Abstract
Radiolucent carbon-fiber-reinforced (CFR) polyethyl-ether-ether-ketone (PEEK) has been established in spinal instrumentation for oncological reasons. Laboratory data reported comparable bacterial adhesion as titanium. Thus, using of CFR-PEEK spinal instrumentation for spondylodiscitis bases on artifact-free imaging to evaluate therapeutic success. Studies comparing the rate of pedicle screw loosening and relapse of spondylodiscitis following titanium versus CFR-PEEK instrumentation do not exist so far. This study evaluates the rate of pedicle screw loosening and recurrence of spondylodiscitis after CFR-PEEK instrumentation for spondylodiscitis compared to titanium. We conducted a prospective single center study between June 2018 and March 2019 on consecutive 23 patients with thoracolumbar spondylodiscitis. Imaging data was evaluated for screw loosening at a minimum of three months after surgery. A matched-pair analysis was performed using spondylodiscitis cases between 2014 and 2016 using titanium instrumentation for equal localization, surgery, and microorganism class. Among 17 cases with follow-up imaging, six cases (35%) showed screw loosening while only 14% (two patients) with titanium instrumentation were loosened (p = 0.004). In both groups the most frequent bacterium was Staphylococcus aureus, followed by Staphylococcus epidermidis. From the S. aureus cases, one infection in both groups was caused by methicillin resistant species (MRSA). No difference was found in the rate of 360° fusion in either group due to matching criteria. As opposed to other indications CFR-PEEK screws show more loosening than titanium in this series with two potentially underlying reasons: a probably stronger bacterial adhesion on CFR-PEEK in vivo as shown by a statistical trend in vitro and instrumentation of spondylytic vertebrae. Until these factors are validated, we advise caution when implanting CFR-PEEK screws in infectious cases.
Introduction
Carbon-fiber-reinforced (CFR) polyethyl-ether-ether-ketone (PEEK) material has been used successfully for many years in the orthopedic field1 and for spinal fusion techniques2,3,4. Results from cadaver and clinical studies exist which show that CFR-PEEK rods can withstand equal stiffness and resistance to motion compared to titanium rods for lumbar fusion5,6. For the use of CFR-PEEK instrumentation in primary spinal tumors a low intraoperative complication rate and low rate of pedicle screw loosening has been shown7. The advantage of CFR-PEEK material is its radiolucency which reduces artifacts on computed tomography (CT) and magnetic resonance imaging (MRI)8, thus making this material more amenable to follow-up analysis. Moreover, CFR-PEEK screws show less radiation interference than titanium screws9, so radiotherapy can be applied more precisely. Therefore, CFR-PEEK seems as a promising option for the instrumentation of spinal tumors and metastases, as artifact-free postoperative imaging is essential for radiotherapy planning, for precise application of radiotherapy and for the evaluation of tumor progression.
Artifact-free postoperative images are not only of interest in the oncological field, but also in the treatment of spinal infection. Treatment of spondylodiscitis consists of spine immobilization and antibiotic therapy10. To treat patients with spondylodiscitis without intraspinal empyema or neurological deficit conservatively or surgically is still an ongoing international debate11. The advantage of surgical treatment using spinal instrumentation is an earlier mobilization of patients and lower complication rates in modern series12. Evaluation of therapeutic success is important before termination of the antibiotic therapy. Successful treatment of spondylodiscitis is evaluated by clinical findings such as infection parameters, pain reduction, and by follow-up imaging13.
The idea to use CFR-PEEK instrumentation for spondylodiscitis cases origins from the ability to gain artifact-free follow-up imaging, enhancing the evaluation of therapeutic success. Laboratory data showed that CFR-PEEK material harbors statistically no unfavorable bacterial adhesion as titanium does14.
However, the use of CFR-PEEK instrumentation in the treatment of spondylodiscitis has not been established yet. Because of the existing laboratory and clinical results for oncological indications, we hypothesized that CFR-PEEK instrumentation is equal or even superior to titanium instrumentation for spondylodiscitis due to artifact-free follow-up imaging.
This study therefore evaluates the first consecutive series of spondylodiscitis patients who underwent CFR-PEEK instrumentation regarding pedicle screw loosening and relapse of spondylodiscitis.
Methods
Study design
We conducted a prospective single-center study between June 2018 and March 2019. Patients with thoracic and lumbar spondylodiscitis treated with CFR-PEEK instrumentation were included. Follow-up X-rays, CT- and MRI-scans were evaluated with regard to screw loosening and relapse of spondylodiscitis by a board of seven neurosurgeons. Screw loosening was defined as the manifestation of a halo sign around at least one pedicle screw in the follow-up imaging. A matched-pair analysis was conducted using retrospective data of spondylodiscitis cases between January 2014 and December 2016 using titanium instrumentation. Main criteria for matching were (with decreasing importance) (1) localization: lumbar versus thoracic spine, (2) posterior fusion versus 360° fusion, (3) spinal levels, (4) type of germs detected intraoperatively or by biopsy.
In both groups pedicle screw placement was performed in the same way by 3D-fluoroscopy or CT based navigation.
This study was designed to evaluate safety and efficacy of CFR-PEEK instrumentation for spondylodiscitis. It was conducted according to STROBE guidelines.
Ethical approval
The study was approved by the ethical committee of our university (“Ethikkommission der Technischen Universität München”, chair: Prof. Dr. G. Schmidt, reference number 229/18s) and conducted in accordance to the Declaration of Helsinki. Written informed consent of all patients was received.
Statistical methods
To analyze demographic data, the number of 360° fusion/posterior fusion and the reasons for revision surgery Fisher’s exact test was used. To compare patients’ median age student’s-t-test was applied. The number of cases with screw loosening was analyzed using the Chi-square test. p < 0.05 was regarded as significant. Statistical analysis was performed by GraphPad Prism 5.0 (La Jolla, California).
Results
Patient characteristics
Between June 2018 and March 2019, 24 patients were included in the CFR-PEEK group. One patient was excluded from the study seeing as no bacteria was detected and histopathology definitively excluded an infection. In total, 23 cases with CFR-PEEK instrumentation for spondylodiscitis were analyzed. They were matched with 23 cases with titanium instrumentation. In 16 patients the lumbar spine was affected by spondylodiscitis, in seven patients the thoracic spine. Of the patients with lumbar spondylodiscitis in twelve cases the infection was monosegmental, while in four cases two or more segments were affected. Of the patients with thoracic spondylodiscitis, in five cases it was monosegmental, while in two cases two or more segments were involved. As this was a matching criterion there was no difference between the CFR-PEEK and the titanium group regarding the localization and number of affected segments. Concerning demographic parameters, no statistically significant difference could be detected (Table 1).
Microbiological spectrum
The most frequently detected bacterium in both groups was Staphylococcus aureus of which one case in both groups was a methicillin resistant species (MRSA), followed by Staphylococcus epidermidis in both groups (Fig. 1).
Bacterial spectrum of CFR-PEEK group and matched titanium group. Spectrum of germs causing spondylodiscitis in the CFR-PEEK group and matched titanium group. y-axis designating the number how often a germ was detected. Cand. albicans = Candida albicans; Citro. koseri = Citrobacter koseri; Cut. acnes = Cutibacterium acnes; E. coli = Eschericha coli; E. faecalis = Enterococcus faecalis; E. faecium = Enterococcus faecium; MRSA = methicillin resistant Staphylococcus aureus; MSSA = methicillin sensible Staphylococcus aureus; Staph. epi. = Staphylococcus epidermidis; Steno. malt. = Stenotrophomonas maltophilia; Strep. ang. = Streptococcus anginosus; Strep. virid = Streptococcus viridans; Pseu. aerug. = Pseudomonas aeruginosa; No. = number.
All patients were treated by two weeks of i.v.-antibiotics, followed by ten weeks of oral antibiotics based on the antibiogram of the detected bacterium.
Surgery
Number of instrumented levels depended on the degree of bony destruction. If this was moderate pedicle screws were placed in the vertebra above and below the affected disc (seven cases with lumbar spondylodiscitis). If there was a significant destruction at least two levels above and below were instrumented (nine cases with lumbar spondylodiscitis, in all cases with thoracic spondylodiscitis). In cases with bechterew’s disease three levels above and below were instrumented.
There was no statistically significant difference in the number of 360° fusion and posterior fusion in both groups (p = 0.758) as it was a matching criterion. In the CFR-PEEK group in seven cases vertebral body replacement was performed, in six cases XLIF (extreme lateral intercorporal fusion) and in two cases ALIF (anterior intercorporal fusion). In the eight cases of merely posterior fusion in six cases discectomy and intercorporal fusion with an antibiotic medium and autologous bone was performed. In the titanium group ten cases received intervertebral body replacement and three cases ALIF. In the ten cases with just posterior fusion in eight cases discectomy and intercorporal fusion with an antibiotic medium and autologous bone was performed. 360° fusion, however, was used more frequently in both groups compared to all cases with spondylodiscitis and titanium instrumentation from January 2014 till December 2016. However, this difference, was not statistically significant (p = 0.067) (Table 2).
In the CFR-PEEK group in four cases an intraoperative revision of pedicle screws was necessary because of malpositioning, in two cases an accidental durotomy was reported and in one case a pedicle screw broke with the tip left in the pedicle. In the titanium group in six patients intraoperative correction of pedicle screws was necessary and two accidental durotomies occurred.
Screw loosening
In the CFR-PEEK and titanium group, six and nine patients were lost to follow-up, respectively (Table 3). In the CFR-PEEK group screw loosening occurred significantly more frequently than in the titanium group (35% versus 14%, p = 0.004) (Table 3, Figs. 2, 3). Mean time of diagnosis of screw loosening was 110 days (± 81) for the CFR-PEEK and 71 days (± 11) for the titanium group. Microbiological spectrum of patients with screw loosening did not differ from the spectrum of the total cohort of patients. As well for CFR-PEEK as for titanium material S. aureus was the most frequent bacterium detected within spondylodiscitis cases with loosened screws. S. epidermidis was the second most common (Table 4). In the CFR-PEEK group in one case Cutibacterium acnes was found in sonication of the explanted material. Initially spondylodiscitis of this case was caused by MSSA (Table 5). In the titanium group no bacterium was detected in sonication of the explanted material (Table 6).
Example 1 of CT-imaging of screw loosening after CFR-PEEK instrumentation. CT-scan of a case of screw loosening of the CFR-PEEK group 128 days post-surgery. CFR-PEEK instrumentation was performed on L4–5–S1–S2 and involved the affected vertebrae on L4/5. L4 and L5 pedicle screws show a halo as a sign for loosening. (a,b) coronal plane; (c) sagittal plane, (d) axial plane of L4.
Example 2 of CT-imaging of screw loosening after CFR-PEEK instrumentation. CT-scan of a case of screw loosening after CFR-PEEK instrumentation T5–6–9–10–11–12 involving the affected vertebra on level T11. There a halo signs around the screws of T11 and 12 on the right side as a sign for loosening. (a) coronal plane; (b) sagittal plane, (c,d) axial plane of T11 and T12.
In the CFR-PEEK group in five out of six cases with screw loosening 360° fusion was performed, while in the titanium group this was performed in only one out of the two cases (p = 0.464).
Revision surgery
Operative revision was performed in eight cases of the CFR-PEEK group, but only in two cases of the titanium group (p = 0.071; Table 7). Reasons for reoperation in the study group included screw loosening in five cases, of which in two cases a relapse of spondylodiscitis had occurred. Relapse of spondylodiscitis was defined as a combination of recurrent back pain, elevated serum inflammation markers and radiographic signs in MRI- and CT-scans, such as new contrast enhancement of the discs and progressive bony destruction of the endplates. In two cases wound healing disorders and in one case a cerebrospinal fluid (CSF) leak had occurred (Table 7). In both cases of revision surgery due to relapse of spondylodiscitis no bacterium was detected by sonication of the explanted material. Both cases of the matched group were reoperated due to screw loosening (Table 7).
Discussion
The results from this study of 23 consecutive cases with CFR-PEEK instrumentation for spinal infection suggest that screw loosening occurs more frequently when using CFR-PEEK than titanium for instrumentation of spinal infection.
To the best of our knowledge this is the first study comparing CFR-PEEK instrumentation with titanium in the clinical practice for spondylodiscitis.
Screw loosening and bacterial adhesion
In Europe, S. aureus and coagulase negative Staphylococcus species are the most frequent causes of spondylodiscitis15,16.
In our study, these bacterial species were also the most frequently detected species in the total cohort of patients as well as in the cohort with screw loosening.
At first sight, in vitro laboratory results might show no statistically significant difference in the adhesion of methicillin sensitive S. aureus (MSSA), methicillin resistant S. aureus (MRSA) and S. epidermidis, on CFR-PEEK compared to titanium14. However, if absolute numbers of bacterial adhesion are compared, it becomes clear that bacterial adhesion on titanium is approximately half of the adhesion on CFR-PEEK for all three species (titanium: MRSA: 3.3 × 106 colony forming unit (cfu); S. aureus: 3.6 × 106 cfu; S. epidermidis: 4.0 × 106 cfu; CFR-PEEK: MRSA: 8.7 × 106 cfu; S. aureus: 7.6 × 106 cfu; S. epidermidis: 6.4 × 106 cfu). The reason that this difference did not reach a statistically significant level may be due to a small number of samples; yet, a statistical trend was shown.
In addition, it was shown in the same work, that the surface of CFR-PEEK has a roughness that is almost twice that of titanium (CFR-PEEK 0.38 ± 0.10 µm, titanium 0.20 ± 0.04 µm). It appears conclusive that bacteria can adhere better to a rougher surface. Hitherto, data from this study may provide a reason why there is more screw loosening after CFR-PEEK instrumentation: bacteria can adhere better to CFR-PEEK and promote the loosening process; an observation that seems to be even stronger in vivo than in vitro.
Not all fiber-reinforced materials share the same characteristics. In the recent years several studies have shown the advantages of bioactive glass-fiber reinforced implants17. These implants can be modified in the way to promote ossification18 and to inhibit bacterial growth and biofilm formation19.
Interestingly, in the CFR-PEEK group there were two cases with relapse of spondylodiscitis, while in the matched titanium group no recurrence was reported. In both cases of relapse spondylodiscitis no bacterium was detected during revision surgery despite sonication. This suggests that there might be another virulence factor aside from bacterial adhesion which in combination with CFR-PEEK promotes infection relapse and screw loosening. The idea that there might be more wear particles generated by CFR-PEEK compared to standard material which could cause tissue inflammation was rejected by studies from the orthopedic field20,21.
In only one case the sonication of the explanted material a bacterium could be detected. This bacterium was Cutibacterium acnes on CFR-PEEK material from a spondylodiscitis initially caused by MSSA. This implicates that antibiotic treatment successfully eliminated the initial causative bacterium in all cases despite the rough surface.
Cutibacterium acnes was identified as one of the main species causing pedicle screw loosening because of subclinical low-grade infection22,23,24.
So low-grade infection of CFR-PEEK may play a role in screw loosening.
Screw loosening and fusion techniques
In a series of 22 primary spinal tumor cases who underwent spinal fusion using CFR-PEEK only one case of screw loosening was found at 12 months after operation7. In our study, for patients with spinal infection screw loosening of CFR-PEEK screws was found more frequently and was registered at an earlier time point (mean 110 ± 81 days) compared to the cases with screw loosening found in the series of tumor cases7. Spinal metastases or primary spinal tumors are usually located in the vertebral body, whereas spondylodiscitis affects the intervertebral disc and adjacent endplates. So in tumor surgery the affected vertebrae are not instrumented with pedicle screws7, while in our series of spondylodiscitis cases pedicle screws were placed in the infected vertebrae.
Figure 2 shows the CT-scan of a patient, who initially suffered from spondylodiscitis of level L4/5, which was instrumented with CFR-PEEK screws on the levels L4–5 and an intercorporal fusion from lateral (XLIF). After one month instrumentation was extended to S1 and S2 with CFR-PEEK screws because of screw loosening of L5. However, three months afterwards he presented again with screw loosening of L4 and L5 (Fig. 2), what lead to explant of the whole material. Figure 3 shows the CT-scan of a patient who initially suffered from spondylodiscitis of the level T7/8. CFR-PEEK instrumentation involved T5–6–9–10 and a vertebral body replacement of T7/8 from transthoracal. After one month he presented with spondylodiscitis of T10/11. Instrumentation was extended with CFR-PEEK screws to T11–12 and intercorporal fusion was performed from lateral. Six weeks later he presented with extreme screw loosening of T11 and T12 (Fig. 3). Screws of T1 and T12 were removed and instrumentation was extended with titanium screws to L1–2–3. These two cases of severe spondylodiscitis show that screw loosening was especially found where screws were inserted in the affected vertebrae. From our point of view this is a risk factor for screw loosening of CFR-PEEK screws, especially in severe cases of spondylodiscitis. A reason for this may be an ongoing inflammatory reaction leading to a segmental instability.
Screw loosening and mechanical loading
Inadequate anterior support and osteoporosis are commonly regarded as risk factors for screw loosening after spinal instrumentation25. However, number of patients with osteoporosis was even higher in the titanium group than in the CFR-PEEK group. Regarding 360° fusion procedures the groups were equal as it was a matching criterion. Within the cases where screw loosening was observed in the CFR-PEEK group, 83% underwent 360° fusion; suggesting that inadequate anterior support did not play a role in screw loosening after CRF-PEEK instrumentation for spondylodiscitis.
Compared to 108 spondylodiscitis patients with titanium instrumentation operated from January 2014 until December 2016, 360° fusion techniques were used significantly more frequently with CFR-PEEK instrumentation. When compared to titanium CFR-PEEK has a lower elastic module which is similar to that of cortical bone26,27,28. Due to this higher elasticity 360° fusion was used more frequently with CFR-PEEK instrumentation. An elastic module similar to that of cortical bone was considered to be an advantage of CFR-PEEK because it avoids a mechanical mismatch during ossification2. In cadaver studies it was shown that CFR-PEEK pedicle screws resisted a similar cyclic loading compared to titanium screws29. Successful results of CFR-PEEK cages exist including cases of spinal infection2. However, in one third of the reported cases an additional instrumentation with titanium or steel from posterior was performed.
Strengths
One strength of our study is its novelty and its consecutive series. Follow-up results of a similar cohort of patients treated with CFR-PEEK instrumentation for spondylodiscitis do not exist so far.
Limitations
One limitation of our study is the fact that it was not a randomized controlled trial. However, with the matched analysis we provided a high level of objective comparability in one center with one treatment standard for spondylodiscitis. Another limitation is the fact that this is a study of a small number of cases. However, even this small number has an impact on clinical practice. It is unlikely that there will be a comparable prospective study with a higher number of cases. Since we know that there is more screw loosening with CFR-PEEK screw in spondylodiscitis cases, this will not be justifiable from an ethical point of view.
Conclusion
As opposed to other indications like tumor, degeneration and fractures, CFR-PEEK screws show a higher rate of screw loosening than titanium screws in spinal infections in this series with two potentially underlying reasons. A probably stronger bacterial adhesion on CFR-PEEK in vivo as shown by a statistical trend in vitro and instrumentation of spondylytic vertebrae, especially in severe cases of spondylodiscitis.
Until these factors are validated, we advise caution when implanting CFR-PEEK screws in infectious cases; if doing so, instrumentation of spondylitic vertebrae should be avoided.
Data availability
Data supporting the findings of this study are available within this article.
References
Li, C. S., Vannabouathong, C., Sprague, S. & Bhandari, M. The use of carbon-fiber-reinforced (CFR) PEEK material in orthopedic implants: A systematic review. Clin. Med. Insights Arthritis Musculoskelet. Disord. 8, 33–45. https://doi.org/10.4137/CMAMD.S20354 (2015).
Heary, R. F., Kheterpal, A., Mammis, A. & Kumar, S. Stackable carbon fiber cages for thoracolumbar interbody fusion after corpectomy: Long-term outcome analysis. Neurosurgery 68, 810–818. https://doi.org/10.1227/NEU.0b013e3182077a9f (2011) (discussion 818–819).
Brantigan, J. W. & Steffee, A. D. A carbon fiber implant to aid interbody lumbar fusion. Two-year clinical results in the first 26 patients. Spine 18, 2106–2107. https://doi.org/10.1097/00007632-199310001-00030 (1993).
Rousseau, M. A., Lazennec, J. Y. & Saillant, G. Circumferential arthrodesis using PEEK cages at the lumbar spine. J. Spinal Disord. Tech. 20, 278–281. https://doi.org/10.1097/01.bsd.0000211284.14143.63 (2007).
Bruner, H. J. et al. Biomechanics of polyaryletherketone rod composites and titanium rods for posterior lumbosacral instrumentation. Presented at the 2010 joint spine section meeting. Laboratory investigation. J. Neurosurg. Spine 13, 766–772. https://doi.org/10.3171/2010.5.SPINE09948 (2010).
Krieg, S. M. et al. Topping-off technique for stabilization of lumbar degenerative instabilities in 322 patients. J. Neurosurg. Spine 32, 1–7. https://doi.org/10.3171/2019.8.SPINE181434 (2019).
Tedesco, G., Gasbarrini, A., Bandiera, S., Ghermandi, R. & Boriani, S. Composite PEEK/carbon fiber implants can increase the effectiveness of radiotherapy in the management of spine tumors. J. Spine Surg. 3, 323–329. https://doi.org/10.21037/jss.2017.06.20 (2017).
Ringel, F. et al. Radiolucent carbon fiber-reinforced pedicle screws for treatment of spinal tumors: Advantages for radiation planning and follow-up imaging. World Neurosurg. 105, 294–301. https://doi.org/10.1016/j.wneu.2017.04.091 (2017).
Nevelsky, A., Borzov, E., Daniel, S. & Bar-Deroma, R. Perturbation effects of the carbon fiber-PEEK screws on radiotherapy dose distribution. J. Appl. Clin. Med. Phys. 18, 62–68. https://doi.org/10.1002/acm2.12046 (2017).
Herren, C. et al. Spondylodiscitis: Diagnosis and treatment options. Dtsch. Arztebl. Int. 114, 875–882. https://doi.org/10.3238/arztebl.2017.0875 (2017).
Rutges, J. P., Kempen, D. H., van Dijk, M. & Oner, F. C. Outcome of conservative and surgical treatment of pyogenic spondylodiscitis: A systematic literature review. Eur. Spine J. 25, 983–999. https://doi.org/10.1007/s00586-015-4318-y (2016).
Shiban, E. et al. A retrospective study of 113 consecutive cases of surgically treated spondylodiscitis patients. A single-center experience. Acta Neurochir. (Wien) 156, 1189–1196. https://doi.org/10.1007/s00701-014-2058-0 (2014).
Kowalski, T. J., Berbari, E. F., Huddleston, P. M., Steckelberg, J. M. & Osmon, D. R. Do follow-up imaging examinations provide useful prognostic information in patients with spine infection?. Clin. Infect. Dis. 43, 172–179. https://doi.org/10.1086/505118 (2006).
Shiban, E. et al. Analysis of bacterial adherence of clinical relevant staphylococcus species on carbon/PEEK composite vs standard titanium: Possible implications in spinal instrumentation for spondylodiscitis. Data presented as a poster at: EANS; Oktober 21–25, 2018; Brussels, Belgium (2018).
Murillo, O. et al. The changing epidemiology of bacteraemic osteoarticular infections in the early 21st century. Clin. Microbiol. Infect. 21(254), e251-258. https://doi.org/10.1016/j.cmi.2014.09.007 (2015).
Doutchi, M. et al. Changing trends in the epidemiology of vertebral osteomyelitis in Marseille, France. New Microbes New Infect. 7, 1–7. https://doi.org/10.1016/j.nmni.2015.04.008 (2015).
Piitulainen, J. M., Mattila, R., Moritz, N. & Vallittu, P. K. Load-bearing capacity and fracture behavior of glass fiber-reinforced composite cranioplasty implants. J. Appl. Biomater. Funct. Mater. 15, e356–e361. https://doi.org/10.5301/jabfm.5000375 (2017).
Vallittu, P. K. et al. Biomaterial and implant induced ossification: In vitro and in vivo findings. J. Tissue Eng. Regen. Med. 14, 1157–1168. https://doi.org/10.1002/term.3056 (2020).
Cabal, B. et al. A new biocompatible and antibacterial phosphate free glass-ceramic for medical applications. Sci. Rep. 4, 5440. https://doi.org/10.1038/srep05440 (2014).
Utzschneider, S. et al. Inflammatory response against different carbon fiber-reinforced PEEK wear particles compared with UHMWPE in vivo. Acta Biomater. 6, 4296–4304. https://doi.org/10.1016/j.actbio.2010.06.002 (2010).
Steinberg, E. L. et al. Carbon fiber reinforced PEEK Optima–a composite material biomechanical properties and wear/debris characteristics of CF-PEEK composites for orthopedic trauma implants. J. Mech. Behav. Biomed. Mater. 17, 221–228. https://doi.org/10.1016/j.jmbbm.2012.09.013 (2013).
Richards, B. R. & Emara, K. M. Delayed infections after posterior TSRH spinal instrumentation for idiopathic scoliosis: Revisited. Spine 26, 1990–1996. https://doi.org/10.1097/00007632-200109150-00009 (2001).
Richards, B. S. Delayed infections following posterior spinal instrumentation for the treatment of idiopathic scoliosis. J. Bone Joint Surg. Am. 77, 524–529. https://doi.org/10.2106/00004623-199504000-00004 (1995).
Leitner, L. et al. Pedicle screw loosening is correlated to chronic subclinical deep implant infection: A retrospective database analysis. Eur. Spine J. 27, 2529–2535. https://doi.org/10.1007/s00586-018-5592-2 (2018).
Galbusera, F. et al. Pedicle screw loosening: A clinically relevant complication?. Eur. Spine J. 24, 1005–1016. https://doi.org/10.1007/s00586-015-3768-6 (2015).
Niinomi, M., Liu, Y., Nakai, M., Liu, H. & Li, H. Biomedical titanium alloys with Young’s moduli close to that of cortical bone. Regen. Biomater. 3, 173–185. https://doi.org/10.1093/rb/rbw016 (2016).
Ciappetta, P., Boriani, S. & Fava, G. P. A carbon fiber reinforced polymer cage for vertebral body replacement: Technical note. Neurosurgery 41, 1203–1206. https://doi.org/10.1097/00006123-199711000-00040 (1997).
Qin, W., Li, Y., Ma, J., Liang, Q. & Tang, B. Mechanical properties and cytotoxicity of hierarchical carbon fiber-reinforced poly (ether-ether-ketone) composites used as implant materials. J. Mech. Behav. Biomed. Mater. 89, 227–233. https://doi.org/10.1016/j.jmbbm.2018.09.040 (2019).
Lindtner, R. A., Schmid, R., Nydegger, T., Konschake, M. & Schmoelz, W. Pedicle screw anchorage of carbon fiber-reinforced PEEK screws under cyclic loading. Eur. Spine J. 27, 1775–1784. https://doi.org/10.1007/s00586-018-5538-8 (2018).
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Contributions
A.J. conception of the study, acquisition, analysis and interpretation of data, writing, review and revision of the manuscript; E.S. conception of the study; S.M.K. conception of the study, analysis and interpretation of data, writing, review and revision of the manuscript, corresponding author; B.M. conception of the study, interpretation of data, writing, review and revision of the manuscript.
Corresponding author
Ethics declarations
Competing interests
ES reports personal fees (as a consultant) from Nevro (Redwood City, USA) and from Brainlab (Munich, Germany). SMK is a consultant for Nexstim Plc (Helsinki, Finland) and Spineart Deutschland GmbH (Frankfurt, Germany) and received honoraria from Medtronic (Meerbusch, Germany) and Carl Zeiss Meditec (Oberkochen, Germany). SMK and BM received research grants and are consultants for Brainlab AG (Munich, Germany). BM received honoraria, consulting fees, and research grants from Medtronic (Meerbusch, Germany), icotec ag (Altstätten, Switzerland), and Relievant Medsystems Inc. (Sunnyvale, CA, USA), honoraria, and research grants from Ulrich Medical (Ulm, Germany), honoraria and consulting fees from Spineart Deutschland GmbH (Frankfurt, Germany) and DePuy Synthes (West Chester, PA, USA), and royalties from Spineart Deutschland GmbH (Frankfurt, Germany). AJ reports no conflicts of interest.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Joerger, AK., Shiban, E., Krieg, S.M. et al. Carbon-fiber reinforced PEEK instrumentation for spondylodiscitis: a single center experience on safety and efficacy. Sci Rep 11, 2414 (2021). https://doi.org/10.1038/s41598-021-81960-8