Introduction

Vascular stiffening is an important health indicator. It influences health in a manner that is independent of blood pressure1. The American Heart Association recommends that vascular stiffening be used as a predictor of cardiovascular disease risk1. Vascular stiffness is inversely proportional to compliance and reflects arterial resistance to stretching and recoiling in response to blood ejection from the heart. In a young or otherwise healthy vascular system, compliant arteries help smooth out the vascular pulsatility induced by blood pressure. As compliance decreases, blood pressure causes structural changes within the arterial walls that adversely affect their function2,3.

Another change that occurs with reduced vascular compliance is increased pulse wave velocity (PWV)4. PWV is the speed of the pressure wave traveling from the heart towards the periphery. Simultaneous measurement of arterial pulse in a proximal and a distal arterial sites e.g. common carotid and femoral gives the travel time of the pressure waveform. The distance between the measurement sites can be measured over the body surface. Hence, PWV is calculated as distance divided by travel time between two measurement sites using applanation tonometry5.

When a pressure wave reaches a vascular bifurcation point, a portion of the wave is reflected back towards the heart. These reflected waves eventually augment the incident wave coming from the heart, thereby increasing the pulse pressure. While we often think of a single reflected wave, in reality many reflected and re-reflected waveforms are present in the signal. Due to the increased PWV in a stiff arterial system, the reflected waves travel faster, thereby reaching the incident wave earlier in systole and increasing the local pulse pressure even further. Both forward and reflected waves leave a peak or an inflection point on the pressure waveform, we will refer to these peaks as the systolic and the reflected peaks, respectively. The time difference between systolic and reflected peaks is the reflection time6,7. The reflection time can be measured using non-invasive technologies such as magnetic resonance imaging (MRI), transcranial Doppler (TCD) and photoplethysmograph (PPG). PPG utilises light emitters and detectors to measure blood volume changes in periphery (e.g. fingertip, ankle, wrist or earlobe). A PPG estimate of PWV, called stiffness index (SI), is calculated by dividing subject height by the reflection time8,9,10,11.

In the brain, increased arterial stiffness can damage the arterial endothelium in small arteries and can, over time, produce a pathology known as small vessel disease (SVD)2. The brain, in particular, is highly vulnerable to SVD, as the blood flow to the sub-surface brain structures is compromised12,13,14. Based on a comprehensive review, cerebral SVD is believed to be one of the main causes of vascular cognitive impairments and accounts for a quarter of all acute ischaemic strokes15.

Cerebrovascular compliance is commonly assessed non-invasively using indices derived from blood pressure and flow readings. TCD and MRI are the modalities most widely used to acquire cerebrovascular compliance and cerebrovascular health indices. TCD measures blood flow velocity; i.e., the speed at which the blood cells move inside a vessel. While it is possible to estimate compliance using TCD blood flow velocity16, the method relies on estimates of vascular cross-sectional area, arterial inflow and venous outflow, which are not readily measurable. The TCD indices such as the pulsatility index, \(\text {PI}_{{\rm TCD}}\), and augmentation index, \(\text {AI}_{{\rm TCD}}\), are the commonly reported TCD cerebrovascular health indicators17,18,19. Of these two, \(\text {PI}_{{\rm TCD}}\) is the most commonly used TCD index due to its ease of calculation. However, \(\text {AI}_{{\rm TCD}}\) has been shown to have a stronger correlation with cerebrovascular health indicators such as age and cardiorespiratory fitness (CRF)20. The MRI pulsatility index (\(\text {PI}_{\rm MRI}\)) is the commonly adopted MRI-based cerebrovascular health measure14. While MRI has a high spatial resolution for localized measurements, it is less commonly used as it is both costly and time consuming, making it unsuitable for simple and routine monitoring of vascular health in the general population.

Near-infrared spectroscopy (NiRS) operates on same principals as PPG and its output is proportional to regional blood volume changes inside the brain based on the absorption of near-infrared light, mainly by oxygenated blood inside the arteries21,22. NiRS offers high temporal resolution and is sufficiently sensitive to changes in blood volume during the cardiac cycle. Therefore, NiRS allows local studies of cerebrovascular behaviour and possible assessment of cerebrovascular health21. The recently proposed pulse relaxation function (PReFx) is the first NiRS based approach to the indexing of cerebrovascular compliance21. PReFx measures the deformation of the blood volume waveform during the systolic relaxation phase due to the presence of reflected waves. It has been demonstrated to correlate with CRF and inversely correlate with age21,23,24. PReFx is an area measurement in the averaged NiRS signal. We are motivated to use an index that can be easily applied to NiRS signals and also has evidence in other modalities.

In our previous work using a mathematical model of the pressure wave, we have shown that the reflection time is proportional to PWV which supports PPG SI as a PWV estimator. Our model also shows that the reflection time is proportional to vascular compliance and vascular resistance which are influenced by vascular stiffness. In fact, it has been shown that the reflection time is more than the time it takes for the blood waveform to travel the distance between the measurement site and the reflection site. The waveform faces an additional delay, which is a function of resistive and capacitive properties of the vascular system beyond the reflection site, before turning back from the reflection site. Therefore, the reflection time is a sum of two time delays: one is a line delay controlled by PWV and physical distance between the measurement and reflection sites; the other is a load delay forced by the vascular properties beyond the reflection site7. All these parameters are known to change with ageing and vascular health. Based on this model, we have introduced a new timing index (TI) as the inverse of the reflection time, similar to PPG SI but without the height parameter. In a previous experimental study we have shown that TI calculated using TCD blood flow velocity waveforms correlates significantly with vascular health factors such as age and CRF20. Here, we propose NiRS timing index (\(\text {TI}_{{\rm NiRS}}\)) as a measure of cerebrovascular health. We define the NiRS timing index as the inverse of the time between the systolic and reflected waveform peaks in the NiRS signal. Then, we will examine the correlation of the proposed index with other health indices including age, CRF and the most currently used vascular health measures derived from other imaging modalities, namely TCD augmentation index and MRI pulsatility index.

Methods

Protocol and data collection

Thirty eight adults (23 female, 15 male, age range = 24–67 years, mean age = 41.7 years) were recruited from the Newcastle, Australia. They provided informed consent prior to assessment. The study protocol was approved by the University of Newcastle Human Research Ethics Committee and is registered in the Australian and New Zealand Clinical Trials Registry (ACTRN12619000144112). All experiments were performed in accordance with relevant guidelines and regulations. TCD recordings from these participants were previously reported in20. Each participant attended the Hunter Medical Research Institute and the University of Newcastle Callaghan campus over two consecutive days and participated in three scanning sessions. Participants were asked to refrain from consuming caffeine before their scans. Height, weight, age, sex and resting heart rate were recorded for each participant and a physical activity questionnaire was completed.

TCD ultrasound (DopplerBox X; Compumedics DWL, Singen, Germany) was used to record resting-state cerebral blood flow velocity from the right and left middle cerebral arteries. Participants were also scanned on a 3T MRI scanner (Magnetom Prisma, Siemens Healthineers, Erlangen, Germany), equipped with 64-channel receive only head coil, while a standard built-in dual channel body coil was used for RF transmission. Blood flow was quantified using a phase contrast flow quantification sequence (TR = 26.5ms, TE = 6.9 ms, slice thickness = 5mm, matrix \(256 \times 256\)) of a single excitation with a velocity encoding value of 120cm/s on middle cerebral arteries. Finally, participants were instructed to relax and sit still for 300 s of resting-state NiRS recording. A frequency-domain NiRS (ISS Imagent, Champaign, Illinois, USA) was used at a 110MHz modulation frequency and sampling rate of 39.0625 Hz. The setup montage was designed to cover the majority of the frontal lobe and consisted of four detectors, each crossed with 16 time-multiplexed sources (eight of which operated at 690 nm and the other eight at 830 nm), making a total of 64 channels. Participants were asked to wear an Equivital EQ02 LifeMonitor sensor belt during this session, which recorded ECG and NiRS signals simultaneously.

During the TCD imaging session, resting-state heart rate was measured using an heart rate monitoring device (Omron HEM-7320). Participants assumed a sitting position for approximately five minutes during headpiece setup before recording commenced. During TCD recording, the heart rate was measured once per minute three times. The mean of these three heart rate measurements was used as the resting-state heart rate.

Calculating the indices

MRI pulsatility index

MRI phase contrast images were processed using the scanner software (Siemens Syngo) by manually placing a region of interest around the middle cerebral artery and quantifying the flow. The MRI pulsatility index, \(\text {PI}_{\rm MRI}\), was then obtained by subtracting the minimum flow from the maximum flow and dividing it by the mean flow over the cardiac cycle14; i.e.,

$$\begin{aligned} PI _{\rm MRI} =\frac{F_{\max} -F_{\min} }{\overline{F}}, \end{aligned}$$
(1)

where F indicates the flow and the bar sign indicates the mean value. Three subjects did not have an MRI scan and among the remaining 35 subjects, \(\text {PI}_{\rm MRI}\) was calculated either for both middle cerebral arteries (\(n=21\)) or for a single middle cerebral artery (MCA) (\(n=14\)) based on data quality. The indices from right and left MCAs were then averaged for each subject.

TCD augmentation index

Calculation of the augmentation index, \(\text {AI}_{{\rm TCD}}\), requires identification of three characteristic points on an averaged Doppler waveform. Peak systole (\(V_{{\rm sys}}\)), peak diastole (\(V_{{\rm dia}}\)) and the peak of the reflected wave (\(V_{{\rm refl}}\)) are the parameters required to calculate the augmentation index18:

$$\begin{aligned} \text {AI}_{{\rm TCD}}=\frac{V_{{\rm refl}}-V_{{\rm dia}}}{V_{{\rm sys}}-V_{{\rm dia}}}. \end{aligned}$$
(2)

Thus, the timing of the diastolic peaks was first determined automatically in MATLAB using a peak detector function. Then, each signal was averaged with respect to its diastolic peaks, resulting in an averaged signal covering the duration of a single cardiac cycle. The systolic and reflected peaks were set manually on the averaged signal. The TCD data for two subjects were excluded from the analysis due to noise. The \(\text {AI}_{{\rm TCD}}\) index was calculated using data from both MCAs (and averaged) for 34 subjects and from a single MCA for two subjects, depending on data quality.

Evaluating cardiorespiratory fitness

The CRF was determined using the non-exercise test model proposed in25. It involves a self-reported physical activity score, age, sex, body mass index and resting heart rate. See the supplementary materials for the physical activity questionnaire and cardiorespiratory fitness formula.

NiRS indices

Optical light intensity NiRS signals were used to derive two different indices: the pulse relaxation function (PReFx) and the proposed timing index, \(\text {TI}_{{\rm NiRS}}\). The pre-processing of the recordings was done similarly for both methods and was adopted from21 using the “\(p\underline{}pod\)” package26 which is a MATLAB-based software module developed by the same group. Specifically, channels with source detector distances of 20–60 mm were selected and went through a quality-control process passing 1283 channels out of the original \(38\times 64=2432\) channels. These channels were then normalized by dividing by their mean values and then bandpass-filtered to 0.5-5.0 Hz to capture the dynamics of the pulse signal and remove any low-frequency elements and moving artefacts. The resulting signal was averaged with respect to the ECG R peaks, and then rotated around the x-axis (see Fig. 1). The result is called the arterial pulsation signal, s. The rotation was performed so that the systolic peak would appear as a positive peak and the shape of the signal would look similar to that of a blood pressure, flow or volume signal. This facilitates understanding of the waveform. The arterial pulsation signals were visually inspected and channels that did not have a pulse shape were removed (i.e., low blood volume at the beginning and end representing diastole, and an increase in between representing systole). Some 964 channels passed the pre-processing stage for further analysis. Channels with source detector distances between 25 and 50 mm proved to be most likely to generate usable waveforms.

Figure 1
figure 1

(a) Ten-second-long raw NiRS signal (top/blue) and simultaneously recorded ECG lead I (bottom/red). (b) Averaged and flipped NiRS s signal. The images are generated using MATLAB (version R2020b, www.mathworks.com/).

PReFx

The pulse relaxation function is calculated based on the methods explained in21. In brief, \(s_{{\rm D1}}\) and \(s_{{\rm D2}}\) peaks occur as minima in the s signal in the first and the second cardiac cycles and \(s_{{\rm S}}\) occurs as the first local maximum after \(s_{{\rm D1}}\) (see Fig. 2). \(\left\{ s_{{\rm D1}},s_{{\rm D2}}\right\} \) and \(s_{{\rm S}}\) are intended to represent the diastolic and systolic peaks, respectively. However, \(s_{{\rm S}}\) is not necessarily a systolic peak, see Fig. 2b where \(s_{{\rm S}}\) is the reflected peak and the systolic peak indicated by an arrow. Then, the area enclosed by the s signal between \(s_{{\rm S}}\) and \(s_{{\rm D2}}\) is calculated as A and inserted into

$$\begin{aligned} \mathrm{PReFx}=\frac{A}{B}-0.5. \end{aligned}$$
(3)

where B is the area of a rectangle formed by \(s_{{\rm S}}\) and \(s_{{\rm D2}}\) at its diagonal ends (i.e. the rectangle formed by blue and red lines in Fig. 2). Note that if the drop in volume between peaks was linear, PReFx would equal zero. After communicating with the authors of21, channels with PReFx value outside the interval of \(\left[ -0.1,0.4\right] \) were removed and other channels were manually excluded from the calculations where reflected peaks were selected by the algorithm (see Fig. 2b for an example). A total of 245 additional channels were removed in this process. Finally, for each subject with more than ten PReFx values, an averaged PReFx value was calculated as the index of vascular compliance. We were able to calculate PReFx for 29 of the 38 subjects.

Figure 2
figure 2

Examples of calculating PReFx for two different s signals. The signal s is shown in black, the \(s_{{\rm S}}\), \(s_{{\rm D1}}\) and \(s_{{\rm D2}}\) peaks are marked with black circles, and the blue and red rectangles are formed with \(s_{S }\) and \(s_{D2 }\) at their diagonals. The shaded area, A, is the area enclosed by s between \(s_{S }\) and \(s_{D2 }\) and the red sides of the rectangle. The calculated PReFx values for (a) and (b) are 0.136 and 0.081, respectively. The images are generated using MATLAB (version R2020b, www.mathworks.com/).

\(\text {TI}_{{\rm NiRS}}\)

We define the timing index as the inverse of the time between the forward and reflected waveform peaks. In order to find the timing of these peaks, named \(t_{sys }\) and \(t_{refl }\) for peak systole and peak reflection, respectively, for a given signal, s, a completely automated algorithm is proposed (Sample data and codes available at https://github.com/Rashid-Afkhami). The algorithm uses the second derivative to locate key points in the signal. This is based on similar approaches that have been proposed for locating peaks and inflection points in the pressure and PPG signals27,28. Firstly, the second derivative, \(s_2\), of s is calculated. Next, the following time points are defined on the s and the \(s_2\) signals: \(\left\{ t_{p1 }, t_{p2 }, ...\right\} \), the time of the peaks on the s signal and \(\left\{ t_{zc1 }, t_{zc2 }, ...\right\} \) the time of zero-crossings on the \(s_2\) signal. If the number of \(s_{2}\) zero-crossings is less than 4, the channel is excluded. Otherwise,

$$\begin{aligned} t_{sys }=\min \left( t_{p1 },t_{zc2 }\right) . \end{aligned}$$
(4)

Then, to define \(t_{refl }\) the algorithm checks only the time period between \(t_{zc3 }\) and \(t_{zc4 }\). If s has a positive peak, \(t_{refl }\) is set as the time point corresponding to this peak, \(t_{pi }\), otherwise, it is set as the time when \(s_2\) reaches its minimum, \(t_{min }\) (all in the specified time period between \(t_{zc3 }\) and \(t_{zc4 }\)) i.e.,

$$\begin{aligned} t_{refl }={\left\{ \begin{array}{ll} t_{pi } &{} {\text {if}}\,s\,\text {contains a positive peak}\\ t_{min } &{} \text {else} \end{array}\right. } . \end{aligned}$$
(5)

After all channels are processed for a subject, channels with \(t_{sys }\) and \(t_{refl }\) outside their mean \(\pm 1.5\) std (i.e., standard deviation) or outside a predefined threshold were also excluded from further calculations. The predefined threshold for \(t_{sys }\) was any value \(<125\) ms and for \(t_{refl }\) it was any value >500 ms. Then, if ten or more channels remained, the \(t_{sys }\) and \(t_{refl }\) values for each channel were used to calculate the timing index as:

$$\begin{aligned} TI _{NiRS }=\frac{1}{t_{refl } - t_{sys }}. \end{aligned}$$
(6)

The timing index values of different channels were then averaged for each participant. Of the original 38 subjects, we were able to calculate a TI for 32 of them. Examples of the proposed algorithm are shown in Fig. 3 when the algorithm is applied to in vivo and simulated signals. Three main types of the arterial pulsation signals, s, are shown in this figure where the detected \(t_{{\rm sys}}\) and \(t_{{\rm refl}}\) correspond to either a peak point or an inflection point on the recorded signal. The simulated signals are formed by adding three waveforms, an incident forward-traveling waveform, a reflected waveform and a re-reflected waveform (which is also traveling towards the periphery). The purpose of the simulated waveforms is to better visualize where the underlying waveforms reach their peak and how the algorithm performs in terms of detecting the peak times. For each simulated case the underlying waveforms are arranged in a form that the resulted observed signal (black line) resembles an in vivo case. Using a NiRS device only the composite signal s is detected and the wave components shown in Fig. 3b,d and f are for illustrative purposes only.

Figure 3
figure 3

Three main types of in vivo arterial pulsation signals are shown on the left column (a), (c) and (e) and an equivalent simulated signal is shown on the right column (b), (d) and (f), respectively. Black lines are the observed recorded blood volume signal, s and the red lines are the second derivatives, \(s_2\). The gray lines represent the underlying forward, reflected and re-reflected waveforms. The x-axis represents time in seconds. The images are generated using MATLAB (version R2020b, www.mathworks.com/).

Statistical analysis

Statistical analysis was carried out in MATLAB using the inbuilt corr function. Lilliefors test of normality with a 5% significance level was used to assess the data distribution. Age and \(\text {PI}_{\rm MRI}\) were found to be non-normally distributed (\(p=0.0072\) and 0.0020, respectively). All other distributions were normal. Pearson or Spearman correlation coefficients were calculated for normally-distributed and non-normally-distributed data, respectively. False discovery rate correction was used to adjust significance levels with an initial \(\alpha =0.05\).

Results

The correlations of the proposed NiRS timing index and existing NiRS PReFx index with the indices derived from other imaging modalities, age and CRF are reported in Table 1. Corresponding scatter-plots are shown in Fig. 4. Least-squares linear models have not been provided for Fig. 4g,h due to small slopes. Significant correlations after correcting for multiple comparisons are marked with an asterisk. Note that PReFx is an index of vascular compliance whereas \(\text {TI}_{{\rm NiRS}}\) is an index of vascular stiffness that is inversely proportional to compliance. Therefore, the two indices show opposite signs when correlated with the same factors. The proposed systolic peak and reflected wave peak detection algorithm for \(\text {TI}_{{\rm NiRS}}\) detected 228 channels (out of 964 input channels) with a systolic peak corresponding to an inflection point (mean participant age for these channels = 42.5 years). Both \(\text {TI}_{{\rm NiRS}}\) (the peak detection algorithm) and PReFx indices are affected by the sampling frequency of 39.0625 Hz (i.e. \(\approx \) 25 ms sampling interval) used in this study; a faster sampling rate would allow a higher resolution and more accurate results.

Table 1 Correlation matrix with correlation coefficients (and p-values). Asterisks indicate statistically significant correlations after correction for multiple comparisons using a false discovery rate
Figure 4
figure 4

Scatter plots of the correlations with least-squares fit line. (a), (b), (c) and (d) represent \(\text {TI}_{{\rm NiRS}}\) scatter plots with CRF, age, \(\text {AI}_{{\rm TCD}}\) and \(\text {PI}_{\rm MRI}\), respectively. (e), (f), (g) and (h) represent PReFx scatter plots with CRF, age, \(\text {AI}_{{\rm TCD}}\) and \(\text {PI}_{\rm MRI}\), respectively. Grey bars show standard deviation. The images are generated using MATLAB (version R2020b, www.mathworks.com/).

Discussion

The NiRS indices are designed to measure vascular stiffness, which is well correlated with age1 and cardiorespiratory fitness29. Thus, a common approach with any vascular index is to compare its relationship with age and CRF30 and here we have done the same. The PReFx correlation coefficient reported here for age is consistent with existing studies; however, the correlation with CRF is lower than previously reported. PReFx correlations with age and CRF in the adult population have so far been studied by the same research group using two different datasets21,23. Reported correlations with age are \(r=-0.39\) (mean age = 69.87 years)21 and \(r=-0.43\) (mean age = 47.8 years)23; a correlation of \(r=-0.60\), mean age = 58.8 years was reported in24 for a combination of the two datasets. Our reported correlation, \(r=-0.40\) (\(p=0.030\)), although consistent with the previous values, is not significant. Reported correlations with CRF are \(r=0.416\)21 and \(r=0.32\)23, however, our correlation with CRF of \(r=0.24\) (\(p=0.203\)) was not significant. In calculation of CRF using the non-invasive method25, it is common to remove the gender factor where the result is called eCRF21. \(\text {TI}_{{\rm NiRS}}\) and PReFx had similar correlations with eCRF as they had with CRF, i.e. \(r=-0.52\) and \(r=0.22\), respectively. Given that the mean age for our sample is 41.7 years, we suspect that the PReFx measure has a lower correlation in younger subjects than older ones. Based on the performance of the PReFx algorithm on the current data, we have identified two potential sources that can distort the PReFx calculation and potentially cause lower correlations. Firstly, there are cases where the PReFx algorithm detects reflected peaks instead of systolic peaks as the \(s_{S }\) points. The PReFx algorithm assumes the first local maximum after the minimum peak, i.e. \(s_{S }\), to be the systolic peak, which is not accurate when the true systolic peak is in the form of an inflection point (e.g., see Figs. 2b, 3b). Based on our analysis these cases account for almost one-fourth of all channels (228 of the 964 studied channels) and are more likely to occur in older participants in which the reflected wave moves faster and reaches the incident wave sooner than in younger participants. The second problem occurs in channels with a notch in the signal after the systolic and reflected peaks, which is likely caused by closure of the aortic valve (similar to the dicrotic notch seen in the pressure waveform31). This notch can appear as a large deformation in the signal (see Fig. 2b for an example), which will change PReFx values independently of vascular health factors. Among the aforementioned two potential causes of low correlation for PReFx, the first was eliminated in the results reported in Table 1 by discarding channels where \(s_{S }\) was assigned to the reflected peak. However, if all the channels are included in the calculations, it will remove the significant correlation of PReFx with age; i.e., \(r=-0.19\).

While no previous NiRS studies have reported a timing index, the correlations of \(\text {TI}_{{\rm NiRS}}\) with age and CRF match those of the TCD reported by our group20. That is, correlations of \(r=0.70\) and \(r=0.50\) between TCD TI and age in two different datasets and a correlation of \(r=-0.79\) between TCD TI and CRF. The proposed NiRS timing index, \(\text {TI}_{{\rm NiRS}}\), also shows significant correlations with age (\(r=0.53\), \(p=0.002\)) and CRF (\(r=-0.44\), \(p=0.011\)). \(\text {TI}_{{\rm NiRS}}\) correlated more strongly with both age and CRF than PReFX. This suggests that TI may be a more sensitive index than PReFX to age and lifestyle related cerebrovascular changes in middle-aged participants.

Based on a previously proven association between \(\text {PI}_{\rm MRI}\) and SVD14, we expect NiRS indices to correlate with the MRI pulsatility index which is an indicator of vascular health. PReFx showed no significant correlation with \(\text {PI}_{\rm MRI}\) (\(r=-0.01\), \(p=0.978\)), whereas \(\text {TI}_{{\rm NiRS}}\) correlated significantly with \(\text {PI}_{\rm MRI}\) (\(r=0.45\), \(p=0.012\)). \(\text {PI}_{\rm MRI}\) indicates vascular health by measuring the pulsatility of the blood flow. High pulsatility means that fast traveling reflected waves join incident waves and contribute to increased maximum flow peaks. Similarly, \(\text {TI}_{{\rm NiRS}}\) reaches greater values when the reflected wave peak is closer to the systolic peak and pulsatility is high. PReFx also changes based on the location of the reflected wave; high PReFx values correspond to channels with high deformation from a perfect sinusoidal waveform, i.e., the existence of a reflected peak. However, the presence of a dicrotic notch may also affect the PReFx index by distorting the wave shape while having no effect on PI or TI. Thus, in channels with a prominent dicrotic notch the PReFx may be high while pulsatility is low, which may explain the lower correlation for PReFx.

\(\text {TI}_{{\rm NiRS}}\) showed a significant correlation with the TCD augmentation index (\(r=0.46\), \(p=0.010\)), while the correlation of PReFx with \(\text {AI}_{{\rm TCD}}\) was not significant (\(r=-0.26\), \(p=0.173\)). The augmentation index is a widely accepted index used in both blood pressure and flow velocity (TCD) signals, and is a strong predictor of vascular aging and stiffening18,19,32,33. The AI captures the augmentation of the incident wave by the reflected wave, with smaller values reflecting no augmentation when the arteries are healthy and the propagation speed is low. TI indexes this delay between the incident and reflected waves directly, which is why \(\text {AI}_{{\rm TCD}}\) and \(\text {TI}_{{\rm NiRS}}\) are significantly correlated. The PReFx also correlates with AI as larger PReFx values can correspond to late reflected waves leaving a peak on the systolic-to-diastolic segment of the arterial pulsation signal and thus increasing the area under the curve. Nevertheless, the correlation may be weakened when the presence of dicrotic notch increases PReFx regardless of the reflected wave.

Overall, NiRS blood volume waveforms are believed to closely resemble the changes in blood pressure21 which is also supported by the use of light emitters in PPG and its morphological similarity to the arterial blood pressure waveform34,35. Thus, we believe that the proposed NiRS timing index is affected similarly by vascular health parameters in the same way as the timing index derived from the blood pressure waveforms7. As blood pressure can not be readily measured in cerebral vessels, a NiRS-based timing index can provide an alternative approach.

Conclusion

In this paper, we presented a novel NiRS based cerebrovascular stiffness index, \(\text {TI}_{{\rm NiRS}}\). The \(\text {TI}_{{\rm NiRS}}\) correlates significantly with age, CRF and other cerebrovascular health indices derived from TCD and MRI data, indicating it has high performance in tracking changes in the cerebrovascular system. NiRS offers a potentially valuable means of indexing vascular health and has superior cost, portability and widespread implementation potential compared with existing techniques.