Introduction

The diagnosis of open-angle glaucoma is based on intraocular pressure (IOP), optic nerve head (ONH) parameters, and visual field (VF) analysis. The investigation of ONH is an important part of glaucoma diagnosis and management [1]. Estimating the cup-to-disc ratio (CDR) is one of the reference parameters for diagnosing glaucoma and is also a main indicator for assessing glaucomatous changes over time [2, 3]. Therefore, it is essential to carefully document CDR. Non-mydriatic retinal camera (NMRC) is a helpful tool for evaluating ONH parameters [4]. However, evaluating CDR using slit-lamp biomicroscopy and stereoscopic photography remains subjective with moderate to high intra- and inter-observer variability [5, 6].

Optical coherence tomography angiography (OCT-A), as an extension of the OCT, allows measuring and quantifying the macular and peripapillary areas, and provides vessel densities along with structural information. It is well established with previous studies that OCT-A has excellent repeatability and reproducibility [7,8,9,10]. In addition, the reduction of the vessel density (VD) and blood-flow index in glaucomatous eye is well documented by numerous studies [11, 12]. The aim of this study was to compare the morphologic findings of ONH including CDR measured by NMRC with OCT-A and their associations with peripapillary VD.

Materials and methods

The data obtained from the present study were acquired with the approval of the Institutional Clinical Research Ethics Committee of Health Sciences University, Gulhane Medical Faculty, and the study was completed in agreement with the principles of the Declaration of Helsinki. The study was designed retrospectively. Patients with a diagnosis of primary open-angle glaucoma (POAG), glaucoma suspect (GS), or healthy subjects (HS) that presented to our glaucoma clinics between 2018 and 2019 and had complete ophthalmological examination including visual acuity, intraocular pressure, slit-lamp biomicroscopy, and non-dilated fundoscopy along with the VF testing (Humprey Field Analyzer II, Carl Zeiss Meditec, Inc), OCT scans of the RNFL (Spectralis, Heidelberg Engineering, Heidelberg, Germany), OCT-A scans (Angiovue, Optovue Inc., Fremont, CA, USA), and NMRC (Nonmyd WX; Kowa Optimed Inc., Japan) images at the same visit were screened. The diagnosis of the POAG and GS were made in accordance with the European Glaucoma Society’s Terminology and Guidelines for Glaucoma textbook [1]. Increased IOP (>21 mmHg), presence of acquired characteristic glaucomatous changes of the ONH, presence of glaucomatous VF defect, and open angle were the features of the POAG. GS diagnosis was made with the presence of at least one suspicious changes in VF, ONH, and RNFL, and with normal or increased IOP. The exclusion criteria were age under 18 years, refractive error of >±3 diopters, presence of any other ophthalmological pathology that could confound the assessment results (e.g., diabetic or hypertensive retinal diseases, amblyopia, optic nerve anomalies, optic neuropathies other than POAG, and age-related macular degeneration), OCT-A signal strength of <70, and poor NMRC quality.

NMRC imaging

ONH was photographed twice by a glaucoma expert (MTK) and analyzed by two glaucoma experts (MTK and ACY) who were blinded to the patients’ clinical data and OCT-A scans. Kowa 3D NMRC (Nonmyd WX; Kowa Optimed Inc., Japan), which is the stereoscopic fundus camera was used to take non-mydriatic fundus stereographs and both right and left parallactic images simultaneously through a single optical system handling light paths in two directions. [13] Kowa 3D NMRC can display a depth map, which shows the depth of the disc separated by color, a 3D display of the optic disc, diagnostic parameters. In accordance with the algorithm provided by Kowa Optimed Inc., all measurements were conducted after correcting for magnification by refractive error, axial length, and corneal curvature. After photographs were taken, two images with binocular parallax were obtained then, two images were automatically superimposed on the monitor by built-in software, and a stereoscopic 3D image was generated using polarized 3D glasses. The examiner stereoscopically observed the outline of the optic disc displayed on a monitor. The edge of disc is plotted using a computer mouse, and the cup is selected in a similar way, to determine their extent. After that, all other analysis was automatically performed by built-in software. The vertical CDR (VCDR), cup area to disc area ratio (CDAR), rim area (RA), disc area (DA), and cup volume (CV) values were recorded for both photographs after the images were analyzed.

The patients were divided into three groups according to the DA measurement of NMRC: small (area < 1.6 mm2), medium (area = 1.6–2.0 mm2), and large (area > 2.0 mm2) optic discs. [14]

OCT-A ONH scanning

OCT-A scans were obtained using an AngioVue OCT-A system (Optovue Inc., Fremont, CA, USA). This device has an A-scan rate of 70,000/s using a light source of 840 nm. En face images were acquired focusing the ONH (4.5 × 4.5 mm) using the Angio Disc QuickVue module. Each scan consisted of 400 × 400 A-scans with two following B-scans at each fixed location. To reduce motion artifacts, each scan comprised one orthogonal horizontal and vertical scan.

The device provides data on anatomic Bruch membrane opening (BMO)-based neuroretinal rim and ONH, retinal nerve fiber layer thickness (RNFLT), and peripapillary VD in a single scan. After scanning, VCDR, CDAR, RA, DA, and CV values along with the small VD of the whole retinal peripapillary capillary plexus (RPCP) were recorded.

Statistical analyses

Quantitative variables were described as mean and standard deviation, and qualitative variables as percentages. The test-retest variability of the two NMRC photographs of each subject was calculated using the Kappa measure of agreement method. The paired t-test was used to compare the results obtained from the two devices. The magnitude of the disagreement between the two raters of NMRC images and between the NMRC and OCT-A data was estimated as the mean absolute differences in the VCDR, CDAR, RA, DA, and CV values. The inter-rater variability of the NMRC and the agreement between the corresponding NMRC and OCT-A data was calculated with the intraclass coefficient (ICC) [15]. Absolute ICCs based on the mixed model analysis of variance were used in the present study. An ICC value of <0.4 was accepted to indicate poor agreement, 0.4–0.75 represented fair to good agreement, and a value of >0.75 indicated excellent agreement [14, 16]. The ICC calculations were executed before and after the subjects were classified as having small, medium, and large optic discs. The ICC calculations were also performed after the subjects were divided into POAG, GS, and HS groups. The Shapiro–Wilk test was used to determine whether the differences between the corresponding NMRC and OCT-A data were normally distributed. Bland–Altman plots were used to assess the agreement between NMRC and OCT-A in ONH parameters. One-sample t-test was performed for the differences before the Bland–Altman plots were created [17]. Linear regression analyses were undertaken to determine the significance of proportional biases. To analyze the associations of ONH parameters with the small VD of RPCP, after checking the collinearity, age- and IOP-corrected multiple regression models were created. The regression results revealed coefficients (B), 95% confidence interval, and P values. Statistical analyses were conducted using SPSS v. 21 (SPSS, Inc., Chicago, IL, USA). A P value of <0.05 was considered statistically significant.

Results

Of the 312 patients screened, 262 (57% female, mean age 59.4 ± 18.3 years) met the inclusion criteria, and only one eye of each patient (the one with better OCT-A quality) was enrolled for the statistical analysis. A total of 138 eyes were diagnosed with POAG, 63 had a GS, and 61 had no-glaucoma (HS). GSs consisted of 59 eyes with suspected ONH damages without VF and RNFL defects, and 4 with RNFL defect without VF defect and ONH changes. The demographic and clinical data of the subjects were given in Table 1.

Table 1 Demographic and clinical data of the study subjects.

Table 2 presents the mean VCDR, CDAR, DA, RA, and CV values of OCT-A and NMRC and the results of the paired t-test. The devices significantly differed from each other in terms of the DA, RA, and CV measurements (P < 0.001). The difference was the greatest for the CV measurement. The VCDR, CDAR, DA, RA, and CV measurements of the two devices were all positively correlated (R = 0.711, 0.734, 0.746, 0.445, and 0.262, respectively.) The correlations were weaker for RA and CV. In accordance with the paired t-test and Pearson’s correlation test results no further statistical analyses were executed for the DA, RA, and CV measurements.

Table 2 Mean values of the OCT-A and NMRC measurements of the ONH parameters.

The overall mean value of the small VD of RPCP was 48.09%. This parameter was calculated as 46.06% for the subjects with a DA of <1.60 mm2, 48.94% for 1.60–2.00 mm2, and 48.12% for >2.00 mm2. There was no significant difference according to the disc sizes (P > 0.05, ANOVA). The mean small VD values of RPCP of the POAG, GS, and HS were 46.18%, 51.01%, and 52.60%, respectively. The VD of POAG was significantly lower than that of GS (P < 0.001, post-hoc Tukey, ANOVA) and NG (P < 0.001, post-hoc Tukey, ANOVA). There was no significant difference between the GS and HS (P = 0.674, post-hoc Tukey, ANOVA).

Agreement between NMRC and OCT-A

The reliability of the measurements of two NMRC images was excellent (κ = 0.902 and P = 0.764 for the VCDR and κ = 0.856 and P = 0.632 for the CDAR). Excellent agreement in VCDR and CDAR estimates was observed between the two raters of the NMRC data (ICC = 0.891 for CDAR and 0.875 for CDAR). The results of ICC analyses are presented in Tables 3 and 4. There was an excellent agreement for the VCDR and CDAR measurements between the two devices (ICC = 0.801 and 0.793, respectively). The agreement was greater for the larger ONHs in all corresponding measured parameters (Table 3). The agreements were also greater for the POAG patients than GS and HS (Table 4).

Table 3 Intraclass correlation coefficient (ICC) between OCT-A and NMRC.
Table 4 Intraclass correlation coefficient (ICC) between OCT-A and NMRC.

Figure 1 presents the scatter plot of OCT-A against NMRC for VCDR and CDAR, as well as the Bland–Altman plots of these parameters. The one-sample t-test results were insignificant in relation to differences (P = 0.193 for the CDAR and 0.095 for the VCDR). The Bland–Altman plots showed the mean bias ± SD between the measurements of VCDR and CDAR assessed by OCT-A and NMRC as 0.038 ± 0.128 and −0.0001 ± 0.135, respectively. The limits of agreement (LOA) were −0.212–0.288 for VCDR, and −0.264–0.264 for CDAR. The slopes of the VCDR and CDAR regression lines were calculated as 0.30 and 0.31, respectively, indicating that OCT-A tended to produce higher rates than NMRC in larger VCDR and CDAR, and vice versa. The linear regression analyses of proportional bias were statistically significant (P < 0.001 for VCDR and P = 0.012 for CDAR). The wide spans of LOA (0.500 for VCDR and 0.564 for CDAR) and the statistically significant proportional biases suggested poor agreement between the two devices.

Fig. 1: The scatter and Bland-Altman plots of the study subjects.
figure 1

Scatter plot of OCT-A against NMRC for VCDR (column a, top) and CDAR (column b, top) and the Bland–Altman plots for the agreement between OCT-A and NMRC for the VCDR (column a, bottom) and CDAR (column b, bottom). Scatter plots were given with the regression line and the 95% limits of agreements Bland–Altman plots were given with the mean of the difference, 1.96 SDs, and the regression lines. OCT-A optical coherence tomography angiography, NMRC non-mydriatic retinal camera, VCDR vertical cup-to-disc ratio, CDAR cup-to-disc area ratio, SD standard deviation.

Associations with the VD of RPCP

In the multivariate analysis of the VD of global RPCP, we created a model in which we searched for the associations with the age, IOP, VCDR, and CDAR (of the OCT-A device). In this model, we excluded the CDAR due to collinearity with the VCDR, therefore, we created a similar model in which we included CDAR instead of VCDR. The results of these two regression models are given in Table 5. The VCDR and CDAR of OCT-A were found to be strongly associated with the small VD of RPCP (B = −23.051 for VCDR and −20.700 for CDAR).

Table 5 The linear regression models for the RPCP small VD.

Discussion

The evaluation of the ONH parameters is of paramount importance in glaucoma management [3, 18,19,20,21]. There are numerous methods to investigate the morphological findings of ONH to identify early glaucomatous changes [22]. Stereophototography has been widely used for decades and was once considered as a gold standard technique for this purpose. NMRC, which is a stereoscopic fundus camera used to take non-mydriatic fundus stereographs, can display a depth map that shows the depth of the disc separated by color, provides a 3D display of the optic disc, and automatically calculates diagnostic parameters [20, 23,24,25,26]. OCT-A, as an extension of the OCT technology, is relatively new in the ophthalmology practice and has recently started to be used at an increasing frequency in glaucoma management. Previous studies on OCT-A described reduction in VD and the flow index of ONH, peripapillary area, and macular region [12, 27,28,29,30,31,32]. In the present study, we first investigated the morphological findings of ONH using OCT-A and NMRC, and then compared the results obtained from these two devices. Secondly, we explored the associations between peripapillary VD based on the ONH parameters assessed by these methods.

The agreement between OCT-A and NMRC was found to be poor in the present study. Even though ICC revealed excellent agreement between NMRC and OCT-A in terms of VCDR and CDAR (ICC = 0.801 and 0.793, respectively), the Bland–Altman plots showed poor agreement with a wide span of LOA (0.500 and 0.528, respectively) and statistically significant proportional bias values. Our results were similar with the previous studies which investigated the agreement between OCT and stereoscopic photography [14, 33]. In addition, we investigate whether there was any agreement between the two devices in terms of RA, DA, and CV; however, we found the agreement to be poor even in relation to VCDR and CDAR. Chan et al. compared three different devices: OCT, Heidelberg retina tomography (HRT), and stereoscopic photography. They reported that the disagreement in the CDR measurements between these modalities altered the five-year risk estimation of glaucoma progression in untreated ocular hypertension eyes [33]. Chan showed the importance of the same device we used for the management of GSs, emphasizing that it could alter treatment decision. Mwanza et al. also determined that OCT and stereoscopic photography had disagreement concerning the horizontal and vertical CDR measurements [14]. Both Chan’s and Mwanza’s studies suggest that the cause of this disagreement could be related to the different disc margin determination, and the assessment of ONH in stereoscopic photography relies on defining the visible disc margin by an expert; on the other hand, the OCT devices they used in their studies, similar to our OCT-A device, recognize BMO, which is accepted as the true anatomical outer border of the neuroretinal rim as the disc margin. Reis et al. also described that the disc margin detected on photographs fell outside BMO in all sectors but the inferotemporal sector where the opposite situation was observed. Also the fact that the RA and CV measurements differ between two devices means that the delineation of the cup and the margin between the cup and rim is different. This proves that selecting different algorithms that determines the disc margin, visible or BMO based, would alter all ONH parameters calculations.

The analyses of ICC revealed that disc size was very important for the agreement between the devices. The agreement was inferior in smaller optic discs and better for the larger ones. Jayasundera et al. and Correnti et al. reported similar results in their comparison of HRT, stereoscopic photography, and digital stereoscopic optic disc camera [34, 35]. This could be the reason why the agreement between the two devices in terms of the VCDR and CDAR measurements was better for the POAG patients than GSs and HS.

The main and most important outcome of this study was that the ONH parameters assessed by OCT-A were found to be closely related to the peripapillary VD independent of age and IOP. Enders et al. investigated the association between the ONH parameters assessed by OCT with VD, and they also found out that CDR and VD were strongly and negatively correlated [36].

The first limitation of this study is that comparing ONH parameters between NMRC and OCT-A could cause potential bias because only the OCT-A could calculate the BMO-based neuroretinal rim, however we performed this study because we also aimed to investigate the relation of peripapillary VD with ONH parameters, which is calculated with different algorithms. In this study we found associations between the VD and the ONH parameters, but it is also essential to search for the associations of the VF changes with the vessel densities. Lastly, using sectoral VD data and stratifying patients into subgroups in accordance with the severity of the glaucoma should be kept in mind while planning future studies in this matter.

In conclusion, in this retrospective study, we found that there was disagreement between NMRC and OCT-A in terms of the ONH measurements, including VCDR and CDAR. The agreement was better in larger disc sizes and patients with POAG. The measurements of the ONH parameters obtained from OCT-A were closely related to the peripapillary VD. Future prospective follow-up studies are needed to provide a better understanding of the alterations of VD through the changes in ONH parameters, which could be detected by NMRC and OCT-A.

Summary

What was known before

  • OCT-A is a new technology for the assessment of the VD and in glaucoma, OCT-A is very helpful to manage the patients.

What this study adds

  • OCT-A could also give reliable information about the ONH parameters. But the data of the different devices such as NMRC should not be used interchangeably.