Abstract
A 22 year old man with incomplete quadriplegia (C6-7) was unable to perform either a sliding or a pivot transfer. He was instrumented with an implanted functional neuromuscular stimulation (FNS) system, radio frequency-linked to a belt-worn controller. The system activated eight muscles selected from among quadriceps, hamstrings, posterior portion of the adductor magnus, gluteus maximus, and erector Spinae, bilaterally. The two-stage implantation procedure included electrode implantation with percutaneous leads followed by stimulator implantation and removal of the percutaneous leads. All implants were well tolerated with no adverse effects. The subject was able independently to put on the external controller portion of the system and to perform a standing pivot transfer with only standby assistance. An unexpected outcome of the FNS system use was increased voluntary upper body strength that resulted in improvement of the sliding transfer from ‘inability’ to ‘independent’.
Similar content being viewed by others
Article PDF
References
Guttmann L (1973) Spinal Cord Injuries. Comprehensive Management and Research. Blackwell Scientific Publications, Oxford.
Sorrentino S (1987) Mosby's Textbook for Nursing Assistants. Mosby Co, St Louis, MO.
Kantor C, Andrews B J, Marsolais E B et al (1993) Report on a conference on motor prostheses for workplace mobility of paraplegic patients in North America. Paraplegia 31: 439–456.
Chaplin E, Habasevich R (1992) Ambulation with functional electrical stimulation (FES) in individuals with paraplegia and incomplete quadriplegia. Neurology 42 (Suppl 3): 638.
Petrofsky J S, Phillips C A, Douglas R, Larson P (1986) A computer controlled walking system: The combination of an orthosis with functional electrical stimulation. J Clin Eng 11: 121–133.
Marsolais E B, Kobetic R, Chizeck H J, Jacobs J L (1991) Orthoses and electrical stimulation for walking in complete paraplegia. J Neurol Rehabil 5: 13–22.
Keith M W, Peckham P H, Thrope G B et al 1988 Functional neuromuscular stimulation neuroprostheses for the tetraplegic hand. Clin Orthop 233: 25–33.
Marsolais E B, Kobetic R (1986) Implantation techniques and experience with percutaneous intramuscular electrodes in the lower extremities. J Rehabil Res Dev 23: 1–8.
Scheiner, Polando G, Marsolais E B (1994) Design and clinical application of a double helix electrode for functional electrical stimulation. IEEE Trans Biomed Eng 411.
Salvati E A, Robinson R P, Zeno S M et al 1982 Infection rates after 3175 total hip and total knee replacements performed with and without a horizontal unidirectional filtered airflow system. J Bone Joint Surg Am 64: 525–535.
Smith B, Peckham P H, Keith M W, Roscoe D D (1977) an externally powered, multichannel, implantable stimulator for versatile control of paralyzed muscle. IEEE Trans Biomed Eng 34: 499–508.
Rushton D H, Brindley G S, Polkey C K, Browning G V (1989) Implant infections and antibiotic impregnated silicone rubber coating. J Neurol Neurosurg Psychiatry 52: 223–229.
Osman S G, Marsolais E B (1994) Endoscopic electrode implantation for enhanced stimulation of the hamstring branches of the sciatic nerve. Arthroscopy 10.
Granat M H, Ferguson A C B, Andrews B J, Delargy M (1993) The role of functional electrical stimulation in the rehabilitation of patients with incomplete spinal cord injury observed benefits during gait studies. Paraplegia 31: 207–215.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Marsolais, E., Scheiner, A., Miller, P. et al. Augmentation of transfers for a quadriplegic patient using an implanted FNS system. Case report. Spinal Cord 32, 573–579 (1994). https://doi.org/10.1038/sc.1994.91