Abstract
To report the usefulness of intraoperative real-time adjustment of intraocular lens (IOL) tilt during the intrascleral fixation with intraoperative optical coherence tomography (iOCT) as a clinical evaluation and investigate the factors contributing to IOL tilt using iOCT as an experimental evaluation. Retrospective cohort study and experimental research. As a clinical evaluation, the medical records of 43 eyes of 41 patients who underwent intrascleral IOL fixation combined with real-time iOCT observation were retrospectively reviewed. As an experimental evaluation, in order to investigate the factors contributing to IOL tilt, the four experiments were performed using iOCT. The mean IOL tilt angle (°) at the end of surgery and 3 months after surgery were 1.81 ± 1.15 and 2.10 ± 1.66, respectively (p = 0.46). No apparent intra- or postoperative complications occurred during the follow-up period. The experimental evaluation indicated that the IOL tilt was influenced by the insertion angle of the haptic in the vertical direction. The mean IOL tilt angle (°) was 1.94 ± 0.09, 4.67 ± 0.11, 8.90 ± 0.11, and 15.78 ± 0.85 when the insertion angle of the haptic was 0°, 10°, 27.5°, and 45° in the vertical direction, respectively (p < 0.01). Clinical and experimental IOL tilt assessment using iOCT is interactively useful for better quality surgery and better postoperative outcome.
Similar content being viewed by others
Introduction
Recently, there have been increases in the frequencies of crystalline and intraocular lens (IOL) dislocation1. Consequently, the number of patients requiring IOL scleral fixation has also increased. Representative types of IOL scleral fixation include IOL suturing and intrascleral IOL fixation, but postoperative problems can occur with suturing, including displacement of the IOL due to looseness or breakage of the sutures2,3,4. Therefore, in recent years, intrascleral fixation is preferred over suturing5.
There are various methods of intrascleral IOL fixation6,7,8, but the frequency of performance of the Yamane intrascleral fixation technique has increased because of the ease of use9. Follow-up studies and arrangement methods relating to the Yamane intrascleral fixation technique have all reported favorable results10. However, with the Yamane intrascleral fixation technique, accurate setting of the needle insertion angle is difficult, and considerable practice is needed to achieve competence. Moreover, the large IOL tilt can develop unexpectedly11, resulting in postoperative refractive error11 and postoperative ocular coma-like aberrations12.
Various methods can be applied for overcoming the IOL tilt with the Yamane intrascleral fixation technique, including verification of the IOL tilt by checking the Purkinje-Sanson mirror image, surgical techniques using stabilizers13, and adjustment based on the amount of haptic pulled-out11. However, irrespective of the method used, intraoperative, real-time accurate verification of the angle of IOL tilt is difficult. Moreover, the cause of IOL tilt has not been investigated in detail to date.
Intraoperative optical coherence tomography (iOCT) has recently been found useful for safe and accurate surgery and providing information that is not visible under microscopy; moreover, the application of iOCT has increasingly gained popularity14. Although the most effective procedure for manipulating iOCT during intrascleral IOL fixation is still under debate, one useful technique for real-time adjustment of IOL tilt using iOCT was recently reported15.
In this study, we followed-up the method of intraoperative real-time adjustment of the IOL tilt with iOCT15 as a clinical evaluation and investigated the factors contributing to IOL tilt using an iOCT system (RESCAN 700; Carl Zeiss Meditec, Oberkochen, Germany) as an experimental evaluation.
Methods
Clinical evaluation
In total, 43 eyes of 41 patients who underwent Yamane intrascleral IOL fixation at our hospital between March 2021 and September 2022 were included in this study. The subjects were observed for 3 months after surgery. Medical records were retrospectively analyzed. The study was conducted in accordance with the Declaration of Helsinki, and approved by the Ethics Committee of Kobe University Graduate School of Medicine (No. B210074, June 25, 2021.). The need for informed consent was waived by the committee because of the retrospective, observational design of the study. Nonetheless, patients were able to withdraw consent at any time, which could be accessed on the hospital homepage as an opt-out choice. The following parameters were extracted from the medical records and used for statistical analysis: sex, age, preoperative best-corrected visual acuity (BCVA), BCVA at 1 month after surgery, BCVA at 3 months after surgery, preoperative intraocular pressure (IOP), IOP at 1 month after surgery, IOP at 3 months after surgery, preoperative corneal endothelial cell density (CECD), CECD at 1 month after surgery, CECD at 3 months after surgery, IOL tilt angle at the end of surgery and 1 month and 3 months after surgery, surgically induced astigmatism 3 months after surgery, refractive difference from the predicted value 3 months after surgery, and occurrence of intra- or postoperative complications.
Surgical procedure
All surgeries were performed at our hospital by one experienced vitreoretinal surgeon (H.I.). Sub-Tenon anesthesia was administered using 4 ml of mixed 2% lidocaine and 0.5% levobupivacaine. The 27-gauge pals plana vitrectomy (27GPPV) with a wide-angle non-contact viewing system (Resight®; Carl Zeiss Meditec AG, Jena, Germany) was performed using the Constellation Vision System (Alcon Laboratories, Inc., Fort Worth, TX, USA). Three cannulas were created with conjunctival displacement and oblique-angled sclerotomies in the inferotemporal, superotemporal, and superonasal quadrants 3.0–4.0 mm posterior to the limbus. For eyes without the history of vitrectomy, following core vitrectomy, the vitreous gel was visualized after injecting triamcinolone acetonide (TA) (MaQaid; Wakamoto Pharmaceutical, Tokyo, Japan) during midperipheral vitrectomy, and a complete 360° vitrectomy was performed for the peripheral vitreous gel. Posterior vitreous detachment was intentionally done if needed. After vitrectomy, for eyes with IOL dislocation, a side port corneal incision was created with a 20-gauge MVR blade (MVR-Lance; Alcon Laboratories, Inc., Fort Worth, TX USA). The dispersive viscoelastic materials (VISCOAT 0.5; Alcon Japan Ltd., Tokyo, Japan) were injected to fill the anterior chamber. Then, the IOL was positioned into the anterior chamber. A 2.4-mm bent transconjunctival single-plane sclerocorneal incision or corneal incision was made with a 2.4-mm slit knife (MSR24; Mani, Inc., Tochigi, Japan) at the 11 o’clock position. The IOL was resected in half and removed from the main incision. For eyes with crystalline lens dislocation, if it was located at the bottom of the fundus, after floating it as far as the iris surface using perfluorocarbon liquid, it was removed by phacoemulsification. If the crystalline lens was fixable using capsule expander, after fixation, it was removed by phacoemulsification and aspiration.
Next, the IOL (NX-70S; Santen Pharmaceutical Co., Ltd. Osaka, Japan) was inserted into the anterior chamber. Using a Yamane double-needle stabilizer(Geuder AG, Heidelberg, Germany) , two 30-gauge needles (TSK ultra-thin wall needle; Tochigi Seiko, Tochigi, Japan) were inserted into the sclera at the 4 and 10 o’clock meridians. Using Maxgrip-type ILM forceps(Alcon Laboratories) , the haptics of the IOL were then inserted into the lumens of the two 30-gauge needles, and pulled out for externalization. The position of the IOL was then adjusted with microscopic observation, and the dispersive viscoelastic substances were removed by irrigation and aspiration (IA) as much as possible; thereafter, TA was dispersed inside the anterior chamber. The TA inside the anterior chamber was deposited on the surface of the IOL, and iOCT (RESCAN 700) therefore displayed it as a high-intensity line. While continuing iOCT observation, the IOL tilt in the vertical and horizontal directions was adjusted, and the amount of haptic pulled-out was adjusted. Finally, after trimming the unnecessary haptics, the ends of the haptics were cauterized using an ophthalmic cautery device (Accu-Temp Cautery; Beaver Visitec, Waltham, MA, USA) to make a flange. The flange was pushed back and fixed into the scleral tunnels. The final IOL tilt was assessed intraoperatively by observing the iOCT, and the absence of an obvious tilt was confirmed. TA was then removed from the anterior chamber by IA, completing the surgery. (Supplemental Digital Content 1 demonstrates the method of intraoperative real-time adjustment of the IOL tilt with iOCT.)
Experimental evaluation
In order to investigate the factors contributing to IOL tilt and decentration, the experiments detailed below were performed. Both ends of the IOL haptic were fixed to a stand at positions 1 mm from the tip, with the optical surfaces positioned horizontally. The stand used was then positioned under the surgical microscope (RESCAN 700) so that the distance between the haptic tips was 13 mm (Fig. 1). The inter-haptic distance was measured using the CALLISTO eye function mounted in the surgical microscope (RESCAN 700). Thereafter, the images of the IOL line were taken in vertical and horizontal directions by iOCT, because iOCT generates images in which the anterior surface line of the IOL has high intensity in the atmosphere. The IOL tilt and decentration were analyzed as follows, taking as the standard the IOL position at that time-point. (Supplemental Digital Content 2 demonstrates how the IOL is imaged and moved under iOCT during the experimental evaluation.)
-
Experiment 1: Investigation of the effects of inconsistency in the distance from the limbus to the 30-gauge needle insertion position on tilt and decentration
The haptic on one side was fixed, and the other haptic was moved so that the distance between the haptic tips as shown by CALLISTO photography was 12, 13, 14, and 15 mm respectively. iOCT and CALLISTO photography were used for the analysis of IOL tilt and decentration at each time-point (Fig. 2A).
-
Experiment 2: Investigation of the effects of inconsistency in the 30-gauge needle insertion angle on tilt and decentration: Investigation of inconsistency in the vertical direction.
With observation from the horizontal direction, the angle of the haptic on one side was set at 0°, 10°, 27.5°, and 45° respectively, on the basis of the reference angle (0°), and iOCT and Callisto photography were used for the analysis of IOL tilt and decentration at each time-point (Fig. 2B).
-
Experiment 3: Investigation of the effects of inconsistency in the 30-gauge needle insertion angle on tilt and decentration: Investigation of inconsistency in the horizontal direction.
With observation from the vertical direction, the angle of the haptic on one side was set at 0°, 10°, 20°, and 30° respectively, on the basis of the reference angle (0°), and iOCT and CALLISTO photography were used for the analysis of IOL tilt and decentration at each time-point (Fig. 2C).
-
Experiment 4: Investigation of the effects of inconsistency in the amount of IOL pulled-out on tilt and decentration.
As mentioned above, both ends of the IOL haptic were fixed to a stand 1 mm from the tip first. The stand was then placed under an operating microscope (RESCAN 700) with a distance of 13 mm between the fixed positions of the haptic. Based on the fixed position of the haptic at this point, we moved the haptic on one side outward by 1, 3, and 5 mm, respectively, while taking care not to shift the haptic position from here. iOCT and CALLISTO photography were used for the analysis of IOL tilt and decentration at each time-point, respectively (Fig. 2D).
Each of the above measurements were performed five times using the IOL (NX-70 + 20.0D), and the values were used in the analysis.
The method of IOL tilt and decentration evaluation
For clinical evaluation, IOL images in vertical and horizontal directions were taken by using iOCT (RESCAN 700; Carl Zeiss Meditec) at the end of the surgery and swept-source anterior segment optical coherence tomography (AS-OCT) (SS-1000 CASIA; Tomey Corporation, Nagoya, Japan) during the follow-up period. For experimental evaluation, IOL images in vertical and horizontal directions were taken by using iOCT (RESCAN 700).
For the evaluation of IOL tilt, image software (Adobe Photoshop®: Adobe, California, USA) was used. IOL images were imported in jpeg format, and the images were enlarged with the magnification tool (”Fit on screen”). Then, the reference line was taken. For clinical evaluation, the line was drawn to be the line joining the straight line though the iris stroma on both sides by using the “line tool”, and a similar line was drawn along the anterior surface of the IOL using the same tool. The angle between reference line and the IOL anterior surface line was defined as the IOL tilt angle(Fig. 3A) . In the experimental evaluation, the horizontal direction of the screen was assumed to be the reference line, and the inclination of the IOL anterior surface line was measured as the IOL tilt angle. The mean of the IOL tilt angle in the horizontal and vertical directions was defined as the mean IOL tilt angle, which was used in the analysis.
For the measurement of IOL decentration in experimental evaluation, the image software (Adobe Photoshop®: Adobe, California, USA) was used. IOL images were imported in jpeg format and enlarged with the magnification tool (”Fit on screen”). The midpoints of the lines (top, bottom, left, and right) were connected using the line tool, and the intersection of the two midlines was set as the center of the frame. Then, the glued parts of the IOL haptics were connected with a straight line, and the midpoint of the line were regarded as the center of the IOL. The length of the measured line segment, or decentration distance (mm), was calculated from the ratio of the length of the line segment to the length of the frame line (pt), which corresponds to 13 mm. The IOL decentration distance (mm) was calculated from the decentration distance under the basic conditions and each condition(Fig. 3B) .
Statistical methods
For all variables, we reported the mean value and the standard deviation (SD). BCVA was converted to logarithmic minimum angle of resolution (log MAR) for statistical analysis. Friedman’s test and Wilcoxon t-test with Bonferroni correction for post hoc test were performed to assess changes in BCVA, IOP, CECD, and IOL tilt and decentration. The Mann–Whitney U-test and Kruskal–Wallis test for continuous variables were used to compare the parameters between groups. Statistical analyses were performed using SPSS software (version 24.0; IBM Corporation, Armonk, NY, USA). Statistical significance was considered at p < 0.05.
Results
Clinical evaluation
Table 1 summarizes the perioperative demographic data of the patients. Overall, 30 men and 11 women were included in this study. The mean age was 70.0 ± 13.8 years. All patients were followed-up for 3 months after surgery. Mean preoperative BCVA (logMAR), BCVA at 1 month after surgery, and BCVA at 3 months after surgery were 0.25 ± 0.45, 0.13 ± 0.32, and 0.13 ± 0.33, respectively (p < 0.01). Mean preoperative IOP (mmHg), IOP at 1 month after surgery, and IOP at 3 months after surgery were 15.0 ± 5.2, 14.1 ± 5.2, and 14.1 ± 3.8, respectively (p = 0.11). Mean preoperative CECD (cells/mm2), CECD at 1 month after surgery, and CECD at 3 months after surgery were 2102.1 ± 638.3, 1807.1 ± 700.9, and 1746.7 ± 736.5, respectively (p = 0.11). Mean IOL tilt angle (°) at the end of surgery and 1 month and 3 months after surgery were 1.81 ± 1.15, 2.11 ± 1.32, and 2.10 ± 1.66, respectively (p = 0.46) (Table 2). The mean surgically induced astigmatism 3 months after surgery was 0.93 ± 0.79 diopters, whereas the mean refractive error from the predicted value 3 months after surgery was − 0.48 ± 0.75 diopter. No apparent intra- or postoperative complications occurred during the follow-up period.
Experimental evaluation
In experiment 1, the mean IOL tilt angle (°) was 0.94 ± 0.09, 1.85 ± 0.05, 2.60 ± 0.10, and 3.49 ± 0.16 when the inter-haptic distance was 12 mm, 13 mm, 14 mm, and 15 mm, respectively (p < 0.01) (Fig. 4A). The IOL decentration (mm) was 0.24 ± 0.05, 0.08 ± 0.03, 0.30 ± 0.05, and 0.51 ± 0.05 when the inter-haptic distance was 12 mm, 13 mm, 14 mm, and 15 mm, respectively (p < 0.01) (Fig. 4B). In experiment 2, the mean IOL tilt angle (°) was 1.94 ± 0.09, 4.67 ± 0.11, 8.90 ± 0.11, and 15.78 ± 0.85 when the angle of the haptic was 0°, 10°, 27.5°, and 45°, respectively (p < 0.01) (Fig. 5A). The IOL decentration (mm) was 0.06 ± 0.03, 0.22 ± 0.04, 0.58 ± 0.05, and 1.11 ± 0.03 when the angle of the haptic was 0°, 10°, 27.5°, and 45°, respectively (p < 0.01) (Fig. 5B). In experiment 3, the mean IOL tilt angle (°) was 0.37 ± 0.10, 0.99 ± 0.02, 1.57 ± 0.13, and 2.81 ± 0.27 when the angle of the haptic was 0°, 10°, 20°, and 30°, respectively (p < 0.01) (Fig. 6A). The IOL decentration (mm) was 0.59 ± 0.01, 0.29 ± 0.04, 0.09 ± 0.04, and 0.40 ± 0.06 when the angle of the haptic was 0°, 10°, 20°, and 30°, respectively (p < 0.01) (Fig. 6B). In experiment 4, the mean IOL tilt angle (°) was 0.16 ± 0.59, 2.24 ± 1.00, and 3.59 ± 0.26 when the amount of the IOL pulled out was 1 mm, 3 mm, 5 mm, respectively (p < 0.01) (Fig. 7A). The IOL decentration (mm) was 0.13 ± 0.09, 1.49 ± 0.10, and 2.08 ± 0.16 when the amount of the IOL pulled out was 1 mm, 3 mm, and 5 mm, respectively (p < 0.01) (Fig. 7B).
Discussion
For clinical evaluation, we performed a follow-up study on the real-time, intraoperative adjustment technique for IOL tilt by using iOCT during the Yamane intrascleral IOL fixation, which was first reported by Fukumoto15. In Fukumoto’s study, the IOL tilt was 4.58° ± 1.86° at 1 month after surgery, which was similar to the results of the present study (2.11° ± 1.32° at 1 month after surgery and 2.10° ± 1.66° at 3 months after surgery). In contrast, Yamane, et al. reported that the postoperative IOL tilt was 3.4° ± 2.5° at 36 months after surgery without the usage of iOCT in their first report about flanged intrascleral IOL fixation16. In several studies without the use of iOCT, the postoperative IOL tilts were 8.80° ± 3.86°, 3.52° ± 3.00°, and 4.32° ± 3.12° at the end of the follow-up period17,18,19. These results show that the postoperative mean IOL tilt was not different between the technique with/without correcting IOL tilt in real time by means of the iOCT. However, the standard deviation of the IOL tilt seems to be much smaller than that in previous reports without iOCT. In addition, in the case of successful intracapsular IOL implantation, the postoperative IOL tilt has been reported to be 5.21° ± 1.47°17. This result shows that both the postoperative IOL tilt and its standard deviation were comparable between the present study and usual intracapsular IOL fixation. Additionally, Kimura reported that even in normal eyes, the crystalline lens exhibited an average tilt of 5.25° under mydriatic conditions, while the normally inserted IOL also had an average tilt of 4.65° under the same conditions20. These results align with our own findings. These collective results suggest that real-time IOL tilt correction using iOCT has the potential to further reduce IOL tilt in all subjects, thus preventing significant unintended tilt. As a result, this approach offers the possibility of achieving IOL tilt levels similar to those seen in a physiological crystalline lens, even in the absence of a normal crystalline lens capsule.
In this study, the postoperative refractive difference from the predicted value was also investigated, and it was found to be − 0.48 ± 0.75 diopters 3 months after the surgery. On the other hand, the postoperative refractive difference in the three previous reports with flange method were − 0.21 ± 0.9916, 0.86 ± 6.5917, and 0.03 ± 0.9318 diopters, and the results of our study tend to have higher postoperative refractive difference. In these three previous reports, a fixed value of 2 mm was used for the distance of the scleral insertion site of the 30-gauge needle from the limbus. As the distance from the limbus was constant, it was probable that postoperative refractive difference from the predicted value would not readily occur if there was no IOL tilt. In our study, the scleral insertion site of the 30-gauge needle was determined using a Yamane double-needle guide13. As the needle insertion angle is constant, and inconsistency in the distance from the limbus does not readily develop, this offers the advantage of suppressing the postoperative IOL tilt. On the other hand, depending upon the ocular axial length and the corneal diameter, the distance of the insertion site from the limbus is not constant because the size of stabilizer is constant. It is possible that this inconstancy affected the final induction of postoperative refraction difference. These findings suggest that, for insertion of a 30-gauge needle, in order to suppress postoperative IOL tilt and refractive difference, it may be important to insert the needle at a constant angle and set constant values for the distance of insertion from the limbus.
Clinical experience and the above results of clinical evaluation support the hypothesis that insertion angle and position are important to minimize postoperative IOL tilt, but which of the various angular and positional factors are particularly important to prevent postoperative IOL tilt and decentration are still unknown. To clarify these clinical questions, an experimental evaluation of the factors affecting IOL tilt and decentration after intrascleral IOL fixation using the flange method was also performed using iOCT system similar to that detailed above.
We found that all four of the parameters were linked with IOL tilt and decentration, and in particular, deviation in the insertion angle of the 30-gauge needle in the vertical direction had a major influence. This finding is in good agreement with our clinical experience and is considered to constitute useful data not only for physicians who are learning how to perform intrascleral IOL fixation, but also for those who already have considerable experience. Of the four parameters, three parameters other than amount of haptics pulled-out were difficult to correct after insertion of a 30-gauge needle, suggesting the importance of correct insertion during surgery. However, as the surgeon determines the insertion angle while observing the eyeball from the front, that is, from the corneal side during surgery, it is particularly difficult to perform surgery with an accurately defined insertion angle as viewed laterally. Thus, especially when the surgery is performed by inexperienced surgeons, stabilizers should be used, and other measures including iOCT should be taken to ensure that the insertion angle is constant.
The ability to adjust the amount of haptic pulled-out after insertion is one of the key techniques to minimize IOL tilt as clinically possible11. However, our results suggest that the IOL tilt that can be adjusted by the amount of haptic pulled-out has limits and is not very effective. Additionally, the uneven pulled-out amount has been found to increase the amount of IOL decentration. Furthermore, the extension of the IOL for surgical intrascleral fixation can alter the higher-order aberrations of IOL21. This indicates that it is important to set the angle and position correctly during insertion without resorting to this procedure.
The limitations of this study are as follows. First, the clinical evaluation has a small sample size and is a retrospective study. Additionally, a control group without the use of iOCT was not established. It is necessary to increase the number of cases and conduct a prospective study in the future. Furthermore, the experimental evaluation is based on model results and may not directly correlate with actual surgical outcomes. Therefore, it is important to understand that the results of the experimental evaluation in this study may not directly match clinical experiences. However, this study provides a detailed analysis of one of the issues associated with the rapidly spreading Yamane intrascleral IOL fixation technique, such as postoperative unexpected large IOL tilt, and holds significant value.
In conclusion, we reported the utility of iOCT for clinically minimizing postoperative IOL tilt and, as an experimental evaluation, assessed factors that contribute to IOL tilt. By making full use of the latest intraoperative examination equipment such as iOCT, it is possible to deepen our understanding of surgical techniques and improve surgical results.
Data availability
The datasets generated during and/or analyzed during the current study are available from H. Imai, the corresponding author, on reasonable request.
References
Mönestam, E. Frequency of intraocular lens dislocation and pseudophacodonesis, 20 years after cataract surgery: A prospective study. Am. J. Ophthalmol. 198, 215–222 (2019).
Vote, B. J., Tranos, P., Bunce, C., Charteris, D. G. & Da Cruz, L. Long-term outcome of combined pars plana vitrectomy and scleral fixated sutured posterior chamber intraocular lens implantation. Am. J. Ophthalmol. 141, 308–312 (2006).
Luk, A. S. W., Young, A. L. & Cheng, L. L. Long-term outcome of scleral-fixated intraocular lens implantation. Br. J. Ophthalmol. 97(10), 1308–1311. https://doi.org/10.1136/bjophthalmol-2013-303625 (2013).
McAllister, A. S. & Hirst, L. W. Visual outcomes and complications of scleral-fixated posterior chamber intraocular lenses. J. Cataract Refract. Surg. 37, 1263–1269 (2011).
Czajka, M. P. et al. Sutureless intrascleral fixation using different three-piece posterior chamber intraocular lenses: A literature review of surgical techniques in cases of insufficient capsular support and a retrospective multicentre study. Acta Ophthalmol. 98, 224–236 (2020).
Gabor, S. G. & Pavlidis, M. M. Sutureless intrascleral posterior chamber intraocular lens fixation. J. Cataract Refract. Surg. 33, 1851–1854 (2007).
Agarwal, A. et al. Fibrin glue-assisted sutureless posterior chamber intraocular lens implantation in eyes with deficient posterior capsules. J. Cataract Refract. Surg. 34, 1433–1438 (2008).
Ohta, T., Toshida, H. & Murakami, A. Simplified and safe method of sutureless intrascleral posterior chamber intraocular lens fixation: Y-fixation technique. J. Cataract Refract. Surg. 40, 2–7 (2014).
Suren, E. et al. Evaluation of the findings of patients who underwent sutureless flanged transconjunctival intrascleral intraocular lens implantation with or without pars plana vitrectomy. J. Ophthalmol. 2021, 4617583 (2021).
Yilmaz, S., Yildiz, A. M. & Avci, R. A comparative study of knotless versus knotted transscleral suture-fixated intraocular lens implantation. Indian J. Ophthalmol. 70, 131–137 (2022).
Kurimori, H. Y., Inoue, M. & Hirakata, A. Adjustments of haptics length for tilted intraocular lens after intrascleral fixation. Am. J. Ophthalmol. Case Rep. 10, 180–184 (2018).
Oshika, T. et al. Influence of tilt and decentration of scleral-sutured intraocular lens on ocular higher-order wavefront aberration. Br. J. Ophthalmol. 91, 185–188 (2007).
Yamane, S., Maruyma-Inoue, M. & Kadonosono, K. Needle stabilizer for flanged intraocular lens fixation. Retina 39, 801 (2019).
Ehlers, J. P. et al. The discover study 3-year results: Feasibility and usefulness of microscope-integrated intraoperative OCT during ophthalmic surgery. Ophthalmology 125, 1014–1027 (2018).
Fukumoto, R., Inoue, M., Ishida, T., Koto, T. & Hirakata, A. Adjustment of intraocular lens tilt during intrascleral fixation assisted by intraoperative OCT. J. Cataract. Refract. Surg. 47, 1308–1313 (2021).
Yamane, S., Sato, S., Maruyama-Inoue, M. & Kadonosono, K. Flanged intrascleral intraocular lens fixation with double-needle technique. Ophthalmology 124, 1136–1142 (2017).
Torii, T. et al. Comparison of clinical outcomes between intracapsular implantation and intrascleral fixation using the same model of intraocular lens. Clin. Ophthalmol. 14, 3965–3974 (2020).
Jujo, T. et al. 27-gauge trocar-assisted sutureless intraocular lens fixation. BMC Ophthalmol. 21, 8 (2021).
Do, J. R. et al. A 1-year prospective comparative study of sutureless flanged intraocular lens fixation and conventional sutured scleral fixation in intraocular lens dislocation. Ophthalmologica 244, 68–75 (2021).
Kimura, S. et al. Assessment of tilt and decentration of crystalline lens and intraocular lens relative to the corneal topographic axis using anterior segment optical coherence tomography. PLoS ONE 12, e0184066 (2017).
Matsuki, N., Inoue, M., Itoh, Y., Nagamoto, T. & Hirakata, A. Changes in higher-order aberrations of intraocular lenses with intrascleral fixation. Br. J. Ophthalmol. 99, 1732–1738 (2015).
Funding
This research received no external funding.
Author information
Authors and Affiliations
Contributions
Data acquisition, H.I. and Y.S.; Data analysis and interpretation, H.I., Y.S., and H.Y.; Figures and tables preparation, H.I. and Y.S.; Writing of main manuscript text, H.I.; Critical review and revision of the manuscript, Y.I, H.Y., W.M., A.M, S.K., and M.N. All authors have read and agreed to the published version of the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Sotani, Y., Imai, H., Iwane, Y. et al. Usefulness of intraoperative optical coherence tomography to minimize the intraocular lens tilt during the intrascleral fixation: a clinical and experimental evaluation. Sci Rep 13, 12065 (2023). https://doi.org/10.1038/s41598-023-39294-0
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-023-39294-0
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.