Introduction

Due to the rapid growth in the super-aging population worldwide, an increasing number of patients have been suffering from such cardiovascular related diseases as myocardial and cerebral infarction, the major cause of which is arteriosclerosis, well known as a ā€œsilent killerā€1,2. Prevention, early diagnosis and treatment of vascular dysfunction as well as cardiovascular complications are therefore of great importance; noninvasive methods are preferred, where possible. Currently, there are several noninvasive vascular function tests performed in clinical practice, such as flow-mediated dilation (FMD)3,4,5, pulse wave velocity (PWV)6,7, ankle-brachial pressure index (ABI)8 and carotid echocardiography9. However, almost all of these tests cannot provide any specific information about arterial mechanical properties themselves, despite their importance and usefulness in clinical practice. This is due mainly to the technical difficulty of quantitatively acquiring representative measures of mechanical properties.

Among these function tests, the FMD test is at least commonly utilized and, in fact, is the only assessment technique for the acquisition of vascular endothelial function. The latter is one of the most important causative factors of arteriosclerotic cardiovascular diseases, and is considered to be a possible potential factor in producing future cardiovascular disease events3,4,5,10. Additionally, it is noteworthy that vasodilatory functions can be strongly associated with peripheral circulatory injuries. One important clinical problem can be the creation of pressure ulcers and the tests can be useful here as part of a treatment to improve sacral skin blood circulation, which is especially important for people with spinal cord injury11,12,13.

With the FMD test, the inner diameter of a brachial artery is measured using ultrasonic equipment before (baseline) and after a 5-min supra-systolic cuff occlusion of the forearm. Vasodilatation after release of the cuff occlusion occurs following an acute increase in blood flow, typically induced via circulatory arrest in the arm. This hyperemia increases shear stress of the inner vessel wall which is transferred to the endothelial cells, inducing nitric oxide production (well known as an endothelium-derived vascular relaxation factor), and then a subsequent vasodilatation occurs4. The increase in arterial diameter, as a consequence of the reactive hyperemia, is compared to the baseline diameter and expressed simply as a percentage of this baseline diameter (%_FMD), which is clinically revealed to be more than 6ā€“7% (up to around 25%) in normal young and old subjects3,10.

Although the FMD test can have clinical utility, there are several technical and operational issues to be addressed in terms of more precise and continuous measurement of arterial diameter, more appropriate attachment of an ultrasonic probe, more compact ultrasonic device with greater portability3,14. This is a hospital-only test, and it would be more desirable and useful for a person to perform a test similar to FMD by him-/herself without the use of ultrasonography, at home as with blood pressure measurement. This would allow checking the vascular function at any time, and even if the subject has an early arteriosclerosis such home-based check could be helpful for a follow-up examination to introduce lifestyle improvements and/or drug therapeutic measures to manage the vascular condition. In this connection a convenient method to evaluate the FMD without using an ultrasound technique was developed, named as ā€œenclosed zone FMD (ezFMD)ā€15,16. This was based on a conventional cuff-oscillometric sphygmomanometer to predict vasodilatation of the radial artery by measuring cuff-pressure oscillations. Although this method is easy and appropriate to use at home, it is difficult to acquire quantitatively direct radial arterial diameter movements as well as arterial mechanical properties.

Taking the vasodilatation mechanism into account, as mentioned above, it is naturally conceivable that changes in arterial elastic properties would occur in reaction to hyperemia. However, we have not found literature reporting the arterial biomechanical properties quantitatively with hyperemia.

The hypothesis of the present study is therefore that arterial elasticity would change following reactive hyperemia. This paper therefore describes the investigation of the impact on possible changes in arterial elasticity with hyperemia. We have recently developed a smartphone-based compact and easy-to-operate device capable of noninvasive measurement of elastic properties (volume elastic modulus (Ev) and volume change ratio (Ī·v)) in finger and radial arteries (see further explanations and overview of the system shown in Figure A2 in the Appendix (a-1)). These measures are expressed as an exponential function of mean blood pressure (MBP) or transmural pressure (Ptrā€‰=ā€‰MBPā€‰āˆ’ā€‰(externally applied pressure or cuff pressure, Pc))17, based on the combination of transmittance- or semi-transmittance-type infrared photo-plethysmography and volume-oscillometric sphygmomanometry (VOS)17,18,19,20,21. Therefore, following on from the above background, this paper describes a first approach for studying the impact of changes in elastic properties of these peripheral arteries with reactive hyperemia. This has implications for the possibility of creating a convenient assessment of endothelial activity as a substitute for the FMD test.

Results

Examples of collected data are shown in Fig.Ā 1. This shows: Ev(Ptr) vs Ptr (upper left panel; (a-1) and lower left; (b-1)) and Ī·v(Ptr) vs Ptr relationships (upper right panel; (a-2) and lower right; (b-2)) before (data points of solid circle; pre-mFMD) and after cuff occlusion (data points of solid diamond; post-mFMD). These were obtained in the finger (upper part: Sub. No. 11) and the radial artery (lower part: Sub. No. 05) of two subjects. Schematic illustrations of the measurement scene in the finger (left upper side) and in the wrist (left lower) are included in this Figure. Solid and grey curves indicate approximately exponential curves for Ev(Ptr) and Ī·v(Ptr) with coefficients of determination, R2, before and after cuff occlusion, respectively. Each exponential equation is also inset into each Figure. In the examples, five parameters (Ī·v0, Ev0, Ī±, Kstif and kdist) for pre- and post-mFMD and %_parameters (%_Ī·v0, %_Ev0, %_Ī±, %_Kstif and %_kdist) were as follows (see also Sub. No. 05 and 11 of Table 1):

  • Finger_pre-mFMD; (0.601, 153, 0.0334, 5.11 and 1.24)

  • Finger_post-mFMD; (0.737, 116, 0.0302, 3.51 and 1.40)

  • Finger_%_changes; (22.6, āˆ’Ā 24.2, āˆ’Ā 9.58, āˆ’Ā 31.2 and 12.4)

  • Radial_pre-mFMD; (0.934, 62.4, 0.0482, 3.01 and 1.50)

  • Radial_post-mFMD; (1.103, 56.0, 0.0395, 2.21 and 1.83)

  • Radial_%_changes; (18.8, āˆ’Ā 10.2, āˆ’Ā 18.0, āˆ’Ā 26.4 and 21.8)

Figure 1
figure 1

Examples of Ev(Ptr)āˆ’ā€‰Ptr and Ī·v(Ptr)ā€‰āˆ’ā€‰Ptr curves before (pre-mFMD) and after cuff occlusion (post-mFMD) on the finger and the radial arteries. Two examples of Ev(Ptr)ā€‰āˆ’ā€‰Ptr curves (upper left panel; (a-1) and lower left; (b-1)) and Ī·v(Ptr)ā€‰āˆ’ā€‰Ptr curves (upper right panel; (a-2) and lower right; (b-2)) before (data points of solid circles; pre-mFMD) and after cuff occlusion (data points of solid squares; post-mFMD) obtained in the finger (a, upper part: Sub. No. 11) and the radial artery (b, lower part: Sub. No. 05). Schematic illustrations of the measurement scene in the finger (left upper side) and in the wrist (left lower) are included in this Figure. Solid and grey curves indicate approximately exponential curves for Ev(Ptr) and Ī·v(Ptr) with coefficients of determination R2 before and after cuff occlusion, respectively. Each exponential equation is also inserted in this graph. Ev indicates volume elastic modulus; Ptr, transmural pressure; Ī·v, volume change ratio; mFMD, modified flow-mediated vasodilatation.

Table 1 Summary of the values of five parameters together with coefficients of determination (R2) for pre- and post-mFMD and those for all of the %_parameters calculated from each parameter value acquired throughout the finger and the radial artery elasticity measurements in each subject.

Table 1 summarizes the values of five parameters (Ī·v0, Ev0, Ī±, Kstif and kdist) together with coefficients of determination (R2) for pre- and post-mFMD, the mean, standard deviation (SD) and standard error of the mean (SEM) values for all %_parameters calculated from the data acquired throughout the finger and the radial elasticity measurements in each subject. Each participantā€™s gender, age and body mass index (BMI), together with SBP/MBP/DBP and PR obtained by the ArterioChecker, are also presented. The elasticity measurements of the finger artery of Sub. No. 05 and of the radial artery of Sub. No. 02, No. 04 and No. 11 were not performed due to the subjectsā€™ wishes and no data were listed for them in Table 1.

FigureĀ 2 shows bar graphs indicating the mean percentage changes in the five parameters of Ī·v0, Ev0, Ī±, Kstif and kdist (%_Ī·v0, %_Ev0, %_Ī±, %_Kstif and %_kdist) obtained in the finger (a) and the radial artery (b) of all volunteer subjects. Vertical lines in each parameter indicateā€‰Ā±ā€‰SEM, and the p-values of their changes from the baseline are also shown. Based on the statistical analysis, U-values and 1% or 5% significance levels were obtained respectively to be Uā€‰=ā€‰34.0 and pā€‰<ā€‰0.01 (pā€‰=ā€‰0.000) for %_Ī·v0, Uā€‰=ā€‰52.0 and pā€‰<ā€‰0.01 (pā€‰=ā€‰0.001) for %_Ev0, Uā€‰=ā€‰70.0 and pā€‰<ā€‰0.05 (pā€‰=ā€‰0.049) for %_Ī±, Uā€‰=ā€‰28.0, pā€‰<ā€‰0.01 (pā€‰=ā€‰0.000) for %_Kstif, and Uā€‰=ā€‰26.0 and pā€‰<ā€‰0.01 (pā€‰=ā€‰0.000) for %_kdist in the finger artery, and to be Uā€‰=ā€‰42.0 and pā€‰<ā€‰0.01 (pā€‰=ā€‰0.008) for %_Ī·v0, Uā€‰=ā€‰43.5 and pā€‰<ā€‰0.05 (pā€‰=ā€‰0.012) for %_Ev0, Uā€‰=ā€‰50.0 and NS (pā€‰=ā€‰0.052) for %_Ī±, Uā€‰=ā€‰12.0 and pā€‰<ā€‰0.01 (pā€‰=ā€‰0.000) for %_Kstif, and Uā€‰=ā€‰12.0 and pā€‰<ā€‰0.01 (pā€‰=ā€‰0.000) for %_kdist in the radial artery.

Figure 2
figure 2

Summarized bar graphs of mean % changes in five parameters in both finger and radial arteries. Bar graphs indicating the mean % changes in five parameters of Ī·v0, Ev0, Ī±, Kstif and kdist (%_Ī·v0, %_Ev0, %_Ī±, %_Kstif and %_kdist) obtained in the finger (a) and the radial artery (b) of all volunteer subjects. Vertical lines in each parameter indicateā€‰Ā±ā€‰SEM, and the p-values of their changes from the baseline are also shown. Ī·v0 indicates the volume change ratio at Ptr (transmural pressure)ā€‰=ā€‰0; Ev0, volume elastic modulus at Ptrā€‰=ā€‰0; Ī±, exponential coefficient related to a stiffness index; Kstif, effective stiffness index of arterial wall; kdist, distensibility (or vasodilatation) index of arterial wall.

Discussion

The aim of the present study was, first of all, to ask whether expected changes of arterial elastic properties in terms of stiffness and distensibility (or vasodilatation indices) of arterial wall in reaction to hyperemia can be obtained to be consistent with the clinically stated reaction in an FMD test, and, secondly, whether the present method is feasible to be used at home as a convenient assessment method of endothelial activity thereby substituting for such an FMD test.

With respect to the changes of elastic properties both in the finger and in the radial artery measured with the ArterioChecker, as shown in an example of Fig.Ā 1, the Ev(Ptr) curves determined approximately as an exponential function of Ptr in the pre-mFMD test shifted downwards in the post-mFMD, indicating the decrease of vascular stiffness in reaction to hyperemia over the range of Ptr regions. On the other hand, the Ī·v(Ptr) curves in the pre-mFMD shifted upwards in the post-mFMD, indicating the increase of vascular distensibility (i.e., vasodilatation responses at any Ptr levels) in reaction to hyperemia. In detail, it was clearly demonstrated, as shown in Fig.Ā 2, that the mean percentage changes of the parameters Ī·v0 (%_Ī·v0), Ev (%_Ev0), Ī± (%_Ī±), Kstif (%_Kstif) and kdist (%_kdist) in both finger and radial arteries showed, respectively, a statistically significant increase, a statistically decrease, a statistically decrease, a statistically significant decrease and a statistically significant increase. As a whole, it should be noted that the Ī·v0 value, the vasodilatation index in an arterial unloaded state that would be less reflected by smooth muscle22,23,24, showed a significant increase (pā€‰<ā€‰0.01), and that the Kstif value, the stiffness index of the arterial wall, showed a significant decrease (pā€‰<ā€‰0.01), while the kdist value, the vasodilatation index, showed a significant increase (pā€‰<ā€‰0.01) as compared to the baseline in reaction to hyperemia in both finger and radial arteries: In more detail, however, although the parameter values as well as the magnitudes of these percentage changes in increase or decrease varied considerably from person to person, the direction of the increase or decrease of the changes were almost the same, as shown in Table 1 and Fig.Ā 2. These findings suggest that either the finger- or radial-measurement could be adopted to evaluate the vascular endothelial function. It is also noted, however, that the operation of the method by use of the finger measurement is practically easier than that by use of the radial measurement. The reasons for this is that the attachment of the photo-plethysmographic sensor inside the wrist cuff precisely positioned over a radial artery is practically not so easy and required a little getting used to.

Taking these results into consideration, it is clear that peripheral vascular elastic properties in terms of stiffness (or rigidity) and distensibility (or vasodilatation degree) of the vascular wall, composed of elastic and collagen fibres and smooth muscle22,23,24, behave appropriately in reaction to hyperemia. These expected changes are consistent with clinical findings in an FMD test obtained in normal people3,10. It should be further noted with this method that the magnitudes of %_Ī·v0 or %_Ev0 and of %_Ī± could be reflected dominantly by the behaviour of elastic fibres and by that of collagen fibres and with smooth muscle, respectively.

It has been recommended that in an FMD test, in order to acquire a reliable assessment of vascular function, it is preferable to make continuous measurements of the brachial arterial diameter, using ultrasonography, over about a 2-min period, or more, until the peak vasodilatation is reached, although the reaction time following the cuff release may differ considerably among individuals3,10,25. The effect of this is that the %_FMD varies considerably from person to person, as is similar to the %_parameters determined by the new method. Furthermore, since the elasticity measurement time is approximately 30ā€“40Ā s, the measurement actually began 1Ā min after the cuff release. It should therefore be noted that the results obtained by the ArterioChecker might include some transient factors in reaction to hyperemia, which might also be one of the causative factors in data variability. Although this drawback due to the measurement technique is unavoidable, it is highly beneficial to be able to assess more reliably the vascular function through the use of various parameters as chosen and reported here.

In this study, the elasticity measurement was carried out in the finger and the radial artery as measuring sites distal from the upper-arm cuff occlusion used to arrest the circulation for the assessment of reactive hyperemia. In contrast, a brachial diameter measurement is performed at a measuring site proximal from the forearm cuff occlusion in a conventional FMD test. This difference in cuff occlusion site may not be a very important issue in assessing the response of reactive hyperemia. This is because, in fact, a peripheral vascular endothelial function test, so-called ā€œreactive hyperemia peripheral arterial tonometry (RH-PAT)ā€, using a fingertip with photo-plethysmography has also been commonly utilized under the clinical validation through a comparison study with a conventional FMD test26,27. Thus, in this respect the present assessment method, i.e., a modified FMD (mFMD) test, would be acceptable as one of the peripheral endothelial function tests.

Photo-plethysmography used in the present elasticity measurement is well known as a simple and convenient means to detect relative volume changes in biological tissue, but it cannot quantify the vascular volume (and its change) and of course the vessel diameter (and its change) as well. Therefore, as mentioned in the Appendix (a-2), the arterial volume change ratio (Ī·v(Ptr)ā€‰=ā€‰Ī”Va/\(\overline{V}_{{\text{a}}}\)(Ptr)) and arterial volume ratio (\(\overline{V}_{{\text{a}}}\)(Ptr)/V0) were also used to explain arterial elastic properties. The arterial volume change ratio (Ī·v(Ptr)) is related to the normalized pulse volume due to pulse pressure and the arterial volume ratio (\(\overline{V}_{{\text{a}}}\)(Ptr)/V0) concerns the mean intra-arterial pressure (P)ā€“volume (V) relationship (or strictly speaking, dynamic Pā€“V relationship), which reflects essentially the nonlinear visco-elastic behaviour of a blood vessel itself20,22,23,24,28,29,30.

Taking these points into consideration, we can calculate the pressure (Ptr)ā€‰āˆ’ā€‰diameter ratio (\(\overline{D}_{{\text{a}}}\)(Ptr)/D0) relationship using equation (A-7) (see Appendix (a-2)) when replacing \(\overline{V}_{{\text{a}}}\)(Ptr)/V0 (=ā€‰\(\uppi \cdot \overline{D}_{{\text{a}}}^{2}\)(Ptr)/4/\(\uppi \cdot \overline{D}_{0}^{2}\)/4)ā€‰=ā€‰\(\overline{D}_{{\text{a}}}^{2}\)(Ptr)/\(\overline{D}_{0}^{2}\)) with \(\overline{D}_{{\text{a}}}\)(Ptr)/D0 as,

$$\overline{D}_{{\text{a}}} (P_{{{\text{tr}}}} )/D_{0} = \surd\,\,(k_{{{\text{dist}}}} {-} \, (k_{{{\text{dist}}}} {-}{ 1})) \cdot {\text{exp}}({-}\alpha \cdot P_{{{\text{tr}}}} )$$
(1)

where \(\overline{D}_{{\text{a}}}\)(Ptr) is arterial diameter at Ptr, and D0 diameter at Ptrā€‰=ā€‰0. FigureĀ 3a shows examples of \(\overline{D}_{{\text{a}}}\)(Ptr)/D0ā€‰āˆ’ā€‰Ptr curves calculated by Eq.Ā (1) using the parameters kdist and Ī± of mean values obtained in the radial artery of all participants before (pre-mFMD: the measured mean values were kd_preā€‰=ā€‰1.59 (SEMā€‰=ā€‰0.0546), Ī±preā€‰=ā€‰0.0487) and after cuff occlusion (post-mFMD: those were kd_postā€‰=ā€‰2.14 (SEMā€‰=ā€‰0.115), Ī±postā€‰=ā€‰0.0443). In this graph, the values of Ī±pre and Ī±post are constant, and the curves with the values of kd_pre and kd_post as well as thoseā€‰Ā±ā€‰SEM are shown. If the D0 (or V0) values before and after cuff occlusion are nearly the same, %_\(\overline{D}_{{\text{a}}}\)(Ptr)/D0ā€‰āˆ’ā€‰Ptr relationships can also be calculated as shown in Fig.Ā 3b: It is naturally noted that %_\(\overline{D}_{{\text{a}}}\)(Ptr)/D0 values are largely dependent upon the distending pressure of the artery, that is, Ptr or MBP when Pcā€‰=ā€‰0. The mean %_\(\overline{D}_{{\text{a}}}\)(Ptr)/D0 value at higher Ptr region is about 16% (mean %_\(\overline{V}_{{\text{a}}}\)(Ptr)/V0 value is about 25.6%). The value of about 16% increase in the diameter is considered to be comparable to the value of %_FMD (from 6ā€“7 to around 25%) obtained in normal people3,10. This agreement seems to support the value of this method for noninvasive assessment of peripheral vascular endothelial function.

Figure 3
figure 3

Diameter ratio (\(\overline{D}_{{\text{a}}}\)(Ptr)/D0)ā€‰āˆ’ā€‰Ptr and %_\(\overline{D}_{{\text{a}}}\)(Ptr)/D0ā€‰āˆ’ā€‰Ptr relationships. (a) Diameter ratio (\(\overline{D}_{{\text{a}}}\)(Ptr)/D0)ā€‰āˆ’ā€‰Ptr curves calculated by Eq.Ā (1) using the measured parameters kdist and Ī± of mean values obtained from all subjects before (pre-mFMD: kd_preā€‰=ā€‰1.59 (SEMā€‰=ā€‰0.0546), Ī±preā€‰=ā€‰0.0487) and after cuff occlusion (post-mFMD: kd_postā€‰=ā€‰2.14 (SEMā€‰=ā€‰0.115), Ī±postā€‰=ā€‰0.0443). In this graph, the values of Ī±pre and Ī±post are constant, and the curves with the values of kd_preā€‰Ā±ā€‰SEM and kd_postā€‰Ā±ā€‰SEM are also shown. (b) Mean %_\(\overline{D}_{{\text{a}}}\)(Ptr)/D0ā€‰Ā±ā€‰SEM vs Ptr relationships calculated from the diameter ratio (\(\overline{D}_{{\text{a}}}\)(Ptr)/D0)ā€‰āˆ’ā€‰Ptr curves of pre- and post-mFMD. Ī±pre and Ī±post indicate exponential coefficients before and after occlusion, respectively; kd_pre and kd_post, distensibility indices of arterial wall before and after occlusion, respectively; Ptr, transmural pressure; mFMD, modified flow-mediated vasodilatation; SEM, standard error of the mean.

The FMD with Doppler ultrasound has currently emerged as the most popular clinical method for assessing vascular endothelial function in a noninvasive manner. However, for the accurate and reliable assessment of vascular function it is essential, firstly, to have access to appropriate ultrasonic equipment and this must be in conjunction with the high level of skill required for the precise and continuous measurement of vessel diameter. Further to these technical issues, it is noted that the vessel diameter measured is dependent upon mean blood pressure (MBP) at that time: Strictly speaking, it should therefore be considered that the value of %_FMD should be compensated for the same level of MBP. Our new method does not need to consider this point and so this is an important feature, because the measures of Ev(Ptr) and Ī·v(Ptr) can be determined simply as a function of transmural pressure (Ptrā€‰=ā€‰MBPā€‰āˆ’ā€‰Pc) and the five parameters of Ī·v0, Ev0, Ī±, Kstif and kdist derived experimentally and analytically are independent of MBP.

Compared with a conventional ultrasound assessment of endothelial cell function by an FMD test in the brachial artery, our new method described herein is much simpler and more convenient to use without any advanced skill being required either in the finger or in the radial artery; it is similar to dealing with a home-sphygmomanometer. As mentioned in the Introduction section, if a new technology with similar performance to the FMD test can be used by anyone at home, it could be useful for early detection of vascular dysfunction. Such a technology could also be helpful for a person with early arteriosclerosis to carry out a follow-up examination as part of his/her lifestyle improvement or monitoring their therapeutic drug programme. Further comparison experiments with the FMD test are now needed using a larger number of subjects, including patients with cardiovascular diseases. Nevertheless, the present experimental results obtained at this stage of development show the potential for offering a useful and convenient assessment method for peripheral vascular functional reactivity. This method has a considerable advantage of simultaneous acquisition of both degree of vasodilatation and arterial elasticity as well as of behaviour of elastic and collagen fibres and smooth muscle as components of the vessel wall. It is also important and interesting to investigate the effects of vascular tone due to smooth muscle using a nitroglycerin-mediated dilation with the present method.

Methods

Evaluation parameters related to arterial elasticity (see Appendix (a-1)ā€“(a-3))

A brief review of the arterial elastic properties from a biomechanical viewpoint is given in the Appendix (a-1). In this study we adopted the volume elastic modulus (Ev(Ptr)) and the volume change ratio (Ī·v(Ptr)\()\) measured photo-plethysmographically to determine as an exponential function of arterial distending pressure (or transmural pressure (Ptrā€‰=ā€‰MBPā€‰āˆ’ā€‰Pc; (MBP is calculated by a well-known formula as DBPā€‰=ā€‰(3Ā·MBPā€‰āˆ’ā€‰SBP)/231)) as shown in Table 2. The measurement principle based on the photo-plethysmography technique is presented in the Appendix (a-2). Using experimentally and analytically derived measures (see also Appendix (a-2)), the five constant parameters, which were related to elastic property factors and independent of MBP (or Ptr), were used as elasticity evaluation indices; these are summarised in Table 2, where physiological interpretations of these parameters are explained in detail (see also Appendix (a-3)).

Table 2 Expression of volume change ratio (Ī·v(Ptr)) and volume elastic modulus (Ev(Ptr)) as an exponential function of transmural pressure (Ptr) derived photo-plethysmographically and used as arterial elastic measures in the present study, with five evaluation parameters related to arterial elasticity and vasodilatation and their physiological interpretations.

Apparatus

A detailed description of the smartphone-based instrument (named ā€œArterioCheckerā€) for the finger and the radial artery used in this study has been reported elsewhere17. Briefly, for the elasticity measurement in the finger or the radial artery, a finger cuff (width of 30Ā mm) or a wrist cuff (width of 70Ā mm) was used, inside of which was embedded a photo-plethysmographic sensor to detect the light intensity (I), i.e., photo-plethysmogram (PPG): This sensor consisted of a near-infrared light emitting diode (LED: SMC810; peak emission wavelength 810Ā nm, Epitex Inc., Kyoto, Japan) as a light source and a photo-diode (PD: S6775-01, Hamamatsu Photonics K.K., Hamamatsu, Japan) as a photodetector.

For a finger-measurement, the occlusive cuff was wrapped around the basal phalanx of a right index finger. The LED and the PD were positioned inside the cuff on opposite sides of the finger. Then, for a wrist-measurement, the occlusive cuff was wrapped around the left wrist, inside of which the LED and the PD were fixed 20Ā mm apart from the center of each element so as for a radial artery to be placed between them.

Bluetooth communication was made between the ArterioChecker and the smartphone (iPhone) operated by an experimental app with appropriate software. This was usable by iPhone 7 and later devices with iOS 13.3 or subsequent upgrades (Apple Inc., Cupertino, USA). The program had a number of functions: Firstly, it allowed transmission of all of the analog-to-digital converted signals (cuff pressure Pc and PPG signals) from the ArterioChecker to the iPhone. Secondly; it carried out the control of the measurement start/stop process of the iPhone, with real-time display of signal waveforms; Thirdly, it carried out signal processing and calculations of various quantities. These were as follows: PPG raw signal, SBP/MBP/DBP, pulse rate (PR), Ptr (MBPā€‰āˆ’ā€‰Pc), Ī·v(Ptr) (equation (A-2) in the Appendix (a-2); same as below), Ev(Ptr) (equation (A-3)), \(\overline{V}_{{\text{a}}}\)(Ptr)/V0 (equation (A-6)), Kstif (equation (A-10)) and kdist (equation (A-9)). Data were stored in the iPhone and also stored and managed on a cloud server provided for this study. The five parameters Ī·v0, Ev0, Ī±, Kstif and kdist thus obtained were used for the data analysis.

Overview of the elasticity measurement system for the finger and radial artery was given in Figure A2 in the Appendix (a-2).

Participants and ethics

A total of 15 volunteers (eight males and seven females) participated in this pre-clinical human study. All were Japanese, 20ā€“74Ā years old, living their normal daily life in Tokyo, Osaka, Kobe and Sapporo cities. Three of the 15 participants (No. 07, No. 08 and No. 15 in the Results section of Table 2) were suffering from grade I hypertension and had been taking antihypertensive agents every day. The other participants from the age of 20 to 55 had no current cardiovascular disease and did not take any prescription medications, although two participants (No. 09 and No. 13), who have had normal daily lives, appeared to have hypotension.

Written informed consent was obtained from the participants after they were provided with a complete description of the study, inclusive of measurement procedures. This study was approved by an institutional review board of the ethics committee of Asahikawa Medical University (reference approval number 19039) and conducted according to the principles expressed in the Declaration of Helsinki. This was not a replicated study.

Measurement procedures

All experiments were carried out in a room temperature of around 24 to 25 ā„ƒ and humidity of approximately 50%. Throughout the measurements, the participant remained as motionless as possible, sitting for one minute at rest in a chair with his/her left hand placed on a desk at heart level before starting the measurements. A brachial cuff used for both BP measurement and arterial occlusion, together with the wrist cuff for the radial elasticity measurement, were attached to the participantā€™s left upper arm and the wrist, respectively. These, together with the finger cuff for the finger elasticity measurement, were attached to the participantā€™s right arm and the finger, respectively. Before the experiment in each participant the brachial BP was measured by a conventional sphygmomanometer (NISSEI DSK-1051, authenticated by the European Society of Hypertension; Japan Precision Instruments, Inc., Japan), and SBPc/DBPc and pulse rate (PRc) values were recorded as a reference.

As mentioned in the Introduction section, the usual FMD test starts by measuring the inner diameter of the brachial artery before and after a 5-min supra-systolic cuff occlusion of the forearm. However, since the measuring site was the finger or the radial artery, the brachial cuff occlusion was carried out in this study to allow for the possibility of similar effects of reactive hyperemia in the peripheral vasculature, as would be with peripheral vascular endothelial function testing26,27. We will therefore call this a modified FMD (mFMD) test hereafter.

Prior to the elasticity measurement in the finger and the radial artery, three repeated tests were carried out to evaluate the repeatability of the five parameters in each participant. Mean relative errors (percentage changes from the averaged values) of the parameters Ī·v0, Ev0, Ī±, Kstif and kdist were respectively less than 3.0, 5.0, 4.0, 5.0 and 2.0% for the finger and less than 5.0, 5.0, 3.0, 5.0 and 5.0% for the radial artery, indicating acceptable repeatability of each parameter.

The measurement procedures were as follows: After 1Ā min of rest at the end of BP measurement, the radial (or finger) elasticity measurement was made as a control (baseline) condition three times to confirm almost the same Ev(Ptr) and Ī·v(Ptr) data, obtaining averaged data as a pre-mFMD test. Then, the connection of the brachial cuff was changed from the sphygmomanometer to a handmade cuff-inflation/deflation device and the cuff pressure was increased approximately 50Ā mmHg higher than individual SBPc level during 5Ā min to completely arrest the circulation in the arm. The rising/falling times of the cuff pressure from 0/200 to 200/0Ā mmHg were less than 0.5Ā s. In 1Ā min after release of the cuff occlusion, the radial (or finger) elasticity measurement was done only once to obtain the data as a post-mFMD test.

After taking a 10Ā min break at the end of the radial (or finger) elasticity measurement, the finger (or radial) elasticity measurement was carried out with the same procedure as for the radial (or finger) measurement. Note that the order of the finger and the radial measurement was counter-balanced for each individual. If a subject did not wish to proceed with the subsequent supra-systolic cuff occlusion for the elasticity measurement due to discomfort and/or feeling unwell, either the finger or the radial measurement was done accordingly.

Data analysis

The measured Ev(Ptr) and Ī·v(Ptr) data that approximated to an exponential function (i.e., Ev(Ptr)ā€‰=ā€‰Ev0Ā·exp(Ī±Ā·Ptr); Ī·v(Ptr)ā€‰=ā€‰Ī·v0Ā·exp(āˆ’ā€‰Ī±Ā·Ptr)) with greater than 0.9 of the coefficient of determination (R2), were adopted for analysis in each subject. Actually, it is noted that the R2 values were over 0.9 in all of the elasticity measurements both in the finger and in the radial artery.

The changes in the five parameters, Ī·v0, Ev0, Ī±, Kstif and kdist, as a consequence of the reactive hyperemia, were compared to the baseline parameters and calculated as a percentage of these baseline parameters (%_Ī·v0 (=ā€‰((Ī·v0_postā€‰āˆ’ā€‰Ī·v0_pre)/Ī·v0_pre)ā€‰Ć—ā€‰100), %_Ev0 (=ā€‰((Ev0_postā€‰āˆ’ā€‰Ev0_pre)/Ev0_pre)ā€‰Ć—ā€‰100), %_Ī± (=ā€‰((Ī±postā€‰āˆ’ā€‰Ī± pre)/Ī±pre)ā€‰Ć—ā€‰100), %_Kstif (=ā€‰((Ks_postā€‰āˆ’ā€‰Ks_pre)/Ks_pre)ā€‰Ć—ā€‰100) and %_kdist (=ā€‰((kd_postā€‰āˆ’ā€‰kd_pre)/kd_pre)ā€‰Ć—ā€‰100)) in each and in all of the subjects. Subscripts ā€œpreā€ and ā€œpostā€ mean respectively before (baseline) and after cuff occlusion. The statistical significance of the mean percentage changes from the baseline in each parameter obtained in all subjects was also examined by the nonparametric Mannā€“Whitney U-test for the comparison between two conditions, i.e., pre-mFMD and post-mFMD. For data that passed tests, data were expressed as meansā€‰Ā±ā€‰SEM. All p-values are two-sided, and pā€‰<ā€‰0.05 and pā€‰<ā€‰0.01 were considered as a statistically significant difference. The analyses were carried out using Microsoft Excel and IBM SPSS Statistics 19.0 (Stats Guild Inc., Chiba, Japan).

Role of the funding source

The funder of the study had no role in study design, data collection, data analysis, data interpretation, writing of the report, or the decision to submit for publication.