Introduction

Low-field magnetic resonance imaging (MRI) has the potential to revolutionize accessibility of MRI for patient diagnosis and neuroimaging1,2,3,4,5. The term “low-field” describes MRI scanners that operate at reduced fields (1–100 mT) compared to standard clinical MRI scanners (1.5–3 T). The lower magnetic fields can be generated using permanent magnets and therefore require less power, space, and accompanying infrastructure than clinical-field scanners that use cryogenic superconducting magnets6,7. The signal-to-noise ratio (SNR) of MRI scales approximately quadratically with applied field strength, requiring compromises in resolution or scan time to recover the signal. Recently, however, improvements in hardware and advanced image reconstruction (such as Deep Learning and post processing) have led to development of a new generation of low-field scanners8,9. Despite the lower SNR, advancements in both hardware and software have enabled low-field scanners that provide diagnostically relevant information within reasonable scan times, albeit with lower spatial resolution than clinical field strengths. Already, low field MRI scanners have been used to identify neurological pathology associated with strokes, hemorrhage, brain tumor, traumatic brain injury, and COVID-1910,11. The pathology is identifiable in T1-weighted, T2-weighted, T2 fluid-attenuated inversion recovery, and diffusion-weighted sequences using intrinsic tissue contrast at low-field. The scanners are highly portable, and, in some cases, the power requirements are low enough that a conventional wall socket can be used as a power source. As a result, low-field scanners have the potential to increase the accessibility of MRI in resource limited areas of the world12. Low-field MRI can also open up new avenues for point-of-care medical imaging. For example, low-field MRI can be performed at a patient’s bedside in intensive care units, enabling timely imaging to be performed in critically-ill patients who are too difficult to move13. We note that the range of magnetic field strengths below 0.1 T is sometimes referred to as “ultra-low-field” or “very-low-field” to distinguish from other commercial devices operating at intermediate fields9. The work described herein focuses on 64 mT and is referred to as “low-field” MRI for simplicity.

In 2020 the US Food and Drug Administration approved a portable low-field scanner operating at 64 mT for neuroimaging (Hyperfine Swoop, Guilford, CT, USA). The 64 mT MRI scanner operates at a field that is nearly 47 × smaller than 3 T clinical MRI. It can be powered via a standard wall socket and does not require cryogens. Portable MRI using these scanners can dramatically change the standard neuroimaging workflow. For instance, 64 mT MRI has been used to evaluate brain injury in intensive care units and perform neuroimaging on highly contagious patients with COVID-1910. Low-field MRI at 64 mT has also been used to assess intracerebral hemorrhages14, intracranial midline shift in stroke patients15, and hypoxic ischemic brain injury after cardiac arrest16. As low-field imaging becomes widespread, new imaging procedures specifically suited this magnetic field strength are needed, one of which is the exploration, evaluation, and characterization of contrast-enhanced MRI for low-field applications.

Contrast-enhanced MRI is currently used in about 25% of all MRI examinations17 at clinical field strengths. Contrast agents that have been traditionally used for clinical field strengths may be suboptimal in the low field-regime, since MRI contrast generation depends on a complex combination of factors and can change considerably depending on field strength. Therefore, we expect that novel low-field contrast agents will play an emerging role in low-field MRI. Contrast agents are typically categorized as either positive or negative contrast agents, depending on whether they increase or decrease signal, respectively, in MR images. Positive T1 contrast agents work by shortening the characteristic longitudinal T1 relaxation time, thus creating regions with increased signal. Negative T2/T2* contrast agents operate by reducing the T2 or T2* time associated with transverse relaxation and decrease signal. Positive T1 contrast agents have an advantage compared to negative contrast agents since they operate via signal increase. T2/T2* agents can obscure underlying contrast in MR image features from signal reduction. Additionally, by shortening the T1 time, it is possible to decrease the repetition time of experiments, allowing for an increased number of scans per unit time. T1 agents are also considerably less ambiguous than dark markers, since dark contrast can be caused by a variety of signal reducing mechanisms, such as air/water interfaces (i.e., bubbles) or accumulation of biogenic iron.

Candidates for effective T1 contrast agents should have high longitudinal relaxivities, meaning they efficiently produce T1 relaxation. High relaxivity is preferable because it means that smaller quantities of contrast agent can be used to create perceptible bright markers on an MR image. Besides high relaxivity, it is important that T1 contrast agents have a ratio of transverse relaxivity to longitudinal relaxivity that is on the order of one. That way, transverse relaxivity does not dominate the net relaxation effect produced by a contrast agent.

At clinical field strengths, contrast-enhanced MRI is dominated by the use of Gd-based chelates17. However, there have been recent concerns regarding the safety of Gd-based contrast agents, specifically with regards to toxicity and long-term deposition in the brain18,19. Iron oxide nanoparticles have been previously explored for contrast-enhanced MRI at clinical field strengths, specifically for imaging of the liver and spleen20,21. SPIONs are potentially less toxic than Gd since iron oxides can be metabolized by the body. At clinical fields, though, their high magnetic susceptibility generally favors increased transverse relaxivity over longitudinal relaxivity, so they have largely been thought of as negative contrast agents. Alternatively, SPIONs with “ultra-small” diameters of less than 4 nm can exhibit enhanced positive T1-weighted contrast at clinical field strengths22,23,24,25,26; however, these particles require substantial expertise to synthesize.

At field strengths below standard clinical imaging fields, SPIONs have shown exciting promise as T1 agents. Preliminary in vivo studies have been performed at 64 mT using ferumoxytol, which is an FDA-approved SPION-based treatment of iron deficiency anemia9,27. In these studies, patients received ferumoxytol for anemia treatment and were later imaged using 64 mT MRI. Contrast-enhanced cerebral vasculature was observed as a result of the ferumoxytol injections. Ferumoxytol has also been explored for off-label use as a contrast agent at 0.25 T and can generate comparable signal enhancement to a Gd-chelate with a lower net concentration of metal (Fe versus Gd)28. At 0.13 mT, the longitudinal relaxivity of SPIONs has been measured to be 615 L mmol−1 s−1, which is two orders of magnitude larger than Gd-based agents at clinical fields29. There are also possibilities for new contrast mechanisms using low-field MRI. For instance, a novel susceptibility-based positive contrast technique that operates using the unique nonlinear magnetization of SPIONs at 6.5 mT has been reported30. Also, low-field nuclear magnetic resonance curve (NMRD) relaxometry has been used to differentiate between intracellular and extracellular distribution of ferumoxytol in tumor associated macrophages31.

Here, we report on the characterization of SPIONs as T1 contrast agents using a commercially available and FDA-approved MRI scanner at a field strength of 64 mT. We measure the properties of SPION-based agents, such as the FDA-approved ferumoxytol. We also investigate monodispersed, carboxylic acid-coated SPIONs in order to understand the size-dependent properties of iron oxides and correlate those with structural and magnetic features. We find that SPION-based contrast agents exhibit favorable qualities as T1 contrast agents at low-field and outperform the longitudinal relaxivity of a commercially available Gd-based agent, gadobenate dimeglumine, by nearly 9 × at room temperature and approximately 8 × at physiological temperatures. We find that longitudinal relaxivities of SPIONs at 64 mT are nearly an order of magnitude larger than at a clinical field strength of 3 T, measuring up to 67 L mmol−1 s−1. Moreover, at 64 mT, the ratio of the transverse relaxivity to the longitudinal relaxivity is of order one, indicating that transverse relaxation does not dominate SPION induced relaxation. Together with the earlier works mentioned above9,27,28,29,30,31, this work suggests that SPION-based contrast agents could play a crucial role in contrast-enhanced MRI at low-fields.

Results

Structural and magnetic properties of contrast agents

Ferrimagnetic SPIONs with different sizes and coatings were used to explore T1 contrast for 64 mT MRI. The efficacy of contrast agents is directly related to their physical and magnetic properties. Thus, we characterized the properties of each contrast agent with electron microscopy and superconducting quantum interference device (SQUID) magnetometry to better understand their potential performance as T1 contrast agents. Figure 1a,b show transmission electron microscopy (TEM) images that represent the two general types of iron oxide particles used in this study. The first type are spherical, monodispersed SPIONs stabilized by a coating of carboxylic acid. Figure 1a shows an example of these particles. We acquired these types of particles (commercial vendors listed in Methods) with 4 sizes (4.9 nm, 8.5 nm, 12.9 nm and 15.7 nm) in order to investigate size-dependent contrast properties of the SPIONs. These particles have isolated iron oxide cores that are well-separated due to a combination of steric and electrostatic forces in aqueous solution. The second general class of SPIONs used in the study are therapeutic agents containing clusters of iron oxide cores with polydisperse diameter embedded in a polymer network. An example of this type of contrast agent, Feraheme (ferumoxytol), is shown in Fig. 1b. Ferumoxytol is an FDA-approved and commercially available pharmaceutical agent used for treatment of iron deficiency disorders. It is comprised of iron oxide cores embedded within a carbohydrate coat32,33. We also investigated a discontinued SPION-based imaging agent, Feridex (ferumoxides), which is made of an iron oxide-dextran complex20,34. In addition to SPIONs, we also procured an FDA-approved, Gd-based MRI contrast agent, Multihance (gadobenate dimeglumine), in order to compare the SPIONS to a commercially available Gd-chelate. Gadobenate dimeglumine (Gd-BOPTA) is a Gd-based chelate complex used for contrast-enhanced imaging of the central nervous system and magnetic resonance angiography17. Gd-BOPTA exhibits longitudinal relativities similar to other commercially available Gd-based contrast agents at clinical field strengths (1.5–3 T)35.

Figure 1
figure 1

(a) Transmission electron microscope (TEM) image of 12.9 nm of carboxylic coated SPIONs. (b) TEM image of ferumoxytol. (c) Magnetization as a function of applied field for ferumoxytol, 15.7 nm Fe3O4 SPIONS, and Gd-BOPTA with vertical bars to show field regions corresponding to 64 mT (low-field MRI) and 3 T (standard clinical field). (d) Expanded region of magnetization versus applied field showing low-field regime.

Figure 1c,d show room-temperature (21.5 °C) measurements of the magnetization as a function of applied magnetic field for two of the iron oxide-based agents, 15.7 nm carboxylic acid-coated nanoparticles and ferumoxytol. Figure 1d shows a magnified region of the same data in 1c from 0 to 0.1 T. The figures also show the magnetization of Gd-BOPTA for comparison. Both types of SPIONs have nonlinear magnetization curves that are characteristic of superparamagnetic particles. At zero applied field, the SPIONs show no remnant magnetization. As the applied field increases, their magnetization rapidly increases and then saturates above 2 T. At high fields, ferumoxytol shows a saturation magnetization that is comparable to bulk magnetite at room temperature (bulk Fe3O4 has MS = 92 Am2/kg)36. The 15.7 nm particles show a saturation magnetization that is 46% of bulk magnetite. This reduction in magnetization is often observed in magnetite nanoparticles and is caused by crystalline defects within the particles37,38. The other carboxylic acid-coated particles also have a reduced saturation magnetization compared to bulk magnetite, ranging from 40.3 to 74.6 Am2/kg. Gd-BOPTA shows paramagnetic behavior, as expected for a Gd-chelate, and exhibits a linear increase in magnetization as a function of applied magnetic field. The magnetization of SPIONs is much higher than Gd-based agents across the measured field range (0–7 T). This is also true for the low-field regime, where the SPIONs have already reached 40% to 50% of saturation magnetization at 64 mT (Fig. 1d). Magnetization data was collected for each of the contrast agents at 21.5 °C, approximately the same temperature used for MRI experiments. For the SPIONs, each magnetization curve was fit to a single Langevin function and used to extract the effective magnetic diameter of the particles (details in Supplemental Information). Table 1 has a summary of the physical and magnetic properties of the SPIONs.

Table 1 Magnetic and structural properties.

Magnetic resonance imaging at 64 mT and 3 T

Contrast agents were characterized with MRI at 64 mT and 3 T to compare performance between low-field and clinical field regimes. Measurements on contrast agents were performed using MRI phantoms for imaging procedures. The 64 mT MRI scanner used for low-field imaging is shown in Fig. 2a. The image shows the scanner with the retractable 0.5 mT critical boundary (FDA requirement to meet pacemaker safety at 0.5 mT) extended out above the permanent magnet assembly and head coil. Figure 2b shows a close-up image of the head coil with a phantom for imaging an array of samples. Samples were prepared in 50 mL centrifuge tubes using dilutions of contrast agents in agarose gel and arranged in the phantom, which was custom 3D printed by Hyperfine (Guilford, CT, USA) and filled with water. Figure 2c shows a typical T1-weighted fast spin echo (FSE) MR image of an axial cross-section of the phantom. In this image, the phantom is filled with an array of 8.5 nm SPION solutions containing different nominal concentrations of Fe (“nominal” refers to the Fe concentration specified by the manufacturer’s label).

Figure 2
figure 2

The 64 mT MRI scanner (Hyperfine, Guilford, CT) with retractable 0.5 mT field line extended above the unit. (b) A close-up image showing an array of samples arranged within a water-filled phantom and situated within the head coil for imaging. (c) A T1-weighted fast spin echo image showing a cross section of the phantom, containing a series of 8.5 nm SPION samples (TI = 900 ms, TE = 5.96 ms).

Figure 3 shows MRI data at 64 mT and 3 T for three contrast agents—15.7 nm carboxylic acid-coated nanoparticles, ferumoxytol and Gd-BOPTA. A comparison of T1-weighted MRI scans at 64 mT show a pronounced difference in relaxation properties of iron oxide-based contrast agents and Gd-BOPTA. Figure 3a shows longitudinal relaxation curves using signal intensity extracted from a 64 mT inversion recovery FSE sequence (details in Methods). Each sample contained 0.06 mmol/L nominal concentration of either iron or gadolinium metal. The signal at each inversion time represents the average signal value within a region of interest (ROI) containing a cross-section of the specific sample. Error bars correspond to the standard deviation of the signal in the ROI. Solid lines were generated by fitting the data to a general inversion recovery equation (Supplemental Information). At 64 mT, the 15.7 nm SPIONs are most efficient at reducing T1, followed by the ferumoxytol. The Gd-BOPTA has the slowest rate of T1 relaxation of the three contrast agents. The significant reduction in T1 for the iron-oxide based agents compared to Gd-BOPTA suggests that the SPIONs are more efficient at relaxing longitudinal magnetization at 64 mT. Figure 3b shows a corresponding curve at 3 T, which was also collected using a spin echo inversion recovery sequence (details in Methods). The 3 T data is markedly different from 64 mT, showing little difference between each of the contrast agents. The difference between longitudinal relaxation observed at low fields, therefore, is not observed at higher fields commonly used for clinical imaging.

Figure 3
figure 3

Inversion recovery curves for ferumoxytol, 16 nm SPIONS and Gd-BOPTA at (a) 64 mT and (b) 3 T. (c) Normalized image intensity from the 64 mT MR image corresponding to TI = 700 ms, error bars represent standard deviation of normalized intensity. (d) Normalized image intensity from the 3 T MR image corresponding to TI = 2000 ms, error bars represent standard deviation of normalized intensity. (e) T1 maps for corresponding ROI’s (regions of interest) at 64 mT and (f) T1 maps at 3 T.

The difference between contrast agents can be further visualized by inspecting individual spin echo images used to form the inversion recovery relaxation curves. Figure 3c shows 64 mT spin echo images (inversion time, TI = 700 ms) for each contrast agent and corresponding plots of the normalized image intensity. At 64 mT, the ferumoxytol and 15.7 nm particles have signal intensities that are roughly 5 × and 6 × larger than the Gd-BOPTA, because they relax longitudinal magnetization much more efficiently. Consequently, the 15.7 nm SPIONs and ferumoxytol appear bright relative to Gd-BOPTA. Figure 3d shows MR images (TI = 2000 ms) taken at 3 T, along with plots of the normalized intensity images. The 3 T data set shows little variation in contrast as a function of added agent.

T1 maps at 64 mT and 3 T offer another means for quantitative comparison between the contrast agents (Fig. 3e,f). For a nominal metal concentration of 0.06 mmol/L, the 15.7 nm SPIONs and ferumoxytol have T1 values of 266 ms and 353 ms, respectively, at 64 mT. These values are 29% and 38% of the value for Gd-BOPTA at 64 mT, which is 923 ms. At 3 T, however, the difference between T1 values among contrast agents is less pronounced. These values are 1347 ms for ferumoxytol, 1716 ms for the 15.7 nm SPIONs, and 1460 ms for Gd-BOPTA.

Relaxivity measurements at 64 mT and 3 T

To further quantify the difference between contrast agents, we measured the longitudinal and transverse relaxivities at 64 mT and 3 T. The relaxivity (r1 for longitudinal and r2 for transverse) is given by

$$\frac{1}{{T_{1,2} }} = \frac{1}{{T_{1,2}^{\prime } }} + r_{1,2} \cdot \left[ {CM} \right],$$
(1)

where [CM] is the concentration of metal (either Fe or Gd), T1,2 is the characteristic relaxation time of a solution containing [CM] and T1,2 is the relaxation time for a solution without contrast agent35. The relaxivity was calculated by measuring the relaxation times T1 and T2 for a series of samples containing different concentrations of metal (Fe for SPIONs and Gd for Gd-BOPTA). Then, the data was fit using the linear relationship described in Eq. (1) to extract the relaxivity, which is the slope of [CM] versus (1/T1,2).

The longitudinal relaxivity, transverse relaxivity, and the ratio of r2/r1 for all contrast agents at both field strengths, 64 mT and 3 T, are plotted in Fig. 4. The data are also summarized in Tables 2 and 3. For the carboxylic acid-coated SPIONs, the longitudinal and transverse relaxivities scale approximately with their physical core sizes. So, larger particles tend to have higher relaxivities. At 64 mT, iron oxide-based contrast agents exhibit much higher r1 values than the Gd-BOPTA. The 15.7 nm SPIONs have the highest r1 at 67 L mmol−1 s−1, which is a factor of 8.7 × higher than the paramagnetic Gd-BOPTA. Ferumoxytol has an r1 of 36.8, which is a 4.8 × increase compared to Gd-BOPTA. In general, the longitudinal relaxivities of iron oxide-based agents at 64 mT are an order of magnitude higher than those measured at 3 T, which range from 1.2 to 6.9 L mmol−1 s−1. Therefore, the iron oxide-based contrast agents are very effective at reducing T1 times at 64 mT. However, the property of high r1 does not, by itself, mean that a contrast agent will be an effective T1 agent. The contrast agent must also have a transverse relaxivity that is on the same order as its longitudinal relaxivity, so that transverse relaxivity does not dominate the net relaxation. At 64 mT, the SPIONs show a considerable reduction in transverse relaxivity compared to the values at 3 T. We used the longitudinal and transverse relaxivities to calculate the ratio r2/r1 for each contrast agent. This ratio is an important indicator of efficiency as a T1 agent and a lower value of r2/r1 means that the T1 reduction effect of the contrast agent will not be dominated by T2/T2* decay of signal35. The 64 mT ratios for all of the SPION-based contrast agents range from 1.2 to 2.4, which compares well to the value of Gd-BOPTA at 1.03 (Fig. 4c). The r2/r1 ratios for SPIONs are dramatically smaller than those measured at 3 T, which are an order of magnitude higher (Fig. 4d). A material is a good candidate for a T1 contrast agent if the ratio r2/r1 can be minimized while maintaining a high r1. At 64 mT, the SPIONs exhibit large values of r1 and low values of r2/r1, suggesting that they perform very well as T1 agents at 64 mT. Moreover, they exhibit r1 values that can exceed the Gd-BOPTA by up to 8.7 × while simultaneously showing a low r2/r1. This indicates that they may be a very useful alternative to Gd-chelates in contrast-enhanced MR procedures where positive contrast is preferred.

Figure 4
figure 4

(a) Longitudinal relaxivity, r1, for each contrast agent at 64 mT and 3 T. (b) Transverse relaxivities, r2, at 64 mT and 3 T. (c) Ratio r2/r1 for each contrast agent at 64 mT (d) Ratio r2/r1 for each contrast agent at 3 T. All measurements made at 21.5 °C.

Table 2 Relaxivity properties, 64 mT MRI.
Table 3 Relaxivity properties, 3 T MRI.

NMRD Measurements

To further understand field dependent relaxivity of the contrast agents, we measured the nuclear magnetic resonance dispersion (NMRD) curves of each contrast agent using fast-field cycling relaxometry. The relaxivity profiles for the contrast agents are shown in Fig. 5 at 21.5 °C and 37 °C. The solids lines for the SPIONs show fits to theory describing proton relaxation in the presence of superparamagnetic nanoparticles39,40. These fits capture the qualitative features of the data and have the shape typically observed for solvent molecules in the presence of SPIONs. The fit for the Gd-BOPTA was performed using Solomon-Bloembergen-Morgan theory, which describes proton relaxation in the presence of paramagnetic ions41,42,43. The relaxivity peaks present from 2 to 10 MHz originate from the rise in the thermal average of the electron magnetic moment of the SPIONs with increasing magnetic field followed by a decrease at frequencies on the order of the inverse of the correlation time for translational diffusion44,45. Translational diffusion is the dynamic process modulating the dipole–dipole interaction between the average magnetic moment of the SPIONs and the magnetic moment of water protons, and is the dominant T1 relaxation process at these field values. Its correlation time (\({\tau }_{D}=\frac{{d}^{2}}{D}\)) depends on the distance of closest approach (d) between electron and proton spins and it is thus related to the SPION core size, and on the diffusional coefficient (D). At low magnetic fields, a further contribution to water proton relaxation is present due to the dipole–dipole interaction between the (non-averaged) electron magnetic moment of the SPIONs and the magnetic moment of water protons. This interaction is typically modulated by the Néel correlation time (\({\tau }_{e}\)), which is the characteristic time associated with flipping the magnetization direction of the SPION’s net magnetic moment between easy axis directions. Figure 5 shows that the frequencies of the peak maxima decrease as the diameter of the nanoparticles is increased, in agreement with the diffusion-driven mode of relaxation described above.

Figure 5
figure 5

Relaxivity profiles obtained from the nuclear magnetic resonance dispersion (NMRD) data recorded at (a) 21.5 °C and (b) 37 °C. Solid lines show fits to theory for proton relaxation. The large vertical lines at 2.72 MHz correspond to the Larmor frequency of protons at 64 mT.

The relaxivity profiles show that the SPIONs exhibit enhancement of r1 across a range of Larmor frequencies ranging from 2 up to 20 MHz. The largest particles, 15.7 nm in diameter, have a peak in the NMRD curve that is nearly coincident with the proton Larmor frequency at 64 mT. It should be noted, though, that the proton Larmor frequency (which is proportional to the magnetic field, being equal to γIB0, where γI is the proton gyromagnetic ratio) is plotted on a log scale and so the peaks are rather broad and encompass a wide range of frequencies. Values for longitudinal relaxivity corresponding to 64 mT for each curve were extracted by interpolating from the theoretical fit. These are displayed in Table 2. The longitudinal relaxivities measured at 21.5 °C generally agree with the values measured using 64 mT MRI. The values measured at 37 °C (physiological temperature) show a reduction in relaxivity compared to 21.5 °C. Still, the SPIONs have higher relaxivities at this temperature compared to Gd-BOPTA. At 37 °C, the 15.7 nm SPIONs show an 8.3 × improvement and ferumoxytol shows a 5.6 × enhancement compared to the Gd-BOPTA.

Discussion

The relaxation properties of the SPIONs measured in this study have favorable properties for applications as T1 contrast agents for low-field MRI. They exhibit high r1 values, which can be nearly an order of magnitude larger than Gd-BOPTA for the largest particles measured here (15.7 nm). The SPIONs also have r2/r1 values that are of order one at 64 mT, meaning that effects of transverse relaxation will not dominate the net relaxation. Magnetometry shows that the SPIONs have a much larger magnetization than Gd-BOPTA at 64 mT, which contributes to their enhanced relaxivities compared to the Gd-chelate. The size range of SPIONs used in this study were chosen to represent a typical range of commercially available particles (diameters listed on label were 5 nm, 10 nm, 15 nm, and 20 nm; actual diameters were measured to be 4.9 nm, 8.5 nm, 12.9 nm, and 15.7 nm). It may be possible to achieve higher r1 values than those measured here by further increasing the size of particles past 15.7 nm or engineering the particles to have a higher magnetization. The 15.7 nm particles have a saturation magnetization of 42.4 Am2/kg, which is notably smaller than the bulk magnetite and can be attributed to defects and mixed phases of iron oxide37,38. Further optimization of longitudinal relaxivity may be achieved by synthesizing particles in the presence of oxygen to remove defects38 or by custom synthesis of larger particles.

The NMRD data suggests that SPION-based agents may also prove useful as T1 agents across the low-field regime, besides 64 mT. The peaks in the relaxivity profiles encompass 1H Larmor frequencies ranging from 1 to 30 MHz, which corresponds to field strengths of 23–705 mT. SPION-based agents may be tuned based on their size and magnetization so that they suit particular field strengths. At higher field strengths, though, the transverse relaxivities are also expected to increase, which will also increase the ratio of r2/r1. So, the design of an effective SPION-based agent for a particular field should account for the balance between effects related to both longitudinal and transverse relaxivity. The engineering of agents for particular fields can be guided by theory for proton relaxation from SPIONs. The SPIONs explored here all have NMRD curves that could be qualitatively modeled by pre-existing theories39,40, which were used to generate the fits to relaxivity profiles in Fig. 5. This suggests that there is a framework for understanding essential physics associated with T1 contrast at low fields and enables a pathway for design of contrast agents with high r1 and low r2/r1.

While synthesis of novel contrast agents has exciting promise, the pathway for clinical application of new contrast agents is arduous. However, the results reported here on the FDA-approved ferumoxytol have immediate clinical relevance. Ferumoxytol is currently FDA-approved for the treatment of iron deficiency but has been used off-label for clinical research on SPION-based MRI contrast. Arnold et al. have reported preliminary in vivo results at 64 mT showing contrast-enhanced cerebral vasculature in patients who received ferumoxytol injections as treatment for iron deficiency anemia9,27. These results suggest that ferumoxytol may be useful as a blood pool agent for low-field cerebral angiography. The measurements described in this manuscript show that ferumoxytol has a longitudinal relaxivity that is 5 × larger than Gd-BOPTA as measured by both 64 mT MRI and NMRD, suggesting that it has potential to be more sensitive than Gd-chelates. Since both ferumoxytol and the 64 mT MRI scanner are FDA-approved, there are immediate pathways available for exploring clinical applications of ferumoxytol as a low-field contrast agent. Further in vivo studies will be necessary, though, to determine whether or not the enhanced relaxivity of ferumoxytol compared to Gd-BOPTA confers sufficient advantage to be used as a T1 agent. Gd-chelates and ferumoxytol have very different pharmacokinetic properties, so ferumoxytol may find applications in imaging routines where Gd agents have unfavorable characteristics. For instance, ferumoxytol could be favorable in situations where blood pool agents, rather than Gd-based extracellular agents, are useful46. SPION-based agents also have some appeal from a toxicological standpoint since iron oxide particles are biocompatible and potentially less toxic than gadolinium32.

Since the measurements in this study were performed on phantoms, there are some limitations to translating the relaxivities reported here to hypothetical performance in vivo. For instance, the samples used in the study were embedded in agarose since it conferred long term stability, ensuring that identical samples could be compared between different MRI scanners. While agarose provides a hydrated, tissue-mimicking environment, it is not nearly as complex as in vivo conditions. In clinical applications, contrast agents are introduced into biological environments where a variety of pharmacokinetic factors, such as aggregation or clearance time, can have an effect on relaxivity. The relaxivity of contrast agents will also change depending on the surrounding fluid environment. It is well known, for instance, that relaxivity of contrast agents varies depending on whether they are immersed in blood, water, or plasma35. In order to understand how environmental changes may affect r1, we calculated theoretical relaxivity curves for isolated ferumoxytol particles in media with diffusion coefficients corresponding to different bio-fluids and tissues, such as blood, cerebrospinal fluid, white matter, and grey matter (Supplemental Information). The calculations were performed using the same model as the fits to NMRD data. The theoretical curves show an enhancement of relaxivity in tissues with lower diffusion coefficients, meaning that r1 may show a diffusion-based enhancement for certain types of tissue.

Also, the 64 mT MRI measurements were performed at laboratory room temperature (21.5 °C). The performance of the contrast agents will change at physiological temperatures (37 °C). The NMRD data taken at 37 °C, however, suggest that the SPION-based contrast agents have significant longitudinal relaxivity compared to Gd-BOPTA at physiological temperature. While the phantom environments measured in this study do not simulate in vivo conditions perfectly, the measurements still provide valuable information for understanding the potential of SPION-based agents at low-field.

In addition, the toxicity profiles of SPION-based contrast agents will affect their biological uptake and will impact applications as contrast agents. Toxicity can be influenced by a variety of parameters, such as nanoparticle size or surface coating. Therefore, if high relaxivity SPIONs are developed specifically for future low-field applications, in vitro or in vivo toxicity tests will be needed to gauge the safety and efficacy of new nanoparticle formulations. Previous studies on the toxicity of SPION formulations for biological applications and can be used to guide future design efforts for low-field contrast agents47,48,49.

In conclusion, we have characterized SPION-based contrast agents with several techniques with the aim of evaluating their effectiveness as positive contrast T1 agents for 64 mT MRI. We focused on the field strength of 64 mT because of the emergence of an FDA approved MRI scanner at this field. Different types of SPIONs were measured using 64 mT MRI and compared to 3 T MRI. Measurements with 64 mT MRI show that SPIONs have much higher longitudinal relaxivity (9x) compared to a commercially available Gd-based agent at room temperature while also exhibiting low r2/r1 ratios. NMRD data was recorded at several temperatures, including physiological temperature, to better understand the field-dependent and temperature-dependent properties of the contrast agents. We found that at physiological temperatures, ferumoxytol, an FDA-approved SPION-based therapeutic agent, showed a 5.6 × improvement in r1 over Gd-BOPTA. Carboxylic-acid coated SPIONs with a diameter of 15.7 nm did even better, showing an 8.3 × enhancement. The relaxivity profiles could be fit to existing theories for proton relaxation in the presence of SPIONs. Taken together, the measurements suggest that SPIONs can play a potentially important role as positive contrast T1 agents for emerging applications in low field-MRI.

Methods

Materials

Carboxylic acid-coated iron oxide nanoparticles were purchased from Sigma Aldrich and MK Nano. Specifically, the 4.9 nm, 8.5 nm, and 15.7 nm particles were acquired from Sigma Aldrich (the labels report diameters of 5 nm, 10 nm, and 20 nm, respectively). The 12.9 nm particles were acquired from MK Nano (listed as 15 nm particles). Agarose (BioReagent for molecular biology, low electroendosmosis) was also purchased from Sigma Aldrich.

MRI sample preparation

Samples were prepared by diluting contrast agents in an agarose gel (prepared with mass fraction of 1% agarose in H2O) based on the concentrations listed on the manufacturers label. Samples were prepared with concentrations of 0.03 mmol/L, 0.06 mmol/L, 0.12 mmol/L, 0.25 mmol/L, 0.5 mmol/L and 1 mmol/L of metal (Fe for the SPIONs and Gd for Gd-BOPTA).

Transmission electron microscopy (TEM)

TEM was performed using a Tecnai T12 Spirit BT microscope with a LaB6 filament. Samples were prepared by diluting nanoparticles in a 1:1 volume ratio mixture of isopropyl alcohol and H2O. The diluted samples were added dropwise to the top of Formvar/carbon-coated copper grids purchased from Ted Pella. Physical core sizes of particles were analyzed using TEM images and ImageJ50.

3 Tesla MRI

An Agilent preclinical scanner was used for 3 T MRI. T1 measurements were made using an inversion recovery sequence with a 256 × 256 matrix and a 128 mm × 128 mm field of view. Images were acquired from 6 axial slices with a thickness of 2 mm and a gap spacing of 4 mm. The inversion times used for the sequence were 50 ms, 75 ms, 100 ms, 125 ms, 250 ms, 500 ms, 1000 ms, 1500 ms, 2000 ms, 3000 ms, and 6000 ms. The repetition time was 10000 ms and the echo time was 13.92 ms. T2 measurements were performed using a spin echo sequence with the same resolution, field of view, and slice parameters as the T1 measurements. Echo times used for T2 acquisition were 14 ms, 28 ms, 56 ms, 112 ms, and 224 ms. The repetition time was 10000 ms. Measurements were made using a temperature-controlled phantom. Temperature control was achieved using a closed flow loop of a perfluorocarbon coolant and monitored by a fiber optic sensor. The temperature was set to 21.5 °C in order to match the lab temperature of the 64 mT MRI measurements.

64 mT MRI

A Hyperfine Swoop scanner with hardware version 1.8 and software version rc8.3.1, was used to acquire T1 and T2 measurements at 64 mT. All scans used an 8-channel receive, 1-channel transmit head coil. The. T1 measurements were made using a research version of the Hyperfine proprietary T1-weighted inversion recovery 3D fast spin echo (FSE) sequence with a 220 mm × 180 mm × 180 mm field of view, an in-plane resolution of 1.6 mm × 1.6 mm, and a slice thickness of 5 mm. The inversion times used for the sequence were 100 ms, 200 ms, 300 ms, 400 ms, 500 ms, 600 ms, 700 ms, 800 ms, 900 ms, 1100 ms, 1300 ms, 1500 ms, 1800 ms, 2100 ms, and 2500 ms. The repetition time was 3000 ms and the echo time was 5.96 ms. T2 measurements were performed using a research version of the Hyperfine proprietary T2-weighted 3D FSE sequence with a 220 mm × 180 mm × 180 mm field of view, an in-plane resolution of 1.5 mm × 1.5 mm, and a slice thickness of 5 mm. Echo times used for T2 acquisition were 37 ms, 111 ms, 184 ms, 259 ms, 333 ms, 407 ms, 480 ms, 554 ms, 628 ms, and 702 ms. The repetition time was 3000 ms. Measurements were performed at an ambient lab temperature of 21.5 °C.

Concentration measurements

The Fe concentration of each SPION-based contrast agent was measured using inductively coupled plasma optical emission spectroscopy (ICP-OES). ICP-OES was performed using a Perkin Elmer Optima 8300 ICP-OES optical system with a segmented-array charge-coupled device detector. Briefly, samples were digested using nitric acid, then further diluted to generate samples for measurement with ICP-OES. A detailed description of sample preparation is given in the Supplemental Information.

T1/T2 mapping and relaxivity analysis

Samples were prepared with nominal concentrations (diluted according to concentration on manufacturer’s label) of 0 mmol/L, 0.03 mmol/L, 0.06 mmol/L, 0.12 mmol/L, 0.25 mmol/L, 0.5 mmol/L and 1 mmol/L. For measurements at 3 T, the 0 mmol/L sample was not measured because the 3 T sample holder accommodates one fewer sample than 64 mT. The concentration of the SPIONs were measured using ICP-OES and the nominal concentrations were rescaled to reflect the actual concentration of the samples (see Supplemental Information for more details). Samples were prepared by embedding contrast agents in an agarose solution (prepared with mass fraction of 1% agarose in H2O), which was allowed to set into a semi-solid gel at room temperature. An agarose medium was chosen for imaging because it provided long term colloidal stability, so that the same samples could be scanned at 64 mT and 3 T. Diluting the samples in other solutions did not confer the same degree of colloidal stability (see Supplemental Information). T1 was calculated for each voxel using LMFIT in Python for the inversion-recovery model,

$$S_{i} = S_{o} \left| {1 - \left( {1 + d} \right)e^{{ - {\raise0.7ex\hbox{${TI}$} \!\mathord{\left/ {\vphantom {{TI} {T_{1} }}}\right.\kern-0pt} \!\lower0.7ex\hbox{${T_{1} }$}}}} + e^{{ - {\raise0.7ex\hbox{${TR}$} \!\mathord{\left/ {\vphantom {{TR} {T_{1} }}}\right.\kern-0pt} \!\lower0.7ex\hbox{${T_{1} }$}}}} } \right|,$$
(2)

with T1 the target value to fit, inversion time TI, repetition time TR, scale factor for imperfect inversion d, the nominal signal intensity for a voxel S0, and measured signal intensity Si. At 64 mT, T2 maps were acquired using a research version of the Hyperfine proprietary T2-weighted 3D FSE sequence. The T2 map was calculated in the Hyperfine protocol using SciPy optimize curve_fit in Python for the model,

$$S_{i} = S_{o} e^{{ - {\raise0.7ex\hbox{${TE}$} \!\mathord{\left/ {\vphantom {{TE} {T_{2} }}}\right.\kern-0pt} \!\lower0.7ex\hbox{${T_{2} }$}}}} ,$$
(3)

with T2 the target value for the fit, echo time TE, nominal signal intensity without relaxation S0, and measured signal intensity Si. At 3 T, the T2 maps were found by using LMFIT for each voxel using Eq. (3).

After T1 and T2 were measured for all concentrations and field strengths, the data was fit using Eq. (1) to calculate the relaxivities. Unfortunately, the experimental protocol at 64 mT could not accurately measure T1 values less than the first inversion time of 100 ms. At high concentrations (0.25–1 L mmol−1 s−1), some of the SPION-based agents with larger physical core sizes had T1 values lower than 100 ms. So, we instituted a selection rule to exclude cases where the concentration of contrast agents was too high for accurate determination of measurement times. The rule states that if the T1 value calculated using a fit of Eq. (2) was less than the first inversion time (100 ms for 64 mT), then the data was excluded from the relaxivity fit (see Supplemental Information for more details). The data corresponding to these concentrations was then excluded from fitting the relaxivities at other conditions (r2 at 64 mT, r1 at 3 T and r2 at 3 T) to maintain consistency across all field strengths and relaxation types. Out of 42 measurements (6 concentrations for 7 different agents), this rule eliminated 11 data sets from relaxivity analysis and fitting. The data sets that were eliminated corresponded to samples with large nanoparticle sizes at high concentrations, since these tended to relax signal very rapidly. A table has been included in the Supplemental Information to indicate the samples that were excluded on this basis.

SQUID magnetometry

The magnetization of samples was measured using a Quantum Design SQUID MPMS 3 magnetometer. Samples were immobilized in a water-soluble hydrogel and cured with UV light to prevent magnetic field-induced chaining and aggregation of particles during measurement.

NMRD

Water relaxation profiles were acquired with a Stelar Spinmaster FFC2000-1 T relaxometer by measuring the water proton relaxation rates as a function of the applied magnetic field (0.01–40 MHz proton Larmor frequency). The field-cycling technique is used to detect the magnetization decay/recovery curves by acquiring the free induction decay signals from the sample after exposure to a given magnetic field for 16 different intervals of time51. By changing the magnetic field, longitudinal relaxation rates can be determined at all frequencies permitted by the instrument. For magnetic fields lower than 0.35 T, a 1 T pre-polarization field was applied before the measurement to generate sufficiently large signals for NMRD acquisition. The relaxation measurements, obtained from the fit of the magnetization decay/recovery curves against a mono-exponential function, were affected by an error below ± 1%. Samples were prepared with a nominal metal concentration of 1 mmol/L (using concentration specified by manufacturer) in an agarose gel (prepared with mass fraction of 1% agarose in H2O). The relaxivity profiles were obtained by normalization of the measured relaxation data to the metal concentration (measured by ICP-OES) after subtracting the diamagnetic relaxation rate contribution from the agarose (NMRD profiles of the agarose are included in the Supplemental Information). The measurements were performed at 21.5 °C for comparison with the MRI data and 37 °C to evaluate relaxivity at physiological temperature. Additional measurements were performed at 15 °C and 25 °C and are included in the Supplemental Information.