Abstract
Closed reduction and internal fixation with three cannulated compression screws is a common method for treating femoral neck fractures in young and middle-aged patients. Protocols including the inverted triangle configuration and dispersion of the screws still needed further supports. The purpose of this study was to explore a novel oblique triangle configuration (OTC) of three screws in fixing femoral neck fractures based on the morphology of the femoral neck isthmus (FNI). The computer-aided design modules were used to explore the ideal spatial configuration with largest triangle by three parallel screws. A univariate evaluation model was established based on the oval-like cross-section of the FNI. When the three screws were positioned by the OTC, Inverted Equilateral Triangle Configuration (IETC), and the Maximum Area Inverted Isosceles Triangle Configuration (MA-IITC) respectively, the proportion of area and circumference in the cross-section of FNI and the changing trend of proportion were compared under various torsion angles, eccentricity, and cross-sectional area of FNI. The area and circumference ratios of the parallel screws using the OTC method were significantly higher than in the IETC and MA-IITC groups. In the univariate evaluation model, the OTC area ratio and circumference ratio remained stable under the different femoral neck torsion angles, FNI cross-sectional area, and eccentricity. The OTC method provided an ideal spatial configuration for the FNA fixation with the largest area using three parallel screws. The position of the posterior screw was also away from the metaphyseal artery, potentially reducing the possibility of vascular injury and screw penetrating.
Similar content being viewed by others
Introduction
Femoral neck fracture is a common fracture of the lower limbs due to high-energy injuries in young and middle-aged patients1,2. Because of the limited blood supply of femoral neck, it is prone to avascular necrosis of femoral head and nonunion after improper internal fixation3,4. Multiple studies have demonstrated that the percutaneous, minimally invasive fixation of femoral neck fractures using cannulated compression screws (CCSs) remained the preferred choice in terms of the lower displacement rate4,5. The parallel screw fixation is currently a common choice in treating femoral neck fractures4,6,7. Furthermore, techniques included the inverted triangle, cortical support, and separation have been proposed to improve the stability of internal fixation and reduce the risk of nonunion and necrosis postoperatively8,9. Parallel implant and cortical support were proved feasibly to achieve using computer navigation technology10,11. However, the inverted triangle pattern and separated technics were controversial since there was no specific method to evaluate the outcomes.
The morphology of the femoral neck isthmus (FNI) has a significant impact on the spatial configuration of the implanted screws when performing the internal fixation of the femoral neck fracture. The biomechanical analysis also proved that the different alignment or dispersion of the screws had a significant influence on the stability of the internal fixation12. Anatomical studies had shown that the cross-section morphology of the femoral neck isthmus was approximately oblique elliptical and affected by the torsion angle, cross-sectional area, and eccentricity of the femoral neck13,14. This study was aimed to explore the optimal spatial configuration of the cannulated three parallel screws for femoral neck fracture in the principles of cortical support and the maximum area of the inscribed triangle ellipse15,16,17. Under the computer-aided model, the area and perimeter of the novel configuration (OTC) were compared with the group of traditional inverted isosceles triangle configuration (IETC) and equilateral triangle configuration. The tendency of the area and circumference ratio of different triangular configurations was analyzed in various torsion angles, cross-sectional areas, and eccentricities of ellipse models.
Materials and methods
This study was performed after obtaining the approval of the Institutional Review Board (IRB) of the General Hospital of the Chinese People’s Liberation Army. A torsional ellipse model was established using the AutoCAD 2007 software (Autodesk Inc., America) to assess the properties of various configuration models13. In order to reduce the individual bias, the morphological data of the FNI cross-section applied in the configuration models was based on our previous study of 200 cases of femoral neck morphology variations14,18. The parameters affecting the morphology of the femoral neck were set in this torsional ellipse model and listed as follows: the torsion angle of the ellipse was 30°, the long and short axis was 17.50 mm and 14.00 mm, and the eccentricity was 0.6018.
In this study, the femoral calcar screw (FCS) was first implanted according to the protocol of parallel inverted triangles. Because of the anatomical torsion of the femoral neck, two operating models with and without torsion angle variables were performed. On the cross-section of FNI, the FCS was first implanted at the bottom of the femoral neck cavity on anteroposterior X-ray and the centerline of the femoral neck on lateral X-ray. The second model was that the FCS was implanted at the lowest point of the femoral neck. Because of the existence of the femoral neck torsion, the screw in the second model was positioned backward relative to the centerline of the femoral neck on lateral X-ray.
In the torsional ellipse model, we first determine the cross-section of the femoral neck isthmus. The largest inscribed circle in the ellipse was drawn, and an equilateral triangle was inscribed in the circle, the endpoints were marked as “a, b, c”. Point A (position of the femoral calcar screw) was determined parallel to the major axes of the ellipse as an endpoint of the oblique triangle. The other two points are made on the ellipse according to the endpoint “b, c” parallel to the major and minor axes of the ellipse, forming an inscribed triangle on the ellipse (Fig. 1)16,19. The largest inscribed triangle on the ellipse was approached under the CAD model.
Then, according to the IETC method, the position of the front and posterior screws on the ellipse were determined. And then, the position of the point B was moved in the ellipse and gradually approached the optimal position of the front and posterior screws with the largest IITC area (Fig. 2). Finally, the triangle area and perimeter of the three configurations were measured.
We set the position of the femoral screws in two operation models as A1 and A3 points respectively and define the midpoint A2 between A1 and A3 on the ellipse. We fixed these vertices to determine the other two vertices B and C, respectively, by the method of inscribing a triangle with the largest area on the ellipse (Fig. 3). The area and perimeter among the Oblique Triangle (OT), Inverted Equilateral Triangle (IET), and Maximum Area Inverted Isosceles Triangle (MA-IIT) were measured.
Under the univariate evaluation model (variables included the torsional angle, area, and eccentricity), a torsional ellipse was first used to simulate the cross-sectional shape of the femoral neck isthmus13. The torsional angle of the ellipse in the first model was set at 10°, 20°, 30°, 40°, and the area and eccentricity are fixed. The second model was to adjust the ellipse area (the semi-major axis was 12.5, 15, 17.5, and 20 mm), and the torsion angle and eccentricity remain the same. The third model was to adjust the ellipse eccentricity (0.4, 0.5, 0.6, 0.7). The largest inscribed triangle (OT), IET, and MA-IIT of the known fixed vertices in the ellipse, and the area and perimeter of these three triangles were drawn. The area and perimeter of the three triangles, and the proportion of the area and perimeter of the ellipse in the three models were compared.
Results
Based on the ellipse model established in this study, the mean area and circumference of the ellipse were 769.69 mm2 and 99.57 mm. The area and perimeter proportions of OT, IET, and MA-IIT using the points A1, A2, and A3 as vertices in the ellipse were shown in Fig. 4. And the results show that the values of OT were significantly greater than IET and MA-IIT, and there was little difference between IET and MA-IIT groups. The area and perimeter of OT in this model remained stable with the position change of point A (A1, A2, A3), the area and perimeter of IET and MA-IIT gradually decrease when point A moves from front to back (Fig. 4). With point A moves, the position of the other two vertices changed in the same direction. And the triangle area moved according to the clockwise pattern (ellipse inclined to the right) (Fig. 5).
In the univariate evaluation model, the area ratio and perimeter ratio of OT, IET, and MA-IIT in the ellipse model was shown in Fig. 6. The area and perimeter ratio of OTC was stable under the given variables (torsional angle, area, and eccentricity) in the torsional ellipse model. The area ratio and circumference ratio of IET and MA-IIT decreased with the increase of the torsion angle and eccentricity of the ellipse, while parameters of OT were not affected by the area of the ellipse (Fig. 6). Moreover, there was little difference between IET and MA-IIT in both the area ratio and circumference ratio.
Discussion
The data from this study proved that the parallel screws using oblique triangle configuration can obtain maximum inscribed area on the FNI cross-section compared with the other methods. The area and circumference of the OTC were significantly larger than traditional IET and MA-IIT configuration and may reduce the potential risks of traditional IET and MA-IIT configuration caused by the cortex penetrating and insufficient cortical support of the screws19,20,21.
Biomechanical and clinical studies have revealed that internal fixation with the parallel inverted triangle, large dispersion, and cortical support showed better mechanical stability and clinical outcomes22,23,24,25,26. The classic screw configuration was based on the rounded cross-section of the femoral neck. However, the cross-section of femoral neck is reported approximately elliptical in the normal population27. In this study, we constructed an oblique elliptical femoral neck isthmus using CAD software to simulate the structure of the femoral neck and used mathematical methods to prove that the OTC with parallel screws showed theoretical advantages in the fixation of femoral neck fractures. And it is more in line with the principles of screw placement: parallel, inverted triangle, dispersion, cortical support.
Lindequist et al. proved that the optimal support of the femoral neck cortex can be achieved when the distance between the screws and the femoral neck cortex was less than 3 mm15. This study found that the area and circumference parameters of the three screws show the lowest change rate during the position change in the OT configuration. While the area and circumference of IET and MA-IIT decreased with the position of the femoral calcar screw from front to back. As the position of the femoral calcar screw changes (clockwise), the positions of the other two screws also change in the clockwise direction. When the femoral calcar screw was implanted on the midline of the femoral neck on the lateral image, the position of the posterior superior screw would be far away from the posterior and superior femoral neck cortex. The position of the inferior screw was reported related to the incidence of stress subtrochanteric fracture, and the probability of stress subtrochanteric fracture may be higher when it is positioned below the level of the lesser trochanter17,28. Compared to the equilateral triangle configuration, the OTC and inverted triangle configuration with a single inferior screw may reduce the potential risk of subtrochanteric fractures. Meanwhile, the inferior screw needs to avoid being excessive inferiorly placed, preventing the penetration of cortical bone.
The screw distance and the triangle circumference of configuration were commonly used to evaluate the degree of screw dispersion19,29. The overall cross-sectional area of the internal fixation is positively related to the bending section modulus and the torsion section modulus16,30. The larger the overall cross-sectional area of the internal fixation indicated the more dispersed stress of the screws in the femoral head, and the less displacement of the fractured end31. The results of this study showed that the area ratio of OT configuration was not affected by the three factors of the inclined ellipse, and remains a stable value (0.413). Therefore, the three parallel screws with oblique triangle configuration showed the largest dispersion and were supposed to be an ideal screw space configuration in the internal fixation of femoral neck fractures.
The main branches of the superior, inferior, and anterior epiphyseal arteries in the epiphysis formed an approximately triangle arterial network32,33. When using an oblique triangle configuration, screws pass through the hypovascular area of the femoral neck, reducing potential iatrogenic injury of the epiphyseal artery. The inferior retinacular artery (IRA) is the main residual artery post-fracture, especially in displaced type. The IRA passes through the femoral neck between the area of the lesser trochanter and quadrate tubercle and subsequently enters the femoral head at the posteromedial area of the femoral neck32. When the femoral calcar screw was implanted on the midline of the femoral neck under the lateral X-ray fluoroscopy, the screw would be far away from the posteromedial area of femoral neck cortex, reducing the incidence of penetration of the screw and the IRA injury.
Tang et al. proposed the “triangular stability theory” of the proximal femur and pointed out that the proximal femur is stabilized by a structural mechanical model formed by the medial, lateral, and upper sides26,34. The medial side forms the oblique support of the cantilever configuration of the proximal femur, which greatly reduces the bending stress and deflection of the structure. The lateral side could significantly reduce the sliding and deflection of the femoral neck under physiological load, and achieve the stability of the fracture end. The upper side connects the medial and the lateral side which could resist the bending moment generated by the physiological load. When the femoral neck fractures, the upper and medial edges of the proximal femur were destroyed at once. The medial edge structure relies on the medial support of the femoral calcar screw to achieve reconstruction. The upper two screws in the oblique triangle configuration provide the best mechanical strength, reconstruct the upper side structure of the proximal femur, and significantly improve the mechanical stability of the fractured end. The configuration of three screws in the OTC model of femoral neck fracture was established to reduce the strain in the internal fixation component, maintain the stability of the fracture end, and promote fracture healing (Fig. 7).
This study has limitations. First, the univariate calculation model using the FNI cross-section provided an idealized experimental condition. Meanwhile, further study on the reconstruction of the entire cavity of femoral neck is required to better simulate the biomechanical stability of screw configuration. Second, parameters included the circumference and cross-section area were reported positively related to rotational stability of the internal fixation. While there was little clinical evidence in evaluating the relationship between these parameters and postoperative stability, which also requires further research to verify the association between the various triangle configuration and postoperative outcomes.
In conclusion, the OTC method provided an ideal spatial configuration with the largest area using three parallel screws for the FNA fixation. The position of the posterior femoral calcar screw was implanted far away from the metaphyseal artery, potentially reducing the possibility of vascular injury and screw penetrating. The OTC method might have potential advantages in mechanical properties in FNF fixation when obtaining the largest triangular area by three screws, while the effectiveness remains biomechanical studies to verify before clinical application.
References
Shao, J. et al. [Epidemiological characteristics of geriatric hip fractures in Hebei Province between 2010 and 2011]. Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chin. J. Reparative Reconstr. Surg. 33, 1287–1290, https://doi.org/10.7507/1002-1892.201804058 (2019).
Hagino, H. et al. Survey of hip fractures in Japan: Recent trends in prevalence and treatment. J. Orthop. Sci. 22, 909–914. https://doi.org/10.1016/j.jos.2017.06.003 (2017).
Stockton, D. J. et al. High rate of reoperation and conversion to total hip arthroplasty after internal fixation of young femoral neck fractures: a population-based study of 796 patients. Acta Orthop. 90, 21–25. https://doi.org/10.1080/17453674.2018.1558380 (2019).
Slobogean, G. P., Sprague, S. A., Scott, T., McKee, M. & Bhandari, M. Management of young femoral neck fractures: Is there a consensus?. Injury 46, 435–440. https://doi.org/10.1016/j.injury.2014.11.028 (2015).
Duffin, M. & Pilson, H. T. Technologies for young femoral neck fracture fixation. J. Orthop. Trauma 33(Suppl 1), S20-s26. https://doi.org/10.1097/bot.0000000000001367 (2019).
Giordano, V. et al. How do orthopedic surgeons manage displaced femoral neck fracture in the middle-aged patient? Brazilian survey of 78 orthopaedic surgeons. Revista brasileira de ortopedia 54, 288–294. https://doi.org/10.1055/s-0039-1691761 (2019).
Oñativia, I. J. et al. Outcomes of nondisplaced intracapsular femoral neck fractures with internal screw fixation in elderly patients: a systematic review. Hip Int. 28, 18–28. https://doi.org/10.5301/hipint.5000532 (2018).
Guo, J., Dong, W., Qin, S. & Zhang, Y. Definition of ideal configuration for femoral neck screw fixation in older people. Sci. Rep. 9, 12895. https://doi.org/10.1038/s41598-019-48258-2 (2019).
Kauffman, J. I. et al. Internal fixation of femoral neck fractures with posterior comminution: a biomechanical study. J. Orthop. Trauma 13, 155–159. https://doi.org/10.1097/00005131-199903000-00001 (1999).
Duan, S. J. et al. Robot-assisted percutaneous cannulated screw fixation of femoral neck fractures: Preliminary Clinical results. Orthop. Surg. 11, 34–41. https://doi.org/10.1111/os.12430 (2019).
Wang, M., Zhang, K. & Yang, N. Development and application of computer assisted optimal method for treatment of femoral neck fracture. Technol, Health Care 26, 31–41. https://doi.org/10.3233/thc-173857 (2018).
Zdero, R., Keast-Butler, O. & Schemitsch, E. H. A biomechanical comparison of two triple-screw methods for femoral neck fracture fixation in a synthetic bone model. J. Trauma 69, 1537–1544. https://doi.org/10.1097/TA.0b013e3181efb1d1 (2010).
Bonneau, N. et al. A three-dimensional axis for the study of femoral neck orientation. J. Anat. 221, 465–476. https://doi.org/10.1111/j.1469-7580.2012.01565.x (2012).
Zhang, R. Y. et al. The oval-like cross-section of femoral neck isthmus in three-dimensional morphological analysis. Orthop. Surg. 13, 321–327. https://doi.org/10.1111/os.12914 (2021).
Lindequist, S. & Törnkvist, H. Quality of reduction and cortical screw support in femoral neck fractures: An analysis of 72 fractures with a new computerized measuring method. J. Orthop. Trauma 9, 215–221. https://doi.org/10.1097/00005131-199506000-00006 (1995).
Feng, X. et al. Axial perspective to find the largest intraosseous space available for percutaneous screw fixation of fractures of the acetabular anterior column. Int. J. Comput. Assist. Radiol. Surg. 10, 1347–1353. https://doi.org/10.1007/s11548-015-1149-6 (2015).
Augat, P., Bliven, E. & Hackl, S. Biomechanics of femoral neck fractures and implications for fixation. J. Orthop. Trauma 33(Suppl 1), S27-s32. https://doi.org/10.1097/bot.0000000000001365 (2019).
Zhang, R. Y. et al. Three-dimensional morphological analysis of the femoral neck torsion angle-an anatomical study. J. Orthop. Surg. Res. 15, 192. https://doi.org/10.1186/s13018-020-01712-8 (2020).
Zhu, Q., Shi, B., Xu, B. & Yuan, J. Obtuse triangle screw configuration for optimal internal fixation of femoral neck fracture: An anatomical analysis. Hip Int. 29, 72–76. https://doi.org/10.1177/1120700018761300 (2019).
Hoffmann, J. C. et al. Is the cranial and posterior screw of the “inverted triangle” configuration for femoral neck fractures safe?. J. Orthop. Trauma 33, 331–334. https://doi.org/10.1097/bot.0000000000001461 (2019).
Shin, K. H., Hong, S. H. & Han, S. B. Posterior fully threaded positioning screw prevents femoral neck collapse in Garden I or II femoral neck fractures. Injury 51, 1031–1037. https://doi.org/10.1016/j.injury.2020.01.032 (2020).
Shehata, M. S. A. et al. Sliding hip screws versus cancellous screws for femoral neck fractures: A systematic review and meta-analysis. Eur J Orthop Surg Traumatol 29, 1383–1393. https://doi.org/10.1007/s00590-019-02460-0 (2019).
Fracture fixation in the operative management of hip fractures (FAITH): An international, multicentre, randomised controlled trial. Lancet. 389, 1519–1527, https://doi.org/10.1016/s0140-6736(17)30066-1 (2017).
Selvan, V. T., Oakley, M. J., Rangan, A. & Al-Lami, M. K. Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study. Injury 35, 136–141. https://doi.org/10.1016/s0020-1383(03)00059-7 (2004).
Li, J. & Wang, M. Optimum configuration of cannulated compression screws for the fixation of unstable femoral neck fractures: finite element analysis evaluation. BioMed Res Int 2018, 1271762. https://doi.org/10.1155/2018/1271762 (2018).
Li, J., Zhao, Z., Yin, P., Zhang, L. & Tang, P. Comparison of three different internal fixation implants in treatment of femoral neck fracture-a finite element analysis. J. Orthop. Surg. Res. 14, 76. https://doi.org/10.1186/s13018-019-1097-x (2019).
Kate, B. R. Anteversion versus torsion of the femoral neck. Acta Anat. 94, 457–463. https://doi.org/10.1159/000144576 (1976).
Sheehan, S. E., Shyu, J. Y., Weaver, M. J., Sodickson, A. D. & Khurana, B. Proximal femoral fractures: What the orthopedic surgeon wants to know. Radiographics 35, 1563–1584. https://doi.org/10.1148/rg.2015140301 (2015).
Li, J. et al. Finite element analysis of different configurations of fully threaded cannulated screw in the treatment of unstable femoral neck fractures. J. Orthop. Surg. Res. 13, 272. https://doi.org/10.1186/s13018-018-0970-3 (2018).
Speck, O., Steinhart, F. & Speck, T. Peak values of twist-to-bend ratio in triangular flower stalks of Carex pendula: a study on biomechanics and functional morphology. Am. J. Bot. 107, 1588–1596. https://doi.org/10.1002/ajb2.1558 (2020).
Lees, M. J. et al. Bone density and cross-sectional geometry of the proximal femur are bilaterally elevated in elite cricket fast bowlers. J. Clin. Densitom. Off. J. Int. Soc. Clin. Densitom. 21, 399–405. https://doi.org/10.1016/j.jocd.2017.06.005 (2018).
Zhao, D. et al. Epiphyseal arterial network and inferior retinacular artery seem critical to femoral head perfusion in adults with femoral neck fractures. Clin. Orthop. Relat. Res. 475, 2011–2023. https://doi.org/10.1007/s11999-017-5318-5 (2017).
Kubik, J. F., Bornes, T. D., Klinger, C. E., Dyke, J. P. & Helfet, D. L. The effect of calcar femoral neck plating on vascularity of the femoral head and neck. Bone & joint open 2, 611–617. https://doi.org/10.1302/2633-1462.28.bjo-2021-0099.r1 (2021).
Li, J., Yin, P., Zhang, L., Chen, H. & Tang, P. Medial anatomical buttress plate in treating displaced femoral neck fracture a finite element analysis. Injury 50, 1895–1900. https://doi.org/10.1016/j.injury.2019.08.024 (2019).
Author information
Authors and Affiliations
Contributions
R.Y.Z., C.Y., X.Y.S., and P.F.T. conceived the study. R.Y.Z., J.X.Z., and J.T.L. wrote the manuscript. R.Y.Z., X.Y.S., and J.T.L. performed the study. R.Y.Z., J.X.Z., J.T.L., Z.Z., and L.C.Z. contributed to the data collection and interpretation of the results.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zhang, R.Y., Li, J.T., Zhao, J.X. et al. The oblique triangle configuration of three parallel screws for femoral neck fracture fixation using computer-aided design modules. Sci Rep 12, 325 (2022). https://doi.org/10.1038/s41598-021-03666-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-03666-1
This article is cited by
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.