Abstract
Using anatomy-based fitting, we can determine the place-specific map with individualized center frequencies for each electrode contact that is a closer match to the natural pitch-place of the cochlea. The primary objective of this study is to evaluate the tonal presentation across the electrode array and to calculate the semitone difference between each adjacent pair of contacts according to their anatomy-based map. The secondary objective is to determine the distancing of the contacts that would result in an equal semitone difference with a uniform tonal presentation. A total of 167 ears were included in this retrospective study. The frequencies across the electrode arrays were found to be unequally presented. The semitonal condensations were higher in the apical inter-contact spaces compared to the basal inter-contact spaces, being 3.0–2.3 semitones/mm (Kruskal Wallis test, p < 0.000). The anatomy-based spacing of the electrode contacts was larger in the basal inter-contact spaces compared to the apical inter-contact spaces, ranging from 1.92 to 1.48 mm. In conclusion, the current electrode designs do not have uniform tonal representation throughout the electrode array. There is a more condensed tonal presentation in the apical electrodes than in the basal electrodes, resulting in a lower tonal resolution in the apical region.
Similar content being viewed by others
Introduction
Cochlear implantation is the mainstay treatment for patients with sensorineural hearing loss who do not benefit from hearing aids1. Since the first cochlear implantation surgery, there has been a continuous effort to improve the design and function of the device2,3, and the satisfaction of the cochlear implant (CI) patients4.
CIs are designed to replace the normal cochlea, particularly with hearing preservation arrays—they bypass the damaged or dysfunctional areas1,5. Therefore, to better understand and improve the designs of the CI, there is a need to understand how a normal cochlea function. The normal cochlea has a tonotopic representation of the frequencies, making each part most sensitive to a specific frequency, which is called its characteristic frequency6. The tonotopicity of the cochlea makes it more sensitive to higher frequencies in the basal region and lower frequencies in the apical region7.
The programming of CIs includes assigning specific frequency range to each electrode contact. This programming map can either be assigned using a default frequency map or, in the recent practice, to be personalized according to the patient’s cochlear anatomy8. The latter is known as anatomy-based frequency mapping9.
In default mapping, the frequencies assigned to each electrode contact follow a frequency table with fixed frequencies for all patients. Although these frequencies were based on the Greenwood equation to follow the tonotopicity of the cochlea, this default map is not personalized. Recent studies have argued that using default frequency mapping for speech processor programming results in a mismatch between the stimulating frequency presented by some of the electrode contacts and the characteristic frequency that corresponds to the real place of those contacts in the cochlea10. There are three reasons for the occurrence and degree of mismatch. First, there are diverse ranges of cochlear duct lengths11,12. Second, electrode lengths differ from one array type to the other13. Third, electrode contacts can be positioned in a different site of the cochlea depending to the degree of surgical insertion14,15.
To date, all electrode arrays have constant, and algebraically equal, spacing between their contacts. This method of spacing would results in a non-homogenous presentation of frequencies along the electrode array, in terms of the number of semitones. We assume that similar semitone differences would result in more uniform hearing representation. However, due to the logarithmic nature of the frequency differences, the electrode spacing should not be algebraically equally divided. To enable similar tonal resolution, electrode arrays need to have a smaller electrode spacing in the apical region compared to the basal region.
The primary objective of this study is to evaluate the condensation of the tonal presentation across the electrode array and to calculate the semitone differences between adjacent pairs of electrodes according to their anatomy-based frequency map. The secondary objective is to determine the spacing of the electrode contacts that results in an equal semitone difference with uniform tonal presentation.
Methodology
The medical data of patients in the CI registry who underwent cochlear implantation at a tertiary referral center for neuro-otology and CI surgery were retrospectively reviewed. Before conducting this study, the approval of the Institutional Review Board of King Saud University Hospital was obtained (Ref. No. 20/0091/IRB, Project No. E-20-5387).
Patient selection
The preoperative computed tomography (CT) scans of 167 ears of 100 patients who underwent cochlear implantation at King Abdullah Ear Specialist Center were reviewed. No limitation was set for the chosen subjects regarding their age, sex, or number of implantations. All implanted electrodes that were compatible with the software used in this study were included. Patients with inner ear anomalies, cochlear ossification, cochlear otosclerosis, an otic capsule involving temporal bone fractures, or any bone diseases involving the temporal bone were excluded.
Imaging
Images were obtained using a 512-slice multidetector-row CT scanner (General Electric Healthcare, Milwaukee, WI). The following scanning parameters were used: axial plane, 0.625 mm slice thickness, 230 mAs, 140 kV, rotation time 1 s, with 0.3 mm reconstruction in the axial and coronal views.
Anatomy-based prediction of frequency allocation for each electrode
The preoperative CT images were uploaded to the Otoplan V.03 DICOM viewer and surgical planning software (CAScination AG, Bern, Switzerland). This software predicts the personalized, anatomy-based frequency allocation of each electrode according to the imaging-based cochlear duct length (CDL) and the length of the electrode array.
Semitonal differences were defined as the differences between the anatomy-based frequency of two adjacent electrode contacts and were measured in semitones.
The inter-contact spacing refers to the physical distance between the two adjacent electrode contacts.
Semitone condensation was calculated by dividing the semitone differences between adjacent electrode contacts by their spacing distance. Tonal resolution was defined as 1 over semitone condensation (Fig. 1).
Anatomy-based electrode spacing was determined by dividing the frequency range of the anatomy-based array frequency coverage into equal semitonal distances. After determining the frequencies that would result in equal semitonal distances, the corrected Greenwood equation was used to determine the spacing of each pair of adjacent electrodes16,17,18,19.
In this study all the included electrode types were from the same manufacturer, Med El (Innsbruck, Austria), that is compatible with the Otoplan software and has twelve contacts in each electrode array type.
The total lengths of the electrode arrays were 24, 28, and 31 mm for the Form-24, Flex-28, and Standard types of electrode arrays, respectively. These electrodes have an active stimulating range of 18.7, 23.1 and 26.4 mm for the Form-24, Flex-28, and Standard types of electrode arrays, respectively. The inter-contact spacing of these electrode arrays is 1.7 mm, 2.1 mm, and 2.4 mm for the Form-24, Flex-28, and Standard types of electrode arrays, respectively.
In Med El devices, discussed in this article, the most apical electrode is referred to as the first while the most basal one is the twelfth one.
Data collection and analysis
Data collection was performed using the Excel software (version 16.3; Microsoft, Seattle, WA). The statistical analyses were performed using IBM SPSS software (version 23.0; IBM Corp., Armonk, NY). Statistical significance was set at p < 0.05. The inter-rater reliability was evaluated using Cronbach’s alpha. The means of the measurements taken by the two reviewers were used for further analyses.
Results
A total of 167 implanted ears in 100 patients (60 males, 40 females) were included. Overall, 105 (62.8%) ears received the Form-24 electrode array, 53 (31.7%) received the Flex-28, and nine (5.3%) received the Standard electrodes.
It was found that the two reviewers' electrode frequency readings were homogenous and had a high level of internal consistency (Cronbach's alpha = 0.96).
The semitone differences between the two most apical electrode contacts were the lowest in the Form-24 electrode array type. The semitone differences in the Form-24 electrodes ranged from 4.9 semitones in the most apical contacts to 4.1 semitones in the last two basal electrodes, 6.7–4.8 semitones in the Flex-28, and 8.9–5.4 semitones in the Standard electrode array types. The semitone differences were significantly higher in the apical spaces and decreased towards the most basal inter-contact spaces (Kruskal Wallis test, Chi-Square value from 108 to 111, p < 0.000) (Supplementary File 1, Fig. 2).
The longest electrode array type, the standard electrode, demonstrated the highest range of tonal representation followed by the shorter Flex-28, and Form-24 (Fig. 3).
The semitonal condensation was higher in the apical electrodes and decreased towards the basal electrode contact spaces, starting from 3.05 semitone/mm at the first apical space, and decreasing to 2.3 semitone/mm at the last basal space. The semitonal condensation of the inter-contact spaces along the array was found to be significantly different (Kruskal Wallis test, Chi-Square value 823.1, p < 0.000). The longer, Standard, electrode array type demonstrated the highest semitonal condensation followed by the Flex-28 and lastly Form-24 in the apical region. This relation was reversed in the basal region, with the Standard electrode demonstrating the lowest semitonal condensation (2.25 semitone/mm) compared to the Flex-28 (2.3 semitone/mm) and the highest condensation was seen in Form-24 (2.4 semitone/mm). The difference in the semitonal condensation between the three electrode array types was found to be significant (Kruskal Wallis test, Chi-Square value 33.1, p < 0.000) (Fig. 4).
The anatomy-based inter-contact spacing revealed that for an equal inter-contact semitonal condensation there is a need for a 1.48 mm inter-contact space in the most apical region and this distance needs to increase to 1.92 mm in the most basal region of the electrode array. The anatomy-based inter-contact spacing were significantly different across the electrode arrays (Kruskal Wallis test, Chi-Square value from 88 to 119, p < 0.000) (Table 1, Fig. 5).
Two-way ANOVA showed that both the type of electrode array and the inter-contact space number are significantly affecting the estimated spacing for equal semitonal differences. (F value 36,277 & 2738, partial Eta squared showing an effect size of 0.97 & 0.93 respectively and p < 0.001 for both) The interaction of the two factors also showed to be significantly affecting the estimated spacing. (F = 9.04, partial Eta squared showing an effect size of 0.80 with p < 0.001) It must be mentioned that the Levene’s test shows that the variances are not homogenous. (F value of 35.86, p < 0.001).
Discussion
This study demonstrated the variability in the tonal representation across the electrode, using anatomy-based frequency mapping. Apical electrode spaces demonstrated greater tonal condensation with higher ranges of semitone differences between adjacent electrode contacts compared to basal electrode spaces.
Significant improvements in the understanding of cochlear anatomy have been achieved following the development of the synchrotron radiation phase-contrast imaging technique7. Using the findings of previous studies, corrections have been made to the equations by which anatomy-based spacings are determined7. In this study, OC map was used to determine the anatomy-based frequency allocations. It must be mentioned that uncertainty still exists on where along the signal pathway, the cochlear stimulation happens. For the lateral wall electrode arrays that were included in this study, the OC is assumed to be the first site of stimulation by the electrical signal due to the more proximity of the OC to the laterally placed array20.
Previous studies have found that the frequency-to-place mismatch differs across the electrode arrays and is greater in apical electrode contacts21. Having perfect frequency-to-place matching would result in worse tonal resolution in the apical region, as demonstrated by the current results. Of the included electrode designs, the highest anatomy-based frequency resolution in the basal spaces was found in the standard electrodes, and the lowest anatomy-based frequency resolution in the apical spaces was also found in the standard electrodes. In the apical region, semitonal condensation was highest in the standard electrodes, followed by the Flex-28 and Form-24 arrays; however, this order was reversed in the basal electrode region. This phenomenon can be explained by the deeper insertion of the Standard electrode causing a lower frequency representation in a more sensitive apical regions of the cochlea that has a condensed tonal representation. While in the more basal area the condensation is more affected by the electrode spacing which is higher in the Standard electrode causing a lower condensation.
The anatomy-based inter-contact spacing findings demonstrated that more space is needed in the basal electrode contacts compared to the apical spaces to achieve equal and uniform inter-contact semitonal resolution across the electrode array (Fig. 6). One key consideration is that the magnitude of anatomy-based electrode spacing is dependent on personalized cochlear duct length. Interestingly, there was small variability in the anatomy-based spacing within each CI electrode array type. This creates the opportunity to develop a new electrode design with constant inter-contact spacing for use across a wide range of cochlear duct lengths. It must be kept in mind however that although these variabilities are small on millimetric spacing scale, they are actually not that small, when you transfer them back onto a semitonal distances. Which means that although using this new spacing would improve the tonal representation, but it still would lack a perfect uniform tonal spacing due to the individual differences in the CDL.
As the anatomy-based frequency allocations of the electrode contacts were used to calculate the inter-contact spacing, the total active stimulating length of the electrode with the anatomy-based spacing was assumed to be constant.
Although we included only one manufacturer’s electrode arrays in this study, the spacing method suggested in this study can be applied to the other manufacturers’ arrays as well. The calculations must consider the number of the contacts and the active stimulating length of the array. Currently there is no electrode array design that has an anatomy-based spacing of the contacts and all contacts are spaced equally in their physical distance rather than the tonal distances of the cochlea. Which means all the arrays used have an ununiform tonal representation (Fig. 7).
A limitation of this study is that the clinical significance of anatomy-based spacing needs to be further studied after electrode spacing is changed. It is also possible to study this spacing method using virtual electrode contact spacing without the need to change the electrode design22,23. Furthermore, the anatomy-based spacing were measured on the assumption of full surgical insertion, which is not always possible. The final configuration of the electrode inside the cochlea can also affected by the spatial geometry of the cochlea resulting in different localization of the contact than expected by preoperative imaging. Even with these limitations, generally concluding, the more apical electrode contacts would still need to be closer to each other compared to the basal electrode contacts to enable a more uniform frequency presentation.
Other factors that affect the location of the stimulation are the spread of excitation in monopolar stimulation mode, which can lead to off-site and cross-turn stimulation in areas which can limit any approach aiming to improve the frequency to place matching.
A number of strategies have aimed to solve the spread of excitation effect, including continuous interleaved sampling (CIS), n-of-m and advanced combination encoder, and precompression for spread of excitation strategies24,25,26,27.
Another matter of importance is the fact that an apically closer spacing of electrode contacts might result in signal interaction due to the spread of the excitation28. This can limit the benefit of these contacts. However, the spread of excitation is affected by numerous variables including the programming parameters that can be calibrated in a way to preserve contact functioning without signal interaction29,30.
Conclusion
Uniform algebraic electrode contacts spacing results in an unequal tonal resolution across electrode arrays, with better resolution in the basal compared to the apical inter-electrode spaces. The anatomy-based electrode contact spacing would require wider spacing in the basal electrode contacts and narrower spacing in the apical electrode contacts. The small variability of the anatomy-based contact spacing despite the variability in patient cochlear duct lengths would enable the potential development of a constant electrode contact spacing that would fit most patients.
Data availability
The data used in this study are available upon requests send to the corresponding author’s email: dr.isra.aljazeeri@gmail.com.
References
Lenarz, T. Cochlear implant: State of the art. GMS Curr. Top. Otorhinolaryngol. Head Neck Surg. 16, 1. https://doi.org/10.3205/cto000143 (2018).
Mudry, A. & Mills, M. The early history of the cochlear implant: A retrospective. JAMA Otolaryngol. Head Neck Surg. 139, 446–453. https://doi.org/10.1001/jamaoto.2013.293 (2013).
Wilson, B. S. & Dorman, M. F. Cochlear implants: A remarkable past and a brilliant future. Hear Res. 242, 3–21. https://doi.org/10.1016/j.heares.2008.06.005 (2008).
Liebscher, T. et al. Electrode translocations in perimodiolar cochlear implant electrodes: Audiological and electrophysiological outcome. Z. Med. Phys. 31, 265–327. https://doi.org/10.1016/j.zemedi.2020.05.004 (2021).
Wilson, B. S. & Dorman, M. F. Cochlear implants: Current designs and future possibilities. J. Rehabil. Res. Dev. 45, 695–730. https://doi.org/10.1682/jrrd.2007.10.0173 (2008).
Olson, E. S., Duifhuis, H. & Steele, C. R. V. Békésy and cochlear mechanics. Hear Res. 293, 31–43. https://doi.org/10.1016/j.heares.2012.04.017 (2012).
Li, H. et al. Three-dimensional tonotopic mapping of the human cochlea based on synchrotron radiation phase-contrast imaging. Sci. Rep. 11, 4437. https://doi.org/10.1038/s41598-021-83225-w (2021).
Fu, Q. J. & Shannon, R. V. Frequency mapping in cochlear implants. Ear Hear. 23, 339–348. https://doi.org/10.1097/00003446-200208000-00009 (2002).
Helpard, L. et al. An approach for individualized cochlear frequency mapping determined from 3D synchrotron radiation phase-contrast imaging. IEEE Trans. Biomed. Eng. 68, 3602–3611. https://doi.org/10.1109/TBME.2021.3080116 (2021).
Kalkman, R. K., Briaire, J. J., Dekker, D. M. & Frijns, J. H. Place pitch versus electrode location in a realistic computational model of the implanted human cochlea. Hear Res. 315, 10–24. https://doi.org/10.1016/j.heares.2014.06.003 (2014).
Alanazi, A. & Alzhrani, F. Comparison of cochlear duct length between the Saudi and non-Saudi populations. Ann. Saudi Med. 38, 125–129. https://doi.org/10.5144/0256-4947.2018.125 (2018).
Spiegel, J. L. et al. Variation of the cochlear anatomy and cochlea duct length: Analysis with a new tablet-based software. Eur. Arch. Otorhinolaryngol. 279, 1851–1861. https://doi.org/10.1007/s00405-021-06889-0 (2022).
Dhanasingh, A. & Jolly, C. An overview of cochlear implant electrode array designs. Hear Res. 356, 93–103. https://doi.org/10.1016/j.heares.2017.10.005 (2017).
Landsberger, D. M., Svrakic, M., Roland, J. T. Jr. & Svirsky, M. The relationship between insertion angles, default frequency allocations, and spiral ganglion place pitch in cochlear implants. Ear Hear. 36, e207–e213 (2015).
Canfarotta, M. W. et al. Validating a new tablet-based tool in the determination of cochlear implant angular insertion depth. Otol. Neurotol. 40, 1006–1010. https://doi.org/10.1097/MAO.0000000000002296 (2019).
Greenwood, D. D. Critical bandwidth and the frequency coordinates of the basilar membrane. J. Acoust. Soc. Am. 33, 1344–1356 (1961).
Greenwood, D. D. A cochlear frequency-position function for several species—29 years later. J. Acoust. Soc. Am. 87, 2592–2605 (1990).
Greenwood, D. D. Critical bandwidth and consonance in relation to cochlear frequency-position coordinates. Hear. Res. 54, 164–208 (1991).
Paouris, D., Kunzo, S. & Goljerová, I. Validation of automatic cochlear measurements using OTOPLAN® software. J. Pers. Med. 13(5), 805 (2023).
Landry, T. G., Wise, A. K., Fallon, J. B. & Shepherd, R. K. Spiral ganglion neuron survival and function in the deafened cochlea following chronic neurotrophic treatment. Hear. Res. 282(1–2), 303–313 (2011).
Aljazeeri, I. et al. Anatomy-based frequency allocation in cochlear implantation: The importance of cochlear coverage. Laryngoscope. 132, 2224–2231. https://doi.org/10.1002/lary.30004 (2022).
Hughes, M. L., Stille, L. J., Baudhuin, J. L. & Goehring, J. L. ECAP spread of excitation with virtual channels and physical electrodes. Hear. Res. 306, 93–103. https://doi.org/10.1016/j.heares.2013.09.014 (2013).
Padilla, M., Stupak, N. & Landsberger, D. M. Pitch ranking with different virtual channel configurations in electrical hearing. Hear. Res. 348, 54–62. https://doi.org/10.1016/j.heares.2017.02.013 (2017).
Wilson, B. S. et al. Better speech recognition with cochlear implants. Nature 352(6332), 236–238 (1991).
Vandali, A. E., Whitford, L. A., Plant, K. L. & Clark, G. M. Speech perception as a function of electrical stimulation rate: Using the Nucleus 24 cochlear implant system. Ear. Hear. 21(6), 608–624 (2000).
Ziese, M. et al. Speech understanding with the CIS and the n-of-m strategy in the MED-EL COMBI 40+ system. ORL J. Otorhinolaryngol. Relat. Spec. 62, 321–329 (2000).
Bolner, F., Magits, S., van Dijk, B. & Wouters, J. Precompensating for spread of excitation in a cochlear implant coding strategy. Hear. Res. 395, 107977 (2020).
Rader, T., Nachtigäller, P., Linke, T., Weißgerber, T. & Baumann, U. Exponential fitting of spread of excitation response measurements in cochlear implants. J. Neurosci. Methods 391, 109854. https://doi.org/10.1016/j.jneumeth.2023.109854 (2023).
Biesheuvel, J. D., Briaire, J. J., Kalkman, R. K. & Frijns, J. H. M. The effect of stimulus level on excitation patterns of individual electrode contacts in cochlear implants. Hear. Res. 420, 108490. https://doi.org/10.1016/j.heares.2022.108490 (2022).
Hughes, M. L. & Stille, L. J. Effect of stimulus and recording parameters on spatial spread of excitation and masking patterns obtained with the electrically evoked compound action potential in cochlear implants. Ear Hear. 31(5), 679–692. https://doi.org/10.1097/AUD.0b013e3181e1d19e (2010).
Acknowledgements
This study was supported by the Deanship of Science Research, King Saud University, Riyadh, Saudi Arabia. It is one of the recommended ideas for research projects by Saudi ORL. The authors also acknowledge the work of Dr.Nezar Hamed and Mr.Taha Sharif in collecting the data for this manuscript.
Funding
This project received no fundings.
Author information
Authors and Affiliations
Contributions
I.A. conceptualization, design, data analysis and writing the manuscript. A.H. supervision and substantively revised the manuscript. All methods were carried out in accordance with relevant guidelines and regulations. Informed consent was obtained from all subjects and/or their legal guardians.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Aljazeeri, I.A., Hagr, A. Application of anatomy-based spacing of electrode contacts for achieving a uniform semitonal resolution: A novel concept in cochlear implant electrode design. Sci Rep 14, 2645 (2024). https://doi.org/10.1038/s41598-024-53070-8
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-024-53070-8
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.