Abstract
The effect of arterial stiffening on elevated pulsatile left ventricular afterload patients with aortic stenosis (AS) is pronounced beyond systemic hypertension. Circulatory afterload pulsatile efficiency (CAPE) is a marker of vascular function, defined as the ratio of steady state energy consumption (SEC) to maintain systemic circulation and pulsatile energy consumption (PEC). Twenty patients aged 80 ± 7 years were assessed at baseline and a median of 60 days post transcatheter aortic valve replacement (TAVR), with pulsatile vascular load calculated using simultaneous radial applanation tonometry derived aortic pressure and cardiac magnetic resonance phase-contrast imaging derived ascending aortic flow. Eight out of 20 patients had a reduction in PEC post TAVR, and the reduction of PEC correlated strongly with the number of days post TAVR (R = 0.62, P < 0.01). Patients assessed within the 100 days of TAVR had a rise in their PEC when compared to baseline (0.19 ± 0.09 vs 0.14 ± 0.08 W, P = 0.04). Baseline PEC correlated moderately with baseline SEC (R = 0.49, P = 0.03), and a high baseline PEC was predictive of post TAVR PEC reduction (R = 0.54, P =0.01). Overall, no significant differences were found between baseline and post TAVR for systolic aortic pressure (131 ± 20 vs 131 ± 20 mmHg), systemic vascular resistance (1894 ± 493 vs 2015 ± 519 dynes.s/cm5), aortic valve ejection time (337 ± 22 vs 324 ± 34 ms) or aortic characteristic impedance (120 ± 48 vs 107 ± 41 dynes.s/cm5). Improved flow profiles after TAVR likely unmask the true vascular properties by altering ventriculo-valvulo-arterial coupling, leading to downstream vascular remodelling secondary to flow conditioning, and results in eventual improvement of pulsatile afterload as reflected by our proposed index of CAPE.
Similar content being viewed by others
Introduction
Aortic stenosis (AS) is a degenerative process with progressive calcification of the valve causing left ventricular (LV) outflow obstruction, from which increased LV afterload and subsequent LV hypertrophy, fibrosis and failure ensue if left untreated [1, 2]. As a disease affecting predominantly the elderly with often co-existing hypertension, the histopathological process of AS shares some similarity to that of atherosclerosis [3], with the resulting increased arterial stiffness serving as an independent predictor of mortality [4]. Treatment of AS with transcatheter aortic valve replacement (TAVR) provides an opportunity to study the effect of valvular obstruction relief on vascular loading, without the complicating vascular disruption arising from conventional surgical aortic valve replacement (SAVR).
The traditional view that AS is a fixed, valvular LV outflow obstruction has recently been challenged by the concept that LV afterload is a series circuit involving both the valve and the vasculature [5], with an emphasis on ventriculo-valvulo-arterial coupling. Quantification of the vascular load in AS is nebulous, as current methods of afterload assessment typically focus on the LV or the valve itself in isolation, and often underestimate the effect of the vascular load [6]. Separation of pulsatile and steady state arterial load is rarely performed in the assessment of vascular load, especially in the context of AS.
To date, LV afterload indices in AS that incorporate vascular load components such as the valvulo-arterial impedance (Zva) derived using transthoracic echocardiography (TTE) [7], instantaneous impedance from cardiac magnetic resonance (CMR) using time domain analysis (Zva-ins) [8], or energy loss index corrected for ascending aortic cross-sectional area [9], all have varying degrees of valvular contribution. A prior study also demonstrated that an acute transfer of load may occur immediately after TAVR, indicating that vascular parameters may have been modified by upstream valvular obstruction from AS [10], presumably from a reduction in transvalvular flow.
A seemingly forgotten method to assess the vascular component of the afterload is the pulsatile energy loss within the circulation first proposed by O’Rourke et al [11]. Pulsatile load arises from the concept that a proportion of the external work imparted by the LV to systemic circulation is lost to arterial stiffness due to the requirement to maintain the pulsatile nature of circulation, and this energy loss per unit time can be expressed as pulsatile power [11], a hemodynamic parameter that we hypothesise to be theoretically independent of upstream obstruction. The ratio of the pulsatile energy consumption (PEC) to the steady state energy consumption (SEC) by the circulatory system can be represented by the novel index of circulatory afterload pulsatile efficiency (CAPE), which is a surrogate for comparing pulsatile afterload as a measurement of total afterload.
As the valvular obstruction is relieved by TAVR, reverse vascular remodelling may take place independently to myocardial remodelling of the LV, which is likely a time-dependent process. We hypothesise a gradual flow mediated vascular remodelling as the vasculature re-adapts to improved flow profile post TAVR, with ongoing reduction in PEC and therefore improvement in CAPE when compared to baseline. In this study, we assess the response of pulsatile circulatory efficiency following TAVR and its association with time using non-invasive methods.
Methods
Study population
Twenty patients with moderate to severe symptomatic AS planned for TAVR were studied. Exclusion criteria included the presence of more than moderate concomitant aortic regurgitation, mitral, tricuspid, or pulmonary disease, the presence of un-revascularized severe coronary disease, previous surgery involving the aortic valve or proximal aorta and general contraindications to CMR. All TAVR procedures were discussed at a local hospital multidisciplinary heart team meeting prior to proceeding.
All patients underwent initial demographic and brachial blood pressure measurements, TTE, and simultaneous radial arterial applanation tonometry (AT) and CMR at baseline and post TAVR (median of 60 days, range 28–812 days). All patients provided informed written consent and the study was approved by the local hospital research ethics committee (HREC 2019/ETH03530).
Study protocol and follow-up
Simultaneous AT and CMR data acquisition method was adapted from our pilot study previously [12]. In brief, a modified high-fidelity arterial tonometer (Millar Instruments, Houston, TX, USA) was applied to the study patients’ radial pulse during CMR [13], and real-time aortic pressure waveform was derived from radial AT waveform using a validated transfer function after calibration to brachial blood pressure cuff measurement using the SphygmoCor 8.1 system (AtCor Medical, Sydney, Australia) [14, 15].
CMR studies were performed in a 3T magnet (Siemens Magnetom, Erlangen, Germany) using a previously described protocol [12]. Ascending aortic cross-sectional area, volume and velocity flow profiles were obtained at the level of the main pulmonary artery bifurcation using 2D through plane phase-contrast imaging with velocity encoding (Venc) of 3m/s. Images were acquired at a voxel size of 1.8×.1.8 mm at a slice thickness of 6mm. Repetition time was 37.12 ms and echo time was 2.47 ms at a flip angle of 20°. Post processing analyses were performed using CVI 42 (Circle Cardiovascular Imaging, Calgary, Alberta, Canada). Patients were asked to perform breath hold to minimize the effect of changing intrathoracic pressure to flow profile. Arrhythmia causing high beat-to-beat variability that resulted in gating errors and produced non-physiological flow profiles were rejected and re-acquired after visual assessment.
Transthoracic echocardiographic images were obtained using Affinity ultrasound machine (Phillips Healthcare, Eindhoven, the Netherlands). Quantification of AS was determined using Doppler echocardiography (apical 5-chamber-view) in accordance with American Society of Echocardiography guidelines [16].
Hemodynamic data analysis
Volume flow impedance parameters and vascular load indices estimation were calculated by frequency domain analysis using simultaneously obtained aortic pressure and flow profile as outlined in a previously published method [12, 14, 15]. In brief, for impedance spectrum estimation, flow velocity and pressure waves were decomposed into their component harmonics using a fast Fourier transform (FFT), and FFT pressure amplitudes were divided by FFT flow velocity amplitudes for frequencies up to 10 Hz. Any harmonics were excluded if the modulus was less than 0.6mmHg for pressure, or less than 1 cm/s for flow velocity when adjusted for cross-sectional area of the ascending aorta, as in Nichols et al. [17]. To account for the difference in heart rate between flow and pressure measurements, pressure and flow harmonics were normalized and interpolated linearly into the nearest integer frequencies, that is 1, 2, 3,..., 10 Hz. The characteristic impedance of the ascending aorta (Zc) was calculated as the average of impedance moduli between 2–10 Hz. Systemic vascular resistance (SVR) is calculated as the ratio between mean pressure and mean flow at steady state (zero Hz frequency).
Calculation of steady state and pulsatile power
Steady state energy consumption (SEC) was defined as the product of mean pressure and mean flow over the course of the cardiac cycle per time unit. Pulsatile energy consumption (PEC) was defined as the sum of the product of the pressure and flow moduli adjusted by the cosine of the phase angle for first 7 harmonics on frequency domain analysis:
Here, \(P_n\) and \(Q_n\) are the pressure and flow at the nth harmonic respectively, with \(\cos \theta _n\) being the cosine of the phase angle at the nth harmonic. The absolute value is taken in the conversion of a vector to a directionless magnitude for the calculation of power as energy expenditure.
Circulatory afterload pulsatile efficiency (CAPE) was defined by the following equation and expressed as a percentage:
A higher CAPE denotes a more efficient pulsatile circulation as less of the total circulatory energy is being dissipated as pulsatile energy loss.
Quality of life measurements
Seventeen of 20 study patients underwent a 6-minute walk test and answered a 12-Item Kansas City Cardiomyopathy Questionnaire (KCCQ-12) for evaluation of functional capacity and quality of life respectively, on the study day at baseline or post TAVR.
Statistical analysis
All data are presented as mean ± SD unless otherwise specified. To test for normality, the Shapiro–Wilk test was used, and a paired two-tailed Student’s t-test or Wilcoxon Signed-Rank test was performed depending on the normality of the data to assess for differences between baseline and post TAVR measurements. Pearson or Spearman’s Rho test were performed to assess for correlation between ordinal variables where appropriate, accounting for the presence of potential outliers. Significance level was set at P < 0.05. Data were analyzed using SPSS (version 27, IBM Inc., Armonk, NY, USA).
Results
Baseline demographic and clinical characteristics of the 20 study patients are summarized in Table 1. Study patients were relatively elderly with a mean age of 80 ± 7 years, mildly overweight (body mass index 28.0 ± 5.2 kg/m2), with 13 out of 20 having pre-existing hypertension requiring pharmacological treatment. Four patients were demonstrated to have evidence of myocardial scarring on late gadolinium enhancement on CMR. There was a significant improvement in KCCQ-12 summary score (and 6-minute walk distance) following TAVR, but the clinical improvement was not predicted by any baseline hemodynamic parameters, including PEC and CAPE (all P > 0.05).
Vascular afterload in AS
The study cohort had symptomatic moderate to severe AS with mean aortic valve gradient of 37 ± 12 mmHg, with a calculated aortic valve area by the continuity equation on TTE of 0.92 ± 0.27 cm2, which correlated well with the planimetry aortic valve area on CMR at 0.97 ± 0.21cm2 (R = 0.73, P < 0.01). All patients had procedural success with TAVR and a significant reduction in their aortic valve pressure gradients (Table 2). At baseline, 1 patient had a reduced left ventricular ejection fraction of <55%, and 2 patients had a stroke volume index of <35 ml/m2. Three patients had adjustments to their antihypertensive regimen (1 patient required up-titration of antihypertensives, 2 patients reduced their antihypertensives) prior to undergoing repeat study, and 2 patients required pacemaker implantation but were not pacing dependent at the time of the repeat study visit.
The cohort demonstrated a higher-than-expected Zc (120 ± 48 dynes.s/cm5) of the proximal aorta when compared to reported values from similar prior studies [14], with a mildly elevated SVR (1894 ± 493 dynes.s/cm5). SEC and PEC were calculated to be 0.99 ± 0.24 W and 0.15 ± 0.09 W, respectively, with CAPE calculated to be 85 ± 8%.
Responses following TAVR
Repeat hemodynamic measurements were performed at 149 ± 206 days post TAVR. As expected, there were significant improvements in aortic valve-related TTE parameters as summarised in Table 2. On volumetric assessment via CMR, there was significant improvement in peak flow rate, but no differences were observed in stroke volume, cardiac output, or myocardial mass.
With respect to vascular load parameters, there were modest reduction of SVR from 1894 ± 493 dynes.s/cm5 to 2015 ± 519 dynes.s/cm5, and Zc from 120 ± 48 dynes.s/cm5 to 108 ± 41 dynes.s/cm5, but neither were statistically significant at P = 0.22 and P = 0.23 respectively.
When evaluated as a cohort, there were no significant difference between baseline SEC (0.97 ± 0.24 W baseline, 0.95 ± 0.36 W post TAVR) or PEC (0.15 ± 0.09W baseline, 0.17 ± 0.09 W post TAVR).
Predictors for improvement in pulsatile energy loss
The change in PEC and CAPE after TAVR for each patient is demonstrated in Fig. 1. The patients were divided into two sub-groups according to changes in PEC post TAVR for further analysis. Eight out of 20 patients demonstrated a numerical reduction in their PEC following TAVR, with associated improvement in CAPE. The 8 patients who had a reduction in PEC underwent repeat assessment post TAVR after a significantly longer duration than the remaining 12 patients, with a mean duration post TAVR of 284 ± 280 days compared to 59 ± 28 days (Table 3). As a group, there was no overall change in PEC upon comparing baseline with post TAVR. However, when divided into a short-term group (<100 days post TAVR, N = 13) and a long-term group (>100 days post TAVR, N = 7), there was a significant rise in post TAVR PEC in the short-term group when compared to baseline (0.19 ± 0.09 vs 0.14 ± 0.08 W, P = 0.04), which is not observed in the long-term group (0.13 ± 0.10 vs 0.17 ± 0.12 W, P = 0.33).
In the cohort of 8 patients, there were no significant baseline predictors for reduction of PEC. In the post TAVR studies, a reduction of PEC was associated with a significant shorter aortic valve ejection time (304 ± 30 ms vs 338 ± 31ms, P = 0.03). The PEC reduction cohort also had a higher CAPE when compared to those whose PEC did not reduce (79 ± 5% vs 88 ± 9%, P = 0.02).
Correlation between PEC and other variables and its relationship post TAVR
Spearman rho correlations was performed for PEC with other hemodynamic parameters at baseline, as well as the change in PEC post TAVR from baseline with other parameters at baseline to look for predictors (Figs. 2 and 3). At baseline, PEC was positively correlated to SEC (R = 0.49, P = 0.03) and stroke volume (R = 0.48, P = 0.04), but not with other hemodynamic parameters including Zc, SVR, aortic meant gradient, aortic valve ejection time, central aortic pressure (all P > 0.05). Post TAVR, the change in PEC (defined as Post TAVR PEC minus baseline PEC) was strongly correlated with the number of days post TAVR (R = 0.61, P < 0.01), and moderately correlated with baseline PEC (R = 0.54, P = 0.01). No other baseline parameters predicted the reduction in PEC. The number of days post TAVR also correlated with an improvement in CAPE post TAVR (R = 0.46, P = 0.04).
Discussion
With a specific focus on the pulsatile elements to the afterload, we evaluated pulsatile energy consumption (PEC) and its relationship with steady-state energy consumption (SEC), and proposed CAPE as a novel surrogate index for pulsatile vascular function. The main finding of our study is that time post TAVR is the most powerful predictor of improvement in circulatory afterload pulsatile efficiency (CAPE). There was a rise in PEC in the short-term (<100 days) post TAVR, but this appears to be a transient process as the PEC and CAPE improves with time, resulting in reduced pulsatile afterload.
We found a significant correlation between baseline PEC and SEC, but PEC was not directly correlated with other traditional afterload parameters including aortic systolic pressure, SVR and Zc of the aorta as we might have expected. Whilst the clinical benefit of TAVR in the form of symptomatic relief was readily observed in the short term, measurable improvements in hemodynamic indices requires gradual vascular remodelling with time.
Previous studies have attempted to find a single parameter that encompasses the global left ventricular afterload in AS by incorporating the valvular and vascular components together. The index most widely used is the valvulo-arterial impedance (Zva), termed by Hachicha et al, which had since been demonstrated to be a strong predictor of adverse outcomes in patients with AS [7, 18]. However, Zva had been shown to predominantly focus on the stenotic valve and underestimate the relative contributory values SVR and compliance to overall LV afterload [19], and its derivation process does not adequately consider the pulsatile nature of the circulation [8, 12],
On the other hand, the traditional “gold standard” arterial stiffness parameter of carotid-femoral pulse wave velocity (PWV) does not directly consider the effect of the aortic valve volume flow profiles, and sampling is performed is well downstream of the aortic valve and free of valvular contribution to its assessment, making it less relevant in the assessment of global LV afterload in AS. This is supported by the prior demonstration that arterial stiffness as measured by PWV is elevated in asymptomatic AS patients but does not correlate with symptoms or severity of AS, despite being an independent predictor of poorer outcomes [20]. Although an initial small scale study suggested improvement in aortic stiffness following SAVR [21], multiple subsequent larger studies have found that SAVR increased arterial stiffness in the longer term [22,23,24], possibly due to trauma to the vaso-vasorum during transection of the aorta and its associated aortic remodelling upon recovery [25]. The effect of TAVR on arterial stiffness seems less clear, with multiple studies using different assessment modalities demonstrating conflicting findings [10, 26,27,28,29].
Our study had found that 8 out of the 20 patients in our study demonstrated a reduction in PEC and an improvement in CAPE. Interestingly, the reduction of PEC is closely associated with the number of days post TAVR, with an initial elevation in PEC. This suggests that there may be an acute rise in pulsatile vascular load in the short-to-medium term, with attenuation over the medium-to-long term. This is consistent with prior studies on arterial stiffness either invasively with catheterisation in the short-term [10], or echocardiographic methods [28, 29]. Our finding that higher baseline PEC is predictive of the degree of PEC reduction post TAVR is consistent with prior findings that reduction in PWV following TAVR was largest in those who had the highest PWV at baseline [30].
In our study, there was a trend towards reduction of LV mass post TAVR (P = 0.06). Prior larger studies focusing on severe AS demonstrated that LV mass improved at 1 year, with concurrent reduction of extracellular volume [31]. We found that PEC reduction and improvement in CAPE seemed to occur independently to LV mass reduction, which suggests that the vascular remodelling process may be separate to myocardial remodelling and may occur even in patients with symptomatic moderate AS after TAVR treatment.
We hence postulate that true arterial stiffness and pulsatile energy loss will only become apparent after relief of the upstream mechanical obstruction at the level of the aortic valve, and TAVR as a less traumatic treatment modality minimizes the effect of vascular damage subjected by SAVR. This will in turn explain the rise in vascular loading parameters in the short-to-medium term, as the vasculature conditioned by chronic low flow state is now exposed to unobstructed flow and a much sharper rise to peak velocity with a normalized pulse profile. In support of the concept proposed by Plunde and Bäck [25], the apparent rise of arterial stiffness and associated pulsatile vascular load following TAVR is not reflective of a rapid change of the intrinsic property of the vessel wall, but rather the changes in ventriculo-valvulo-vascular coupling. Nevertheless, the improvement of PEC observed in the medium-to-long suggests that improved valvular flow dynamics likely eventually results in downstream vascular remodelling.
We found that there was a significant 10% reduction in aortic valve ejection time in those patients who had a reduction in PEC post TAVR when compared to their aortic valve ejection time at baseline, but they did not have higher baseline transvalvular gradients or smaller aortic valve area. Ejection time has been associated with the degree of AS [32, 33], and relief of AS mediated LV obstruction by TAVR can result in a reduction in ejection time in the immediate post-op period [34]. Furthermore, a prior study has demonstrated that a reduction in LV ejection time post SAVR is predictive of elevated arterial stiffness measured by cardio-ankle vascular index [35], suggesting that the effect of wave reflection from the peripheral arterial tree may become more pronounced after relief of valvular obstruction by TAVR. This is likely because ejection time accounts for the changes in pulse contour following TAVR and includes vascular function elements in the peripheral circulation impacted upon by wave reflection indices [25].
Consequently, our finding that PEC reduction is closely associated with aortic valve ejection time reduction also suggests that aortic valve ejection time is a complex marker that incorporates both the valvular properties as well as the vascular properties, and likely serves as a better simple predictor of downstream pulsatile vascular load and arterial stiffness than transvalvular gradients.
Another approach to evaluate global afterload in AS has been to explore the myocardial energy expenditure [36], which had been done using positron emission tomography to measure the myocardial oxygen consumption [37, 38]. More recently, a CMR-based method had also demonstrated the calculation of myocardial energy consumption in AS [39, 40]. However, both methods again fail to account for the pulsatile nature of the circulation, and each study found that TAVR failed to improve the myocardial efficiency of a proportion of patients in the medium term, which was defined as the ratio between steady-state circulatory energy consumption and the apparent myocardial energy consumption [38, 40]. We postulate that the sub-group of patients whose myocardial efficiency fail to improve is likely due to an elevated pulsatile vascular load that is not being accounted for, particularly if their baseline PEC is elevated, and their CAPE has not had adequate time to improve with adequate time since the TAVR.
Hence, the application of a hybrid method as suggested by our study accounting for pulsatile energy consumption with the novel index of CAPE will help to further elucidate the process myocardial and vascular energy efficiency and help to identify individuals who do not have optimal hemodynamic responses post TAVR. Furthermore, the unification of watt as the unit of energy consumption or power loss would help to compare previously separated concepts of myocardial and vascular function due to inconsistent units of measurement.
Study limitations
Our study has several limitations. We studied a group of patients with moderate to severe symptomatic AS with a relatively heterogenous transvalvular gradients and aortic valve area, in line with real-life indications for TAVR rather than a pure severe AS group. Our study patients had a variable follow-up period, which was partially due to institutional lockdown arising from the COVID-19 pandemic. Hemodynamic loading conditions can also be variable pending on the study patient’s fluid status on the day of the study.
Our CMR acquisition time was limited by the tolerance of our elderly study cohort. Sequences for planning and aortic flow lead to prolonged study durations, and patients found it difficult to tolerance prolonged scanning to allow the calculation of pulse wave velocity, and equipment limitation prevented us from performing simultaneous carotid and femoral tonometry for gold standard pulse wave velocity in the confined environment of the CMR with the interference of the magnetic field. Accurate correlation with CMR and/or tonometry-derived pulse wave velocity and other direct measurement of aortic stiffness would have added to the strength of our study. Our study also did not assess the regression of myocardial fibrosis or extracellular volume on CMR due to patient tolerance factors, which would have added to the strength of our analysis.
Frequency domain analysis is sensitive to noise and prone to the introduction of error [10, 41]. However, we have attempted to minimie noise by setting up a stand-alone, operator-independent acquisition system with high-fidelity pressure and flow data output. We have also chosen to sample the ascending aortic flow at the level of the pulmonary artery bifurcation, which allowed adequate distance from the flow artefact produced by the supravalvular metal struts of a self-expanding TAVR to dissipate.
Our study cohort was small and sub-group analysis needs to be interpreted with caution. Hard clinical endpoints such as mortality or morbidity were unable to be assessed given the follow-up period and small sample size, and we were unable to demonstrate any association between hemodynamic predictors and clinical outcomes on KCCQ-12 or 6-min walk test. Larger scale longitudinal studies with repeated measurements and correlation with clinical outcomes is required for further elucidation of CAPE post TAVR.
Conclusions
The transient rise in pulsatile vascular load post TAVR in the short-to-medium term appears to attenuate over time and result in medium-to-long-term improvement in CAPE. The improved flow profiles after correction of mechanical obstruction in AS with TAVR likely unmask the true vascular properties by altering the ventriculo-valvulo-arterial coupling and may result in downstream vascular remodelling secondary to improved flow conditioning.
References
Lindman BR, Dweck MR, Lancellotti P, Généreux P, Piérard LA, O’Gara PT, et al. Management of asymptomatic severe aortic stenosis: evolving concepts in timing of valve replacement. JACC Cardiovasc Imaging. 2020;13:481–93.
Otto CM, Nishimura RA, Bonow RO, Carabello BA, Erwin JP, Gentile F, et al. 2020 ACC/AHA guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2021;143:e72–227.
Otto CM, Kuusisto J, Reichenbach DD, Gown AM, O’Brien KD. Characterization of the early lesion of “degenerative” valvular aortic stenosis. Histological immunohistochem Stud Circ. 1994;90:844–53.
Vlachopoulos C, Aznaouridis K, Stefanadis C. Prediction of cardiovascular events and all-cause mortality with arterial stiffness: a systematic review and meta-analysis. J Am Coll Cardiol. 2010;55:1318–27.
Lindman BR, Otto CM, Douglas PS, Hahn RT, Elmariah S, Weissman NJ, et al. Blood pressure and arterial load after transcatheter aortic valve replacement for aortic stenosis. Circ Cardiovasc Imaging. 2017;10:1–12.
Hungerford SL, Adji AI, Hayward CS, Muller DWM. Ageing, hypertension and aortic valve stenosis: a conscious uncoupling. Heart Lung Circ. 2021;30:1627–36.
Hachicha Z, Dumesnil JG, Pibarot P. Usefulness of the valvuloarterial impedance to predict adverse outcome in asymptomatic aortic stenosis. J Am Coll Cardiol. 2009;54:1003–11.
Soulat G, Kachenoura N, Bollache E, Perdrix L, Diebold B, Zhygalina V, et al. New estimate of valvuloarterial impedance in aortic valve stenosis: a cardiac magnetic resonance study. J Magn Reson Imaging. 2017;45:795–803.
Briand M, Dumesnil JG, Kadem L, Tongue AG, Rieu R, Garcia D, et al. Reduced systemic arterial compliance impacts significantly on left ventricular afterload and function in aortic stenosis: Implications for diagnosis and treatment. J Am Coll Cardiol. 2005;46:291–8.
Yotti R, Bermejo J, Gutiérrez-Ibañes E, Pérez Del Villar C, Mombiela T, Elízaga J, et al. Systemic vascular load in calcific degenerative aortic valve stenosis: Insight from percutaneous valve replacement. J Am Coll Cardiol. 2015;65:423–33.
O’rourke MF. Steady and pulsatile energy losses in the systemic circulation under normal conditions and in simulated arterial disease. Cardiovasc Res. 1967;1:313–26.
Hungerford SL, Adji AI, Bart NK, Lin L, Namasivayam MJ, Schnegg B, et al. A novel method to assess valvulo-arterial load in patients with aortic valve stenosis. J Hypertens. 2021;39:437–46.
Beck DT, Martin JS, Nichols WW, Gurovich AN, Braith RW. Validity of a novel wristband tonometer for measuring central hemodynamics and augmentation index. Am J Hypertens. 2014;27:926–31.
Namasivayam M, Adji A, Lin L, Hayward CS, Feneley MP, O’Rourke MF, et al. Non-invasive quantification of ventricular contractility, arterial elastic function and ventriculo-arterial coupling from a single diagnostic encounter using simultaneous arterial tonometry and magnetic resonance imaging. Cardiovasc Eng Technol. 2020;11:283–94.
Adji A, Kachenoura N, Bollache E, Avolio AP, O’Rourke MF, Mousseaux E. Magnetic resonance and applanation tonometry for noninvasive determination of left ventricular load and ventricular vascular coupling in the time and frequency domain. J Hypertens. 2016;34:1099–108.
Bonow RO, Brown AS, Gillam LD, Kapadia SR, Kavinsky CJ, Lindman BR, et al. ACC/AATS/AHA/ASE/EACTS/HVS/SCA/SCAI/SCCT/SCMR/STS 2017 appropriate use criteria for the treatment of patients with severe aortic stenosis: a report of the American College of Cardiology Appropriate Use Criteria Task Force, American Association for Thoracic Surgery, American Heart Association, American Society of Echocardiography, European Association for Cardio-Thoracic Surgery, Heart Valve Society, Society of Cardiovascular Anesthesiologists, Society for Cardiovascular Angiography and Intervent. J Am Soc Echocardiogr. 2018;31:117–47.
Nichols WW, Conti CR, Walker WE, Milnor WR. Input impedance of the systemic circulation in man. Circ Res. 1977;40:451–8.
Hachicha Z, Dumesnil JG, Bogaty P, Pibarot P. Paradoxical low-flow, low-gradient severe aortic stenosis despite preserved ejection fraction is associated with higher afterload and reduced survival. Circulation. 2007;115:2856–64.
Katsanos S, Yiu KH, Clavel MA, Rodés-Cabau J, Leong D, Van Der Kley F, et al. Impact of valvuloarterial impedance on 2-year outcome of patients undergoing transcatheter aortic valve implantation. J Am Soc Echocardiogr. 2013;26:691–8.
Saeed S, Saeed N, Grigoryan K, Chowienczyk P, Chambers JB, Rajani R. Determinants and clinical significance of aortic stiffness in patients with moderate or severe aortic stenosis. Int J Cardiol. 2020;315:99–104.
Nemes A, Galema TW, Geleijnse ML, Soliman OII, Yap SC, Anwar AM, et al. Aortic valve replacement for aortic stenosis is associated with improved aortic distensibility at long-term follow-up. Am Heart J. 2007;153:147–51.
Musa TAL, Uddin A, Fairbairn TA, Dobson LE, Sourbron SP, Steadman CD, et al. Assessment of aortic stiffness by cardiovascular magnetic resonance following the treatment of severe aortic stenosis by TAVI and surgical AVR. J Cardiovasc Magn Reson. 2016;18:37.
Barbetseas J, Alexopoulos N, Brili S, Aggeli C, Marinakis N, Vlachopoulos C, et al. Changes in aortic root function after valve replacement in patients with aortic stenosis. Int J Cardiol. 2006;110:74–9.
Chirinos JA, Akers SR, Schelbert E, Snyder BS, Witschey WR, Jacob RM, et al. Arterial properties as determinants of left ventricular mass and fibrosis in severe aortic stenosis: findings from ACRIN PA 4008. J Am Heart Assoc. 2019;8:e03742.
Plunde O, Bäck M. Arterial stiffness in aortic stenosis and the impact of aortic valve replacement. Vasc Health Risk Manag. 2022;18:117–22.
Vavuranakis M, Vrachatis DA, Boudoulas H, Papaioannou TG, Moldovan C, Kariori MG, et al. Effect of transcatheter aortic valve implantation on the ascending aorta’s elasticity. Clin Res Cardiol. 2012;101:895–9. 10111. 2012
Terentes-Printzios D, Gardikioti V, Aznaouridis K, Latsios G, Drakopoulou M, Siasos G, et al. The impact of transcatheter aortic valve implantation on arterial stiffness and wave reflections. Int J Cardiol. 2021;323:213–9.
Müller C, Goliasch G, Schachinger S, Kastl S, Neunteufl T, Delle-Karth G, et al. Transcatheter aortic valve replacement (TAVR) leads to an increase in the subendocardial viability ratio assessed by pulse wave analysis. PLoS ONE. 2018;13:e0207537.
Vizzardi E, Sciatti E, Bonadei I, D’Aloia A, Gelsomino S, Lorusso R, et al. Effects of transcatheter aortic valve implantation on ascending aorta wall elastic properties: Tissue Doppler imaging and strain Doppler echocardiography study. Int J Cardiol Heart Vessel. 2014;4:198.
Goudzwaard JA, Disegna E, de Ronde-Tillmans MJAG, Lenzen MJ, de Jaegere PPT, Mattace-Raso FUS. Short-term changes of blood pressure and aortic stiffness in older patients after transcatheter aortic valve implantation. Clin Inter Aging. 2019;14:1379.
Treibel TA, Kozor R, Schofield R, Benedetti G, Fontana M, Bhuva AN, et al. Reverse myocardial remodeling following valve replacement in patients with aortic stenosis. J Am Coll Cardiol. 2018;71:860–71.
Gamaza-Chulián S, Díaz-Retamino E, Camacho-Freire S, Ruiz-Fernández D, Gutiérrez-Barrios A, Oneto-Otero J. Acceleration time and ratio of acceleration time to ejection time in aortic stenosis: new echocardiographic diagnostic parameters. J Am Soc Echocardiogr. 2017;30:947–55.
Bache RJ, Wang Y, Greenfield JC. Left ventricular ejection time in valvular aortic stenosis. Circulation. 1973;47:527–33.
Pagoulatou S, Stergiopulos N, Bikia V, Rovas G, Licker MJ, Müller H, et al. Acute effects of transcatheter aortic valve replacement on the ventricular-aortic interaction. Am J Physiol Hear Circ Physiol. 2020;319:1451–8.
Plunde O, Franco-Cereceda A, Bäck M. Cardiovascular risk factors and hemodynamic measures as determinants of increased arterial stiffness following surgical aortic valve replacement. Front Cardiovasc Med. 2021;8:754371.
Sörensen J, Harms HJ, Aalen JM, Baron T, Smiseth OA, Flachskampf FA. Myocardial efficiency: a fundamental physiological concept on the verge of clinical impact. JACC Cardiovasc Imaging. 2020;13:1564–76.
Hansson NHS, Sörensen J, Harms HJ, Kim WY, Nielsen R, Tolbod LP, et al. Myocardial oxygen consumption and efficiency in aortic valve stenosis patients with and without heart failure. J Am Heart Assoc. 2017;6:e004810.
Güçlü A, Knaapen P, Harms HJ, Vonk ABA, Stooker W, Groepenhoff H, et al. Myocardial efficiency is an important determinant of functional improvement after aortic valve replacement in aortic valve stenosis patients: a combined PET and CMR study. Eur Hear J Cardiovasc Imaging. 2015;16:882–9.
Lee C-B, Goubergrits L, Fernandes JF, Nordmeyer S, Knosalla C, Berger F, et al. Surrogates for myocardial power and power efficiency in patients with aortic valve disease. Sci Rep. 2019;9:1–10.
Nordmeyer S, Lee CB, Goubergrits L, Knosalla C, Berger F, Falk V, et al. Circulatory efficiency in patients with severe aortic valve stenosis before and after aortic valve replacement. J Cardiovasc Magn Reson. 2021;23:1–14.
Qureshi MU, Colebank MJ, Schreier DA, Tabima DM, Haider MA, Chesler NC, et al. Characteristic impedance: Frequency or time domain approach? Physiol Meas. 2018;39:014004.
Acknowledgements
We would like to thank Dr. Mark Butlin from Macquarie University for his contribution to maintain our working equipment, and the staff at the Advanced Cardiac Imaging Centre at St Vincent’s Hospital, Sydney for their support of the study.
Funding
NS is supported by a Cardiac Society of Australia and New Zealand Research Scholarship and an Australian Government Department of Education Research and Training Programme grant.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
DWMM has been an advisory board member for Medtronic and Boston Scientific; has been a consultant to Abbott Vascular, Medtronic; has received research grant support from Abbott Vascular and Medtronic; and is a proctor for Medtronic and Abbott Vascular.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Song, N., Adji, A.I., Hungerford, S.L. et al. Pulsatile energy consumption as a surrogate marker for vascular afterload improves with time post transcatheter aortic valve replacement in patients with aortic stenosis. Hypertens Res 46, 730–741 (2023). https://doi.org/10.1038/s41440-022-01127-4
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/s41440-022-01127-4