Introduction

Telemedicine has already become a common tool in image-based medical specialty such as ophthalmology.1 This new method of service delivery offers a number of substantial advantages over conventional face-to-face consultations. This includes cost effectiveness,2, 3 improved service accessibility in remote areas,4, 5 and local health-care workers gaining more experience in ophthalmology through participation in telemedical consultations.1, 4, 6

There are several major components in a teleophthalmology system. These consist of image capture, digitisation, transmission, storage, interpretation, and finally a reply. Image capture is the beginning and perhaps the most vital part of the system. Poor quality images may lead to poor diagnosis. Image acquisition is particularly a problem in ophthalmology as the eye can only be examined thoroughly with expensive specialised optical instruments. For teleophthalmology, these instruments need to be modified for high-resolution digital capture, which further adds to the cost and complexity. On the other hand, these instruments need to be more affordable, portable, and less complex to widen the service access.1

The standard slit-lamp biomicroscope is the benchmark device for anterior segment ocular examination. However, the complexity of the instrument together with the cost prohibits widespread use of it in remote communities. A portable and simplified instrument would be more suitable for the health-care workers in distant locations. A portable slit-lamp video camera has been developed at the Lions Eye Institute (LEI) in Perth, Western Australia (Figure 1). This hand-held instrument has a fixed width slit beam with two selectable alignments. A third ‘red reflex’ coaxial illumination is also selectable. Only four switches and one trigger are needed to control this digital slit-lamp camera. The switches adjust the intensity of the background illumination intensity, background illumination colour (blue or white), slit beam intensity, and the direction of the slit beam. The trigger activates image capture. There is a liquid crystal display panel on the unit to aid focusing. A frame grabber attached to a personal computer digitises the video images. The image resolution of the digital image is 640 × 480 pixels. A health-care worker with little background computer knowledge can be trained to use this camera to capture the standard images for this study in few hours.

Figure 1
figure 1

Image of the hand held, portable digital slit-lamp (PSL) video camera developed at the Lions Eye Institute (LEI) in Perth, Western Australia. This hand held instrument has a fixed width slit beam with two selectable alignments. A third ‘red reflex’ coaxial illumination is also selectable.

Most research in the image capture component of teleophthalmology to date has been aimed at retinal pathologies. Although most of the published works have been descriptive in nature, there have been some comparative trials that compared images from video fundus cameras to 35 mm film in detection of diabetic retinopathy.7, 8, 9, 10 The results so far have been encouraging. Relatively little work has been done on anterior segment telemedical examinations. A feasibility study on ocular surface assessment using telemedical equipment was performed by Shimmura et al.11 Threlkeld et al12 examined the accuracy of telemedical anterior segment ocular examination. The use of a video slit-lamp in examination of postoperative cataract and glaucoma patients was investigated by Murdoch et al.13 Schiffman et al14 has described several years of experience using a teleophthalmology system with several different peripherals including video slit lamps. All these investigators used a combination of a standard slit-lamp biomicroscope and a CCD video camera attachment. Also, in most of these studies the video slit lamps were operating in a real-time mode.

There are two general system architectures in telemedicine. In a real-time or teleconferencing consultation, the local operator, the patient, and the remote specialist are all present and communicating. Video-conferencing equipment is one of the most common forms of real-time telemedicine. While in a store and forward system, digital images, video, audio, and clinical data are captured and ‘stored’ on the client computer; then at a convenient time, transmitted securely (‘forwarded’) to a clinic at another location where they are studied by relevant specialists. The opinion of the specialist is then transmitted back.

To the best of our knowledge, there has been no comparison study of a store and forward still image system against traditional method of service delivery in anterior segment ophthalmology. There has also been no formal comparison of digital anterior segment ophthalmic images against 35 mm slides. Using traditional examination and photographic methods as a comparison, this study aimed to investigate the accuracy of this new dedicated telemedicine camera in identifying and grading anterior segment eye diseases. Although the images were not transmitted electronically using telecommunication technology between computers, the design of this study simulated a store and forward telemedicine environment.

Methods

Patients attending the clinic of an anterior segment ophthalmologist at the Lions Eye Institute were recruited into this trial. The Patients were not selected on the basis of presence or absence of pathology. Because of the technical difficulties with the equipment, only 343 eyes had digital images taken and only 338 eyes had 35 mm slides taken.

Each patient was examined by an experienced anterior segment consultant ophthalmologist using a standard slit-lamp biomicroscope (Haag Streit BQ900, Bern, Switzerland). The presence or absence of 33 specific findings were noted on a standardised scoring sheet (Table 1). Lens opacities were graded as 0, 1+, 2+, 3+, and 4+. A health worker with little experience in ophthalmology imaged the subject's eyes with the Zeiss (Carl Zeiss Jena GmbH) 40SL/P slit-lamp camera (ZSL) and the LEI portable slit-lamp camera (PSL) before or after the examination. Two standardised views were taken with the ZSL. A third retroillumination view was taken, if the patient had pupil dilation. Figure 2 summarises the instrument setup for each of these views. 30–40 s of video image were then captured from each eye with the PSL. Two frames (three in dilated patients) with the same standardised attributes were extracted from these video clips. The digital images are then saved and transferred to other computers through a CD-ROM.

Table 1 List of anterior segment abnormalities detected in this study
Figure 2
figure 2

Diagram of portable digital slit-lamp (PSL) instrument setup. The diagram illustrates the various positions of the camera and the direction of the illumination beam as seen from above the patient for three different left eye standard views.

The same ophthalmologist assessed the still images captured from the two cameras in a masked fashion. The digital images were assessed several weeks after the initial clinic visit. One month after the digital image assessment sessions, 35 mm slides were presented. The assessor noted the findings using the same scoring sheet as that used in the clinical assessment stage. To give a measure of inter-reader agreement, two other ophthalmologists graded the digital images from the first 100 patients.

Steps were taken to match the image attributes between the two image formats. First, the relative magnification of the images were matched. On average, the diameter of the cornea was around 54% of the diagonal dimension of either type of image. This was achieved by using the 10 times magnification setting on the ZSL. The focal length of the PSL was adjusted to match this magnification. Second, in both image formats, the slit beam was fixed at 1.5 mm in width and angled 30° temporally from the camera axis. Third, to balance the stability disadvantage of the PSL, digital still images were extracted from video clips rather than capturing still images directly at the time of examination. Images with the correct focus and slit location could be carefully selected from a video clip. As each frame could be played slowly one by one, the effect of camera shake could be minimised. Fourth, the images were presented in similar visual angular sizes. Digital images were viewed on a 17-inch monitor in unchanged resolution adjusted to give a diagonal length of 15 cm and viewed at 40 cm from the screen. 35 mm slides were projected to a diagonal length of 37.5 cm and viewed at 100 cm from the screen.

Images from the PSL were compared with clinical and photographic gold standards. Traditional face-to-face ophthalmic examination by an ophthalmologist was designated as the clinical gold standard. 35 mm transparencies taken with the ZSL were designated as the photographic gold standard. Although 35 mm slit-lamp slides are not frequently used for diagnostic purposes, this imaging format is commonly used for record keeping and progress monitoring of anterior segment diseases. The comparison between the digital images and the 35 mm slides was mainly for image quality validation.

Unweighted κ statistics was used as a measure of agreement between each diagnostic modality. Sensitivity and specificity values were also calculated. The interpretation of unweighted κ statistics are as follows: 0–0.20, slight agreement; 0.21–0.40, fair agreement; 0.41–0.60, moderate agreement; 0.61–0.80, good agreement; and 0.81–1, excellent agreement. In general, most of the signs were not present in large numbers in this group of patients. The most common signs were blepharitis and cataracts. As κ statistics are grossly influenced by the prevalence of the disease,15 these measures were not calculated in categories where there were less than 25 true positives.

Results

A total of 196 patients (392 eyes) participated in the trial. The average age of the patients was 62.5 years old (range: 15–94) and 54% of them were female. Thirty-nine patients had pupil dilation.

Table 2 summarizes the comparison between imaging modalities and the clinical findings. Digital images showed moderate to excellent agreement to clinical findings (κ 0.45–0.82) in all areas other than lid pathologies. In particular, lens signs other than posterior subcapsular opacities were well demonstrated with digital images (unweighted κ 0.54–0.75, sensitivity 76–92%, and specificity 92–99%). Gross corneal abnormalities were identified well as shown by the high sensitivity and κ values in corneal grafts (κ 0.82, sensitivity 88%, and specificity 98%), contact lens (sensitivity 67% and specificity 100%), corneal vascularisations (sensitivity 69% and specificity 99%), and corneal edemas (sensitivity 67% and specificity 99%). Small and subtle corneal signs were less well detected (for example, sensitivity of 0% for keratic precipitate and epitheliopathy). Conjunctival signs correlated moderately well with the clinical findings (κ 0.45–0.60, sensitivity 38–57%, and specificity 98–100%). Iris abnormalities were moderately well identified (sensitivity 60%). Lid pathologies were not well detected (κ 0.23 and sensitivity 0–50%). Cells and flare in the anterior chamber were not identified in the digital images (sensitivity 0%).

Table 2 35 mm film and digital images findings vs clinical assessment

The results from the 35 mm transparencies were in general similar to those from the digital images. There was a wider variation in the concordance to clinical findings (κ 0.14–0.85). The trends in the statistics were similar to those found in the comparison between digital image results and clinical impressions. All the lens findings except posterior subcapsular cataracts correlated very well with the clinical gold standard (sensitivity 59–77%, specificity 86–99%, and unweighted κ 0.43–0.65). Gross corneal abnormalities were also detected well. Pterygia (κ 0.720.60 and sensitivity 7054%) were better identified with 35 mm slide films. Digital images were significantly better at detection of cortical cataracts (unweighted κ 0.62 0.43 and sensitivity 92%59). Lid and anterior chamber signs were once again poorly detected (sensitivity 0–31%).

When the findings from the digital images were compared with those from 35 mm films, the κ, sensitivity, and specificity figures generally show the same profiles as those mentioned above. Table 3 shows the comparison between results seen in digital images and those seen in 35 mm transparencies. The two image formats generally showed more agreement to each other than to the clinical findings. κ values are 0.34–0.90 when the two image formats are compared to each other; whereas it is only 0.23–0.82 for digital images, and 0.14–0.85 for 35 mm images, when compared to clinical findings.

Table 3 Digital images findings vs 35 mm film findings

The assessments of the digital images from two other ophthalmologists generally showed a good agreement with the findings from the first ophthalmologist. These statistical results are shown in Table 4. Owing to the fact that only the first 100 patients were seen, the prevalence of some of the signs was very low. The statistical figures were good for the categories that showed a good agreement between the digital findings from the first ophthalmologist and clinical gold standard. For example, the agreement in lens signs other than posterior subcapsular cataracts was good to excellent (κ 0.62–0.82). There was also a good agreement in gross corneal signs like corneal grafts (sensitivity 83–60%, specificity 99–99%, and κ 0.87–0.70) and pterygia (sensitivity 53–67%, specificity 99–98%, and κ 0.62–0.67). However, in the categories that showed inadequate digital detection rates such as blepharitis, the agreement was poor (κ 0.12–0.30).

Table 4 Digital images findings: second ophthalmologist and third ophthalmologist vs first ophthalmologist

Discussion

The results from this study were limited by low prevalence in many categories of anterior segment abnormalities in the subjects. Statistical measures of κ, sensitivity and specificity have diminished meaning when there are less than five subjects with the condition. Despite the low prevalence of some disease signs, it was clear that certain signs were much better detected than others. It was also evident that both imaging formats produce comparable diagnoses.

The similarity in the results from both image formats illustrated that portable digital slit-lamp images had similar diagnostic quality to that of 35 mm slides from a traditional slit-lamp camera when used in the same setup as used in this study. There were better statistical agreements when the findings from the two types of images were compared with each other, than when either is compared to the clinical findings. The significance of this outcome can be better appreciated when the digital resolutions of the two image formats are listed. The digital images used in this trial had 640 × 480 pixels or 0.3 megapixels. On the other hand, 35 mm slide typically has a 10–20 megapixel resolution when digitised. Hence, there is a 50-fold difference in resolution between the two formats. It is clear that poor detection of some subtle signs seen in this study was not due to a lack of image resolution.

It has to be noted that the PSL was designed to be a simple instrument with few configuration options. The ZSL is usually used with a variety of illumination, camera angle, and magnification options. For the sake of uniformity, these capacities were not used in this study.

Although both formats showed moderate to excellent agreement with clinical findings in lens and gross corneal signs, more subtle diseases were not detected well. In general, standardised views of still images in either format did not compare well with clinical examinations in diagnostic accuracy. Most notable was the poor detection of even gross eyelid and conjunctival disorders. This was because of the fact that both cameras, with the study setup, did not offer a full view of both lids or both sides of the sclera in one frame. Also, both instruments had a shallow depth of focus. While the corneal surfaces were focused, the lids, especially the lashes, were out of focus. At least four more views would be needed, if the lids and sclera were to be imaged adequately. There were several other reasons for the fact that these standardised still images compared poorly with clinical assessment. First, small signs that were not adjacent to the narrow slit beam (for example, posterior subcapsular cataracts and small corneal scars) were not detected. Second, the assessor's depth perception was reduced due to the lack of movement and stereopsis in the images. Third, flare and cells detection requires high magnification, which was not available to either image formats in this study. Fourth, certain examination manoeuvers such as fluorescein staining or eyelid eversion was only available during clinical examination. Lastly, there was no patient history given to the image assessors.

The design of the study simulated a store and forward telemedicine environment. Store and forward telemedicine has several advantages over real-time telemedicine. A stored–forward system provides much higher image resolution, which is essential in ophthalmology. The maximum resolution in a teleconferencing video system is only 352 × 288 pixels. Also, it is much more time efficient for the clinicians. The time delay is acceptable for the majority of cases. Furthermore, the setup and running cost is less in a stored–forward system.

However, a store and forward system is highly dependant on the quality of the image captured by the remote operator. Although the regulated views used in this study were shown to be limited, there is still a need for a standardized method to capture anterior segment images other than for research purposes. The remote camera operators need to be taught to capture a series of controlled views as they may not have enough knowledge to decide which views to take. This is especially important in anterior segment ophthalmology telemedicine. Image capturing for stored–forward telemedicine is simpler when the target of interest is essentially single planed like the retina or skin. In these cases, the camera operator only has to be concerned with aligning the landmarks in two dimension and focus on the plane of interest. As the eye is a three-dimensional structure with an optical medium, there are many focal planes of interest during image capture. As demonstrated in this study, two to three still images captured from two focal planes are inadequate for accurate diagnosis. It is clear that many more images from different focal planes and different locations should be captured from each eye.

As part of the study, the PSL camera also captured 30–40 s of video from each of the patient's eyes. In these video clips, the slit beam scanned over the eyeball and eyelid several times in successive levels of focus. Each video consisted of 450–600 frames. Furthermore, a moving picture also gives a much better sense of depth. These clips should provide much more diagnostic information. Schiffman et al16 has noted the advantage of store and forward video clips over stored–forward still images.

Despite the shortcoming of using two to three standard views, the moderate to good agreement (unweighted κ) between digital picture grading and clinical grading of lens pathology was very encouraging. The PSL images were shown to be significantly better for cortical opacity detection than images from the ZSL. This is most likely due to the advantage of having a selectable coaxial illumination built into the PSL. Detection of posterior subcapsular cataracts should improve if more image views of the lens were captured. The slit beam often missed the posterior pole of the lens and the camera was never focused on the posterior capsule of the lens. If the sensitivity for posterior subcapsular cataracts is adequate, the instrument could be used as a screening tool for cataracts

Conclusion

The image quality of the portable slit-lamp camera, for anterior segment diagnosis, has been shown to be comparable to a 35 mm slit-lamp camera when used in the setup described in this study. The study team will compare the findings from the movie clips with the clinical gold standard in the next phase of the trial.