Abstract
Clinicians often perform pumping of infusions with a syringe (PIS) to quickly deliver fluid or blood transfusion to patients, especially during an emergency. Despite the efforts of the clinicians, critically ill patients are prone to acquire catheter-related bloodstream infections. Although clinicians have reported the possibility of PIS contamination, no group of researchers has studied nor confirmed this possibility. Here, we examined whether PIS can cause bacterial contamination of the fluid inside the syringes, using microbiological tests, including the analysis Escherichia coli DH-5 alpha growth by measuring the absorbance at OD600. We confirmed that contamination of fluid in the barrel was almost proportional to the applied volume of bacterial fluid. Aliquots of DH-5 alpha artificially applied on the surface of the gloved hand of an examiner, the plunger or the inner side of the barrel of a syringe could permeate inside the syringe. Furthermore, disinfection with ethanol before PIS almost successfully prevented bacterial multiplication. Our findings suggest that PIS can cause intraluminal contamination when performed with unsterilized hands, and that previous disinfection with ethanol can effectively prevent PIS-induced contamination. These results highlight the risk of PIS-induced contamination and the importance of disinfection in the daily clinical practice.
Similar content being viewed by others
Introduction
In the perioperative practice at an emergency department or intensive care unit, clinicians, such as emergency physicians, anesthesiologists, intensivists, and nurses, encounter critically ill patients requiring rapid diagnosis and resuscitation, such as those undergoing a cardiovascular surgery, or suffering from a hypovolemic shock caused by traumatic, massive, and acute gastrointestinal bleeding, or a septic shock caused by acute pan-peritonitis1. In order to treat these critically ill patients, healthcare providers often perform pumping infusions with a syringe (PIS) to intravenously and rapidly deliver a large volume of fluid or perform blood transfusions2. While performing PIS, a disposable sterile syringe should be carefully loaded at the lumen side of the intravenous line of the patient to avoid contamination of the solution by direct contact with the hands of the clinician or unsterilized gloves, according to the recommended good clinical practices3,4. However, when clinicians treat critically serious patients with refractory shock vital signs, appropriate aseptic maneuvers including hand washing or disinfection with ethanol tend to be left aside because of the emergency and shortage of time. Catheter-related blood stream infections (CRBSI) account for about twenty percent of total nosocomial bacterial infections5, which can substantially decline the perioperative health status of the patients6,7,8,9,10. In general, the four main causes of CRBSI are the presence of skin organisms, contaminated catheter hubs and infusates, and hematogenous infections from distant sites11,12. Although CRBSI mostly occurs in patients who received a central rather than a peripheral venous catheter, the latter can also cause CRBSI and related severe complications13. On the other hand, even though most clinicians believe that PIS itself constitutes a risk of contamination, no report has yet demonstrated that it can cause intraluminal contamination, or that patients who receive PIS during perioperative treatments are more likely to contract CRBSI than those who do not. A previous study showed that the use of a "blood conservation system," in which the syringe had a plastic cover that prevented its physical contact with the hand of the clinician during blood sampling, has lower chances of causing intraluminal contamination than that of a three-way stopcock catheter14; however, the precise mechanism of contamination was not elucidated. A few studies have shown PIS-induced contamination by hazardous drugs, such as chemotherapeutic agents15,16, but these reports mainly focused on the risk for health care providers, rather than on the risk of contaminating the reagents to be injected during syringe loading or patient administration. Taken together, PIS may be a risk factor for intraluminal contamination, but it had not been confirmed by microbiological investigations until now.
Here, we examined whether PIS can cause bacterial contamination of the fluid in a syringe barrel and trigger bacterial trans-catheter infections, by performing microbiological analyses of E. coli cultures. We found that artificially applied bacteria on a plunger, a syringe barrel, and on the surface of the hand of an examiner penetrated into the intraluminal fluid. Furthermore, we also found that disinfection with ethanol before PIS prevented bacterial contamination. These findings implicate that PIS itself can cause intraluminal contamination. Our data also indicate that disinfection is effective for prevention of bacterial contamination. These results reconfirm the importance of disinfection when PIS is required during perioperative, emergency, and intensive care clinical practice.
Methods
Bacterial cultures
To establish an assay system suitable for this study that could not be easily contaminated with other bacteria, we adopted an specific bacterial culture system. DH-5 alpha (E. coli DH-5 alpha Competent Cells, Takara Bio, Kusatsu City, Shiga, Japan) is a commercially available E. coli strain with high transformation efficiency for the incorporation of exogenous DNA that is frequently used for genetic manipulations in molecular cell biology research. DH-5 alpha was multiplicated in LB Miller medium (2.5%, Nacalai Tesque, Kyoto City, Kyoto, Japan); however, LB is not a selective medium for E. coli amplification.
Pumping infusions with a syringe (PIS)
All of the procedures were performed in a safety cabinet to ensure the safety of the examiner and to prevent bacterial contamination from other sources. Briefly, the bacterial contamination of each of the following samples was determined (Fig. 2a): 2 mL of sterilized 2.5% LB Miller medium placed into the barrel of a new sterilized 10-mL plastic syringe (Terumo, Shibuya Ward, Tokyo, Japan) (negative control; NC), 10 µL of LB Miller medium containing DH-5 alpha 1–2 × 106 cells/mL with 2 mL of sterilized LB Miller medium placed into the barrel of a new sterilized 10-mL plastic syringe (positive control; PC), 40 µL of LB Miller medium containing 1–2 × 106 cells/mL of DH-5 alpha applied on the plunger with a micropipette (plunger contact; PL), 40 µL of LB Miller medium containing 1–2 × 106 cells/mL of DH-5 alpha applied on the inner side of the barrel (IB), 2 mL of LB Miller medium placed into the barrel and LB Miller medium containing 1–2 × 106 cells/mL of DH-5 alpha applied on sterilized gloves before PIS (contaminated gloves; GL), and 2 mL of LB Miller medium placed into the barrel and LB Miller medium containing 1–2 × 106 cells/mL of DH-5 alpha, which were first applied on gloves, and then the gloves were sprayed with 80% ethanol and finally allowed to dry before PIS (disinfection of contaminated gloves with ethanol; GL + EtOH) (Fig. 2b). Also, we examined the effect of pulling the plunger down and up in different directions: the “vertical” method in which the plunger was pulled vertically, and the “inclined” method, in which the plunger was forcefully pushed down in an inclined direction with the tip of a syringe almost closed so that intraluminal LB Miller medium leaked through the gap between the edge of the plunger and a barrel (Fig. 3b) by originating a negative pressure in the barrel. Each syringe plunger was moved back and forth 10 times between the “3 and 12 mL” marks in the barrel, for 4 s per stroke, according to different methods of syringe usage (Fig. 2b). Then, each LB Miller medium sample in the barrel was transferred into a different sterilized test tube and cultured for about 19 h at 37 °C while shaking the solution (BR-43FL, Taitec, Nagoya City, Aichi, Japan).
Measurement of OD600
Absorbances at 600 nm (OD600) of 96 µL aliquots per sample, for all the samples were measured with xMark Microplate Reader (Bio-Rad, Hercules, CA, USA). Each OD600 was calculated as follows:
Statistical analysis
All the tests were performed in triplicate using Prism 7 software for Mac OS X (GraphPad software, San Diego, CA, USA). Statistical data are presented as mean ± standard error of the mean.
Results
Establishment of experimental PIS conditions without any other bacterial contamination
For this study, we set up different experimental conditions. DH-5 alpha, a highly competent E. coli strain, was grown in LB medium. To confirm the sterility of the growth medium, we performed two tests. For the first one, we soaked our fingers, covered with sterile nitrile gloves, into the medium; while for the second one, we immersed our bare fingers into the LB medium. Both media were incubated at 37 °C overnight and their OD600 values were measured. There are several methods to determine bacterial grow, among them colony forming unit and OD600 values has been widely used in research articles and guidelines on infections3,4. In our study, we have adopted OD600 values as bacterial grow indicator because is faster and easier to performed while also providing accurate results.
Insertion of sterile nitrile gloves did not result in any bacterial growth in the medium, as opposed to the use of bare fingers (Fig. 1a, p = 0.33). Based on this result, we performed all the remaining procedures with sterile gloves.
In order to confirm that the DH-5 alpha bacterial growth was almost proportional to the OD600 values, we cultured DH-5 alpha at a very low concentration (1 × 105 cells/mL, OD600 = almost zero) and measured OD600 at sequentially consecutive time points. We observed that the OD600 values increased in time along with bacterial growth, between OD600 values of 0 and 2 (Fig. 1b). Based on these observations, we established as our experimental conditions those in which the OD600 values increased along with the number of DH-5 alpha, in the range of OD600 values between 0 and 2.
PIS contaminated the fluid inside the loaded syringes
To determine whether PIS could induce bacterial contamination, we tested several scenarios that could lead to contamination: 40 µL of DH-5 alpha droplets were applied at the plunger (PL), 40 µL were applied with a micropipette inside the barrel around the “10 mL” indicator (IB), 200 µL were applied on top of gloved hands (contaminated gloves; GL, Fig. 2a). For PL and IB conditions, we performed ten strokes between the “3 and 12 mL” indicators for about 4 s per stroke by grasping the edge of the plunger (Fig. 2b). For GL and GL + EtOH conditions, we grasped the whole plunger with the treated gloves during the strokes (Fig. 2a). We determined that PL, GL, and in particular IB conditions caused bacterial contamination (Fig. 2c). In addition, to examine whether the degree of PIS-induced contamination in the intraluminal medium was dependent on the number of strokes after adding the aliquot of bacteria in the syringe, these procedures were repeated with several volumes of DH-5 alpha-containing medium. This assay determined that the bacterial contamination was almost proportional to the applied volume (R2 = 0.9254, Fig. 2d). These results suggested that PIS that contained bacteria in the plunger, lumen, and hand of the examiner caused bacterial contamination in an originally aseptic intraluminal fluid.
Disinfection with ethanol before PIS prevented PIS-induced contamination
Next, to determine whether disinfection by spraying the syringe with ethanol before PIS prevented bacterial contamination, we compared both GL and GL + EtOH conditions (200 µL of DH-5 alpha-containing medium applied on sterile gloved hands followed by 80% of ethanol spray, Fig. 2a). We noted that addition of EtOH prevented further, although not statistically significant, GL-induced contamination, compared to GL conditions, (Fig. 3a, p = 0.074, ANOVA post-hoc Dunnett's multiple comparison test).
Finally, clinicians often experience the leakage of intraluminal reagent into the rubber edge of a plunger during PIS, especially when PIS is performed quickly (Fig. 3b) and the plunger is rapidly and forcefully pulled at an inclined angle, because the negative pressure produced inside the barrel of a syringe can easily draw air through the gap between the plunger and the barrel. We hypothesized that this leakage could introduce bacteria attached around the plunger into the syringe and thus cause intraluminal contamination. To examine this, we performed strokes at a normal angle (vertical) and at an inclined angle with leakage at the edge of a plunger (inclined). We observed that the inclined procedure did not result in increased intraluminal contamination (Fig. 3c). Taken together, a simple and easy disinfection of the syringe with ethanol before PIS was able to reduce the risk of intraluminal contamination; however, the precise mechanism of bacterial contamination is still unknown.
Discussion
In this study, we examined whether the artificial application of bacterial culture on a syringe plunger or lumen, or the glove of the examiner can cause the penetration of the bacteria into the fluid inside the syringe, using a microbiological methodology. We demonstrated for the first time that PIS itself could cause trans-catheter bacterial contamination by any of the conditions examined in this study. Furthermore, we also confirmed that previous disinfection with ethanol could prevent PIS-induced bacterial contamination.
Several previous studies have highlighted the risk of contamination in clinical practice from different aspects17,18,19,20. Stucky et al. emphasized the influence of pollutants in the clinical environment on the risk of contamination inside a syringe19. In the present experiments, we showed that PIS caused intraluminal contamination, although the contamination induced by considerably contaminated gloves was not significantly increased compared to loading bacterial droplets on the plunger or the inner side of the barrel. We also showed that previous disinfection of the syringe by spraying ethanol almost prevented the contamination. Therefore, these results provided a very important take-home lesson for all clinicians. During the perioperative period and critical care, where PIS is often performed for patient resuscitation, the incidence of CRBSI is one of the highest risk factors for a longer intensive care unit stay, that could lead to devastating complications, such as refractory severe sepsis, infectious endocarditis, and death7,21. Therefore, to minimize the possible risk of contamination all health care providers are required to apply good clinical practices as hygienically as possible. As this study demonstrated, previous disinfection is a very simple and easy procedure that would be beneficial for the prevention of iatrogenic complications.
Our study also presented scientific validity data on the role of the plastic cover outside the plunger of the syringe on preventing contamination by preventing physical contact of the plunger with external materials, shown in Fig. 3d. Closed-type peripheral arterial catheters, which are commonly used in intensive care unit and operation room for real-time blood pressure monitoring and easy blood access, have adopted this type of covered plastic syringes, but no study has been reported on its efficiency to prevent contamination. Our study shows that the plastic cover would be effective in the prevention of intraluminal contamination. Further experiments using the covered disposable plastic syringe may enable us to better understand the mechanism of contamination of the fluid in the barrel.
We observed a relatively large variability between the results of our experimental replicates (Fig. 2c). The reason is unknown; however, when we tested if the way of moving the plunger could affect the degree of contamination, no significant differences were obtained (Fig. 3c). It could be possible that a very small amount of DH-5 alpha-containing medium was pulled into the gap between the plunger and the barrel and that the resulting contamination depended on the amount of trapped volume. The mechanism of this phenomenon is still unknown, but one conceivable hypothesis is that the syringes used in our study were plastic disposable ones, and they might have been deformed by the slight heat produced during forced and repeated strokes, resulting in a small gap sufficient for penetration of bacteria into the intraluminal medium. To verify this hypothesis, further investigation on PIS-induced contamination, such as using glass-made or heat-resistant plastic syringes, is required.
Our study had some limitations. First, not all of the PIS procedures, in which a plunger was moved for 4 s per stroke, were perfectly performed at the same speed, pressure, and friction force between the plunger and barrel. The absorbance values obtained in each experiment, which greatly differed in number, indicated that a slight difference in the pressure and friction might deform the plastic lumen, resulting in very small bacterial penetration. Second, our results do not propose a strict warning that PIS with bare hands would cause bacterial contamination of loaded fluid or reagents in the syringe, and could lead to CRBSI with high possibility, because our study was just a simple experiment. Although several samples showed high absorbance in our experiment, they were considerably smaller than that of positive control (data not shown). This implies that the number of bacteria that penetrated the barrel might be very low, and these penetrations might be negligible in clinical practice. Finally, it is yet unknown whether PIS-induced contamination itself can actually cause CRBSI in clinical medicine. In this sense, several related factor such as the detailed mechanism and growth pattern of PIS-induced contamination, the amount of bacteria outside the syringe or in the fluid inside the barrel that is sufficient to cause infection, the extent of the penetrated bacteria increase, and the extent to which the immunologic conditions of the patient receiving PIS get compromised for infection, remain unknown. Thus, future investigations on the detail of exacerbation of intraluminal contamination induced by PIS, possibly into infectious complications in animal models, such as rodents, are required.
In conclusion, we showed for the first time that PIS itself caused intraluminal contamination when bacteria are attached outside the barrel of a syringe, and that previous disinfection of gloves prevented bacterial contamination. Our results confirmed the importance of keeping the hands of the clinicians antiseptic during management of the catheter of a patient, according to the recommended good clinical practice, and performing a disinfection prior and during surgery, emergency medical treatments, and intensive care to protect the health of the patient.
Data availability
All data generated or analyzed during this study are included in this published article.
References
Kalkwarf, K. J. & Cotton, B. A. Resuscitation for hypovolemic shock. Surg. Clin. N. Am. 97, 1307–1321. https://doi.org/10.1016/j.suc.2017.07.011 (2017).
Harvey, G., Foster, G., Manan, A., Thabane, L. & Parker, M. J. Factors affecting pediatric isotonic fluid resuscitation efficiency: A randomized controlled trial evaluating the impact of syringe size. BMC Emerg. Med. 13, 14. https://doi.org/10.1186/1471-227X-13-14 (2013).
O’Grady, N. P. et al. Guidelines for the prevention of intravascular catheter-related infections. Clin. Infect. Dis. 52, e162–e193. https://doi.org/10.1093/cid/cir257 (2011).
Koh, D., Gowardson, J., Rickard, C. M., Brown, A. & Robertson, I. Prospective study of peripheral arterial catheter infection and comparison with concurrently sited central venous catheters. Crit. Care Med. 36, 397–402. https://doi.org/10.1097/CCM.0b013e318161f74b (2008).
Nagao, M. A multicentre analysis of epidemiology of the nosocomial bloodstream infections in Japanese university hospitals. Clin. Microbiol. Infect. 19, 852–858. https://doi.org/10.1111/1469-0691.12083 (2013).
Daschner, F. D., Frey, P., Wolff, G., Baumann, P. C. & Suter, P. Nosocomial infections in intensive care wards: A multicenter prospective study. Intensive Care Med. 8, 5–9. https://doi.org/10.1007/BF01686847 (1982).
Donowitz, L. G., Wenzel, R. P. & Hoyt, J. W. High risk of hospital-acquired infection in the ICU patient. Crit. Care Med. 10, 355–357. https://doi.org/10.1097/00003246-198206000-00001 (1982).
Duggan, J. M., Oldfield, G. S. & Ghosh, H. K. Septicaemia as a hospital hazard. J. Hosp. Infect. 6, 406–412. https://doi.org/10.1016/0195-6701(85)90057-x (1985).
Craven, D. E., Kunches, L. M. & Lichtenberg, D. A. Nosocomial infection and fatality in medical and surgical intensive care unit patients. Arch. Intern. Med. 148, 1161–1168. https://doi.org/10.1001/arch- (1988).
Pittet, D., Tarara, D. & Wenzel, R. P. Nosocomial bloodstream infection in critically ill patients. Excess length of stay, extra costs, and attributable mortality. JAMA 271, 1598–1601. https://doi.org/10.1001/jama.271.20.1598 (1994).
Mermel, L. A., McCormick, R. D., Springman, S. R. & Maki, D. G. The pathogenesis and epidemiology of catheter-related infection with pulmonary artery Swan-Ganz catheters a prospective study utilizing molecular subtyping. Am. J. Med. 16, 197S-205S. https://doi.org/10.1016/0002-9343(91)90369-9 (1991).
Crnich, C. J. & Maki, D. G. The promise of novel technology for the prevention of intravascular device-related bloodstream infection. II. Long-term devices. Clin. Infect. Dis. 34, 1362–1368. https://doi.org/10.1086/340105 (2002).
Tsuboi, M. et al. Comparison of the epidemiology and microbiology of peripheral line- and central line-associated bloodstream infections. Am. J. Infect. Control 47, 208–210. https://doi.org/10.1016/j.ajic.2018.08.016 (2019).
Oto, J. et al. Comparison of bacterial contamination of blood conservation system and stopcock system arterial sampling lines used in critically ill patients. Am. J. Infect. Control 40, 530–534. https://doi.org/10.1016/j.ajic.2011.08.006 (2012).
Favier, B., Gilles, L., Latour, J. F., Desage, M. & Giammarile, F. Contamination of syringe plungers during the sampling of cyclophosphamide solutions. J. Oncol. Pharm. Pract. 11, 1–5. https://doi.org/10.1191/1078155205jp147oa (2005).
Smith, S. T. & Szlaczky, M. C. Syringe plunger contamination by hazardous drugs: A comparative study. J. Oncol. Pharm. Pract. 20, 381–385. https://doi.org/10.1177/1078155214526428 (2014).
Clarke, T. N. & Human, R. P. Contamination of hypodermic needles in anaesthetic practice. Anesthesia 39, 549–551. https://doi.org/10.1111/j.1365-2044.1984.tb07359.x (1984).
Rupp, M. E. et al. Outbreak of bloodstream infection temporally associated with the use of an intravascular needleless valve. Clin. Infect. Dis. 44, 1408–1414. https://doi.org/10.1086/517538 (2007).
Stucki, C., Sautter, A. M., Favet, J. & Bonnabry, P. Microbial contamination of syringes during preparation—The direct influence of environmental cleanliness and risk manipulations on end-product quality. Am. J. Health Syst. Pharm. 66, 2032–2036. https://doi.org/10.2146/ajhp070681 (2009).
Abe, S. et al. Bacterial contamination upon the opening of injection needles. JA Clin. Rep. https://doi.org/10.1186/s40981-018-0197-7 (2018).
Almuneef, M. A. et al. Rate, risk factors and outcomes of catheter-related bloodstream infection in a paediatric intensive care unit in Saudi Arabia. J. Hosp. Infect. 62, 207–213. https://doi.org/10.1016/j.jhin.2005.06.032 (2006).
Acknowledgements
We thank Dr. Kohtaro Takei at Yokohama City University Graduate School for providing all the experimental devices, equipment, and reagents for the study; Eriko Saitoh, Yuki Kawaguchi, and Junpei Matsubayashi at Yokohama City University Graduate School for technical assistance during the experiments; Dr. Tomoko Seki, Dr. Toru Sawamoto, Dr. Rimako Umebachi, Dr. Sayuri Inagaki, and Dr. Asami Omata at Odawara Municipal Hospital for peer reviewing the article; and Masahide Aoyagi at Kyowa Medical Corporation for providing experimental advice.
Funding
Grants-in-Aid for Scientific Research, Japan (KAKENHI-20K09296, Takashi Tagami).
Author information
Authors and Affiliations
Contributions
Y.K. designed the research, performed experiments, and drafted the manuscript. T.T. designed the whole study, revised the manuscript, and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Kawakami, Y., Tagami, T. Pumping infusions with a syringe may cause contamination of the fluid in the syringe. Sci Rep 11, 15421 (2021). https://doi.org/10.1038/s41598-021-94740-1