Introduction

Magnetic resonance force microscopy (MRFM) is a scanning probe technique based on coupling a soft magnetic tipped cantilever to the spins of a sample, and measuring the tiny force arising from spin manipulation through Magnetic Resonance techniques. The spatial selectivity in MRFM is provided by the strong field gradient generated by the magnetic tip, similar to external field gradients used in conventional MRI. MRFM is nowadays considered one of the most viable routes towards three-dimensional imaging of biomolecules or nanostructures with atomic resolution. Although this ambitious goal has not been reached yet, several milestones have already been demonstrated, including mechanical detection of a single electron spin1 and 3D nuclear spin imaging of a virus with a few nanometres resolution2.

Present limiting factors on MRFM resolution are the thermal-noise-limited force sensitivity of the cantilever, and a surface-induced frequency and force noise, which is poorly understood3. Besides a better characterization of the latter effect, further improvements require softer and less dissipative mechanical sensors and if possible reduction of operating temperature. So far, MRFM has been demonstrated at temperatures as low as 280 mK4. Cooling below 100 mK would both improve the force sensitivity and increase the Boltzmann spin polarization, thereby increasing the signal. The last consideration does not hold for small nuclear spin ensembles, for which statistical polarization becomes dominant for small spin ensemble5, unless temperatures below 1 mK can be achieved.

A known issue for ultralow temperature operation is the need for a suitable cantilever detection technique. Interferometric detection is an option, but exceedingly low power would be needed to avoid cantilever overheating, unless advanced dielectric mirrors are used to reduce photon absorption in the mirrors. We have recently demonstrated a superconducting quantum interference device (SQUID)-based detection technique, capable of measuring the thermomechanical fluctuations of a cantilever down to 25 mK6. The scheme is based on detecting the flux change induced in a pick-up coil by the motion of the magnet attached to the cantilever.

In this article, we demonstrate for the first time that this detection system can be readily integrated in a real MRFM experiment. As a case study, we have mechanically detected paramagnetic centres located close to the surface of the chip supporting the superconducting pick-up coil, and subsequently we have performed a MRFM saturation-recovery experiment. Paramagnetic centres appear to be almost ubiquitous on thin film surfaces and interfaces6, with typical density of the order of 0.1–1 nm−2, and have been recently a subject of various studies, because they are possibly related to the 1/f magnetic noise observed in SQUIDs at millikelvin temperatures7,8,9,10. The same magnetic noise appears to be a major source of dephasing in some superconducting qubits11,12, and in single dopant qubits13,14. Several different mechanisms for 1/f noise generation have been recently proposed, based on electron hopping15,16 or spin diffusion17. Our measurements support a picture in which spin diffusion has a key role in the dynamics of surface paramagnetic centres at millikelvin temperatures.

Results

Experimental scheme

A general scheme of our experiment is shown in Figure 1. An ultrasoft microfabricated silicon cantilever with a micron-size spherical magnetized particle attached to its end (from now on, the magnet) is approached vertically to the sample chip surface. The cantilever stiffness k is in the 10−4 N m−1 range and the magnet moment μ is of the order 10−11 A m−1. Assuming the z-axis perpendicular to the chip surface, both the magnetic moment of the magnet and the cantilever fundamental mode deflection are oriented along the x-axis. The motion of the magnet is detected by a superconducting pick-up coil, which is connected to a SQUID amplifier6. The xy position is aligned at room temperature so that the cantilever will still be several tens of microns away from the superconductor lines, when it approaches the surface. The magnet thus interacts mostly with the oxidized silicon surface of the chip.

Figure 1: Experimental scheme.
figure 1

(a) Schematic representation of the experimental setup. An ultrasoft cantilever (stiffness k≈10−4 N m−1) with a ferromagnetic particle, the magnet, attached on its end (magnetic moment μ≈10−11 A m−1) is approached in a vertical configuration to the surface of the detection chip. A superconducting niobium pick-up coil deposited on the chip surface is connected to a dc SQUID. The pick-up coil detects the motion of the magnet through the position-dependent magnetic flux φ coupled by the magnet into the coil. The magnet generates a quasistatic field which couples to the sample, constituted by paramagnetic electron spins close to the Si–SiO2 surface of the chip. A rf current Irf flowing in the superconducting niobium microwire generates the B1 field at microwave frequency, which excites the electron spins in the sample. (b) Optical microscope picture of the assembled setup for the first experimental run, with a multiturn pick-up coil and without the microwire. The scale bar is 200 μm.

We measured the frequency of the cantilever fundamental mode under different experimental conditions, varying the temperature and the distance from the sample, and exposing the sample to microwave radiation. Far from the surface, the cantilever frequency is around 3 kHz. The quality factor is about 3×104 and is reduced to about 1000 close the the surface at very low temperature. A detailed report and discussion on the dissipation measurements will be the subject of a future paper. We present data from two experimental runs with slightly different cantilevers, magnets and samples (details of which are described in the Methods).

Static frequency shift

In our first experiment, we accurately measured the resonant frequency of the cantilever fundamental mode fr as a function of both temperature and distance from the surface, constituted by a 300-nm thick SiO2 layer on a Si substrate18. We excited the cantilever with a piezoelectric actuator and performed both ringdown measurements and frequency sweeps. Representative experimental results are shown in Figure 2a, where the frequency shift induced by the sample surface is plotted as a function of the distance d between the magnet centre and the surface. The power-law dependence and the persistence of a relatively large effect at distances of the order of a few microns rule out short-range van der Waals interactions and point to either electrical or magnetic interactions. Figure 2b shows the frequency shift as a function of temperature for four different distances between the magnet centre and the surface. The data of each curve are normalized to their maximum value measured at low temperature to provide an easy comparison of the curves. A Curie-like 1/T dependence is clearly observed for temperatures higher than 100 mK. This feature is common to all curves, suggesting that the cantilever is indeed coupled to a paramagnetic spin system, for which the polarization increases with decreasing temperature. The low-temperature saturation in Figure 2b, together with the anomaly at low distance and low temperature in Figure 2a could indicate either a full polarization of the spins, or a saturation of the effective spin bath temperature.

Figure 2: Static frequency shift as a function of magnet-surface distance and temperature.
figure 2

(a) Resonant frequency shift Δfr=frfr0 in experiment 1 as a function of the distance between magnet centre and sample surface, for two representative bath temperatures, T=11 mK and T=450 mK. Here, fr0 is the resonant frequency measured when the magnet is far away from the surface. The straight line shows a log–log linear fit of the 450 mK data, with slope coefficient n=−5.0±0.2. (b) Resonant frequency shift Δfr in experiment 1, as a function of temperature, for four different distances between magnet centre and sample surface,6.5 μm (blue), 4.7 μm (green), 2.5 μm (red), 1.8 μm (black), respectively. To provide an easy comparison of the curves, the data are normalized to the maximum of each curve. The dashed violet line represents the 1/T Curie-like behaviour.

To account for these observations, we built a simple model to estimate the effect of a layer of paramagnetic spins on the sample surface. Assuming that the longitudinal relaxation time is much longer than the cantilever period, the spring constant induced in the cantilever by an individual electron spin with magnetic moment μe is given by ks=−μePBxx, where P=tanh(μeB0/kBT) is the Boltzmann polarization factor and Bxx is the second derivative of the magnet field B0 with respect to magnet position. As illustrated in Figure 3, ks is positive for spins close to the magnet and negative for spins far away from the magnet. Integration over a uniform surface layer with spin density σ gives the total spring constant change, from which the frequency shift can be calculated.

Figure 3: Model of the magnet coupled to a layer of electron spins.
figure 3

(a) Scheme of the magnet on the cantilever approaching the surface. It illustrates the fact that surface electron spins are polarized by the magnet dipole field in different directions, and therefore contribute to the cantilever spring constant with both positive and negative sign, depending on the position. (b) Simulated change of the cantilever spring constant ks induced by a spin located on the surface (black thick line) and magnetic resonance frequency fB0/2π (orange line) as a function of the coordinate x, for y=0. The parameters correspond to the conditions of experiment 2, with the centre of the magnet located at a distance d=3.6 μm from the surface above the origin of the xy plane. The vertical stripes indicate the location of the resonant slices determined by the labelled microwave frequencies.

According to the model, we expect a dependence on distance as dn, with −6<n<−5 slightly dependent on distance and temperature. Experimental data of Figure 2a are consistent with this prediction. Furthermore the model predicts the observed high-temperature 1/T dependence in Figure 2b. A fit of the data in the 1/T region yields a density of (0.6±0.1) spins per nm2. This figure is reasonably close to the estimates of paramagnetic defects densities in different thin film structures using other techniques9,19, in particular it is consistent with recent estimates of the density of dangling bond paramagnetic centres, so called Pb centres, on Si–SiO2 interfaces using single dopant techniques13. As we are probing a Si–SiO2 interface, the magnet is likely coupled to this latter kind of defects. In principle, electron spins or nuclear spins in the bulk could be coupled to the magnet as well, but for an undoped silicon substrate their effect has been estimated to be at least two orders of magnitude smaller.

It is worth mentioning that several past studies on non-contact interaction between a cantilever and a surface, focusing mostly on non-contact friction, found effects due to an interaction of electrical nature, rather than magnetic20,21. However, non-contact electrical effects are usually found to decrease with decreasing temperature and show a weaker power-law dependence on distance (n≈1–3) than that we have measured. Therefore, our claim that the measured frequency shift is due to a magnetic interaction with a layer of spins is not in contradiction with previous works. In fact, our measurements are performed at much lower temperature, and we use a relatively large magnet, so that magnetic effects are strongly enhanced.

Microwave-induced frequency shift

In the second experiment, we have applied magnetic resonance pulses to selectively address different spin subsets. In this experiment we used a different detection coil chip, fabricated in a different laboratory with different fabrication protocols. A frequency shift measurement as a function of temperature has been performed as in experiment 1, for a fixed distance d=3.6 μm, which yields a spin density of σ=(0.20±0.05) spins per nm2 slightly lower compared with experiment 1. Spin manipulation was enabled by a niobium microwire, 15 μm wide, deposited at 50 μm from the detection coil, through which we could deliver an effective microwave power at low temperature up to −17 dBm in a frequency range up to 4 GHz. In this experiment, the magnet on the cantilever was located at a relatively large distance of 70 μm from the wire.

The microwave field B11 μT generated in the vicinity of the magnet was used to saturate the polarized spins, using the so-called CERMIT saturation-recovery protocol22. The microwave power is switched on for a given time, of the order of the longitudinal relaxation time T1, ranging from a few tens of milliseconds up to a few seconds. The microwave frequency frf=ωrf/2π defines the resonant slice of spins satisfying the resonant condition ωrfB0, where γ is the gyromagnetic ratio, within a bandwidth determined by the homogeneous broadening. The microwave pulse causes saturation and hence suppression of the polarization of the resonant spins. Subsequently the microwave power is switched off and the thermal equilibrium polarization is allowed to recover. According to standard Bloch equations, the timescale of the polarization suppression process is of the order of T1 in the undersaturated regime, and 1/(γ2B12T2) in the saturation regime, where T2 is the transverse relaxation time, whereas the timescale of the recovery process is given by T1. We note that in our scheme B1 is parallel to the z axis, and is not purely transverse to the inhomogeneous dipole field B0 generated by the magnet, as in conventional magnetic resonance. Although this would clearly be an issue in a pulsed magnetic resonance scheme, it is not a major concern for a saturation experiment. All spins in the resonant slice will be saturated, except a small fraction of them which feel an almost vertical magnetic field B0 field.

To monitor the frequency shift during the saturation-recovery transients, the cantilever is self-oscillated in a phase-locked loop with a 50-Hz bandwidth. We maintained a constant cantilever distance d=3.6 μm and therefore a constant field B0 throughout the following measurements, although we could address different resonant slices by varying the microwave frequency.

The resulting saturation-recovery curves for different microwave frequencies frf and a nominal power of −23 dBm are shown in Figure 4. The measurements were done at a bath temperature of 30 mK. It is apparent that addressing different resonant spins yields qualitatively different behaviours. For frf=1.25 GHz the frequency shift is positive, and both transients are slightly non-exponential, with a faster initial behaviour followed by a slower relaxation. For frf=4.2 GHz we find a negative signal and exponential relaxation. The physical origin of the sign change is easily understood by examining Figure 3. Frequencies around 4 GHz select high field spins close to the magnet, which give a positive contribution to the frequency shift, whereas frequencies around 1 GHz select low-field spins far away from the magnet, giving a negative contribution. Temporarily removing the polarization of these spins gives the signal with the signs we have observed. Finally, at an intermediate frequency of 2.7 GHz we observe a mixed behaviour, with an initially positive transient, which then develops into a net negative frequency shift. In general, the recovery relaxation is characterized by a slow time constant of the order of 1–2 s, slightly dependent on the resonant slice.

Figure 4: Microwave-induced frequency shift saturation and recovery.
figure 4

Cantilever frequency shift as a function of time for three different frequencies of the applied microwave pulses, respectively, from the top to the bottom 4.2 GHz (red), 2.7 GHz (blue) and 1.25 GHz (green). Here, a microwave power of −23 dBm (corresponding to a field ≈1 μT) is switched on at t=0 and switched off at t=1 s. The shaded region indicates the time during which the microwave power is on. At 1.25 GHz the microwave signal induces a positive transient with non-exponential features, whereas the 4.2 GHz pulse induces a negative exponential transient. At an intermediate frequency of 2.7 GHz a double transient with opposite sign is observed both when switching on and off the microwave.

Discussion

The observed behaviour at intermediate frequency can hardly be explained in terms of a simple Bloch equation model. Instead, we interpret these data as evidence of spin diffusion mediated by flip-flop processes, which spreads the local suppression of polarization throughout the sample. Spin diffusion also explains the non-exponential behaviour at 1.25 GHz. We have checked the faster saturation transient at 1.25 GHz for shorter microwave pulses, and found that the amplitude of the fast frequency response is roughly independent of the microwave power for powers higher than −23 dBm. This rules out direct spurious cross-talk and suggests that the fast response is instead related with the saturation of local spins, followed by a slower diffusion to non-resonant spins.

We now discuss whether a spin diffusion mechanism is expected to be significant in our experiment. A theory of spin diffusion in an inhomogeneous field was developed in the framework of NMR experiments23. A key point is that a flip-flop process between two spins in the presence of a magnetic field gradient is not energy conserving, hence the diffusion requires that the spin dipolar bath takes up the excess energy. An important consequence is that diffusion can be quenched by sufficiently large gradients24. Application of the model to real situations is generally tricky25, requiring large simulations, but some basic features can be estimated by simple arguments. The diffusion process is controlled by the diffusion constant D=Wa2. Here, a−1/2 is the mean distance between neighbouring spins, and W is the rate of a flip-flop process between two spins. W can be estimated as W≈δ/30, with the resonant line width δ given approximately by26:

For our spin density, we estimate a diffusion constant D≈20 μm2 s−1. A local perturbation of the polarization from its equilibrium value will diffuse across the sample to an approximate diffusion length LD=(DT1)1/2. If we assume the measured recovery relaxation time τ ≈1 s as an approximate estimation of T1, the corresponding diffusion length becomes LD≈4.5 μm, which is of the order of the characteristic length of the spin system probed by the magnet. This means that spin diffusion can indeed spread the polarization transient throughout the sample. A simplified model supporting this picture qualitatively is described in the Supplementary Note, and relevant numerical simulations are presented in Supplementary Figures S1 and S2. On the other hand, if T1 was much shorter, the diffusion length would be suppressed as direct or indirect spin-phonon relaxation would become dominant. As we expected a temperature dependence of T1, we measured the relaxation time τ in the recovery transient as a function of temperature (Fig. 5). We find that the non-exponential features of the transient are suppressed at higher temperatures and the relaxation rate 1/τ is increasing approximately according to a power law. A fit of the experimental data with the function A+BTn gives for the exponent the value n=2.5±0.2. Here, a finite value of A accounts for a possible saturation of the spin equilibrium temperature. The temperature dependence of 1/τ is roughly comparable with previous EPR studies on Pb centres27, which have determined 1/T1≈10 s−1 at 400 mK and a power-law exponent 2.3<n<3.5.

Figure 5: Recovery relaxation rate as a function of temperature.
figure 5

Relaxation rate of the frequency-shift signal in the recovery transient (Fig. 4) 1/τ as a function of the bath temperature T. The solid line represents the best fit of the data with the function A+BTn, which gives for the exponent of the temperature-dependent term the value n=2.5±0.2.

Because spin diffusion is spreading polarization across the sample, it is hard to define the number of spins generating the frequency shift signals shown in Figure 4. However, we can try to estimate an effective number of spins at least for the resonant slice at 1.25 GHz, which shows evidence of saturation of the resonant spins. Referring to Figure 3, the thickness of the resonant slice can be estimated from the spin resonance linewidth δ as Δx=δ/2π∂f/∂x, where f is the spin magnetic resonance frequency (orange line). Approximating the resonant slice with a circular annulus of radius x and thickness Δx, the total number of spins in the slice becomes N≈2πσxΔx. Each spin contributes with ks (black line in Fig. 3) to the total spring constant change Δk. The expected magnitude of the microwave-induced cantilever frequency shift, under the assumption that the resonant spins are saturated, is then given by:

For the 1.25 GHz slice we find Δx≈43 nm, N≈3.1×105 and an expected frequency shift Δfr≈45 mHz. This rough estimation is in reasonable agreement (within a factor 2) with the experimentally measured signal shown in Figure 4, in particular with the amplitude of the fast frequency response. This indicates that the microwave-induced frequency shift is consistent with the static frequency shift, from which we have inferred the spin density σ, giving further support to the hypothesis that the static effect is indeed due to a magnetic interaction with the paramagnetic spins.

We now discuss the possible connections of our findings with the 1/f magnetic noise observed in thin-film devices such as SQUIDs and superconducting qubits. Recent models point to fluctuations of unpaired paramagnetic electron spins as a likely origin of this excess noise. In particular, a model of 1/f noise from spin diffusion in paramagnetic defects located in the superconductor-insulator interface has been proposed17. In this model, the emergence of 1/f noise is related to long-range low-frequency spin correlations driven by spin diffusion, whose typical scale is determined by the diffusion length. Diffusion constants in the range of 10–100 μm2 s−1, as estimated in our experiment, imply correlation times of seconds on a 10-μm length scale, which should couple to the typical size of most thin film devices very well. Thus, we suggest that Pb centres in the insulator interface could be partly responsible for the magnetic noise generation as well. This is in contrast with ref. 17, where spin densities two orders of magnitude smaller were assumed for paramagnetic centres in the insulator, leading to the conclusion that their role is negligible. Another relevant point is the enhancement of spin diffusion by cooling down to a temperature of the order of 100 mK, due to the observed strong dependence of competing relaxation mechanisms on temperature (Fig. 5). This is in agreement with the typical feature of magnetic 1/f noise in SQUIDs and qubits, which increases sharply only when the temperature is reduced below 1 K7,8.

We conclude with the implications of these results for MRFM at ultralow temperature. The interaction of paramagnetic centres in the Si–SiO2 interface with the magnetic tip and with the spins in a sample leads to undesired cantilever damping and shortened spin coherence time, respectively. To advance further towards single atom imaging by MRFM, it will be necessary to either develop methods to shield the experiment from the Pb centres, for example, by applying a background magnetic field in combination with low temperatures to freeze Pb centres dynamics, or to resort to substrates that contain a sufficiently low concentration of paramagnetic defects.

Methods

Experimental apparatus

In the first experiment, the cantilever was a micromachined IBM-type ultrasoft silicon beam28 with length, width and thickness 120 , 5 and 100 nm, respectively. The magnet was a 3.3-μm diameter spherical particle from a commercial neodymium-alloy powder (Magnequench, type MQP-S-11-9) with saturation magnetization μ0Mr=(1.3±0.1) T29. The particle is glued to the free end of the cantilever by means of a platinum electron beam deposition and is subsequently magnetized in a 5T field at room temperature. Its magnetic moment is estimated from the volume and the saturation magnetization to be μ≈2×10−11 A m−1. In the second experiment, the length of the cantilever was 90 μm and the magnet was 5.2 μm in diameter, with magnetic moment μ≈8×10−11 A m−1.

We also used two different detection-sample silicon chips. The first chip supports a multiturn niobium coil fabricated in a multilayer process, with inner and outer size of 220 μm and 660 μm, respectively, a total number of 44 turns and an estimated inductance of 0.6 μH. After fabrication of the coil, a 300-nm thick insulating layer of silicon dioxide is deposited on the top of the chip by rf sputtering18. In the second experiment, we used a different chip, containing a single loop square Nb pick-up coil, with 100 μm size, and a 5 μm niobium microwire for microwave excitation. There was no intentional oxide layer in addition to the native oxide.

The flux change in the pick-up coil is measured by a two-stage SQUID amplifier with input inductance Li=1.6 μH and a constant flux noise of 0.6 below 500 mK6. In the second experiment we used an intermediate superconducting transformer to match the pick-up coil and SQUID inductance. The overall sensitivity of the SQUID-based displacement detector was a few pm/ in both experiments.

The cantilever was mounted on a piezo-motor, which allows to approach the chip surface with 170 nm steps. During the microwave saturation-recovery measurements, a piezo-tube actuator was used to drive the cantilever fundamental mode and a phase-locked loop was used to track the cantilever resonant frequency. The position of the cantilever was aligned at room temperature so that the magnet approaches the sample on the dielectric in a position, which provides a reasonable coupling between the magnet motion and the coil, while maintaining the magnet more than 20 μm away from the superconducting lines. The whole assembly was mounted on a vibration isolation stage, which is thermalized to the mixing chamber of a cryofree dilution refrigerator with a base temperature of 10 mK.