Introduction

Currently, late preterm births (infants born at 34 to 36 weeks of gestation) account for ~70% of all preterm births. These infants, which are still physiologically and developmentally immature, have higher risks of morbidity and mortality compared with infants born at term.1, 2 Because fetal lung maturation occurs late in gestation, respiratory morbidities constitute a major concern for late preterm infants. Anomalies in the process of fetal lung fluid reabsorption and surfactant deficiency are the main factors involved in respiratory disorders of late preterm infants.3, 4, 5

Some studies have recently investigated the effectiveness of early sustained lung inflation (SLI) in preterm infants to prevent mechanical ventilation. This strategy, consisting of the administration of a positive inflation pressure for a period of time that is significantly more prolonged than the normal inspiratory time (15 to 25 s), is considered an alternative to clear fluid-filled lungs and to achieve an early functional residual capacity (FRC).6, 7, 8, 9, 10, 11, 12

The current evidence is not sufficient to support the efficacy and safety of SLI in preterm infants; furthermore, no reports of its use in late preterm infants are available. The aim of our study was to evaluate the effectiveness of SLI in reducing the need for respiratory support in late preterm infants.

Methods

This randomized controlled trial was conducted in a tertiary-level neonatal intensive care unit (NICU) in Milan between September 2013 and June 2014. We enrolled all consecutive inborn infants with a gestational age of 34 to 36 weeks after obtaining parental consent. Infants with major congenital anomalies, or those for whom it was not possible to obtain parental consent, were excluded. The study was approved by the Institutional Review Boards of the Fondazione IRCCS Ca’ Granda, Ospedale Maggiore Policlinico, Milan, Italy. Eligible infants were randomly allocated to one of the two following treatment groups using sealed opaque envelopes:

  • Group 1 (SLI group) – Infants assigned to this group were placed at birth under a radiant heat source and, after oropharyngeal and nasal suctioning, a prophylactic pressure-controlled (25 cmH2O) inflation was sustained for 15 s using a neonatal mask and a T-piece ventilator, followed by the delivery of 5 cmH2O CPAP. In the following 6 to10 s, CPAP was discontinued in the absence of signs of inadequate respiratory effort (that is, apnea or gasping) or heart rate >100 beats per min (b.p.m.). In the presence of signs of inadequate respiratory effort and/or whenever the heart rate was between 60 and 100 b.p.m. despite CPAP, a SLI maneuver with the same parameters was repeated. If the heart rate was <100 b.p.m. after the second SLI maneuver, the infant was resuscitated according to the recommendations of the American Academy of Pediatrics (AAP).

  • Group 2 (Control group) – Infants assigned to this group were assisted according to the recommendations of the AAP, providing the initial steps in stabilization (warming and maintaining normal temperature, positioning, clearing secretions only if copious and/or obstructing the airway, drying, and stimulating). We progressed beyond the initial steps in the presence of inadequate respiratory effort (defined as above) and/or a HR <100 b.p.m. according to the AAP recommendations.

In both groups, respiratory support in the delivery room was given using a T-piece ventilator (Neopuff Infant T-Piece Resuscitator, Fisher & Paykel, Auckland, New Zealand). To avoid pressure leakage, we used a properly sized neonatal mask that adequately covered both the mouth and nostrils of the infants (size 0 or 1 face mask, Ambu UltraSeal, Ballerup, Denmark). The flow rate was set at 8 to 10 lmin−1. Infants in both groups who developed respiratory distress were admitted to the NICU; during an observation period of 30 min, the respiratory rate, pulse oximeter oxygen saturation (SpO2), and the severity of respiratory distress measured using the Silverman score were monitored. The need for oxygen supplementation was assessed to maintain an SpO2 between 88 and 95%. At the end of the observation period, if the infant needed a fraction of inspired oxygen (FiO2)0.30 to maintain an SpO2 between 88 and 95% and/or the Silverman score was5, a non-invasive respiratory support (nasal CPAP of 5 to 6 cmH2O, Infant Flow SiPAP System – Viasys Healthcare, Yorba Linda, California, USA; High Flow Therapy, flow 4 to 6 LPM, Vapotherm, Stevensville, Maryland, USA) was initiated. Infants requiring FiO2>0.40 to maintain an SpO2 of 88 to 93% and/or with a Silverman score>6 were treated with surfactant (200 mg/kg; Curosurf, Chiesi Farmaceutici SpA, Parma, Italy) with the INtubation-SURfactant administration-EXtubation procedure. Infants were intubated when the pH was <7.20 with partial pressure of carbon dioxide (PaCO2)>70 mmHg or if FiO2>0.40 after surfactant treatment was needed to maintain an SpO2 of 88 to 93%, or if infants had frequent episodes of apnea (>4 episodes in 1h or >2 episodes requiring bag-and-mask ventilation) despite adequate non-invasive support. Infants were extubated when they met all of the following criteria: FiO2<0.40, mean airway pressure <6 cmH2O, PaCO2<65 mmHg with a pH>7.20, and PaO2>50 mmHg.

The primary outcome of the study was to evaluate the effectiveness of SLI in reducing the need for respiratory support in late preterm infants compared with control infants. Secondary outcomes were NICU admission, NICU admission for respiratory disease, length of stay, and exclusive breastfeeding at discharge.

Perinatal and neonatal data were collected for each infant as follows: gestational age, birth weight, sex, mode of delivery (vaginal or cesarean section), Apgar at 1st and 5th minute, antenatal steroid treatment, main diseases of pregnancy, NICU admission, Silverman score, diagnosis of respiratory disease (respiratory distress syndrome, transient tachypnea of the newborn, pneumonia, and persistent pulmonary hypertension), type of respiratory support, total length of respiratory support, surfactant administration, air leak syndromes, exclusive breastfeeding at discharge, length of stay, and mortality. The diagnosis of respiratory distress syndrome was based on the occurrence of typical signs (need for oxygen supplementation, tachypnea, intercostal retractions, and grunting), and findings on chest radiographs (decreased lung expansion, reticulogranular pattern of the lung, and air bronchograms), and the exclusion of other causes of respiratory failure. The diagnosis of transient tachypnea of the newborn was based on chest radiographs indicative of retained fluid (for example, streakiness, fluid in fissure), an oxygen requirement >40%, and spontaneous improvement over two to five days. Pneumonia was defined by an abnormal chest radiograph in the presence of two of the following: extra-pulmonary clinical signs of sepsis, abnormal non-specific (screening) laboratory tests, positive blood culture, positive tracheal culture, or perinatal history indicating exposure (prolonged rupture, maternal fever, so on). Diagnosis of persistent pulmonary hypertension was based on an echocardiogram.

The study was not blinded. The decision to start respiratory support was made by clinicians other than the investigators involved in the study according to specific guidelines, and researchers assessing the study end points were blinded to the nature of the study treatments. The staff involved in the study is a skilled team of specialists in neonatal resuscitation and the SLI maneuver is a well-established procedure at our institution and has been used in previous clinical trials. Comprehensive education and training were undertaken to ensure technical proficiency and protocol compliance through audits with clinical personnel.

Statistical analysis

From 2009 to 2012, 15% of 2503 late preterm infants born at our center needed respiratory assistance (invasive or non-invasive ventilation). In calculating the sample size, we followed the confidence interval approach.13, 14 We calculated a sample size of 92x2 newborns assuming a reduction in the need for respiratory assistance in this population from 15 to 5%, that is, a risk ratio of 0.33 with a 95% confidence interval from 0.13 to 0.88. We compared the characteristics between the SLI and the control group using Student’s t-test or the Mann–Whitney test for quantitative variables and the χ2-test for categorical variables. We calculated the risk ratios and 95% confidence intervals of primary and secondary outcomes for the SLI group in reference to the control group. All analyses were performed with version 9.2 SAS software (SAS Institute, Cary, NC, USA, 1999).

Results

In the period between September 2013 and June 2014, we enrolled a total of 185 late preterm infants. The number of infants deemed eligible for the study and the number of whom were randomly assigned to receive the SLI procedure or standard assistance in the delivery room are shown in Figure 1. The baseline characteristics of infants at birth were equally distributed between the two groups (Table 1). More than 90% of infants were born by cesarean section in both groups; an emergent cesarean section, in the absence of labor, was performed in 21% of cases. Twins accounted for over 50% of the study population.

Figure 1
figure 1

Diagram showing the flow of participants through the trial.

Table 1 Characteristics of infants in SLI and Control group

Table 2 shows the primary and secondary outcomes of the study. There was no difference in the need for any type of respiratory support between the infants treated with SLI at birth and the control group (10.6 vs 8.7%, P=0.64). Six (60%) out of the 10 infants in the SLI group who required respiratory support had a diagnosis of respiratory distress syndrome and four (40%) had a diagnosis of transient tachypnea of the newborn. Eight infants in the control group required respiratory support, two (25%) with a diagnosis of respiratory distress syndrome, and six (75%) with a diagnosis of transient tachypnea of the newborn. Similarly, NICU admission, NICU admission for respiratory distress, length of stay, and exclusive breastfeeding at discharge did not differ between the two groups. All of the 185 infants were followed until discharge, and none died during the in-hospital stay.

Table 2 Primary and secondary outcomes

Three infants (3.2%) developed a pneumothorax in the SLI group compared with no infants in the control group (P=0.08). The clinical characteristics of infants who developed pneumothorax are detailed in Table 3. None of these infants required vigorous resuscitation at any time. In two cases, a chest tube was placed for tension pneumothorax.

Table 3 Clinical characteristics of infants who developed pneumothorax

Discussion

This study evaluated the effectiveness of SLI at birth in reducing the need for respiratory support in late preterm infants compared with controls. Our results demonstrated that a prophylactic SLI at birth does not affect the need for respiratory support in late preterm infants.

Respiratory assistance in the delivery room is an aspect of respiratory care in preterm infants that has been thoroughly investigated in recent years. The critical event is the aeration of the fluid-filled lung; at birth, a pressure gradient is required to overcome airflow resistance and move lung fluid distally through the airway to the terminal airspaces, where the opening pressure is necessary to overcome the surface tension and open the alveoli. While term infants begin to establish the FRC with their first breath after birth, preterm infants are hampered because of poor respiratory drive, limited muscle strength, high chest wall compliance, and surfactant system immaturity.15 The SLI has been suggested as an additional approach to promote lung fluid clearance and aeration, as it mimics the first breaths of life in healthy term infants that are characterized by a prolonged inspiratory time due to a long time constant of the fluid-filled lung, a prolonged expiratory phase and high inflation pressure. The effect of SLI as an initial ‘prolonged’ inflation is to move the air/liquid interface distally, with direct lung aeration and clearance of lung fluid from the major airways immediately after birth. This process is completed by interstitial lymphatic and venous reabsorption in the following hours.16, 17

The effect of SLI to facilitate lung aeration and improve postnatal ventilation has been investigated in animal and human studies.6, 7, 8, 9, 10, 11, 12, 18, 19, 20, 21, 22 In preterm infants, the evidence to support its efficacy and safety remains limited, and no reports of its use in late preterms are available. We chose to study this population of preterm infants as they account for an increasing proportion of respiratory-associated short-term morbidities with a need for specialized care and prolonged hospital stays.5, 23, 24

Colin et al.25 suggested that the functional consequences of late preterm birth include difficulty in protecting FRC, vulnerability to airway collapse, and increased airway resistance. Moreover, rapid clearance of the fetal lung fluid plays a key role in the transition to air breathing; after late preterm birth, this process can be delayed because this is dependent upon developmentally regulated epithelial sodium channels, especially in the absence of spontaneous labor and related hormonal changes, as frequently occurs in elective cesarean sections.26 Because the delayed absorption of lung fluid is an important cause of respiratory failure in late preterms, we speculated that SLI could reduce the need for respiratory care in these infants, although no previous experimental or human studies on the effectiveness of SLI in this population are available.

We estimated a sample size by considering a reduction in the need for respiratory support from 15 to 5% and a number needed to treat of 10, which is clinically relevant. We think that a less significant reduction would not justify a prophylactic SLI whose risks and benefits are not yet completely known.

We applied the sustained inflations using a face mask. As recently demonstrated by van Vonderen et al.,27 the effect of SLI on FRC gain may be impaired by substantial leakiness and a limited volume may consequently be delivered. In studies that reported the efficacy of SLI in creating inspiratory volume and FRC, the sustained inflation was applied via an endotracheal tube, bypassing the glottis and minimizing leaks.7, 22 However, van Vonderen et al.27 observed that, even in infants in whom no leak was present, the delivered volumes achieved by the sustained inflation were very low. It is possible that closure of the vocal cords in both breathing (that is, between breaths) and apneic infants could cause obstruction; furthermore, obstruction occurring at the pharyngeal level by a relatively large tongue or obstruction of the nose and mouth could occur with improper mask technique. In our study, lung volumes were not recorded, but it is reasonable to believe that the problems described by van Vonderen et al.27 may have influenced our results. We believe that it would be helpful in further studies to evaluate lung mechanics during the SLI and to assess the optimal parameters of the maneuver (time and pressure), actual lung recruitment and its variations over time.

There are some limitations to our study. We performed SLI in all late preterm infants; therefore, we also performed SLI in infants with appropriate lung fluid reabsorption and adequate FRC. Although this approach was used in the largest multicenter randomized controlled trial to date,21 in late preterm infants with adequate respiratory effort and partially aerated lungs, SLI at birth could make them more susceptible to over-distention and air leaks. In our study, pneumothorax occurred in 3.2% (n=3) of infants in the SLI group compared with no infants in the control group (P=0.08); all pneumothorax events occurred beyond the first 12 h of life. This result should be taken into consideration, although the difference was not statistically significant and the number of cases is limited. In the recent study by Lista et al.,21 most of the pneumothoraces occurred beyond the first 48 h of life, rather than soon after the SLI maneuver. A possible reason for this finding as suggested by Keszler28 could be that the late pneumothoraces could be related to the reluctance to intubate and provide surfactant in the intervention group of this unblinded study in which the need for mechanical ventilation was the primary outcome, rather than the intervention itself. No other studies have shown an increase in air leaks in the intervention group treated with SLI.29

Another limitation of the study is that it is a non-blinded study and a single site trial. We tried to minimize this bias through adequate training of care providers, by using clinicians who had the decision to start respiratory support other than the investigators involved in the study, and by blinding the researchers that assessed the study end points to the nature of the study treatments.

Conclusion

Delivery room sustained inflation is an attractive concept in the respiratory assistance of preterm infants at birth. To the best of our knowledge, no reports of its use in late preterm infants are available. Our results demonstrated that prophylactic SLI at birth does not reduce the need for respiratory support in these preterm infants. Possible future trials should exclude infants with adequate respiratory effort and focus on late preterm infants who require resuscitation intervention at birth due to inadequate respiratory effort and thus likely to benefit from recruitment maneuver. Moreover, trials should include close monitoring of adverse events, especially with regard to the risk of overdistension and air leak syndromes. Associated measures of respiratory mechanics could be useful to evaluate the efficacy and individualize the SLI maneuver at birth.