Abstract
Percutaneous drains have provided a minimally invasive way to treat a wide range of disorders from abscess drainage to enteral feeding solutions to treating hydronephrosis. These drains suffer from a high rate of dislodgement of up to 30% resulting in emergency room visits, repeat hospitalizations, and catheter repositioning/replacement procedures, which incur significant morbidity and mortality. Using ex vivo and in vivo models, a force body diagram was utilized to determine the forces experienced by a drainage catheter during dislodgement events, and the individual components which contribute to drainage catheter securement were empirically collected. Prototypes of a skin level catheter securement and valved quick release system were then developed. The system was inspired by capstans used in boating for increasing friction of a line around a central spool and quick release mechanisms used in electronics such as the Apple MagSafe computer charger. The device was tested in a porcine suprapubic model, which demonstrated the effectiveness of the device to prevent drain dislodgement. The prototype demonstrated that the miniaturized versions of technologies used in boating and electronics industries were able to meet the needs of preventing dislodgement of patient drainage catheters.
Similar content being viewed by others
Introduction
Minimally invasive percutaneous drainage catheters have revolutionized the treatment of a wide variety of diseases, ranging from infectious abdominal complications, such as perforated appendicitis or diverticulitis, to decompression of obstructed urinary tracts from kidney stones to providing enteral access for patients with neurodegenerative disorders. The use of percutaneous drainage catheters has converted previously lethal illnesses treated with morbidly invasive procedures to chronic, palliative, or often curable disease states by providing a small outlet or inlet to previously inaccessible regions of the body. Unfortunately, a common complication plagued by all percutaneous drainage catheters is accidental catheter dislodgement (ACD)1,2,3.
Percutaneous drainage catheters may either be partially or completely dislodged from the patient. This typically takes place when a pull force is applied to the external component of the device. This may occur during sleep, normal daily activities, or self-removal in the case of patients with altered mental status (Fig. 1)1,2,4,5 Partial catheter dislodgement may result in decreased drainage catheter output while the dislodgement may not be as externally apparent to the patient or healthcare provider (Fig. 1C).
The risks associated with ACD depend on the type of catheter dislodged, but are typically related to tissue trauma during the dislodgement event, interruption in necessary device function, and risks related to reinsertion3 Each of these complications are exacerbated by disparities in healthcare access and patient distance from requisite specialty care. In addition to these physical harms, management of patients who experience ACD poses significant financial burdens. ACD is associated with costs related to emergency department utilization, radiologic imaging, specialty procedural care, inpatient admission, and prolonged length of hospital stay6,7
The current strategies for catheter fixation are inadequate8 and there is a paucity of data describing the forces encountered by a drainage catheter during normal use. A force-body diagram was utilized in order to understand the individual elements of a standardly secured drainage catheter (Fig. 2). The standard of care for drain securement relies in part on two primary elements: internal and external fixation components. Internal fixation components are intrinsic to the type of catheter placed, which include an end pigtail—with resistive force provided by the tensile strength of an inner securement string and by the elastic modulus of the tube—or from a retention balloon with resistance based on the size and pressure of the balloon. External fixation is operator dependent, with providers typically using a drain stitch or “sandal” stitch, which secures the catheter to the patient’s skin. Beyond standard of care, multiple groups have developed skin securement devices, such as the StatLock, which clamp around the drainage catheter and utilize an adhesive component to stick to the patient’s skin, adding an in-line linear resistive component. Further, some groups have developed a break-away device at the drainage catheter hub that allows the catheter to linearly decouple when a dislodging force is applied.
Current catheter securement modalities have several drawbacks. The most commonly used strategy of a drain suture induces tissue trauma and does not prevent the transmission of accidental pull forces to the internal portion of drain catheters. Other proprietary external catheter fixation mechanisms improve on patient comfort but sacrifice the security of both external and internal positioning, often by allowing some degree of sliding of the catheter tubing. Several groups have described the use of a linear breakaway device to reduce dislodgement rates7. However, the lack of a valve in currently available breakaway devices results in bodily fluid staining clothes or patients, and makes thoracic applications implausible given the risk of pneumothorax. The ideal drainage securement device may have elements that reduce the amount of extended tubing hanging from the patient, increase the friction safety factor when a catheter is pulled, and incorporate a valve-containing breakaway device that decouples following application of a multidirectional dislodgement force.
In searching for solutions, the authors took inspiration from other industries. Increasing the friction of a line in a contained space has long been solved in the boating industry through the use of capstans where winding of a non-elastic line around a central column dramatically increases its resistive frictional force with a relatively small holding force (Fig. 3D). Quick release mechanisms are employed across daily life including gas station line safety break mechanisms or the MagSafe power cord (Apple Inc., Cupertino, CA), which disconnects from the laptop computer when the charging cord is pulled (Fig. 3C).
Utilizing these inspirations, in this study, ex vivo and in vivo models were used to determine the forces that characterize ACD. The authors analyzed the forces necessary to dislodge a drainage catheter and developed a novel, 3D-printed securement device specifically designed to limit the possibility of ACD.
Materials and methods
Device development and characterization
Computer-aided design (CAD) software (Fusion 360, Autodesk Inc., San Rafael, CA), a Form 3 3D printer (Formlabs Inc., Somerville, MA), and a MakerBot Replicator 3D printer (MakerBot, Brooklyn, NY) were used in the iterative design and production of the device. We based development on a design thinking methodology and research infrastructure as previously described by our group9,10,11. The proposed device has three main components: (1) a skin-level catheter securement device which manages the external length of drainage catheter tubing and increases the frictional resistive force, (2) an adhesive component which affixes the skin level catheter securement device to the patient’s skin, and (3) a valved Luer lock compatible quick release mechanism that is interposed between the drainage catheter hub and drainage bag tubing (Fig. 3a-b, supplemental video 1), which detaches when a sufficient pull force is applied to the external catheter tubing and ensures the critical, internal drainage catheter component remains in place (Fig. 4).
Using 3D printing, the proposed skin-level catheter securement and quick-release device were generated. The components were fabricated using a combination of thermoplastic polyurethane (TPU), polylactic acid (PLA), and photopolymer resin. The flexible TPU base of the catheter securement component enabled flush skin-level apposition of the device and secured drainage catheters with a variable external length ranging from 4 to 30 cm near the site of percutaneous entry. The skin level catheter securement device has a surface area of 32.5 cm2. The drainage catheter is coiled around the central spool of the securement device and keeps the drainage catheter tubing within 6 mm of the patient’s skin. The skin-level catheter securement device was hand sutured to the adhesive component of the Hollister ostomy appliance, which was chosen for its adhesion, conformability, porosity, and fluid resistance. The quick-release component, composed of two halves, is designed to be connected to any Luer fitting at the terminal end of the drainage catheter, and the second half is designed to connect to the Luer fitting of any stopcock or drainage bag tubing. Neodymium ring magnets were utilized as the coupling force between the two halves of the quick release. A stop valve is incorporated into the quick release device to prevent fluid leakage when the quick release components are uncoupled during a dislodgement event.
Force body diagram analysis
Using the force body diagram (Fig. 2), important contributing forces to ACD in the current standard of care were identified, including the force required to break a standard 2–0 polypropylene drain suture (Prolene, Ethicon, Inc., Raritan, NJ), the force required to break the pigtail-forming string within the drainage catheter, and the force required to remove the adhesive component of an ostomy appliance (Hollister Inc., Libertyville, IL), an appropriate surrogate for a non-woven skin adhesive. Appropriate force analysis equations were generated from this analysis. We then added the Capstan component to mathematically model the impact of our proposed skin-level securement device.
Ex Vivo component force analysis of drain dislodgement
The component forces noted above which resist drainage catheter dislodgement were determined empirically. Additionally, the force required to remove a drainage catheter secured by a linear fixation device, the force required to decouple the proposed quick release component, and the resistive force of the proposed skin level catheter securement device were also tested. Forces were applied with a McMaster-Carr Tension and Compression Digital Force Gauge (McMaster-Carr, Princeton, NJ) by one operator until device failure and data was exported to Excel (Microsoft, Redmond, WA). At minimum, three trials from each component were obtained. By then employing our mathematical model with the necessary variables established empirically, the idealized peak force that a drainage catheter can resist before dislodgement with the presently reported skin-level securement device and the proposed skin-level securement was calculated. The idealized peak force was compared to the experimentally derived force required to fully remove a drainage catheter to validate whether the capstan equation applied to the proposed device in practice.
In vivo assessment of drain dislodgement
The study used three female Yorkshire pigs, with weights between 40 and 50 kg, for testing of the device. Animals remained under general anesthesia for the duration of the experiment and were euthanized at the end of the experiment with inhaled Isoflurane 1–2% followed by 100 mg/kg Phenobarbital IV. All experimental protocols for this portion of the study were approved by the Institutional Animal Care and Use Committee (IACUC) at the Massachusetts General Hospital (MGH) under protocol #2021N000227, and all methods were completed in accordance with IACUC guidelines and regulations. Methods and results are reported in accordance with ARRIVE guidelines. 10.2 Fr Dawson Mueller (DM) drainage catheters (Cook Medical, Bloomington, IN) were aseptically inserted into the porcine bladder using standard trocar technique and ultrasound guidance. Drainage catheter pigtails were formed in the standard fashion. Drainage catheters were secured in three ways: without any securement, with a standard 2–0 polypropylene drain suture, and with the described skin-level securement adhered to the animal’s skin. These trials were conducted in sequence on the same pigs to reduce confounding. Initial catheter position was confirmed either fluoroscopically or with computed tomography (CT). A tensile force great enough to fully dislodge the catheters was applied with a digital force gauge and data were exported to Excel. Initial and post-pull catheter position was determined fluoroscopically.
Results
Force body diagram analysis
The forces on a drainage catheter can be decomposed into two directions, x and y, as a sum of the force on the drain itself (Fd), tension on the suture (Ts), tension on the pigtail (Tp), and any linear adhesive normal force (Nl adhesive * u) in the x or y direction. (Fig. 2).
At stability (F = ma = 0),
In the x-direction, we find the system can be modeled as:
In the y-direction, similarly, we find the system can be modeled as:
The simplified model has several limiting assumptions. Forces on the drain followed a stepwise break function, i.e., the adhesive is in line with the suture/pig-tail drain, and that each of these subcomponents have a certain breakpoint at which they no longer contribute. Force on the drain at any point is dynamic, both externally and internally, resulting in each force being a function of time.
The addition of a capstan component in the skin level catheter securement device changes these formulas as follows with the frictional force between catheter and securement device now described with an exponential relationship \(e^{\mu \alpha }\), previously \(\left( {Nl_{adhesivey} *\mu_{adhesive} } \right)\).
At stability (F = ma = 0),
In the x-direction, we find the system can be modeled as:
In the y-direction, similarly, we find the system can be modeled as:
Functionally, applying the capstan equation to the drain force body diagram analysis provides a multiplicative effect of the adhesive towards the sum force of the Ts (suture) and pigtail internal string (Tp). This predicts a significantly higher resistive force for a coiled adhesive design over a standard linear one.
Ex vivo component force analysis of drain dislodgement
In order to apply the simulation, we empirically measured the above specific forces, Ts, Tp, and the linear adhesive normal force of an ostomy appliance. The 2–0 polypropylene drainage suture failed at 32.0 ± 5.9 N (n = 4) (Fig. 5a), the pigtail failed at 16.5 ± 3.6 N (n = 3) (Fig. 5b), and the ostomy appliance adhesive failed at 64.3 ± 11.7 N (n = 4) for parallel to skin pull forces (Fig. 5d). Assuming a linear system, the theoretical maximum force resisted by the standard of care is the sum of the drainage catheter suture and pigtail forces: 48.8 ± 7.1 N. We then analyzed the ex vivo predicted components of using a coiled instead of linear friction model. With one revolution around the spool, it was able to resist 78.5 ± 8.4 N (n = 5) forces (Fig. 5f). Conversely, a linear fixation system resisted 19.1 ± 1.8 N (n = 3) prior to the drain slipping out through the linear fixation system (the adhesive itself did not fail) (Fig. 5e). The quick release mechanism decoupled at 17.5 ± 2.7 N (n = 4) (Fig. 5c), providing an adequate safety margin to dislodgement. An average drainage bag can be expected to hold between 500 to 1000 mL, or roughly 5–10 N of expected force when hanging by gravity. An ideal quick release mechanism was thus determined to decouple 15–20 N allowing for a ~ 2 × safety factor to inappropriate decoupling during daily use, while also allowing for a 3–5 × safety factor compared to the force required to break either a catheter drain stitch or remove an adhesive device.
In vivo assessment of drain dislodgement
In the in-vivo model, without any catheter securement, the drainage catheter was displaced with unregistered forces (< 1 N). A catheter secured with a drain stitch and internal pigtail was dislodged with a peak force of 47.7 ± 4.6 N (n = 4). A drainage catheter fully secured by the proposed device is pictured in Fig. 6a–c (sans cap for illustrative purposes). When a pull force was applied, the catheter wound more tightly around the device’s central column, as slack was removed with tension. Eventually, the force became large enough for the quick-release unit to disconnect. The critical internal pigtail component of the drainage catheter remained in unchanged position throughout the pull event (Fig. 6a–c). In contrast, the pigtail was observed to have displaced under fluoroscopy when secured with only a drain suture (Fig. 6d–f). Overall, the proposed securement device detached at 15.1 ± 0.7 N (n = 3) and resulted in no migrated or dislodged catheters (Fig. 7). In the suprapubic model, urine continued to drain from the bladder while the catheter was coiled within the skin level securement device, which was engineered to prevent catheter kinking.
Discussion
In this proof-of-concept study, we have taken steps to better understand the conditions under which drainage catheters dislodge. We used this data to inspire a new approach to limiting drain dislodgement, which was successful when tested in a swine model.
With over 3.2 million percutaneous drainage catheters placed in the US annually, ACD represents a significant source for patient morbidity and mortality as well as a burden on healthcare systems. Prior reports illustrate an ACD incidence of 4.5%–12.6% for abdominal drains3,12,13, 10.2%-12.8% for percutaneous gastrostomy catheters2,7,14, and 6.6%-9.2% for thoracic drains6,15,16, with a single prospective RCT subanalysis reporting rates of ACD for thoracic drains as high as 28%-42%17. Including the average cost of ambulance transportation, adult emergency room visits, and non-invasive imaging, the U.S. healthcare system conservatively estimated to spend over $2.5 billion on the management of dislodged percutaneous drainage catheters annually. This figure does not include the significant cost associated with repositioning and replacement procedures or potential hospital admission. Beyond the costs of replacement, significant morbidity is caused by ACD; dislodgement of percutaneous gastrostomy catheters, especially prior to formation of a mature gastrocutaneous fistula, may lead to peritonitis requiring surgical consultation4,7 and dislodgement of a percutaneous chest or abdominal drainage catheter may result in abscess/pneumothorax reaccumulation, hemodynamic compromise, or death3. This study establishes the forces that lead to dislodgement of drainage catheters secured with standard of care and proposes a novel catheter securement and quick-release system that better mitigates this ubiquitous complication.
While several external and internal fixation strategies exist, a method to sufficiently maintain drainage catheter position during a dislodgement event has not gained widespread use. A sufficient force will always overcome a catheter fixation device. This necessitates the use of a quick-release component to interrupt dislodgement forces from being transmitted to the critical internal component of a catheter. Applying lessons from the electronics and fuel industries to develop a valved approach builds on the lessons by other linear non-valved groups that have shown benefit to reduce dislodgement. The proposed valved quick release allows a solution for even externally short drainage catheters, where external length may be less than 4 cm.
Current fixation modalities leave a significant length (5–15 cm) of external drainage catheter unmanaged, thus increasing the risk of partial or complete dislodgement. This risk would be maintained even with the use of a quick-release device as pull forces proximal to the quick-release would still result in dislodgement. Given the enormity of ACD with these devices on the market, a new approach that secures the entire external drain close to the skin and allows for the interruption of dislodgement forces is imperative. Contrary to other fixation strategies, the herein described securement device delivers a low-profile solution to this problem. By employing inspiration from the sailing industry in applying and miniaturizing the capstan equation, we aim to mitigate external forces on the drain and pigtail while managing the entire external drain length. The empirical results of a spiral coil demonstrated a fourfold increase in drain safety compared to a linear adhesive model. This coiled design further has the advantage of maintaining drain integrity without kinking the drain by bending or compressing it. Numerous iterations of prototypes and calibration of the quick release mechanism enabled us to successfully achieve our goal of withstanding normal daily forces, defined as the 5-10N required to hold a 0.5-1L drainage bag of fluid, while ensuring consistent catheter position. Additionally, the design of both the skin level securement and quick release components have demonstrated a low learning curve throughout initial ex vivo tests with patients, nursing staff, and clinicians. Following one demonstration, patients, nursing staff, and clinicians are facile at implementing this device. User feedback has been central to the iterative development of the device.
In this study, there are a number of limitations. The differences in cutaneous and subcutaneous tissue between swine and humans provide different resistances to the dislodgement of catheters. Furthermore, none of the experimental conditions fully embody the breadth of catheter dislodgement situations. The experiments represent a comparative moment in time between solutions rather than the longitudinal conditions that would be experienced by patients. Future studies are needed to demonstrate longitudinal effectiveness and safety in patients. Additionally, the materials and production techniques used were selected based on cost-effectiveness and availability, and future iterations will employ more tailored adhesives, more flexible and conforming materials, and better tolerances.
Within these parameters, we have demonstrated that a three-component device providing skin-level securement adhered to a patient’s skin and a quick-release detachment mechanism is a compelling solution to ubiquitously dislodged catheters. Further, the holding force amplification and drain management achieved by the spool design suggest more devices should consider utilizing the capstan principle when securing medical interventions. While we are initially using drainage catheters, this strategy can be translated to many percutaneous devices, including vascular catheters, whose dislodgement may result in bleeding or interruptions in critical medications. Prevention of dislodged percutaneous catheters is of significant benefit to patients.
Data availability
The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.
References
Moureau, N. Impact and safety associated with accidental dislodgement of vascular access devices: A survey of professions, settings, and devices. J. Assoc. Vasc. Access 23, 203–215 (2018).
Feil, M. Dislodged gastrostomy tubes: Preventing a potentially. Fatal Comp. 14, 9 (2017).
Lorente, L., Huidobro, M. S., Martín, M. M., Jiménez, A. & Mora, M. L. Accidental catheter removal in critically ill patients: A prospective and observational study. Crit. Care 8, R229 (2004).
Huang, S., Engstrom, B., Lungren, M. & Kim, C. Management of dysfunctional catheters and tubes inserted by interventional radiology. Semin. Interv. Radiol. 32, 067–077 (2015).
Collares, F. B., Faintuch, S., Kim, S. K. & Rabkin, D. J. Reinsertion of accidentally dislodged catheters through the original track: What is the likelihood of success?. J. Vasc. Interv. Radiol. 21, 861–864 (2010).
Asciak, R. et al. Chest drain fall-out rate according to suturing practices: A retrospective direct comparison. Respiration 96, 48–51 (2018).
Rosenberger, L. H. et al. reducing accidental dislodgement of the percutaneous endoscopic gastrostomy: a prospective trial of the “safetybreak” device. Surg. Innov. 23, 62–69 (2016).
Macmillan, T. et al. SecurAcath for securing peripherally inserted central catheters: A nice medical technology guidance. Appl. Health Econ. Health Policy. 16, 779–791 (2018).
Succi, M. D. et al. Initial experience integrating a hands-on innovation curriculum into a radiology residency program and department. J. Am. Coll. Radiol. 17, 1329–1333 (2020).
Succi, M. D., Uppot, R. N., Gee, M. S., McLoud, T. C. & Brink, J. A. Medically engineered solutions in health care: A technology incubator and design-thinking curriculum for radiology trainees. J. Am. Coll. Radiol. 15, 892–896 (2018).
Som, A., Lang, M., Di Capua, J., Chonde, D. B. & Cochran, R. L. Resident-led medical student radiology research interest group: An engine for recruitment, research, and mentoring—radiology in training. Radiology 300, E290–E292 (2021).
Gupta, S., Gulati, M., Shankar, K. U., Rungta, U. & Suri, S. Percutaneous nephrostomy with real-time sonographic guidance. 4.
Carrafiello, G. et al. Complications of percutaneous nephrostomy in the treatment of malignant ureteral obstructions: Single–centre review. Radiol. Med. (Torino) 111, 562–571 (2006).
Galazzi, A. et al. Accidental removal of devices in intensive care unit: An eight-year observational study. Intensive Crit. Care Nurs. 54, 34–38 (2019).
Hooper, C. E., Welham, S. A. & Maskell, N. A. Pleural procedures and patient safety: A national BTS audit of practice: Table 1. Thorax 70, 189–191 (2015).
Hooper, C., Maskell, N., & on behalf of the BTS audit team. British thoracic society national pleural procedures audit 2010.Thorax 66, 636–637 (2011).
Rahman, N. M. et al. Effect of opioids vs NSAIDs and larger vs smaller chest tube size on pain control and pleurodesis efficacy among patients with malignant pleural effusion: The TIME1 randomized clinical trial. JAMA 314, 2641 (2015).
Acknowledgements
The photograph in Figure 3c was produced by Ashley Pomeroy and accessed from wikimedia commons: Own work, CC BY-SA 4.0, https://commons.wikimedia.org/w/index.php?curid=64776055. Figure 3d was produced by George Hodan and released under a CC0 Public Domain license, accessed from: https://www.publicdomainpictures.net/es/view-image.php?image=434883&picture=cuerda-enrollada-en-un-cabrestante-de-ya. We would like to acknowledge the feedback of the Departments of Surgery, specifically Dr. Cristina Ferrone, and the Department of Radiology. Animal portion of this study was supported by the MGB MESH Incubator, Schlaeger Research Fellowship Grant, and the RSNA resident research award (AS). We’d also like to acknowledge the mentorship of Dr. Raul Uppot in this work.
Author information
Authors and Affiliations
Contributions
M.R., J.D., J.K., and A.S. conceptualized and wrote the manuscript main text. M.R., J.D., J.K., A.A., H.B., G.L., A.S., Z.I., R.A., P.S., S.Z., D.D., S.K., M.S., and A.T. contributed to the review and revision of the manuscript. Z.I., R.A., P.S., S.Z., D.D., S.K., M.S. provided advice and iterative support. A.S., and A.T. additionally served as joint supervisors.
Corresponding authors
Ethics declarations
Competing interests
Dr. Di Capua and Dr. Som are listed inventors on a patent pending application (#63306932) which describes the securement device evaluated by this manuscript. As the intellectual property was created as part of an employment role, the employing institution, Massachusetts General Hospital, holds full ownership and rights to the intellectual property, in accordance with local legislation. The authors disclose no other financial or non-financial competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Supplementary Video 1.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Russo, M., Di Capua, J., Anlage, A. et al. Preventing inadvertent drain removal using a novel catheter securement device. Sci Rep 13, 16130 (2023). https://doi.org/10.1038/s41598-023-37850-2
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-023-37850-2
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.