Introduction

Eyelid tension is an important factor in the pathogenesis of many acquired lid diseases. Adequate upper lid tension is important in maintaining normal lid-tear resurfacing1 mechanisms and in preventing ocular exposure. Upper eyelid tension has been shown to decrease with age.2 Floppy eyelid syndrome3 is an example of functional loss where upper lid tension is reduced. Reduced lower eyelid tension is a significant factor in the pathogenesis of involutional ectropion and entropion.4 In patients with watery eyes, lower eyelid tension may be reduced.5, 6

Lid tension has been evaluated using mechanical7, 8, 9, 10 and ultrasonic techniques11 to quantify upper lid tension, but lower eyelid tensometry (LET) has previously been evaluated using subjective techniques only.12, 13 To our knowledge, there are no published data on the magnitude of human lower eyelid tension.

In physical terms, tension is defined as the first derivative of the force over displacement function of any elastic material under stress. For the purpose of this study, the measurement principle is based on displacing the lid while recording exerted force as a function of displaced distance.

Determination of lid tension is clearly relevant in determining the natural history of lid disease, and in ocular plastic surgery. The aim of this study was to establish normative data and intra- and inter-occasion repeatability data for LET for displacement in the postero-anterior (PA), temporal, and nasal directions in younger normal subjects, to examine the effect of age on PA eyelid tension, and to establish the effect of gender on eyelid tension in age-matched subjects.

Materials and methods

Instrumentation

A novel eyelid tensometer designed and built to measure eyelid tension objectively was used in this study. The instrument comprises an eyelash clamp device, two stepper motors, two force transducers, an electronic interface box, and computer for instrument control and data acquisition, and a patient fixation device (Figure 1). The instrument was validated and calibrated in an earlier study14 by measuring the force and displacement values using a series of weights covering the measurement range of 5 mN (milliNewtons) to 2 N. Validation and calibration were carried out to correct for systematic errors, to verify measurement accuracy, to confirm linearity of the force transducers across the measurement range, and to demonstrate that any effect of hysteresis was negligible.

Figure 1
figure 1

Eyelid tensometer instrument.

Subject selection

A total of 32 Caucasian subjects were selected to participate in the study. The younger age group (Group 1, n=12) comprised six males and six females with a mean age of 29 years (range 23–37). The older age group (Group 2, n=20) comprised 10 males and 10 females with a mean age of 74 years (range 61–81). The subjects had no pre-existing ocular pathology, reported no eye symptoms and were using no topical medication. The investigation was conducted in accordance with the tenets of the Declaration of Helsinki. Written consent was obtained following explanation of the study procedure. Approval for the procedures from the University of New South Wales Ethics Committee was obtained prior to the investigation.

Study design

Group 1: LET was carried out on each subject on two occasions, several days apart. Each measurement was carried out twice within each occasion. The lash clamp was not removed between different eyelid movements unless the subject expressed discomfort. Group 2: PA tension only was measured on each subject on a single occasion. Two measurements were taken for each eye.

Procedures

The subject was seated at the tensometer, and the machine action explained. The side of LET to be performed first was decided by the toss of a coin. The subject was then asked to bite firmly onto the tensometer bite bar, and to rest the forehead securely onto the forehead rest of the tensometer. The chinrest was adjusted so that the subject was comfortable and the head completely steady on the tensometer. Topical anaesthesia was not used and subjects were asked to relax their lids as far as possible during the assessment. The lower eyelid lashes were clamped in the mid-pupillary line using the eyelash gripper (Figure 2), and eyelid tension neutralised by observation initially by the operator, and then electronically.

Figure 2
figure 2

Eyelash clamp attached to the patient's lower eyelid during eyelid tensometry.

For Group 1, LET was performed in the PA, temporal, and nasal directions, with a maximum force of 0.7 N, a maximum displacement of 3.5 mm, and with the speed of displacement being 1 mm/second. That is, the lash clamp was moved in the PA, temporal, and nasal directions to displace the lid away from the globe anteriorly (PA), and perpendicularly to the visual axis for the temporal and nasal directions of tension, respectively. Forward force displacement data were captured for analysis. The second eye was then similarly measured. On the second occasion, the LET measurements were repeated using the contralateral eye first. Eyelid tension in mN/mm was calculated as the slope of the best-fit linear curve over the full displacement range.

For both groups over this displacement range, tension measurements were adequately fitted using a linear curve fit, since increased and nonlinear tension were observed with displacements larger than 3.5 mm.

Data analysis

There were differences in the handling of the data between the two subject groups. The younger group was utilised for multiple and repeated measurements to establish repeatability of the eyelid tension measurements, gender effects for a range of directions and to establish the effects of laterality. These findings informed the data collection for the older age group. Thus, duplicate measurements were taken for a single displacement direction (PA) on a single occasion in the older age group, as PA proved to be the most highly correlated direction for LET. Group 1: Associations between measurements made on occasions 1 and 2 and between right and left eyes were examined using Pearson's parametric correlation. The coefficient of repeatability was estimated as 2.20 × the within-subject standard deviation.15 Comparisons between males and females were made using a one-way ANOVA. Group 2: The mean PA tension for the right and left eyes was estimated. Gender and age effects between groups 1 and 2 were examined using a grouped t-test.

Results

General results

For all younger subjects (Group 1), the mean tension for temporal displacement (temporal tension) was 7.8±2.9 mN/mm, nasal displacement (nasal tension) was 13.0±4.6 mN/mm, while PA tension was 11.2±5.3 mN/mm. Nasal displacement was significantly higher than temporal (P<0.005). In the older subject group (Group 2), mean PA tension was 11.1±3.4 mN/mm. Differences in PA tension with age were not significant (Table 1).

Table 1 Mean postero-anterior (PA) eyelid tension with age and gender

Laterality effects

There was moderate correlation between PA measurements from the right and left eyes, where r=0.48, P=0.001. Nasal and temporal tension measurements were not well correlated between the right and left eyes (P>0.05). The order of testing did not influence the eyelid tension: in other words, whether the right or left eye was tested first did not influence the test result.

Interoccasion repeatability

Measurements of PA tension in Group 1 appeared to be highly correlated between occasions (right eye, r=0.82, P<0.005 (Figure 3); left eye, r=0.84, P<0.05). Agreement between the two repeats is shown in Figure 4. The coefficient of repeatability, defined as 2.20 × the within-subject standard deviation15 for PA measurements, was 1.8 mN/mm, which is the value below which the difference between two repeated measurements would be expected to fall with 95% probability.

Figure 3
figure 3

Graph showing interoccasion variability in PA tension in Group 1. A good correlation is shown between measurements taken on two occasions, r=0.821, P=0.001.

Figure 4
figure 4

Graph showing agreement between two repeated measures for PA displacement in Group 1. Difference between the two measurements is plotted against the mean. The solid lines show the coefficient of repeatability for two repeated PA displacement measurements. Bias is −0.19 mN/mm and 95% confidence intervals ±1.77 mN/mm.

Nasal and temporal displacement

Individual subject data for nasal and temporal displacement in Group 1 were not well correlated with each other (P>0.05), and the correlation between measurements on two occasions did not reach statistical significance.

Gender effects

Group 1: For measurements taken on a single occasion, younger males were found to have significantly higher eyelid tension (in all directions) than younger females (ANOVA, P=0.03, Figure 5). Mean PA tension in females was 10.7±1.8 mN/mm and in males 14.8±5.1 mN/mm (P=0.03) (Table 1). Group 2: In the older subjects, gender did not appear to influence PA tension. Mean PA tension in females was 10.0±2.4 mN/mm and in males 12.2±4.0 mN/mm (P=0.15, Table 1).

Figure 5
figure 5

Graph showing mean and standard deviation for male and female tensions in Group 1 for PA, nasal and temporal displacement.

Discussion

This study presents for the first time normative, reproducibility, and gender difference data in LET. Both younger and older adult populations were studied. Previous studies in eyelid tensometry have measured the force required to move the upper eyelid, in contrast to the lower eyelid, in a PA direction, or the force generated by the levator muscle.7, 8, 9, 10 To our knowledge, lower eyelid tension has not been measured previously except by clinical4, 12, 13 means.

Mean lower eyelid tension for all subjects ranged between 7.8 and 14.8 mN/mm and nasal displacement tension was significantly higher than temporal displacement. Mean PA tension for all subjects ranged between 10.0 and 14.8 mN/mm, and the magnitude of this tension measurement is consistent with that reported for the upper eyelid by Vihlen and Wilson.2

PA displacement in the present study was slightly less than nasal tension but higher than temporal tension. The fact that the PA tension was higher than temporal tension may reflect the reflex orbicularis contraction when the eyelid is moved away from the globe during PA displacement. Theoretically, the orbicularis oculi will be contracting on both sides of the lash clamp in the PA movement, while on temporal displacement, only the side of the orbicularis opposite to the direction of tensometry (the nasal orbicularis) may be contracting. On the other hand, we speculate that the nasal tension may be higher than the temporal tension because the greater bulk of the orbicularis is found lateral to the midpupillary line,16 the site at which the eyelid was grasped by the tensometer lash clamp. However, this result is unexpected, as the normal movement of the lower eyelid during blinking is in the nasal direction, suggesting that the tension generated by the nasal fibres should be greater than that generated by those of the temporal side.16 The inter-eye correlation in PA measurement was moderate, as has been previously reported for upper eyelid measurements.2

We have demonstrated that there is a high interoccasion correlation in lower eyelid tension for PA displacement, but not for either nasal or temporal displacement. The poorer repeatability with lateral (nasal and temporal) displacements may reflect bending of the lashes when moving the eyelid perpendicular to the visual axis/lashes. The variances for repeated occasion measurements in PA tension were similar and, as expected, the main variability in the current measurement is attributed to the between-subject variability. Coefficient of repeatability data will assist in gauging meaningful differences between population groups using this technique. Our data would suggest that measurement of lower lid PA tension offers reasonable discriminatory ability.

Consistently higher lid tension for all displacements was reported in younger males compared with younger females. A similar gender difference in upper lid tension has been reported.17 The reason for these differences is unknown, but the higher tension in males may reflect the greater muscle bulk in males compared with females. There was also greater variability in PA measurements in males. Interestingly, this finding was not reflected in the older group, although the power to detect differences was low (0.244). While not statistically significant, we would speculate that the greater muscle bulk in males may reduce with age. Based on the mean standard deviation, and using the coefficient of repeatability as the estimate of minimum detectable difference between groups, power of 80% and α=0.05, a sample size of 89 subjects per group of young and older males would be required to confirm the age effect in males. Thus a total sample of 178 subjects (89 young and 89 old) would be required to establish that eyelid tension decreases with age in males.

It was curious and reassuring that once the subject had his/her bite on the bite bar established, along with the forehead and chin firmly resting on the forehead rest and chin rest respectively, there was almost never any demonstration of blepharospasm or withdrawal. We suspect that this type of absolute head fixation, largely engendered by the effect of the bite bar, has a very reassuring effect on the subject, and may thereby prevent unwanted voluntary or involuntary subject movement, and in particular eyelid movement. It is certainly different from the clinician's experience when using the customary slit lamp, where the patient frequently moves quite regularly. By inspecting the slope of lid tension on the tensometer computer screen, and glancing at the subject's position in space and in particular in relation to the forehead rest, it was possible to recognise the rare occasion, which suggested that the patient had withdrawn and produced a falsely high tensometry reading. The subjects uniformly responded willingly to instructions given both before and during their time on the lid tensometer.

The findings of this study may have relevance to surgeons involved in eyelid plastic surgery, particularly since the lower eyelid is more frequently operated upon than the upper eyelid in populations exposed to high levels of ultraviolet radiation. Examples of such conditions requiring lower eyelid surgery include involutional, tarsal or cicatricial ectropion, involutional entropion, lower eyelid neoplasms, some types of watery eyes,6 as well as eyelid abnormalities associated with thyroid orbitopathy. Ophthalmic surgeons have always operated on the lower eyelid based on a clinical assessment of the laxity of both medial and lateral canthal tendons. We have recently reported on a clinical classification of canthal tendon laxity.18, 19 Clinical assessment also involves an estimation of the eyelid's PA elasticity using the ‘snap back test’20 and an estimate of eyelid distraction.20 The tension of the lower eyelid must be adjusted accurately at surgery to avoid excessive horizontal eyelid tightening, which could cause lower eyelid ptosis. In the future, an awareness of normal and abnormal lower eyelid tension, as measured by an objective test, may allow the surgeon to plan eyelid surgery more effectively.

Other applications of eyelid tensometry could include the assessment of patients with upper eyelid disorders, such as those having floppy eyelid syndrome, superior limbic keratoconjunctivitis (SLK), and some patients with contact lens difficulties.21 Specifically, eyelid tension may influence contact lens centration, movement, and rotational stability. In these clinical settings, one would expect to see a variation in eyelid tension, the eyelid being looser in the floppy eyelid syndrome and involutional ectropion, and being increased in some patients with contact lens difficulties and SLK. In addition, the impact of eyelid tension on corneal topography has been raised as a significant issue, especially in refractive surgery. Eyelid retraction by the eyelid speculum during refractive surgery may alter corneal power by a significant amount compared with the presurgical measurement, and this may impact the refractive outcome.22

Eyelid tension in older normal adults was shown in this pilot study to demonstrate no significant reduction compared with younger normal adults, although the power to detect differences was low (0.299). This lack of effect might suggest that the presumed collagen failure of the lower eyelid with increasing age23 does not significantly contribute to its reduction. By contrast, it may be that the effect of UV radiation24 in stiffening the anterior, or weather-exposed, lamella of the lower lid accounts for this. However, testing a range of ages using a larger sample size would be appropriate to investigate further these effects. Our LET information will be used in further studies for comparison with LET measured in patients with both anatomical and functional nasolacrimal duct obstruction, since the multifactorial nature of the functional cases17, 25 may allow us insight into more directed and precise management of this cause of watery eyes.

In summary, this study has reported normative data for tension in the lower eyelid. It has demonstrated lower eyelid tension, and specifically PA displacement tension, can be accurately measured and is repeatable. While it has also shown that eyelid tension in young males is higher than in young females, gender differences in PA tension were not apparent in older subjects. In this pilot study, PA tension does not appear to reduce significantly with age.