Abstract
Laser spectroscopy outperforms electrochemical and semiconductor gas sensors in selectivity and environmental survivability. However, the performance of the state-of-the-art laser sensors is still insufficient for many high precision applications. Here, we report mode-phase-difference photothermal spectroscopy with a dual-mode anti-resonant hollow-core optical fiber and demonstrate all-fiber gas (acetylene) detection down to ppt (parts-per-trillion) and <1% instability over a period of 3 hours. An anti-resonant hollow-core fiber could be designed to transmit light signals over a broad wavelength range from visible to infrared, covering molecular absorption lines of many important gases. This would enable multi-component gas detection with a single sensing element and pave the way for ultra-precision gas sensing for medical, environmental and industrial applications.
Similar content being viewed by others
Introduction
Ultrasensitive detection of molecular gases has important applications in atmospheric and planetary science1,2, breath diagnostics3, and combustion monitoring4. Comparing with semiconductor and electrochemical sensors, laser absorption spectroscopy (LAS) offers better selectivity5. The sensitivity of LAS depends on absorption path length and line strength that is weak in the near infrared (NIR) and much stronger in the mid-infrared (MIR). With tunable diode laser absorption spectroscopy (TDLAS) and multi-pass Herriott cells to achieve long effective path length, sub-ppm (parts-per-million) and ppt (parts-per-trillion) level noise-equivalent concentration (NEC) have been achieved, respectively, in the NIR and MIR6,7. One of the most sensitive spectroscopic techniques is the noise-immune cavity-enhanced optical heterodyne molecular spectroscopy (NICE-OHMS), which could achieve ppq (parts-per-quadrillion) level NEC in the NIR8,9. However, there is a trade-off between sensitivity and dynamic range. With a cavity finesse of 30,000, the dynamic range is limited to about four orders of magnitude10. In addition, for cavity-enhanced systems with long effective path lengths (e.g., several kilometers for NICE-OHMS), complex cavity-locking using, for example, the Pound-Drever-Hall technique is needed and it would be difficult to maintain the performance over a long period of time in a real-world environment.
Photoacoustic spectroscopy (PAS) and photothermal spectroscopy (PTS) measure, respectively, the pressure and refractive index (RI) change owing to optical absorption. PAS does not require a long optical path length and has demonstrated ppt-level NEC for acetylene by combining an optical power build-up cavity with cantilever-enhanced detection in the NIR11. With hollow-core optical fiber (HCF) gas cells of moderate optical path lengths (i.e., meters), PTS has demonstrated ppb-level NEC in the NIR12,13. Operating in the NIR would allow the use of mature fiber-optic components, which have the advantage of compactness, remote interrogation, and multi-point detection14,15,16,17.
Here, we report mode-phase-difference (MPD) PTS for high performance gas detection. The MPD between two transverse modes of a gas-filled HCF is measured interferometrically and taken as the photothermal (PT) signal. The MPD is sensitive to gas absorption in the hollow-core but insensitive against external environmental perturbations, enabling remarkably high signal-to-noise ratio (SNR). With meters-long HCF gas cells operating in the NIR telecom band, we demonstrate all-fiber optical acetylene detection with ppt level NEC and an unprecedented dynamic range of seven orders of magnitude.
Results
Theory
MPD-PTS uses a dual-mode HCF as the gas cell. Figure 1a shows a single-ring (SR) anti-resonant (AR) HCF used in this research. It has a broad transmission band from below 850 to beyond 1700 nm and supports a fundamental LP01-like mode and a second-order LP11-like mode, as shown, respectively, in Fig. 1b, c.
The basics of MPD-PTS may be explained intuitively by using the illustrations in Fig. 2. A modulated pump laser beam propagating in a dual-mode HCF is absorbed by trace molecules in the hollow-core, which heats up the gas and perturbs the RI distribution. The heating profile follows the intensity distribution of the pump, which depends on the fractional power of the pump in the two modes. The pump intensity would have an approximate Gaussian distribution for the LP01 mode and a two-lode distribution for the LP11 mode. Ideally, all the pump power should be launched into the LP01 mode (Supplementary Note 2) and the heating profile in the cross-section would be approximately Gaussian and invariant along the HCF. Figure 2a shows a general scenario where both LP01 and LP11 modes are present in the HCF. The pump intensity, the temperature (and hence the RI) distribution in the hollow-core varies periodically along the HCF owing to coherent mixing of the two modes and the spatial period equals to the modal beat length lb (Supplementary Note 2). Figure 2b, c show, respectively, the variations of pump intensity and temperature along the HCF. If a probe laser beam is now propagating simultaneously in the HCF, it will be modulated by the RI change. The probe also has LP01 and LP11 modes, and the PT phase modulations for the two probe modes are different, depending on the overlap integral of the mode fields with the RI perturbation, as shown in Fig. 2c. According to the Beer–Lambert law and under the assumption of weak absorption, i.e., \(\alpha ( {\lambda _{{\mathrm{pump}}}})CL\; < <\; 1\), the differential PT phase modulation is proportional to gas concentration C and given by
where Δϕi (i = 01 or 11) represents phase modulation for LP01 or LP11 mode of the probe, α(λpump) the absorption coefficient for a relative concentration of 100%, λpump the wavelength of the pump, L the length of the sensing HCF, Ppump the average pump power over L, η the fractional pump power in the LP01 mode and f the pump modulation frequency. The differential phase modulation coefficient k* is a function of η and f. The detailed formulations of differential modulation and calculation of k* are presented in Supplementary Note 2. For η = 90% and f < 10 kHz, k* is determined to be 1.7 (±0.2) × 10−7 rad cm ppm−1 m−1 mW−1 for balance gas of argon at 1.5 bar. It is ~ 20% of the phase modulation coefficient of the fundamental mode for the SR-AR-HCF, which means that the MPD is still reasonably sensitive to gas absorption in the hollow-core. This is because that the fields of the probe LP01 and LP11 modes overlap differently with the non-uniform heating profile and the phase sensitivities of the two modes to gas absorption are quite different. The variation in the MPD can be conveniently detected with a proper in-line dual-mode fiber interferometer, as shown in Fig. 2d.
However, sensitivity of the MPD to external perturbation can be much smaller than that of the fundamental mode phase. Owing to the very small diameter of the HCF, the environmental (e.g., temperature and pressure) changes would result in uniform RI change across the hollow-core region and hence affect the phases of the LP01 and LP11 modes similarly, meaning that the MPD is insensitive to environmental perturbation. The sensitivities of the LP01 mode phase and the MPD to external perturbation (X) may be expressed as18,19
where \({\mathrm{\Delta }}n = n_{01} - n_{11}\) with n01 and n11 representing, respectively, the effective RI of probe LP01 and LP11 modes. The sensitivity ratio \(\zeta = \Delta \phi _{01}/\delta \phi\) is on the order of \(n_{01}/{\mathrm{\Delta }}n\), indicating that the MPD is much less sensitive to external perturbation than the fundamental mode phase. Numerical simulation with COMSOL Multiphysics (Supplementary Note 3) shows that the values of ζ are, respectively, ~280 for temperature and ~170 for pressure for the SR-AR-HCF at 1550 nm. To summarize, the MPD is reasonably sensitive to gas absorption in the hollow-core but very insensitive to external disturbances, which enhances the SNR significantly and makes extremely sensitive gas sensors possible.
Experimental setup
Two gas cells are made using the SR-AR-HCF shown in Fig. 1a. One uses a 4.67-m-long HCF with both ends mechanically spliced to single mode fiber (SMF) pigtails. A small air-gap is kept at the spliced joints for gas filling into the HCF. The other uses a 0.74-m-long HCF with both ends mechanically spliced to SMF pigtails, and 34 lateral microchannels are fabricated along the HCF for fast gas filling. Lateral offsets are introduced at the splicing joints to excite and to collect light from the LP01 and LP11 modes to form an in-fiber dual-mode interferometer to probe the modulation in the MPD. The structures, fabrication processes, and properties of the HCF gas cells can be found in Methods.
The experimental system is shown in Fig. 3. The pump source is a 1.53-μm distributed feedback laser and its wavelength is modulated sinusoidally at ~9.5 kHz via the internal signal generator of lock-in amplifier (LIA) and at the same time scanned slowly across the P(13) absorption line of acetylene at ~ 1532.830 nm. The amplitude of wavelength modulation is set to be ∼2.2 times the linewidth of the P(13) line to maximize the second harmonic (2 f) signal20. The probe source is an external-cavity diode laser and its wavelength is tuned to a quadrature point of the dual-mode interference fringe at ~1550 nm. The pump and probe beams are launched into the HCF from the opposite directions via two wavelength-division multiplexers (WDM1 and WDM2), and the probe beam coming out from WDM2 is detected by a photodetector (PD). The 2 f component of the MPD modulation is demodulated by the LIA and transferred to a computer via a data acquisition (DAQ) card for further processing.
Lower detection limit and long-term stability
Experiments were first conducted with the 4.67-m-long HCF gas cell filled with the calibrated gas sample of 1 ppm acetylene in argon. The gas was pressurized into the hollow-core with ~1.95 bar pressure at the input joint, whereas the output joint is open to atmosphere. The details of the experiments can be found in Supplementary Note 5. Figure 4a shows the 2 f lock-in output (the PT signal or 2 f signal) when the pump is tuned across the P(13) line of acetylene for different pump power levels delivered into the HCF. The 2 f lock-in output when the pump is tuned away from the absorption line to ~1532.57 nm is also recorded and regarded as background noise. The PT signal increases linearly with pump power, whereas the standard deviation (s.d.) of the noise remains almost unchanged, as shown in Fig. 4b. At the pump power of ~108 mW, the average SNR over 10 measurements is ~14809 for a lock-in time constant of 1 s with 18 dB/Oct slope, giving a NEC of ~68 ppt for SNR of unity (i.e., 1σ).
Allan–Werle deviation analysis21,22 is conducted with the noise data over a period of 2 hours, and the results are shown in Fig. 4c. The 1σ NEC obtained from the Allan–Werle plot is ~43 ppt and ∼15 ppt, respectively, for 10 and 100 s averaging time. The system is capable of averaging over a much longer time of up to a few thousand seconds owing to superb noise cancellation ability of the MPD detection technique, which would enable sub-ppt level acetylene detection. This result is more than two orders of magnitude better than the previous optical fiber gas sensors12.
The system stability, which is important for practical applications, is tested by repeatedly scanning the pump wavelength across the P(13) line and the results for 1 ppm C2H2 with a pump power of 108 mW are shown in Fig. 5. The p–p value of DC compensated 2 f signal varies ~0.8% over a period of 3-hours (see Supplementary Note 5).
Dynamic range and response time
With the same experimental setup but the 0.74-m-long HCF gas cell, the dynamic range of the sensing system is tested by filling different concentrations of acetylene into the hollow-core. Figure 6a shows the 2 f signal for 4, 10, 20 ppm acetylene in nitrogen, whereas Fig. 6b shows the p–p value of the 2 f signal as functions of acetylene concentration from 4 ppm to 4% acetylene in nitrogen at atmospheric pressure. An approximately linear relationship is obtained for acetylene concentration up to ~1% and significant non-linearity starts to appear beyond this value. The 1σ NEC is evaluated by filling the calibrated 1 ppm acetylene in argon at the pressure of ~1.5 bar in to the HCF and found to be 2 ppb for 1 s lock-in time constant. Alan–Werle deviation analysis (see Supplementary Note 5) obtains NEC of ~500 ppt for an averaging time of ~100 s, giving a dynamic range of ~2 × 107, nearly two orders of magnitude larger than the state-of-the-art gas detection systems12.
The response time is tested by filling HCF gas cell, via the microchannels, with nitrogen, 1000 ppm acetylene in nitrogen and then nitrogen. For this experiment, the pump laser wavelength is fixed to the center instead of scanning across the P(13) line. The 2 f signal from the LIA is shown in Fig. 7. The response time t90, which is defined as the time taken to reach 90% of the applied concentration23, is ~44 s, showing that the introduction of microchannels breaks limitation of slow response in HCF gas sensors24.
Discussion
The PT signal (i.e., 2 f signal) increases linearly with pump power, whereas the noise level stays almost unchanged. This leaves rooms for further improving the detection sensitivity by simply increasing the pump power level. The overlap of mode fields with fiber material (silica) in the SR-AR-HCF is extremely small, meaning that the HCF would have a high damage threshold. In fact, transmission of laser beams with average power as high as 70 W has been demonstrated with a SR-AR-HCF25. If we could deliver ~1 W pump power into the SR-AR-HCF by, for example, using a large-mode-area fiber without an air-gap at the fiber joint, it would be possible to achieve sub-ppt level acetylene detection. For 1 W pump power launched into SR-AR-HCF filled with 5% C2H2 at ambient conditions, the highest temperature rise is calculated to be ~14 °C, well below acetylene’s autoignition temperature of 305 °C according to DIN EN 14522-2005. The pump power density for 1 W in the hollow-core is ~105 W/cm2, far below the breakdown value ~1012 W/cm2 of air, which is similar to that of C2H226.
The long-term stability of the system benefits from two factors: (1) the SMFs are not parts of the interferometer for the MPD detection and hence disturbance on the SMFs would not affect the phase difference between the interfering modes, resulting in a better stability over the traditional two-beam interferometers such as Mach–Zehnder. (2) Comparing with the fundamental mode phase, the MPD is much insensitive to external disturbance. For the stability test results in Fig. 5, the probe wavelength is simply tuned to a quadrature point of the interference fringe. Stable operation over a much longer term may be achieved by locking the probe laser wavelength to a fringe quadrature via servo-control, and by using polarization maintaining SMFs for light transmission and a dual-mode sensing HCF capable of maintaining polarization as well as lobe-orientation27.
The dynamic range of the system is limited by the non-linear transfer function of the interferometer as well as the non-linear dependence of the phase modulation beyond the weak absorption approximation12. With 110 mW pump power and the modulation frequency of 9.5 kHz, the differential phase modulation for the 0.74-m-long SR-AR-HCF filled with 1% acetylene in nitrogen is calculated to be ~0.14 rad, well within the 1% linear dynamic range of the interferometer transfer function (that is, \(\delta \varphi \le 0.1\pi\)). Nevertheless, according to the Beer–Lambert law, the pump power is absorbed significantly and reduced to ~50% at the output of the HCF gas cell, suggesting weak absorption assumption becomes invalid, which bears the primary responsibility for the observed non-linear behavior (See Supplementary Note 6).
The response time of the present system is not actually limited by the HCF gas cell, but by the time taken to fill the gas chamber. We placed the 0.74-m-long HCF gas cell inside a tubular chamber with the diameter of 25 mm and length of 1 m, and time is needed to fill gas into the chamber. With multiple microchannels fabricated along the HCF, the response time limited by gas diffusion into the hollow-core can be as short as 3 s24.
The performances of some state-of-the-art spectroscopic laser gas sensors are summarized in Table 1. For ease of comparison, we list the noise-equivalent absorption (NEA), which is independent gas types and absorption strength. The values of NEA can be converted from NEC through equation \({\mathrm{NEA}} = {\mathrm{NEC}} \cdot \alpha ( {\lambda _{pump}})\). The NEA of the MPD-PTS is more than two orders of magnitude better than the state-of-the-art PTS system with a bandgap HCF12 and comparable to some of the most sensitive spectroscopic techniques. However, the MPD-PTS system is much simpler and in an all-fiber format. The dynamic range is orders of magnitude larger than any of the previous techniques. The MPD-PTS system can operate under the condition of ambient pressure and temperature.
In summary, we developed a MPD PT spectroscopic technique for ultrasensitive gas detection over extremely large dynamic range. The detection of MPD minimizes the effect of environment perturbation while maintaining a sufficiently large PT modulation signal, enabling orders of magnitude of enhancement of SNR. With a 4.67-m-long HCF, we achieved detection limit down to low ppt level and 0.8% instability over a period of 3 hours. With a 0.74-m-long HCF, we demonstrated dynamic range of over 7 orders of magnitude and response time of ~ 44 s. The broadband transmission of the SR-AR-HCFs28,29,30 will allow pumping at different wavelengths from visible to MIR, which, in combination with an ultra-low-noise fiber-optic probe interferometer in the NIR, would enable cost-effective, precision all-fiber sensors with multi-gas detection capability. The technique could also be applied to liquid-core fiber sensors and other dual-mode waveguide systems for the study of light-matter interaction.
Methods
Fabrication and characterization of HCF gas cells
The 4.67-m-long HCF was made by mechanically splicing the SR-AR-HCF to an input SMF pigtail and an output SMF pigtail. For gas filling purpose, a gap of 1–2 µm was kept between the input SMF and the HCF, as well as between the HCF and the output SMF. A lateral offset of a few μm is introduced between the HCF and the probe input SMF, to excite the two probe modes (LP01 and LP11) simultaneously. The input mechanical joint is sealed within a T-shaped glass tube with a gas inlet and outlet. The HCF is also offset aligned to the output SMF and the relative lateral positions are adjusted by use of translation stages to achieve a reasonable fringe contrast for the interference between the two probe modes. For the 0.74-m-long HCF gas cell, the SR-AR-HCF is mechanically spliced to SMFs in the input and the output ends with lateral offsets to form an all-fiber modal interferometer. Microchannels are drilled, by use of a femtosecond infrared laser, from the side of the HCF, and gas filling into the hollow-core is achieved via these microchannels. For more details, please refer to Supplementary Note 4.
Preparation of gas samples
Gas samples with different concentrations are prepared by mixing calibrated gas samples at atmospheric pressure with two mass flow controllers. For gas concentration below 1000 ppm, the gas samples are prepared by mixing 1000 ppm acetylene in nitrogen with high-purity nitrogen with different ratios. For gas concentration above 1000 ppm, the samples are prepared by mixing high-purity nitrogen with 99.99% acetylene.
Data availability
The data that support the findings of this study are available from the authors upon reasonable request.
References
Laj, P. et al. Measuring atmospheric composition change. Atmos. Environ. 43, 5351–5414 (2009).
Mahaffy, P. R. et al. The sample analysis at Mars investigation and instrument suite. Space Sci. Rev. 170, 401–478 (2012).
Smith, D. & Španěl, P. The challenge of breath analysis for clinical diagnosis and therapeutic monitoring. Analyst 132, 390–396 (2007).
Allen, M. G. Diode laser absorption sensors for gas-dynamic and combustion flows. Meas. Sci. Technol. 9, 545 (1998).
Liu, X. et al. A survey on gas sensing technology. Sensors 12, 9635–9665 (2012).
Claps, R. et al. Ammonia detection by use of near-infrared diode-laser-based overtone spectroscopy. Appl. Opt. 40, 4387–4394 (2001).
Weibring, P. et al. Ultra-high-precision mid-IR spectrometer II: system description and spectroscopic performance. Appl. Phys. B 85, 207–218 (2006).
Ye, J., Ma, L. S. & Hall, J. L. Ultrasensitive detections in atomic and molecular physics: demonstration in molecular overtone spectroscopy. JOSA B 15, 6–15 (1998).
Zhao, G. et al. Shot-noise-limited Doppler-broadened noise-immune cavity-enhanced optical heterodyne molecular spectrometry. Opt. Lett. 43, 715–718 (2018).
Foltynowicz, A. et al. Noise-immune cavity-enhanced optical heterodyne molecular spectroscopy: current status and future potential. Appl. Phys. B 92, 313 (2008).
Tomberg, T., Hieta, T., Vainio, M. & Halonen, L. Cavity-enhanced cantilever-enhanced photo-acoustic spectroscopy. Analyst 144, 2291–2296 (2019).
Jin, W., Cao, Y. C., Yang, F. & Ho, H. L. Ultra-sensitive all-fibre photothermal spectroscopy with large dynamic range. Nat. Commun. 6, 6767 (2015).
Lin, Y. C., Jin, W., Yang, F., Tan, Y. Z. & Ho, H. L. Performance optimization of hollow-core fiber photothermal gas sensors. Opt. Lett. 42, 4712–4715 (2017).
Shemshad, J., Aminossadati, S. M. & Kizil, M. S. A review of developments in near infrared methane detection based on tunable diode laser. Sens. Actuat. B Chem. 171, 77–92 (2012).
Hodgkinson, J. & Tatam, R. P. Optical gas sensing: a review. Meas. Sci. Technol. 24, 012004 (2012).
Culshaw, B., Stewart, G., Dong, F., Tandy, C. & Moodie, D. Fibre optic techniques for remote spectroscopic methane detection-from concept to system realisation. Sens. Actuat. B Chem. 51, 25–37 (1998).
Ho, H. L., Jin, W. & Demokan, M. S. Sensitive, multipoint gas detection using TDM and wavelength modulation spectroscopy. Electron. Lett. 36, 1191–1193 (2000).
Dangui, V., Kim, H. K., Digonnet, M. J. & Kino, G. S. Phase sensitivity to temperature of the fundamental mode in air-guiding photonic-bandgap fibers. Opt. Express 13, 6669–6684 (2005).
Pang, M., Xuan, H. F., Ju, J. & Jin, W. Influence of strain and pressure to the effective refractive index of the fundamental mode of hollow-core photonic bandgap fibers. Opt. Express 18, 14041–14055 (2010).
Reid, J. & Labrie, D. Second-harmonic detection with tunable diode lasers—comparison of experiment and theory. Appl. Phys. B 26, 203–210 (1981).
Allan, D. W. Statistics of atomic frequency standards. Proc. IEEE 54, 221–230 (1966).
Werle, P. O., Mücke, R. & Slemr, F. The limits of signal averaging in atmospheric trace-gas monitoring by tunable diode-laser absorption spectroscopy (TDLAS). Appl. Phys. B 57, 131–139 (1993).
Heilig, A. et al. Gas identification by modulating temperatures of SnO2-based thick film sensors. Sens. Actuat. B Chem. 43, 45–51 (1997).
Hoo, Y. L., Liu, S., Ho, H. L. & Jin, W. Fast response microstructured optical fiber methane sensor with multiple side-openings. IEEE Photonic Tech. L. 22, 296–298 (2010).
Michieletto, M. et al. Hollow-core fibers for high power pulse delivery. Opt. Express 24, 7103–7119 (2016).
Miziolek, A. W. & Sausa, R. C. Photochemical Ignition Studies. I. Laser Ignition of Flowing Premixed Gases. Technical Report BRL-TR-2644 (US Army Ballistic Research Laboratory, 1985).
Osório, J. H. et al. Tailoring modal properties of inhibited-coupling guiding fibers by cladding modification. Sci. Rep. 9, 1376 (2019).
Gao, S. F., Wang, Y. Y., Ding, W. & Wang, P. Hollow-core negative-curvature fiber for UV guidance. Opt. Lett. 43, 1347–1350 (2018).
Ding, W., Wang, Y. Y., Gao, S. F., Hong., Y. F. & Wang, P. Theoretical and experimental investigation of light guidance in hollow-core anti-resonant fiber. ACTA Phys. SIN-CH ED 67, 12 (2018).
Yu, F. & Knight, J. C. Spectral attenuation limits of silica hollow core negative curvature fiber. Opt. Express 21, 21466–21471 (2013).
Modugno, G., Corsi, C., Gabrysch, M. & Inguscio, M. Detection of H2S at the ppm level using a telecommunication diode laser. Opt. Commun. 145, 76–80 (1998).
Nelson, D. D. et al. Characterization of a near-room-temperature, continuous-wave quantum cascade laser for long-term, unattended monitoring of nitric oxide in the atmosphere. Opt. Lett. 31, 2012–2014 (2006).
Kassi, S. et al. Looking into the volcano with a Mid-IR DFB diode laser and cavity enhanced absorption spectroscopy. Opt. Express 14, 11442–11452 (2006).
Crosson, E. A cavity ring-down analyzer for measuring atmospheric levels of methane, carbon dioxide, and water vapor. Appl. Phys. B 92, 403–408 (2008).
Engel, G. S., Drisdell, W. S., Keutsch, F. N., Moyer, E. J. & Anderson, J. G. Ultrasensitive near-infrared integrated cavity output spectroscopy technique for detection of CO at 1.57 μm: new sensitivity limits for absorption measurements in passive optical cavities. Appl. Opt. 45, 9221–9229 (2006).
Rothman, L. S. et al. The HITRAN 2004 molecular spectroscopic database. J. Quant. Spectrosc. Radiat. Transfer 96, 139–204 (2005).
Acknowledgements
This work was supported by Hong Kong SAR government GRF grant (15220617), National Natural Science Foundation of China through NSFC grant (61827820, 61535004), The Hong Kong Polytechnic University (SB95, 1-ZVG4 and 4-BCD1) and the Program for Changjiang Scholars and Innovative Research Team in University (under grant-IRT_16R02). We thank professor Y.P. Wang and his team for allowing us to use the femtosecond laser micro-machining facility at Shenzhen University to fabricate microchannels on HCFs.
Author information
Authors and Affiliations
Contributions
The work was conducted at HK PolyU. P.Z. built the systems and conducted the experiments. H.B. and H.L.H. assisted in building the systems and conducting the experiments. P.Z. and Y.Z. performed numerical modeling. P.Z., Y.Z., W.J. and S.F. analyzed the results and prepared the manuscript. S.G., Y.W. and W.P. designed and fabricated the SR-AR-HCF. W.J. conceived the idea of SNR enhancement and developed the mathematical formulations. W.J. and S.F. coordinated the project.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Peer review information Nature Communications thanks Karol Krzempek and the other, anonymous, reviewer(s) for their contribution to the peer review of this work. Peer reviewer reports are available.
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Source Data
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zhao, P., Zhao, Y., Bao, H. et al. Mode-phase-difference photothermal spectroscopy for gas detection with an anti-resonant hollow-core optical fiber. Nat Commun 11, 847 (2020). https://doi.org/10.1038/s41467-020-14707-0
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41467-020-14707-0
This article is cited by
-
Ultrafast near-infrared pyroelectric detector based on inhomogeneous plasmonic metasurface
Light: Science & Applications (2024)
-
Dual-comb optomechanical spectroscopy
Nature Communications (2023)
-
Dual-band light-induced thermoelastic spectroscopy utilizing an antiresonant hollow-core fiber-based gas absorption cell
Applied Physics B (2023)
-
Dual-comb photothermal spectroscopy
Nature Communications (2022)
-
Recent Advances in Spectroscopic Gas Sensing With Micro/Nano-Structured Optical Fibers
Photonic Sensors (2021)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.