To the Editor:
Intravitreal aflibercept has been used in the UK for treatment of a variety of retinal vascular disorders and from April 2020 has had market authorisation for use in a pre-filled syringe (PFS).
We report a series of five eyes in four patients who developed transient CRAO immediately post injection using the PFS (see Table 1). These occurred shortly after the adoption of the PFS in our departments. All patients had previously received intravitreal injections (ranibizumab or aflibercept) without complication.
Following these events, we investigated the volume of fluid expressed when using a standard 1 ml syringe compared to the new aflibercept PFS.
Six injectors drew water into an empty aflibercept PFS and a standard 1 ml syringe, primed them to 0.05 ml and then expressed the contents into a receptacle on a weighing scale (Hooiswee, Shenzen, China). The mass of water was recorded (0.05 ml having a mass of 0.05 g. Results reported as volumes in ml). This was repeated 20 times for each syringe (Total of 120 repeats for each syringe). Data were analysed with MS Excel with StatPlus.
Epoxy resin casts were taken of each syringe and Vernier callipers used to measure the internal diameter.
The mean volume of water expelled per repeat was 0.063 ml (SD 0.009 ml) for the PFS and 0.054 ml (SD 0.002 ml) for the 1 ml syringe (difference 0.009 ml, p = 8.3 × 10−20, 95% CI 0.007–0.011) (see Fig. 1). The maximum volume expressed was 0.064 ml for the 1 ml syringe and 0.084 ml for the PFS.
The PFS expressed a greater range of volumes than the 1 ml syringe with 21% of PFS repeats expressing 0.07 ml or more.
Changes in volume with errors in syringe plunger position
The measured internal diameter was 6.4 mm for the PFS and 4.6 mm for the 1 ml syringe. (For comparison, the internal diameter of the ranibizumab PFS was also measured as 4.6 mm). Therefore, the calculated internal area (πr2) was 32 mm2 for the aflibercept PFS and 16.6 mm2 for the 1 ml syringe.
Any error in alignment with the 0.05 ml mark will result in an error in the volume expressed. The size of this error is determined by the linear error in plunger alignment multiplied by the internal area of the syringe.
Thus, any unit error in plunger alignment will result in approximately a twofold greater error in volume for the PFS as compared to the 1 ml syringe.
Pallikaris et al. [1] demonstrated a linear relationship between IOP rise and volume injected. For the majority of patients, the observed errors in volume are unlikely to have any clinically significant impact. The risk of transient CRAO, even with higher injected volumes is still low [2]. There will be a subset of patients however for whom these higher volumes may be more significant (e.g. advanced glaucoma, low ocular perfusion pressure). Given that the wider diameter syringe results in a doubling of the volume error with each unit error in plunger alignment, it is particularly important with the aflibercept PFS to pay close attention to the plunger position.
References
Pallikaris I, Kymionis G, Ginis H, Kounis G, Tsilimbaris M. Ocular rigidity in living human eyes. Investig Ophthalmol Vis Sci. 2005;46:409–14.
VEGF Inhibition Study in Ocular Neovascularization (V.I.S.I.O.N.) Clinical Trial Group, D’Amico DJ, Masonson HN, Patel M, Adamis AP, Cunningham ET Jr, Guyer DR, et al. Pegaptanib sodium for neovascular age-related macular degeneration: two-year safety results of the two prospective, multicenter, controlled clinical trials. Ophthalmology. 2006;113:992–1001.e6.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Gallagher, K., Raghuram, A.R., Williams, G.S. et al. Pre-filled aflibercept syringes—variability in expressed fluid volumes and a case series of transient central retinal artery occlusions. Eye 35, 2899–2900 (2021). https://doi.org/10.1038/s41433-020-01211-4
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/s41433-020-01211-4
This article is cited by
-
Accuracy and residual volume of intravitreal injection syringes in administration of very small volumes
Eye (2024)
-
Accuracy of intravitreal injection volume for aflibercept pre-filled syringe and BD Luer-Lok one-milliliter syringe
International Journal of Retina and Vitreous (2022)
-
Incidence of severe rise in intraocular pressure after intravitreous injection of aflibercept with prefilled syringes
Scientific Reports (2022)
-
Prefilled Eylea Syringe: our recent experience
Eye (2021)
-
Aflibercept
Reactions Weekly (2021)