Abstract
In times of collective concern about pandemics, body-shield resuscitation barrier devices are more and more considered to protect against transmission of different pathogens between rescuers and patients. The objective of this experimental study was to investigate the characteristics of blood drops dispersed on the surface of four different foils suitable for blanketing patients during resuscitation. We analyzed run-off characteristics of blood stains depending on surface properties of polyvinyl chloride, polyethylene, polyethylene terephthalate and aluminum-coated polyethylene terephthalate. Confocal fluorescence microscopy revealed less cellular density and lack of fibrin networks in blood stains on the four foil surfaces than on paper towel. Delayed clotting went along with larger areas of contamination indicating a greater likelihood of coming into contact with potential germs but a smaller chance of contracting an infection. Space blankets as obligatory components of first aid kits are readily available for rescuers and serve as a mechanical barrier between rescuers and patients during resuscitation.
Similar content being viewed by others
Introduction
Unprotected rescuers are exposed to numerous pathogens during cardiopulmonary resuscitation (CPR), either through direct contact with the patient’s blood, body fluids or with droplets and aerosols1. Despite the ubiquitous risk of infection with human immunodeficiency virus (HIV), hepatitis B virus (HBV) and hepatitis C virus (HCV) through blood, current attention is focused mainly on transmission of viruses by aerosols and droplets during CPR1.
During the corona virus disease (COVID-19) pandemic, the fear of contracting severe acute respiratory syndrome-associated corona virus (SARS-CoV-2) has become the prime reason for lay rescuers to not perform CPR and for bad resuscitation outcome2,3. Consequently, the European Resuscitation Council omitted ventilation attempts by lay rescuers and recommended compression-only CPR and public-access defibrillation (PAD) during the COVID-19 pandemic4. Although the potential for contracting HIV, HBV or HCV via saliva through mouth-to-mouth ventilation is extremely low, viruses can be contracted by direct contact with blood and wounds5. Since even the smallest volumes of blood in infected persons may contain infectious viruses, the quantity of exposed blood plays a role in relation to the likelihood of rescuers being infected through direct contact6.
Basic infection-control measures such as plastic shields as a barrier between patient and rescuer are expected to diminish contaminations from vomit, blood and body secretions7. Presumably, the strong hydrophobicity of plastic foils may have additional effects on infectious secretions beside the barrier function8. Foils may even reduce revulsion, thus increasing the readiness of lay rescuers to perform CPR9.
The object of this experimental study was to investigate the dispersion of fresh blood stains on the surface of four different foils suitable for blanketing patients during CPR.
Results
Foil fragments of either PVC, PE, PET, and Al-PET were investigated according to clotting and run-off speed from 30 to 100 µl drops of untreated blood. Six blood drops freshly drawn from a co-author were investigated per foil and volume (total 48 runs).
CFM investigation
Real time live confocal imaging revealed fibrin formation only in the paper towel controls but not in the foils (Fig. 1).
Run-off area and run-off speed
The diameter of 30 µl and 100 µl drops and lengths of run-off drainage marks at 2, 7, 10, 20 and 30 min were measured (Fig. 2). Overall, lengths of drainage marks and run-off speed were higher with 100 µl drops than with 30 µl drops. Drainage marks in 30 µl drops ranged between 8 and 75 mm with few alterations after 10 min. Similarly, in 100 µl drops drainage marks exceeded 300 mm on all tested foils within the first 10 min. After an initial continuous drainage of 100 µl drops the further run-off occurred gradually in intervals showing few alterations between 20 and 30 min. Run-off speed within 2 min ranged between 2.7 and 2.8 mm/s. Clotting was significantly delayed and the ratio between end drop area and run-off area as surrogate of presumed infectivity was diminished to approx. 1% on all tested foils (Table 1).
Discussion
To our knowledge this is the first experimental study to compare different materials suitable for barrier resuscitation regarding their potential to diminish exposure to blood. As the quantity of exposed blood differs within the run-off figure of a blood stain, the likelihood of infection depends on direct contact with either blood drops or large run-off areas. Even before the pandemic began, fear of infection was a significant reason for lay rescuers to not perform CPR10. The use of personal protective equipment (PPE) to minimize the risk of infection was of utmost importance during the pandemic11. Unfortunately, accessibility to PPE is limited to professional rescue teams. Thin foils were established in first aid kits or at public defibrillators for the purpose of protecting lay rescuers performing resuscitation.
Our experimental results revealed that clotting was delayed on all tested materials as shown by the lack of fibrin networks in comparison to paper towels. Initial blood drops differed in area and diameter according to the different hydrophobicity of foils. We consider hydrophobicity and delayed clotting to be related to surface properties of the test materials8,12. Since the prosthetic blood interaction depends on the nature of the surface, the wide choice of additives, ingredients, plasticizers, fillers, and stabilisers used with PVC affects blood coagulation12. Lowkis and Szymomowicz reported that fresh, human citrate blood on PVC had a markedly prolonged thromboelastographic coagulation time and time of clot formation to standard elasticity. The authors discussed the potential effect of electric charge on the adhesion of human blood platelets13. Corresponding with our findings, coating with aluminum oxide (Al2O3) was reported to further increase anticoagulation effects14.
Our experimental investigation showed that, on the one hand, delayed clotting on different foil surfaces goes along with larger areas of contamination, which increases the chance of coming into contact with potential germs. On the other hand, the lesser content of germs per area decreases the likelihood of contracting an infection. The foil as mechanical barrier between rescuer and patient provides an additional protective effect. Length of run-off marks was more pronounced on both sides of the space blanket, indicating that smooth surface texture influenced dispersion of blood more than did composition of foils.
There should be a diminished risk of contamination from blood when the foil is used as barrier between rescuer and patient. It is likely that the faster drainage of blood on the space blanket leads less germ load to be expected on exposed sites, thus reducing the risk of contamination for rescuers. In addition, survival of human coronaviruses was reported to be significantly shorter on aluminum surfaces than on PVC surfaces15.
Blanketing the patient with a foil can protect against contamination by blood, vomit and secretions. Whereas PVC and PE foils may not be readily available to lay rescuers, space blankets as basic components of first aid kits are. Ideally, space blankets used for thin film barrier resuscitation would need a transparent center to facilitate finding the correct pressure point for resuscitation. For professional rescuers foils with a predesigned central perforation could help ensure air-tight installation of mask or tube with filter when ventilation is attempted. Eventually, invasive procedures with direct blood contact e.g. venipuncture or chest drainage are allocated to professional staff additionally equipped with PPE11. Presumably, covering the patient with a space blanket can reduce the risk of contamination and may prompt an even greater willingness of lay rescuers to start resuscitation. This, however, deserves further clinical investigation.
There are several limitations to be accounted. We performed preliminary investigations comparing blood on glass slides and on test slides to detect potential interference from the interventions. Blood was not evenly spread in run-off figures. This was particularly obvious on space blankets with their package-dependent longitudinal and transverse folds. This made dispersion of blood more unpredictable. We are aware that infectivity depends on exceeding the lower threshold of germ load and that it does not necessarily increase in a linear manner with increasing germ load. For the sake of simplification, let us state that we postulated germs to be distributed equally in blood fluid and germ load to linearly correlate with the estimated area of contamination.
In conclusion, thin film barrier resuscitation could protect against pathogens transmitted by skin contact, aerosol or droplet infection. On the one hand, delayed clotting might increase the risk of contamination from direct contact. On the other hand, the lesser concentration of germs in the run-off area that goes along with increased fluid surface and decreased volume might indicate less risk of infection. Applications and mode of handling by lay rescuers need to be investigated to determine the optimal choice of barrier material. It would be worthwhile to see whether a barrier between patient and rescuer might even increase the willingness of lay rescuers to start CPR.
Materials and methods
Study design and ethics considerations
Run-off characteristics of blood drops on various foil surfaces were macroscopically observed. In addition, confocal fluorescence microscopy (CFM) was used to microscopically analyze cellular density and coagulation activity of blood stains on different foil surfaces. Investigations were conducted at the Anesthesiology Research Laboratory, Medical University of Innsbruck, Austria. Study design and manuscript presentation followed the SQUIRE 2.0 Revised Standards for Quality Improvement Reporting Excellence to improve the quality, safety, and value of healthcare16. As the experimental design focused on physical properties of plastic foils and not on patients and users the institutional guidelines of the OE Clinical Trial Center (OE CTC), Medical University of Innsbruck, did not advocate the need for ethics approval.
Context
Surface properties of four different materials potentially applicable for blanketing patients during CPR were investigated.
Intervention
Specifics of the team involved anaesthesiology and critical care (Lederer W, Hermann M; Bilgeri V) and mountain rescue (Isser M, Klinger A).
Test materials
We investigated four easily accessible foils, waterproof, low-weight, low-bulk with varying transparency and surface structure.
-
(1)
Plasticised polyvinyl chloride (PVC) foil (d50 universalpresenning; JUFOL GmbH, Krumbacher Str. 9, D-86154, Augsburg, Germany), smooth surface, with an area of 4000 × 5000 cm and a thickness of 50 μm.
-
(2)
Polyethylene (PE) foil (Henry Schein Inc., 135 Duryea Rd., Melville, NY 11747, USA), rough and structured surface, with an area of 120 × 120 cm.
-
(3)
Polyethylene terephthalate (PET) surface and
-
(4)
1% aluminum-coated PET blanket (Al-PET) surface of a space blanket (First Aid Blanket; M06299; Franz Kalff GmbH, D-53881 Euskirchen, Germany), smooth surface, with an area of 160 × 210 cm, 0.012 mm in thickness, approx. 50 g in weight.
-
(5)
Bleached pulp, absorbent, rather bulky with rough and structured surface (Paper towel; Zentrale Handelsgesellschaft GmbH, D-77656 Offenburg, Germany) served as control.
Measuring device
CFM (Zeiss Observer Z1, Zeiss, Oberkochen, Germany) in arrangement with a spinning disc confocal system (UltraVIEW VoX, Perkin Elmer, Waltham, MA, USA) was used to perform high-resolution imaging from blood contamination on four surfaces of selected barrier materials. The technique combined a focused laser for excitation with a pinhole for detection for three-dimensional imaging. It provided high-sensitivity, low-phototoxicity and low noise hybrid detection17.
Experimental setup
In a self-constructed physical model we attached four longitudinal 200 × 6000 mm stripes of PVC, PE, PET, and Al-PET to a frame. Three freshly drawn blood drops per stripe were applied on the upper edge using an Eppendorf pipette. Then the frame was brought to a vertical position and the run-off characteristics of the blood stains were observed at 2-, 5-, 7-, 10-, 20- and 30-min intervals. The investigation was performed for 30 µl drops and for 100 µl drops and repeated once.
In a different experimental setting we used CFM to microscopically visualize changes in stained blood drops after run-off on the surface of the four tested materials. For preparation of the blood live stain mixture we added 10 μg/ml fluorescently labeled wheat germ agglutinin Alexa Fluor 555 (Molecular Probes, Waltham MA, USA), Hoechst (Molecular Probes) and 4.5 μg/ml fluorescent fibrin-binding peptide solution (FFBP; Genic Bio, Shanghai, China) to 1 ml of fresh whole blood18. Similar to the initial experiment the run-off areas from 30 µl drops and 100 µl drops were investigated. At the end of the test run 24 × 50 mm specimens of the four shields were taken, mounted on microscope slides (Menzel-Gläser, Vienna, Austria) and microscopically examined with real time live confocal imaging at three wavelengths (405/488/561 nm). Blood drops trickled on paper towel served as control. Density and dispersion of fibrin networks and red blood cells were compared. Observations were repeated three times per foil. Investigations were made at room conditions of 31.4% relative humidity and 23.3 °C temperature.
Study of the interventions
The run-off characteristics of 30 µl and 100 µl untreated blood drops on PVC, PE, PET, and Al-PET surfaces were analyzed by measuring lengths and width of drainage marks (in mm) 2, 7, 10, 20 and 30 min after the frame was brought to a vertical position. The initial run-off speed was calculated (in mm/s) during the first two minutes of observation. After 30 min we measured run-off mid-width and length as well as initial and end drop diameters of each blood stain. The corresponding areas for the entire run-off (mid-width × length), and the initial drop and the end drop ((diameter/2)2 × π) were calculated (in mm2).
Assuming equal distribution of germs in blood fluid during infection, the germ load per area is higher in accumulated blood drops than in dispersed blood after run-off. In order to estimate the relative germ load in the run-off area (Arun) we calculated the reduced area of the end drop related to the area of the initial drop. The ratio of the area of the end drop (Aend) to the area of the initial drop (Ainit) indicates the reduced area of contact (AC) with high germ load (1). AC values were multiplied by 100 to express results as a percentage.
The ratio of initial drop area to run-off area was calculated to indicate the area gain (AG) of increased contamination (2). AG was expressed as a percentage.
Area equivalent (AE) ratio was calculated by ratio multiplication as the product of AC multiplied by AG. Based on the assumption that germs are equally distributed in blood fluid, the sum germ load on run-off surfaces and in the end drop equals germ load in the initial drop. Thus, the AE ratio indicates the reduced germ load per area as surrogate of presumed infectivity (3). The hypothetical germ load after run-off was expressed as a percentage.
Via CFM we visualized fibrin and cellular structures in real time using a 10 × objective for the overview and a 40 × water immersion objective with a numerical aperture of 1.4 for a more detailed view. On PET and Al-PET we used bright field microscopy (objective 40 × water immersion) to visualize erythrocytes. We used HOECHST staining to visualize nuclei of leukocytes and lymphocytes. Extrinsic pathway of coagulation was induced by adding STAR-TEM and EXTEM reagents from rotational thromboelastometry (ROTEM).
Measures
Rationale for choosing the four materials was their availability in the out-of-hospital setting. In accordance with sample size estimation six test runs were performed per foil and per selected volume of 30 µl and 100 µl blood. With respect to material and volume of the initial blood drop we observed little variation in the data. Prior to the investigation we performed preliminary investigations comparing blood on glass slides and on test slides to detect potential interference from the interventions. Elements that contributed to, influenced and altered efficiency were humidity and room temperature.
Statistical analysis
Descriptive statistics were applied using the statistical package for the social sciences (IBM SPSS Statistics Standard Version 27, Armonk, NY, USA) to determine mean and standard deviation. The estimated sample size was minimum six runs per group to achieve a power of 80% (power = 0.8) and a level of significance of 5% (alpha = 0.05) with one-way analysis of variance for means between 250 and 30 and standard deviation of 25 using an online sample size calculator (https://homepage.univie.ac.at/robin.ristl/samplesize.php?test=wilcoxon).
Data availability
All data generated or analysed during this study are included in this published article.
References
Fragkou, P. C. et al. Transmission of Infections during Cardiopulmonary Resuscitation. Clin. Microbiol. Rev. 34(4), e0001821. https://doi.org/10.1128/CMR.00018-21 (2021).
Baldi, E., Contri, E., Savastano, S. & Cortegiani, A. The challenge of laypeople cardio-pulmonary resuscitation training during and after COVID-19 pandemic. Resuscitation 152, 3–4. https://doi.org/10.1016/j.resuscitation.2020.04.040 (2020).
Marijon, E. et al. Out-of-hospital cardiac arrest during the COVID-19 pandemic in Paris, France: A population-based, observational study. Lancet Public Health. https://doi.org/10.1016/S2468-2667(20)30117-1 (2020).
Nolan, J. P. et al. European Resuscitation Council COVID-19 guidelines executive summary. Resuscitation 153, 45–55 (2020).
Arend, C. F. Transmission of infectious diseases through mouth-to-mouth ventilation: Evidence-based or emotion-based medicine?. Arq. Bras. Cardiol. 74(1), 73–97 (2000) (English, Portuguese).
Thompson, S. C., Boughton, C. R. & Dore, G. J. Blood-borne viruses and their survival in the environment: Is public concern about community needlestick exposures justified?. Aust. N. Z. J. Public Health. 27(6), 602–607 (2003).
Mehta, H. J., Patterson, M. & Gravenstein, N. Barrier enclosure using a Mayo stand and plastic sheet during cardiopulmonary resuscitation in patients and its effect on reducing visible aerosol dispersion on healthcare workers. J. Crit. Care. 59, 108–111. https://doi.org/10.1016/j.jcrc.2020.05.007 (2020).
Fu, L., Li, J., Wang, G., Luan, Y. & Dai, W. Adsorption behavior of organic pollutants on microplastics. Ecotoxicol. Environ. Saf. 217, 112207. https://doi.org/10.1016/j.ecoenv.2021.112207 (2021).
Lederer, W. & Isser, M. Barrier resuscitation by lay rescuers during COVID-19 pandemic. Med. Hypotheses https://doi.org/10.1016/j.mehy.2021.110648 (2021).
Savastano, S. & Vanni, V. Cardiopulmonary resuscitation in real life: The most frequent fears of lay rescuers. Resuscitation 82(5), 568–571 (2011).
Seto, W. H. et al. Advisors of Expert SARS group of Hospital Authority. Effectiveness of precautions against droplets and contact in prevention of nosocomial transmission of severe acute respiratory syndrome (SARS). Lancet 361(9368), 1519–1520 (2003).
Kurian, G. & Sharma, C. P. Surface modification of polyvinyl chloride towards blood compatibility. Bull. Mater. Sci. 6(6), 1087–1091 (1984).
Lowkis, B. & Szymonowicz, M. Effect of an electric charge on athrombogenic properties of polyvinyl chloride foil. Polim. Med. 13(3–4), 151–160 (1983).
Huang, Z., Zhang, W., Yu, J. & Gao, D. Nanoporous alumina membranes for enhancing hemodialysis. J. Med. Dev. 1(1), 79–83 (2007).
Kampf, G., Todt, D., Pfaender, S. & Steinmann, E. Persistence of coronaviruses on inanimate surfaces and their inactivation with biocidal agents. J. Hosp. Infect. 104(3), 246–251 (2020).
SQUIRE 2.0 Revised Standards for Quality Improvement Reporting Excellence. http//:SQUIRE|SQUIRE 2.0 Guidelines (squire-statement.org).
Chen, J. T. et al. Confocal laser scanning microscopy: I. An overview of principle and practice in biomedical research. Acta Anaesthesiol. Taiwan 42, 33–40 (2004).
Weiss, N. et al. FITC-linked Fibrin-Binding Peptide and real-time live confocal microscopy as a novel tool to visualize fibrin(ogen) in coagulation. J. Clin. Transl. Res. 3(2), 276–282 (2017).
Author information
Authors and Affiliations
Contributions
H.M. and L.W. contributed to the concept and design, analysis, interpretation of the data and writing of the manuscript. I.M., B.V. and K.A. contributed to literature research, interpretation of the data and writing of the manuscript. All authors approved the final version.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Hermann, M., Isser, M., Bilgeri, V. et al. Thin foil body-shield resuscitation barrier device to protect from blood: an experimental study. Sci Rep 12, 13573 (2022). https://doi.org/10.1038/s41598-022-17915-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-022-17915-4
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.