Introduction

Substantial attention has been drawn to the inorganic-organic perovskite-based solar cells, which currently achieve a certified high light conversion efficiency of 20.1%1. The combination of several excellent optoelectronic properties, such as very low exciton binding energy2,3, highly mobile charge carriers4,5,6 and efficient charge transportation to selective contact layers3,5,7,8, makes perovskite “a game changer”9 for photovoltaic devices and “a new avenue of research”10. As a fundamental issue, the carrier diffusion in perovskite is a major factor affecting the design and performance of the devices. However, this topic is still under debate at moment. It was shown that the charge diffusion distance in tri-iodine perovskite, CH3NH3PbI3, is ~100 nm, studied by transient fluorescent spectroscopy11,12. On the other hand, many high efficient perovskite solar cells based on CH3NH3PbI3 were made with perovskite layers thicker than this distance13,14,15. It is also investigated by impedance spectroscopy, photoinduced time-resolved microwave conductance (TRMC) and electron beam-induced current (EBIC) method, which hint a much longer charge transfer distance within perovskite layer16,17,18. A study on single crystal even give an extremely long diffusion length above 175 μm19. In addition, the diffusing balance between electrons and holes is not clear either. It was regarded that this balance is well maintained, while some reports say that the diffusion of holes is more/less efficient than electrons17,20.

Beside the diffusion issue, some experimental observations are also in conflict. E.g. the fluorescent lifetime of the CH3NH3PbI3 are dramatically varied in reports. In Xing’s report, the lifetime is 4.5 ns11, while in Stranks’ report, it is 9.6 ns12. Some other experiments show that the lifetime for CH3NH3PbI3 should be much longer than that. In reports by Yamada, the lifetime under low excitation light intensity can be 140 ns21. In single crystal, it is even longer than 100 μs under low excitation intensity19. This is an important parameter when calculating the charge diffusion distance by one-dimensional diffusion model11,12. It seems that all these conflicts need a better explanation.

To clarify these conflicts, we performed a study of directly observing the charge transfer in perovskite with various thicknesses and with an electron/hole transfer layer, by means of time-resolved transient fluorescence. It shows that the charge diffusion in CH3NH3PbI3 is of distance at micrometer scale, which obviously longer than film thickness. The study also explains why former studies provide short diffusion lengths. The results show that hole diffusion is faster than electron within perovskite thin film.

Results

Absorption coefficient of CH3NH3PbI3

All perovskite films discussed here were prepared by a two-step dipping procedure similar to a report22 and our study recently23 on flat glass substrates. Figure 1 shows the absorption coefficient of CH3NH3PbI3 derived from the absorption spectrum (see details in Supplementary Information, SI). This spectrum, which is in line with former reports, covers the entire UV and visible range up to 760 nm11,24. At 517 nm (the wavelength of pump light), a coefficient of 1.2 × 105 cm−1 is slightly higher than the reference11, corresponding to a penetration depth of 84 nm.

Figure 1
figure 1

Plot of absorption coefficient.

Absorption coefficient towards wavelength for CH3NH3PbI3 thin film prepared via a two-step deposition method. Insets are the top view photos of perovskite samples of a thin (yellow brown, left) and a thick (dark brown, right) one on glass substrates.

Thickness dependence of lifetime

The thickness of four prepared perovskite films are determined by a profilometer and listed in Table 1. An insulating polymer poly(methylmethacrylate) (PMMA) layer was coated atop the neat perovskite films for all photoluminescence (PL) decay measurement to passivate their moisture sensitivity25. By excitation at 517 nm, their transient fluorescent decay for the peak emission wavelength are shown in Fig. 2. The lifetimes for each thickness are also listed in Table 1. For a brief comparing, the curves are fitted by stretch exponential decay function26. The lifetimes show thickness dependency. For the films of 63 nm and 156 nm, their decays are 2.8 and 12.6 ns, which is similar to the reports11,12. For the two thick films of 254 and 310 nm, they have quite identical fluorescence decay as 90 ns and 91 ns. This means that the fluorescent decay is thickness-dependent in thin films, which disappears in thick ones.

Table 1 Summary of the stretched exponential fitting for the PL decays in Fig. 2.
Figure 2
figure 2

Thickness-dependent time-resolved PL data.

PL decay curves of CH3NH3PbI3 of different thicknesses depending on varied PbI2 concentration (63 nm, 0.3 M; 156 nm, 0.5 M; 254 nm, 0.8 M; and 310 nm, 1.1 M, respectively) upon excitation at 517 nm, 90 nJ/cm2. The solid lines are the stretched exponential fits to the corresponding results.

CH3NH3PbI3 characterization

The thickness-dependent fluorescence lifetime is analogous to previous reports on perovskite crystal grain size26. To investigate the influence of crystallite nature on PL properties, a series of scanning electron microscopy (SEM) and X-ray diffraction (XRD) measurement were performed on perovskite films of various thickness. Top-view SEM images of four samples mentioned above are shown in Fig. 3a–d. The thinnest film (made by 0.3 M PbI2) in Fig. 3a is a thin layer of individual nanocrystallites with plenty of voids or pinholes. The average grain size is about few tens to ~100 nm. When films became thicker, as shown in Fig. 3b–d, the undesired voids evidently decreased, generating much more compact morphology. Meanwhile, larger crystallites were obtained in thick film, e.g. in Fig. 3d, the crystal size is of ~250 nm. This evolution of grain growth is in agreement with previous reported thermally annealed perovskite films13,15. The corresponding XRD patterns (Fig. 1 in SI) clearly show the perovskite structure (14.66°, 27.09°, 31.82°) with the presence of residual unreached PbI2, in keeping with a previous study about PbI2 deposited on flat glasses22. Moreover, the relative amount of PbI2 decreases when thickness increases. Since the fabrication procedures for each sample are the same, this tendency should be attributed to the thickness for different samples.

Figure 3
figure 3

SEM images.

Top-view SEM images of CH3NH3PbI3 deposited via changing PbI2 precursor concentration. (a) 0.3 M; (b) 0.5 M; (c) 0.8 M; and (d) 1.1 M.

PL decays with and without quenchers

To examine the charge transfer properties of the CH3NH3PbI3 film, transient fluorescence experiments were performed by measuring the PL decay in perovskite film with or without a selected electron or hole acceptor. We made two samples with low and high PbI2 concentrations, as listed in Table 2. Figure 4 shows the corresponding PL results of a thick perovskite film (390 nm). As shown in Fig. 4a and Table 2, the perovskite/PMMA film has a long lifetime of 170 ns, which we will explain in discussion. When the film was coated with a charge transfer layer, PCBM, e.g., fast fluorescent quenching happens (Fig. 4b). The decay is as fast as 1.24 ns, which means highly efficient electron transfer to the interface. It is even faster, when Spiro-OMeTAD is coated above perovskite films, as shown in Fig. 4c. The decay is 0.17 ns, close to the instrument response of streak camera. For the thinner film of 95 nm, PL decays show the same trend (Fig. 2 in SI). The neat perovskite film has a lifetime of 12.4 ns, which decreases to 0.40 ns and 0.16 ns for PCBM and Spiro-OMeTAD coated samples, respectively.

Table 2 Summary of parameters in Fig. 4.
Figure 4
figure 4

Time-resolved PL decays of perovskite CH3NH3PbI3 coated with different layers (red circles).

(a) spin-coated CH3NH3PbI3 perovskite film (390 nm); (b) film covered by PCBM; and (c) covered by Spiro-OMeTAD, excited at 517 nm, 90 nJ/cm2. The black and blue solid lines are the fits to the PL results by rate equation and one-dimensional diffusion model, respectively.

Discussion

The thickness-dependent lifetimes indicate that the fluorescent quenching is neither local, nor to the surfaces. Then this quenching is more like a boundary related effect. We observed that the size of the grain become larger when the films are thicker. In addition, the films become compact with less defects. Then the abundant surface area, void and defects in thin film should be responsible for the quenching. In thick films, the boundary effect become insignificant. In our study, it is above 250 nm, as shown in Fig. 2. This grain size independency was also shown in D’Innocenzo’s report, in which the lifetime of ~100 ns was found for grains between 0.2–2 micron26. Smaller grains presenting dramatically reduced lifetime was also presented. This proves that when the grain size becomes too small, it will produce reduced lifetime. However, in an optimized real device of 300–500 nm, the small size grain is avoided and the lifetime is not sensitive to detailed morphology. On the contrary, in semi-transparent device and thin films for photophysical study, small size grain exists with reduced lifetime. Therefore, towards its real applications, we prefer to take film thickness as a basic parameter to describe the lifetime dependency, instead of grain size and defects, which had already been optimized in cells with high efficiency by many groups.

The SEM and XRD studies reveal the mechanism of this dependency. The SEM images manifest the evolution of grain growth, from the level of below 100 nm to ~250 nm and the diminishing of voids or pinholes. We can rationally assume that long lifetime exists when large size of crystallite are the main species in film and the boundaries between crystallites have a significant impact on the fluorescence characteristic27,28, esp. when the crystal size is small and not compact. Some groups showed that a proper amount of PbI2 species can fill perovskite grain boundaries, eliminate defect states and thus slow down the carrier relaxation, whereas a large amount of excessive PbI2 is detrimental to charge transport29,30. This is in according with our XRD results for thick and thinner perovskite films, respectively. We believe this is the quenching mechanism of thickness dependent fluorescent lifetime. In short, when the perovskite layer become thicker, it has larger crystallite size with reduce overall grain boundary area and much less defects due to reduced PbI2 at boundary. Both the factors finally make the fluorescence emitted by thick film independent to thickness.

It has been well established that excitons in perovskite are nearly fully ionized because of low binding energy2,31,32. So the charge diffusion directly relate to their lifetime. Therefore, the film thickness dependent fluorescent lifetime becomes an important issue here. When the film is thin, the lifetime is short due to the boundary defects. This means that to find out the unaffected charge diffusion distance, the real lifetime needs to be established in advance. As mentioned earlier, in the thicker film of ~280 nm, a long lifetime of 140 ns can be found at lowest pump intensity21. In addition, when the perovskite are in large crystal of ~1 μm, its lifetime is also at ~100 ns timescale26. These results are very similar to our observation in thick films. It reasonably suggests that the native lifetime for CH3NH3PbI3 without considerable boundary defect is at ~100 ns timescale, though small differences exist among research groups. To our knowledge, for reports whose lifetime is ~100 ns, they are shown as thicker films or larger grain sizes.

We performed a step forward experiment to verify the grain size dependent fluorescent decay. For thick film of 345–390 nm, grain size of 150–350 nm were prepared, shown in SI Fig. 4. We found that grain size has little effects on lifetime. All of them present lifetime of ~200 ns, as shown in SI Fig. 5 and SI Table 1. Therefore, we can conclude that the grain size should have no significant effect on charge diffusion. The grain size has large tunable range when it is compact and with less defects. This also explains that high performance perovskite solar cells can be repeated among labs, in spite of morphological variation.

The one-dimensional diffusion model are describe in the SI. The fittings produce diffusion constant, D. The charge diffusion distance LD, is calculated by the equation , where τPL is PL lifetime of 390 nm in this study by simple rate equation21,31,32. We take the duration when fluorescence decays to 1/e of initial intensity as the diffusion time for electrons and holes. As summarized in Table 2, we obtain the electron diffusion coefficient of 0.18 cm2 s−1 and corresponding diffusion length of ~1.7 μm. This confirms the observation by EBIC method17. As a comparing, when taking lifetime of the thinner film, 95 nm, we can calculate the corresponding diffusion constant of 0.06 cm2 s−1 and diffusion length of 273 nm. This result is close to other reports based on thin films11,12,33,34. For diffusion of hole, the D is found 2.3 cm2 s−1 in perovskite/Spiro-OMeTAD film, which is one order larger than the electron diffusion coefficient. The corresponding charge diffusion distance is calculated as ~6.3 μm. For the thin film of 95 nm, this distance is 459 nm, as listed in Table 2. It should be remarked here that both the electron and hole’s diffusion distances obtained are much longer than the thickness of the perovskite films ever made with top solar energy conversion efficiency.

There are several points should be addressed here. The first is that though the charge diffusion distance are thickness dependent, it is much longer than the film thickness, even for the thin film less than 100 nm. Therefore, both in thin semi-transparent devices and black thick devices, the high cell performance can both be achieved. The long diffusion distances provide a large tunable range for preparing perovskite working layers, both for thickness and morphology. The second is that the charge transfer balance between the electrons and holes is not exactly shown in our study. However, it is less important for currently developed devices, which are usually 300–500 nm. The last one is that the lifetime of CH3NH3PbI3 may not be an exact number but a range around 100 ns or larger, since the crystal grain size, defects, preparation procedure, post treatment, pump energy and fitting methods et al. can not exactly be the same. However, the lifetime variation will not make the diffusion distance lower than micrometer level.

The plain films provide a simple model to study the charge diffusion inside perovskite working layers. Another widely applied cell structure is with mesoporous scaffold such as TiO2 and Al2O3. Due to the restricted growth for crystals and large interface area, the charge diffusion inside the structure is much more complicated. E.g. the fluorescent lifetime varies to the size of mesoscopic pores, the type of scaffold and capping layer35,36. High speed charge transfer to TiO2 of 200 fs was found through ultrafast spectroscopy, which will benefit these devices which possess large interface area37. At molecular level, the interface shows oriented permanent dipoles indicating the existence of ordered perovskite layer38. In addition, the perovskite inside scaffold shows co-existence of medium range crystalline and local structural coherence39. These studies reveal the significant differences of these mesoporous cells compared to the planner structured devices. In spite of these differences, it has been reported that the charge diffusion in mesoporous devices has lower coefficient than the layered cells40. It is also reported that the existence of capping layer may recover the longer charge diffusion distance36. Base on these study, we can summarize that the charge diffusion distance inside the mesoporous layer is smaller, but may benefit from the large interface area because of the highly efficient charge extraction5,37.

In conclusion, we found that the fluorescent lifetime of spin-coated CH3NH3PbI3 perovskite thin film depends on the film thickness. The lifetime increases towards the increment of film thickness, till ~250 nm. The lifetime finally increases to >100 ns. The fluorescent quenching in thin film is due to the defects at grain boundary. Therefore we take a thick film of 390 nm to study the charge diffusion in CH3NH3PbI3. After coating charge-transfer layer, PCBM or Spiro-OMeTAD and applying one-dimensional diffusion model, we can obtain the charge diffusion distance of 1.7 μm for electrons and 6.3 μm for holes. For thin film of 95 nm, a result of short diffusion distance similar to other reports is found. This study resolves the current conflict between the measured short charge diffusion distance and thick working layer in high efficient devices. The result shows that in case of thin and thick films, CH3NH3PbI3 both can provide long charge diffusion distance for best cell performance.

Methods

Sample preparation

All samples were fabricated on glass slide substrates. First, glass substrates were cleaned sequentially by ultrasonic bath in detergent water, deionized water, acetone and ethanol for 15 min, respectively and then exposed to oxygen plasma for 15 min to achieve optically smooth films. The CH3NH3PbI3 perovskite were fabricated with a two-step sequential deposition method under nitrogen atmosphere. The pre-cleaned glass substrates were spin-coated a PbI2 solution (6000 rpm, 0.3 M, 0.5 M, 0.8 M and 1.1 M) of N,N-dimethylformamide (DMF) at ambient temperature to obtain layers of different thicknesses. After drying at 60 C in ambient environment for 6 h, the films were dipped into CH3NH3I solution in 2-propanol (15 mg/mL) at 65 C for 90 s and then rinsed with 2-propanol. For the samples fabricated by using various concentration of CH3NH3I solution (10 mg/mL, 15 mg/mL and 20 mg/mL), PbI2 was kept at 1.0 M. After CH3NH3PbI3 annealing at 100 C for 40 min, Spiro-OMeTAD (10 mg/mL), PCBM (10 mg/mL) or PMMA (20 mg/mL) was spin-coated at 2000 rpm for 60 s atop the CH3NH3PbI3 perovskite films.

Characterization details

XRD patterns were obtained using a Philips X’PERT-MRD x-ray diffractometer system with a Cu K radiation source (λ = 0.1541 nm) at 45 kV and 40 mA. SEM images were collected using a Hitachi S-4800 microscope, with a working bias of 10 KV. Sample thicknesses were measured using a Veeco Dektak 150 profilometer. Ultraviolet-visible (UV-vis) absorption measurements were recorded with an Agilent 8453 UV–vis Spectroscopy System at room temperature.

Time-resolved photoluminescence

The time-resolved fluorescence spectra were recorded with a high resolution streak camera system (Hamamatsu C10910). We used an amplified mode-lock Ti: Sapphire femtosecond laser system (Legend, Coherent) and a two-stage optical parametric amplifier (OperA Solo, Coherent) to generate the pump beam with a repetition rate of 1 KHz. All the samples were excited by 517 nm at room temperature. The excitation fluence on the sample surface was in the range from 9 nJ/cm2 to 1.2 μJ/cm2 per pulse.

Additional Information

How to cite this article: Li, Y. et al. Direct Observation of Long Electron-Hole Diffusion Distance in CH3NH3PbI3 Perovskite Thin Film. Sci. Rep. 5, 14485; doi: 10.1038/srep14485 (2015).