Abstract
Miniaturization and electrochemical performance enhancement of electrodes and microelectrode arrays in emerging long-term implantable neural stimulation devices improves specificity, functionality, and performance of these devices. However, surgical site and post-implantation infections are amongst the most devastating complications after surgical procedures and implantations. Additionally, with the increased use of antibiotics, the threat of antibiotic resistance is significant and is increasingly being recognized as a global problem. Therefore, the need for alternative strategies to eliminate post-implantation infections and reduce antibiotic use has led to the development of medical devices with antibacterial properties. In this work, we report on the development of electrochemically active antibacterial platinum-iridium electrodes targeted for use in neural stimulation and sensing applications. A two-step development process was used. Electrodes were first restructured using femtosecond laser hierarchical surface restructuring. In the second step of the process, atomic layer deposition was utilized to deposit conformal antibacterial copper oxide thin films on the hierarchical surface structure of the electrodes to impart antibacterial properties to the electrodes with minimal impact on electrochemical performance of the electrodes. Morphological, compositional, and structural properties of the electrodes were studied using multiple modalities of microscopy and spectroscopy. Antibacterial properties of the electrodes were also studied, particularly, the killing effect of the hierarchically restructured antibacterial electrodes on Escherichia coli and Staphylococcus aureus—two common types of bacteria responsible for implant infections.
Similar content being viewed by others
Introduction
Over the past several decades, millions of patients across the world have depended on implantable neurostimulation (or neuromodulation) devices for treatment of various cardiac and neurological disorders1,2,3,4,5,6,7,8,9. These devices function by inhibition, stimulation, modification, or alteration of neural tissue through targeted delivery of an artificial electrical stimulation to specific neurological sites in the body9. The electrical stimulation occurs through transfer of an external electrical signal from a neurostimulator10 or an implantable pulse generator (IPG)11 to an implanted electrode or electrode array and then causing a change in response within the neural activity11,12,13.
Electrodes and electrode arrays3,14 are fundamental to the functionality of neurostimulation devices and thus are designed and manufactured with specific sizes, geometries, profiles, material composition, surface properties, as well as electrical, electrochemical, and mechanical properties to match the biological requirements of their intended applications3,15,16,17,18,19. Electrodes or electrode arrays for sensing applications are often required to exhibit low impedance, while high charge storage and charge injection capacity and specific capacitance are desired for neurostimulation applications and for cardiac rhythm management devices, respectively12,14,20,21. These specific requirements have resulted in transformations in these devices and their capabilities. In recent years, there has also particularly been a trend towards device miniaturization since smaller implantable devices are desired for compatibility with normal human activities and to enhance patient comfort22,23.
Enhancing the electrochemical surface area (ESA) of the electrodes not only improves the electrochemical performance of the electrodes by increasing charge storage/injection capacity and specific capacitance and reducing impedance, but also enables a decrease in geometric surface area (GSA) of the electrodes to achieve miniaturization14,24,25. In a recent work12, we demonstrated that the electrochemical performance of femtosecond-laser hierarchically-restructured Pt10Ir alloy electrodes can be tuned to yield unprecedented electrochemical performance that significantly exceeds those reported in the literature.
Surgical site-infections impact 2–4% of all patients undergoing surgery, and are the primary cause of hospital readmission post-surgery, and account for ~ 20% of all hospital acquired infections26,27,28. Notably, over half of these infections can be attributed to indwelling medical devices, with specific infection rates varying based on the type of device29,30,31,32. Once the infection occurs, the bacteria are often able to form biofilm structures, a 3D superstructure containing both living bacterial cells and a secreted extracellular polysaccharide layer which provides protection to the bacterial cells imbedded within23,33,34. These biofilms are notoriously resistant to traditional small molecule antimicrobials, a problem which is heightened by the overall increase in antimicrobial resistance35. The problem of antimicrobial resistance is recognized as a major threat to human health by both the CDC and the WHO36. Taken together, the combination of these emerging factors highlights the need to develop new and alternative strategies to combat infections, such as development of medical devices with novel antimicrobial properties37,38,39.
For safe and long-term application of implantable devices, it is necessary to endow the implant with antimicrobial properties to resist microbial-induced infections40,41. However, in the case of devices used in neurostimulation applications, the electrodes and electrode arrays that are at the interface of the device and the biological environment, e.g. the widely used electrode materials such as Pt10Ir alloy and coatings such as Titanium nitride (TiN) or Iridium oxide (IrO2)12,42,43,44 do not exhibit significant antimicrobial activity40,41. Bacteria and biofilm generation on long-term implantable medical devices for neurostimulation applications, e.g., implantable cardiovascular devices, can occur frequently and cause serious infections, which can lead to death40,41. Therefore, new methodologies for preventing and eliminating bacteria and biofilm formation are critical and needed.
Various biocompatible antimicrobial materials, e.g. copper oxide (CuO), are capable of inhibiting and/or even eliminating bacteria and biofilms40,41,45,46. However, these antimicrobial materials are normally not good electrical conductors and have poor mechanical properties, therefore, they cannot be directly used as electrodes in neurostimulation applications. Creating an antimicrobial surface coating on long-term implantable electrodes can be a promising approach for establishing an antimicrobial barrier, which actively prevents and eliminates the formation of bacteria and biofilms. It should also be noted that antibacterial coatings may be applied universally regardless of whether they are deposited on implantable electrodes or leads. Therefore, this study provides pivotal insights to use the same coating method to offer antibacterial properties and infection prevention for the leads by coating them with similar antibacterial coatings.
Hierarchically restructured Pt10Ir electrodes used in neurostimulation applications consist of topographic features spanning a variety of length scales including coarse-scale rough structures (~ 1–100 µm) and a finer nanostructure subset (~ 5–100 nm) on top of the coarse structures (Fig. 1) essential for the high electrochemical performance of these electrodes12. Standard coating approaches would render a subset of the increased surface area achieved by hierarchical surface restructuring inaccessible, thus reducing the desired electrochemical performance. Therefore, to minimize the impact of the antimicrobial coatings on the performance of these electrodes, the coating must be thin and conformal. Low thickness is essential for minimal impact on the hierarchical structure, and electrochemical performance such as charge storage capacity, specific capacitance and impedance. Moreover, good conformality is essential for the complete antimicrobial coverage of the complex hierarchical structure of these electrodes. Furthermore, to release the optimal amount of antimicrobial ions to the surgical site, it is important to precisely control the thickness of the antimicrobial coating: (1) the coating should be thick enough to actively prevent biofilm formation during the healing period, and (2) the coating should be thin enough to be fully dissolved after completion of healing to minimize the antimicrobial ions’ impact on the long-term health. It should be noted that this study focuses on demonstrating the feasibility of using atomic layer deposition (ALD) technique to deposit conformal and thin antibacterial coatings on complex hierarchically restructured surfaces of electrodes for neurostimulation and cardiac rhythm management devices. Investigation of the optimum ALD film thickness will be a future work.
However, achieving these two very essential attributes of the surface coating—i.e. low thickness and conformality—is extremely challenging for conventional coating techniques, such as physical vapor deposition (PVD) and chemical vapor deposition (CVD), due to their line-of-sight attributes and/or difficulty of atomic-level thickness control47. On the contrary, use of an atomic layer deposition (ALD) technique is ideal (and perhaps the only known suitable technique) to achieve thin and conformal coatings with excellent thickness-control and repeatability in a manufacturing environment due to the intrinsically self-limiting coating mechanism of the ALD technology47.
Objectives
It is hypothesized that the ALD technique can efficiently deposit conformal antibacterial thin films on the nano-structures induced on the surface of hierarchically restructured Pt10Ir electrodes, as illustrated in Fig. 2. To test this, a two-step development process was used in this work. Electrodes were first hierarchically restructured using femtosecond-laser hierarchical surface restructuring (HSR™) technology12. In the second step of the process, ALD was utilized to deposit thin and conformal copper oxide (CuxO) films on the hierarchical surface structure of the electrodes to impart antibacterial properties to the electrodes. Subsequently, antibacterial properties and electrochemical performance of the ALD coated electrodes were evaluated. Antibacterial properties of the electrodes were studied, particularly, on the killing effect of the antibacterial electrodes on Escherichia coli and Staphylococcus aureus—two common types of bacteria responsible for infections on implanted devices. This work will demonstrate that the CuxO ALD thin films deposited on hierarchically restructured Pt10Ir electrodes exhibit significant antibacterial properties while having minimal impact on the electrochemical performance of the electrodes. In addition, morphological, compositional, and structural properties of the ALD CuxO coated hierarchically restructured Pt10Ir electrodes were studied using multiple modalities of microscopy and spectroscopy.
Materials and methods
Femtosecond laser hierarchical surface restructuring
Full details of hierarchical surface restructuring technique are outlined in a previous work12. In summary, a series of flat 0.3 mm thick Pt10Ir electrodes (6 and 10 mm diameter disc-shaped electrodes for electrochemical measurements and antibacterial studies, respectively, and 1 cm by 1 cm square electrodes for microscopy and spectroscopy) were hierarchically restructured and laser cut using a femtosecond laser system. The laser system used was a diode pumped Yb:YAG solid state laser (Coherent StarFemto, Santa Clara, CA) that generates 300 fs pulses with a central wavelength of 1030 nm. The experiments were performed in air, under ambient conditions. Surface patterns were created via a graphical editor (Visual Laser Marker provided by Coherent), tied into the axis controls, and the beam path was directed using a deflection head. Electrodes were mounted onto a porous ceramic vacuum plate (PhotoMachining, Inc., Pelham, NH) mounted on a tip-tilt stage (Edmunds Optics, Barrington, NJ) on an XYZ-translation stage. The electrodes were leveled to within 5 µm delta across the surface using an optical non-contact displacement transducer (Micro Epsilon, Ortenburg, Germany). Electrodes were brought directly under the deflection head to minimize the incident angle.
ALD coating process
There are mainly two types of copper oxides: cupric oxide (CuO) and cuprous oxide (Cu2O)48,49,50, which will be referred to as CuxO hereafter. Cu2O and CuO are semiconductors with significant electronic conductivity48,49,50, which is beneficial as potentially conductive antibacterial coatings for electrodes in neurostimulation applications. CuxO also has significant antibacterial properties40,41,45,46. Copper (Cu) is an essential element to human health51, and also human tissue shows much larger resistance to Cu compared to micro-organisms52,53,54. Cu ions also play a crucial role in modulating wound healing process55,56, therefore from a biosafety and biocompatibility perspective, CuxO, in small quantities57, is a safe choice to be applied on long-term implantable electrodes for neurostimulation and cardiac rhythm management device applications. There are various deposition techniques for thin film deposition of CuxO, e.g. spin coating58,59, sputtering60,61,62, pulsed laser deposition63,64, solution processing65,66, and ALD48,49,50. ALD is a sequential and self-limiting chemical deposition process47 where the substrate is brought to various precursors and co-reactants followed by inert gas purges. Due to the ALD’s self-limiting process, the ALD films are highly conformal, uniform and well controlled even for substrates with a complex morphology47. The previous studies on CuxO ALD deposition are limited48,49,50, and a recent study demonstrated feasibility of depositing CuxO ALD films using plasma-enhanced ALD (PEALD)50. However, due to plasma radical recombination, PEALD is known to have difficulty to form conformal coatings on hierarchical surfaces47,67. We investigated here that thermal ALD using ozone as the co-reactant can efficiently create conformal CuxO ALD films. In this work, CuxO films were grown using a Veeco Fiji G2 system (Veeco-CNT, Waltham, MA, USA). Cu(dmap)2, namely, (bis(dimethylamino-2-propoxide) copper(II)), was used as the Cu-containing precursor (Ascensus-Strem, Bellevue, WA, USA). The precursor source was maintained at 125 °C using a Low Vapor Pressure Delivery (LVPD) module (Veeco-CNT, Waltham, MA, USA). Argon gas was used as the carrier gas with a constant flow rate of 30 sccm. Ozone (O3) was used as the co-reactant. The substrate temperature was maintained at 150 °C. Each ALD cycle consisted of a 2 s Cu precursor pulse and then a 0.075 s O3 pulse. Three types of substrates/electrodes were used: single crystal silicon (Si) wafers, and flat as-rolled un-restructured and smooth Pt10Ir electrodes (Pt10Ir), and flat Pt10Ir electrodes processed using hierarchical surface restructuring technology (HSR-Pt10Ir). In order to distinguish the ALD-coated and uncoated regions of the substrates/electrodes for characterization purposes, Kapton tape was used to block/mask the ALD68 deposition, so that only the unmasked regions could be ALD-coated.
Spectroscopic ellipsometry and atomic force microscopy
A spectroscopic ellipsometer (M-2000, J.A. Woollam, Lincoln, NE, USA) was used to analyze the thickness and optical properties of the ALD coatings. CompleteEASE™ software was used to analyze the ellipsometry data for Si substrate with CuxO coating. A Park Systems AFM (Park NX10, Santa Clara, CA, USA) with pre-mounted tip (OMCL-AC160TS, Santa Clara, CA, USA) was used in non-contact (tapping) mode (NCM) to analyze surface morphology and film thickness on the masked Si wafer. Scanning was started from the uncoated region and moved towards the coated region crossing the boundary to determine the thickness difference across the boundary.
Scanning electron microscopy, EDS analysis, and focused ion beam milling
SEM imaging was performed using a ZEISS Crossbeam 340 (ZEISS, Oberkochen, Germany). Measurements of composition were performed using energy dispersive spectroscopy (EDS) in a Thermo Fisher Scientific Apreo S (Brno, Czech Republic). The Apreo is configured with a thermally assisted field emitter and an Oxford Instruments X-MAX 50 EDS system (High Wycombe, United Kingdom) coupled with Aztec v3.3 software capable of mapping and selected area scans. All EDS data was collected using 10 keV electrons that focused on the Ma-lines for Pt and Ir, the La-line for Cu, and the Ka-lines for O and Si. Samples were masked with Kapton tape so that the interface between the CuxO coated and uncoated regions could be clearly defined. Compositional EDS maps of CuxO coated and uncoated regions were collected on Si substrates and HSR-Pt10Ir electrodes. To investigate the depth of CuxO penetration into the HSR-Pt10Ir electrodes, cross-sections were milled across the coated/uncoated interface using a Tescan S8252X Plasma Focused Ion Beam (FIB) system (Brno-Kohoutovice, Czech Republic) with a Xe plasma. Xe-ions were accelerated with a 30 kV potential difference. Cross-sections were milled and imaged in the dual beam system. Subsequent measurements of composition of the materials were performed using the Apreo system with tilt stage and selected area scans. After tilting so that there was a line-of-sight path between the milled region of interest and the X-ray detector, selected area compositional scans were collected to approximately determine the depth of the Cu-signal in the restructured electrodes.
Grazing-incidence X-ray diffraction and X-ray photoelectron spectroscopy
Grazing-incidence X-ray diffraction (GIXRD) using a Rigaku XRD system (Rigaku SmartLab X-ray diffractometer, Woodlands, TX, USA) with angle of incidence of ~ 0.5°–1° was performed. X-ray generator of 2.2 kW power was used with Cu Kα X-ray tube. Parallel Beam Optics was used out of two Bragg Brentano and Parallel Beam optics. Scanning was done from 30° to 50°. X-ray photoelectron spectroscopy (XPS) compositional analysis was performed using a PHI VersaProbe 5000 (Chanhassen, MN, USA). Monochromatic Al Kα source of 1486.2 eV was used. The XPS spot size was 200 µm and calibration was performed using C–C component of C 1 s peak at 284.8 eV. Cu 2p spectra were measured between 925 to 965 eV. Casa XPS software was used to analyze spectra and Shirley type backgrounds were used.
Electrochemical measurements
Cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) were used to measure charge storage capacity (CSC), specific capacitance and impedance12,14,69,70,71. Five different sets of electrodes were produced and used for CV and EIS measurements: Pt10Ir electrodes processed using hierarchical surface restructuring technology (HSR-Pt10Ir) tested as control and labelled as “Control”, and four sets of HSR-Pt10Ir electrodes coated with CuxO ALD coatings at 5.0 nm, 9.5 nm, 12.5 nm, and 24 nm (HSR-Pt10Ir-5 nm labeled as A1, HSR-Pt10Ir-9.5 nm labelled as A2, HSR-Pt10Ir-12.5 nm labelled as A3, and HSR-Pt10Ir-24 nm labelled as A4, respectively). The voltage in the CV tests was restricted and determined by the water window representing the potential-range where oxidation or reduction currents will not lead to formation of hydrogen or oxygen at the electrode/tissue interface (− 0.6–0.8 V vs. a Ag/AgCl reference electrode)12,14. Both CV and EIS tests were performed in a three-electrode Teflon® plate cell (Fig. 3), comprising a Ag/AgCl reference electrode (ALS-Co Ltd., RE-1B, Tokyo, Japan), a coiled Pt counter-electrode, and the working electrodes.
The geometric surface area (GSA) of the working electrodes in the cell was 0.09 cm2. The electrolyte used was a commercially available phosphate-buffered saline (PBS) solution (Blood Bank Saline, Azer Scientific, Morgantown, PA). All potentials were recorded with respect to the Ag/AgCl reference electrode. All CV tests were performed at room temperature and at a 50 mV/s voltage sweep rate (ν) between potential limits of − 0.6 V and 0.8 V, beginning at open-circuit potential (OCP) and sweeping in the positive direction first. As outlined earlier, potential windows were selected to ensure water electrolysis did not occur. EIS measurements were performed at OCP and measured over a frequency range of 0.1–105 Hz using a 10 mV root-mean-square (Vrms) sinusoidal excitation voltage amplitude about a fixed potential between − 0.6 V and 0.8 V. All CV and EIS measurements were performed using a Gamry potentiostat (5000E interface, Warminster, PA) and the vendor supplied software. All data reported for CV and EIS are an average of 3 electrodes per condition, tested 3 times, i.e., a total of 9 measurements. Specific capacitance was calculated using CV/EIS data and common Randles model.
Inductively coupled plasma mass spectrometry
Electrodes were tested for static release of copper ions into solution using inductively coupled plasma mass spectrometry (ICP-MS) as described previously72. Each electrode was submerged in 10 ml liquid medium and incubated at 37 °C with shaking (225 RPM) for a total period of 70 min. At 10-min intervals, 0.5 ml aliquots from these samples were removed and mixed with 5 ml 1% nitric acid, which was subsequently injected into an Agilent 7900 ICP-MS for ion release analysis. Ion concentrations were determined by comparison to a standard curve ranging from 62.5 to 4000 ppb using Aristar Copper ICP standard. All samples were tested in at least duplicates and data reported represent the average and standard deviation or range of results.
Antibacterial measurements
All experiments were carried out at physiological temperature of 37 °C. Model bacterial strains of E. coli and S. aureus were used to differentiate activity against Gram-negative and Gram-positive strains, respectively. Two types of liquid media were used: Luria–Bertani broth (LB; Difco LB Broth, Miller) and nutrient broth (NB; Difco Nutrient Broth) bacterial growth media. All electrodes were tested in at least duplicate and data reported represent the average and standard deviation or range of results. Antibacterial analysis includes two types of characterizations: (1) Bacterial surface and contact activities: electrodes were tested for bacterial adhesion and contact-dependent antibacterial activity. Cultures of bacteria were inoculated from single colonies and allowed to grow overnight to stationary phase. An aliquot of this overnight culture was diluted 1:250 in fresh media and allowed to grow at 37 °C with shaking until the optical density at the wavelength of 600 nm (OD600) of the culture was in the range of 0.2–0.7, indicating mid-log growth phase. Next, an aliquot of the culture was diluted to 1 × 105 CFU/ml in fresh growth medium and 10 μl of this diluted culture was deposited on a sample surface and incubated for 60 min at 37 °C73. Then, any remaining liquid was aspirated and the surface of the electrode was swabbed to isolate any remaining viable bacteria. The aspirated liquid was added to 3 ml of growth medium and allowed to incubate overnight at 37 °C with shaking. Next, the swab was streaked onto a petri dish containing culture medium, and then the swab was subsequently used to inoculate 3 ml of fresh culture medium. These cultures were then subjected to an 18-h incubation at 37 °C (with shaking for the liquid cultures). Bacterial growth was determined using direct visualization for the petri dish and the turbidity of the liquid cultures was measured using OD600; (2) Modified Kirby Bauer—zone of inhibition (ZOI) test: electrodes were tested for their ability to kill bacteria on semi-solid growth media surfaces using a modified Kirby-Bauer disk diffusion assay as described previously74. Briefly, electrodes were laid coated-side down onto solid nutrient agar pre-inoculated with E. coli (D31) or S. aureus (ATCC: 27660). The petri dish was allowed to incubate for 18 h at which point the diameter of the zones, if present, were measured. A zone of inhibition (ZOI) around the electrodes represents the failure of the bacteria to grow due to Cu ion release, with larger zones representing a greater degree of growth inhibition and/or bacterial killing.
Results and discussion
Morphology and microstructure
For morphology and microstructural characterization purposes, 950 cycles of ALD were conducted. Ellipsometry measurements of coated Si wafers measured the film to be at 21.4 nm ± 0.4 nm, indicating an average growth rate of ~ 0.022 nm/cycle. Thickness of the CuxO films were also measured to be at ~ 24 nm via AFM scanning of ALD-coated Si wafer that was masked with Kapton tape after removing the Kapton tape (Fig. 4).
The AFM measured film thickness was in reasonable agreement with the ~ 21.4 ± 0.4nm thickness measured by ellipsometry. It should be noted that the coating was visible on the Si wafer with naked eye appearing in a purple color tune. The coating uniformly changes the contrast in the SEM micrographs of the Si wafer suggesting an overall uniform coating.
Figure 5 (top row) shows the SEM micrographs of uncoated HSR-Pt10Ir electrodes at various magnifications. Figure 5 (bottom row) shows the SEM micrographs of ALD coated HSR-Pt10Ir electrodes after 950 ALD cycles at ~ 24 nm CuxO film at various magnifications. Comparison of the uncoated versus ALD-coated electrodes shows that the ALD coating process has not changed the overall hierarchical surface structure of the electrodes. Figure 6 shows high magnification SEM micrographs of the nano-structured regions of the uncoated (Fig. 6a) versus ALD-coated (Fig. 6b) Pt10Ir electrodes, suggesting that the ALD-coating conformally and uniformly coats the surface, and the fine nano-structured features of the electrodes become very slightly blunted due to the ~ 24 nm ALD-coating coverage.
Composition and structure
Figures 7, 8, and 9 show the SEM image and corresponding elemental EDS maps of an ALD-coated Si wafer, an ALD-coated Pt10Ir electrode, and an ALD-coated HSR-Pt10Ir electrode, respectively. All three substrates were masked with Kapton tape. The EDS maps indicate that: (a) Pt and Ir universally exist as expected for Pt10Ir electrodes; (b) Cu exists only on the CuxO coated regions of the electrodes; (c) Si content is lower on the CuxO-coated region of the Si sample, corresponding to the Si X-rays that are generated being blocked/absorbed by the Cu-containing coating on the coated region; (d) oxygen exists on both coated and uncoated regions, where in the coated side the oxygen distribution matches well with the Cu distribution, as expected for the CuxO coating, and on the uncoated side, the oxygen distribution matches well with the Si distribution as expected for the oxygen-containing silicone adhesive residues of the Kapton tape. Collectively, the EDS maps of all substrates qualitatively indicate that the ALD process can uniformly deposit CuxO on the substrates and electrodes. Additionally, quantitative EDS analysis of the film on Si wafer suggests 1:1 ratio of Cu to O, suggesting that the coating is dominantly CuO, consistent with a previous study of using ozone as co-reactant75 as well as the XPS results discussed below.
Figure 10 shows the XPS Cu-2p spectra of Si wafer, Pt10Ir electrode and HSR-Pt10Ir electrode coated with CuxO. The spectra indicates ~ 933.7 eV (Cu 2p3/2), and ~ 953.7 eV (Cu-2p1/2) peaks, and also their satellite peaks at ~ 941.5 eV and ~ 943.6 eV (Cu 2p3/2 satellite) as well as ~ 962.3 eV (Cu 2p1/2 satellite), respectively. The satellite peaks are distinguishing features of CuO75,76,77,78, which do not show up for metallic Cu or Cu2O75,76,77. Figure 11 shows the GIXRD spectra of CuxO ALD coated Si wafer, again suggesting CuO49,79. Thus, based on EDS, XPS, and GIXRD, crystalline CuO is the dominant phase in the ALD deposited films.
Electrochemical performance
Figure 12a,c display representative cyclic voltammograms and Electrochemical Impedance Spectroscopy (EIS) data for HSR-Pt10Ir electrodes (control), and HSR-Pt10Ir electrodes coated with CuxO ALD coatings at 5.0 nm (A1), 9.5 nm (A2), 12.5 nm (A3), and 24 nm (A4), respectively. Figure 12b shows the plots of CSCtotal and specific capacitance as a function of ALD coating thickness. Figure 12d shows the Nyquist plots of phase angle response of control, A1, A2, A3, and A4 electrodes plotted in the 0.1–105 Hz frequency range. The EIS data (Fig. 12c) is demonstrated in bode-plot format in which the logarithm of the impedance is plotted as a function of the logarithm of frequency (plotted in the 0.1–105 Hz frequency range). At high frequencies (greater than approximately 1 kHz), impedance magnitudes showed resistive behavior representing solution resistance. The CV results indicate that the total charge storage capacity (CSCtotal) of the electrodes (with uncoated “control” electrode being the baseline) slightly increases after CuxO ALD deposition. This increase in CSCtotal could likely be attributed to the semiconducting properties of CuO48,49,50. Specific capacitance of the CuxO coated electrodes, however, seems to have decreased slightly compared to the uncoated control HSR-Pt10Ir electrode. The reason(s) for this slight decrease in capacitance is not yet entirely understood but it could be attributed to the slight decrease in the double layer capacitance elements in the hierarchical surface structure of the electrodes’ surface (the peaks and valleys) being slightly blunted due to increase in thickness of the CuxO ALD-coating coverage from ~ 5 nm (A1) to ~ 24 nm (A4), as shown in the SEM and FIB micrographs of Figs. 5 and 6. The EIS bode plots suggest that the overall impedance behavior of the electrodes as a function of frequency seem to have remained unchanged (within experimental scatter). The response of phase angle, which is the arctangent of the ratio of the imaginary and real parts of the measured impedance, is shown in the Nyquist plots of phase angle response plotted as a function of frequency (log) (Fig. 12d). At frequencies above 100 Hz, the phase angles for all the electrodes were close to zero, which indicates the purely resistive nature of the system. At frequencies below 100 Hz, the electrodes had none-zero phase angle values decreasing to ~ − 60° as DC frequency was approached indicating the presence of reactance. Overall, the CV and EIS measurements suggest that the CuxO ALD coatings have insignificant impact on the overall electrochemical performance of the electrodes post-ALD.
Antibacterial properties
CuO is known to act as an antibacterial80,81. The mechanism for CuO’s antibacterial capability is complex and multifold but is directly linked to the interaction of the Cu ions with the bacterium. Some possible mechanisms are: (1) Cu ions can change the charge of the surrounding medium to break and rupture cell wall membrane82; (2) Cu ions can alter DNA molecules and disrupt the biochemical process in bacteria83. When the concentration of Cu ion reaches a threshold in the culture medium, the bacteria would be killed with sufficient time83,84,85. Thus, to check the Cu release of the ALD-coated electrodes, ICP-MS measurements were conducted on Pt10Ir and HSR-Pt10Ir electrodes (ALD-coated and uncoated). Figure 13 shows Cu ion concentration as a function of time in LB, NB, and H2O. As expected, the uncoated Pt10Ir and HSR-Pt10Ir electrodes show no evidence of released Cu ions over time due to the absence of any CuO coating on those electrodes. It should be noted that the charge state of the Cu ions is not determined in this work, and they may exist as Cu1+, Cu2+, or a mixture of these.
Figure 13 indicates that Cu ion concentration in ALD-coated Pt10Ir electrodes for all three types of media continues to slowly increase with increasing electrode immersion time up to ~ 40 min after which the Cu ion concentration reaches a plateau of ~ 0.2 ppm. With water and LB as the media, the Cu ion concentration of ALD coated HSR-Pt10Ir electrodes rapidly increases as a function of time and reaches ~ 1.9 ppm after 70 min. In NB media, the Cu ion concentration of ALD coated HSR-Pt10Ir electrodes increase but at a slower rate and reach ~ 1.7 ppm after 70 min. This is strong evidence that hierarchical surface restructuring of the electrodes prior to ALD coating can increase the Cu ion release rate by nearly one order of magnitude due to the increase in surface area of the electrodes after restructuring12. The ion release experiments also clearly suggest that the CuO coating can release Cu ions into the surrounding medium.
Bacterial growth on a swabbed petri dish and the turbidity (OD600) of the liquid culture inoculated after surface sampling (in LB media) are shown in Figs. 14 and 15. The results interestingly indicate that uncoated Pt10Ir electrodes show no inhibition of S. aureus growth but do exhibit inhibition of E. coli growth, suggesting that Pt10Ir itself is intrinsically antibacterial86 against E. coli but not against S. aureus in direct contact. On the other hand, the coated Pt10Ir electrodes showed similar behavior as their uncoated counterparts, which may be attributed to the fact that the coated Pt10Ir electrodes did not rapidly release sufficient Cu ions (see Fig. 13) to reach the required threshold concentration to inhibit the growth of S. aureus. This could be attributed to a reduced accessible surface area in comparison to the conformal coatings on the restructured electrodes.
Figure 14 also shows that the uncoated HSR-Pt10Ir electrode inhibited the growth of both E. coli and S. aureus, which might be the combination of antibacterial Pt10Ir86 surface and the nanostructured morphology resulting from hierarchical surface restructuring87,88. Sharp surface nanomorphology has been previously shown to be capable of killing Gram-negative (e.g., E. coli) and Gram-positive (e.g., S. aureus) bacteria in contact, possibly through physically rupturing the bacterial cell wall87,88. Thus, hierarchical surface restructuring of electrodes creates a hierarchical surface structure that induces bacterial-killing characteristics in contact while demonstrating high electrochemical performance.
Figure 15 shows the OD600 for the surface-contacting liquid culture, again indicating that only the uncoated and CuO-coated Pt10Ir electrodes did not inhibit S. aureus growth on contact and all HSR-Pt10Ir electrodes are antibacterial in direct contact against both E. coli and S. aureus. This is consistent with the results of bacterial growth on swabbed petri dishes (Fig. 14). Thus, only based on the study of bacterial surface and contact activities as in Fig. 14, the effect of hierarchical surface restructuring combined with CuO coating cannot be differentiated.
Considering the previous results, a modified Kirby Bauer method to test zone of inhibition (ZOI) was used to test the effect of CuO coatings deposited on hierarchically restructured electrodes. ZOI assays in NB are shown in Fig. 16, indicating zero ZOI for uncoated HSR-Pt10Ir electrodes and significant ZOI for coated HSR-Pt10Ir electrodes: 19.3 ± 0.2 mm for E. coli and 16 ± 1.5 mm for S. aureus. This implies that the CuO coating, i.e., the Cu ion release, is solely responsible for the formation of ZOIs. Thus, while hierarchically restructured electrodes can kill E. coli and S. aureus in direct contact mainly due to the sharp nanomorphology (Figs. 14, 15), the CuO coating on hierarchically restructured electrodes is capable of releasing sufficient antibacterial Cu ions to kill E. coli and S. aureus in the surrounding medium that is not directly contacting with the electrode surface (Fig. 16).
Concluding remarks
In this report, we introduced a two-step process using femtosecond laser hierarchical surface restructuring (HSR™) and atomic layer deposition (ALD) for development of electrochemically active antibacterial platinum-iridium (Pt10Ir) electrodes targeted for use in cardiac rhythm management device, and neurostimulation and sensing/recording applications. We showed that after hierarchical surface restructuring, the surface of Pt10Ir electrodes undergo topographical transformations that result in electrodes with superior electrochemical performance. These topographic features span a variety of length scales including coarse-scale rough microstructures and a finer nanostructure subset on top of the coarse structures that impart high electrochemical performance to the electrodes. Through multiple modalities of microscopy and spectroscopy for surface and subsurface characterization of the ALD-coated hierarchically restructured electrodes, we demonstrated that the electrodes were coated with ultra-thin (~ 24 nm) and conformal CuO films without changing the overall hierarchical surface structure of the electrodes. The results of electrochemical measurements on ALD-coated hierarchically restructured electrodes indicated that the total charge storage capacity of the electrodes slightly and favorably increases after CuO ALD deposition, possibly attributed to the semiconducting properties of CuO. The antibacterial properties of the ALD-coated hierarchically restructured electrodes were also studied, particularly, the killing effect of the electrodes on E. coli and S. aureus—two common types of bacteria responsible for implant infections. It was shown that uncoated hierarchically restructured electrodes can kill E. coli and S. aureus in direct contact likely due to the sharp nanomorphology of the hierarchical structures on the surface of the electrodes. More importantly, it was demonstrated that the ALD-deposited CuO coating on hierarchically restructured electrodes is capable of releasing sufficient antibacterial Cu ions to kill E. coli and S. aureus in the surrounding medium that is not directly contacting with the electrode surface. Thus, this study shows that antibacterial ultra-thin and conformal CuO films can be deposited via ALD on hierarchically restructured Pt10Ir electrodes with minimal impact on their electrochemical properties, rendering the electrodes antibacterial. Since the threat of antibiotic resistance is significant and is increasingly being recognized as a global problem, the findings reported in this study are insightful and paramount for design and manufacturing of neurostimulation and cardiac rhythm management electrodes with antibacterial properties to eliminate post-implantation infections and reduce antibiotic use.
Data availability
All datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
References
Mulpuru, S. K., Madhavan, M., McLeod, C. J., Cha, Y.-M. & Friedman, P. A. Cardiac pacemakers: Function, troubleshooting, and management: Part 1 of a 2-part series. J. Am. Coll. Cardiol. 69, 189–210. https://doi.org/10.1016/j.jacc.2016.10.061 (2017).
Stevenson, I. & Voskoboinik, A. Cardiac rhythm management devices. Aust. J. Gen. Pract. 47, 264–271 (2018).
Ordonez, J., Schuettler, M., Boehler, C., Boretius, T. & Stieglitz, T. Thin films and microelectrode arrays for neuroprosthetics. MRS Bull. 37, 590–598. https://doi.org/10.1557/mrs.2012.117 (2012).
Theogarajan, L. Strategies for restoring vision to the blind: Current and emerging technologies. Neurosci. Lett. 519, 129–133. https://doi.org/10.1016/j.neulet.2012.02.001 (2012).
Wilson, B. S. et al. Better speech recognition with cochlear implants. Nature 352, 236–238. https://doi.org/10.1038/352236a0 (1991).
Stellbrink, C. & Trappe, H.-J. The follow-up of cardiac devices: What to expect for the future?. Eur. Heart J. Suppl. 9, I113–I115. https://doi.org/10.1093/eurheartj/sum071 (2007).
Halperin, D., Heydt-Benjamin, T. S., Fu, K., Kohno, T. & Maisel, W. H. Security and privacy for implantable medical devices. IEEE Pervasive Comput. 7, 30–39. https://doi.org/10.1109/MPRV.2008.16 (2008).
Maisel, W. H. Safety issues involving medical devices: Implications of recent implantable cardioverter-defibrillator malfunctions. JAMA 294, 955–958. https://doi.org/10.1001/jama.294.8.955 (2005).
Krames, E., Peckham, P. H. & Rezai, A. R. Neuromodulation: Comprehensive Textbook of Principles, Technologies, and Therapies Second. (Academic Press is an imprint of Elsevier, 2018).
Razavi, B. et al. Real-world experience with direct brain-responsive neurostimulation for focal onset seizures. Epilepsia 61, 1749–1757. https://doi.org/10.1111/epi.16593 (2020).
Eljamel, S. & Slavin, K. Neurostimulation: Principles and Practice (Wiley, 2013).
Amini, S. et al. Femtosecond laser hierarchical surface restructuring for next generation neural interfacing electrodes and microelectrode arrays. Sci. Rep. 12, 13966. https://doi.org/10.1038/s41598-022-18161-4 (2022).
Bergey, G. K. Neurostimulation in the treatment of epilepsy. Exp. Neurol. 244, 87–95. https://doi.org/10.1016/j.expneurol.2013.04.004 (2013).
Cogan, S. F. Neural stimulation and recording electrodes. Annu. Rev. Biomed. Eng. 10, 275–309. https://doi.org/10.1146/annurev.bioeng.10.061807.160518 (2008).
Boehler, C., Stieglitz, T. & Asplund, M. Nanostructured platinum grass enables superior impedance reduction for neural microelectrodes. Biomaterials 67, 346–353. https://doi.org/10.1016/j.biomaterials.2015.07.036 (2015).
Guenther, T., Lovell, N. H. & Suaning, G. J. Bionic vision: System architectures—A review. Expert Rev. Med. Devices 9, 33–48. https://doi.org/10.1586/erd.11.58 (2012).
Carlson, M. L., Driscoll, C. L., Gifford, R. H. & McMenomey, S. O. Cochlear implantation: Current and future device options. Otolaryngol. Clin. North Am. 45, 221–248. https://doi.org/10.1016/j.otc.2011.09.002 (2012).
Epstein, L. J. & Palmieri, M. Managing chronic pain with spinal cord stimulation. Mount Sinai J. Med. A J. Transl. Personal. Med. 79, 123–132. https://doi.org/10.1002/msj.21289 (2012).
Bradley, K. The technology: The anatomy of a spinal cord and nerve root stimulator: The lead and the power source. Pain Med. 7, S27–S34. https://doi.org/10.1111/j.1526-4637.2006.00120.x (2006).
Daubinger, P., Kieninger, J., Unmussig, T. & Urban, G. A. Electrochemical characteristics of nanostructured platinum electrodes—A cyclic voltammetry study. Phys. Chem. Chem. Phys. 16, 8392–8399. https://doi.org/10.1039/c4cp00342j (2014).
Norlin, A., Pan, J. & Leygraf, C. Investigation of electrochemical behavior of stimulation/sensing materials for pacemaker electrode applications II. Conducting oxide electrodes. J. Electrochem. Soc. https://doi.org/10.1149/1.1933372 (2005).
Ko, W. H. Early history and challenges of implantable electronics. ACM J. Emerg. Technol. Comput. Syst. 8, 8. https://doi.org/10.1145/2180878.2180880 (2012).
Bazaka, K. & Jacob, M. Implantable devices: Issues and challenges. Electronics 2, 1–34. https://doi.org/10.3390/electronics2010001 (2012).
Green, R. A. et al. Laser patterning of platinum electrodes for safe neurostimulation. J. Neural Eng. 11, 056017. https://doi.org/10.1088/1741-2560/11/5/056017 (2014).
Green, R. A., Lovell, N. H., Wallace, G. G. & Poole-Warren, L. A. Conducting polymers for neural interfaces: Challenges in developing an effective long-term implant. Biomaterials 29, 3393–3399. https://doi.org/10.1016/j.biomaterials.2008.04.047 (2008).
Ban, K. A. et al. American college of surgeons and surgical infection society: Surgical site infection guidelines, 2016 update. J. Am. Coll. Surg. 224, 59 (2017).
Anderson, D. J. et al. Strategies to prevent surgical site infections in acute care hospitals: 2014 update. Infect. Control Hosp. Epidemiol. 35, 605–627. https://doi.org/10.1086/676022 (2014).
Fuglestad, M. A., Tracey, E. L. & Leinicke, J. A. Evidence-based prevention of surgical site infection. Surg. Clin. North Am. 101, 951–966. https://doi.org/10.1016/j.suc.2021.05.027 (2021).
VanEpps, J. S. & Younger, J. G. Implantable device-related infection. Shock 46, 597–608. https://doi.org/10.1097/shk.0000000000000692 (2016).
Trampuz, A. & Widmer, A. F. Infections associated with orthopedic implants. Curr. Opin. Infect. Dis. 19, 349–356. https://doi.org/10.1097/01.qco.0000235161.85925.e8 (2006).
Filipović, U., Dahmane, R. G., Ghannouchi, S., Zore, A. & Bohinc, K. Bacterial adhesion on orthopedic implants. Adv. Colloid Interface Sci. 283, 102228. https://doi.org/10.1016/j.cis.2020.102228 (2020).
Pietrocola, G. et al. Colonization and infection of indwelling medical devices by Staphylococcus aureus with an emphasis on orthopedic implants. Int. J. Mol. Sci. 23, 5958 (2022).
Khatoon, Z., McTiernan, C. D., Suuronen, E. J., Mah, T. F. & Alarcon, E. I. Bacterial biofilm formation on implantable devices and approaches to its treatment and prevention. Heliyon 4, e01067. https://doi.org/10.1016/j.heliyon.2018.e01067 (2018).
Nagaiah, H. P., Kandaswamy, K., Priya, A., Kasthuri, T. & Pandian, S. K. in Understanding Microbial Biofilms (eds Das, S. & Kungwani, N. A.) 335–353 (Academic Press, 2023).
Li, Y., Xiao, P., Wang, Y. & Hao, Y. Mechanisms and control measures of mature biofilm resistance to antimicrobial agents in the clinical context. ACS Omega 5, 22684–22690. https://doi.org/10.1021/acsomega.0c02294 (2020).
Antibiotic resistance threats in the United States, 2019. (2019).
Yoneyama, H. & Katsumata, R. Antibiotic resistance in bacteria and its future for novel antibiotic development. Biosci. Biotechnol. Biochem. 70, 1060–1075. https://doi.org/10.1271/bbb.70.1060 (2006).
Laxminarayan, R. et al. Antibiotic resistance—The need for global solutions. Lancet Infect. Diseases 13, 1057–1098. https://doi.org/10.1016/s1473-3099(13)70318-9 (2013).
Barie, P. S. & Eachempati, S. R. Surgical site infections. Surg. Clin. North Am. 85, 1115–1135. https://doi.org/10.1016/j.suc.2005.09.006 (2005).
Mostafavi, E. et al. Antimicrobial surfaces for implantable cardiovascular devices. Curr. Opin. Biomed. Eng. 23, 100406. https://doi.org/10.1016/j.cobme.2022.100406 (2022).
Shahid, A. et al. The prospects of antimicrobial coated medical implants. J. Appl. Biomater. Funct. Mater. 19, 22808000211040304. https://doi.org/10.1177/22808000211040304 (2021).
Taylor, G. et al. Investigation of iridium, ruthenium, rhodium, and palladium binary metal oxide solid solution thin films for implantable neural interfacing applications. Surf. Coat. Technol. https://doi.org/10.1016/j.surfcoat.2021.127803 (2021).
Taylor, G. et al. Electrochemical enhancement of reactively sputtered rhodium, ruthenium, and iridium oxide thin films for neural modulation, sensing, and recording applications. Electrochim. Acta https://doi.org/10.1016/j.electacta.2021.139118 (2021).
Page, N. et al. The effect of deposition parameters on microstructure and electrochemical performance of reactively sputtered iridium oxide coatings. Mater. Today Commun. https://doi.org/10.1016/j.mtcomm.2021.102967 (2021).
Dizaj, S. M., Lotfipour, F., Barzegar-Jalali, M., Zarrintan, M. H. & Adibkia, K. Antimicrobial activity of the metals and metal oxide nanoparticles. Mater. Sci. Eng. C 44, 278–284. https://doi.org/10.1016/j.msec.2014.08.031 (2014).
Khezerlou, A., Alizadeh-Sani, M., Azizi-Lalabadi, M. & Ehsani, A. Nanoparticles and their antimicrobial properties against pathogens including bacteria, fungi, parasites and viruses. Microb. Pathog. 123, 505–526. https://doi.org/10.1016/j.micpath.2018.08.008 (2018).
Johnson, R. W., Hultqvist, A. & Bent, S. F. A brief review of atomic layer deposition: From fundamentals to applications. Mater. Today 17, 236–246. https://doi.org/10.1016/j.mattod.2014.04.026 (2014).
Tamm, A., Tarre, A., Verchenko, V., Seemen, H. & Stern, R. Atomic layer deposition of superconducting CuO thin films on three-dimensional substrates. Crystals 10, 650 (2020).
Avila, J. R., Eddy, C. R. & Wheeler, V. D. Comprehensive characterization of copper oxide atomic layer deposition using water or ozone with enhanced bis-(dimethylamino-2-propoxide) copper delivery. J. Vac. Sci. Technol. A 38, 042403. https://doi.org/10.1116/6.0000248 (2020).
Lenef, J. D. et al. Plasma-enhanced atomic layer deposition of p-type copper oxide semiconductors with tunable phase, oxidation state, and morphology. J. Phys. Chem. C 125, 9383–9390. https://doi.org/10.1021/acs.jpcc.1c00429 (2021).
Olivares, M., Pizarro, F., Speisky, H., Lönnerdal, B. & Uauy, R. Copper in infant nutrition: Safety of World Health Organization provisional guideline value for copper content of drinking water. J. Pediatr. Gastroenterol. Nutr. 26, 251 (1998).
Ren, G. et al. Characterisation of copper oxide nanoparticles for antimicrobial applications. Int. J. Antimicrob. Agents 33, 587–590. https://doi.org/10.1016/j.ijantimicag.2008.12.004 (2009).
Gunawan, C., Teoh, W. Y., Marquis, C. P. & Amal, R. Cytotoxic origin of Copper(II) oxide nanoparticles: Comparative studies with micron-sized particles, leachate, and metal salts. ACS Nano 5, 7214–7225. https://doi.org/10.1021/nn2020248 (2011).
Heinlaan, M., Ivask, A., Blinova, I., Dubourguier, H.-C. & Kahru, A. Toxicity of nanosized and bulk ZnO, CuO and TiO2 to bacteria Vibrio fischeri and crustaceans Daphnia magna and Thamnocephalus platyurus. Chemosphere 71, 1308–1316. https://doi.org/10.1016/j.chemosphere.2007.11.047 (2008).
Ermini, M. L. & Voliani, V. Antimicrobial nano-agents: The copper age. ACS Nano 15, 6008–6029. https://doi.org/10.1021/acsnano.0c10756 (2021).
Kornblatt, A. P., Nicoletti, V. G. & Travaglia, A. The neglected role of copper ions in wound healing. J. Inorg. Biochem. 161, 1–8. https://doi.org/10.1016/j.jinorgbio.2016.02.012 (2016).
Maryam Zivari, F. et al. The investigation of the cytotoxicity of copper oxide nanoparticles on peripheral blood mononuclear cells. Nanomed. Res. J. 5, 364–368. https://doi.org/10.22034/nmrj.2020.04.008 (2020).
Isherwood, P. J. M., Abbas, A., Bowers, J. W., Grew, B. & Walls, J. M. Deposition of cupric oxide thin films by spin coating. Mater. Res. Innov. 18, 95–98. https://doi.org/10.1179/1433075x14y.0000000204 (2014).
Nalbant, A., Ertek, Ö. & Okur, İ. Producing CuO and ZnO composite thin films using the spin coating method on microscope glasses. Mater. Sci. Eng. B 178, 368–374. https://doi.org/10.1016/j.mseb.2013.01.010 (2013).
Dulmaa, A., Vrielinck, H., Khelifi, S. & Depla, D. Sputter deposition of copper oxide films. Appl. Surf. Sci. 492, 711–717. https://doi.org/10.1016/j.apsusc.2019.06.263 (2019).
Nayan, N. et al. Sputter deposition of cuprous and cupric oxide thin films monitored by optical emission spectroscopy for gas sensing applications. Procedia Chem. 20, 124–129. https://doi.org/10.1016/j.proche.2016.07.023 (2016).
Shukor, A. H., Alhattab, H. A. & Takano, I. Electrical and optical properties of copper oxide thin films prepared by DC magnetron sputtering. J. Vacuum Sci. Technol. B 38, 012803. https://doi.org/10.1116/1.5131518 (2019).
Rajendiran, S., Meehan, D. & Wagenaars, E. Plasma-enhanced pulsed laser deposition of copper oxide and zinc oxide thin films. AIP Adv. 10, 065323. https://doi.org/10.1063/5.0008938 (2020).
Farhad, S. F. U., Cherns, D., Smith, J. A., Fox, N. A. & Fermín, D. J. Pulsed laser deposition of single phase n- and p-type Cu2O thin films with low resistivity. Mater. Design 193, 108848. https://doi.org/10.1016/j.matdes.2020.108848 (2020).
Kim, S. Y. et al. p-Channel oxide thin film transistors using solution-processed copper oxide. ACS Appl. Mater. Interfaces 5, 2417–2421. https://doi.org/10.1021/am302251s (2013).
Liu, A. et al. In situ one-step synthesis of p-type copper oxide for low-temperature, solution-processed thin-film transistors. J. Mater. Chem. C 5, 2524–2530. https://doi.org/10.1039/C7TC00574A (2017).
Oviroh, P. O., Akbarzadeh, R., Pan, D., Coetzee, R. A. M. & Jen, T.-C. New development of atomic layer deposition: Processes, methods and applications. Sci. Technol. Adv. Mater. 20, 465–496. https://doi.org/10.1080/14686996.2019.1599694 (2019).
Forouzmehr, M. et al. Selective atomic layer deposition on flexible polymeric substrates employing a polyimide adhesive as a physical mask. J. Vac. Sci. Technol. A 39, 012405. https://doi.org/10.1116/6.0000566 (2021).
Taylor, G. et al. Investigation of iridium, ruthenium, rhodium, and palladium binary metal oxide solid solution thin films for implantable neural interfacing applications. Surf. Coat. Technol. 426, 127803. https://doi.org/10.1016/j.surfcoat.2021.127803 (2021).
Taylor, G. et al. Electrochemical enhancement of reactively sputtered rhodium, ruthenium, and iridium oxide thin films for neural modulation, sensing, and recording applications. Electrochim. Acta 394, 139118. https://doi.org/10.1016/j.electacta.2021.139118 (2021).
Page, N. et al. The effect of deposition parameters on microstructure and electrochemical performance of reactively sputtered iridium oxide coatings. Mater. Today Commun. 29, 102967. https://doi.org/10.1016/j.mtcomm.2021.102967 (2021).
Goderecci, S. S. et al. Silver oxide coatings with high silver-ion elution rates and characterization of bactericidal activity. Molecules 22, 1487 (2017).
Chrom, C. L., Renn, L. M. & Caputo, G. A. Characterization and antimicrobial activity of amphiphilic peptide AP3 and derivative sequences. Antibiotics https://doi.org/10.3390/antibiotics8010020 (2019).
Fordham, W. R. et al. Silver as a bactericidal coating for biomedical implants. Surf. Coat. Technol. 253, 52–57. https://doi.org/10.1016/j.surfcoat.2014.05.013 (2014).
Alnes, M. E., Monakhov, E., Fjellvåg, H. & Nilsen, O. Atomic layer deposition of copper oxide using copper(ii) acetylacetonate and ozone. Chem. Vap. Depos. 18, 173–178. https://doi.org/10.1002/cvde.201106959 (2012).
Poulston, S., Parlett, P. M., Stone, P. & Bowker, M. Surface oxidation and reduction of CuO and Cu2O studied using XPS and XAES. Surf. Interface Anal. 24, 811–820. https://doi.org/10.1002/(SICI)1096-9918(199611)24:12%3c811::AID-SIA191%3e3.0.CO;2-Z (1996).
Tahir, D. & Tougaard, S. Electronic and optical properties of Cu, CuO and Cu2O studied by electron spectroscopy. J. Phys. Condens. Matter 24, 175002. https://doi.org/10.1088/0953-8984/24/17/175002 (2012).
Lee, S. W. et al. Improved Cu2O-based solar cells using atomic layer deposition to control the Cu oxidation state at the p–n junction. Adv. Energy Mater. 4, 1301916. https://doi.org/10.1002/aenm.201301916 (2014).
Holzschuh, H. & Suhr, H. Deposition of copper oxide (Cu2O, CuO) thin films at high temperatures by plasma-enhanced CVD. Appl. Phys. A 51, 486–490. https://doi.org/10.1007/bf00324731 (1990).
Hans, M. et al. Role of copper oxides in contact killing of bacteria. Langmuir 29, 16160–16166. https://doi.org/10.1021/la404091z (2013).
Halbus, A. F., Horozov, T. S. & Paunov, V. N. Strongly enhanced antibacterial action of copper oxide nanoparticles with boronic acid surface functionality. ACS Appl. Mater. Interfaces 11, 12232–12243. https://doi.org/10.1021/acsami.8b21862 (2019).
Stohs, S. J. & Bagchi, D. Oxidative mechanisms in the toxicity of metal ions. Free Radical Biol. Med. 18, 321–336. https://doi.org/10.1016/0891-5849(94)00159-H (1995).
Kim, J.-H., Cho, H., Ryu, S.-E. & Choi, M.-U. Effects of metal ions on the activity of protein tyrosine phosphatase VHR: Highly potent and reversible oxidative inactivation by Cu2+ ion. Arch. Biochem. Biophys. 382, 72–80. https://doi.org/10.1006/abbi.2000.1996 (2000).
Lin, Y.-S.E., Vidic, R. D., Stout, J. E., McCartney, C. A. & Yu, V. L. Inactivation of Mycobacterium avium by copper and silver ions. Water Res. 32, 1997–2000. https://doi.org/10.1016/S0043-1354(97)00460-0 (1998).
Dadi, R., Azouani, R., Traore, M., Mielcarek, C. & Kanaev, A. Antibacterial activity of ZnO and CuO nanoparticles against gram positive and gram negative strains. Mater. Sci. Eng. C 104, 109968. https://doi.org/10.1016/j.msec.2019.109968 (2019).
Vaidya, M. Y., McBain, A. J., Butler, J. A., Banks, C. E. & Whitehead, K. A. Antimicrobial efficacy and synergy of metal ions against Enterococcus faecium, Klebsiella pneumoniae and Acinetobacter baumannii in planktonic and biofilm phenotypes. Sci. Rep. 7, 5911. https://doi.org/10.1038/s41598-017-05976-9 (2017).
Tripathy, A., Sen, P., Su, B. & Briscoe, W. H. Natural and bioinspired nanostructured bactericidal surfaces. Adv. Colloid Interface Sci. 248, 85–104. https://doi.org/10.1016/j.cis.2017.07.030 (2017).
Jiang, Y. et al. Hydrophilic nanoparticles that kill bacteria while sparing mammalian cells reveal the antibiotic role of nanostructures. Nat. Commun. 13, 197. https://doi.org/10.1038/s41467-021-27193-9 (2022).
Acknowledgements
The Veeco Fiji G2 ALD equipment was acquired through NSF CBET MRI 1428500 to support the ALD coating experiments by Khosla and Feng. Caputo and Ammerman were supported in part by NSF CHE MRI 2018320. This work was carried out in part at the Singh Center for Nanotechnology, part of the National Nanotechnology Coordinated Infrastructure Program, which is supported by the National Science Foundation grant NNCI-2025608. Lastly, illuminating discussions with Dr. Greg Taylor from Lawrence Livermore National Laboratory is greatly acknowledged.
Author information
Authors and Affiliations
Contributions
All authors contributed to conducting experiments, data analysis and manuscript preparation.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Khosla, H., Seche, W., Ammerman, D. et al. Development of antibacterial neural stimulation electrodes via hierarchical surface restructuring and atomic layer deposition. Sci Rep 13, 19778 (2023). https://doi.org/10.1038/s41598-023-47256-9