Introduction

Gait impairments are prevalent among individuals with neurological disorders1, affecting more than 60% of these patients2. These impairments often show common gait abnormalities, such as slower gait speed, shorter stride length, and poor body balance3. Given that gait function is integral to daily life activities4, disturbances in gait often represent the initial stages towards loss of mobility and independence2,5. As such valid, reliable, and sensitive assessments of walking function are crucial in quantifying gait impairments in individuals with neurological disorders, thereby enabling these impairments to be therapeutically addressed.

A variety of clinical tests exist to assess walking function. They range from assessments of walking ability during different activities to timed measures determining how many meters one can walk in a given time or how long it takes to walk a given distance. Each of these measures has specific advantages and limitations. However, gait speed (derived from timed measures) is probably the most established clinical outcome to monitor walking function, as it can be considered as a key feature of locomotor control and is easy to assess in clinical practice6. The main drawback of timed measures is their inability to assess specific individual deficits or gait characteristics underlying a gait impairment.

Locomotion is controlled at multiple levels of the central nervous system (CNS), i.e. cortical, subcortical, brainstem and spinal networks7,8. Various domains, such as pace, rhythm, variability, asymmetry, postural control, etc.9,10,11 can partly describe these physiological systems. Assessing these different domains of locomotor function by gait analysis allows the construction of individual gait profiles that are unique to each person, similar to “fingerprints”, particularly after spinal cord injury (SCI)9. For gait analysis, optical motion capture (OMC) systems are considered the gold standard12. However, these systems come with the disadvantages of high cost, the need for expert operation, time-consuming setup and the fact that it is restricted to a laboratory setting13, which does not make them practical for everyday clinical use.

In recent years, inertial measurement units (IMU) have been used as a possible alternative to objectively assess gait14 and could be a valuable extension of standard functional gait tests. Compared to OMC, IMUs are compact, inexpensive, easy to operate and can be combined with functional walking tests, making them convenient for clinical applications or gait assessments even in the home environment13. Even a sparse IMU setup of only two IMUs attached to the lower extremities can provide reliable spatiotemporal gait parameters in individuals without neurological movement disorders15. In individuals with mild walking impairments, sensor setups have been developed that show high validity and good test–retest reliability in several neurological diseases16,17,18. However, a previous study has shown that the detection of gait events by such IMU setups is unreliable in individuals with slow walking speeds (< 1.2 m/s) and short stride lengths (< 1.0 m)19, often observed in individuals with severe gait impairments. A new sensor setup and algorithm for IMU-based gait analysis (ZurichMOVE, https://zurichmove.com) has recently been validated in individuals with severe SCI, walking at slow speeds (mean walking speed of 0.76 ± 0.17 m/s) in a laboratory environment20. The IMU modules consist of a tri-axial accelerometer, a tri-axial gyroscope, and a tri-axial magnetometer that record at a sampling rate of 200 Hz. Due to indoor magnetic field distortions, the magnetometer data was excluded from the data analysis20. The setup involves two sensors attached laterally just above each ankle, and the developed algorithms provide valid spatiotemporal parameters even for individuals with severe walking impairments walking at very low speeds. Using such IMU setups in combination with timed measures would allow us to understand individuals’ gait characteristics and interpret mechanisms that lead to changes in gait performance over consecutive assessments. Nevertheless, prior to combining an IMU-setup and the implemented algorithms with a timed measure like the 2mWT, their reliability must be established.

This study aimed to investigate whether the ZurichMOVE sensor setup and the newly developed algorithms can reliably assess gait characteristics of individuals with SCI performing a 2mWT. Firstly, test–retest reliability of the sensor setup was evaluated in two measurements performed within a week. Subsequently, we describe whether the setup is sufficiently sensitive to discern various gait profiles among individuals with comparable walking performance.

Methods

Participants

Individuals with SCI were recruited at the Spinal Cord Injury Center of the Balgrist University Hospital. The inclusion criteria were acute/subacute (1–6 months) or chronic (> 6 months) SCI, age ≥ 18 years and able to walk at least 10 m with/without braces, assistive devices at a minimum walking speed of 0.17 m/s.

The exclusion criteria were current orthopaedic issues, major psychosis, depression, a history of severe heart conditions, and other neurological diseases.

Procedure

The study has been approved by the Ethical Committee of the Canton of Zurich (BASEC 2020-01473) and was conducted in accordance with Good Clinical Practice (GCP) guidelines and the Declaration of Helsinki. Prior to enrolment, written informed consent has been obtained from all participants. Individuals were invited to participate on 2 test days, separated by 1 to 7 days. On the first day, a familiarisation run of the 2-min walk test (2mWT) was performed. After a break of at least 30 min, another 2mWT was performed. On the second test day, the 2mWT was repeated. In addition, the lower extremity motor score (LEMS) was assessed on the first day.

Sensor setup

Spatiotemporal gait parameters were assessed by two IMU sensors (ZurichMOVE, https://zurichmove.com) (measuring 35 × 35 × 12 mm and weighing 18 g)21. More details on the IMU sensor modules are described in a previous study21. They were attached to the ankles just above the lateral malleolus with an elastic Velcro strap (Fig. 1).

Figure 1
figure 1

Positions of the sensors attached laterally above the ankles.

Assessment of gait profiles

Gait profiles were created consisting of sensor-derived parameters that could be divided into spatiotemporal measures (gait cycle parameters expressed as mean of multiple strides) and dynamic features (representing the inconsistency of spatiotemporal measures across the strides20. Spatiotemporal gait parameters except smoothness were calculated using a previously developed algorithm20 and they were extracted for each cycle, namely: stride velocity (m/s), stride time (s), step time (s), swing time (s), relative swing time (%), stance time (s), relative stance time (%), double support (s), relative double support (%), stride length (cm), horizontal foot displacement in frontal plane (cm), vertical foot displacement in sagittal plane (cm) and smoothness (see Supplementary Table 1 for the detailed description of each parameter). The algorithm identified turning steps as steps that deviated more than 30° from the main movement direction. One step before the turn, the steps taken during the turn, and one step immediately after the turn were excluded from the analysis. Smoothness was calculated based on analytical models from Balasubramanian et al.22. Dynamic features of gait were deduced by calculating symmetry and variability parameters from all sensor-derived spatiotemporal parameters. Symmetry (symmetry index: \(SI= \frac{L-R}{max}\times 100\))23 and variability ((coefficient of variance: \(COV= \frac{\sigma }{\mu }\)), σ = standard deviation; μ = mean)) were calculated. Gait parameters were grouped in functional domains according to the literature9,10,24,25. This resulted in five gait domains with the corresponding parameters defined in Table 1. To generate a visual representation of the gait profiles, we normalized the data by converting them into z-scores, using the data of the first day as a basis for this transformation. During this process, for each subject, we calculated the mean values across all their gait cycles, then used these z-scores to create a radar graph, providing a clear visual depiction of the data.

Table 1 Allocation of gait parameters into gait domains.

2-min walk test

The 2mWT was performed according to the Guidelines of the American Thoracic Society26, except that the hallway length was 35 m (instead of 30 m). The decision to alter the length was prompted by the narrowness of the corridor. By extending it to a length of 35 m, more space was available for the turns. For safety reasons, individuals were accompanied by an examiner walking behind each individual to allow them to set the pace.

Braces and/or habitual assistive devices were permitted but must be kept similar across the assessments. The individuals were instructed to walk as far as possible but safely within two minutes. They were allowed to take rest breaks if needed, but time continued running during the break.

The participants were asked to wear closed, comfortable shoes. Shoes with high heels (> 3 cm) were excluded. Shoes were identical for both assessments.

Statistics

Sample size calculation

Sample size calculation for the assessment of test–retest reliability was conducted using the formula proposed by Walter et al.19. The formula was implemented in an online calculator (https://wnarifin.github.io/ssc/ssicc.html). The sensor-derived parameters were validated with an infrared marker-based motion capture system20 which is considered the gold standard in modern gait analysis due to their high level of precision27 Based on the high accuracy of the sensor-derived parameters as demonstrated in this validation study, we anticipated that the true intraclass correlation coefficients (ICC) would be as well excellent with a value of approximately 0.95, and a minimal acceptable ICC set at 0.85. The significance level was chosen as 0.05 at a power of 0.8. Repetitions were set to 2. Based on these considerations, the minimal required sample size for the study was determined to be n = 25.

Test–retest reliability

ICCs (two-way mixed effect, total agreement) were calculated to determine test–retest reliability of all parameters between the first and second day28. ICC values have been interpreted according to the recommendation of Koo et al.29: < 0.5: “poor”, 0.5–0.74: “moderate”, 0.75–0.9 “good”, and > 0.9 “excellent”. Bland–Altman plots have been created to estimate the absolute agreement between the two measurements.

The standard error of measurement ((\(SEM=SD\times \sqrt{(1-ICC)}\)), SD = standard deviation)30 and the Minimal Detectable Change (\(MDC=SEM\times 1.96\times \sqrt{2}\))31 were calculated based on the respective formulas.

Consistency of gait profiles

To assess the consistency of the gait profiles between days 1 and 2, the sum of the absolute z-score values was calculated and then compared using a t-Test.

Ethical approval

The Ethics Committee of the Canton of Zurich (BASEC 2020-01473) approved this study. All participants gave written informed consent before data collection began.

Results

Participant characteristics

Twenty-eight individuals with moderate to severe gait impairments have been recruited for this study and assessed. Data of two individuals had to be excluded from the analysis due to technical problems with the sensors. Data of 26 individuals (age: 57.1 ± 14.6) with SCI (AIS C: 7; AIS D: 19) were analysed in this study. Further demographic characteristics of the study participants are depicted in Table 2.

Table 2 Detailed demographic characteristics of the individuals with spinal cord injury who participated in the study.

Test–retest reliability

In total, 8608 strides of 26 individuals with SCI were included in the analysis. The mean distance in the 2mWT did not differ between the test days (p = 0.086) and was 105 m ± 54 m (min–max: 20–234 m) on the first day and 109 m ± 54 m (min–max: 25–241 m) on the second day. On average, the testing days were spaced 3.6 ± 1.65 days apart. The number of strides did not differ between the tests (p = 0.322), with an average of 165 strides (min–max: 79–252) in the first and 166 strides (min–max: 79–258) in the second 2mWT. The gait parameters showed good to excellent ICCs (0.84–0.99 for spatiotemporal measures and 0.77–0.99 for dynamic measures). More details about the averaged gait parameters, ICCs, SEM and MDC can be found in Table 3. No significant difference in any parameters was found between days 1 and 2. The Bland–Altman plots of the parameters showed a reasonable agreement between the days (Fig. 2).

Table 3 Test–retest reliability and minimal detectable change of spatiotemporal parameters.
Figure 2
figure 2

Bland–Altman plots to assess test–retest reliability. The averages of the two measurements were plotted against the differences. The solid lines represent the means and the dashed lines the 95% limits of agreement.

The Bland Altman plots of the dynamic parameters can be found in Supplementary Fig. 1.

Consistency of gait profiles

The comparison of the absolute summed z-scores between day 1 (11.07 ± 6.13) and day 2 (10.37 ± 5.1) revealed no significant difference (p = 0.1228). Figure 3A shows the gait profile of 4 individuals who covered almost the same distance in the 2mWT. All four individuals show a similar profile on both testing days, illustrating the within subject consistency of the gait profiles. However, gait profiles also illustrate that gait characteristics may differ between individuals with comparable walking distance in the 2mWT (Fig. 3B). This is in accordance with the substantial left–right asymmetry in the LEMS (left: 10, right: 25) found for subject 02, whereas subject 13 achieved the maximal score of 25 with both legs.

Figure 3
figure 3

Radar chart illustrating the 15 gait characteristics organised by domains. (A) Shows the consistency of the gait profile from day 1 (D1) to day 2 (D2) for four individuals with a similar performance in the 2-min walk test. (B) Shows different gait strategies used by individuals, albeit same walking distances. Two show similar gait profiles (right subplot) and two show profoundly different gait profiles (left subplot).

The z-scores for all individuals, calculated for day 2 are presented in Supplementary Table 2.

Discussion

Gait impairments are frequently seen in neurological diseases. Therefore, it is essential to objectively assess walking impairments in order to offer appropriate physical therapy interventions, measure the success of treatments, and monitor disease progression.

The present study aimed to assess the reliability of IMU-based gait parameters in individuals with SCI. The main findings of the present study are that in individuals with SCI, (1) the presented setup showed good to excellent test–retest reliability for all extracted gait parameters, (2) individuals with SCI show consistent gait profiles between two measurements within 1 week, and (3) the sensor setup is suitable to detect patient-specific gait characteristics and distinguish patient-specific gait impairments.

Test–retest reliability of spatiotemporal parameters

It has been demonstrated previously that spatiotemporal gait parameters assessed with IMUs show high validity and good test–retest reliability in individuals with Parkinson’s disease16, multiple sclerosis17, and stroke18 with mild gait impairments. Typically, the algorithms used to calculate spatiotemporal parameters rely on detecting gait events such as heel strike or foot off. Different methods are used to detect these events, and based on their complexity, they show different robustness against individual differences and disturbed gait32. Typically, the robustness of gait event detection algorithms decreases with poorer walking ability of an individual, as they usually rely on fixed thresholds, which makes them prone to fail in poor walkers20. The algorithm used in this study uses adaptive thresholds to detect each individual’s gait cycles and gait events20. This approach resulted in good to excellent reliability (ICC: 0.77–0.99) of spatiotemporal and dynamic parameters, even in individuals with severe gait impairments. ICCs were higher than 0.75 for all parameters, corresponding to good to excellent reliability29. It is known that the walking protocols used to collect the gait data have an influence on its reliability6. Walking speed and cadence for instance can be reliably assessed over 10 m in SCI33, therefore requiring only a few steps to be accurate. However, a higher number of steps is needed to assess dynamic parameters reliably. In elderly and those with Parkinson’s disease, continuous walking protocols with at least 15 strides are recommended to assess reliable dynamic parameters34,35. We used a continuous walking protocol (2mWT) that resulted in assessments of at least 79 strides per individual, allowing for a detailed and reliable analysis of different aspects of walking function. To further increase the reliability of spatiotemporal and especially dynamic parameters, the IMU-based assessment could also be done during a 6-min walk test, which is also considered to be reliable in SCI36. However, longer testing protocols and the related exhaustion should be avoided in the case of poor walkers.

Consistency of gait profiles

Recovery of gait is of high relevance for individuals with incomplete SCI37. Restoration of gait consists of improvements in functional (e.g. speed, endurance) as well as qualitative (e.g. kinematic, biomechanical) aspects38. However, typical clinical measures such as timed measures only assess functional gait recovery and do not inform about movement quality. This is a limitation for clinical trials in neurological disorders, as outcome measures focusing purely on functional gait recovery can miss important information on gait recovery or disease progression. For example, it has been shown that variability parameters for step time and stance time are able to distinguish between healthy controls and individuals with MS in the absence of clinical gait impairments as determined by the EDSS39. Also, for fall prediction, spatiotemporal and dynamic parameters are more sensitive than functional outcomes like walking speed40. This highlights the potential of sensor-derived gait profiles as a sensitive tool to characterise and monitor gait parameters in individuals with gait impairments. We could demonstrate that gait profiles assessed with IMUs were reliable in SCI, which is a prerequisite for their clinical implementation. In addition, the derived gait profiles were able to reveal different gait patterns in individuals with comparable walking performance. Studies in stroke survivors have identified several altered kinematic gait patterns including decreased knee flexion and decreased dorsiflexion during the swing phase compared to healthy controls. In order to achieve sufficient foot clearance despite these kinematic alterations, patients often develop compensation strategies (e.g., pelvic hiking and circumduction)41. Our data shows that individuals with SCI with a similar 2mWT performance can also utilize different gait strategies. For example, Fig. 3B illustrates that subject 02 uses more compensatory movements (reflected in a more asymmetrical gait pattern) than subject 13. This can be explained by the profound left–right lower limb muscle strength asymmetry (assessed by the LEMS) of subject 02, while subject 13 has identical (maximal) muscle strength in both legs. Thus, IMU-based gait analysis allows us to identify patient-specific deficits that cannot be determined by a simple timed walking test. This is of high relevance for longitudinal assessments in SCI. Measurements of gait over time after injury can inform us about primary adaptations in motor behaviour as response to injury and, over time, demonstrate mechanisms of improvement (i.e., compensation vs regeneration).

It has been shown in individuals with a neurological disease that the walking speed differs when measured in clinical vs community environments42. These results align with a study performed with individuals with Parkinson’s disease, where a significant difference has been found for all gait parameters between laboratory and free-living environments10. Therefore, it is crucial to move the assessment of gait out of the laboratory and into more real-life environments such as at home and in the community. IMUs allow to capture gait over extended periods and can help to provide a more comprehensive picture of individuals’ gait deficits6. However, the automatic recognition of walking bouts during everyday life remains challenging, especially for patients with a distinct pathological gait pattern. Further, the threshold for determining the minimum number of steps required to classify an activity as walking remains unknown9. IMUs could be used in large, multicentre, cohort, and longitudinal studies, as well as in trials investigating different treatments, where they can facilitate the collection and improve the quality of gait data.

Limitations

A limitation of this study is the absence of reference data from healthy individuals, which hinders the ability to compare and quantify the pathological nature of gait patterns in individuals with SCI. Given that the primary objective of this study was to assess reliability and sensitivity of gait parameters in detecting inter-individual differences, reference data collection from healthy individuals was not pursued. However, acquiring such data will be the next crucial step towards enabling the quantification of the extent of pathology in the gait patterns of individuals with spinal cord injuries.

It's worth noting that the z-scores were computed to visualize the spatiotemporal parameters among individuals with SCI and to highlight the potential of sensor-derived parameters in identifying distinct walking strategies. However, the calculated z-scores should not be used to compare different spatiotemporal parameters against each other, as normal distribution was not given in all scores.

Conclusion

This study demonstrated good to excellent test–retest reliability for all gait parameters of an IMU-based gait analysis in individuals with SCI. Based on these findings, IMU-based gait analysis is clinically feasible (simple handling, time effective) to complement gait assessments while providing detailed and objective measures of gait characteristics and changes over time.