Abstract
Variability in the x-ray tube current used in computed tomography may affect quantitative features extracted from the images. To investigate these effects, we scanned the Credence Cartridge Radiomics phantom 12 times, varying the tube current from 25 to 300 mA∙s while keeping the other acquisition parameters constant. For each of the scans, we extracted 48 radiomic features from the categories of intensity histogram (n = 10), gray-level run length matrix (n = 11), gray-level co-occurrence matrix (n = 22), and neighborhood gray tone difference matrix (n = 5). To gauge the size of the tube current effects, we scaled the features by the coefficient of variation of the corresponding features extracted from images of non-small cell lung cancer tumors. Variations in the tube current had more effect on features extracted from homogeneous materials (acrylic, sycamore wood) than from materials with more tissue-like textures (cork, rubber particles). Thirty-eight of the 48 features extracted from acrylic were affected by current reductions compared with only 2 of the 48 features extracted from rubber particles. These results indicate that variable x-ray tube current is unlikely to have a large effect on radiomic features extracted from computed tomography images of textured objects such as tumors.
Similar content being viewed by others
Introduction
The field of radiomics, in which quantitative image features are used to determine the tumor phenotype, has been demonstrated to have various potential roles in clinical decision-making. These include classifying tumors (e.g., benign or malignant), determining mutation status, improving patient risk stratification, predicting appropriate treatment strategies, and monitoring treatment response to improve outcome predictions1,2,3,4,5,6,7,8,9. However, radiomic features and results are sensitive to a variety of noise sources. For example, inter-scanner variations in image features can be relatively large9,10. Similarly, details of the imaging protocol, such as pixel size, can significantly affect the values of the calculated features11.
To maximize the amount of useful information obtained from computed tomography (CT) images in radiomics (and avoid incorrect interpretation of results), researchers must understand all sources of noise. This understanding can help with the development of solutions, such as image preprocessing, to mitigate the effects of the noise. In retrospective studies, understanding of the noise sources could be used to guide which image data are analyzed (e.g., only images with a pixel size within a specified range). In prospective studies, noise analysis could be used to guide the creation of harmonized imaging protocols in which the most important parameters are controlled. Examples of noise sources that have been examined in previous work include different CT scanners, pixel size, image spacing, and reconstruction kernels10,12,13,14,15.
The impact of tube current on diagnostic tasks has been covered extensively in the literature16,17,18,19,20,21, but only preliminary data are available regarding the effects of tube current in radiomics studies. Fave et al. simulated the effect of tube current on measured image features by adding Gaussian noise to patient CT images22. They observed no significant effect on image features (compared with inter-patient variations) but acknowledged that their noise model was very basic and did not properly reflect the changes in noise as the tube current was reduced. Larue et al. found that optimizing the number of gray-levels in images to improve prognostic value did not adversely affect feature stability. Further, they found that feature values were not correlated to tube currents or to slice thickness after resampling23. A study from Mahmood et al. raised questions about the robustness of image texture features. They imaged an anthropomorphic phantom on scanners from three manufacturers, and no features had CCC values greater than 0.924. A more reliable answer would involve scanning an object using different tube currents and then assessing the impact of this variation on the calculated image features.
Mackin et al. recently described a texture phantom that can be used to assess the impact of the imaging device or protocol on extracted image features10. In the current study, we experimentally examined the effects of tube current on quantitative image features by scanning this texture phantom using a range of tube current values.
Methods and Materials
The credence cartridge radiomics (CCR) phantom as described by Mackin et al.10 was used to study the effects of tube current on radiomic features. The CCR phantom has ten cartridges with various textures. In the current study, we analyzed four of the cartridges: solid acrylic, cork, rubber particles, and wood. These were selected to give a full range of textures from minimal to highly varied, similar to the texture of non-small cell lung cancer tumors (Fig. 1).
The phantom was imaged on a GE LightSpeed VCT scanner (GE Healthcare, Waukesha, WI) and a Toshiba Aquilion ONE scanner (Toshiba Medical Systems, Tustin, CA) using a range of tube current settings. The GE LightSpeed images were acquired in helical mode at 120 kVp, 0.969 pitch, STANDARD reconstruction kernel, 50-cm display field of view, and 2.5-mm image thickness. Each voxel was 0.98 × 0.98 × 2.5 mm3. Twelve scans were acquired at 25, 50, 75, 100, 125, 150, 175, 200, 225, 250, 275, and 300 mA∙s. The Toshiba Aquilion ONE scans were acquired in helical mode at 120 kVp, 1.4 pitch, FC18 reconstruction kernel, 50-cm display field of view, and 5-mm image thickness. Each voxel was 0.98 × 0.98 × 5.0 mm3. Ten scans were acquired at 25, 50, 75, 100, 125, 150, 175, 200, 225, and 250 mA∙s.
The images were imported into IBEX, a freely available radiomics software program25. The location of a small radiopaque marker on the edge of the phantom was manually identified, and its image coordinates were entered into an in-house Python script (Python version 2.7), which created a rectangular region of interest (ROI) of 8 × 8 × 2 cm3 for each cartridge in each scan. The same ROI file was used for all images from a particular scanner. Results for smaller (2 × 2 × 2 cm3) voxels are included in the supplementary materials. Forty-eight features were calculated in IBEX: 10 intensity histogram, 22 gray-level co-occurrence matrix26, 11 gray-level run length matrix27,28, and 5 neighborhood gray tone difference matrix features29 (Table 1). These features were selected because they are commonly used in radiomics studies30,31. As noted below, features were calculated with one and four CT numbers per bin. The texture features were calculated for each slice in the ROIs and then combined, a procedure referred to as 2.5D25.
Reducing the tube current used in CT scans will increase the noise in the images. In this study our primary concern is not measure the size of tube current effect in the phantom materials. Our primary concern is to gauge the size of the effect tube current may have on radiomics studies of patients. Therefore, we used a metric that scales the effect seen in phantom materials by the variability in patients. If a feature is highly variable in patients, a small tube current effect is unlikely to weaken a radiomic feature. On the other hand, a large tube current effect is likely to weaken a radioimic feature when the patient variability is small. To gauge the size of the effects of variable tube current relative to the variability in patients, we normalized the extracted feature values by the coefficient of variation for the same features extracted from CT images of non-small cell lung cancer tumors. This patient-normalized feature, \({\hat{f}}_{i}\), was defined as
where f i is the feature value for a given tube current value i (mA∙s) and f0 is the feature value at the baseline tube current value (300 mA∙s for the GE scanner and 250 mA∙s for the Toshiba scanner). σT,i and µT,i are the standard deviation and mean, respectively, of the features from the non-small cell lung cancer tumors. The numerator of \(\hat{{f}_{i}}\) is the fractional difference for a tube current scan and the baseline, and the denominator is the coefficient of variation for the same feature calculated on 106 non-small cell lung cancer tumors. This normalization assesses the variability caused by tube current relative to inter-patient differences.
The patients in this normalization cohort were part of a clinical trial approved by the Institutional Review Board. Informed consent for participation in the trial was obtained for all patients, and all procedures were performed in accordance with the Declaration of Helsinki on Ethical Issues. Additional informed consent for this retrospective study was waved by the Institutional Review Board. The gross tumor volumes from the end-of-exhale phase of the planning CT images were used as the ROIs for feature extraction. The features were extracted after applying a threshold of 900 HU. The end-of-exhale phase is considered the most stable22,32,33. The mean and median tumor ROI volumes were 96 and 42 cm3 respectively (range 5–568 cm3). ROIs smaller than 5 cm3 were excluded from the normalization cohort. This patient cohort was part of a prior radiomics study which details its clinical characteristics9.
Features from the phantom and patients were extracted in four ways: (1) no preprocessing, (2) intensity rescaling (10-bit depth rescaling), (3) Butterworth smoothing, and (4) Butterworth smoothing and intensity rescaling. These preprocessing techniques were chosen on the basis of work by Fave et al., who showed that preprocessing affects the significance of features in prognostic models34. Rescaling the intensity from the initial 4096 bins (12-bit) to 1024 bins (10-bit) combines 4 CT numbers per bin.
Data availability
The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.
Results
The effects of tube current on image intensity histograms for the acrylic, cork, rubber particle, and sycamore wood cartridges from images acquired on the GE scanner are shown in Fig. 2 (results for images acquired on the Toshiba scanner are shown in Supplementary Fig. 1). Results did not vary substantially between the two scanners used. Therefore, all figures show results obtained using the GE LightSpeed VCT scanner, and results obtained using the Toshiba Aquilion ONE scanner are shown in corresponding supplementary figures. The acrylic cartridge image from the 25 mA∙s scan had a much greater dispersion of intensities than did the image from the 300 mA∙s scan. The differences between the intensity histograms for these two scans were not as apparent in the images of the other three materials. Preprocessing the images by rescaling, smoothing, or both rescaling and smoothing had only a marginal effect on the dispersion differences.
The effects of tube current on the intensity features, shown in Fig. 3 for the GE scanner (results for the Toshiba scanner are shown in Supplementary Fig. 2), are consistent with the effects seen in the intensity histograms. The values used in this figure were scaled according to Equation 1 to produce patient-normalized feature differences. Reducing the tube current values produced the largest changes in the acrylic cartridge features. As with the intensity histograms, rescaling the intensities, smoothing the images, or doing both had little effect on the patient-normalized feature values.
Grey-level co-occurrence matrices for the acrylic and rubber particles cartridges for images acquired on the GE scanner are shown as images in Fig. 4 (results for images acquired on the Toshiba scanner are shown in Supplementary Fig. 3). Gray-level co-occurrence matrices record the frequency of image intensity values being adjacent each other, or “co-occurring”, in the image. The difference in the dispersion of the CT numbers for the 25 and 300 mA∙s scans produced obvious changes in the gray-level co-occurrence matrix for the acrylic cartridge. This dependence is also evident in the features derived from the gray-level co-occurrence matrices, shown in Fig. 5 (results for images acquired on the Toshiba scanner are shown in Supplementary Fig. 4). The tube current dependence was greatest for the acrylic cartridge. A more moderate effect was seen in the sycamore wood cartridge and little effect was seen in the heterogeneous cork and rubber particles cartridges.
A high-level look at the 48 features from the four feature groups is shown in Fig. 6, for images acquired on the GE scanner (results for images acquired on the Toshiba scanner are shown in Supplementary Fig. 5). In this figure, the patient-normalized feature values are color-coded into one of four categories indicating the degree to which the feature values depend on the tube current value used when the images were acquired. The features extracted from the most homogeneous materials, acrylic and sycamore wood, were much more dependent on tube current than were the features extracted from the more textured materials, cork and rubber particles. More specifically, 25 of 48 features extracted from the acrylic cartridge were strongly dependent on tube current (\({\hat{f}}_{i} > 2\)). In contrast, \({\hat{f}}_{i} > 0.5\) was observed for only two of the features extracted from the rubber particles cartridge.
Discussion
We investigated the effect of reducing tube current on materials with varying amounts of texture and found that objects with more intrinsic texture were not substantially affected by tube current changes. Reducing the tube current of a CT scan increases the image noise, and therefore increases the spread of CT numbers. We found that the impact of this noise was more apparent for homogeneous materials (e.g., acrylic) than for textured materials. The increased spread observed in the intensity histogram for the acrylic cartridge at 25 mA∙s compared with 300 mA∙s implies that image intensity features for this cartridge are expected to be dependent on tube current. For the most textured cartridges, rubber particles and cork, dispersion at the lower tube current values was minimal, and thus, features extracted from these types of materials are not expected to be dependent on tube current.
Indeed, the expected feature dependence was observed: feature dependence in acrylic varied whereas this dependence in rubber particles and cork – which have texture more similar to NSCLC tumors – stayed constant. This is shown through the plots of the patient-normalized feature differences (Figs 3 and 5), in which values for cork and rubber particles stayed near 0 for all points whereas values for acrylic approached 2 or -2 or even extended beyond the displayed range for many features. These findings suggest that when a material has minimal texture, such as acrylic, the texture feature values are heavily dependent on noise, i.e., tube current values. Materials that did have texture showed little effect from tube current variation. Additionally, image preprocessing (intensity rescaling or smoothing) did not substantially change this relationship. Overall, these results are consistent with the simpler model that was examined by Fave et al.22. The results presented here indicate that retrospective radiomics studies should not be significantly affected by variations in tube current. In other words, there should be no need to exclude patient data from retrospective studies on the basis of differences in tube current alone.
These results also indicate that harmonizing tube current values between scans need not be a big concern when planning prospective radiomics studies. This is important because most scanners now modulate mA∙s to control the overall noise level, and these settings may be determined locally (e.g., on the basis of radiologist feedback). Thus, it may be difficult to harmonize the tube current between institutions. Instead, harmonization can focus on details of the image reconstruction such as pixel size and reconstruction kernel, which can both be achieved in a second reconstruction which is specifically designed for radiomics studies and which does not increase the radiation dose to the patient.
The current study has a few limitations. Phantom materials are not perfect surrogates for tissue, and some quantitative features extracted from some human tissues might be more sensitive to changes in the tube current values. It is possible that some regions in tumors might be more sensitive to the tube current than other regions. In addition, the acquired scans had an image thicknesses of 2.5 or 5.0 mm and all ROIs had a volume of 128 cm3. These factors may reduce the impact of tube current owing to the reduced effect of noise. We repeated our study using smaller, 8 cm3, voxels, and found that the results for the smaller ROIs were similar for both the GE and Toshiba scanners (supplementary Figs 6 and 7). Also, the results for 2.5 and 5.0 image thicknesses were similar to each other. Additionally, the current study evaluated only the effects of tube current on measured radiomic features within a defined volume. The effects of tube current on physician delineation were not evaluated. With more noise at lower tube current values, it may be more difficult to determine ROI bounds within patients. A change in the contouring would affect the radiomic features measured35. Finally, we did not evaluate many of the other parameters of the imaging protocol, such as tube voltage (kV). These should be investigated in future studies.
Conclusion
To provide noninvasive and relatively inexpensive biomarkers, radiomic studies may rely on images acquired using diagnostic imaging or radiation therapy simulation protocols. Retrospective reconstruction of medical images where the imaging procedure is performed one time but the images are reconstructed multiple times using a radiomics protocol in addition to the standard protocol may help to standardize the images. For a CT scan, however, some parameters, including the pitch, tube voltage, and tube current, cannot be changed for retrospective reconstruction. Most CT scans of adults use 120 kVp. It seems unlikely that the pitch of helical scans will affect quantitative imaging features. Thus, the tube current, which influences both the overall noise in the image and the radiation dose to the patient, is most concerning of these three parameters. Our finding that radiomic features are robust to changes in tube current in CT studies of tumor-like materials indicates that variations in the tube current used while imaging patients is unlikely to weaken the study. It is unlikely that radiomic image features calculated from CT images of textured objects (such as tumors) are significantly affected by x-ray tube current.
References
Fried, D. V. et al. Stage III Non-Small Cell Lung Cancer: Prognostic Value of FDG PET Quantitative Imaging Features Combined with Clinical Prognostic Factors. Radiology 278, 214–222, https://doi.org/10.1148/radiol.2015142920 (2016).
Fried, D. V. et al. Potential Use of (18)F-fluorodeoxyglucose Positron Emission Tomography-Based Quantitative Imaging Features for Guiding Dose Escalation in Stage III Non-Small Cell Lung Cancer. International journal of radiation oncology, biology, physics 94, 368–376, https://doi.org/10.1016/j.ijrobp.2015.10.029 (2016).
van Rossum, P. S. et al. The incremental value of subjective and quantitative assessment of 18F-FDG PET for the prediction of pathologic complete response to preoperative chemoradiotherapy in esophageal cancer. Journal of Nuclear Medicine 57, 691–700 (2016).
Scrivener, M. et al. Radiomics applied to lung cancer: a review. Translational Cancer Research 5, 398–409 (2016).
Aerts, H. J. et al. Defining a Radiomic Response Phenotype: A Pilot Study using targeted therapy in NSCLC. Sci Rep 6, 33860, https://doi.org/10.1038/srep33860 (2016).
Aerts, H. J. W. L. et al. Decoding tumour phenotype by noninvasive imaging using a quantitative radiomics approach. Nat Commun 5, 4006 (2014).
Coroller, T. P. et al. CT-based radiomic signature predicts distant metastasis in lung adenocarcinoma. Radiotherapy and Oncology 114, 345–350, https://doi.org/10.1016/j.radonc.2015.02.015 (2015).
Mattonen, S. A., Palma, D. A., Haasbeek, C. J., Senan, S. & Ward, A. D. Early prediction of tumor recurrence based on CT texture changes after stereotactic ablative radiotherapy (SABR) for lung cancer. Medical physics 41, 033502 (2014).
Fave, X. et al. Delta-radiomics features for the prediction of patient outcomes in non–small cell lung cancer. Scientific Reports 7, 588 (2017).
Mackin, D. et al. Measuring Computed Tomography Scanner Variability of Radiomics Features. Investigative radiology 50, 757–765, https://doi.org/10.1097/RLI.0000000000000180 (2015).
Mackin, D. et al. Harmonizing the pixel size in retrospective computed tomography radiomics studies. PLoS ONE (2017).
Zhao, B. et al. Reproducibility of radiomics for deciphering tumor phenotype with imaging. Scientific reports 6, 23428, https://doi.org/10.1038/srep23428 (2016).
Zhao, B., Tan, Y., Tsai, W. Y., Schwartz, L. H. & Lu, L. Exploring Variability in CT Characterization of Tumors: A Preliminary Phantom Study. Translational oncology 7, 88–93 (2014).
Shafiq‐ul‐Hassan, M. et al. Intrinsic dependencies of CT radiomic features on voxel size and number of gray levels. Medical physics 44, 1050–1062 (2017).
Basu, S. et al. In Systems, Man, and Cybernetics (SMC), 2011 IEEE International Conference on. 1306–1312 (IEEE).
Alikhani, B., Jamali, L., Raatschen, H.-J., Wacker, F. & Werncke, T. Impact of CT parameters on the physical quantities related to image quality for two MDCT scanners using the ACR accreditation phantom: A phantom study. Radiography (2017).
Peng, W. et al. A CONSORT-compliant prospective randomized controlled trial: radiation dose reducing in computed tomography using an additional lateral scout view combined with automatic tube current modulation: Phantom and patient study. Medicine 96, e7324, https://doi.org/10.1097/MD.0000000000007324 (2017).
Lee, S. et al. Comparison of image quality and radiation dose between combined automatic tube current modulation and fixed tube current technique in CT of abdomen and pelvis. Acta radiologica 52, 1101–1106, https://doi.org/10.1258/ar.2011.100295 (2011).
Solomon, J. B., Li, X. & Samei, E. Relating noise to image quality indicators in CT examinations with tube current modulation. AJR. American journal of roentgenology 200, 592–600, https://doi.org/10.2214/AJR.12.8580 (2013).
Tang, K. et al. Effect of low tube voltage on image quality, radiation dose, and low-contrast detectability at abdominal multidetector CT: phantom study. Journal of biomedicine & biotechnology 2012, 130169, https://doi.org/10.1155/2012/130169 (2012).
Ledenius, K. et al. Effect of tube current on diagnostic image quality in paediatric cerebral multidetector CT images. The British journal of radiology 82, 313–320, https://doi.org/10.1259/bjr/24404354 (2009).
Fave, X. et al. Preliminary investigation into sources of uncertainty in quantitative imaging features. Computerized medical imaging and graphics: the official journal of the Computerized Medical Imaging Society 44, 54–61, https://doi.org/10.1016/j.compmedimag.2015.04.006 (2015).
Larue, R. T., Defraene, G., De Ruysscher, D., Lambin, P. & Van Elmpt, W. Quantitative radiomics studies for tissue characterization: a review of technology and methodological procedures. The British journal of radiology 90, 20160665 (2017).
Mahmood, U., Apte, A. P., Deasy, J. O., Schmidtlein, C. R. & Shukla-Dave, A. Investigating the Robustness Neighborhood Gray Tone Difference Matrix and Gray Level Co-occurrence Matrix Radiomic Features on Clinical Computed Tomography Systems Using Anthropomorphic Phantoms: Evidence From a Multivendor Study. Journal of computer assisted tomography 41, 995–1001 (2017).
Zhang, L. et al. IBEX: an open infrastructure software platform to facilitate collaborative work in radiomics. Medical physics 42, 1341–1353, https://doi.org/10.1118/1.4908210 (2015).
Haralick, R. M. & Shanmugam, K. Textural features for image classification. IEEE Transactions on systems, man, and cybernetics, 610–621 (1973).
Galloway, M. M. Texture analysis using gray level run lengths. Computer Graphics and Image Processing 4, 172–179, https://doi.org/10.1016/S0146-664X(75)80008-6 (1975).
Tang, X. Texture information in run-length matrices. IEEE transactions on image processing 7, 1602–1609 (1998).
Amadasun, M. & King, R. Textural features corresponding to textural properties. IEEE Transactions on systems, man, and Cybernetics 19, 1264–1274 (1989).
Zwanenburg, A., Leger, S., Vallières, M. & Löck, S. Image biomarker standardisation initiative-feature definitions. arXiv preprint arXiv:1612.07003 (2016).
Kalpathy-Cramer, J. et al. Radiomics of lung nodules: A multi-institutional study of robustness and agreement of quantitative imaging features. Tomography: a journal for imaging research 2, 430 (2016).
Fried, D. V. et al. Stage III non–small cell lung cancer: prognostic value of FDG PET quantitative imaging features combined with clinical prognostic factors. Radiology 278, 214–222 (2015).
Seppenwoolde, Y. et al. Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. International Journal of Radiation Oncology* Biology* Physics 53, 822–834 (2002).
Fave, X. et al. Impact of image preprocessing on the volume dependence and prognostic potential of radiomics features in non-small cell lung cancer. Translational Cancer Research 5, 349–363 (2016).
Parmar, C. et al. Robust Radiomics feature quantification using semiautomatic volumetric segmentation. PLoS ONE 9, e102107, https://doi.org/10.1371/journal.pone.0102107 (2014).
Acknowledgements
The authors thank Kelly Tharp for designing and constructing the phantom. We would also like to thank Erica Goodoff of MD Anderson’s Department of Scientific Publications for her careful editing of this manuscript. Rachel Ger is supported by the Rosalie B. Hite Graduate Fellowship in Cancer Research and the American Legion Auxiliary Fellowship in Cancer Research awarded by the UTHealth Graduate School of Biomedical Sciences. This work was supported by the National Cancer Institute of the National Institutes of Health under award number R03CA178495 and award number R21CA216572. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.
Author information
Authors and Affiliations
Contributions
Project conception and design was by D.M., C.D.1., A.K.J. and L.C. The data collection was performed by D.M., P.C., C.D.1, C.D.2, R.G., S.B. The software programming and analysis was performed by D.M., L.Z., J.Y., X.F., R.G. The manuscript was written by D.M., R.G., and L.C., and it was reviewed by C.D.1, X.F., and A.K.J.
Corresponding author
Ethics declarations
Competing Interests
The authors declare that they have no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Mackin, D., Ger, R., Dodge, C. et al. Effect of tube current on computed tomography radiomic features. Sci Rep 8, 2354 (2018). https://doi.org/10.1038/s41598-018-20713-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-018-20713-6
This article is cited by
-
Radiomics reproducibility challenge in computed tomography imaging as a nuisance to clinical generalization: a mini-review
Egyptian Journal of Radiology and Nuclear Medicine (2023)
-
The impact of inter-observer variation in delineation on robustness of radiomics features in non-small cell lung cancer
Scientific Reports (2022)
-
A radiomics approach for lung nodule detection in thoracic CT images based on the dynamic patterns of morphological variation
European Radiology (2022)
-
Robustness of CT radiomics features: consistency within and between single-energy CT and dual-energy CT
European Radiology (2022)
-
Preliminary report on harmonization of features extraction process using the ComBat tool in the multi-center “Blue Sky Radiomics” study on stage III unresectable NSCLC
Insights into Imaging (2022)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.