Abstract
Saphenous veins (SVs) are frequently employed as bypass grafts. The SV graft failure is predominantly seen at the valve site. Avoiding valves during vein harvest would help reduce graft failure. We endeavored to detect SV valves, tributaries, and vessel size employing upright computed tomography (CT) for the raw cadaver venous samples and in healthy volunteers. Five cadaver legs were scanned. Anatomical analysis showed 3.0 (IQR: 2.0–3.0) valves and 13.50 (IQR: 10.00–16.25) tributaries. The upright CT completely detected, compared to 2.0 (IQR: 1.5–2.5, p = 0.06) valves and 9.5 (IQR: 7.5–13.0, p = 0.13) tributaries by supine CT. From a total of 190 volunteers, 138 (men:75, women:63) were included. The number of valves from the SF junction to 35 cm were significantly higher in upright CT than in supine CT bilaterally [upright vs. supine, Right: 4 (IQR: 3–5) vs. 2 (IQR:1–2), p < 0.0001, Left: 4 (IQR: 3–5) vs. 2 (IQR: 1–2), p < 0.0001]. The number of tributaries and vessel areas per leg were also higher for upright compared with supine CT. Upright CT enables non-invasive detection of SV valves, tributaries, and vessel size. Although not tested here, it is expected that upright CT may potentially improve graft assessment for bypass surgery.
Similar content being viewed by others
Introduction
For decades, the saphenous vein (SV) has served as the most reliable bypass conduits for coronary and peripheral artery revascularization. Over 250,000 coronary artery bypass grafts and over 80,000 lower extremity vein grafts are performed annually in the United States alone1. The average rate of CABG has been reported to be 44 per 100,000 individuals2, and almost 80% of bypass grafts include saphenous veins (SVs)2,3. However, SV graft failure during follow-up is not uncommon3,4, wherein most stenotic lesions are focal and often occur either at the valve sites or anastomotic regions5,6,7. Therefore, it has been proposed that non-inclusion of venous samples with valve sites may reduce SV graft failure. This concept has led to the development of valvulotomy8; however, this approach did not improve prognosis9.
The valves of venous systems with venous muscle pumps allow the blood to return to the heart against gravity and generally demonstrate expansion at the valve site during in the upright position 10. We have developed an upright CT (Fig. 1A,B) to negate the postural effect of gravity11 and render the clinical imaging as close to the physiological state as possible 12,13. Employing the upright CT, we aimed to detect SV valves, tributaries, and venous areas and compare them with those obtained using standard supine CT imaging.
Materials and methods
Cadaver study
Cadaveric specimens were obtained from the Willed Body Program (a whole-body donation program which written informed consent was obtained from doner or next of kin at Department of Anatomy, Keio University School of Medicine) and were investigated in accordance with institutional regulations. Lower extremities, with no scars and no signs of vascular disease, were obtained from raw cadavers. Lower extremities were disarticulated at the hip joint and the external iliac artery, external iliac vein, and the surrounding soft tissue remaining attached to the specimens. The distal position of saphenous vein was cannulated with a 22-Gauze catheter at above ankle level, and phosphate buffered saline (PBS) with 10% polyethylene glycol 14 was infused under continuous natural dripping by gravity during the scan to reproduce the venous reflex. The CT images were acquired as ex-vivo imaging of the disarticulated leg. Thereafter, the leg specimens were fixed in formaldehyde and the saphenous vein was dissected and made a long single incision along the venous path from the cannulation point to the saphenous-femoral junction (SF junction) (Supplemental Fig. 1). The SV valves and tributaries positions were macroscopically identified as the reference standard for CT images.
Volunteer study
This volunteer prospective study was approved by the Institutional Review Board (Keio University Independent Ethics Committee) (Clinical Trial number: UMIN000026586) and written informed consent was obtained from all participants. Healthy volunteers were recruited from a volunteer recruitment company from July 2017 to March 2019; volunteers of over 30 years of age were requested so that they could understand the purpose of the study. Individuals with a history of hypertension, dyslipidemia, diabetes, smoking, or those who had previously undergone cardiac surgery or were currently receiving treatment were excluded.
Image acquisition
A conventional 320-detector row CT (Aquilion ONE, Canon Medical Systems Corporation, Otawara, Japan) was prospectively performed for all volunteers in the supine position. In addition, and upright CT (prototype TSX-401R, Canon Medical Systems Corporation, Otawara, Japan) 11 was performed in the standing position immediately after the conventional procedure. The upright and conventional supine CT scanners were placed next to each other, and the two examinations were consecutively performed. The upright CT system is characterized by up-and-down movements of a transverse 320 row-detector gantry (isotropic 0.5 mm in detector size), with a bore size of 780 mm, a gantry rotation speed of 0.275 s, a maximum vertical speed of 100 mm/s, and a 1200 view at optimal performance. Scanning was performed at 100kVp and at a gantry rotation speed of 0.5 s in the helical scan mode (80-row detector), with a noise index of 24 and helical pitch of 0.8 for the body trunk from the level of superior margin of the external acoustic meatus to the lowest position of the upright CT (37 cm height).
For cadaver study, cadaver legs were placed in plastic moulds and whole samples were scanned in both supine and upright CT scanners. The plastic box containing the sample was fixed on a pedestal (42 cm height) and the tube current for scanning was set at approximately 600 mA to compensate for the noise from the fixtures.
Image reconstruction was performed using Adaptive Iterative Dose Reduction 3D (Canon Medical Systems Corporation, Otawara, Japan), which could reduce the radiation dose.
Image analysis
CT images were transferred to an off-line workstation (SYNAPSE VINCENT, FujiFilm, Tokyo, Japan) and multiplanar reconstruction (MPR) images were developed from the SF junction of the saphenous vein both for cadaveric images and clinical images. The positions and number of valves were counted visually. Simultaneously, the vessel area in cross-sectional images were measured to calculate the dilation ratios (vessel area at valve /vessel area at reference segments). The reference segments were set at the average of the vessel area of the proximal and distal positions of the valves. The dilation ratio was derived in the cadaveric study and applied for the volunteer study to study the valves. The number of tributaries were counted visually. The vessel area was measured of the cross-section image on the 15 cm from the SF junction. For the cadaveric study, two observers analyzed the data with a consensus reading. For the volunteer study, all measurements were performed in a blinded and randomized manner.
Statistical analysis
Intra-observer and inter-observer variability for the assessment of valves, tributaries, and cross-sectional measurements were examined using Bland–Altman analysis and Pearson’s correlation. Data were presented as median (interquartile range [IQR]; i.e., 25th to 75th percentile, or Q1, Q3). Continuous data were compared using a Wilcoxon rank-sum test between the 2 groups. A Spearman Rank Correction coefficient test was used for the assessment of linear correlation of 2 parameters. A 2-sided p value of < 0.05 was considered statistically significant. All analyses were performed using SAS software, version 9.4. (SAS Institute Inc., Cary, North Carolina).
Results
Cadaver study
Five legs were dismembered from cadavers (age range 78–93 years, mean 88.4 ± 6.0 years, 4 left and 1 right leg, 4 female and 1 male). On visual assessment 3.0 (IQR 2.0–3.0) valves and 13.50 (IQR 10.00–16.25) tributaries per leg were observed in the SV (Fig. 1C–E).
The upright CT images accurately identified the valves and tributaries, while the identification was difficult using conventional supine CT. The upright CT image showed 3.0 (IQR 2.0–3.0) valves and 13.50 (IQR 10.00–16.25) tributaries, while supine CT images showed 2.0 (IQR 1.5–2.5, p = 0.06) valves and 9.5 (IQR 7.5–13.0, p = 0.13) tributaries per leg; given a small sample size the difference however was not statistically significant. The expansive venous diameter at the valve site (referred to as dilation ratio) in the upright CT images was 1.56 (IQR 1.44–1.83, range 1.20–4.24), and the minimum dilation ratio of “1.20” was used to confirm the valve site in the volunteer images.
Volunteer study
Of the 190 volunteers, 38 who received the scan with 120 kV were excluded; another 14 volunteers were excluded where complete CT data from the SF junction to 35 cm could not be obtained due to their short statures. We report the CT characteristics of the SV from 138 volunteers (men 75, women 63) who completed the study (Table 1).
Firstly, 40 CT leg images from 10 consecutive volunteers taken from both sides and in both the supine and upright positions, were analyzed by two observers, with a high intra-/inter-observer agreement noted (Supplemental Figs. 2, 3 and 4). The upright images allowed for the identification of valves and tributaries (Fig. 2). It was difficult to characterize valves and tributaries in supine CT images, similar to that seen in cadaveric studies.
The number of valves was significantly higher in upright CT than in supine CT on both sides [upright right side: 4 (IQR: 3–5) vs. supine right side: 2 (IQR: 1–2), p < 0.0001, upright left side: 4 (IQR: 3–5) vs. supine left side: 2 (IQR: 1–2), p < 0.0001] (Fig. 3A,B). The dilation ratio at the valve site was 1.85 (IQR 1.57–2.25). The frequency of valve distribution was maximum between 0 and 5 cm from the SF junction, followed by 5 to 10 cm from the SF junction (Fig. 3C).
The number of tributaries was also higher in the upright CT than in the supine CT [upright right: 12 (IQR: 10–15) vs. supine right: 11 (IQR: 9–13), p < 0.0001; upright left: 12 (IQR: 9–14) vs. supine left: 11 (IQR: 9–13), p < 0.0001] (Fig. 4A).
The area of the cross-sectional image at 15 cm from the SF junction was larger in the upright CT [upright right side: 14 (IQR: 11–17) mm2 vs. supine right: 12 (IQR: 9–16) mm2, p < 0.0001; upright left: 14 (IQR: 11–18) mm2 vs. supine left: 12 (IQR: 9–15) mm2, p < 0.0001] (Fig. 4B).
Discussion
This study showed that the SV valves were clearly visualized in the standing position utilizing upright CT both in cadaveric legs and in-vivo study of healthy volunteers. In the cadaver study, SV valves and tributaries identified by upright CT were confirmed by anatomical analysis. A minimum dilation ratio of “1.20” was found to be the cut-off point to define the level of the valve to confirm the valve site in the in-vivo study. In the volunteer study, upright CT detected a larger number of valves and tributaries as well as a larger size of vessel area compared with supine CT. We propose that upright CT could have an advantage for SV characterization prior to bypass surgeries. We envision that this method can allow for SV valve and tributary assessment in order to locate the optimal part of the saphenous vein (minimum SV valves and tributaries with sufficient diameter) before harvesting the saphenous vein for bypass grafts, although further studies need to be done.
SV grafts have the advantages of having long lengths and being easily available while being more resistant to iatrogenic injury and less susceptible to vasospasm 15. They have considerably longer length compared to the other types of grafts. SV grafts have been most used for non-LAD coronary territories worldwide. It has the major disadvantage of a high graft failure rate2,3, with approximately 40–50% of grafts occluding at 10 years after CABG surgery, of which 10–25% are occluded within the first year post-operatively 4,16,17. In comparison, the occlusion rates of other grafts are lower, for instance the radial artery shows a failure rate of 17–37% 16,18. One of the causes of SV failure is the endothelium dysfunction during the harvest, new harvesting techniques being employed reduce dysfunction and graft failure, such as including bridge, no-touch, endoscopic resections3. Another possible case is the presence of venous valves. Although the mechanisms of primary stenosis of coronary artery bypass grafts and peripheral artery bypass grafts have been suggested to be different, they share thrombosis as the basis of luminal loss 19. In SV grafts, valve sites frequently gather clots 20, which cause valve dysfunction, scarring and develop post-thrombotic syndrome 19; the response of local smooth muscle cells to injury is found to be accelerated in-vitro 21. Clinical studies have confirmed that valve sites frequently cause stenotic lesions when SV graft reoperations are performed 5,6,7,22,23,24. We assumed that knowing the SV valve position and avoiding the valve sites during harvesting may offer another approach to reduce SV graft failure.
To identify the best portion of the SV as a conduit for bypass surgery, the number and position of valves in the SV were reviewed. A Brazilian study characterized 60 veins from 30 adult cadavers and reported that the average number of valves from the medial epicondyle of the femur to the saphenous hiatus were 4.77 and 4.87 on the left and right sides, respectively 25. Similarly, a Japanese study of 26 SVs from 20 adult cadavers found 111 valves (average: 4.27) between the SF junction and the upper patellar margin compared to 63 (average: 2.42) valves between the upper patellar margin and the medial malleolus 26; they also reported that greatest number of valves were observed within 10 cm from the SF junction and between 35 and 45 cm from the SF junction. Because we identified 4 (IQR: 3–5) valves from 0 to 35 cm from the SF junction and the distribution of valves was mainly observed between 0 and 10 cm, the results of our volunteer study were comparable with these autopsy studies; however, the data beyond 35 cm from the SF junction were not available in this study. Therefore, upright CT was able to accurately characterize of SV valves, which could not be easily done using conventional computed tomography.
Another advantage of upright CT is its noninvasive identification of tributaries and vessel areas. Vessel size is an important factor to predict graft failure and a luminal diameter of over 2.0 mm is preferable for SV graft patency and longevity 3,16,27. New harvesting techniques with less invasiveness to the SV may reduce damage compared with conventional open harvesting techniques, however; it is difficult to evaluate the valve sites and vessel sizes in the whole SV. In addition, it is important to be able to ligate all side tributaries 28. Upright CT allows to locate valves, identify tributaries, and accurate measurement vessel sizes in non-invasive manner, and should be incorporated in the intraoperative strategy before performing graft harvesting.
Recently upright (or weight-bearing) CT 29 and MRI 30 have been proposed to evaluate the effect of gravity. However, previous upright CT scanners were equipped with a cone beam CT, and their scan range was limited. Upright MRI has a lower spatial resolution than CT 31. Compared to these machines, the upright CT used in the current study has a wide-scan range with quick motion and high spatial resolution, which is similar to a high-end supine CT in the standing position. It has an ability to evaluate blood distribution of whole body 11, and may contribute to the development of phlebology, which has more room for the development compared with arteriology.
Limitations of the study
Although a novel attempt, this study has several limitations. First, because the upright CT used was a prototype machine, this is a single center study. Second, the protocol of this study was designed to obtain body-trunk images. Thus, the scan range of this prototype machine could not be detected beyond a height of 37 cm, which did not allow for imaging of the entire saphenous vein. However, we could acquire the entire SV with a pedestal to reach a height of 42 cm, which we utilized for the evaluation of the ankles 32. Third, valve insertion sites were identified by defining the venous (expansive) dilation ratio compared to the lumen, with a cut-off point of 1.20 used based on anatomic observation in the cadaveric limb. The application of a workstation is being developed to enable to automatic detection of the valve in future clinical studies. Fourth, we could not identify the valve leaflet directly, because the upright CT scans were performed without contrast medium and the limit of the spatial resolution of CT is 0.4 to 0.6 mm. Invasive OCT with 15 micro-meter spatial resolution 31 has allowed for the recognition of culprit valve lesions after CABG 33. However, it is difficult to identify SV valves in supine CT, even with the use of contrast materials. Thus, this approach, which utilized the difference in venous pressure upon changes in posture34, may have practical use for detecting valves positions.
Conclusions
Upright CT enables detection of SV valve sites, tributaries, and vessel sizes noninvasively. It may potentially assist the strategies for SV harvesting to improve patency of bypass-grafts.
Data availability
The datasets generated during and/or analyzed during the current study are available from the corresponding author upon reasonable request.
References
Bhasin, M. et al. Temporal network based analysis of cell specific vein graft transcriptome defines key pathways and hub genes in implantation injury. PLoS ONE 7, e39123. https://doi.org/10.1371/journal.pone.0039123 (2012).
Head, S. J., Milojevic, M., Taggart, D. P. & Puskas, J. D. Current practice of state-of-the-art surgical coronary revascularization. Circulation 136, 1331–1345. https://doi.org/10.1161/circulationaha.116.022572 (2017).
Caliskan, E. et al. Saphenous vein grafts in contemporary coronary artery bypass graft surgery. Nat. Rev. Cardiol. https://doi.org/10.1038/s41569-019-0249-3 (2019).
Fitzgibbon, G. M. et al. Coronary bypass graft fate and patient outcome: Angiographic follow-up of 5,065 grafts related to survival and reoperation in 1,388 patients during 25 years. J. Am. Coll. Cardiol. 28, 616–626. https://doi.org/10.1016/0735-1097(96)00206-9 (1996).
Bandyk, D. F., Schmitt, D. D., Seabrook, G. R., Adams, M. B. & Towne, J. B. Monitoring functional patency of in situ saphenous vein bypasses: The impact of a surveillance protocol and elective revision. J. Vasc. Surg. 9, 286–296 (1989).
Mills, J. L., Fujitani, R. M. & Taylor, S. M. The characteristics and anatomic distribution of lesions that cause reversed vein graft failure: A five-year prospective study. J. Vasc. Surg. 17, 195–204; discussion 204–196. https://doi.org/10.1067/mva.1993.42796 (1993).
Davies, M. G. & Hagen, P. O. Pathophysiology of vein graft failure: A review. Eur. J. Vasc. Endovasc. Surg. 9, 7–18. https://doi.org/10.1016/s1078-5884(05)80218-7 (1995).
Mills, N. L. & Ochsner, J. L. Valvulotomy of valves in the saphenous vein graft before coronary artery bypass. J. Thorac. Cardiovasc. Surg. 71, 878–879 (1976).
Albäck, A., Kantonen, I., Ihlberg, L. & Lepäntalo, M. Valvulotomy of non-reversed saphenous vein bypass grafts: A randomised, blinded, angioscopy-controlled study. Eur. J. Vasc. Endovasc. Surg. 18, 144–148. https://doi.org/10.1053/ejvs.1999.0863 (1999).
Bergan, J. J. et al. Chronic venous disease. N. Engl. J. Med. 355, 488–498. https://doi.org/10.1056/NEJMra055289 (2006).
Jinzaki, M. et al. Development of upright computed tomography with area detector for whole-body scans: Phantom study, efficacy on workflow, effect of gravity on human body, and potential clinical impact. Invest. Radiol. 55, 73–83. https://doi.org/10.1097/rli.0000000000000603 (2020).
Yamada, Y. et al. Differences in lung and lobe volumes between supine and standing positions scanned with conventional and newly developed 320-detector-row upright CT: Intra-individual comparison. Respiration 99, 598–605. https://doi.org/10.1159/000507265 (2020).
Narita, K. et al. Pelvic floor morphology in the standing position using upright computed tomography: Age and sex differences. Int. Urogynecol. J. https://doi.org/10.1007/s00192-020-04335-z (2020).
Yamazaki, S. et al. Three-dimensional demonstration of the lymphatic system in the lower extremities with multi-detector-row computed tomography: A study in a cadaver model. Clin. Anat. 26, 258–266. https://doi.org/10.1002/ca.22179 (2013).
Parang, P. & Arora, R. Coronary vein graft disease: Pathogenesis and prevention. Can. J. Cardiol. 25, e57–e62. https://doi.org/10.1016/s0828-282x(09)70486-6 (2009).
Goldman, S. et al. Long-term patency of saphenous vein and left internal mammary artery grafts after coronary artery bypass surgery: Results from a Department of Veterans Affairs Cooperative Study. J. Am. Coll. Cardiol. 44, 2149–2156. https://doi.org/10.1016/j.jacc.2004.08.064 (2004).
Hess, C. N. et al. Saphenous vein graft failure after coronary artery bypass surgery: Insights from PREVENT IV. Circulation 130, 1445–1451. https://doi.org/10.1161/circulationaha.113.008193 (2014).
Motwani, J. G. & Topol, E. J. Aortocoronary saphenous vein graft disease: Pathogenesis, predisposition, and prevention. Circulation 97, 916–931. https://doi.org/10.1161/01.cir.97.9.916 (1998).
ten Cate-Hoek, A. J., Henke, P. K. & Wakefield, T. W. The post thrombotic syndrome: Ignore it and it will come back to bite you. Blood Rev. 30, 131–137. https://doi.org/10.1016/j.blre.2015.09.002 (2016).
Mackman, N. New insights into the mechanisms of venous thrombosis. J. Clin. Invest. 122, 2331–2336. https://doi.org/10.1172/jci60229 (2012).
Kikuchi, S. et al. Smooth muscle cells of human veins show an increased response to injury at valve sites. J. Vasc. Surg. 67, 1556-1570.e1559. https://doi.org/10.1016/j.jvs.2017.03.447 (2018).
Szilagyi, D. E., Elliott, J. P., Hageman, J. H., Smith, R. F. & Dall’olmo, C. A. Biologic fate of autogenous vein implants as arterial substitutes: Clinical, angiographic and histopathologic observations in femoro-popliteal operations for atherosclerosis. Ann. Surg. 178, 232–246. https://doi.org/10.1097/00000658-197309000-00002 (1973).
Tullis, M. J., Primozich, J. & Strandness, D. E. Jr. Detection of “functional” valves in reversed saphenous vein bypass grafts: Identification with duplex ultrasonography. J. Vasc. Surg. 25, 522–527. https://doi.org/10.1016/s0741-5214(97)70263-5 (1997).
Vesti, B. R., Primozich, J., Bergelin, R. O. & Strandness, E. Jr. Follow-up of valves in saphenous vein bypass grafts with duplex ultrasonography. J. Vasc. Surg. 33, 369–374. https://doi.org/10.1067/mva.2001.111744 (2001).
Portugal, I. B., Ribeiro Ide, L., Sousa-Rodrigues, C. F., Monte-Bispo, R. F. & Rocha, A. C. Distribution of saphenous vein valves and its practical importance. Rev. Bras. Cir. Cardiovasc. 29, 564–568. https://doi.org/10.5935/1678-9741.20140038 (2014).
Shinohara, H., Morisawa, S., Toshima, M. & Mizukami, S. Distribution of valves in the great saphenous vein; its clinical implications. Okajimas Folia Anat. Jpn. 67, 219–221. https://doi.org/10.2535/ofaj1936.67.4_219 (1990).
Souza, D. S. et al. Improved patency in vein grafts harvested with surrounding tissue: Results of a randomized study using three harvesting techniques. Ann. Thorac. Surg. 73, 1189–1195. https://doi.org/10.1016/s0003-4975(02)03425-2 (2002).
Souza, D. S. et al. Harvesting the saphenous vein with surrounding tissue for CABG provides long-term graft patency comparable to the left internal thoracic artery: Results of a randomized longitudinal trial. J. Thorac. Cardiovasc. Surg. 132, 373–378. https://doi.org/10.1016/j.jtcvs.2006.04.002 (2006).
Hirschmann, A., Pfirrmann, C. W., Klammer, G., Espinosa, N. & Buck, F. M. Upright cone CT of the hindfoot: Comparison of the non-weight-bearing with the upright weight-bearing position. Eur. Radiol. 24, 553–558. https://doi.org/10.1007/s00330-013-3028-2 (2014).
Alyas, F., Connell, D. & Saifuddin, A. Upright positional MRI of the lumbar spine. Clin. Radiol. 63, 1035–1048. https://doi.org/10.1016/j.crad.2007.11.022 (2008).
Nakahara, T. et al. Coronary artery calcification: from mechanism to molecular imaging. JACC Cardiovasc. Imaging 10, 582–593. https://doi.org/10.1016/j.jcmg.2017.03.005 (2017).
Kaneda, K. et al. Three-dimensional kinematic change of hindfoot during full weightbearing in standing: An analysis using upright computed tomography and 3D–3D surface registration. J. Orthop. Surg. Res. 14, 355. https://doi.org/10.1186/s13018-019-1443-z (2019).
Koeda, Y., Itoh, T., Fusazaki, T., Nakamura, M. & Morino, Y. A unique stenosis in saphenous vein graft visualized by optical coherence tomography. Heart Vessels 29, 278–281. https://doi.org/10.1007/s00380-013-0362-x (2014).
Stick, C., Hiedl, U. & Witzleb, E. Venous pressure in the saphenous vein near the ankle during changes in posture and exercise at different ambient temperatures. Eur. J. Appl. Physiol. Occup. Physiol. 66, 434–438. https://doi.org/10.1007/bf00599617 (1993).
Acknowledgements
This study was supported by Canon Medical Systems, Japan Society for the Promotion of Science (JSPS KAKENHI: Grant Number JP17H04266), Uehara Memorial Foundation, and Canon Medical Systems (Otawara, Japan) to JM.
Author information
Authors and Affiliations
Contributions
T.N. made contribution to the conception and design of the study, acquisition, analysis and interpretation of cadaver study data and volunteer study data, and drafting the article. Mi.Y., Y.Yo., made contribution to the acquisition of cadaver study data and volunteer study data, and revising the manuscript critically for important intellectual content. Y.Ya. and K.N. made contribution to the acquisition of volunteer study data and revising the manuscript critically for important intellectual content. N.I., Ma.Y. H.S. and J.N. made contribution to revising the manuscript. M.J. made contribution to the conception and design of the study and revising the manuscript.
Corresponding author
Ethics declarations
Competing interests
This study was supported by the Japan Society for the Promotion of Science (JSPS KAKENHI: Grant Number JP17H04266), Uehara Memorial Foundation, and Canon Medical Systems (Otawara, Japan). Masahiro Jinzaki received a grant from Canon Medical Systems. Canon Medical Systems has loaned an upright computed tomography machine to Keio University. However, Canon Medical Systems is not involved in the design and conduct of the study; in the collection, analysis, and interpretation of the data; nor the preparation, review, or approval of the manuscript. All other authors have no relationships with the industry or other entities.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Nakahara, T., Yamada, M., Yokoyama, Y. et al. Saphenous vein valve assessment utilizing upright CT to potentially improve graft assessment for bypass surgery. Sci Rep 11, 11602 (2021). https://doi.org/10.1038/s41598-021-90998-7
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-90998-7
This article is cited by
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.