Introduction

Radiological imaging has become increasingly important for diagnosis, clinical assessment, and treatment planning1. In recent years, technology has advanced from two-dimensional conventional radiographs to three-dimensional digital imaging such as computed tomography (CT), magnetic resonance imaging (MRI), and cone-beam computed tomography (CBCT)2.

CBCT is now an important part of diagnosis and treatment in fields like maxillofacial surgery, orthodontics, endodontics, implant planning, and evaluating patients after treatment3,4,5,6,7. Nevertheless, the presence of metals or alloys in implants, posts, cores, crowns, bridges, and amalgam fillings decreases image quality in the area adjacent to them due to reduced contrast caused by beam hardening, scattering effect, partial volume or edge effect, aliasing artifacts, and ring artifacts8,9,10. Due to increased life expectancy and the consequent increase in partial or complete edentulism, implant-supported prostheses have also grown in popularity10.

In their studies, Torres11 and Sheikhi12 found the physical linear measurements to be more than radiographic CBCT measurements. Linear measurements of alveolar bone and maxillofacial structures by CBCT were proven accurate, trustworthy, and independent of the presence of dental implants by Ekrish13 and Amarnath14. Beam hardening artifacts were observed in the presence of metal implants in studies conducted by Schulze15 and Dreanert16.

The current study sought to evaluate the accuracy of CBCT as a measuring tool and quantify the artifact, if any, in the presence of implants and implant-supported prostheses by measuring and comparing linear distances in transverse, sagittal, and vertical planes in the physical model to radiographic measurements, as well as evaluating metal artifacts produced by CBCT of the mandible with or without implants or prostheses.

The null hypothesis stated that the accuracy of CBCT images is unaffected by the presence of implants and implant-supported prostheses.

Materials and methods

The research was conducted at the RUHS College of Dental Sciences in Jaipur, in the Department of Prosthodontics. The study received ethical committee clearance no. EC/PG-015/2019. The study used dry edentulous mandible and endosteal tapered SLA-treated (4 × 10 mm) titanium implants with titanium abutments (Superline Dentium implant system; Dentium South Korea). For assessment of accuracy, two points were marked on the buccal cortical plate of the body of a dried edentulous mandible, corresponding to the right mandibular canine (A and D) and mandibular first molar (C and F), both near the crest and base, respectively. Third point B and E were subsequently marked between points A and C and D and F respectively. Similarly, points a to f, corresponding to points A to F on the buccal cortical plate, were marked on the lingual cortical plate. This was referred to as ‘Absolute Control (C0)’ and was used for physical linear measurement (PLM). The radio-opaque gutta-percha balls of 0.5 mm diameter were then affixed using cyanoacrylate adhesive through points A to F and a to f on the buccal and lingual plates, respectively. This was referred to as radiographic control (C1), for radiographic linear measurements (RLM) in different spatial configurations. These reference points provide a standardized and reproducible framework for measuring and comparing PLM with RLM. The specific locations of the points were chosen to provide a reference for measurements in the transverse, sagittal, and vertical planes, which would be important for assessing the accuracy of CBCT images in the study (Figs. 1, 2).

Figure 1
figure 1

Flow-chart depicting the study design for assessment of dimensional accuracy and artifacts in CBCT at different stages. C0 = Absolute Control, C1 = Radiographic Control, TG1 = Single Implant placed at 43, TG2 = Implant-supported Porcelain fused to metal (PFM) crown with the abutment at 43, TG3 = Two Implants placed at 43 and 46, TG4 = Two Implant-supported PFM crowns at 43 and 46, TG5 = Implant-supported four-unit PFM prosthesis from 43 to 46, TG6 = Implant-supported four-unit PFM prostheses from 43 to 46 and 33 to 36 (Implants placed at 43, 46, 33 and 36), TG7 = Full arch implant-supported prosthesis.

Figure 2
figure 2

Reference points for accuracy measurement. (a) Buccal reference points for accuracy measurement at the crest of the mandible (A, B, C) and the base of the mandible (D, E, F). (b) Lingual reference points for accuracy measurement at the crest of the mandible (a, b, c) and the base of the mandible (d, e, f).

Furthermore, seven test groups (TG1 to TG7) were created (Fig. 3). A styrofoam customized base was used to standardize the position of the mandible, with the base parallel to the horizontal plane and centered within the CBCT field of view (FOV). CBCT images were acquired using a CBCT machine (Care Stream CS3D-9000 machine; Care Stream Health Inc). The image volume was reconstructed with isotropic isometric 250 × 250 × 250 μm voxels after receiving a scout image to include the region of interest.

Figure 3
figure 3

Stages of study model. (a) 4 × 10 mm titanium implant placed at 43 (TG1). (b) Implant-supported PFM crown with the abutment at 43 (TG2). (c) 4 × 10 mm titanium implants placed in the study model at 43 and 46 (TG3). (d) Implant-supported PFM single crowns at 43 and 46 (TG4). (e) Implant-supported four-unit PFM prosthesis from 43 to 46 (TG5). (f) Implant-supported four-unit PFM prostheses from 43 to 46 and 33 to 36 with 4 × 10 mm titanium implants placed in canine and first molar region bilaterally at 33, 43, 36, and 46 (TG6). (g) Full-arch implant-supported prosthesis (TG7).

Three independent observers performed PLM with a digital vernier caliper with a sensitivity of 0.01 mm. To determine dimensional accuracy, PLM was compared to RLM in all three planes: transverse, sagittal, and vertical.

RLM were performed in multi-planar oblique CBCT sections using the CBCT software measurement tool (Carestream 3D imaging, version 3.10.8; Carestream Health, Inc). RLM was performed on radiographic points in the transverse, sagittal, and vertical planes corresponding to physical points.

The artifacts were examined at locations A, B, a, and b in the superior axial plane. The images were integrated at a scale of 0.25 mm, 10.8 mm, and 21.8 mm and saved using the CBCT workplace screenshot tool. The screenshot was transformed to a 16-bit grayscale image using image software (The image J program, version 1.530; National Institute of Health and the Laboratory for Optical and Computational Instrumentation (LOCI)—University of Wisconsin) (grayscale range 0–255). Concentric circles with diameters of 6 mm, 10 mm, 15 mm, and 20 mm were drawn around the center of the implant at the canine and molar regions to define the region of interest. A reference line (Ro) was drawn across the center of the implant and parallel to the mandible's buccal and lingual cortical plates. Intersecting lines were made with respect to Ro at 0°, 65°, 90°, 115°,180°, 245°, 270°, and 295°, intersecting the concentric circles to yield points of interest (Fig. 4). Macro (macro record tool of Image-J software version 1.530, National Institute of Health and the Laboratory for Optical and Computational Instrumentation (LOCI), University of Wisconsin) was used to delineate 32, 1-mm square regions of interest (ROI) for both implants placed at 43 and 46. Finally, the recorded Macro was run on an unmarked image to capture the average grayscale values of the square ROI mentioned previously (Figs. 5, 6). The artifacts in CBCT were recorded in the study at C1 and TG1 through TG7. The artifacts of the superior axial plane were then compared to those of the inferior axial plane (which corresponded to the appearance of the superior border of the mental foramen).

Figure 4
figure 4

CBCT image depicting the region of interest for implant at canine and molar region.

Figure 5
figure 5

CBCT image depicting points ROI for the implant. (a) At canine (1–32). (b) At canine and molar region (1–64).

Figure 6
figure 6

3-D CBCT image showing artifacts with implants at 33, 36, 43, and 46 and full arch prosthesis. (a) Occlusal view. (b) Frontal view.

For osteotomy, conventional implant drilling protocols were followed for the Dentium surgical kit (SuperLine & Implantium Surgical Kit; Dentium). All Porcelain fused to metal (PFM) prostheses were fabricated by the lost wax technique using a base metal casting alloy.

The data obtained was compiled on an MS Office Excel sheet (v. 2016). The Statistical Package for the Social Sciences (SPSS) version 21.0 (IBM) was used to analyze the data. A one-way ANOVA test was used for intergroup comparison. The intra-group comparison was made using the post-hoc Tukey test. For all the statistical tests, p < 0.05 was considered statistically significant, keeping α error at 5% and giving 95% power to the study.

Results

The dimensional accuracy of CBCT as a measuring instrument was determined by comparing the RLM of C1 with the PLM (C0) in the transverse, sagittal, and vertical planes. The observed p-values after the paired t-test were statistically non-significant (p > 0.05) in all three planes.

In the transverse plane, the PLM (C0) was 9.23 mm, 10.77 mm, and 10.63 mm for A-a, B-b, and C–c in the superior plane, while the minimum and maximum RLM for test groups observed were 9.37 mm (TG1) and 9.63 mm (TG7), 10.57 mm (TG1) and 10.93 mm (TG6), and 9.37 mm (TG4 and TG5) and 10.73 mm (TG6) for A-a, B-b, and C-c, respectively. Similarly, PLM(C0) for D-d, E-e, and F-f in the inferior plane was 8.30 mm, 8.12 mm, and 7.19 mm, respectively, as compared to the observed minimum and maximum RLM values of 9.37 mm (TG1, TG4, and TG5) and 8.37 mm (TG7), 7.83 mm (TG5) and 8.03 mm (TG1 and TG7), 6.83 mm (TG7) and 7.13 mm (TG4 and TG5) for D-d, E-e, F-f thus, showing statistically significant results in the superior axial plane (Table 1).

Table 1 Comparative assessment in transverse plane: A one-way ANOVA comparison followed by post-hoc Tukey HSD test.

The sagittal plane PLM for A-B, B-C, D-E, and E-F was 15.47 mm, 12.25 mm, 15.72 mm, and 10.83 mm, respectively, with minimum and maximum test group RLM values of 15.53 mm (TG4, TG6, and TG7) and 15.93 mm (TG5), 12.13 mm (TG4) and 12.27 mm (TG2), 16.03 mm (TG6) and 16.27 mm (TG3), 10.83 mm (TG6 and TG7) and10.97 mm (TG1) for A-B, B-C, D-E, E-F. PLM and test RLM at A-B, B-C, and D-E in the sagittal plane were statistically significant (p < 0.01) (Table 2).

Table 2 Comparative assessment in the sagittal plane: using one-way ANOVA test.

The vertical plane PLM (C0) for A-D, B-E, and C-F was 15.43 mm, 14.36 mm, and 15.35 mm, respectively, with a minimum and maximum value of test group RLM observed of 15.23 mm (TG1) and 15.63 mm (TG4, TG5, and TG7), 13.37 mm (TG6) and 14.33 mm (TG2, TG4, and TG5), 14.93 mm (TG4 and TG5), and 15.47 mm (TG1), respectively, for A-D, B-E, and C-F. There was a statistically significant difference between PLM and RLM at A-D, B-E, and C-F (p < 0.01) (Table 3). An intra-group analysis of test CBCT also revealed statistically significant differences across groups.

Table 3 Comparative assessment in Vertical plane: using one-way ANOVA test.

Quantification of artifacts of CBCT in the presence of implants or prostheses showed that with an increase in the number of implants or prostheses, the mean grayscale values increased. The highest mean grayscale value was seen in full-arch prostheses (170 ± 90.90), while the lowest grayscale value was seen in the single implant with prosthesis (98.03 ± 84.47) (Table 4). The mean grayscale change was greatest in the vicinity of the implant and implant prosthesis, with the greatest (64.64 78.097) at 3 mm and the least (8.63 81.269) at 10 mm (Table 5).

Table 4 Comparison of grayscale values of the test group.
Table 5 Mean grayscale change at various distances from the center of implant: a comparison using one-way ANOVA followed by post-hoc Tukey HSD test.

The canine and molar implant artifacts followed geometric distribution patterns. The grayscale changes at a 3 mm distance from the center of the canine implant (1 mm from the surface) had the highest values at 90° (94.33 ± 38.43) (lingual) and 270° (98.62 ± 41.722) buccally. Similarly, the highest grayscale values were observed around the molar implant at 90° (87.33 ± 49.808) and 270° (122.19 ± 74.682). In the region between canine and molar, reduced grayscale values were observed, i.e., at canine 180° (− 17.14 ± 62.041) and molar 0° (− 26.10 ± 61.173) due to the influence of implant or prosthesis (Table 6).

Table 6 Mean grayscale change at 3 mm distance at various angular regions of interest: a one-way ANOVA comparison followed by post-hoc Tukey HSD test.

The grayscale changes at a 5 mm distance from the center of the canine implant had the highest values at 90° and 270°. The highest grayscale values were observed around the molar implant at 115° and 270° (Table 7).

Table 7 Mean grayscale change at 5 mm distance at various angular regions of interest: A one-way ANOVA comparison followed by post-hoc Tukey HSD test.

The grayscale changes at a 7.5 mm distance from the center of the canine implant had the highest values at 115° and 270°. The highest grayscale values were observed around the molar implant at 90° and 270° (Table 8).

Table 8 Mean grayscale change at 7.5 mm distance at various angular regions of interest: A comparison using one-way ANOVA followed by post-hoc Tukey HSD test.

The grayscale changes at a 10 mm distance from the center of the canine implant had the highest values at 90° and 270°. The highest grayscale values were observed around the molar implant at 90° and 270° (Table 9).

Table 9 Mean grayscale change at 10 mm distance at various angular regions of interest: a one-way ANOVA comparison followed by post-hoc Tukey HSD test.

In the region between canine and molar, reduced grayscale values were observed, i.e., at canine 180° and molar 0° due to the influence of implant or prosthesis (Tables 6, 7, 8, 9).

Grayscale and change in grayscale(∆G) were evaluated at 0.25 mm,10 mm, and 20 mm Voxel integration scale (VIS). The grayscale value increased at a VIS (10 mm and 20 mm), and statistically significant (p < 0.01) results were observed in mean grayscale and grayscale change (∆G) at different VIS (Tables 10 and 11).

Table 10 Mean grayscale at different voxel integration scales (VIS): a one-way ANOVA comparison followed by post-hoc Tukey HSD test.
Table 11 Mean grayscale change(∆G) at different integration scales (VIS): A one-way ANOVA comparison followed by post-hoc Tukey HSD test.

Discussion

The increasing dental awareness has led to increased implant-supported prostheses in partially and completely edentulous patients. The presence of dental implants and prostheses causes artifacts such as beam hardening, scattering, etc. These artifacts also reduce imaging diagnostic accuracy, making diagnosing peri-implant features difficult17. CBCT is a reliable radiographic tool with accuracy in the 0.163 mm to 0.40 mm range, as shown in our study. The results are supported by Maroua AL18, Ganguly R19, and Menezes et al.20.

As we progressed from the single implant (TG-1) to full-arch implant-supported prosthesis (TG-7), the dimensional accuracy was reduced, confirming the co-relation between metal implants or implant-supported prostheses and the accuracy of CBCT image. The highly significant difference in observations can be explained by the role of artifacts induced by prostheses. The present study's findings prove that the artifacts increase as metal in implants and prostheses increases. An accurate delineation of buccal and lingual cortical plates in the peri-implant or prostheses area becomes difficult. The implant or prostheses produced artifacts strong enough to cause a dimensional error. This can be explained by the direction of the implant artifacts, which are diagonal to the implant axis, causing in-accuracies in transverse and sagittal planes making radiographic bone appearance wider than reality. This result is supported by Grobe A21, who, in an in-vivo study after placement of multiple implants, found a statistically highly significant (p value 0.003) difference between transverse measurements in CBCT and histological sections with a mean overestimation of 0.3 ± 0.04 mm in the CBCT image.

The TG RLM at the inferior axial plane showed no significant dimensional observation compared to C0. This can be explained by the reduced influence of prostheses as distance increases (at the inferior border). Studies have shown a sub-millimeter difference in CBCT measurements compared to the gold standard (caliper measurements)11,22,23,24,25. The present study evaluated and compared measurements in all three planes, endorsing the conclusion that CBCT is a reliable diagnostic tool for linear measurements. However, previous studies focused on one or two implant configurations across multiple models, but none looked at the impact of prostheses on accuracy. The present study found that as we progressed from TG1 to TG 7, the dimensional accuracy of CBCT was reduced. As metals in implants and prostheses increase, accurate delineation of buccal and lingual cortical plates in the peri-implant and implant-supported prostheses area becomes difficult. Also, the artifacts increase in a direction diagonal to the implant axis, causing accuracies in transverse and sagittal planes making bone appear wider than reality, causing the dimensional error21.

Beam hardening is the most significant metal-induced artifact of CBCT, causing grayscale change8,26. This is because lower energy X-ray beams from polychromatic X-rays are scattered and absorbed by denser metal and even lighter titanium. In comparison, higher energy beams pass through the surface/body of the metal object and are recorded as a high-energy beam by the detector8,27. Beam hardening causes metallic structure image distortion by differential absorption, known as a cupping artifact.

The lowest greyscale values observed in TG2 can be explained by the cupping effect, which is less with an 'only' implant (TG1) than for an implant with a prosthesis (TG2) (Table 10). The current study showed differential absorption of x-rays because of the metallic structure resulting in decreased grayscale values at the immediate periphery, as also shown by a study by Kou28, who found the least grayscale value change in the proximity of a single dental implant prosthesis. The single implant with prostheses showed a higher decrease in grayscale values in the periphery than the only implant due to the additional cupping effect of increased metal content. Because of the artifact additive effect, the grayscale values increased with the number of implants8,28.

Grobee A21 supports Benic GI9 fontanelle29, who found significant artifacts up to 3.5 cm from the implant. The artifacts were present both in proximity and away from implants. The current results have emphasized the change in grayscale, with the diminishing influence of metal as the distance from the implant or prostheses increases.

The pattern of artifacts around the implant was observed at 3 mm, 5 mm, 7.5 mm, and 10 mm from the center of the implant, which showed a higher standard deviation at 0° and 180° in both the canine and molar region at 3 mm distance, indicating a greater amount of artifacts variation and worse image quality at these angular regions of interest (ROI)30. Schulze8, Benic9, and Pauwels31 also found a similar decrease in grayscale values at the interproximal region between two implants or prostheses. Benic9 recorded the greatest grayscale change at points perpendicular to the mandibular axis, corroborating with the present study. While Benic9 found reduced grayscale values in mesial and distal regions of a single implant, the present study recorded negative grayscale changes in the interproximal area and distal to molar region, but not at 0° (mesial) in the canine region where only reduced grayscale change values were recorded. Also, a study by Pauwels R31 concluded that the area between two metal objects was most affected by artifacts, which concurs with the present study.

The mean grayscale change at the superior axial plane was higher (50.72) than at the inferior axial plane (33.36), indicating a greater influence of implant prosthesis. Likewise, the grayscale value increased at a higher VIS with significant observations in mean grayscale values and changes in grayscale values as a VIS indicator accuracy is proportional to VIS. The findings endorse previous observations of Candemill Ap32, who concluded that smaller voxel sizes were preferred for higher accuracy.

The present study revealed higher but statistically non-significant grayscale values when the canine region (126.04 ± 93.26) was compared with the molar region (120.17 ± 86.48). The results indicated that artifacts produced are independent of the area or place of implant or prosthesis placement. This result concurs with earlier studies by Benic GI9 and Fontenele29, who also did not find any correlation between the position of the implant and artifacts.

The Limitation of the present study is a simplified in-vitro situation where the role of soft tissue around the mandible is eliminated. Since the study was carried out on a dry mandible, the reported results should be considered optimal, as precision and reliability are most likely lowered in clinical settings because of factors such as patient movement, adjoining soft tissue, and other intra-oral factors such as restorations that may influence linear measurements in a CBCT scan. Furthermore, lower VIS should be used to reduce artifacts and improve accuracy, as higher VIS increases artifacts. Moreover, the influence of different CBCT equipment/software for image reconstruction could also be studied.

Conclusion

The study aimed to assess the accuracy of CBCT as a measuring tool and quantify metal artifacts produced by CBCT of the mandible in the presence of implants and implant-supported prostheses. The results showed that the presence of implants and prostheses can cause artifacts, such as beam hardening and scattering, which can reduce the accuracy of the images. These artifacts were found to increase with the number of implants and the size of the prostheses. The artifacts were more pronounced in the presence of more metal, such as in full-arch implant-supported prostheses. The study also found that the artifacts were more pronounced in the superior axial plane than in the inferior axial plane. Furthermore, lower VIS should be used to reduce artifacts and improve accuracy, as higher VIS increases artifacts. Therefore, caution should be exercised when interpreting CBCT images with implants and prostheses, and additional diagnostic imaging modalities may be necessary to obtain more accurate information. Overall, this study provides important insights into the use of CBCT in implant dentistry and highlights the need for further research to improve imaging quality and reduce artifacts in CBCT images.