Introduction

Wearable biomedical devices have witnessed a growing interest in recent years. Indeed, these devices allow long-term non-invasive monitoring of people in the comfort of their daily routine1,2,3, providing information that is valuable to enable early diagnosis and effective treatment of patients, or that can be used to support a healthy lifestyle. Biomedical wearables typically aim at measuring vital signs or the bioelectric activity of different organs, such as the brain, the muscles and the heart. These signals will be named in the context of this paper as “bio-signals”.

Cardiovascular diseases are worldwide the first cause of death, causing 31% of all fatalities every year4. A valuable tool to diagnose these diseases is the electrocardiogram (ECG), a non-invasive surface measurement of the heart’s electrical potentials. To perform ECG measurements, patients are generally connected with wires to a bulky biopotential acquisition system. This makes long-term monitoring impossible. Wearable solutions for continuous monitoring of ECG would thus be particularly valuable to diagnose various cardiovascular diseases in an early stage. For this reason, the focus of this work is on a wearable patch to measure ECG.

Continuous monitoring of physiological signals require wearable devices to be conformable to the shape of the body, in order to improve the quality of physiological signal monitoring and to ensure maximum comfort of use. Advancements in fabrication techniques and new materials that make possible to build flexible and conformal electronics are thus an important enabler to build wearables for biomedical applications5,6,7.

Many flexible materials including polysilicon8, hydrogenated amorphous Silicon (a-Si:H)9, organic semiconductors10,11, and metal oxides12 have been investigated to serve as semiconductors in thin film transistors (TFTs) in the last few decades. Hosono’s group gave a fundamental contribution to the field of flexible TFTs demonstrating amorphous indium gallium zinc oxide (a-IGZO) TFTs deposited on a flexible polyethylene terephthalate substrate at room temperature13. These a-IGZO TFTs14,15,16 fabricated on flexible large-area substrates are a very interesting platform to build wearable bio-signal acquisition systems due to their flexibility, lightweight, and low cost.

Present wearable bio-signal monitoring devices can be broadly classified in three categories:

  1. (1)

    Wearable devices in which all active components are rigid Silicon integrated circuits (ICs). These devices are built by encapsulation and packaging techniques exploiting flexible or elastic layers, by integration in fabrics and/or by using flexible interconnects. These wearables can implement complex electronic functionality thanks to the Silicon ICs, at the cost of a reduced comfort level for the user, due to the presence of many rigid parts. Examples of such IC-based wearables include ECG sensors with electrodes integrated into a cotton T-shirt17, cardiac sensors where a rigid IC is encapsulated in elastic layers18,19, stretchable “e-tattoos” for ECG and temperature monitoring20, and a portable electrocorticography (ECoG) system with flexible microelectrodes array21. Multimodal bio-signal acquisition has been demonstrated too, with stretchable sensors capable of monitoring several combinations of electroencephalograpgy (EEG), ECG, electromyography (EMG), electrooculography (EOG), and photoplethysmography (PPG)22,23,24.

  2. (2)

    Wearable devices in which sensors and/or front end elements are based on flexible materials/substrates while the rest of the system is implemented with rigid Silicon ICs. In this type of hybrid solutions, flexible electronic devices and sensors are used for interfacing with the human body, while rigid electronics (conventional Silicon ICs) are used for the signal processing and communication functions such as amplification, filtering, digitization and data transmission. In this way, the flexibility and conformability of the flexible substrate are combined with the good signal acquisition and processing capability of rigid components. In this approach, however, the manufacturing steps and costs increase, due to the processing required to combine flexible and rigid parts. Besides, the rigid components may result in reduced patient comfort and in disturbance to the measurements due to movements. Examples of this category of wearable sensors include screen-printed electrodes for ECG25,26, electrophysiological sensing arrays27, self-powered ultra-flexible electronics using organic electrochemical transistors28, implantable sensors for the brain29, flexible printed electrodes for EEG30, a pulse oximeter based on organic materials compatible with flexible substrates31, an ultrasonic device conformable to the skin and capable of monitoring blood pressure32, and epidermal electronics for multimodal physiological monitoring33.

  3. (3)

    Wearable devices in which all components are built using flexible electronics. The materials exploited in these flexible devices are lightweight and fully conformable to human body, ensuring maximum user comfort. Examples of such wearable sensors include flexible ECG monitoring devices34,35,36, a prosthetic skin with stretchable multi-electrode arrays for nerve stimulation37, organic electrochemical transistors for ECoG monitoring38, large-area flexible or stretchable sensors able to detect arterial pulses39,40, and an epidermal electronic system capable of monitoring ECG and EMG41. Although all these examples show the feasibility of flexible or conformable wearable devices, the inherently poor electric performance of flexible TFT technologies results in limited electronic functionality for this class of wearables. For instance, all smart bio-signal sensors based only on flexible components lack digitization of the measured signals and data transmission capability34,35,36,37,38,39,40,41.

In an effort to advance the state of the art of wearable bio-signal monitoring devices exclusively based on flexible electronics, we report here for the first time (to the best of the authors’ knowledge) a flexible ECG sensor which exploits TFT-based circuits on flexible foil for the full chain of signal acquisition, amplification and digitization34,35,36,37,38,39,40,41 and is capable to measure ECG signals captured by a pair of electrodes with sufficient signal to noise ratio42 (Fig. 1). Additionally, the digital output is provided as a serial 105.9 kb/s Manchester-encoded bit stream compatible with the NFC wireless transmission standard, so that the measured data could be sent to a smart phone for further elaboration and presentation. The absence of any rigid active component makes the system conformable to the human body and comfortable to use. Moreover, the implementation of the full system with flexible TFTs reduces the extra processing steps and the cost required for the integration of rigid and flexible electronic components, enabling a solution that could become compatible with single use.

Fig. 1: Application scenario.
figure 1

The presented ECG patch in a representative application scenario.

Results

System architecture

A biopotential acquisition system typically consists of two main blocks (Fig. 2a): an analogue frontend (AFE) amplifies the signal while ensuring a suitable noise level, to preserve the shape of the signal, and provides sufficiently high impedance to enable reading the voltage present on the skin with appropriate electrodes. An analogue to digital converter (ADC) follows, to convert the biopotential signal to a digital representation, enabling robust wireless transmission and/or further signal processing. As shown in Fig. 2a, the ADC consists typically of a comparator, a digital to analogue converter (DAC) and suitable logic. To minimize the influence of temperature and technology variability, both the AFE and the ADC are typically designed so that their signal transfer function is determined by the ratio of similar components (e.g. resistors). If variations affect all components proportionally to their value (e.g. the resistance per square increases by a given amount), these ratios tend to stay unchanged. Moreover, in typical Silicon designs, the DAC uses a reference voltage generator (Fig. 2a). This reference voltage is very stable against process, supply and temperature (PVT) variations. The DAC uses this reference to produce output voltages that are also stable against PVT. At the state of the art, it is difficult to make a voltage reference in a-IGZO, because the main components used for this purpose (diodes and bipolar junction transistors) are not available in a-IGZO processes.

Fig. 2: System architecture.
figure 2

a Conventional biopotential acquisition system, b System-level diagram of the ECG acquisition system presented in this work, and c Detailed architecture of the presented system.

For this reason, the design presented in this paper takes a radically different approach, and digitizes the input data in the time domain. The concept is shown in Fig. 2b. After the AFE a reset integrator is used to transform the signal amplitude in duty cycle information (pulse-width modulated, i.e. PWM representation). This means that the duty cycle of the signal produced by the integrator is proportional to the input signal. Digitization is then simply achieved by counting for how many clock cycles the integrator output is high every period. In this case a time reference is needed for digitization, as shown in Fig. 2b. A time reference can be easily generated in our system, which is meant to communicate to a smart phone via NFC. Indeed the NFC carrier can be received on foil and its 13.56 MHz frequency used as time reference, after suitable division in the digital domain43.The approach here described provides a robust digitization strategy, without the need for a voltage reference, and thus is particularly suited to implementation with a-IGZO TFTs on foil.

Circuit design

A more detailed schematic of the proposed flexible patch is provided in Fig. 2c. The system consists of two plastic foils (Fig. 1) on which a-IGZO TFTs are fabricated. The first foil (analogue subsystem) is connected to the electrodes. The AFE captures the ECG signals, and the reset integrator transforms them to a PWM representation. The second foil (digital subsystem), connected to the first one (analogue subsystem), digitizes the PWM signal and serializes the output bits, so that they can be sent to a receiver using radiofrequency (RF) NFC communication.

The analogue frontend (AFE) consists of an input chopper (Fig. 2c, CHin), a two-stage cascaded diode connected load preamplifier (Fig. 2c), a de-chopper (Fig. 2c, CHout), and a low-pass filter (Fig. 2c, LPF). The input signal is AC-coupled before the input chopper, to cancel the large DC electrode offsets44. Due to the low-frequency bio-signals, chopping is used to supress the amplifier flicker noise and offset. The chopping frequency is chosen as a trade-off between noise, preamplifier bandwidth (which has an impact on power consumption) and input impedance. The preamplifiers are designed to have a bandwidth larger than the target chopping frequency, which is 1 kHz. They are built using diode-connected load TFTs, so that their gain is depending on the ratio of the transconductance of input and load devices, improving the robustness to process variations. The detailed schematics of each preamplifier is shown in Fig. 3a. The input TFTs M1,2 have large (15,000 µm2) area to bring the 1/f noise corner close to the chopping frequency. For the diode load transistors M3 and M4 (Fig. 3a), the back gates are connected to the output nodes to increase the output resistance, while the input transistors M1 and M2 have back gates connected to their front gates to increase transconductance and improve noise efficiency. To avoid excessive noise aliasing, the output of the preamplifier is low-pass filtered by another differential amplifier with diode connected load, which has a nominal 100 Hz bandwidth (Fig. 2c, LPF).

Fig. 3: Circuit schematics and working principle.
figure 3

a Schematics of each of the two cascaded diode-load amplifiers in the preamplifier, b Reset integrator, and c Reset integrator working principle.

The AFE is followed by the reset integrator (Fig. 3b). The latter consists of a differential input pair M5,6 with load capacitors CL. Here a two-phase integration, where in the first phase ϕ the capacitor discharge is signal-dependent, and in the second phase \(\overline \phi\) the capacitors are discharged by a constant current, is applied (Fig. 3c). The instants at which the integrator outputs are discharged below a given threshold Vref is detected by two external comparators. Thanks to the constant slope observed in the phase \(\overline \phi\) the time walk difference between the two comparators is minimized and the linearity of the voltage to duty-cycle transformation is improved. Besides, the low-pass behavior of the reset integrator improves the signal to noise ratio of the whole chain. The time delay in threshold crossing between the two branches of the integrator output, Tout, is proportional to the input signal (Fig. 3c). The comparator outputs are then fed to a XOR gate to obtain a PWM representation of the input signal with an additional sign bit. The entire analogue subsystem is extremely minimalistic in terms of transistor count, which benefits to the circuit yield and to the power consumption.

From Fig. 3c we can find the relationship between the pulse width of the integrator output, Tout and the input signal Vin, as:

$$\frac{{T_{{\mathrm{out}}}}}{{V_{{\mathrm{in}}}}} = \frac{{{{{\mathrm{Ag}}}}_{\mathrm{m}}\phi }}{{I_{{\mathrm{b2}}}}}$$
(1)

where, Tout is the output pulse width, Vin the differential input signal, A the gain of the two stage cascaded amplifier, gm the transconductance of the input transistor in the integrator, ϕ the integration period, and Ib2 is the discharging current during the phase \(\overline \phi\).

In the digital subsystem (Fig. 2c), a clock divider is used to generate timing signals from a 13.56-MHz source (CLK). The PWM output is clock gated (and thus quantized in time) with a clock of frequency CLK/32 and then fed to an asynchronous counter which creates a digital multi-bit representation of the PWM output. The typical ECG signal has an amplitude of 1 mV, the target equivalent input noise is in the order of 10 µVrms. Adding a margin of 10 dB to make sure that the quantization noise is non-dominant, and additional 10 dB to accommodate for larger signals, one obtains a target dynamic range for the analogue to digital conversion of 60 dB. This necessitates a 10-bit conversion resolution and hence a 10-bit counter has been selected for our first prototype. The counter outputs are then serialized and Manchester-encoded, generating a 32-bit serial stream including a 12-bit preamble, 10 counter bits and other fixed payload. The serializer and Manchester encoder are driven with a clock of frequency CLK/128, to provide the 105.9 kb/s output bit stream which is compatible with NFC RF transmission. The clock phases needed to drive the analogue subsystem are also generated on foil (Clock Phase Generator).

Electrical characterization

Figure 4a shows the measured transfer curve of the cascaded preamplifier. It achieves a gain of 23.5 dB in 3.7 kHz bandwidth with 12.5 μA current consumption from 10 V supply. The integrated input-referred noise in the 1–100-Hz BW is 8 μVrms with a chopping frequency of 1 kHz.

Fig. 4: Measurement results.
figure 4

a Measured frequency response and input-referred noise of the cascaded preamplifier, b Measured transfer function of the full analogue chain (from Input to Output PWM), c In vivo ECG measured signal reconstructed from the output PWM, d In vivo ECG measured signal reconstructed from Manchester Encoded output, e Manchester Encoded output corresponding to three samples of the in vivo measurement and a zoom-in of the Manchester encoding of the output bits.

Figure 4b shows the measured transfer curve of the full analogue chain, including the reset integrator and PWM generation. The low frequency gain is 31.3 ms/V. Notice that the notches are at multiples of the inverse of the integration period (555 Hz) and thus do not affect the target signal bandwidth. The system is powered using a battery and the total power consumption of the analogue front end is 280 μW from a 10 V supply while it is 15.4 mW for the digital subsystem. The system achieves experimentally 67.4 dB CMRR, 58.9 dB PSRR, and 16.5 MΩ input impedance at 50 Hz with 1 kHz chopping. The achieved spurious free dynamic range (SFDR) is 43.11 dB (Supplementary Fig. 1). The linear relationship between the digital code measured at the output of the digital part and the PWM pulse width is shown in Supplementary Fig. 2.

In vivo characterization

Figure 4c shows an in vivo experiment measuring the cardiac activity of a person, obtained by connecting our system to the test subject using standard wet electrodes placed on the chest. A third body electrode (which for simplicity is not shown in Fig. 1) acts as reference connected to the ground. The signal obtained by low-pass filtering the output PWM spectrum in the band of 1–200 Hz is displayed in Fig. 4c. Figure 4d shows the reconstructed signal obtained by the Manchester-encoded serial digital output.

In Fig. 4e a segment of the 105.9-kb/s Manchester-encoded bit stream corresponding to the in vivo measurement is displayed. The bit stream corresponding to the digitization of one sample is repeated three times in order to reduce errors in the NFC transmission. The sampling frequency of the input signal is 250 Hz, while the Manchester-encoded bit stream is generated at 105.9 kb/s, hence, there is sufficient time to repeat each sampled value three times.

Figure 5 shows the foil micrograph of the presented system. The total area of the foil is 123.4 mm2.

Fig. 5: Foil micrograph.
figure 5

Foil microgrpah of the presented system with dimensions.

Discussion

Table 1 provides a performance summary and comparison to prior literature. This work is the first reported circuit on foil capable to convert a biopotential input to the digital domain. Additionally, it generates a 105.9-kb/s Manchester-encoded bit stream compatible with the NFC standard. Some of the initial steps taken in the direction of bio-signal acquisition systems using flexible TFT technologies have been published in literature for EEG45, EMG46,47, ECG48 and Heart Rate (HR)49 measurements. Comparing our contribution to existing literature on flexible TFTs circuits, in references45,46,47,48 the signal is not digitized. A recently published frontend48 is able to acquire ECG using a flexible TFT-based amplifier, but it also does not digitize the signal. An HR readout on flexible foil49 was able to convert the signal to binary levels, encoding it in the transition times (i.e. PWM output representation), but was lacking a time-sampled digital output representation (i.e. in bits), which is needed to be compatible with most modern wireless transmission systems. Compared to49, we have included in this work a digital subsystem (consisting of a 10 bit counter, serializer with Manchester encoder, clock divider and digital phase generator) which transforms the PWM to a time-sampled digital format. The digital clock signals needed to operate the full system are also generated on the foil. Apart from this, the integrated input noise in49 was 186.3 μVrms which is too high to perform ECG measurements, enabling only HR measurements. Within the analogue frontend we have thus implemented in the present work a low-pass filter after the cascaded preamplifier, which limits in band noise and noise aliasing, and enables 8 μVrms input equivalent noise in the 1–100-Hz bandwidth: a performance that is now compatible with ECG measurements. The input signal is transformed to PWM without digitization in50 too, and that circuit does not provide a voltage, but rather a current input, making it unsuited to measure biopotential signals. Biopotential amplifiers are typically characterized by a figure of merit called Noise Efficiency Factor – NEF51:

$${\mathrm{NEF}} = V_{{\mathrm{ni}},{\mathrm{rms}}}\sqrt {\frac{{2I_{{\mathrm{tot}}}}}{{\pi .U_{\mathrm{T}}.4{\mathrm{kT}}.{\mathrm{BW}}}}}$$
(2)

where, Vni,rms is the total input-referred noise, Itot is the total supply current of the amplifier, BW is the amplifier −3 dB bandwidth, and UT the thermal voltage. As it can be seen from (2) the NEF does not take into account the supply voltage. For this reason, to compare circuits having a different supply voltage VDD, the Power Efficiency Factor PEF was introduced in52.

$${\mathrm{PEF}} = {\mathrm{NEF}}^2V_{{\mathrm{DD}}}$$
(3)
Table 1 Comparison with the state of the art.

The PEF can be used to characterize the trade-off between noise and power consumption, and can be used to compare the power efficiency of different bio-signal acquisition front-ends operating at different supply voltages. In this work, we achieved experimentally NEF and PEF equal to 109.81 and 1.20 × 105 respectively. These figures provide an improvement of 13 and 86% in NEF and PEF respectively, compared to the best NEF and PEF reported for previous bio-signal acquisition front-ends implemented using flexible TFTs45. These advancements are obtained due to both, a better TFT technology and circuit improvements. The self-aligned IGZO technology used in this work has indeed a comparatively lower Hooge factor, thus providing less 1/f noise. Besides, the larger input transconductance obtained thanks to the use of the double gate TFT architecture, benefits the noise and power efficiency of the front-end.

This contribution shows the fundamental steps to demonstrate intelligent plasters for biomedical applications based on flexible electronics which are compatible with the NFC standard. The analogue frontend has a measured input-referred noise of 8 μVrms in the 1–100 Hz band, which is the lowest input-referred noise reported in a-IGZO TFT circuits so far5,6,7,47,49,50,53. The system achieves experimentally 67.4 dB CMRR, 58.9 dB PSRR, and 16.5 MΩ input impedance at 50 Hz with 1 kHz chopping. The output is transformed to the digital domain and provided as a 105.9-kb/s Manchester-encoded serial bit stream. To the best of authors’ knowledge, this is the first demonstrated ECG acquisition system in flexible IGZO TFTs on foil providing a digital serial output bit stream compatible with the NFC standard5,6,7,45,46,47,48,49,50,53. Power consumption is 15.4 mW for the digital and 280 μW for the analogue part. The NEF and PEF are 109.81 and 1.20 × 105 respectively, which are the best values reported to date in literature for IGZO TFT circuits5,6,7,45,46,47,48,49,50,53.

Methods

Technology and TFT electrical characterization

The presented system is fabricated using double gate self-aligned TFTs with a-IGZO53 as semiconductor, which results in n type TFTs only. The schematic cross-sectional view of a-IGZO TFT is shown in Fig. 6a. The layer stack consists of plastic substrate, a buffer layer, back gate (BG) metal M0, back gate dielectric, a-IGZO, top gate dielectric SiO2, top gate (TG) metal M1, interlayer dielectric, and source/drain (S/D) metal M2. The fabrication process starts with the deposition of a buffer layer on a polyimide plastic substrate. Then a BG metal (MoCr) is deposited by physical vapor deposition (PVD). This step is followed by deposition of back gate dielectric (SiO2) by plasma enhanced chemical vapor deposition (PECVD) at 350 °C. The next step involves the deposition of a-IGZO layer (20 nm with 10% ratio of O2/Ar) by DC sputtering and patterning with wet etching. After that, the top gate dielectric (SiO2) is deposited using PECVD at 250 °C, which is followed by TG metal (Mo) layer deposition. Afterwards, the gate stack including metal layer and dielectric is patterned using wet and dry etch in a combined step. Then an interlayer dielectric (SiNx) is deposited by PECVD at 250 °C. Patterning of contact vias using dry etch follows. In the same step BG, TG and S/D are contacted. The last step involves deposition of S/D (Ti-Al-Ti) metal using PVD and patterning by dry etching. Finally, annealing of the samples takes place at 240 °C in N2 for 1 h.

Fig. 6: Device characterization.
figure 6

a TFT cross-sectional view, b Id-Vgs transfer curves of TFTs, c VT distribution and, d Mobility distribution in a sample of 43 test TFTs.

The flexible polyimide substrate for the TFT process is hold on glass during manufacturing and released afterwards. The circuits realized in this way can be bent without significantly modifying their electrical characteristics54.

The transfer characteristics of 43 TFTs, each having aspect ratio of 10/5 (µm/µm), are shown in Fig. 6b. Figure 6c shows threshold voltage (VT) distribution of these TFTs. The median VT is −0.36 V while the threshold standard deviation is 0.21 V. Figure 6d shows the mobility distribution of the TFTs, with has a median of 14.47 cm2/Vs and a standard deviation of 1.30 cm2/Vs.

Electrical characterization setup

To characterize the preamplifier (Fig. 4a), the device under test has been hooked up via a suitable mechanical connector (Yokowo DS CCNL-050-47-FRC) to off-the-shelf voltage buffers (TI OPA445AP), followed by an instrumentation amplifier (TI INA111AP) for differential to single ended conversion. The output of the instrumentation amplifier is connected to the HP35670A Dynamic Signal Analyser to measure the frequency response and the output noise. A 10-mVp–p sine wave with 5V DC bias, and a 5V DC reference were applied to the positive input (Vin+) and negative input (Vin−) of the preamplifier (Fig. 3a), respectively. The supply voltage is 10 V. The gain of the preamplifier measured over six samples varied in the range of 23.1–27.5 dB, with a bandwidth between 3.1 and 4.3 kHz, as shown in Supplementary Fig. 3.

For the full analogue chain characterization (Fig. 4b), a 10-mVp-p sine wave with varying frequency was applied to the input of the system and the output PWM was observed. The output data are acquired using a Digilent® Analog Discovery 2 board. The ratio of the fundamental of the filtered output PWM to the applied input provides each point of the frequency response recorded for the full analogue chain (from input to output PWM).

The in vivo measurements are performed connecting our system to the test subject using wet electrodes (Kendall’s H124SG) placed on the chest with a spacing of ~5 cm. A third body electrode acts as the ground electrode. The output data are acquired using a Digilent® Analog Discovery 2 board for a period of 10 s. The output spectrum is filtered using MATLAB® with a Butterworth filter in the band 1–200 Hz. The ECG signal is reconstructed by dividing the output signal by the previously recorded full chain gain, to obtain the input referred ECG signal. In vivo measurements from six samples are shown in Supplementary Fig. 4.