Abstract
A novel technique based on the remote-focusing concept, using a galvanometer scanner combined with a self-fabricated “step mirror” or “tilted mirror” to transform fast lateral scanning into axial scanning, was reported as a new solution for fast, subcellular, 3D fluorescence imaging.
Fluorescence imaging techniques, such as multiphoton microscopy (MPM) and light-sheet fluorescence microscopy (LSFM), have led to rapid advances in cell biology, developmental biology, neuroscience and many other fields. Thanks to thin optical sectioning, high spatial and temporal resolution, and deep penetration into scattering tissues, these imaging technologies have allowed an unprecedented look into many details of subcellular structure and dynamics in living animals1. The focus of many studies has recently shifted from the single-cell level to the populational level and from two dimensions (2D) to three dimensions (3D). However, the side effects of thin optical sectioning are limited cell yield if imaging is performed in only a single plane. Fast axial scanning is therefore highly desirable. In recent years, many novel techniques have been presented to approach this goal. For example, adding an electrical tunable lens (ETL) in the conjugated pupil plane has become a popular way to perform rapid Z-scanning2,3, and ETLs have been integrated into several commercial systems. Remote focusing by introducing a module with another objective and piston-moving mirror is another elegant method with the additional benefit of aberration compensation4. Spatial/temporal multiplexing is another recent rapid z-scanning technology that has extended the axial scanning speed beyond the kHz range in a semisynchronous fashion5,6,7 (Fig. 1).
To address this grand challenge, Fiolka et al. provide a novel solution based on remote focusing8. Instead of using a piston actuator to drive the movement of the mirror forward and backward, they use a galvanometer scanner combined with a self-fabricated “step mirror” or “tilted mirror” to transform fast lateral scanning into axial scanning. Compared to the previous approaches, the new method has several advantages. First, the lateral scanning device, such as the galvanometer scanner, can achieve speeds ~one order of magnitude higher than the ETL. Thanks to the well-established controlling hardware, the galvanometer also features a more linear response and more reliable performance. Second, more galvanometer options are available from various companies than piston-moving mirror module options. Third, the axial scanning range can be highly customized by designing and fabricating different “step/tilted mirrors”. An especially tempting prospect of this technology is that in principle, it could be combined with other techniques, such as spatial/temporal multiplexing or holographic pattern generation.
With the impressive results that the authors showed here, we believe that this technique may be very useful in many applications, such as the study of intracellular 3D dynamics, multilayer neuron network recording, and functional imaging of fast-moving organisms, such as zebrafish embryos, C. elegans, or fly larvae. At the same time, however, we also envision some open questions and potential for further development of this technique. For example, can step/tilted mirrors be easily fabricated at a reasonable cost so that this technique can be widely applied? Can this technique be conveniently adapted to upgrade existing custom or commercial systems? For future developments, an interchangeable mechanical design allowing users to choose freely among multiple step/tilted mirrors with different scanning ranges may be interesting.
References
Liu, T. L. et al. Observing the cell in its native state: imaging subcellular dynamics in multicellular organisms. Science 360, eaaq1392 (2018).
Fahrbach, F. O. et al. Rapid 3D light-sheet microscopy with a tunable lens. Opt. Express 21, 21010–21026 (2013).
Grewe, B. F. et al. Fast two-layer two-photon imaging of neuronal cell populations using an electrically tunable lens. Biomed. Opt. Express 2, 2035–2046 (2011).
Botcherby, E. J. et al. Aberration-free three-dimensional multiphoton imaging of neuronal activity at kHz rates. Proc. Natl Acad. Sci. USA 109, 2919–2924 (2012).
Wu, J. L. et al. Kilohertz two-photon fluorescence microscopy imaging of neural activity in vivo. Nat. Methods 17, 287–290 (2020).
Beaulieu, D. R. et al. Simultaneous multiplane imaging with reverberation two-photon microscopy. Nat. Methods 17, 283–286 (2020).
Weisenburger, S. et al. Volumetric Ca2+ imaging in the mouse brain using hybrid multiplexed sculpted light microscopy. Cell 177, 1050–1066.E14 (2019).
Chakraborty, T. et al. Converting lateral scanning into axial focusing to speed up three-dimensional microscopy. Light.: Sci. Appl. 9, 165 (2020).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zong, W. New solution for fast axial scanning in fluorescence microscopy. Light Sci Appl 9, 199 (2020). https://doi.org/10.1038/s41377-020-00442-0
Published:
DOI: https://doi.org/10.1038/s41377-020-00442-0