Introduction

Dental implant associated artifacts in CT scans pose a common challenge in the assessment of the craniomandibular region. As lifestyle and dental medicine changed over the last decades, dental implants and fillings are frequently used and therefore are found in most CT examinations, especially in older patients. In a study among emergency patients from 2019 dental artifacts were present in 75% of CT scans1. These artifacts can severely hamper diagnostic assessment of CT scans and disguise tumor metastasis or overlap tumor delineation. Inflammation foci and traumatic injuries of the dental apparatus might be missed. As CT is the standard imaging modality for assessment of head and neck tumors as well as facial/cranial injuries and inflammatory diseases there is a strong need for dental implant associated artifact reduction to allow for diagnostic assessment, especially of the oral cavity2.

Most dental implant associated artifacts result from beam hardening, photon starvation and scattering3 and lead to reduced diagnostic quality of the adjacent soft tissue. The extent of these are influenced by various parameters such as material of the implants, tube voltage and current as well as reconstruction parameters4. Several studies employing dual energy CT systems based on energy integrating detectors have shown that the reconstruction of virtual monoenergetic images (VMI) can help to decrease the extent of implant associated artifacts2,5,6,7,8.

The new emerging technology of photon-counting detector CT (PCD-CT) has raised high expectations9. PCD-CT utilizes a direct conversion of X-ray photons into electronic signal, whereas energy-integrating detector CT (EID-CT) converts X-ray photons in a preceding step into visible light before the visible light is converted into an electronic signal through photo diodes. In recent studies, the inert advantages of photon-counting detector systems (i.e. lower image noise, higher SNR, higher CNR as well as higher spatial resolution with the potential for simultaneous dose reduction) have been demonstrated10,11,12,13,14,15,16,17. Photon-counting CT technology enables lower image noise, spectral information and exact photon energy arrangement, especially of low energy-photons, ultimately leading to the reconstruction of less artifact-burdened images. A phantom study showed a significant reduction in metal artifacts for orthopedic prosthesis compared to EID-CT18. Therefore, PCD-CT is expected to reduce artifacts compared to EID-CT17.

The aim of this study was to assess the clinical use of VMI for dental artifact reduction in PCD-CT as compared to a conventionally, reconstructed polychromatic image (PI).

Materials and methods

The local institutional review board of the University Hospital Bonn (Ethics Committee of the Medical Faculty of the University of Bonn, Application number 159/22) approved this study. Informed consent was waived by the Ethics Committee of the Medical Faculty of the University of Bonn. All scans were acquired during clinical routine and no scan was performed solely for research purposes. The study was conducted in accordance with the Declaration of Helsinki and its amendments.

Patients with multiple dental implants (≥ 5) and associated artefacts receiving a head/neck scan between 12/2021 and 02/2022 were included in this retrospective, single-center study. Implants were located in maxilla and/or mandible. Scans were performed on a clinical PCD-CT (NAEOTOM Alpha, Siemens Healthcare GmbH). Inclusion criteria included age ≥ 18 years, five or more dental implants and the acquisition of a spectral post-processing (SPP) enabled image data set with standard protocol as described below.

Overall, 30 patients were included in the analysis (8 female, 22 male) with an average age of 69 (range: 31–98) years (Fig. 1). Mean DLP was 292.63 mGy*cm and mean CTDIvol was 9.36 mGy.

Figure 1
figure 1

Patient flow chart with exclusion criteria.

Imaging Protocol

A weight-adapted volume of iodine based contrast agent (Accupaque 300 mg/ml, GE Healthcare Buchler GmbH & Co. KG) was applied intravenous with a flow rate of 3 ml/s followed by a bolus of 40 ml of physiologic saline solution. Post-threshold delay was 60 s. Scans were performed in a head-first supine position.

Scan parameters were a tube voltage of 120 kVp with activated automatic tube current modulation, a pitch of 0.8 and a gantry rotation time of 0.5 s. Collimation was 144 × 0.4 mm. Reconstruction parameters were 1.5 mm slice thickness of the reconstruction with an increment of 1 mm. A regular body kernel (Br40; Siemens Healthcare GmbH) as well as Quantum Iterative Reconstruction (QIR Level 3; Siemens Healthcare GmbH, Erlangen, Germany) was used for image reconstruction.

VMI were reconstructed in axial view for 100–190 keV in an interval of 10 keV using dedicated software (syngo.via VB 60, Monoenergetic Plus; Siemens Healthcare GmbH). VMI below 100 keV were not investigated based on adverse results of previous studies2.

Quantitative image analysis

Quantitative assessment of polychromatic and virtual monochromatic images was performed using region of interest (ROI) based attenuation analysis in the most pronounced hypodense and hyperdense artifacts using a conventional clinical DICOM viewer (Deep Unity R20 XX; Dedalus HealthCare GmbH). Furthermore, values and standard deviation of X-ray attenuation in fat tissue and soft tissue areas with and without presence of artifacts were evaluated as shown in Fig. 2. Hereby, differences in attenuation and gradient of attenuation of tissue were measured and corrected attenuation and image noise assessed as proposed before2. The employed methodology has been established previously and is based on single slices only. The small size of the ROI enables to include only one tissue/anatomical region. Readers were explicitly requested to decide on the slice with the most pronounced hypo- and hyperdense artifacts to avoid bias in image preselection. Therefore ROIs are not necessarily in the same axial level. Similarly, the selected slices do not necessarily show all of the dental implants of a patient, as these can appear very unevenly distributed in mandible and maxilla and readers were free to decide on the chosen slices.

Figure 2
figure 2

Placement of ROIs (window width/window level 250/50 HU) in the most pronounced hypodense (white) and hyperdense (green) artifacts, in fat tissue (red) and soft tissue areas with (black) and without (brown) presence of artifacts.

Corrected attenuation was assessed to differ artifact reduction from regular changes in HU for differing VMI energy levels. It was calculated as the difference of muscle tissue impaired by artifacts and without artifact impairment. As image noise is generally higher in images with presence of artifacts, we calculated a corrected image noise as the difference of noise of soft tissue (both muscle and fat) in areas with and without artifacts as proposed before2,7. Corrected attenuation of 0 indicates optimal artifact reduction, values above 0 indicate insufficient artifact reduction and values below 0 an overcorrection. All measurements were performed for polychromatic and virtual monochromatic images from 100 to 190 keV in steps of 10 keV.

Qualitative Image Analysis

Two radiologists with two (YCL) and eleven (DK) years of experience in head and neck CT evaluated the CT images independently regarding artifact extent of hyperdense and hypodense artifacts as well as assessment of soft palate and buccal tissue using a five-point Likert grading scale. The rating of artifacts was defined as follows: (1) excessive artifacts; (2) pronounced artifacts; (3) moderate artifacts; (4) minor artifacts; and (5) artifacts are absent. For assessment of soft palate and buccal tissue the following Likert scale was used: (1) highly restricted diagnostic interpretability; (2) restricted diagnostic interpretability; (3) moderate diagnostic interpretability; (4) minor restrictions on diagnostic interpretability; and (5) unrestricted diagnostic interpretability. For the analysis, the polychromatic image as well as VMI with 100 keV, 130 keV, 160 keV and 190 keV were rated as there are hardly subjective differences in shorter intervals.

Statistical analysis

All statistical analyses were conducted using IBM SPSS Version 27 (IBM Corp.). Graphs were carried out using the software GraphPad PRISM Version 6.02 (GraphPad Software). Quantitative results are stated as mean and standard deviation. Wilcoxon signed-rank test was used for statistical analysis of quantitative image parameters. Qualitative results are expressed as median with interquartile range (IQR). Interrater reliability was assessed using the intraclass correlation coefficient (ICC). ICC estimates and their 95% confidence intervals (CI) were calculated based on a mean-rating (k = 2), consistency, two-way mixed-effects model19. p-values below 0.05 were considered significant.

Results

Quantitative image analysis

VMI130 keV showed the best corrected attenuation compared to PI with the closest value to 0 (PI: 30.48 ± 168.64; VMI130 keV: − 0.54 ± 110.31; p = 0.0026). Corrected attenuation was negative for higher keV VMI ≥ 130 keV, indicating an overcorrection of the artifacts. Mean attenuation of hyperattenuating artifacts decreased in VMI compared to polychromatic images (PI: 474.5 ± 75.9; VMI130 keV: 292.5 ± 54; p < 0.0001). This effect aggravated with VMI of higher keV (Table 1; Fig. 3). Precise ROI values for CI and each VMI are stated in Table 1.

Table 1 Mean attenuation values and standard deviation within defined regions of interest for different reconstructions.
Figure 3
figure 3

Mean attenuation values within defined regions of interest for soft tissue with presence of artifacts and most pronounced hypodense and hyperdense artifacts. p-values below 0.05 are marked with an asterisk.

Qualitative image analysis

Qualitative assessment of soft palate (SP) and buccal tissue (BT) improved compared to conventional image (Median PI: SP 1 (1–2); BT 1 (1–2); Median VMI130 keV: SP 2 (2–4), BT 2 (1–3); p < 0.0001; Supplementary Information 1). Hyperdense artifacts were subjectively reduced, scoring a 2.5 (2–3; p < 0.0001) at 130 keV. For the PI the score was 1 (1–1) and VMI190 keV scored 3 (2–3; p < 0.0001). For hypodense artifacts, PI scored 1 (1–2) and VMI130 keV scored 2(1–2; p < 0.0001). No remarkable difference was observed between VMI160 keV and VMI190 keV (Fig. 4). In some cases hypodense artifacts increased with higher keV VMI (Fig. 5.). Rating results for each criteria by each rater are shown in Fig. 6. Regarding the most pronounced artifacts, only small improvements could be achieved, regardless of the keV, as shown in Fig. 7.

Figure 4
figure 4

Example of an axial view (window width/window level 250/50 HU) of a conventional polyenergetic reconstruction (PI) compared to virtual monoenergetic images (VMI) with 100 keV, 130 keV, 160 keV and 190 keV.

Figure 5
figure 5

Example of an axial view (window width/window level 250/50 HU) of a conventional polyenergetic reconstruction (PI) and virtual monoenergetic images (VMI) with 100 keV, 120 keV and 140 keV. With higher keV increasing hypodense streaks occur.

Figure 6
figure 6

Bar plots show distribution of hypodense artifact extent, hyperdense artifact extent, soft palate diagnostic quality and buccal diagnostic quality rating of polyenergetic reconstruction (PI) and virtual monoenergetic images (VMI) with 100 keV, 130 keV, 160 keV and 190 keV.

Figure 7
figure 7

Example of an axial view (window width/window level 250/50 HU) of a conventional polyenergetic reconstruction (PI) compared to virtual monoenergetic images (VMI) with 100 keV, 130 keV, 160 keV and 190 keV. Regarding these most excessive artifacts only little visible improvements occur, regardless of keV.

Interrater agreement was excellent with an overall ICC value of 0.973 (95% confidence interval: 0.968; 0.977). Respective ICC was 0.933 (95% CI 0.907–0.951) for the extent of hypodense artifacts, 0.983 (95% CI 0.976–0.987) for the extent of hyperdense artifacts, 0.978 (95% CI 0.969–0.984) for assessment of the buccal tissue and 0.980 (95% CI 0.972–0.985) for assessment of the soft palate.

Discussion

This study evaluates the impact of virtual monoenergetic image reconstruction on artifacts from dental implants in photon counting detector CT. The current results indicate that VMI are a useful tool to reduce dental implant artifacts and, thus enable a better assessment of the oral cavity. VMI associated image quality improvement may enable detection of pathologies that would otherwise be concealed by artifacts in conventional CT images.

There are different causes of implant-associated artifacts. Beam hardening, based on the preferential signal-loss of low-energy photons, causes streaking hypodense artifacts20. Photon starvation, which leads to hypodense artifacts, is observed when too few photons can pass a dense structure21. Scatter artifacts, mostly due to incoherent scattering, contribute to a higher X-ray attenuation22,23. High energetic VMIs mainly address the artifacts caused by beam hardening. As low-energy photons are absorbed more rapidly by an object they pass, the mean detected photon energy increases, causing artifacts during image reconstruction24. With the use of high keV VMI, these artifacts can be reduced. High keV images, however, typically show a decrease in contrast. CNR in PCD-CT is higher than in EID-CT15,25,26,27, thus PCD-CT technology allows for higher keV reconstructions in diagnostic assessment.

The effect of VMI associated artifact reduction reached a plateau at VMI160 keV, no further improvement was seen at higher energetic reconstructions. A possible explanation for this phenomenon is that the beneficial effects of VMI are already maximized at VMI160 keV.

In the current study, artifacts were not completely eliminated in VMI but merely decreased. These remaining artifacts might be caused by other effects than beam hardening, for instance photon starvation. Another possible explanation is the selection of patients with at least five dental implants and therefore comparatively extensive artifacts. Supporting this thesis, previous studies with extensive artifacts observed only minor differences especially between VMI with high keV2.

The extent of both hypodense and hyperdense artifacts was reduced in reconstructed VMI which resulted in an improvement of qualitative and quantitative scores. For hyperdense artifacts, the reduction of artifact extent as well as the increase of image quality was highly significant. Although the same tendencies were seen for hypodense artifacts, results were not significant, analogue to studies previously conducted on EID-CT2,5,7.

For quantitative assessment, the method of placement of ROIs for measurement of HU to quantify the extent of artifacts was used. There are various other approaches to determine the extent of artifacts, including voxel measurements and calculation formula28,29. There is currently no consensus regarding a gold standard for the assessment of artifact magnitude in CT. In the current study, a practical and comprehensible method was employed. For qualitative assessment, two radiologists independently scored images using a Likert-scale grading system ranging from one to five, with one being the worst quality and five the best achievable quality. An analogous approach was also employed in previous similar studies2,7.

Important differences in artifact reduction seem to be caused by material composition and material orientation, as it in general determines the extent of artifacts21. The optimal keV setting will most likely depend on material that caused the artifact as well as the extent of artifacts. Although a personalized algorithm taking the specific implant material of each patient into account would be desirable, clinical implementation is not feasible. Thus, a universal solution is necessary for clinical practice. In some rare cases, we experienced new hypodense streaks that aggravate with increasing keV and thus worsen diagnostic quality of the region. This phenomenon could not be observed for most dental implants and fillings. In some cases, there was hardly a difference between subjective impressions of the artifacts between the different VMI, while quantitively difference could be shown to the PI. Overall, for most cases, the best diagnostic quality was observed at 130 keV. Given that, the recommendation of the authors would be to additionally reconstruct VMIs at 130 keV for PCD-CT examinations with extensive dental implant associated artifacts. The results of our studies are in line with previous studies for reduction of dental implant associated artifacts performed on a dual layer dual energy CT, favoring a keV setting between 130 and 160 keV2,7. Studies performed on dual source dual energy CT recommend a slightly lower keV between 100 and 130 keV, especially due to the decrease of contrast at higher keV30,31. In general, most studies favor a VMI energy level between 100 and 140 keV4,32,33,34,35,36. A first phantom study on PCD-CT stated no benefit from higher-energetic reconstructions, contrary to our results37. A recent study which assessed artifacts of cochlear implants in both PCD-CT and EID-CT, found a comparable artifact reduction for VMI of EID-CT and PCD-CT38.

There are several limitations to this study. Study setup was retrospective, single-centered and only included a small number of patients. Further studies with more patients included are necessary to test clinical benefit of VMI, especially once additional artifact reduction techniques are clinically available and can be compared.

Caused by the retrospective character of the study, no exact dental history was available. Therefore, the various composition of the included implants is unknown. However, this resembles a real life situation, where there is usually no knowledge available regarding the material of dental implants. Nevertheless, as different materials cause different artifacts, it would be of interest to what extent the artifacts of each material can be reduced using VMI39. We noticed that the extent of artifacts differed vastly depending on the type of implant. Thus, studies investigating optimal artifact reconstruction depending on implant type are needed.

There is the need for further studies on PCD-CT to address the performance of metal artifact reduction algorithms (MAR) also in combination with VMI. At the time of the end of the investigation, there was no clinical MAR for the PCD-CT available. Even though VMI showed improvements in diagnostic image quality, artifacts were merely reduced, not eliminated. There is still potential to further optimize artifact reduction for extended artifacts.

Conclusion

In summary, the current results show that VMI lower dental—implant associated artifacts of the craniomandibular region in PCD-CT images. VMI reconstructed at 130 keV showed the best results as a tradeoff between artifact reduction and image quality. Higher kev reconstructions lead to lower contrast and the appearance of extensive, new hypodense artifacts.