Introduction

Remote monitoring systems of human healthcare have been widely studied because they can reduce the risk of secondary infection among healthcare workers1. Robotics offers a promising method to mitigate risk and improve patient care effectiveness and quality as a focused remote healthcare technology2. Blood pressure (BP) is an essential medical diagnostic tool in the medical healthcare system because high BP is closely related to several chronic diseases3. In addition, hypertension or high BP is a pre-existing symptom or precursor of possible infection in patients. This increases the demand for BP monitoring systems of patients in their daily lives4. From this point of view, the use of sensing robots in medical healthcare systems has substantial advantages because they can assist healthcare workers in monitoring the vital signs of patients and creating a friendly environment for those patients who may need to be isolated3. For the development of BP monitoring using robots, conventional methods with a cuff-based sphygmomanometer are not feasible because they cannot perform continuous monitoring of patients4. Thus, a cuffless monitoring system combining sensing tools of electrocardiogram (ECG)5 and photoplethysmography (PPG)6 is rational. The pulse arrival time (PAT) parameter, which indicates the time taken for the pulse waveforms to traverse from the heart to a distal site, is often used in a paired sensing system7. PAT is determined using fiducial points, including the R-wave and the systolic peak of the ECG and PPG waveforms8,9. In addition, unlike a sphygmomanometer, an ECG–PPG paired sensing system can continuously monitor BP8,10. Moreover, once such healthcare tools are integrated with sensing robot systems, they can form a bridge in remote or general monitoring between medical personnel and patients receiving outpatient care.

However, an ECG sensing system, an essential component for BP monitoring, requires direct contact with the human skin to collect the cardiac signals; moreover, there is the issue of high noise signals depending on the contact conditions11. Therefore, wet- or sticky-type electrodes are employed for the system12. Although such wet-type electrodes are appropriate for solving the contact issue, they can cause skin irritation and allergy in patients13,14. In addition, it is not appropriate to monitor BP with the robot because, at each measurement, the electrodes need to be replaced or treated. Consequently, the demands for dry electrodes for ECG sensors have dramatically increased, reducing the risk of skin troubles and requiring less treatment during BP measurement15. However, the use of dry electrodes for the ECG sensor causes contact drawbacks16. Hence, methods to impose adhesive properties to dry electrodes for ECG sensors are urgently required. In recent years, conducting polymers such as poly (3,4-ethylene dioxythiophene): polystyrene sulfonate (PEDOT: PSS) have been used as the dry electrode of the ECG sensor after inserting silver/silver chloride (Ag/AgCl) into the conducting polymer matrix to improve the interface with the skin17. In addition, self-adhesive electrodes for the sensor were designed using extra soft or microstructurally patterned polymers18,19,20,21,22,23. Moreover, printing patterned dry electrodes on stretchable polymers has been reported to improve adhesion to the skin24. Although the aforementioned studies contribute to improving the inadequacies of dry electrodes of ECG sensors, additional support systems such as belts- or tapes-tightening to secure the adhesion of the dry electrodes to the skin are still required, which can make frequent monitoring challenging and inconvenience of the patients.

A simplified compact hardware of the system, which encompasses all the aforementioned requirements, is required14. Otherwise, the sensing system would hinder actual robot motion. Therefore, among possible solutions, the architectural properties of origami are strongly considered because origami has peculiar mechanical properties derived from decorative folding patterns25. For example, the mechanical rigidity of the structure is tunable by the customization of folding patterns on the structure. In addition, origami has force direction-shifting properties such as rotating26 or bending27 using a simple axial force. Moreover, the physical dynamics of origami mainly occur on the crease line such that the patterned planes or facets are scarcely affected, which can be utilized as potential location sites of sensing units28.

Origami has many folding patterns and methods for generating various types of 3D structures. Also, origami has shown unique characteristics such as suction for grasping29, and foldability inspired by nature, such as the origami robot inspired by ladybug wings or dual morphing pelican eels30,31. Among complex mechanisms of leech’s adhesive feature, it has an expandable posterior sucker and body32. Its organs expand and shrink appropriately to maintain better adhesion to the victim33,34. This adhering mechanism can be considered as the simultaneous motion of non35 and auxetic36 structures. From this point of view, origami can achieve such motion in a monolithic structure using folding patterns26. Moreover, tuning or customizing processes are available after the development of such nature-inspired origami26.

In this study, a leech-inspired origami (LIO) sensor was developed to monitor biomedical vital signs using a humanoid-sensing robot. The difficulty in continuing essential health services during the COVID-19 pandemics provides a lesson on the needs of remote health care systems. Thus, a robot-assisted system for BP monitoring is an ideal remote healthcare system. Dry serpentine electrodes were three-dimensionally (3D) printed on the LIO sensor for their stability against the expansion and contraction of the LIO sensor. The developed LIO fingertip sensors were integrated into the humanoid robot and demonstrated reliable ECG sensing performance compared to commercial wet ECG electrodes. Finally, a BP monitoring robot system composed of paired ECG and PPG sensing hands was demonstrated by human–robot interaction. Different BP profiles were obtained from the developed sensing robot system through interaction with human subjects under different exercise conditions.

Results and discussion

Design of the LIO sensor

Leeches expand the mouth (sucker) to increase the contact area with the skin, and simultaneously, its body is shrunk to prepare for the pumping activity to suck the blood of the victim, as presented in Fig. 1a, b32. During sucking, its expanded mouth is shrunk, and the body is expanded, as shown in Fig. 1c. Structurally, such a mechanism is considered a combination of simultaneous non-auxetic and auxetic behaviors. Based on this mechanism, an origami composed of a non-auxetic tip with auxetic bodies was developed (Fig. 2a). Once the non-auxetic structure in the origami is expanded, the auxetic part shrinks under compression, as shown in Fig. 2b. During compression, the area of the non-auxetic part in origami increases as observed in Fig. 2c in the x, y plane view, implying the expansion of the non-auxetic part in origami. By contrast, the auxetic part is shrunk, indicating a reduction in the inner space. In addition, the LIO has distinctive features, including rotating when a z-directional force is applied. The details of the structural behaviors of the origami are presented in Supplementary Video 1. The corresponding area changes were measured depending on the compressed heights in Fig. 2d. The area of the non-auxetic structure shows linearly upward trends at a more compressed length, Δh, and at 8 mm of the length, its area at the x, y plane view is 1.6 times larger than the de-compressed state. In the auxetic origami part, the area is reduced to a factor of 0.4 after compression. The rotating behavior of the origami structure is shown in Fig. 2e. The rotating angle of the origami has a linear relationship with Δh, indicating that the force is converted from the z-directional force to torque. With these structural features, the origami structure is expected to form a standby state to pump after compression; the structure starts sucking during decompression like a leech. As shown in Fig. 2f and Supplementary Video 2, the LIO stably contacts and sucks the human finger such that the LIO is still attached to the fingertip when the finger is moved.

Fig. 1: Design of Leech-inspired origami (LIO).
figure 1

Schematics of the sticking mechanism of the posterior leech sucker. a Before the attachment, the shrunk sucker is positioned to target the skin surface. b The sucker part of the leech is stretched out along the radial direction and shrunken the body to attach to the skin. c The leech’s body expands for the pumping to the victim’s blood. The body trunk of the leech was mimicked by auxetic design, and the sucker was mimicked by non-auxetic origami design. The 3D origami composed of non-auxetic tip and the auxetic trunk is proposed in this study.

Fig. 2: The mechanical behavior of the designed LIO structure.
figure 2

a Schematics of the posterior sucker and body of leech and b their corresponding leech-inspired non- and auxetic origamis with the folding features under compression. c The actual images of 3D printed LIO under compression, side views (top), and top views (bottom). d The normalized area changes of LIO. e Rotated angles depending on compressed distances. f Representative photos of the adhesive performance of the designed LIO.

Parametric study to optimize the LIO

As mentioned earlier, origami can generate various 3D structures, even mimicking bio-nature. Beyond structural generation, its structure can be customized to improve specific mechanical properties. For the optimized conformal contact of origami for the development of the ECG sensing unit, the LIO is parametrically studied with tunable factors. The auxetic part of the LIO is composed of a three-layered Kresling origami structure27. Detailed information about the Kresling origami is described in Supplementary Note 1 and Supplementary Fig. 1. When angular designs of triangular patterns are tailored on origami structures, as shown in Fig. 3a, the mechanical properties of the structure are adaptable. In this parametric study, we compared the responses of three different parameters at different angles: (1) Poisson’s ratio, (2) rotation angle, and (3) elastic modulus of the origami structures26,27.

Fig. 3: Parametric study for the optimization of LIO.
figure 3

a Schematics of different leech-inspired origami (LIO) designs. The auxetic origami is tailored by changing the angle of triangular patterns. b Poisson’s ratio of the non-auxetic and auxetic parts of the LIO at the z-directional strain. c Actual images of auxetic origami to represent transverse strain (εtrans) at z-directional compression (εaxial). d Rotated angles at the z-directional compression. e The stress–stain curve of the LIO design with the angular of σ = 30°, β = 40°. f Elastic modulus of LIO designs with different angles. Error bars represent the standard deviation.

Poisson’s ratio reflects the degree of perpendicular strain depending on the axial force. Thus, when the Poisson’s ratio is changed, the volume ratio before and after compression is varied. Consequently, the volume ratio affects the suction strength of the origami structure by generating a vacuum. Such a vacuum is estimated using Eqs. 1 and 2.

$${{{\mathrm{Vacuum}}}}\,{{{\mathrm{\% }}}}:{{P}}_{{{{\mathrm{v\% }}}}} = {{{\mathrm{100\% }}}} - \left( {P_o/P_{{{{\mathrm{atm}}}}}} \right)\, \times\, {{{\mathrm{100\% }}}}$$
(1)
$${{{\mathrm{Ideal}}}}\,{{{\mathrm{gas}}}}\,{{{\mathrm{equation:}}}}\,{{P}}_{{{\mathrm{o}}}} = {{nRT/V}}_{{{\mathrm{o}}}}$$
(2)

where Pv%, Po, Patm, and Vo are the percent of vacuum (vacuum %), the absolute pressure in the origami structure, absolute pressure in the standard atmosphere, and the actual inner volume of the structure, respectively. It can be seen that Pv% is proportional to Vo. Thus, the more negative the Poisson’s ratio of the origami structure, the higher the suction strength. Figure 3b shows the Poisson’s ratio profiles at different z-directional strains (εaxial and axial strain) with various angular designs on origami. The transverse strain (εtrans) is represented by the blue line in Fig. 3c. The lowest Poisson’s ratios with negative values are observed from the auxetic part at a strain of approximately 0.15, indicating a fully shrunken state in the transverse direction. By contrast, the non-auxetic structure shows the highest Poisson’s ratio of approximately 1.7, with positive values at a strain of approximately 0.25. At a fixed α angle, a higher β angular design shows a lower Poisson’s ratio. The boundary conditions for the α and β angles are described in Supplementary Note 1 with Supplementary Fig. 2.

Additionally, we consider that the rotating behavior of the origami structure is an important property. The rotating of origami under compression results in more effective contact with the human skin because of its rubbing effects. Thus, it can be concluded that the rotating feature of origami needs to be investigated for better ECG sensing performance. The rotation angles of the origami structure are shown in Fig. 3d. They also exhibited linear trends at higher angular designs of the auxetic parts. However, the non-auxetic structure is nonrotational with the z-directional force. Thus, the rotation of the entire structure is primarily caused by auxetic elements.

These two physical properties of LIO are derived from the collapsing or folding behavior, which is affected by the rigidity of the origami creases, as presented in Supplementary Fig. 1a, b. Thus, the structural rigidity of the origami was investigated. Compression tests of the origami were performed to obtain the elastic modulus (Fig. 3e). The stress–strain curve has three distinctive peaks at approximately 15, 30, and 45% strains, derived from the multi-stability of the origami26,27. The three different elastic moduli were measured from the various angular origami designs shown in Fig. 3f. Generally, the higher β angular designs of the LIO structure show a greater elastic modulus. The structure with β = 38° has 0.79 ± 0.090, 0.33 ± 0.081, and 0.25 ± 0.054 MPa for regions #1 to #3, respectively, as the lowest values among the auxetic samples. At β = 39°, elastic moduli of 0.98 ± 0.096, 0.39 ± 0.082, and 0.35 ± 0.078 MPa are obtained for regions #1 to #3, respectively. The elastic modulus of region #1 for both designs is more than twice as high as that of regions #2 and #3, owing to the force-dispersing properties of the origami structure or structural rigidity before the collapse initiation26,37. After the first region, the LIO structure gradually collapses during compression, as shown in Supplementary Videos 1 and 2, reflecting a similar elastic modulus at regions #2 and # 3. In the case of the angular designs, at β = 40°, the elastic modulus of region #2 (0.52 ± 0.086 MPa) is lower than that of the other regions (#3, 0.65 ± 0.050 MPa). At β = 41°, the difference (ΔE = E#3 − E#2) of the elastic modulus between regions #2 and #3, 0.73 ± 0.031 MPa, and 1.03 ± 0.032 MPa, respectively, is 0.30 MPa, which is higher than 0.13 MPa at β = 40°. This finding implies that the LIO structure with an angular design of β > 40° does not gradually collapse; the partial collapse of the 2nd origami layer occurs with the 1st layer. This finding implies that the structural rigidity is close to the material rigidity assumed from the stress–strain curve beyond region #3 after the modification of the angular designs of the LIO of β > 40°. LIO with β = 42° does not show collapsing or folding behaviors, as shown in Supplementary Fig. 2b, because the structural rigidity of the LIO is greater than its material property. Based on the parametric study, an origami structure with α = 30° and β = 40° was selected as the optimal candidate for the LIO of the conformal ECG sensor. As the optimal criteria, we considered the angular designs for the maximum Poisson’s ratio and rotation angles of the LIO structure; however, to maintain the better folding behavior of the LIO, an angular design with β = 40° was selected instead of β = 41°. Adhesion force has been measured by using the pull-off force test, as shown in Supplementary Fig. 3c. The LIO with α = 30 ° and β = 40° shows 1.21 ± 0.24 N.

Conductive serpentine for ECG electrode design

In the design of reliable ECG electrodes, the electrode must be robust against the strain when it is expanded. The expansion of the LIO structure is a result of the combination of both x-and y-directional deformations. Although the origami is mainly deformed at the crease line without theoretical plane deformation, an apparent strain or stress distribution exists on the plane in actual cases38. Thus, the design of a stable structure for ECG electrodes with planar deformation is required. Serpentine-patterned electrodes were printed on the non-auxetic plane of the LIO. As demonstrated in Fig. 4a, the non-auxetic part of the LIO is fixed on the body of the patient after expansion. The serpentine design is composed of four different semi-circular lines in a repeatable pattern known as reliable stretchability against planar deformation25.

Fig. 4: 3D printed dry electrocardiogram (ECG) Electrode with conductive serpentines.
figure 4

Schematics of a 3D printing process of serpentine electrodes on the non-auxetic origami surface, and b the folding and unfolding behaviors of non-auxetic origami. Gap distance, d, is wider once the origami is compressed. The direction of force, f, during the folding is highlighted with red arrows. c Relation of gap distance with height changes of z-direction. d Optical microscopic images of the 3D printed electrodes with the different layered number, n. e The different strain values from 0 to 1.8 with the sample of n = 1 ~ 3 (R0 = 3.46 × 10−1 Ω). f The normalized impedance of the 3D printed serpentine electrodes with different printing layers.

In addition, full integration along the non-auxetic part and proper contact of the skin with the printed electrodes need to be secured. Therefore, the serpentine electrodes at the valley of the LIO are connected using a free-standing serpentine path, as shown at the bottom right of Fig. 4b. The distance (d) of the valley at the non-auxetic part increases when the LIO is in contact with the skin. The force direction of d has two directional elements; the length of d is proportionally varied at compressed Δh, as shown in Fig. 4c. At the fully expanded state of the non-auxetic part of the LIO, strain d reaches approximately 1.8. In Fig. 4d, the printed serpentine electrodes with a number of n = 1–3 layers are represented, and their conductivity based on the strain is tested, as shown in Fig. 4f. With n = 3 of the printed electrodes in Fig. 4d (ii), until the strain changes to 1.8, very few bisected or disconnected conductive paths are observed. Thus, it is concluded that the printed serpentine electrodes with n = 3 are appropriate for the ECG electrodes of the LIO sensor. In addition, the impedance of the electrodes with different n values is shown in Fig. 4e. At n = 1, the printed conductive path is disconnected at a strain of 0.15 such that the impedance is drastically increased to create an open circuit. At n = 2, the electrodes were reliable until the strain reached 0.8. Thicker electrodes are expected to show improved conductive reliability when stretched. Once a curved object is stretched, it fails by buckling under tensile loads39. Thus, the critical buckling force was applied to the object owing to the compressive stress perpendicular to the stretching direction39. In the serpentine design, the force is applied to the middle of the curvatures in the patterns. Thus, the design of a structure requiring a higher critical buckling force (σcr) and the global tensile stress to trigger failure by buckling can enhance stretchability. The critical buckling force of the stretched plate is described by Eq. 340

$$\sigma _{{{{\mathrm{cr}}}}} = {{kE}}\left( {{{t}}/{{w}}} \right)^2$$
(3)

where E, t, and w are the elastic modulus, thickness, and width of the plate, respectively. k represents the buckling factor, which is the ratio of length and width (l/w) with a fixed Poisson’s ratio. At n = 1, t/w is approximately 0.4, indicating a lower σcr, resulting in easier failure through buckling than the design with n = 3 (t/w = 1.3). In addition, when t/w > 1, the critical buckling force is predominantly affected by the elastic modulus of the materials, implying that the stretching behavior of the electrodes is converted to bending features instead of mechanical features related to buckling failure40. Thus, the electrodes with n = 3 can be reliably conductive until they are stretched to a strain of 1.8.

ECG sensing humanoid robot with LIO

The LIO with serpentine electrodes was developed for ECG sensors, the so-called LIO sensors. Its skin impedance has been measured. As shown in Supplementary Fig. 4, the skin impedance of electrodes with LIO is about 23% lower than that without LIO at all frequency range, indicating that there’s little contact resistance attributed to LIO folding behavior. As shown in Fig. 5a, two thimble-type LIO sensors with dry ECG electrodes were prepared on the fingertip. The LIO sensors were integrated into the middle fingers of the humanoid robot. Figure 5b shows the ECG sensing position and process of the humanoid robot, which is also presented in Supplementary Videos 3 and 4. By conformal contact with the human chest, the humanoid robot detects the ECG signals and displays the results on the tablet monitor. The summarized operating logic of the humanoid robot is shown in Supplementary Fig. 5. The voice input, ‘check’ is the triggering word to start the ECG sensing logic for the robot. After the detection of the word ‘check’, the robot first starts checking whether obstacles interrupting its arm raising are in front of the robot. The robot then performs the ECG sensing tasks if effective signals are obtained. After 30 s of ECG recording, the robot displays the results on the tablet monitor. Once the sensing fingers approach and push toward the chest, the LIO sensors expand and hold contact during the measurement of the ECG signals. No damaged electrodes were observed on the LIO sensors during the repeated measurements; this was attributed to the structural properties of the LIO-serpentine combination, as shown in Fig. 5c. The measured ECG signals are presented in Fig. 5d, compared with the wet (Ag/AgCl) electrodes. No significant difference exists between them; the voltage range for the ECG signals was similar, and the ECG signal peaks were highly distinguishable. Moreover, the signal-to-noise41 ratio of the measured ECG with the LIO sensor and conventional wet electrodes was 21.7 ± 0.56 and 18.1 ± 0.85 dB, respectively. Thus, it is concluded that the ECG sensing ability of the humanoid sensing robot is comparable to that of conventional wet-type ECG sensors. The repeatability and lifetime42 test have been conducted as shown in Supplementary Fig. 6. Although SNR and ECG output values show slight fluctuation, their values are reliable within the error range of 20 dB and 0.42 mV for the 60 sensing cycles as shown in Supplementary Fig. 6a, b, respectively. Moreover, after storing the LIO sensor in the ambient condition for 30 days, ECG output shows about 0.41 mV as shown in Supplementary Fig. 6c. However, the ECG output is decreased with longer storing time potentially because of dust or contamination of the electrodes.

Fig. 5: Electrocardiogram (ECG) monitoring sensing robot with LIO.
figure 5

a Schematic of the equipped LIO sensor for ECG measurement on the humanoid robot fingertip. b The actual images of the ECG sensing process through human–robot interaction. All photos are produced under the agreement of the participant. c The detailed images of the robot fingers near the contact. The ECG signals are collected near the heart of the human subject. d Comparison of ECG signals from the sensing robot and conventional wet ECG sensor. e Collected ECG signals before and after the exercise.

ECG monitoring with different exercise conditions is shown in Fig. 5e. Two different levels of exercise were tested by climbing 100 and 200 stairs to make the heart beat faster. Before climbing, a heartbeat rate of 86 bpm was detected in the human subject. After climbing 100 stairs, the heartbeat rate increased to 117 bpm. After climbing, more peaks of ECG signals are detected in the same time interval of the signals before climbing, resulting in the 5th peak at approximately 2.5 and 3.3 s for 117 and 86 bpm, respectively. Thus, the peak patterns of the ECG signals reflect the changes in bpm during exercise. After climbing 200 stairs, 122 bpm was measured from the human subject. As indicated by the red dotted line in the graph, the time for the 5th peak of its ECG signals is slightly shifted to 2.3 s. From the results, it can be concluded that the sensing robot detects distinctive ECG signals even with slight changes in the heartbeat. After the exercise, the human body cannot hold fixed gestures because of variable breathing. In this condition, the measurement of reliable ECG signals with a dry electrode is challenging without tightening supports or holders, such as a belt. Nevertheless, the ECG sensing robot detects the signal, which is attributed to the improved contact between the skin and the LIO sensors.

Blood-pressure monitoring humanoid robot

The performance of the paired sensing robot with ECG and PPG for BP monitoring is shown in Fig. 6. The BP was monitored using the LIO-based ECG sensing hand and PPG sensing hand, as shown in Fig. 6a. The underlying principle of BP monitoring is based on the relationship between the time taken for blood to travel from the heart to the distal position43. This “time delay” is utilized as the pulse arrival time (PAT)44. The PAT is calculated from the time difference between the R-peak of the ECG and a characteristic point of the PPG peak. The ECG sensing hand was positioned at the chest of the human subject for the ECG signal near the heart, and PPG sensing hands were placed on the distal side of the subject, as shown in Fig. 6b and Supplementary Video 5. Both sensing hands are shown in Fig. 6c. The time difference between the R-peak and PPG peaks was obtained from the very beginning of the pulse waves, as shown in Fig. 6d. The relationship between BP and PAT is described by Eq. 445

$${{{\mathrm{BP}}}} = \frac{1}{\gamma }\left[ {{{{\mathrm{ln}}}}\left( {\frac{{{{{\mathrm{2}}}}\rho {{DL}}^2}}{{aE_0}}} \right) - 2{{{\mathrm{ln}}}}T} \right]$$
(4)

where T, E0, and γ are the PAT and the coefficient of Young’s modulus for the artery with and without BP, respectively. ρ, a, D, and L are the density of blood in the artery, the wall thickness, the diameter of the artery, and the length between the heart and the distal site. Equation 4 can be simplified to Eq. 54

$${{{\mathrm{BP}}}} = \beta _0 \times {{{\mathrm{PAT}}}} + \beta _1$$
(5)
Fig. 6: Blood pressure (BP) monitoring sensing robot system.
figure 6

a Schematic of the robot hand equipped with PPG sensor on the left and ECG sensor on the right. b The actual images of the sensing process through human–robot interaction for BP measurement. All photos are produced under the agreement of the participant. c Their actual photos of ECG (i) and PPG (ii) sensing robot hands. d The overlaid ECG and PPG graphs to obtain pulse arrival time (PAT). e The diagrams of monitored systolic blood pressure (SBP) obtained from a sphygmomanometer with their corresponding PAT. From the diagram, the linear fit model is used to obtain the estimated SBP. f Comparison of measured and estimated SBP and diastolic blood pressure (DBP) at different postures or exercises (N: Normal, P: Push-up, R: rest, Sq: Squat, Fw: fast walking, and Sp: Sprint).

A least-squares algorithm is usually implemented to estimate the unknown coefficients β0 and β1, which is a calibration process4,7. In the calibration process, the systolic blood pressure (SBP) and diastolic blood pressure (DBP) of the subjects were investigated using a cuff-based digital sphygmomanometer over the brachial artery of the left upper arm. Simultaneously, ECG and PPG sensing of the subject were conducted for improved time synchronization. The measured PAT at different SBP and DBP values are plotted in Fig. 6e and Supplementary Fig. 7a, respectively. The unknown coefficients of Eq. 5 were calculated from the scatterplots.

$${{{\mathrm{SBP}}}} = - {{{\mathrm{50}}}}{{{\mathrm{.23PAT}}}} + {{{\mathrm{168}}}}{{{\mathrm{.12}}}}$$
(6)
$${{{\mathrm{DBP}}}} = - {{{\mathrm{42}}}}{{{\mathrm{.04PAT}}}} + {{{\mathrm{116}}}}{{{\mathrm{.42}}}}$$
(7)

The coefficients of determination for Eqs. 6 and 7 were 0.948 and 0.890, respectively. From the equations, we obtained the estimated SBP and DBP once the PAT was measured from the subject. The Bland–Altman plot46,47 was used to analyze the validity of the measured and estimated BP in Supplementary Fig. 7b, c. The x-axis of the plots indicates the average of the estimated and measured BP, and the y-axis represents the difference between them. The mean difference for SBP and DBP are –0.03 and 0.00 mmHg, respectively (≤5 mmHg, based on the Association for the Advancement of Medical Instrumentation4,47 (AAMI) criteria). The standard deviation of 2.13 and 2.58 mmHg are obtained for SBP and DBP, respectively (≤8 mmHg, AAMI criteria), summarized in Supplementary Table 24,47. Thus, we concluded that the equations from the calibration process are valid for the measurement of the additional BP monitoring of the subject under different exercise conditions. However, more calibration processes must be conducted with more subjects for the universal utilization of BP monitoring. The estimated and measured BP values under different exercise conditions are presented in Fig. 6f. BP monitoring with four different exercises was performed. First, BP monitoring immediately after the push-up case noted as P of the subject showed an increasing trend of the estimated SBP and DBP from 135 to 140 mmHg and 87 to 91 mmHg, respectively. Simultaneously, the measured BP values from upper-arm monitoring show SBP from 136 to 139 mmHg and DBP from 86 to 93 mmHg, respectively. Next, the BP was measured with three different exercise conditions: squat exercise (Sq), fast walk (Fw), and sprint (Sp). Drastic increases in estimated SBP and DBP after Sq, Fw, and Sp ranging from 150 to 167 mmHg and 101–116 mmHg, respectively was observed. By comparison, the measured BP with the upper-arm monitoring device showed 149–166 and 99–115 mmHg for SBP and DBP, respectively. These results imply that the BP estimated using the sensing robot is compatible with cuff-measured BP under different exercise conditions. Moreover, the PAT includes a pre-ejection period with pulse transit time (PTT)3,47,48. The pre-ejection period is the time taken to convert electrical signals into a mechanical pumping force of the heart and isovolumetric contraction to open the aortic valves24,48. We demonstrated BP monitoring with strenuous exercises such as Sq, Fw, and Sp, which are significantly related to ventricular contraction and vascular function, as the reasons for using PAT instead of PTT49,50. Thus, reliable BP monitoring of the LIO sensor was possible with the sensing robot.

We developed a bio-inspired origami sensor for BP monitoring sensing robot applications. By pairing PPG with ECG sensing hands, cuff-less BP monitoring is demonstrated by the human–robot interaction of the sensing robot. BP monitoring robot systems based on PAT recording require reliable ECG signals near the heart and PPG signals at the distal site. In addition, frequent measurement with less labor required by healthcare workers is in high demand. Thus, an ECG sensor that enables stable positioning of dry ECG electrodes on the skin is essential to obtain effective ECG signals and develop a BP monitoring robot system. Based on the advantages of origami, LIO has been developed for BP monitoring robot applications. Such nature-inspired origami allows the improved contact of ECG sensing dry electrodes with the skin and BP monitoring even during the out-of-breath condition of the subject without belts or adhesives. This feature is achieved by the simultaneous dynamics of the auxetic and non-auxetic samples in a single monolithic structure, which is the unique advantage of origami. In addition, its mechanically tunable properties allow for additional parametric studies for optimal LIO design. Moreover, the structural properties of serpentine-shaped ECG dry electrodes complement LIO design, contributing to stable conductivity during conformal contact with the skin of the LIO sensor. Thus, the LIO sensors show reliable ECG sensing even under out-of-breathing conditions of the human body without additional belts or tapes after being integrated into the humanoid robot. In addition, the sensing robot with LIO was implemented for paired detection of ECG and PPG for BP monitoring under different exercise conditions.

Cuffless BP monitoring is widely used because it is more convenient for patients. In addition, the demand for home health aides has increased owing to the aging population and the prevalence of heart-related diseases. Moreover, the previous and current COVID 19 pandemics provide a lesson on the importance of the remote monitoring system for BP. Consequently, the application of sensing robots for cuffless BP monitoring is a remarkable system for remote healthcare. Beyond mere healthcare, these can be a future platform or bridge between medical personnel and remote patients. Furthermore, toward continued innovation and development as the next generation of sensors, biomedical sensing robotic systems have the potential to play an essential role in the new era of remote healthcare. Although we conclude that our BP sensing robot would have a reliable performance in defining BP values, further investigation is required to make a biomedically meaningful implementation. Beyond such an intensive view of the biomedical area, the development of a human–robot interactive BP monitoring system can be a helpful tool to facilitate intensive research on human hypertension.

Methods

Research ethical approvals

All procedures to perform the studies of the bio-medical sensing robot have been approved by the research ethics committee at Simon Fraser University in Canada: Minimal Risk Approval 2020s0193. A male volunteer with no prior medical history of chronic cardiovascular, skin, mental health disease, or physical disability was recruited for participation in this test. The signed consents including consent of photography and videos during the test and consent to be filmed and the footage to be used for publication were obtained from the individual. According to the legal term of Canada, the confidentiality of participants is strictly respected.

Preparation of origami sensors

The fused filament fabrication (FFF) 3D printer with a direct-drive type (Tl-D3 pro, Tenlog, ltd) is used to prepare all origami designs. The filament for the 3D printer is commercial Ninjaflex TPU85A (Fenner Drivers, Inc). The designs of origami structures are prepared using Solidworks, a 3D CAD file, and Cura slicer program (Ultimaker, Ltd.) to generate their g-codes for the 3D printing process. The nozzle size of the hot end is 0.2 mm. The print speed, infill, layer height, temperature, and width are 30 mm s−1, 0.1 mm, 223 °C, and 0.1 mm, respectively. The commercial co-polyester filament (ColorFabb, ltd.) is used for the sensor frames, and the FFF 3D printer with the bowden tube type (Ultimaker 3) is utilized to print. The g-code preparation is the same as above.

Characterization of the 3D printed origami sensor

Mechanical properties of the origami designs were investigated using a universal testing machine, Shimadzu EZ-LX. The compression and pull-off tests were conducted at a loading rate of 2 mm min−1. For the parametric studies of the origami design, the loading force is applied by using a universal testing machine, and all of the necessary information such as length, width, height, and thickness are measured by using a vernier caliper very carefully. For the pumping behavior of the origami design, as shown in Supplementary Video 2 and Fig. 1f, the small hole of the origami design on the bottom is fully sealed by using commercial silicone rubber (Eco-flex 00-30, Smooth-on, Inc.).

Preparation and characterization of serpentine electrodes

The conductive silver paste for serpentine electrodes was prepared by dissolving 0.4 g polyurethane into a blend (1 ml N, N-dimethylformamide (DMF) with 4 ml tetrahydrofuran (THF)) by SpeedMixer DAC150.1 FVZ-K (FlackTek, Inc.), followed by adding 2 g silver flake (Inframat Advanced Materials, US). After thoroughly mixing, the viscosity of the silver paste is controlled by heating at 65 °C for the ambient time. Then, serpentine electrodes are programmed and fabricated by a direct writing printer SHOT mini 100Sx (Musashi Engineering, Inc.) with 1 mm s−1, followed by cured at 80 °C for 1 h. The stretching behavior of the serpentine electrodes is monitored by using an optical microscope (OT-HD, Opti-tekscope, ltd). For the impedance of the electrodes, the electrodes are printed on the flexible plate prepared by using an FFF 3D printer. A pair of the conductive path is connected to digital multimeters and monitored the impedance changes as a function of stretched distance. Also, the skin impedance of dry serpentine electrodes with and without LIO was measured using an LCR meter (Agilent, E4980A, ltd). The measurement was conducted in the frequency range from 20 Hz to 400 kHz.

ECG monitoring

The LIO sensors are equipped to the two fingers of the humanoid robot. The logic for the programming is represented in Supplementary Fig. 5. The ECG sensors are connected to a commercial ECG sensing system (Pulsebit EX, Wellue Ltd.). Also, an entire ECG system is integrated onto Softbank’s Pepper humanoid robot. The duration to record the ECG signals of the human subject is 30 s for one trial. The ECG sensing system has ±3 mV of the input dynamic range and 0.05–40 Hz bandwidth. After measuring the ECG signals by the system, the recorded signals are automatically transferred to the robot via wireless Bluetooth. The prepared LIO sensor, a LIO design with dry serpentine electrodes, is used for the humanoid robot to measure the ECG signals. For the comparison, the conventional wet (Ag/AgCl) electrodes with an ECG sensor are used with the same sensing system at the same position.

Blood pressure monitoring

A commercial PPG sensor (MAX30102, Maxim integrated, Ltd.) is paired with the developed LiO sensor to integrate onto Softbank’s Pepper humanoid robot. The LIO sensor fingertip on the sensing hand (left hand) of the robot is positioned on the subject’s chest, and the PPG sensing hand (right hand) is touching the subject’s fingertip. The waveforms obtained from ECG and PPG sensing hands are collected simultaneously for time synchronization. At the same time, a cuff-type digital sphygmomanometer is employed to obtain systolic and diastolic BP for reference from the left upper arm. For the BP measurement from ECG-PPG sensing hands, the calibration process of BP for the subject is carried out with the linear square model in Supplementary Table 1. The PAT is calculated based on the time difference between two R-wave and systolic wave peaks from ECG and PPG signals.

Ethics information for ECG and BP monitoring

A 36-year-old male student is employed as the human subject for ECG and BP monitoring. His height and weight are 178 cm and 88 kg, respectively. Dry LIO sensors of the humanoid’s fingertip are positioned on his chest for ECG measurement. The subject has performed two different exercises climbing 100 and 200 stairs for the ECG measurement. The subject has carried out four different exercises: 20 times of push-up, rest for 20 min, 30 times of squat, fast walking, and a sprint of 50 m for the BP monitoring with varied heart rates. All tests have been performed after getting written permission from the subject and the approval of research ethics. All Supplementary Videos are produced under the agreement of the participant.