Introduction

Mycobacterium chimaera, a species of the Mycobacterium avium complex, is ubiquitous in the environment and is an opportunistic pathogen capable of causing human infections. In 2015, the infection control team at University Hospital Zurich reported that multiple patients had infections due to M. chimaera following open-heart surgey, with the outbreak linked to LivaNova (formerly Sorin) 3 T heater-cooler units (HCUs)1. Since then, outbreaks of infections due to M. chimaera among patients who received open-heart surgeries and were exposed to 3 T HCUs have also been reported in several European countries (Germany, the Netherland and UK), USA and Australia2,3,4. Infections due to M. chimaera are extremely difficult to treat and patients with such infections have a mortality rate of approximately 50%4,5,6. HCU-associated outbreaks of infections due to M. chimaera have been linked to the manufacturing site7,8. Although 3 T HCUs are also used in China, no such HCU-associated outbreaks have been reported. Here we report a study looking for the presence of M. chimaera in HCUs in a university hospital in China, with a hypothesis that the absence of such outbreaks is due to the absence of the organism in HCUs used in our hospital.

Materials and Methods

Settings

West China Hospital of Sichuan University, Chengdu, China, is a 5,000-bed tertiary hospital and serves as one of the major referral medical centers in western China. Around 2,600 patients received HCU-required open-heart surgeries in 2016 in the hospital. There are five 3 T HCUs, which are the only type of HCU, in the hospital. Three HCUs were manufactured in July 2009 and were introduced into the hospital in 2010 (n = 1) and 2012 (n = 2), while the remaining two were manufactured and were introduced into the hospital in August 2016. The HCUs were cleaned and disinfected every two to three months using the manufacturer’s protocol.

Sampling and culture

We sampled 1 liter of water from each of the five 3 T HCUs at West China Hospital before they underwent disinfection in May 2017. The protocol developed by the European Center for Disease Prevention and Control (ECDC)9 was used for our study. Water samples were concentrated by filtration and were subjected to culture on M7H11 plates (Haibo, Qingdao, China) for up to eight weeks9.

Whole genome sequencing

Mycobacteria isolates were subjected to whole genome sequencing with 300× coverage using the HiSeq X10 Sequencer (Illumina, San Diego, CA). The coverage was calculated based on the estimated genome size and the average output of the sequencer. Reads were trimmed using Trimmomatic10 and were then de novo assembled to contigs using the SPAdes program11 with careful mode turned on. Species identification of mycobacteria isolates was established based on average nucleotide identity (ANI) with the genome of strain DSM 44623T, the type strain of M. chimaera (GenBank accession no. LQOO00000000), using the JSpecies web program (http://imedea.uib-csic.es/jspecies/).

Phylogenetic analysis

To determine the group and the subgroup of our three isolates, the paired-end reads of the isolates were mapped against the chromosome of the reference strain ZUERICH-1 of group 1 (Accession no. NZ_CP015272) using Bowtie 212 and SAMtools13. Raw variants were called using Freebayes (https://github.com/ekg/freebayes). INDELs were discarded, and SNPs were filtered keeping only those with a minimum coverage, quality score and allele frequency of 10-fold, 30 and 0.75, respectively using vcflib (https://github.com/vcflib/vcflib). SNPs due to recombination were filtered using Gubbins14.

Illumina paired-end reads of all M. chimaera genomes available in the NCBI SRA database with a minimum of 200 Mbp (approximately 30× genome coverage) were retrieved (n = 508; for strains with multiple genome sequencing, only the first SRA record was retrieved). Ninety-eight of these genomes of 98 were discarded due to unresolvable problems (e.g. unequal length of sequence and the presence of invalid characters) of format and quality of their reads, resulting in a collection of 410 genome sequences. The raw sequence reads of these 410 strains were mapped against ZUERICH-1 and SNPs were called as described above. A phylogenetic tree of isolates belonging to subgroup 1.1 was constructed using RAxML15 with the GTRGAMMA model and a 1,000-bootstrap test.

Nucleotide sequence accession numbers

Draft whole-genome sequences of the three strains have been deposited into GenBank under the accession no. NTFV00000000, NTFW00000000 and NTFX00000000.

Results

Mycobacteria grew from three HCUs manufactured in 2009 but did not grow from the two manufactured in 2016. One isolate from each of the three culture-positive HCUs, designated WCHMC000001, WCHMC000030 and WCHMC000032, was subjected to whole genome sequencing.

A total of 1.99 to 2.13 Gb clean bases and 6,618,824 to 7,116,445 clean reads were generated for the three isolates, which were then assembled to 6.1 to 6.6 Mb draft genomes with 84 to 112 contigs (77 to 104 were ≥1,000 bp in length) with a 67.46 to 67.64% GC content, respectively. The genome sequences of the three isolates had a 99.14 to 99.52% ANI with the genome of strain DSM 44623T. The three isolates were therefore identified as M. chimaera.

A recent study of the genomic sequence of 250 isolates revealed two major groups of M. chimaera and isolates of the HCU-related outbreaks belonged to group 17. Using 10 single nucleotide polymorphisms (SNPs) for subgroup attribution, group 1 has been further classified into 11 distinct subgroups. Most isolates from water systems of 3 T HCUs in clinical use (Australia, Denmark, Germany, the Netherlands, New Zealand, UK, USA) and some from patients (Australia, Germany, the Netherlands, UK, USA) plus one from HCUs sampled at the LivaNova production site were clustered in the subgroup 1.17. There were a maximum of 6 SNPs difference between our three isolates (Table 1), suggesting that they originated from a very recent common ancestor. WCHMC000001, WCHMC000030 and WCHMC000032 differed from ZUERICH-1 by only 7, 9 and 10 SNPs, respectively, suggesting that the three isolates belonged to group 1. Substitutions of guanine (G) by adenine (A) at positions 113,518 and 209,278 of the DSM 44623T genome (GenBank accession no. LQOO00000000) were specific to subgroup 1.17, both of which were found in our three isolates. This suggests that the three isolates belong to the subgroup 1.1.

Table 1 SNPs among the three isolates and most closely related isolates.

Among the 410 strains from the SRA database, 169 were found to harbour either or both of the subgroup 1.1-specific SNPs. A total of 295 SNP sites were identified from these 169 strains. As revealed by the phylogenetic tree of isolates belonging to the subgroup 1.1 (Fig. 1), our three isolates were most closely related to strains recovered from HCUs or patients in several European countries, Australia, New Zealand and USA with a maximum of just 5 SNPs difference between them (Table 1). The high clonality of isolates of the subgroup 1.1 from different geographical regions suggests a common source, which is most likely from contamination at the manufacturer site.

Figure 1
figure 1

Maximum likelihood tree built from 295 SNP positions of the 169 strains of the subgroup 1.1 mapped to the genome of M. chimaera ZUERICH-1 shown as a circular phylogram. The three isolates in the present study were indicated in red.

Discussion

We recovered M. chimaera from water tanks of three LivaNova 3 T HCUs at a university hospital. To be the best of our knowledge, this is the first report of M. chimaera from HCUs in China. Genome sequencing and phylogenetic analysis revealed that the three isolates were clonal and were placed within strains from multiple countries, suggesting the same common source.

It has been reported that brand new 3 T HCUs direct from the manufacturer can be quickly contaminated by M. chimaera1,8. However, no M. chimaera was recovered from the two recent HCUs manufactured in August 2016. There are several possible reasons. First, brand new 3 T HCUs have been shown to be contaminated after 158 to 358 days16. The two HCUs in our hospital had been in use for 8 months and it may be that sufficient time has not yet passed for them to be colonized to detectable levels by the organism. Second, the manufacturer has modified the post-production process in response to the findings of M. chimaera contamination8, which may reduce the risk of contamination in all new units. Third, the M7H11 media that we used has a higher detection limitation than the Mycobacterial Growth Indicator Tubes and may therefore be suboptimal17.

Unfortunately, there is no facility for blood culture of mycobacteria available in the hospital and there is no follow-up system for patients who received heart surgeries in the region. We are therefore unable to identify any clinical case of M. chimaera infection at present. Nonetheless, information of the global 3 T HCU-associated outbreak of infections due to M. chimaera and the local discovery of such M. chimaera strains has been disseminated to the cardiac surgeons and infectious disease physicians in the hospital. The facility for blood culture of mycobacteria will be established soon in the hospital to allow thorough future investigation of potential cardiac related M. chimaera infections.

Several measures have been proposed to minimize the risksof M. chimaera infections including strict separation of HCUs from the air volume of critical medical areas such as operating rooms, ensuring traceability of HCD use and following the updated manufacturer’s disinfection recommendations5,18. Unfortunately, it is not practical to place 3 T HCUs outside operating rooms in our hospital at present. The devices have been placed further away from patients, guided by smoking tests as described previously19. This practice could reduce patient risk but as patients are still exposed to HCUs in the room they need to be monitored for M. chimaera infections18. Strict separation of HCUs from operating rooms will be considered at the time of renovation. In addition, HCUs have been subjected to cleaning and disinfection using the updated protocol provided by the manufacturer (http://www.livanova.sorin.com/products/cardiac-surgery/perfusion/hlm/3t) since June 2017.