Abstract
Peripheral nerves are often vulnerable to damage during surgeries, with risks of significant pain, loss of motor function, and reduced quality of life for the patient. Intraoperative methods for monitoring nerve activity are effective, but conventional systems rely on bench-top data acquisition tools with hard–wired connections to electrode leads that must be placed percutaneously inside target muscle tissue. These approaches are time and skill intensive and therefore costly to an extent that precludes their use in many important scenarios. Here we report a soft, skin-mounted monitoring system that measures, stores, and wirelessly transmits electrical signals and physical movement associated with muscle activity, continuously and in real-time during neurosurgical procedures on the peripheral, spinal, and cranial nerves. Surface electromyography and motion measurements can be performed non-invasively in this manner on nearly any muscle location, thereby offering many important advantages in usability and cost, with signal fidelity that matches that of the current clinical standard of care for decision making. These results could significantly improve accessibility of intraoperative monitoring across a broad range of neurosurgical procedures, with associated enhancements in patient outcomes.
Similar content being viewed by others
Introduction
Injuries to peripheral nerves during surgical procedures constitute a significant source of morbidity and result in worsened quality of life for many patients.1,2 Iatrogenic peripheral nerve damage is a particularly common and devastating clinical entity that leads to significant pain and compromised functional outcomes.2,3,4 For example, approximately 5% of patients undergoing arthroscopic hip repair suffer from transient neuropraxia.5,6 Permanent damage to nerves is a well-known and debilitating health risk associated with peripheral, cranial, and spinal nerve access.4 In such cases, intraoperative monitoring techniques, primarily involving evoked potentials as well as stimulated and spontaneous electromyography (EMG), provide surgeons with the ability to identify and assess vulnerable nerve sites by probing nerve-muscle activities during surgery.7,8,9,10,11,12 Real-time monitoring strategies offer a powerful set of capabilities for surgeons that can positively affect outcomes, but existing intraoperative systems are large, expensive, and cumbersome; they include data acquisition consoles coupled to sensing electrode leads via multiple, fixed electrical connections. These platforms demand active engagement of trained technicians who are skilled in their operation and in the placement of needle electrodes, thus preventing access to neurosurgical care in settings where such tools are unavailable and/or where the associated costs cannot be supported.13 Furthermore, the wired interfaces create spatial complexity and limit electrode placement14 and the needle electrodes used for recording EMG signals cause discomfort due to penetration through the dermis,8 commonly leading to postoperative pain and/or soft tissue hematomas.15,16 The quality of the EMG signals also depends critically on precise placement of these needles. This set of clinical limitations establishes a clear need for alternative intraoperative monitoring approaches.
An ideal solution would offer multimodal measurement modalities and high signal quality, but in physical formats and functional designs that provide: (1) soft and conformal tissue interfaces compatible with skin across all regions of the body, (2) wireless communication capabilities and on-board power supplies capable of supporting operation continuously during an entire surgical procedure, (3) non-invasive electrodes that interface with the skin, without requiring penetration, specialized skin preparation procedures or precise spatial positioning, (4) easy-to-use interfaces with automated data capture, storage, and real-time analysis, and (5) path to low-cost embodiments. Here, we present a thin, soft biosensing device, referred to as “biostamp”, and demonstrate the utility of its unique features during intraoperative neurosurgical monitoring. The soft mechanical construction and advanced encapsulation strategies allow the use of the biostamp prototype across a broad range of neurosurgical procedures, which has not been demonstrated previously. Furthermore, the soft system-level mechanics enable intimate mechanical coupling of the biostamp to the skin, even on sensitive curvilinear regions of the face, where it can capture electrical activity (e.g., EMG) and movement of targeted muscle groups in response to direct electrical stimulation of nerves, in real-time during critical interventional procedures. Detailed clinical studies on multiple patients establish the performance characteristics and practical advantages of this technology relative to conventional needle and surface-EMG (s-EMG) measurement techniques during surgeries on peripheral, spinal, and cranial nerves.
Results
Soft, wireless, skin-integrated platform for intraoperative monitoring
Recent advances in materials, mechanics designs, and manufacturing methods establish the foundations for classes of thin, mechanically compliant electronic systems that enable multimodal sensing on the surface of the skin at nearly any body location.17,18,19 As demonstrated here, these platforms combine high-performance electronics and biosensors with wireless functionality to achieve high accuracy monitoring of muscle activity in response to nerve impulses and intraoperative stimulation. Figure 1a presents a schematic, exploded view of the design. The functional sub-components distribute across a collection of “islands” that interconnect electrically and mechanically via narrow, filamentary serpentine traces, optimally configured to create low modulus, “spring-like” mechanics in geometries guided by computational modeling of the mechanical and electrical characteristics. Encapsulating this “island-bridge” mesh network above and below with a low modulus, silicone elastomer defines skin-compatible physical properties, as a soft and comfortable interface to the skin (Fig. 1b). The resulting form factor and intimate skin interface are strikingly different than those of conventional wearable devices, in which rigid packaged electronic components mechanically attach to the skin via straps, penetrating pins, tapes, or bands. Stretching, twisting, bending, and other complex modes of deformation can be accommodated without altering functional operation (Fig. 1c). Robust, comfortable coupling to the skin even at tightly curved regions of the anatomy (e.g., ankle) and sensitive parts of the body (e.g., face) are possible, thereby supporting dual s-EMG and motion sensing from multiple high-flexion or contractile muscle groups, in a mode that is mechanically imperceptible to the patient and physically confined without restricting the physician.
An important enhancement of the biostamp designs used here involves embedding the active components in an ultra-low modulus silicone formulation (3–5 kPa) as a core, with a surrounding thin silicone shell that has a slightly higher modulus (50–100 kPa). This core/shell configuration allows additional degrees of freedom and motion of the serpentine interconnect structures, thereby leading to an effective modulus of 390 kPa, much smaller than that of standard designs (890 kPa) (Fig. S1). The average thickness (~2.5 mm), the physical size (3.4 × 6.6 × 0.3 cm3), and the weight (7 g) of such devices are similar to those of a standard gauze patch, qualitatively differentiated from any alternative commercial monitoring system currently available.
The functional electrical components include a Bluetooth® Low Energy Smart radio, flash memory module (32 MB), 3-axis accelerometer and gyroscope, an analog-front end that connects to two Ag/AgCl electrodes (single lead configuration), and a rechargeable battery (15 mAhr; 10 × 25 × 1.6 mm3; wirelessly charging) that enables simultaneous measurement of motion and s-EMG signals for up to ~16 h. The accelerometer has an adjustable sampling rate (12–50 Hz), 2-mg sensitivity, and 12-bit resolution, sufficient to capture fine muscle movements, twitching, and spasms. The analog front-end samples s-EMG signals at 250 Hz with an amplification factor of 21.6 dB and a band-pass filter from 0.5–125 Hz. The low-end corner filtering is a byproduct of the AC-coupled relationship between the Ag/AgCl electrodes and amplifier electronics, while the high-end is set by the analog front-end to satisfy the Nyquist condition of the analog-to-digital converter (ADC).
Quantitative analysis of the mechanical attributes
Figure 1d presents a schematic cross-sectional illustration of the design (as in Fig. 1a, b) in the core/shell configuration described above.20 The shell provides mechanical stability, elastic restoring force, and skin-compatible interface; the core allows the island/bridge construct to mechanically deform freely, thereby minimizing the material strains induced by system deformation. Additionally, optimized choices for the thickness of the core (Silbione, elastic modulus 3 kPa, and thickness, hcore in Fig. 1d), guided by computational mechanics, suppress stresses that can result from mechanical loading of the surface of the skin.21 Simulation results summarize the dependence on hcore for the case of 20% tensile strain at the system level (Fig. 1e, Fig. S1). The interfacial stresses for hcore = 0.5 mm and 1.5 mm lie below values associated with thresholds for sensation in the underlying skin (~2 kPa)20 in most regions; hcore = 2.5 mm enables even further reductions. Figure 1f shows the strain distributions in the first copper interconnect layer from the top, for the case of an applied strain that corresponds to the limit of elastic stretchability (with yield strain of ~0.3% in copper22). In all cases, the strains in the island regions are negligible. The stretchability increases with hcore (Fig. 1f) to levels comparable or larger than the elastic limit of human skin. The results in Fig. 1e and f highlight the advantages of large values of hcore (~2.5 mm). Computed results for bending appear in Figs. S2–S5. Figures S3 and S4 show the distributions of interfacial stresses on the skin for different curvatures (κ = 0.0092, 0.0184, and 0.0276 mm−1), corresponding to different bending angles (α = 30°, 60°, and 90°) as shown in Fig. S2. Even at large flexion angles (~90°), the interfacial stresses for hcore = 2.5 mm are below threshold values for skin sensation (~2 kPa). The corresponding strain distributions in the copper fall within the elastic limit for all three cases (i.e., hcore = 0.5, 1.5, and 2.5 mm).
Quantitative comparisons of usability and signal fidelity to clinical standards
Current state-of-the-art intraoperative monitoring systems provide high-quality signal recordings, but they are large and complex, and the costs of the equipment and of the trained personnel necessary for its operation are prohibitive. Additionally, the wired connections and computer consoles needed to analyze and display data (Fig. S6) are cumbersome for surgeons and support staff in the operating room. A conventional system of this type (Cascade IONM) offers high sampling rates (25.6 kHz), high signal resolution (18-bit), and low noise operation (<2 µVRMS) optimized for EMG recordings, with penetrating needle electrodes or surface electrodes as the measurement interface. By contrast, the biostamp embeds all necessary electronics and electrodes in a single, compact platform that, itself, softly couples to the skin in a straightforward, non-invasive manner, without need for specialized skill or training, to nearly any region of the body. Although the biostamp is not hermetically sealed, the silicone encapsulation layer prevents water ingress, and thereby mitigates risks of moisture exposure to the circuitry during intraoperative monitoring procedures. The overall size (~2 orders of magnitude) and mass (~2 orders of magnitude) of the biostamp are qualitatively smaller than those of conventional hardware used for intraoperative monitoring. The wireless, battery-powered operation and intimate skin interface isolates the system from noise associated with power lines, motion artifacts, and ambient electrical interference, as demonstrated in our previous systems.17,18,19 Although sampling frequencies and levels of resolution (100 μVRMS, 1 kHz, and 16-bits) are somewhat lower than those of conventional systems, the quality of the data determined by signal-to-noise ratio (SNR) analysis, in practical clinical contexts, is comparable (to within ~1 dB) to that of s-EMG research tools (Delsys EMG recording system, Fig. S7). Further improvements in SNR can achieved by employing thin hydrogel layers between the skin and the recording electrodes.
Monitoring of nerve-muscle function during surgeries on the peripheral, spine, and cranial nerves
Standard neurosurgery procedures expose targeted nerves identified for tumor removal, decompression, grafting, or other medical purposes (see “Methods” for more details). In this context, direct nerve electrical stimulation represents a common neurosurgical technique for locating, visualizing, and assessing the health of neural–muscle interfaces. Here, electrical current pulses excite nerves and polarize neuronal cell membranes, thereby producing an action potential in the nerves that leads to muscle contraction.23,24 The induced activation of the muscles generates a corresponding EMG response, typically recorded with penetrating needle electrodes and large-scale data acquisition systems. The minimum current that elicits a measurable EMG signal defines the stimulation current threshold. This response is strongly non-linear, with zero signal below this threshold and a smooth, monotonic increase with current above it, until saturation at some upper limiting value. With the known distance between the stimulation and measurement sites, these data allow determination of the nerve conduction velocity, a metric that highlights and characterizes nerve damage.25 Although threshold levels vary according to different physiological factors and the relative location of the nerves and muscle groups,26,27 comparison of measured thresholds serves as an effective means to compare the performance of the skin-interfaced biostamp devices reported here with clinically established tools.28 This approach simply quantifies the comparisons; clinical use typically involves stimulation at levels significantly above threshold.
The clinical studies involved biostamp devices placed on one or more muscles innervated by the target nerve with consenting patients undergoing needle EMG recordings using a state-of-the-art intraoperative monitoring system during peripheral nerve, spine, or cranial nerve surgical procedures (Tables 1–3, at Northwestern Memorial Hospital). We compared measurements from the biostamp platform to those using two types of EMG electrodes with hard–wired interfaces to conventional data acquisition and conditioning electronics: needle electrodes (with Cascade IONM electronics) and surface electrodes (also with Cascade IONM electronics) (Table S1). In most perpheral surgeries, all three types of recording systemswere applied (Fig. 2a) near the surgical access point to the common peroneal nerve (Fig. 2b) for capture of EMG during stimulation.
As outlined in detail subsequently, such measurements revealed that stimulation current thresholds determined using the biostamp were statistically indistinguishable from those determined using Cascade electronics coupled to needle electrodes and surface electrodes (Fig. 2c). In addition to high-quality electrical measurements, the biostamp supports motion sensing with an integrated accelerometer, as a complementary data stream for detecting muscle activation, where mechanical response serves as the basis of measured signal. In Fig. 2d (top three traces), motion recorded using a tri-axis accelerometer captures small mechanical vibrations caused by contractions of the tibialis anterior muscle group. The mechanical activation profiles observed in the y-planes and z-planes align with the induced EMG response of the subjacent muscle group (Fig. 2d, bottom trace). Multimodal sensing in this manner enables tracking of both electrical and mechanical signatures of muscle activity, to provide redundancy in monitoring of muscle response and to enable compensation for motion-induced artifacts. This capability could also provide insight into the depth of anesthesia.29,30,31 In control experiments, we tested for the presence of motion or other artifacts by positioning the nerve stimulator probe above threshold at a neighboring non-neural tissue site, and showed that stimulation of non-neural tissue does not generate measurable EMG signals (Fig. S8). Although motion detection represents an area of opportunity, the work reported here focuses on EMG because of its use as the current standard of care.
In careful comparative studies in patients undergoing peripheral nerve surgery, biostamp devices detected average current thresholds similar to those reported for conventional monitoring systems with needle electrodes (Fig. 3a; n = 10 patients). The precise placement of the biostamp is less susceptible to noise as the biostamp can detect average threshold currents similar to the conventional needle-based monitoring system. The biostamp and the needle-based system strongly correlated with nearly 95% of the two data sets falling within +0.18 mA and −0.15 mA (Fig. 3b), across n = 55 peripheral nerve surgery subjects.
This finding is important, partly because the morphology of the s-EMG waveforms captured by the biostamp had lower peak amplitudes and different decay envelopes compared to those recorded with needle electrodes in the conventional manner. The differences in amplitude varied across subjects, likely influenced by the position of the surface electrodes and the nature of the skin barrier in the case of the biostamp, and the precise placement of the needle electrodes in the case of the clinical standard. The data envelopes depend on the sampling rates and the frequency-dependent electrical impedance, both of which are different for these two systems.
Examination of representative data (patients 1 and 2) highlight some of the key findings. For patient 1, the biostamp and conventional needle platforms both exhibited current thresholds in the ~1.8–2.2 mA range, and high-quality signals with stimulation levels commonly applied during neurosurgery (Fig. 3c, d). Specifically, the EMG waveforms captured by the biostamp exhibit SNR values of ~10 dB at threshold and reach ~50 dB in response currents employed during routine neurosurgery procedures (Fig. 3e). The needle electrode system offers improved SNR, but in clinical terms, SNR thresholds above 10 dB are equivalent, in the sense that they offer routine ability for threshold detection. For patient 2, the EMG waveforms also showed robust muscle signal patterns in response to current stimulus levels in a safe operating range (Fig. 3f, g). Similar to the measurements in patient 1, the SNR levels for the needle electrode platform (~60 dB) and biostamp (~30 dB) were well above the noise at the threshold current level (biostamp: 1.4 mA, needle: 1.3 mA), and increased with current in an expected manner above threshold (1.3–1.7 mA, Fig. 3h). The differences in signal levels between these two systems largely depend on the electrode placement and stimulation location. Increasing the stimulus current resulted in increasingly large muscle responses, up to EMG amplitudes of ~0.15 mV for the biostamp and ~0.4 mV for the needle electrode platform. In both cases, the SNR was sufficiently high to detect muscle activation, throughout the normal range observed during surgical procedures.
In addition to peripheral nerve surgeries, the biostamp can provide intraoperative monitoring insights during spinal and facial nerve stimulation. As in the other cases, in spinal nerve procedures, current threshold levels for the biostamp were consistent with those of needle electrode platform (Fig. 4a–c). Across all four patients examined, threshold values determined by the two platforms are comparable, ranging from ~1–7 mA (Fig. 4c). As before, EMG waveforms measured by the biostamp and the needle electrode platform had SNRs well above the noise floor (>10 dB). In both cases, the SNR increases with stimulation current level but approached similar values above 2.3 mA (Fig. S9), indicating that the nerve fibers are activated, and thereby causing a saturated muscle response. In facial nerve procedures, the measured current thresholds using the biostamp and needle electrodes were also similar (Fig. 4d–f). Taken together, these findings demonstrate the broad applicability of the biostamp for intraoperative monitoring across multiple, highly sensitive nerve targets.
Discussion
The results presented here demonstrate that wireless, skin-mounted device technologies that exploit soft mechanics, multifunctional electronics, and precision biosensors enable continuous electrical and mechanical monitoring of muscle responses during peripheral nerve, spine, and cranial surgeries. The ultra-soft, miniaturized form factor facilitates direct attachment to a broad range of muscle groups, including challenging regions of the anatomy such as the face. The result is a non-invasive, easy-to-use platform for capturing EMG and motion signals, with capabilities that reproduce the key functionalities of conventional, large-scale electronic platforms, which serve as the current clinical standard of care (Figs. 3, 4, Fig. S2).
Compared to the established intraoperative monitoring systems (for the purpose of this experiment we used a Cadwell Cascade IONM), the physical design of the biostamp offers significant advantages in multimodal sensing (electrophysiology and accelerometry), size, weight, and comfort, without any practically significant sacrifices in signal fidelity or threshold detectability for intraoperative monitoring. The simplicity of operation eliminates the need for personnel with specialized training in needle electrodes and EMG recording. These usability attributes, taken together, with the wireless mode of operation, offer the potential to significantly simplify clinical preparation effort and time in the operating room. In particular, the heavy demands associated with needle electrodes, the complex procedures for insertion into the tissue, and the management of wired cables that must be secured to the body of the patient and routed across the operating room table to adapter boxes and computer control systems could be greatly minimized. Figure S10 provides a series of EMG recordings obtained from the flexor carpi radialis muscle of a human subject to confirm recording stability of the biostamp over the duration of a typical surgery (<10 h). The EMG amplitude and SNRs stimulated by maximum voluntary muscle contraction in recordings remain the same (±0.2 V and 2.8 dB) during 10 h of wearing, which ensures the recording stability throughout the duration of a surgery. In addition, biostamps are placed in multiple body locations to measure electrical potentials simultaneously. Concurrent monitoring of EMG, electrocardiography, and electrooculography offers complimentary diagnostic capabilities during neurosurgical procedures (Fig. S11).
These attractive physical and operational characteristics can serve to drive procedural uptake in hospitals where complex and expensive monitoring equipment is unavailable or untenable. The ease of setup offers significant benefit to medical facilities with untrained personnel, thereby significantly broadening access to intraoperative monitoring. In addition, the soft form factor and ability for attachment to multiple body locations make this type of technology applicable across a variety of surgical procedures and clinical scenarios. For example, these platforms could be used to monitor unintended stretching of the sciatic nerve during hip surgery and the axillary nerve during shoulder surgery, both of which represent common complications of orthopedic procedures. Real-time monitoring of salient nerve health to minimize intraoperative nerve damage during parotidectomies represents another intriguing possibility. Other types of demanding surgeries where peripheral nerves are vulnerable, and where monitoring could be valuable, include procedures targeting the neck, where the spinal accessory nerve is susceptible to damage, as well as the abdomen and pelvis, where the lumbosacral plexus is at risk. The ease of use suggests the possibility for many other clinical applications of these technologies in monitoring nerve muscle integrity.
More generally, the successful application of biostamp in a surgical context foreshadows other modes of operation in different clinical use cases. Specifically, many recently reported “soft” sensor technologies18,19,32 can easily be incorporated into the biostamp platform to capture additional physiological parameters of interest. Examples include blood flow,33 blood pressure,34,35 temperature,36 hydration state,37,38 tissue stiffness,39 mechano-acoustic signatures,17 swelling,40 and many others.41,42,43,44,45,46 Electrical and thermal stimulation represent additional actuation possibilities. These unique sensing and actuation capabilities could enable monitoring over a broad range of surgeries beyond neurophysiological procedures. For example, in reconstructive surgical procedures, the state of health of various types of tissue flaps could be monitored by measuring blood flow in the region of the transplanted tissue and from the feeding arteries. To surveil and prevent skin pressure ulcers in immobile patients, diabetics and those with peripheral neuropathies, these added sensors and actuators could be applied to measure tissue stiffness, hydration, swelling, and temperature, providing early warning signs of tissue breakdown and ulcers. The onboard processor and wireless connectivity in combination with the biostamp sensors and actuators would ultimately allow operation in an automated mode as part of a closed loop system, which senses for skin and muscle injuries, and, in turn, delivers therapy (via drug release, thermal activation, or electrical stimulation).
In summary, soft mechanics, compact size, and wireless modes of operation in advanced skin-mounted electronic sensing technologies have the potential to fundamentally improve the state of intraoperative monitoring during neurosurgical procedures. Based on our clinical studies, the combination of biostamp measurement platforms and traditional stand-alone nerve stimulators can provide surgeons with the opportunity to monitor nerve and muscle function during a wide range of operations in which nerves are at risk of damage. Adoption of this class of wireless wearable technology may not only simplify the state of intraoperative monitoring, but also improve patient outcomes during invasive surgical procedures.
Materials and methods
Design of clinical study
This study was approved by the Northwestern University’s Investigational Review Board (IRB #: STU00201505) and consents were obtained from all patients prior to undergoing surgery and the scientific research study. All patients were administered general anesthesia without long-lasting paralytic agents so that muscle activity could be monitored. Once anesthetized, the biostamp devices were placed on the skin overlying relevant muscles that were also being monitored using standard needle EMG techniques. In several patients, conventional s-EMG electrodes were also used. Nerve–muscle activities were measured once target nerves were exposed surgically and direct current stimulation could be applied, as instructed and overseen by the main surgeon.
Conventional EMG sensing electrodes and stimulation
Needle electrodes (Rhythmlink, 13-mm long, 0.4-mm diameter, 1.5-m leadwire, SP119022, stainless steel) served to monitor nerve activity during the surgeries. A stimulation probe delivered direct current to the targeted site (Prass Standard Flush-Tip Probe, Medtronic Xomed, ~10-cm long, stainless steel with plastic handle). The electrodes electrically connected to an external stimulator box (conventional intraoperative monitoring system). The stimulation pulses consisted of monophasic waveforms (at 2.6 Hz, 200 μs pulse width), at adjustable current levels with control at the level of 0.01 µA. The s-EMG recordings used hydrogel adhesive electrodes with Ag/AgCl backing layer (AMBU/Neuroline Surface Electrodes, Disposable, 700 SERIES). The distance between electrodes on the biostamp was ~5 cm. The needle and s-EMG electrodes were positioned ~5 cm apart, next to the biostamp, to facilitate comparison. The details of the data acquisition system and noise floor for the s-EMG electrodes and the needle electrodes can be found in the Table S1 and Fig. S7.
Surgery to expose and electrically stimulate nerves while recording from muscles
The skin overlying the nerve was infiltrated superficially with local anesthetic and epinephrine. The peripheral nerve was then exposed using standard surgical techniques. Once exposed, a monopolar electrical nerve stimulation device (Prass Standard Flush-Tip Probe, Medtronic Xomed) applied electrical current at different threshold levels either directly to the surface of the nerve or to the surrounding tissues as a control (Fig. S12).
Stimulation current threshold study
To define the thresholds, electrical stimulation was directly applied to the surface tissues of the nerve. The stimulus starts at electrical current levels that are too low to elicit activity in the studied muscle as measured by conventional EMG monitoring system. The stimulus level is then increased gradually until a muscle response could be recorded using the biostamp and the conventional intraoperative monitoring system (Fig. S7) separately. A trained intraoperative monitoring specialist at Northwestern Memorial Hospital determined the threshold current value at the point when the EMG signal waveform was visually distinguishable. This procedure was repeated six times and the stimulus threshold level for eliciting a detectable muscle response was recorded during each trial. The data were collected and then analyzed postoperatively. The reported values correspond to the average of six trials with 1 standard deviation as error bars.
Signal recording
Needle electrode and s-EMG signals were recorded using the conventional intraoperative monitoring system under Free Run EMG mode. The built-in ADC has a sampling rate of 25.6 kHz at a gain of 50 uV/div at 18-bit resolution. The detected EMG signal was time-locked to the applied stimulation pulses. Biostamp s-EMG signals were collected with a custom engineering app interface with a sampling rate of 250 Hz and gain of 12 at 16-bit resolution.
Signal processing
All needle and surface EMG signals were exported from the conventional intraoperative monitoring system without any data processing. Biostamp EMG signals were processed with a high pass 7th order Butterworth filter at 25 Hz. Time scales were manually aligned to synchronize the onset of nerve–muscle response from direct electrical stimulation. The SNR in dB scale is defined as 10 times log10(Variance2Signal/Variance2Noise) for a set period of time. All data were processed using Origin Pro software.
Mechanical modeling and finite element analysis (FEA)
Three-dimensional FEA simulations based on commercial software packages (ABAQUS) guided optimization of the mechanics of the system. The flexible circuit model, made of polyimide (PI, elastic modulus 2.5 GPa), 11.9-μm thick Cu (elastic modulus 119 GPa), and 25.4-μm thick adhesive (elastic modulus 931 MPa), had the cross-section from top to bottom of 25.4 μm PI/Cu/12.7 μm PI/adhesive/Cu/25.4 μm PI/Cu/adhesive/12.7 μm PI/Cu/38.1 μm PI. Together with the core/shell package, the flexible circuits were mounted on a phantom skin (elastic modulus 130 kPa).
Data deposition
All data generated or analyzed during this study are included in this published article (and its Supplementary Information files). All relevant data are available from the authors.
References
Simon, N. G., Spinner, R. J., Kline, D. G. & Kliot, M. Advances in the neurological and neurosurgical management of peripheral nerve trauma. J. Neurol. Neurosurg. Psychiatry 87, 198–208 (2015).
Dyck, P. J. Peripheral Neuropathy (Elsevier Inc., Philadelphia, PA, USA, 2005).
Kakazu, R., Dailey, S. K., Schroeder, A. J., Wyrick, J. D. & Archdeacon, M. T. Iatrogenic radial nerve palsy after humeral shaft nonunion repair: more common than you think. J. Orthop. Trauma 30, 256 (2016).
Koch, G. et al Anatomical risk evaluation of iatrogenic injury to the infrapatellar branch of the saphenousnerve during medial meniscus arthroscopic surgery. Surg. Radiol. Anat. 39, 611 (2016).
Weber, A. E., Harris, J. D. & Nho, S. J. Complications in hip arthroscopy: a systematic review and strategies for prevention. Sports Med. Arthrosc. 23, 187 (2015).
Kneist, W., Kauff, D., Juhre, V., Hoffmann, K. & Lang, H. Is intraoperative neuromonitoring associated with better functional outcome in patients undergoing open TME?: results of a case-control study. Eur. J. Surg. Oncol. 39, 994 (2013).
Falyar, C. R., Shaffer, K. M. & Perera, R. A. Localization of the brachial plexus: sonography versus anatomic landmarks. J. Clin. Ultrasound 44, 411 (2016).
Ajiboye, R. M. et al. Routine use of intraoperative neuromonitoring during ACDFs for the treatment of spondylotic myelopathy and radiculopathy is questionable: a review of 15,395 cases. Spine 42, 14 (2017).
Mian, A. et al. Brachial plexus anesthesia: a review of the relevant anatomy, complications, and anatomical variations. Clin. Anat. 27, 210 (2014).
Deletis, V. & Sala, F. Intraoperative neurophysiological monitoring of the spinal cord during spinal cord and spine surgery: a review focus on the corticospinal tracts. Clin. Neurophysiol. 119, 248 (2008).
Koht, A., Sloan, T.B., Toleikis, J.R. Monitoring the Nervous System for Anesthesiologists and Other Health Care Professionals (Springer, New York, NY, USA, 2012).
Møller, A. R. Intraoperative Neurophysiological Monitoring (Springer, Totowa, NJ, USA, 2006).
Hermann, M., Hellebart, C. & Freissmuth, M. Neuromonitoring in thyroid surgery: prospective evaluation of intraoperative electrophysiological responses for the prediction of recurrent laryngeal nerve injury. Ann. Surg. 240, 9 (2004).
Cypher, D., Chevrollier, N., Montavont, N. & Golmie, N. Prevailing over wires in healthcare environments: benefits and challenges. IEEE Commun. Mag. 44, 56 (2006).
Al-Shekhlee, A., Shapiro, B. E. & Preston, D. C. Iatrogenic complications and risks of nerve conduction studies and needle electromyography. Muscle Nerve 27, 517 (2003).
Lynch, S. L., Boon, A. J., Smith, J., Harper, C. M. & Tanaka, E. M. Complications of needle electromyography: hematoma risk and correlation with anticoagulation and antiplatelet therapy. Muscle Nerve 38, 1225 (2008).
Liu, Y. et al. Epidermal mechano-acoustic sensing electronics for cardiovascular diagnostics and human-machine interfaces. Sci. Adv. 2, e1601185 (2016).
Xu, S. et al. Soft microfluidic assemblies of sensors, circuits, and radios for the skin. Science 344, 70 (2014).
Kim, D.-H. et al. Epidermal electronics. Science 333, 838 (2011).
Lee, C. H. et al. Soft core/shell packages for stretchable electronics. Adv. Funct. Mater. 25, 3698 (2015).
A. Documentation. ABAQUS analysis user’s manual. Materials. Other plasticity models. Simulia 113, 23.3.1 (2010).
Riley, W. F. Mechanics of Materials (John Wiley & Sons, Inc., Indianapolis, IN, USA, 2006).
De Luca, C. J., Adam, A., Wotiz, R., Gilmore, L. D. & Nawab, S. H. Decomposition of surface EMG signals. J. Neurophysiol. 96, 1646 (2006).
Merletti, R., Knaflitz, M. & De Luca, C. J. Electrically evoked myoelectric signals. Crit. Rev. Biomed. Eng. 19, 293 (1992).
Rattay, F., Electrical Nerve Stimulation (Springer, Wien, 1990).
Gorman, P. H. & Mortimer, J. T. The effect of stimulus parameters on the recruitment characteristics of direct nerve stimulation. IEEE Trans. Biomed. Eng. 30, 407–414 (1983).
Keyl, C., Held, T., Albiez, G., Schmack, A. & Wiesenack, C. Increased electrical nerve stimulation threshold of the sciatic nerve in patients with diabetic foot gangrene: a prospective parallel cohort study. Eur. J. Anaesthesiol. 30, 435 (2013).
Tschopp, K. P. & Gottardo, C. Comparison of various methods of electromyographic monitoring of the recurrent laryngeal nerve in thyroid surgery. Ann. Otol. Rhinol. Laryngol. 111, 811 (2002).
Radtke, F. et al. Monitoring depth of anaesthesia in a randomized trial decreases the rate of postoperative delirium but not postoperative cognitive dysfunction. Br. J. Anaesth. 110, i98 (2013).
Sutter, M. et al. Use of multimodal intra-operative monitoring in averting nerve injury during complex hip surgery. J. Bone Joint Surg. Br. 94, 179 (2012).
Thilen, S. R. & Bhananker, S. M. Qualitative neuromuscular monitoring: how to optimize the use of a peripheral nerve stimulator to reduce the risk of residual neuromuscular blockade. Curr. Anesthesiol. Rep. 6, 164 (2016).
Kaltenbrunner, M. et al. An ultra-lightweight design for imperceptible plastic electronics. Nature 499, 458 (2013).
Webb, R. C. et al. Epidermal devices for noninvasive, precise, and continuous mapping of macrovascular and microvascular blood flow. Sci. Adv. 1, e1500701 (2015).
Dagdeviren, C. et al. Conformable amplified lead zirconate titanate sensors with enhanced piezoelectric response for cutaneous pressure monitoring. Nat. Commun. 5, 4496 (2014).
Schwartz, G. et al. Flexible polymer transistors with high pressure sensitivity for application in electronic skin and health monitoring. Nat. Commun. 4, 1859 (2013).
Webb, R. C. et al. Ultrathin conformal devices for precise and continuous thermal characterization of human skin. Nat. Mater. 12, 938 (2013).
Huang, X. et al. Stretchable, wireless sensors and functional substrates for epidermal characterization of sweat. Small 10, 3083 (2014).
Krishnan, S. et al. Multimodal epidermal devices for hydration monitoring. Microsyst. Nanoeng. 3, 17014 (2017).
Dagdeviren, C. et al. Conformal piezoelectric systems for clinical and experimental characterization of soft tissue biomechanics. Nat. Mater. 14, 728 (2015).
Huang, X. et al. Materials and designs for wireless epidermal sensors of hydration and strain. Adv. Funct. Mater. 24, 3846 (2014).
Boutry, C. M. et al. A sensitive and biodegradable pressure sensor array for cardiovascular monitoring. Adv. Mater. 27, 6954 (2015).
Capogrosso, M. et al. A brain–spine interface alleviating gait deficits after spinal cord injury in primates. Nature 539, 284 (2016).
Gao, W. et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 529, 509 (2016).
Khodagholy, D. et al. NeuroGrid: recording action potentials from the surface of the brain. Nat. Neurosci. 18, 310 (2015).
Lochner, C. M., Khan, Y., Pierre, A. & Arias, A. C. All-organic optoelectronic sensor for pulse oximetry. Nat. Commun. 5, 5745 (2014).
Yokota, T. et al. Ultraflexible organic photonic skin. Sci. Adv. 2, e1501856 (2016).
Acknowledgements
We thank the Neurosurgery Medicine at Northwestern Memorial Hospital for conducting clinical patient studies and providing intraoperative monitoring support for this study. Y.L. and R.G. acknowledge support from the Center for Bio-Integrated Electronics at Northwestern University. L.T. acknowledges the support from Beckman Institute Postdoctoral Fellowship at UIUC. Y.M. acknowledges support from the National Natural Science Foundation of China (Grant No. 11402135). Y.H. acknowledges support from NSF (Grant Nos. DMR-1121262, CMMI-1300846, CMMI-1400169, and CMMI-1534120).
Author information
Authors and Affiliations
Contributions
Conception, design, and study direction: Y.L., L.T., M.S.R., R.G., M.K., and J.A.R. Clinical studies: Y.L., M.C., N.D., O.B., T.K., L.T., M.K., and J.A.R. Data analysis: Y.L., R.G., G.S., M.K., Y.M., M.S.R., A.J., A.J.A., J.A.W., L.O., B.M., and J.A.R. Mechanical simulation: Y.M., Y.Z., and Y.H. Manuscript writing: Y.L., A.P., R.G., G.S., L.T., M.S.R., A.J.A., M.K., Y.H., and J.A.R.
Corresponding authors
Ethics declarations
Competing interests
J.A.R. and R.G. hold equity in the company MC10 that makes related wearable devices for medical applications. The remaining authors declare no competing financial interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Liu, Y., Tian, L., Raj, M.S. et al. Intraoperative monitoring of neuromuscular function with soft, skin-mounted wireless devices. npj Digital Med 1, 19 (2018). https://doi.org/10.1038/s41746-018-0023-7
Received:
Revised:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41746-018-0023-7
This article is cited by
-
Shaping high-performance wearable robots for human motor and sensory reconstruction and enhancement
Nature Communications (2024)
-
Wireless facial biosensing system for monitoring facial palsy with flexible microneedle electrode arrays
npj Digital Medicine (2024)
-
Soft and stretchable organic bioelectronics for continuous intraoperative neurophysiological monitoring during microsurgery
Nature Biomedical Engineering (2023)
-
Feasibility of Passive ECG Bio-sensing and EMA Emotion Reporting Technologies and Acceptability of Just-in-Time Content in a Well-being Intervention, Considerations for Scalability and Improved Uptake
Affective Science (2022)