Introduction

Intravenous milrinone, a phosphodiesterase inhibitor and inodilator, has been extensively used in cardiac surgery for the treatment of pulmonary hypertension (PH), particularly during difficult separation from cardiopulmonary bypass (CPB)1,2,3,4. An important drawback of intravenous milrinone is its association with systemic hypotension5,6,7. To avoid this side effect, inhalation has been proposed as an alternative therapeutic route of administration for milrinone8,9,10. Pulmonary drug delivery presents advantages such as rapid absorption, high bioavailability, and high local concentrations11. Consequently, a hypothesis was put forward in the past decade that inhaled milrinone administered prior to CPB would have a protective effect against the exacerbation of PH in cardiac surgical patients12,13 by minimizing CPB-related inflammation14, preventing pulmonary endothelial dysfunction15, and facilitating separation from CPB16. More recently, a multicenter randomized controlled trial demonstrated the clinical efficacy of inhaled milrinone in reducing the degree of PH; however, it was not associated with a reduction in difficulty separating from CPB17. Several factors could explain those results, including suboptimal drug delivery18.

Nebulizers are commonly used drug delivery devices in aerosol therapy for patients with pulmonary diseases. These devices operate by converting liquid formulations into fine breathable droplets. There are three types of nebulizers: jet, ultrasonic and mesh19. Apart from four clinical studies using ultrasonic9,14 or mesh nebulizers17,20, jet nebulizers have been the standard drug delivery devices for inhaled milrinone in adult cardiac patients. Jet nebulizers use a high velocity jet of compressed gas (air or oxygen) flowing through the solution to draw and shear it into aerosol droplets of different range of sizes. Ultrasonic nebulizers use a high frequency vibrating piezoelectric element to produce ultrasonic waves into the solution and breaking it up at the surface into small aerosol droplets. Improvements in nebulization technologies have led to the development of mesh nebulizers using a multiaperture vibrating mesh (up to 10 000) to force the solution through the holes and generate a high fraction of fine aerosol droplets21,22. Mesh nebulizers have gained increasing popularity over the past decade, especially for delivering aerosolized drug therapy to mechanically ventilated patients. They offer many advantages over the two other types of nebulizers – i.e., portability due to a compact design, no secondary airflow, no liquid heating, and high efficiency at delivering drugs to the lungs with little wasted drug23.

This report on inhaled milrinone presents results obtained from a randomized controlled pilot study in cardiac surgical patients undergoing CPB. The objectives of this study were twofold: first, to assess the early plasma concentrations of milrinone in patients after both types of nebulization; and second, to verify whether the systemic exposure in vivo was consistent with the inhaled dosing in vitro. It was hypothesized that given the smaller particle size, milrinone delivered with mesh nebulization would generate higher early plasma levels compared to jet nebulization.

Results

Twelve cardiac surgical patients were recruited between December 2006 and June 2007, with similar characteristics between both groups (Table 1).

Table 1 Patient Characteristics.

In the present study, hemodynamic improvement for each parameter including mean arterial pressure (mAP), mean pulmonary artery pressure (mPAP) and the mAP/mPAP ratio refers to the maximum response after inhalation in variation from baseline (ΔEmax = post-inhalation value – baseline value). Within and between group comparisons are presented in Table 2. The administration of milrinone using mesh nebulization significantly reduced mPAP by 26.2% (26.4 vs. 19.3 mmHg for baseline and post-inhalation, respectively; mean difference, −7.1 mmHg; 95% confidence interval [CI], −10.8 to −3.3, P = 0.005) and increased the mAP/mPAP ratio by 32.1% (2.6 vs. 3.5 for baseline and post-inhalation, respectively; mean difference, 0.8; 95% CI, 0.4 to 1.3, P = 0.005). These hemodynamic improvements were significant after mesh nebulization, but not statistically significant after jet nebulization. No systemic hypotension was reported in our patients.

Table 2 Hemodynamic Monitoring in Cardiac Surgical Patients after Jet (n = 6) and Mesh (n = 6) Nebulization.

In vivo early systemic exposure

Milrinone mean plasma concentrations and individual concentration-time profiles are presented in Table 3 and Fig. 1, respectively. Overall, early systemic exposure of milrinone was significantly higher (2–3 fold) using mesh nebulization.

Table 3 Milrinone Plasma Concentrations in Cardiac Surgical Patients after Jet (n = 6) and Mesh (n = 6) Nebulization.
Figure 1
figure 1

Milrinone plasma concentration-time profiles for 12 cardiac surgical patients after limited sampling (n = 4) following the administration of a 5 mg dose using jet or mesh nebulization.

In vitro inhaled dose

Results for total dose recovered from Setting 1 and Setting 2 are summarized in Table 4. In Setting 1, the mean percentage emitted dose (filter A) was similar with both types of nebulizers (64.0 vs. 68.0% for jet and mesh, respectively; mean difference, 4.1%; 95% CI, −5.4 to 13.5, P = 0.30). However, distribution of residual dose and wasted dose differed within the nebulizer components (cup and T-piece). Residual dose in the nebulizer cup was greater after jet nebulization (29.7 vs. 3.1% for jet and mesh, respectively; mean difference, 26.7%; 95% CI, 24.0 to 29.3; P < 0.001). On the other hand, wasted dose in the nebulizer T-piece was greater after mesh nebulization (1.0% vs. 25.4% for jet and mesh, respectively; mean difference, 24.4%; 95% CI, 15.1 to 33.6; P = 0.004). Mean total dose recovered was 94.7% and 96.5% of nominal dose (5 mg) with jet and mesh nebulization, respectively (mean difference, 1.7%; 95% CI, −3.0 to 6.5; P = 0.37) and a high degree of consistency (CV%, 2.1% and 2.3%, respectively).

Table 4 In Vitro Experiments for Milrinone Dose Recovery.

In Setting 2, the mean percentage of the inhaled dose (filter B) was almost threefold higher with mesh (46.4%) compared to jet (16.6%) nebulization (mean difference, 29.8%; 95% CI, 14.1 to 45.5; P = 0.006). Accordingly, a lower exhaled dose (filter C) was observed with mesh (7.4%) compared to jet (34.1%) nebulization (mean difference, 26.7%; 95% CI, 19.0 to 34.4; P < 0.001). For both types of nebulizers, distribution of the residual dose and wasted dose within the nebulizer cup and T-piece, respectively, were similar to those observed in experiments from Setting 1 (P > 0.05). The mean total dose recovered was 78.6% and 75.1% with jet and mesh nebulization, respectively (mean difference, 3.5%; 95% CI, −9.4 to 16.4; P = 0.49). Consequently, mean backcalculated unrecovered dose in the Y-connector and endotracheal tube was estimated as 16.1% and 21.4% with jet and mesh nebulization, respectively (mean difference, 5.3%; 95% CI, −4.8 to 15.3 P = 0.22).

Discussion

This is the first study reporting on both milrinone early systemic exposure and inhaled dose after nebulization in cardiac surgical patients. In agreement with their respective in vitro inhaled dose, in vivo early plasma concentrations were almost threefold higher with mesh compared to jet nebulization.

The notion of “inhaled aerosol” was introduced for the first time in 1991 by Smaldone24, who defined the inhaled dose as the mass of medication ultimately inhaled into the patient’s airways. Since then, quantification of the inhaled dose in typical clinical settings and conditions has become a popular method for evaluating the performance of devices in pulmonary drug delivery19,25. Inhaled milrinone is still investigational, as it is not yet labeled for the use under discussion.

One of the major benefits of using nebulizers to deliver drugs to the lungs lies in their unique ability to transform drug solutions into fine, breathable droplets. However, like any pulmonary drug delivery system, their performance varies considerably depending on intrinsic factors such as brand and design, additionally to extrinsic factors such as fill volume, flow rate and position in the setting19,23,26,27,28. Furthermore, factors determining drug delivery to the lungs in presence of mechanical ventilation differ significantly from those in spontaneous breathing23,28. Therefore, the choice of inhalation device and appropriate administration techniques according to the type of setting will highly influence the pharmacokinetics of inhaled milrinone (e.g., inhaled dose, input rate, systemic levels, etc.), and eventually drug response, in cardiac patients. In vitro studies were therefore necessary for quantitative assessment of milrinone inhaled dose and proper interpretation of future pharmacokinetic data.

Our in vitro experiments (Setting 1) showed that both jet and mesh nebulizers generated similar emitted doses (Table 4). However, the residual dose in the nebulizer cup was significantly lower with the mesh nebulizer and consistent throughout experiments. Indeed, high stability and minimal residual volume are two well-known properties of mesh nebulizers21,23,28. It is worth mentioning that, although distributed differently within the nebulizer two components (cup and T-piece), approximately 30% of milrinone nominal dose remained trapped within the device in both settings and for both types of nebulizers. This also suggests that the presence of filters placed at different positions in the ventilator circuit did not influence devices performance. In addition, results from Setting 2 showed that milrinone inhaled and exhaled doses together accounted for 50% of nominal dose, for both types of nebulizers. Accordingly, it was estimated that the remaining wasted dose in the ventilator circuit, including the Y-connector and endotracheal tube, would be approximately 20% of the nominal dose.

Evaluation of efficiency in drug delivery yielded an in vitro inhaled dose corresponding to 16.6% of milrinone nominal dose with jet nebulization and 46.4% (almost threefold higher) with mesh nebulization. Other studies testing the same brands of nebulizers have reported similar findings21,27,28. In mechanically ventilated patients, 2–3 fold greater inhaled drug was also observed with the same model of mesh nebulizer compared to another model of jet nebulizer placed in the same setting position21,22,29. From our in vitro experiments, the poor efficiency in delivering a higher inhaled dose with the jet nebulizer may be partially explained by a proportionally higher exhaled dose compared to the mesh nebulizer.

Since inhalation is being considered as an alternative therapeutic route of administration for milrinone in cardiac surgical patients, it was important to determine the early systemic exposure after nebulization in order to verify whether these plasma concentrations were susceptible to exceed the therapeutic range (100–300 ng·ml−1)5,30,31 and cause systemic hypotension5,6,7. In agreement with their respective inhaled dose determined during in vitro experiments, milrinone early systemic exposure was influenced by the type of nebulizer, which resulted in plasma concentrations 2–3 fold greater with mesh compared to jet nebulization. After administration of a 5 mg (50–80 µg·kg−1) dose of milrinone by inhalation using either type of nebulizer, plasma levels observed in our patients were below 100 ng·ml−1, and well below those reported after a 50 μg·kg−1 intravenous bolus dose of milrinone (over 600 ng·ml−1)32. No systemic hypotension was reported in our patients. Significant reduction in blood pressure or mAP following the administration of milrinone has been described when using intravenous milrinone. This is most likely related to the systemic vasodilatory effect of the drug and the elevated peak plasma concentrations (over 200 ng·ml−1)32,33. However, when milrinone is administered by inhalation, plasma levels are lower and no systemic hypotension is reported (as demonstrated in this study)34,35,36. In a recent multicenter randomized controlled trial, we did not observe any difference in terms of arterial pressure change in the group receiving inhaled milrinone vs. placebo17. Furthermore, different hemodynamic effects suggesting selective pulmonary vasodilation and increased atrial contraction were observed17.

The use of a relative hemodynamic parameter to evaluate the severity of PH such as the mAP/mPAP ratio (normal 4 to 1) has been shown to be much more predictive of outcome in various types of cardiac surgical procedures37,38,39,40,41. Induction of anesthesia reduces all pressures including mAP and mPAP, but the mAP/mPAP ratio remains unchanged37. As right ventricular dysfunction develops with PH, the absolute pulmonary artery pressures will tend to normalize but the ratio will remain abnormal. In our study, although there was a significant difference in baseline values for mAP between groups (82.6 mmHg vs. 67.4 mmHg for jet and mesh, respectively), this was corrected by using the mAP/mPAP ratio, which baseline values were not significantly different between groups before drug administration (3.0 vs. 2.6 for jet and mesh, respectively). This ratio is used as a pharmacodynamic measurement of the effect of inhaled pulmonary vasodilators17. For instance, if a vasodilator is active and selective to the pulmonary circulation, mAP will remain unchanged but mPAP will decrease and the mAP/mPAP ratio will increase. On the other hand, if a non-selective vasodilator is administered, both mAP and mPAP will decrease but the mAP/mPAP ratio will remain unchanged. In this study, the administration of milrinone significantly reduced mPAP by 26.2% (−7.1 mmHg) and increased the mAP/mPAP ratio by 32.1% (0.8) after mesh nebulization compared to baseline values. These hemodynamic improvements were not statistically significant after jet nebulization. Therefore, the systemic blood levels generated by inhaled milrinone, although not associated with systemic hypotension, may reflect local pulmonary levels that are effective for reducing the severity of PH as its site of action is located within the pulmonary vascular smooth muscle cells. In other words, our results suggest that it is unlikely that either of the two types of nebulizers would generate sufficiently high plasma levels to induce systemic hypotension in cardiac surgical patients, but mesh nebulization may provide better efficiency in delivering milrinone to the lungs and potentially improve the efficacy of PH treatment.

Because delivering drug to the lungs is a challenging process depending on numerous clinical and device-related factors19,23,26,27,28, in vitro testing can provide the end user with significant information and guidance on appropriate administration techniques for optimal devices performance. In order to acquire useful insight into the pulmonary deposition pattern of inhaled milrinone, our team studied the in vitro particle size distribution using both jet and mesh nebulization42. Gavra et al. found more milrinone particles in the lower airways (mean aerodynamic diameter from 1.4 to 5.4 μm) with mesh nebulization, while greater milrinone particles were collected from the higher airways (mean aerodynamic diameter over 14.1 μm) with jet nebulization. However, milrinone optimal site of deposition in the lungs after inhalation remains unclear. In that study, the inhaled dose was reported as 30% and 60% with jet and mesh nebulization, respectively. Differences in their animal experimental setting and breathing patterns could explain the higher inhaled doses observed compared to those obtained from our in vitro experiments (17% and 46% for jet and mesh, respectively): breathing simulation with single input/output port, constant negative pressure, higher respiratory rate, lower minute volume and lower temperatures (7 °C).

The present study has important limitations

The main one, which constitutes the purpose of additional studies, consists of the absence of a full characterization of milrinone pharmacokinetics and in vivo inhaled dose. The duration of nebulization was very variable and not precisely documented in vivo, which explains why exact timing of blood sampling after inhalation was approximate and may not always represent milrinone maximum concentration. Importantly, the main purpose of this study was to provide preliminary data for a full-scale study on the concentration-effect relationship of inhaled milrinone in cardiac surgical patients.

In conclusion, adequate administration of inhaled drugs relies on careful evaluation of aerosol devices, delivery systems and dosing. Low plasma concentrations of milrinone are observed in cardiac surgical patients following both jet and mesh nebulization. However, mesh nebulization provides better efficiency in delivering aerosolized milrinone to mechanically ventilated patients, resulting in almost threefold increased inhaled dose and systemic exposure compared to conventional jet nebulization. The vibrating mesh nebulizer is more suitable for clinical setting; it offers continuous aerosolized drug therapy, does not require a secondary airflow with additional oxygen which could alter the severity of PH, is reusable, compact and portable compared to the simple jet nebulizer. These data suggest that inhalation represents a promising alternative route of administration for milrinone in cardiac surgery in order to avoid systemic hypotension, and warrants further investigations in a larger scale study.

Materials and Methods

Study design and patients

Following permission from Health Canada (non-objection letter, ref. 108851; November 2, 2006), the study was approved by the Montreal Heart Institute Research Ethics Committee (ref. ICM 06-888; December 6, 2006). The present study is an exploratory substudy (pilot study) of the main research protocol later amended (ref. ICM 06-888; August 5, 2008) and registered at ClinicalTrials.gov (NCT01725776). Written informed consent was obtained from a convenience sample of 12 patients diagnosed with preoperative PH and scheduled for elective cardiac surgery using CPB. Preoperative PH was defined as either systolic pulmonary artery pressure (sPAP) > 35 mmHg or mPAP > 25 mmHg prior to surgery43. Patients with preoperative hemodynamic instability, defined as acute requirement for vasoactive or mechanical support prior to surgery, were excluded. All methods were performed in accordance with the relevant guidelines and regulations.

Surgical protocol

Patients were premedicated with lorazepam (1–2 mg orally) one hour before surgery, morphine (0.1 mg·kg−1 intramuscularly) before entering the operating room and midazolam (0.01–0.05 mg·kg−1 intravenously) at the discretion of the anesthesiologist. In addition to routine monitors44, a five-lead electrocardiogram, radial and femoral arterial catheters, central venous pressure catheter, and a fast-response thermodilution pulmonary artery catheter were placed before induction of anesthesia. Anesthesia was induced with sufentanil (1 μg·kg−1 intravenously) and midazolam (0.04 mg·kg−1), and muscle relaxation was achieved with pancuronium (0.1 mg·kg−1 intravenously). After tracheal intubation, anesthesia was maintained with sufentanil (1 μg·kg−1·h−1) and midazolam (0.04 mg·kg−1·h−1). Hemodynamic monitoring post-induction of anesthesia included mAP, mPAP and the mAP/mPAP ratio before starting inhalation (baseline; 0 min) and after the start of inhalation (at 15, 20, 25, 30, 35, 40, 45 min). Cardiopulmonary bypass was instituted using arterial cannulation of the distal ascending aorta and venous (two-stage or bicaval) cannulation of the right atrium. Blood to crystalloid (4:1) cardioplegia was administered intermittently during CPB with induction and maintenance temperatures ranging from 15 to 29 °C. The patient’s systemic temperature was allowed to drift to 34 °C for coronary artery bypass procedures, and to 32–34 °C for valve and more complex procedures. Weaning from CPB commenced after rewarming to a systemic temperature > 36 °C.

Drug administration

The day before surgery, eligible patients were randomized to receive milrinone by inhalation using either a jet (Airlife Misty Max 10 Nebulizer; Salter Labs, Arvin, CA, USA) or a mesh nebulizer (Aeroneb Professional Nebulizer System; Aerogen Ltd., Galway, Ireland). Randomization was achieved according to a list of random numbers that were assigned to the study devices by a research coordinator who conducted patient recruitment and allocation sequence. The investigators had no access to the randomization list until after data analysis. After induction of anesthesia and baseline transesophageal echocardiography examination, 5 mg (50–80 µg·kg−1) of milrinone (Milrinone Lactate 1 mg·ml−1; Pharmaceutical Partners of Canada Inc., Richmond Hill, ON, CAN) was administered by inhalation before initiation of CPB. Jet nebulization was achieved using a secondary source of airflow (8 L·min−1) from wall oxygen, while mesh nebulization was performed using a battery-powered controller. The assigned nebulizer was connected into the inspiratory limb of the ventilator circuit, immediately before the Y-connector and the endotracheal tube. Milrinone nominal dose, defined as the total drug dose placed in the nebulizer cup, was nebulized until aerosol production was deemed complete (typically 15–20 minutes) after gentle tapping and visualization of the device (for both types of nebulizers). Concomitant medications were administered according to local standards of care, with the exception of milrinone.

In vivo early systemic exposure

Blood samples were drawn to study milrinone early systemic exposure. Arterial blood samples (5 ml) were collected in lithium heparin tubes (Vacuette® heparin tubes; Grenier Bio-One, Kremsmünster, Austria) before starting inhalation (baseline; 0 min) and after the start of inhalation (at 15, 20, 25, 30 min). Samples were kept on ice for a short period of time before centrifugation (3500 rpm; 15 min; 4 °C), and plasma was immediately flash-frozen on dry ice for storage at −80 °C until analysis. Milrinone plasma concentrations were determined by high-performance liquid chromatography using ultraviolet detection (HPLC-UV), as previously described35. The lower limit of quantification (LLOQ) was 1.25 ng·ml−1 with mean intra-assay (n = 6) and inter-assay (n = 5) precisions of < 8%, expressed as coefficients of variation (CV%).

In vitro inhaled dose

This part of the study was designed to closely replicate the typical clinical setting and conditions in the operating room. On three different days and for each type of nebulizer, experiments that mimic in vivo administration of inhaled milrinone to cardiac surgical patients were carried out to determine devices performance in terms of emitted dose and inhaled dose. The experimental setup (breathing system) consisted of a disposable adult breathing circuit connected to an anesthesia workstation (7000 Anesthesia Ventilator Multi-Voltage Electronic; Ohmeda, Madison, WI, USA) with an expandable balloon (3.0 L Rusch® Breathing Bag; Teleflex, Morrisville, NC, USA) placed at the patient interface (i.e., directly connected to the distal end of the endotracheal tube). Representative breathing patterns for our patient population were used including tidal volume (500 ml), respiratory rate (12 breaths per minute), minute volume (6 L per minute), and inspiratory:expiratory ratio (1:2)45,46.

Each nebulizer was composed of a cup (reservoir) and a T-piece. The jet nebulizer was designed with the T-piece positioned on top of the nebulizer cup (Fig. 2A) and vice versa for the mesh nebulizer (Fig. 2B). Low-resistance collecting filters (Vital Signs Inc., Totowa, NJ, USA) were used to collect aerosolized milrinone over the duration of nebulization (Fig. 3). For each type of nebulizer, two series of experiments were carried out using the same setup, but having collecting filters placed at different positions within the breathing circuit (Fig. 3A,B). Milrinone administration was as earlier described for patients.

Figure 2
figure 2

Jet nebulizer (Airlife Misty Max 10 Nebulizer; Salter Labs, Arvin, CA, USA). (A) Mesh nebulizer (Aeroneb Professional Nebulizer System; Aerogen Ltd., Galway, Ireland) (B).

Figure 3
figure 3

In vitro settings for estimation of milrinone dose recovery (mesh nebulizer displayed). Setting 1 (A) was used to determine the emitted dose (filter A). Setting 2 (B) was used to determine the inhaled dose (filter B) and exhaled dose (filter C). The residual dose (nebulizer cup) and wasted dose (nebulizer T-piece) were also measured.

Setting 1 (Fig. 3A) was used to determine the emitted dose, defined as the mass of medication leaving the nebulizer. For this experiment, filter A (emitted dose) was placed immediately at the outflow of the nebulizer T-piece. Setting 2 (Fig. 3B) was used to determine the inhaled dose and exhaled dose, defined as the mass of medication inhaled into the patient’s lungs and the mass of medication wasted during the expiratory phase, respectively. For this experiment, filter B (inhaled dose) was placed at the distal end of the endotracheal tube immediately before the expandable balloon (3.0 L breathing bag), and filter C (exhaled dose) was placed at the expiratory limb of the ventilator circuit immediately after the Y-connector.

At the end of nebulization, the doses remaining within the nebulizer cup (residual dose) and the nebulizer T-piece (wasted dose) were determined. Laboratory film (Parafilm®; Bermis Company, Inc., Neenah, WI, USA) was used to seal the T-piece and each of the nebulizer components (cup and T-piece) was rinsed with sterilized water (2 ml) before collection in a test tube for analysis. Milrinone content was eluted from filters after immersion in 250 ml buffer solution (NaH2PO4 50 mM, pH 3) and 10 min ultrasound sonication (Ultrasonic Cleaner FS 28 H; Fisher Scientific Co., Hampton, NH, USA). Preliminary experiments had confirmed quantitative recovery of milrinone from filters (99.1%, n = 4). Samples were stored at −20 °C until analysis. Milrinone concentrations were determined by HPLC-UV using a simplified version of the assay used for plasma (see above). Six concentrations of milrinone prepared in buffer solution (20 down to 0.01 µg·ml−1) were used to establish calibration curves (r2 = 0.9967, n = 6). Each sample was injected twice in the HPLC with the mean value used for data analysis.

Statistical analysis

For the in vivo experiments, patient characteristics and milrinone concentrations at each time point were expressed as mean (standard deviation). Comparisons between groups (jet vs. mesh) were performed using unpaired Student t-test for continuous variables and Chi-square test for categorical variables. Comparisons within group (post-inhalation vs. baseline) were performed using paired Student t-test.

For the in vitro experiments, mean individual dose recovered was determined and expressed as a percentage of the nominal dose (5 mg). For each replicate, the total dose recovered was obtained by summing individual recoveries. In Setting 2 only, unrecovered doses in the Y-connector and the endotracheal tube could not be directly determined due to the difficulty of disconnecting these components without significant spillage. Instead, the unrecovered dose from those two components was backcalculated by subtracting individual recoveries determined for the other components from the total dose recovered determined in Setting 1.

Statistical analysis was carried out with SigmaPlot® version 11.0 (Systat Software, Inc., San Jose, CA, USA). A P < 0.05 was considered statistically significant.