Correction to: Scientific Reports https://doi.org/10.1038/s41598-022-21122-6, published online 24 October 2022


The original version of this Article contained a repeated error, where the pressure unit was incorrectly given as ‘mbar’ instead of ‘mmHg’.


In the Results and discussion section, under the subheading ‘Intra-catheter pressure’,


“The average pressure variation for Brand A was − 364 ± 42 mbar, − 248 ± 81 mbar for Brand B and − 272 ± 59 mbar for Brand C at 20 cmH2O. When the abdominal pressure was adjusted to 50 cmH2O, the average pressure for Brand A was − 383 ± 50 mbar, − 323 ± 47 mbar for Brand B and − 330 ± 93 mbar for Brand C.”


now reads:


“The average pressure variation for Brand A was − 364 ± 42 mmHg, − 248 ± 81 mmHg for Brand B and − 272 ± 59 mmHg for Brand C at 20 cmH2O. When the abdominal pressure was adjusted to 50 cmH2O, the average pressure for Brand A was − 383 ± 50 mmHg, − 323 ± 47 mmHg for Brand B and − 330 ± 93 mmHg for Brand C.”


Additionally,


“The measured intra-catheter pressure variation for Brand B at 20 cmH2O was equal to − 296 ± 56 mbar (N = 9, SD) for the tests where mucosal suction was perceived by the operator. Conversely, the intra-catheter pressure variation that could be measured at the first flow-stop for Brand B at 20 cmH2O when mucosal suction was not detected by the operator was equal to − 180 ± 64 mbar (N = 6, SD). A similar scenario was seen for Brand C at 50 cmH2O, where the measured intra-catheter pressure variation was equal to − 373 ± 62 mbar (N = 11, SD) when mucosal suction was perceived by the operator, and to − 212 ± 45 mbar (N = 4, SD) when mucosal suction was not perceived by the operator.”


now reads:


“The measured intra-catheter pressure variation for Brand B at 20 cmH2O was equal to − 296 ± 56 mmHg (N = 9, SD) for the tests where mucosal suction was perceived by the operator. Conversely, the intra-catheter pressure variation that could be measured at the first flow-stop for Brand B at 20 cmH2O when mucosal suction was not detected by the operator was equal to − 180 ± 64 mmHg (N = 6, SD). A similar scenario was seen for Brand C at 50 cmH2O, where the measured intra-catheter pressure variation was equal to − 373 ± 62 mmHg (N = 11, SD) when mucosal suction was perceived by the operator, and to − 212 ± 45 mmHg (N = 4, SD) when mucosal suction was not perceived by the operator.”


Furthermore,


“What remains to be understood is whether a pressure variation of, for example − 250 mbar, is sufficient to cause discomfort to the IC users, or even cause microtraumas to the bladder mucosa, and if the speed at which the peak is generated has any relevance.”


now reads:


“What remains to be understood is whether a pressure variation of, for example − 250 mmHg, is sufficient to cause discomfort to the IC users, or even cause microtraumas to the bladder mucosa, and if the speed at which the peak is generated has any relevance.”


Under the subheading ‘In-vivo animal studies’ of the same section,


“During bladder emptying, the pressure difference at first flowstop was equal to − 96 mbar (Fig. 15).”


now reads:


“During bladder emptying, the pressure difference at first flowstop was equal to − 96 mmHg (Fig. 15).”


Finally, the error was also present in Table 1 and in Figures 10, 11, 12, 13 and 15.


The correct and incorrect values of Table 1 appear below.


Table 1


Incorrect:

 

ΔPressure ± SD (mbar)

N

ΔPressure ± SD (mbar)

N


Correct:

 

ΔPressure ± SD (mmHg)

N

ΔPressure ± SD (mmHg)

N


The original Figures 10, 11, 12, 13 and 15, and accompanying legends appear below.


The original Article has been corrected.

Figure 10
figure 10

Examples of intra-catheter pressure sensor measurements. (a) Brand A, (b) Brand B, and (c) Brand C. The numbers on the figures represent specific events during IC: (1) insertion of the catheter through the sphincter and into the bladder, emptying starts; (2) flow-stop with an associated mucosal suction; (3) series of mucosal suction events during repositioning; (4) withdrawal of the catheter out of the bladder. The first mucosal suction pressure drop for each example, as indicated by the numbers “2” is zoomed in next to the pressure profile. In the zoomed in picture, the measured profile is shown in blue whereas a gaussian fitting is depicted in red. Brand A, B, and C were tested 5 times in 3 different porcine LUTs (N = 15, SD). An abdominal pressure of 50 cmH2O was used.

Figure 11
figure 11

Pressures measured with the intra-catheter pressure sensor at first flow-stop. The test was performed at both 20 and 50 cmH2O of abdominal pressure. Each Brand was tested 5 times in 3 different porcine LUTs. The same porcine LUTs where used at both abdominal pressures. Results are reported as individual values, mean and standard deviation. Statistical analysis was performed by means of t-test using Welch´s correction when appropriate.

Figure 12
figure 12

Comparison between the pressure at first flow-stop recorded with the intra-catheter pressure sensor. The results are divided according to whether the mucosal suction phenomenon was perceived by the operator during catheterization, or not. Results are reported as individual values (N = 15, SD). Statistical analysis was performed by means of t-test using Welch´s correction when appropriate.

Figure 13
figure 13

Example of hammering measured with the intra-catheter pressure sensor (Brand C, 20 cmH2O).

Figure 15
figure 15

In-vivo in-catheter pressure analysis. The pressure drop visible after the 150 s mark corresponds to the perceived mucosal suction phenomenon.