Abstract
This study aimed to develop a poly-ε-caprolactone (PCL) material that has a low melting point while maintaining the deformation ability. The new PCL (abbreviated as 4b45/2b20) was fabricated by mixing two types of PCL with different molecular weights, numbers of branches, and physical properties. To investigate the melting point, crystallization temperature, elastic modulus, and elongation at break for 4b45/2b20 and three commercially available masks, differential scanning calorimetry and tensile tests were performed. The melting point of 4b45/2b20 was 46.0 °C, and that of the commercially available masks was approximately 56.0 °C (55.7 °C–56.5 °C). The elastic modulus at 60 °C of 4b45/2b20 was significantly lower than the commercially available masks (1.1 ± 0.3 MPa and 46.3 ± 5.4 MPa, p = 0.0357). In addition, the elongation at break of 4b45/2b20 were significantly larger than the commercially available masks (275.2 ± 25.0% and 216.0 ± 15.2%, p = 0.0347). The crystallization temperature of 4b45/2b20 (22.1 °C) was clinically acceptable and no significant difference was found in the elastic modulus at 23 °C (253.7 ± 24.3 MPa and 282.0 ± 44.3 MPa, p = 0.4). As a shape memory-based thermoset material, 4b45/2b20 has a low melting point and large deformation ability. In addition, the crystallization temperature and strength are within the clinically acceptable standards. Because masks made using the new PCL material are formed with less pressure on the face than commercially available masks, it is a promising material for making a radiotherapy mask that can reduce the burden on patients.
Similar content being viewed by others
Introduction
In radiotherapy for head and neck cancer, a patient-specific thermoplastic mask is created to improve the patient’s positional accuracy because head and neck cancer is often in proximity to critical organs such as the brain stem, spinal cord, and optic nerves1,2,3,4,5,6,7,8. Poly-ε-caprolactone (PCL), which has thermoset properties, is a widely used material in commercially available masks9. The melting point of PCL is generally around 55 °C–60 °C10,11. To create the mask, it is necessary to heat PCL above its melting point, which creates the risk of burns, a physical burden due to the use of sedatives for pediatric patients, and a mental burden due to pressing a high-temperature material against the patient’s face3,12. Thus, the current commercially available masks are of great disadvantage to the patients.
To lower the melting point of PCL, the molecular weight of the material should be decreased13,14. However, when the melting point of PCL is lowered, its deformation ability reduces15. In clinical practice, because the mask is molded by pressing the heated PCL against the patient's face, the pressure applied when creating the mask is increased, or a mask with sufficient accuracy cannot be created when the deformation ability of the PCL is reduced. In recent years, to eliminate the burden on patients caused by heat against their face, many reports have detailed the efficacy of patient-specific masks using three-dimensional print technology (3D printed mask)16,17,18,19,20,21,22,23. However, 3D-printed masks do not have widespread clinical use yet owing to many problems that remain to be solved, such as production technology, cost, time limits, and dose verification20,21. Thermoplastic polyurethane, which is used in body immobilization systems24, is inappropriate for use as a mask because hydrolysis occurs and the stability of the mask decreases when it is immersed in high-temperature water25,26. In contrast, PCL is inexpensive, easy to form, and has a long history of being used as a mask for radiotherapy; moreover, the performance of the immobilization system is guaranteed1,2,3,4,5,6. In addition, the use of PCL has been approved by the United States Food and Drug Administration and can be safely used for radiotherapy. Therefore, a mask made of PCL that has a lower melting point while maintaining the deformation ability would be useful; however, to date, no such product exists.
Uto et al. proposed a method for lowering the melting point by mixing two types of PCL with different molecular weights, numbers of branches, and physical properties27. According to this method, by changing the mixing ratio, a PCL that has an arbitrary melting point between the melting points of other PCLs can be produced. In addition, the melting point can be lowered while maintaining the deformation ability by combining a PCL that has a low melting point with one that has a large deformability. To investigate this possibility, this study aimed to develop a PCL that has a low melting point while maintaining the deformation ability to reduce the burden on the patient.
Methods
Development of the cross-linked PCL
Synthesis of branched PCL macromonomer
PCL was prepared by cross-linking tetra-branched PCL with acrylate end groups in the presence of linear telechelic PCL with acrylate end groups according to a previously reported protocol28,29,30. Briefly, linear and tetra-branched PCL were synthesized by ring-opening polymerization of ε-caprolactone (Tokyo Chemical Industry [TCI] Co., Tokyo, Japan) that was initiated with tetramethylene glycol (Wako Pure Chemical Industries, Osaka, Japan) and pentaerythritol (TCI) as initiators, respectively. Then, acryloyl chloride (TCI) was reacted to the hydroxyl end group of the branched chains to obtain the macromonomers. The structures and molecular weights were estimated by proton nuclear magnetic resonance spectroscopy (1H NMR) (JEOL, Tokyo, Japan). The spectra of the reaction mixture is shown in Fig. 1 and Supplementary figure S1. The average degrees of polymerization of each branch for linear and tetra-branch PCL estimated by 1H NMR were 20 and 45, respectively. The linear PCL with average degrees of polymerization of 20 was abbreviated as 2b20, and the tetra-branched PCL with average degrees of polymerization of 45 was abbreviated as 4b45 (Fig. 2).
Fabrication of cross-linked PCL macromonomers
The 2b20 and 4b45 macromonomers were dissolved in 50 wt% in xylene containing 1.5 wt% (against polymer) benzoyl peroxide (Sigma-Aldrich, St. Louis, MO, USA) and were mixed at a ratio of 33.3 wt% and 66.7 wt%. The solution was injected between glass slides with a 0.2-mm thick Teflon spacer. The samples prepared using thermal polymerization at 80 °C for six hours to obtain the cross-linked PCL were washed using acetone (a good solvent) and shrunk using methanol (a poor solvent), and then dried for 12 h at a reduced pressure. From these processes, a mixed cross-linked PCL, abbreviated as 4b45/2b20, was fabricated.
Physical characterization of the developed PCL
Measurement of melting point and crystallization temperature of the PCLs
To investigate the melting points of 2b20, 4b45/2b20, and 4b45, the thermal properties were measured (temperature range: 0–100 °C) by differential scanning calorimetry (DSC) (DSC-6100, SEIKO Instruments, Chiba, Japan) at a heating rate of 5 °C/min. In addition, to investigate the crystallization temperature, DSC was performed at − 10 °C/min. The melting point and crystallization temperatures were defined as the endothermic and exothermic peaks on DSC.
Measurement of elastic modulus and elongation at break
To investigate the elastic modulus of 4b45/2b20 at heating and room temperature, a tensile test (AG-50kNXplus, Shimadzu, Kyoto, Japan) according to the Japanese industrial standards (JIS) K7161-2 was performed at 60 °C and 23 °C. Additionally, the elongation at the break was measured from the tensile test at 60 °C. The sample was attached using probes with a flat screw grip made of cast iron (Fig. 3). As for the 60 °C conditions, the tensile test was started after heating using the test machine for 15 min at 60 °C. The tensile tests were conducted with a loading rate of 200 mm/min. The elastic modulus was determined using the initial slope of the stress–strain curve on the tensile test, and the elongation at break was defined as the materials’ breakage point. Every measurement was repeated three times to confirm the reproducibility.
Physical characterization of the commercially available masks
Measurement of melting point, crystallization temperature, elastic modulus, and elongation at break
To investigate the physical characterization of the three types of commercially available masks Type-S (CIVCO, Iowa, USA), Portrait (Qfix, PA, USA), and Create (Shenzhen Tengfei Yu Technology Co., Shenzhen, China), analysis was performed using DSC (DSC823e, Mettler-Toledo LLC, Tokyo, Japan). Additionally, tensile test using the Type-S mask was conducted at 60 °C and 23 °C. The heating rate of DSC was 5 °C/min and − 10 °C/min, and the loading rate of the tensile test was 200 mm/min. Every measurement was repeated thrice to confirm the reproducibility.
Statistical analysis
The Mann–Whitney U test was used to compare differences in the elastic modulus and elongation at break between 4b45/2b20 and Type-S. The two-sided p values < 0.05 were significantly different. The statistical analyses were conducted, using EZR version 1.3.6 (Saitama Medical Centre, Jichi Medical University, Saitama, Japan), which is a graphical user interface for R (The R Foundation for Statistical Computing, Vienna, Austria)31.
Results
Melting point and crystallization temperature
The melting point and crystallization temperatures were evaluated using DSC as presented in “Physical characterization of the developed PCL” and “Physical characterization of the commercially available masks”. The melting points of 2b20, 4b45/2b20, and 4b45 were 39.7 °C, 46.0 °C, and 53.0 °C, respectively, and their crystallization temperatures were 2b20, 4b45/2b20, and 4b45 were 18.4 °C, 22.1 °C, and 26.2 °C, respectively. The melting points of the commercially available masks were 56.2 °C (Type-S), 55.7 °C (Portrait), and 56.5 °C (Create), and their crystallization temperatures were 29.7 °C (Type-S), 32.2 °C (Portrait), and 33.2 °C (Create) (Figs. 4 and 5).
Elastic modulus and elongation at break
The elastic modulus and elongation at break were measured using the tensile test as presented in“Physical characterization of the developed PCL” and “Physical characterization of the commercially available masks”. The stress–strain curves on the tensile test of 4b45/2b20 and the commercially available mask (Type-S) at 60 °C are shown in Supplementary figure S2. The elastic muduli at 60 °C were 1.1 ± 0.3 MPa (4b45/2b20) and 46.3 ± 5.4 MPa (Type-S), and the elongation at break were 275.2% ± 25% (4b45/2b20) and 216 ± 15.2% (Type-S). The elastic moduli at 23 °C were 253.7 ± 24.5 MPa (4b45/2b20) and 282.0 ± 44.3 MPa (Type-S). Considerable differences were observed in the elastic modulus at 60 °C (p = 0.0357) and elongation at break (p = 0.0347), but no significant difference was found in the elastic modulus at 23 °C (p = 0.4) (Table 1).
Discussion
The melting point of 2b20 (39.7 °C) was lower than that for 4b45 (53.0 °C). The tendency for the melting point to decrease with the decrease in the molecular weight of PCL was the same as described in previous studies [13,14,33]. In contrast, the melting point of the newly developed PCL (4b45/2b20) was 46.0 °C, which was an intermediate point between 2b20 and 4b45 and was approximately 10 °C lower than the three commercially available masks (55.7 °C–56.5 °C). When the skin temperature is increased to approximately 50 °C, pain is gradually elicited, with grade 1 burns occurring within 3 min and grade 2 burns occurring within 6 min32,33. The PCL melting point of ≤ 50 °C is useful for decreasing the patient burden because the heating temperature can be < 50 °C which is less than that required for a grade 1 burn as just described. Crystallization temperature indicates the temperature around which the material begins to crystallize when creating the mask. Because of the time required for creating the mask when the crystallization temperature is far below room temperature, a crystallization temperature that is around room temperature is preferable for PCL. Therefore, the results shown herein for 4b45/2b20 with a melting point of 46 °C and crystallization temperature of 22.1 °C are useful and within a reasonable range for clinical use.
The elastic modulus of PCL during heating indicates the maximum pressure exerted on the patient’s face when creating the mask. The elastic modulus of 4b45/2b20 was significantly lower than that of the commercially available masks under same conditions. Additionally, the elongation at break indicates the amount of maximum elongation of the material at 60 °C, and that of 4b45/2b20 were significantly larger than the commercially available masks. Linear PCL made of tetramethylene glycol has lower transition temperature (melting point) than tetra-branched PCL and exhibits relatively large deformation ability. As tetramethylene glycol was used for the low melting point PCL of 4b45/2b20, its deformation ability was maintained despite the low melting point. The elastic modulus at 23 °C indicates the strength of the material at room temperature, and no significant differences between 4b45/2b20 and the three commercially available masks were observed (Table 1).
Based on these results, it was confirmed that the 4b45/2b20 has both a low melting point and a large deformation ability as well as clinically acceptable strength. The developed PCLs have an advantage because the handling method is the same as that for commercially available masks, and thus, it can be easily introduced into clinical settings. Additionally, in our previous research, we confirmed using computed tomography images and the similarity index that the PCL created in the same process as this research (“Development of the cross-linked PCL” and “Physical characterization of the developed PCL”) has almost a perfect shape memory performance34. Therefore, even when the mask creation fails for whatever reason (e.g., patient’s movement during mask formation), the shape can be restored to its original shape (i.e., flat) by heating the material again so that the mask can be easily recreated (Supplementary figure S3). Additionally, because the PCL has been approved for cleaning with 70% alcohol based on the recommended cleaning procedures according to the World Health Organization35,36, it can be used safely for at least six months to one year. Thus, the new PCL is promising as a material for a radiotherapy mask that can reduce the burden on patients.
This study has two major limitations. First, the tensile test of the commercially available masks could not be performed in actual conditions for clinical use, which is to heat for few minutes at 75 °C as recommended in the package insert and technical paper9. However, in the tensile test, to ensure uniform temperature of the material, it was necessary to heat the material for approximately 15 min. Because prolonged heating is not expected to be performed in the clinical setting, the commercially available masks are not sufficiently cross-linked, and the material melted during heating at 75 °C. Therefore, in this study, the tensile test at 75 °C could not be performed. To investigate the pressure to the patient’s body under clinical conditions, additional data is needed using a pressure sensor37. Second, this study did not evaluate any mixture other than 2b20 and 4b45. If the melting point is set too low, unintended deformation may occur during transportation depending on the season. Additionally, if the crystallization temperature is too close to room temperature, the time required for the mask to harden increases. The PCL which was mixed with 2b20 and 4b45 at a ratio of 33.3% and 66.6%, has sufficient deformation ability, appropriate melting point, and crystallization temperature. Therefore, the 4b45/2b20 is acceptable for clinical use. However, because the physical characteristics change based on the molecular weight, number of branches, and mixing ratio27,28,29,30, determining the optimal combination by conducting additional evaluations is necessary. In the future, using any reinforcement material also needs to be considered as regards good face-mask adhesion. Although issues, such as price and radiation scattering remain, it is necessary to evaluate by adding additives, such as cellulose to develop better materials.
Conclusions
In this study, we developed a new PCL (4b45/2b20) by mixing two different PCLs and investigated the results of the analysis performed using DSC and tensile tests between the developed PCL and three commercially available masks. The new PCL 4b45/2b20 has a low melting point (46 °C) and large deformation ability, and its crystallization temperature and strength are within the clinically acceptable standards. These results show that 4b45/2b20 will be useful as a material for a thermoset mask for radiotherapy with reduced patient burden.
Data availability
The datasets during and/or analyzed during the current study available from the corresponding author on reasonable request.
References
Bentel, G. C., Marks, L. B., Hendren, K. & Brizel, D. M. Comparison of two head and neck immobilization systems. Int. J. Radiat. Oncol. Biol. Phys. 38, 867–873. https://doi.org/10.1016/s0360-3016(97)00075-8 (1997).
Hong, T. S. et al. The impact of daily setup variations on head-and-neck intensity-modulated radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 61, 779–788. https://doi.org/10.1016/j.ijrobp.2004.07.696 (2005).
Sharp, L. et al. Randomized trial on two types of thermoplastic masks for patient immobilization during radiation therapy for head-and-neck cancer. Int. J. Radiat. Oncol. Biol. Phys. 61, 250–256. https://doi.org/10.1016/j.ijrobp.2004.04.047 (2005).
Karger, C. P., Jäkel, O., Debus, J., Kuhn, S. & Hartmann, G. H. Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system. Int. J. Radiat. Oncol. Biol. Phys. 49, 1493–1504. https://doi.org/10.1016/s0360-3016(00)01562- (2001).
Velec, M. et al. Cone-beam CT assessment of interfraction and intrafraction setup error of two head-and-neck cancer thermoplastic masks. Int. J. Radiat. Oncol. Biol. Phys. 76, 949–955. https://doi.org/10.1016/j.ijrobp.2009.07.004 (2010).
Ahn, P. H. et al. Random positional variation among the skull, mandible, and cervical spine with treatment progression during head-and-neck radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 73, 626–633. https://doi.org/10.1016/j.ijrobp.2008.10.007 (2009).
Fukao, M. et al. Repositioning accuracy of a novel thermoplastic mask for head and neck cancer radiotherapy. Phys. Med. 74, 92–99. https://doi.org/10.1016/j.ejmp.2020.05.005 (2020) ((epub 2020 May 22)).
Tomihara, J., Takatsu, J., Sugimoto, S., Shikama, N. & Sasai, K. Intrafraction stability using full head mask for brain stereotactic radiotherapy. J. Appl. Clin. Med. Phys. 9, 360–370. https://doi.org/10.1002/acm2.13382 (2021) ((epub 2021 Aug 4)).
Ulery, B. D., Nair, L. S. & Laurencin, C. T. Biomedical applications of biodegradable polymers. J. Polym. Sci. B Polym. Phys. 49, 832–864. https://doi.org/10.1002/polb.22259 (2011).
Baptista, C. et al. The effect of temperature and pressure on polycaprolactone morphology. Polymer 191, 122227. https://doi.org/10.1016/j.polymer.2020.122227 (2020).
Moritz, A. R. & Henriques Jr, F. C. Studies of thermal injury: II. The relative importance of time and surface temperature in the causation of cutaneous burns. Am. J. Pathol. 23, 695–720 (1947).
Lendlein, A., Schmidt, A. M., Schroeter, M. & Langer, R. Shape-memory polymer networks from oligo(ε-caprolactone)dimethacrylates. J. Polym. Sci. A Polym. Chem. 43, 1369–1381. https://doi.org/10.1002/pola.20598 (2005).
Tian, G. et al. The effects of PCL diol molecular weight on properties of shape memory poly (ε-caprolactone) networks. J. Appl. Polym. Sci. 136, 47055. https://doi.org/10.1002/app.47055 (2019).
Yang, P., Zhu, G., Shen, X., Yan, X. & Nie, J. Poly (ε-caprolactone)-based shape memory polymers crosslinked by polyhedral oligomeric silsesquioxane. RSC Adv. 6(93), 90212–90219 (2016).
Laycock, S. D. et al. Towards the production of radiotherapy treatment shells on 3D printers using data derived from DICOM CT and MRI: Preclinical feasibility studies. J. Radiother. Pract. 14, 92–98. https://doi.org/10.1017/S1460396914000326 (2015).
Haefner, M. F. et al. 3D-Printed masks as a new approach for immobilization in radiotherapy—A study of positioning accuracy. Oncotarget 9, 6490–6498. https://doi.org/10.18632/oncotarget.24032 (2018).
Asfia, A., Novak, J. I., Mohammed, M. I., Rolfe, B. & Kron, T. A review of 3D printed patient specific immobilisation devices in radiotherapy. Phys. Imaging Radiat. Oncol. 13, 30–35. https://doi.org/10.1016/j.phro.2020.03.003 (2020).
Mattke, M. et al. Individual 3D-printed fixation masks for radiotherapy: First clinical experiences. Int. J. Comput. Assist. Radiol. Surg. 6, 1043–1049. https://doi.org/10.1007/s11548-021-02393-2 (2021) ((epub 2021 May 22)).
Canters, R. A. et al. Clinical implementation of 3D printing in the construction of patient specific bolus for electron beam radiotherapy for non-melanoma skin cancer. Radiother. Oncol. 121, 148–153. https://doi.org/10.1016/j.radonc.2016.07.011 (2016).
Craft, D. F. et al. Material matters: Analysis of density uncertainty in 3D printing and its consequences for radiation oncology. Med. Phys. 45, 1614–1621. https://doi.org/10.1002/mp.12839 (2018).
Dipasquale, G., Poirier, A., Sprunger, Y., Uiterwijk, J. W. E. & Miralbell, R. Improving 3D-printing of megavoltage X-rays radiotherapy bolus with surface-scanner. Radiat. Oncol. 13(1), 203. https://doi.org/10.1186/s13014-018-1148-1 (2018).
Baltz, G. C. et al. 3D-printed headrest for frameless gamma knife radiosurgery: Design and validation. J. Appl. Clin. Med. Phys. 9, 6–15. https://doi.org/10.1002/acm2.12956 (2020) ((epub 2020 Jun 30)).
Verhey, L. J. Immobilizing and positioning patients for radiotherapy. Semin. Radiat. Oncol. 5, 100–114. https://doi.org/10.1054/SRAO00500100 (1995).
Rutkowska, M., Krasowska, K., Heimowska, A., Steinka, I. & Janik, H. Degradation of polyurethanes in sea water. Polym. Degrad. Stab. 76, 233–239. https://doi.org/10.1016/S0141-3910(02)00019-8 (2002).
Pretsch, T., Jakob, I. & Müller, W. Hydrolytic degradation and functional stability of a segmented shape memory poly(ester urethane). Polym. Degrad. Stab. 94, 61–73. https://doi.org/10.1016/j.polymdegradstab.2008.10.012 (2009).
Uto, K., Yamamoto, K., Hirase, S. & Aoyagi, T. Temperature-responsive cross-linked poly (ε-caprolactone) membrane that functions near body temperature. J. Control. Release 110, 408–413. https://doi.org/10.1016/j.jconrel.2005.10.024 (2006).
Ebara, M., Uto, K., Idota, N., Hoffman, J. M. & Aoyagi, T. Shape-memory surface with dynamically tunable nano-geometry activated by body heat. Adv. Mater. 24, 273–278. https://doi.org/10.1002/adma.201102181 (2012).
Uto, K., Ebara, M. & Aoyagi, T. Temperature-responsive poly (ε-caprolactone) cell culture platform with dynamically tunable nano-roughness and elasticity for control of myoblast morphology. Int. J. Mol. Sci. 15, 1511–1524. https://doi.org/10.3390/ijms15011511 (2014).
Uto, K., Aoyagi, T., DeForest, C. A. & Ebara, M. Dynamic alterations of hepatocellular function by on-demand elasticity and roughness modulation. Biomater Sci. 6(5), 1002–1006. https://doi.org/10.1039/C8BM00047F (2018).
Kanda, Y. Investigation of the freely available easy-to-use software “EZR” for medical statistics. Bone Marrow Transplant 48(3), 452–458. https://doi.org/10.1038/bmt.2012.244 (2013).
Hardy, J. D. Thresholds of pain and reflex contraction as related to noxious stimulation. J. Appl. Physiol. 5, 725–739. https://doi.org/10.1152/jappl.1953.5.12.725 (1953).
Henriques, F. C. Jr. Studies of thermal injury; the predictability and the significance of thermally induced rate processes leading to irreversible epidermal injury. Arch. Pathol. (Chic) 43, 489–502 (1947).
Aoyama, T. et al. Physical and dosimetric characterization of thermoset shape memory bolus developed for radiotherapy. Med. Phys. 47, 6103–6112. https://doi.org/10.1002/mp.14516 (2020).
CIVCO. Keeping Patients Safe and Comfortable with CIVCO Thermoplastics. https://civcort.com/ro/Resources/Keeping-Patients-Safe-and-Comfortable-with-CIVCO-Thermoplastics (2020).
World Health Organization. Alcohol and COVID-19: What You Need to Know. https://www.euro.who.int/__data/assets/pdf_file/0010/437608/Alcohol-and-COVID-19-what-you-need-to-know.pdf?ua=1 (2021).
Inata, H. et al. Development of a real-time monitoring system for intra-fractional motion in intracranial treatment using pressure sensors. Phys. Med. Biol. 60, 7229–7243. https://doi.org/10.1088/0031-9155/60/18/7229 (2015).
Acknowledgements
The authors would like to thank Dr. Eri Niiyama and Mr. Sosuke Ouchi for technical support during the PCL creation processes. We thank Mr. Wataru Takahashi of TOYO MEDIC Co., Ltd and Mr. Kazunari Shiohara of Engineering System Co., Ltd., for the helpful advice. We are grateful to Mr. Norio Fukuda of Aichi Center for Industry and Science Technology and Mr. Ryo Takahashi and Ms. Mariko Takeuchi of Kanagawa Institute of Industrial Science and Technology for their useful suggestions. Furthermore, the authors would like to thank Enago (https://www.enago.jp) for the English language review.
Funding
This work was supported by JSPS KAKENHI Grant Number JP19H00439 and JP20H01105 and Children’s Cancer Association of Japan.
Author information
Authors and Affiliations
Contributions
T.A. conceived the study, performed analysis and interpretation of data, and drafted the manuscript, supported by H.S., K.U., and M.E. supported the development of the material. K.U., H.S., M.E., T.K., H.T., K.S., and T.K. were involved in the study design and contributed significantly to the editing of the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Aoyama, T., Uto, K., Shimizu, H. et al. Development of a new poly-ε-caprolactone with low melting point for creating a thermoset mask used in radiation therapy. Sci Rep 11, 20409 (2021). https://doi.org/10.1038/s41598-021-00005-2
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-00005-2
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.