Abstract
Background/Objectives:
Air-displacement plethysmography (ADP) body composition systems utilize a precise pre-test protocol that must be followed for each trial. Previous research has shown that body temperature changes influence the results when using ADP as a measurement. The objective of this study is to determine the effect of post-exercise body temperature changes on body composition results using ADP.
Subjects/Methods:
Forty young adults (18–30 years) participated in the study (23 females, 17 males). Resting heart rate was measured to calculate exercise intensity. First, an ADP test was done according to the instructions outlined by the manufacturer. Upon completion of the ADP test, the subject exercised on a treadmill at moderate intensity (∼65% heart rate reserve) for 30 min to increase body temperature. Another ADP measurement followed the physical activity. Chamber temperature (CT) was measured during each of the trials using a thermistor to assess changes. Fat mass (FM), fat-free mass (FFM), percentage body fat (%BF) and CT were compared pre- and post exercise.
Results:
Paired sample t-tests revealed significant differences (P<0.05) between pre-exercise %BF and post-exercise %BF (21.3±9.8% vs 19.6±10.2%), pre-exercise CT and post-exercise CT (22.7±1.2 vs 23.0±1.1 °C), and pre-exercise FM and post-exercise FM (14.9±7.9 vs 13.8±8.0 kg). FFM showed no significant difference.
Conclusion:
This investigation demonstrates the importance of following the manufacturer’s recommended pre-test protocol as the accuracy of the ADP testing may be compromised, resulting in lower FM and %BF estimations.
This is a preview of subscription content, access via your institution
Access options
Subscribe to this journal
Receive 12 print issues and online access
$259.00 per year
only $21.58 per issue
Buy this article
- Purchase on Springer Link
- Instant access to full article PDF
Prices may be subject to local taxes which are calculated during checkout
Similar content being viewed by others
References
BOD POD body composition tracking system operator's manual 2004, Life Measurement Inc., Concord.
Thomas S, Reading J, Shephard R . Revision of the physical activity questionnaire (PARQ). Can J Sport Sci 1992; 17: 338–345.
Karvonen M, Kentala K, Musta O . The effects of training on heart rate: a longitudinal study. Ann Med Exp Bio Fenn 1957; 35: 307–315.
Tegenkamp MH, Clark RR, Schoeller DA, Landry GL . Effects of covert subject actions on percent body fat by air-displacement plethysmography. J Strength Cond Res 2011; 25: 2010–2017.
Fields DA, Higgins PB, Hunter GR . Assessment of body composition by air-displacement plethysmography: influence of body temperature and moisture. Dyn Med 2004; 3: 1–7.
Brooks GA, Fahey TD, Baldwin KM (eds) Exercise physiology: human bioenergetics and its applications. 4th edn., The McGraw-Hill Companies Inc: New York, NY, USA, 2004, pp 511–532.
Drummond MJ, Parcell AC . Air displacement plethysmography under dehydrated and hyperhydrated conditions. Med Sci Sports Exerc 2004; 36: S72.
Vukovich MD, Peeters BM . Reliability of air-displacement plethysmography in detecting body composition changes after water ingestion and after creatine supplementation. J Exerc Phys 2003; 6: 115–122.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Competing interests
The authors declare no conflict of interest.
Rights and permissions
About this article
Cite this article
Otterstetter, R., Johnson, K., Kiger, D. et al. The effect of acute moderate-intensity exercise on the accuracy of air-displacement plethysmography in young adults. Eur J Clin Nutr 67, 1092–1094 (2013). https://doi.org/10.1038/ejcn.2013.133
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/ejcn.2013.133