Introduction

Dual-energy computed tomography (DECT) has become a well-established and widely used technique in modern medical diagnostics. Image acquisition at two energy levels—typically 70–80 kV and 140–150 kV—takes advantage of the different physical properties of low and high energetic photons to unravel the elemental composition of materials, which would not be resolvable based on CT numbers of voxels defined by single-energy CT (SECT) alone1. The combination of two attenuation measurements enhanced the power to delineate different tissues and paved the way for novel applications, e.g. in cardiovascular, thoracic, bone and abdominal imaging2.

Next to diagnostic benefits, DECT offers the chance of significant dose reduction, in particular by abolishing the obligation of a complete unenhanced scan due to reconstruction of virtual unenhanced images3,4,5,6,7. Nevertheless, comparative risk assessment of SECT and DECT should not only focus on radiation dose and dose rate, but also has to consider differences in biological effectiveness (BE) associated with varying photon energies8,9,10. In case of DECT, BE of the contributing energy spectra may presumably account only for a small fraction of individual risk in comparison to dose reduction opportunities by DECT in clinical routine. Yet, it might be a pivotal point of criticism as to risk assessment by means of the linear no-threshold model which is routinely applied in clinical risk assessment.

The importance of higher BE associated with exposure to low-energy photons has already been addressed in the context of mammography (X-ray tube voltage usually 25–35 kV). Several in vitro experiments comparing low-energy mammography X-rays with higher energetic radiation qualities showed an increase in dicentric chromosomes11, neoplastic cell transformation12,13,14,15, enhanced mutagenicity16 and micronuclei induction17. This raises the question whether the low-energy portion in X-ray spectra of CT might cause measurable differences in biological effectiveness between SE- und DECT which could reduce the apparent benefit of DECT. Against this background there is need for a profound knowledge-base that could assist radiologists in finding the optimal risk–benefit balance for DECT. To the best of our knowledge only one study so far has investigated biological effects of single- versus dual-energy scans and found no significant differences after pulmonary CT angiography comparing single- and dual-energy CT protocols based on the measurement of DNA double strand breaks (DSB) by means of the γH2AX-assay18.

In this study we aimed to gain further insights into the varying biological response possibly associated with very low dose, low-energy X-ray exposure during DECT. In the controlled setting of an ex vivo exposure of peripheral blood cells to adapted modern SECT and DECT scan protocols, we investigated differences in gene expression and the extent of DNA DSBs between the two CT techniques by means of whole transcriptome sequencing and 53BP1/γH2AX co-immunostaining19, respectively.

Results

Transcriptional response to irradiation with varying tube voltages

The overall effect of CT exposure on gene expression was only moderate. Unsupervised hierarchical clustering based on sample-to-sample distances indicated the clear dominance of inter-individual differences over irradiation associated effects. As depicted in Fig. 1 all samples of each individual clustered together, irrespective of whether they had been irradiated or not. In line with this observation, only few genes were found deregulated with respect to corresponding sham irradiated controls (18–109 for 80 kV, 9–67 for 150 kV and 10–21 for dual-energy acquisition setting; Supplementary Table S1). Among the significantly deregulated genes, AEN (Apoptosis Enhancing Nuclease) was shared between all replicates, probands and CT protocols (Fig. 2).

Figure 1
figure 1

Unsupervised hierarchical clustering based on sample-to-sample distances of all 36 RNA sequencing experiments. Naming of samples indicates treatment (Sham, 80 kV, 150 kV and DE) and proband (1, 2, 3). Probands have additionally been highlighted in yellow (proband 1), green (proband 2) and red (proband 3) to the right and at the top of the heatmap. The order of samples in combination with the dendrogram at the top or to the left of the heatmap denote the clusters of samples arising as a result of the hierarchical clustering process. Given the dominance of inter-individual differences of gene expression profiles over CT-induced effects, all samples derived from the respective probands group together in distinct clusters, independent of whether they have been irradiated or not. Within the heatmap sample-to-sample distance based on log2-transformed normalized read count data is displayed by a blue color gradient with dark blue indicating a high degree of similarity.

Figure 2
figure 2

Gene expression plots for AEN for 80 kV single-energy CT (SECT), 150 kV SECT and dual-energy CT (DE) irradiation: Each plot depicts gene counts obtained from RNA sequencing data of the respective proband (G1-3) with post-irradiation gene count displayed to the left and corresponding sham-irradiated gene counts displayed to the right. Colored lines (green, blue and red) indicate the three replicate experiments.

A direct comparison of 80 kV, 150 kV and DECT experiments did not reveal any common gene expression signature characteristic for one of the three tube voltages. In total this direct comparison identified 48, 49 and 45 significantly differentially expressed genes. None of them was shared among all probands (Fig. 3).

Figure 3
figure 3

Venn diagram summarizing the number of differentially expressed genes identified by direct comparison of 80 kV single-energy CT (SECT), 150 kV SECT and dual-energy CT (DE), respectively. The corresponding gene lists can be found in Supplementary Table S2.

A second round of analysis with less stringent criteria (see “Materials and methods”) added BAX, DDB2, EDA2R and FDXR to the list of CT associated candidate genes. Upregulation relative to corresponding sham irradiated controls was verified for all of these five CT associated genes by qRT-PCR (Supplementary Fig. S1). As the good concordance of sequencing and qRT-PCR data suggested a high accuracy of fold-change quantification, we further investigated the dependency of fold-change magnitude on the type of CT exposure. For this purpose we calculated the average fold-change for each gene and treatment (i.e. the average over three individuals and their three replicates). The resulting 15 average fold changes (five genes x three exposure types) were categorized according to whether they represent the lowest, middle or highest fold-change of each gene. The frequency distribution within these categories was non-random (Chi square p = 0.0004), with all fold-changes induced by DECT falling in the lowest fold-change category. Apart from AEN, highest fold-changes for each gene were associated with 80 kV SECT, whereas four out of five middle fold-changes were induced by 150 kV SECT (Fig. 4). The same trend was found when considering probands individually (Chi Square p = 0.0342).

Figure 4
figure 4

Frequency distribution of mean fold changes on a linear scale for the genes EDA2R, FDXR, AEN, DDB2 and BAX. Mean fold changes for each gene and treatment averaged over all probands and replicates were categorized according to whether they represent the lowest, middle or highest fold change of each gene. Exposure groups are indicated by red (80 kV), green (150 kV) and yellow (dual-energy (DE)) coloring of boxes. Mean fold changes given in the boxes refer to qRT-PCR results based on nine measurements (3 probands × 3 technical replicates per proband). See “Results” section for details.

Within the sets of up-regulated genes, functional enrichment analysis identified shared overrepresentation of transcriptional regulatory processes (p ≤ 1.32 × 10−4) in either single-energy level of two probands and of intrinsic apoptotic signaling pathways (p ≤ 5.95 × 10–5) partly linked to the WNT signaling pathway (p ≤ 4.16 × 10−5) in 80 kV and dual-energy setting of two probands.

DNA double strand break frequency after SE- and DECT exposure

DNA double strand breaks (DSBs) were estimated by counting γ-H2AX + 53BP1 colocalizing foci in PBMCs19,20 of three different probands in triplicates (Fig. 5). The mean of the average number of baseline DSB foci per cell (FPC ± SD) detected in the control samples was 0.27 ± 0.11. For all three radiation modalities the CT-induced increase in average DSB FPC values over that of the sham irradiated samples was highly significant (p < 0.001). The average FPC values between the three irradiation modalities was 0.78 ± 0.077 for 80 kV, 0.78 ± 0.069 for 150 kV and 0.73 ± 0.073 for DECT, an insignificant difference (p > 0.15) between DECT and each of the SECT modalities. These data show that there is no difference for the average DSB numbers induced between the two single-energy CT modalities (p = 1) and between these two and the DE mode (p > 0.15). Hence, the different CT energies alone or in combination do not induce different amounts of DSBs indicating a similar biological effect of the different CT settings.

Figure 5
figure 5

Average number of DNA double strand break (DSB) foci observed without (Sham) and after 80 kV, 150 kV and dual-energy CT (DE) irradiation. Each box represents the interquartile range (25th to 75th percentile) of average DSBs per cell observed in three probands and their three replicates. Whiskers indicate the minimum and maximum. The median is shown as horizontal line within the boxes. Relative to control all CT applications led to a highly significant increase in DSB (p < 0.001). Relative to SECT the average number of foci per cell appears slightly lower in samples exposed to DECT, but this difference did not reach significance (p > 0.15).

Scrutinizing the frequency distribution of DSBs analogous to the procedure described in the context of gene expression fold changes did not reveal any significant dependency of DSB frequency on the type of CT exposure.

Discussion

Low energy photons emitted during DECT may inherit the risk of increased biological effectiveness, a fact that has to be taken into account when weighing up benefits and potential risks of DECT diagnostics. In this study, we investigated the consequences of ex vivo exposure of peripheral blood inside a X-ray phantom to 80 kV SECT, 150 kV SECT and DECT irradiation with regard to two biological endpoints: gene expression and DNA double strand break frequency.

Irrespective of the CT protocol applied, CT-associated effects on gene regulation were only moderate when compared to inter-individual differences of gene expression profiles. Nevertheless, we observed shared upregulation of five genes in CT-irradiated samples, namely, AEN, DDB2, FDXR, BAX and EDA2R. These five genes have already emerged as potential radiation biomarkers in studies with exposures to doses ≥ 50 mGy21 or ≥ 0.5 Gy22,23. Moreover, they largely overlap with the candidate genes defined in our previous CT study using 150 kV SECT alone24. Altogether, these findings underscore the potential of these genes as biomarkers for irradiation, whose upregulation can even be triggered by very low doses. Noteworthy, all of these five genes are known targets of the p53 tumor suppressor with impact on cell cycle control, apoptosis and DNA damage repair, activated by cellular damage25,26,27.

While we have found a common set of genes distinguishing irradiated and non-irradiated samples, it was not possible to further specify whether a sample was irradiated with 80 kV SECT, 150 kV SECT or DECT. This suggests that in the experimental settings used in this study, involving exposition to very low X-ray doses in conjunction with varying X-ray spectra, the fraction of low energy photons emitted during DECT diagnostics has no detectable impact on gene regulation. Yet, further studies are needed to verify that this statement holds true for other points of time post irradiation, higher dose values and in vivo studies. One limitation especially in the context of the low doses applied is that it cannot be ruled out that the polyenergetic spectra of X-ray irradiation or the use of a phantom may have masked subtle effects. Unfortunately, this uncertainty cannot be addressed in the current study as local dose and its variability is unknown and diverging spectra may affect the ratio of local to effective dose. Moreover, CTDIvol and DLP values may vary by a systematic deviation of ± 10% as specified by the manufacturer (System owner manual SOMATOM Force; Chapter 1.1.8 “Accuracy of Dose Indications”) along with additional differences depending on the applied spectrum.

Results at the level of gene regulation were mirrored by the outcome of DNA damage assessment based on the focal colocalization of 53BP1 and γ-H2AX DSB damage markers. Estimation of DSB frequency by means of these two dsDNA damage markers is a well-established approach19,28,29. In line with previous in vivo studies18,20 we clearly observed the induction of DSBs by CT irradiation in significant excess over base line rate (p < 0.001), allowing the clear separation of irradiated and non-irradiated samples, which agrees with other SECT DSB studies20,30,31,32. With regard to single- and dual-energy CT irradiation modalities, the DSBs numbers induced were similar to each other (Fig. 4). These results show that SECT or DECT elicit a similar amount of dsDNA damage, suggesting a comparable biological effect for this endpoint. Future studies of other endpoints like e.g., chromosome aberrations or micronuclei33 shall provide further insights into the radio-response of PBMCs to SECT or DECT.

Clearly, most of the potential limitations mentioned in the context of gene expression may also be valid for DNA double strand break analysis. However, unlike for gene expression the optimal time slot for DSB analysis is predefined by DNA repair kinetics, mitigating this matter of concern.

Altogether, the analysis of both gene regulation and DNA DSBs does not provide evidence for an increased biological effect of DECT as compared to 80 kV and 150 kV SECT. On the contrary, in light of the observation that the extent of upregulation of all genes tested by qRT-PCR is lower in DECT relative to 150 kV SECT and in particular to 80 kV SECT one may speculate that e.g. the modulation of radiation quality by the use of tin filtration, routinely added to the high-kilovolt tube during DECT exposures, could result in reduced biological effectiveness. Yet, given the rather small differences in gene upregulation observed in this study, this aspect has to be systematically addressed by future studies.

In conclusion, the results of this study do not suggest any increased biological effectiveness associated with the low energy spectrum of DECT. Yet, despite decreasing dose burden caused by CT scans, the genotoxic effect of CT irradiation still subsists and should remain a crucial aspect of risk assessment before performing CT diagnostics.

Materials and methods

Ex vivo CT exposure and dose estimation

Ethical approval for this study was obtained from the responsible ethical review committee of the Medical Association of Rhineland Palatinate, Germany (reference number: 837.084.17(10918)). All experiments were performed in accordance with relevant named guidelines and regulations. Written informed consent was obtained from all probands. Peripheral blood samples were collected from three healthy male donors (age 30, 32 and 53 years) in EDTA-Monovettes (5 ml; Sarstedt, Nümbrecht, Germany) in triplicates with matching sham control samples. Both, single-energy CT and dual-energy CT were performed utilizing adapted lumbar spine protocols in a 384 (2 × 192)—channel multi-detector row dual source CT scanner (SOMATOM Force, Siemens Medical Solutions, Forchheim, Germany) with blood samples placed in a lumbar spine phantom at room temperature. Acquisition parameters were set to X-ray tube voltages of 80 kV and 150 kV single source single-energy, respectively, and dual source dual-energy (80 kV/150 kV (with Sn filter)). To obtain identical dose values the current was manually set to 950 and 160 mA in SECT and 440 and 220 mA in DECT for 80 kV and 150 kV, respectively. Pitch was 0.6 and rotation time was 0.5 s at a scanning length of 367 mm for all protocols. The applied setting resulted in a Volume CT Dose Index (CTDIvol) of 15.79–18.26 mGy, in a dose length product (DLP) of 606.7–613.8 mGy*cm and in an administered effective dose of 7.0 ± 0.08 mSv as calculated for better context in accordance to ICRP 103 report by means of Monte Carlo model-based dose estimation implemented in ImpactDose (CT Imaging GmbH, Erlangen, Germany). After incubation for 6 h at 37 °C, a point of time chosen based on the results of our previous study24, samples were transferred into PAXgene Blood RNA tubes (PreAnalytiX, Hombrechtikon, Switzerland). The fraction of peripheral blood dedicated to DNA double strand break analysis was incubated for 20 min at 37 °C before lymphocytes were isolated by density centrifugation as detailed below.

Gene expression analysis

Total RNA was isolated using the PAXgene Blood RNA Kit (PreAnalytiX, Hombrechtikon, Switzerland). Agilent 2100 Bioanalyzer System (Agilent, Santa Clara, CA, USA) was used for RNA quantification and quality assessment. RNA-Seq library preparation was conducted for all samples with RNA Integrity Numbers (RIN) equal or above 7 employing the NEBNext Ultra II Directional RNA Library Kit for Illumina following the manufacturers recommendations (NEBNext Ultra™ II Directional RNA Library Prep Kit for Illumina and rRNA depletion option, New England Biolabs Inc., Ipswich, MA, USA, protocol version 1.0 4/17). After quantification using a Qubit 3.0 Fluorometer (Thermo Scientific, Waltham, MA, USA) and quality control by means of Agilent Bioanalyzer system (Agilent 2100 Bioanalyzer, Agilent, Santa Clara, CA, USA), samples were sequenced on the Illumina NextSeq500 sequencing platform (single-end; 1 × 75 bp; high output reagent kit).

Differential gene expression analysis was based on a workflow suggested by Love et al.34. In brief, read mapping to the human reference transcriptome (GRCh37.67), quantifying transcript abundance and correcting for fragment-level gc-biases was performed with Salmon (version 0.8.1)35. Salmon results were imported via tximport (version 1.4.0)36 into DESeq2 (version 1.22.2)34. Overall estimation of similarities and differences between treatments was done by means of unsupervised hierarchical clustering of sample-to-sample distances based on log2 transformed normalized read counts using the default settings of DESeq2. The hierarchical clustering algorithm groups samples based on the similarities of their gene expression profiles and visualizes this clustering in the form of a dendrogram. The number of clusters is not predefined (unsupervised), but arises as a result of sample composition. To identify the number of genes that were commonly deregulated between the different exposure conditions, differentially expressed genes were displayed using a Venn diagram made with InteractiVenn37. Significant differential gene expression was defined by a log2 fold change of FPKM (fragments per kilobase per million mapped fragments) >|0.5| and Padj (FDR-adjusted p-value) < 0.1. This stringent analysis was complemented by an additional search for candidate genes using slackened criteria in order to accommodate to the subtle effects that can be expected in light of the very low doses applied during CT scans, i.e. a log2 fold change of FPKM >|0.5| and an uncorrected p value < 0.05 in at least two of the three probands. All candidate genes fulfilling these criteria were tested by real-time quantitative polymerase chain reaction (qRT-PCR) in all replicates of all probands.

AmiGO 238,39 was used for gene enrichment analysis applying Fisher’s exact test in combination with False Discovery Rate (FDR p < 0.05) correction 40,41.

DNA double strand break analysis by means of γ-H2AX + 53BP1 co-immunostaining

The frequency of DNA DSBs within peripheral blood mononuclear cells (PBMC) was determined by microscopic quantification of focal accumulations of γ-H2AX and 53BP119, two marker proteins indicative for the presence of a DSB. For this purpose, PBMCs were isolated from blood samples by Ficoll Paque Plus (Merck, Darmstadt, Germany) density centrifugation. PBMCs were washed twice with PBS, fixed in 70% ethanol and stored at − 20 °C. Immunofluorescence staining and DSB foci analysis was performed as described previously19,20,42. Using a Zeiss Axioimager Z2 epifluorescence microscope of the MetaSystems (Altlussheim, GER) ISIS imaging system equipped with appropriate single and dual-band filter sets co-localizing γ-H2AX + 53BP1 foci were counted in 100 well separated and morphologically intact cells by an experienced investigator (HS).

RNA-Seq validation using real-time polymerase chain reaction (qRT-PCR)

Aliquots of total RNA (1 µg) were reverse transcribed with the High-Capacity cDNA Reverse Transcription Kit (Applied Biosystems, Life Technologies, Darmstadt, Germany). PCR reactions were performed in a 96-well format using commercial TaqMan assays (AEN, Hs00224322_m1; DDB2, Hs00172068_m1; FDXR, Hs01031617_m1; BAX, Hs00180269_m1; EDA2R, Hs00939736_m1; 18S rRNA, Hs03003631_g1) following the QuantStudio 12 K Flex Real-Time PCR System amplification protocol. Cycle threshold (Ct) values of the five genes were normalized relative to 18S rRNA. The ratio/fold-change (FC) relative to the unexposed sample at the same time point was determined. A FC of one corresponds to a gene expression similar to unexposed samples. A FC higher or lower than one refers to a several-fold over- or under-expression of the gene of interest after exposure relative to the reference. All materials and consumables used were ordered from ThermoFisher/Applied Biosystems (Weiterstadt, Germany).