Abstract
The relationship between the applied torque and the load required to pull off the calipers was established on cadavers. Torques of 0.7, 0.8 and 0.9 Newton-metres were applied to a specially adapted skull traction caliper whilst attached to the skull in the normal position when used clinically, and loads were applied in steps of 2 lb up to a maximum of 60 lb. It was found that at a torque of 0.7 Newton-metres, the caliper became detached at the maximum load, but still held during traction at torques above this value. Taking into account the fall in load with time that may occur clinically and the limiting torque for pull-out with a traction load of 60 lb, it is suggested for complete clinical safety that a torque of 0.9 Newton-metres should be applied. An open-ended torque release spanner was set for 0.9 Newton-metres for clinical use such that the compressive load is sufficient to prevent pull-out even in the unlikely event of 60 Ib being applied for up to 21 days.
Similar content being viewed by others
Article PDF
References
Blackburn J 1938 A New Skull Traction Appliance, South Surgeon 7:16.
Cone W, Turner W G 1937 The treatment of fracture-dislocations of the cervical vertebrae by skeletal traction and fusion. Journal of Bone and Joint Surgery 19:584–602.
Crutchfield W G 1933 Skeletal Traction for Dislocation of Cervical Spine, Report of a Case. South Surgeon 2:156–159.
Crutchfield W G 1954 Skeletal Traction in Treatment of Injuries to the Cervical Spine. Journal of American Medical Association 155:29–32.
Crutchfield W G 1966 Redesigned Crutchfield Skull Tongs. Journal of Neurosurgery 25:656–657.
Grundy D J 1983 Skull Traction and its Complications. Injury 15:173–177.
Vinke T H 1948 A Skull-Traction Apparatus. Journal of Bone and Joint Surgery 30A:522–524.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Minns, R., Sutton, R. Torque pull-out load characteristics of skull traction calipers of ‘ice tong’ type (with particular reference to the cone type of caliper). Spinal Cord 23, 192–196 (1985). https://doi.org/10.1038/sc.1985.33