be moved from a bed to another, and the cameras we used cannot see through the incubator.

While we think that the benefits of using OEP fully overcome these limits, future technical improvements are needed to resolve or, at least, reduce the impact of most of these limitations before it is suitable for routine applications at the bedside of clinically ill newborn infants. For instance, markers could be attached to a shirt or tissue strips to simplify their positioning and removal. Recent advances in TV camera technology could also lead to a miniaturization, so that the system could be placed inside the incubator and moved from an incubator to another.

In conclusion, even if further technological developments are welcome, this study demonstrated that OEP could already be used for the accurate assessment of ΔVL and its compartmental distribution in term and preterm newborns. Thanks to its specific advantages, it has the potential to be used in many research and clinical applications to further improve the understanding of physiology and pathophysiology of respiration in newborns and at the early stage of development.