Abstract
Very high temporal and spatial resolution is mandatory for the diagnosis of arteriovenous malformations (AVM) of the hand. Until now, magnetic resonance imaging (MRI) has not fulfilled both requirements simultaneously. This study presents how the combination of a very fast TWIST MRI (time-resolved angiography with interleaved stochastic trajectories) sequence and iterative reconstructions optimizes temporal as well as spatial resolution. 11 patients were examined at a 3-T MRI scanner with two different TWIST protocols: the standard and the study protocol, acquiring a data set every 5.57 s and 1.44 s respectively. The study data was retrospectively iteratively reconstructed with different regularization factors (0.001, 0.002, 0.004, 0.008). Results were compared using the sign-test. P-values < 0.05 were regarded statistically significant. With a low amount of contrast medium, the temporal resolution of the study protocol enabled the differentiation of arteries from veins in all patients whereas the signal-to-noise ratio (SNR) deteriorated. Depending on the regularization factors, SNR, delineation of arterial feeders and non-involved hand and interdigital arteries, as well as artefact levels varied. Overall, iterative reconstruction with regularization factor 0.004 achieved the best results, consequently showing the ability of MRI as a reliable diagnostic method in AVMs of the hand.
Similar content being viewed by others
Introduction
According to Mulliken et al.1, vascular anomalies are divided into vascular tumours and vascular malformations. The latter ones occur in approximately 0.3–0.5% of the population2. Based on their rheological behaviour, they are subcategorized into high-flow and low-flow malformations3. Although representing the most common type of high-flow malformations, peripheral arteriovenous malformations (AVMs) are very rare and occur in 14.3% of patients suffering from vascular malformations4. Of these, 28.5% of peripheral AVMs are located on the upper and lower extremities, mostly sparing the feet and hands4. AVMs consist of a direct shunting area between arteries and veins. The inner centre of an AVM is the reticulate nidus with fine-spun shunts between tiny arteries and veins. Patients with AVMs may suffer from pain, aesthetical impairment, ulcer, bleeding, and increased cardiac output5,6.
Radiological interventional therapy of AVM aims for the complete occlusion of the nidus and is a well-established treatment option7,8. However, the occlusion of non-involved arterial vessels that provide the blood supply of tissue distal of the AVM has to be prevented to avoid tissue necrosis or amputation in the worst case6. Most importantly, the angioarchitecture and hemodynamics of the AVM have to be fully understood for treatment planning9. As the blood flow in AVMs is very fast and diameters of vessels may be very small, the perfect imaging modality for intervention planning needs to combine an extremely high temporal and a high spatial resolution10,11.
Both can be achieved with conventional digital subtraction angiography (DSA)11,12,13. The disadvantages of DSA are radiation exposure, iodine-contrast medium injection, and the technique’s inherent invasiveness with the risk of bleeding and vessel disruption10. Furthermore, morphologic information about the surrounding tissue is lacking.
A generally accepted imaging alternative is dynamic contrast-enhanced magnetic resonance angiography (MRA)14. After the injection of intravenous gadolinium contrast repeated MRA data sets are acquired15. For the imaging of high-flow vessels, a high acquisition speed is mandatory. Techniques like TWIST (time-resolved angiography with interleaved stochastic trajectories) MRA, which images k-space centre more frequently than peripheral portions, achieve a high temporal resolution15. Contrary to single-phase MRA, TWIST MRA is independent of the perfect contrast bolus timing16. However, side effects of these new techniques accelerating the image rate are blurring, ghosting, and a loss of signal-to-noise ratio (SNR), which interferes with the image quality10 and masks smaller vessels and their hemodynamics especially in peripheral anatomic regions.
Recent studies revealed that iterative reconstructions seem to be beneficial for MRA by enabling fast sequences and improving the visualization and demarcation of micro vessels15,17. AVMs exclusively located on the hand are difficult to image and have never been examined in this setting because of the small size of the feeding and draining vessels and the fact that these are mostly terminal vessels. Accordingly, this study investigated the temporal and spatial resolution of a very fast TWIST study protocol of AVMs involving the hand. Due to the high-flow situation in extremely tiny vessels, the correct depiction of hemodynamics is challenging. Study data sets were retrospectively reconstructed with four different regularization factors and results were compared to our standard TWIST protocol.
Thereby, we examined the value of the study TWIST MRA in improving the temporal resolution in high-flow vessels. Furthermore, we investigated if iterative reconstructions with different regularization factors are able to optimize the study protocol, especially in delineating very small vessels and in diminishing artefacts.
Methods
Institutional Review Board approval for this study was obtained.
From 02/2016 to 10/2016, 11 consecutive patients who presented at our tertiary care Vascular Anomalies Centre (VAC) with AVMs of the hand or the fingers where included in this study and examined according to a predefined protocol with a 16-channel hand/wrist coil at a 3-T MR system (MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany)14.
MRA was indicated for interventional radiological treatment planning or for follow-up imaging after therapy.
All patients gave their written informed consent for the MRI examination including intravenous application of contrast medium. Contraindications for intravenous contrast medium applications like allergies or kidney disease were ruled out. The cumulative applied dose of contrast medium was 0.1 mL/(kg body weight) per patient (Gadobutrol, Gadovist, Bayer vital, Leverkusen, Germany).
Standard and study protocol were executed during the same examination. Primarily the standard TWIST protocol was performed. For every affected level of the upper extremity, like the upper arm, the forearm, or the hand, a separate contrast medium injection was executed. Afterwards the study protocol with one contrast medium administration for the hand was performed. Thus, the maximal single dose of contrast medium was divided through the number of affected levels for the standard protocol plus one injection for the study protocol, resulting in the available contrast medium amount per injection.
Contrast medium was administered by power injector with a flow of 2 mL/s.
According to the routine protocol using parallel imaging with an acceleration factor of 3 (TWIST_3) in phase-encoding-direction, a 3D data set was reconstructed every 5.57 s with a measured voxel size of 0.77 mm × 0.70 mm × 0.83 mm.
The study TWIST protocol was applied with an acceleration factor of 12 (TWIST_12): 4 (phase-encoding-direction) × 3 (in partition direction). A 3D data set was reconstructed using a prototype iterative approach15 each 1.44 s with a measured voxel size of 0.72 mm × 0.72 mm × 0.72 mm. Image calculation was initially performed with the standard view sharing technique. In addition to the product reconstruction, retrospective iterative reconstructions of the same datasets were performed using four different regularization factors (0.001, 0.002, 0.004, 0.008). Thus, five differently reconstructed study protocols were assessed.
Standard TWIST alternates the acquisition of k-space center (A) region and differently undersampled k-space periphery (B) regions, where combining multiple B regions results in a regular undersampling suitable for standard parallel imaging reconstruction, i.e. a sliding window covering one A and multiple B regions. According to Wetzl et al., for iterative reconstruction, consecutive single A – B pairs were reconstructed. Thus, a reduction of the temporal footprint and an increase of undersampling were achieved15. To recover the frames \(\{\widehat{x}_{t}\}\)t=1,…,T for all time points T, a nonlinear, iterative SENSE-type reconstruction with spatio-temporal wavelet regularization was performed:
The abbreviation C represents the number of coils and At the sampling pattern for time t. F is the Fourier transform, Sc the multiplication by the sensitivity of coil c, yt,c the measured data for time t and coil c (from an A and B region), λσ and λτ are the spatial and temporal regularization parameters and Wσ and Wτ are spatial and temporal redundant Haar wavelet transforms 15. The temporal and spatial regularization factors were related by a fixed factor λτ = 5λσ. All regularization factors in the manuscript refer to the spatial factor. The regularization coefficients were scaled by the maximal intensity in each dataset to make the selection of regularization parameters independent of intensity variations across subjects.
Reconstruction was performed at the scanner directly utilizing its graphic processing unit (GPU).
For comparison between TWIST_3, TWIST_12, and TWIST_12 reconstructed with four different regularization factors, the following parameters were analysed on the maximum intensity projections (MIPs):
Delay
Dependent on the extent of the AVM, a region of interest (ROI) was placed in the radial/ulnar artery and the adjacent dominant outflow vein; time-intensity curves were calculated for each ROI and time lag between arterial and venous peak enhancement was determined. Results between the TWIST_3 standard protocol and the study protocol iteratively reconstructed with the regularization factor 0.004 were compared with each other.
SNR
Three different ROIs (area: 0.02 cm2 each) were placed within a major hand artery on TWIST_12 and all iteratively reconstructed data sets at exactly the same position as soon as the vessel was visible. The mean signal intensity was divided by the standard deviation within these ROIs, thus an apparent “vessel SNR” was calculated; the mean SNR out of three ROIs within one data set was calculated. Results of the product were compared with every retrospectively reconstructed study data set.
Two radiologists with 8 and 6 years of experience in the field of vascular malformations evaluated the MIPs of the TWIST_3 as well as the product and the iteratively reconstructed TWIST_12 data sets visually in consensus on an imaging workstation. They were blinded to the patients’ medical history and assessed 5 different categories:
Non-involved interdigital arteries
Delineation of healthy interdigital arteries was assessed in an unaffected finger according to a four-point Likert scale (uniform, nearly uniform, partial delineation, or no delineation).
Non-involved hand arteries
Healthy, non-involved hand arteries, preferably in the deep palmar arch, were evaluated and graded to a four-point scale (uniform, nearly uniform, partial delineation, no delineation).
Arterial feeders
Identifiability and delineation of arteries feeding the AVM nidus were evaluated (yes/no).
Bleeding
Bleeding or blurring artefacts restricting the exact and sharp demarcation of the radial or ulnar artery were assessed on a three-point scale (not present; present, but not diagnostically relevant; present and diagnostically relevant).
Ghosting artefacts
Multiple reflected, displaced blood vessels (ghosting artefact) were assessed on a three-point scale (not present; present, but not diagnostically relevant; present and diagnostically relevant).
Statistical analysis was performed using the sign-test (IBM SPSS STATISTICS 64-bit MS Windows 23.0.0.0); p-values < 0.05 were regarded statistically significant; results of the TWIST_3 and TWIST_12 data sets were compared with each other as well as every retrospectively reconstructed study data set with the product reconstructed.
Ethics approval
This was a retrospective single-center consecutive case series. IRB approval was obtained by the Ethikkommission of University of Regensburg, University Medical Center Regensburg, Franz-Josef-Strauss-Allee 11, 93053 Regensburg, Germany; https://ethikkommission.uni-regensburg.de. Individual consent for inclusion in the review was waived by the Research Ethics Board because of the retrospective character of the study.
We confirm that all methods were carried out in accordance with relevant guidelines and regulations.
Informed consent
Informed consent for MRI was obtained.
Results
11 patients (n = 9 females, n = 2 males) with an average age of 43.9 years (min: 27.3 years, max: 78.3 years) were included in the study. MRI data acquisition was successful in all patients. In 1 patient, the iterative reconstruction with a regularization factor of 0.008 was not available.
TWIST_12 MRA of the hand was acquired with 39.8% on average of a single dose gadolinium (20.0–50.0%) corresponding to a total amount of 7.5 mL of contrast medium. Calculation time for the iterative reconstruction took about 8 min 30 s for 50 volumes with a 320 × 320 matrix size.
Delay
Concerning the standard TWIST_3 protocol, three patients demonstrated contrast-medium enhancement of major arteries and dominant outflow veins simultaneously on the same MR data set, thus making discrimination impossible. Using TWIST_12 and a regularization factor of 0.004, in all patients a differentiation between major arteries and dominant outflow veins was possible with a gap of 1–9 data sets between arterial and venous phase (Fig. 1).
SNR
Using iterative reconstructions, SNR was improved by a mean factor of 2.1 (1.9 – 2.3) compared to the TWIST_12 study protocol without iterative reconstructions. SNR improvement declined with increasing regularization factor (Table 1, Fig. 2, ESM Diagram 1).
Non-involved interdigital arteries
There was no relevant difference in the delineation of normal finger arteries on TWIST_3 and TWIST_12 protocols (Table 1). Iterative reconstructions did not improve uniform delineation of healthy finger arteries. On the contrary, after iterative reconstruction with the regularization factors 0.001 and 0.008, results worsened significantly (p = 0.039); whereas regularization factors 0.002 (p = 0.18) and 0.004 (p = 0.289) led to a statistically not significant deterioration of results.
Non-involved hand arteries
Using the standard TWIST_3, uniform delineation of healthy hand arteries was possible in ten patients. Whereas on TWIST_12 images without iterative reconstruction the delineation of healthy hand arteries declined statistically significantly (p = 0.001) (Table 1), post-processing data with all regularization factors improved these results significantly (regularization factors 0.001 and 0.008: p = 0.002; regularization factors 0.002 and 0.004: p = 0.001).
Arterial feeders
Arterial feeders were detectable in eight patients on TWIST_3 and on TWIST_12 images, respectively. Iterative reconstructions with all regularization factors improved the detectability of arterial feeders in comparison to TWIST_12, but results did not differ significantly (regularization factors 0.002, 0.004, 0.008: p = 0.25; regularization factor 0.001: p = 0.625). The regularization factors 0.002, 0.004, and 0.008 enabled the detection of arterial feeders in all patients (Table 1, Fig. 1).
Bleeding
On TWIST_3 and TWIST_12, no diagnostically relevant bleeding artefact was diagnosed. Results worsened significantly with the regularization factors 0.001 (p = 0.016) and 0.002 (p = 0.031). Increasing the regularization factor helped to extinguish those disadvantages (regularization factor 0.004: p = 0.125; regularization factor 0.008: p = 1.00) (Table 1).
Ghosting artefacts
On TWIST_3 and TWIST_12 MRA, no diagnostically relevant ghosting artefacts were present. With low regularization factors (0.001 and 0.002), ghosting artefacts deteriorated significantly (p = 0.008). Higher regularization factors improved ghosting artefacts (regularization factor 0.004: p = 0.07; regularization factor 0.008: p = 1.00) (Fig. 2, Table 1).
Discussion
Conventional DSA today still presents the gold standard for the imaging of AVMs11. However, this technique has the disadvantages of invasiveness, radiation dose, and iodinated contrast-medium10. MRA represents an alternative non-invasive imaging modality, but the exact delineation of subtle hemodynamic and anatomical detail of AVMs is still challenging12. Two different aspects have to be considered: morphologic information—in terms of a high spatial resolution—is important for three-dimensional understanding of AVMs and thus for treatment planning. Furthermore, the exact analysis of the hemodynamics is crucial12. As AVMs are high-flow malformations consisting of multiple small direct shunts between arteries and veins, a high temporal resolution is mandatory for the understanding of blood flow. Furthermore, it is important to delineate the blood supply of the nidus as well as of tissue located distal to the nidus in order to embolize the nidus without the risk of distal tissue necrosis and potential amputation6. To date, the benefit of TWIST MRA and parallel MR imaging techniques has been studied for information regarding morphologic and hemodynamic parameters of the extracranial carotid circulation. Acquisition speed increased generally, but artefacts restricted the results10.
A well-known disadvantage of these very fast sequences is the loss of signal, resulting in a low SNR as well as blurring of small vessels 18. Wetzl et al. have shown advantages of iterative reconstructions especially in improving the delineation of small vessels of the lung15.
In our institution, a TWIST_3 standard protocol with a 3D data set reconstructed every 5.57 s was established for treatment planning and post-therapeutically imaging of AVMs. Whereas morphologic information was deemed to be adequate in the clinical routine, the temporal resolution with a data set every 5.57 s constituted a lack of information regarding the exact differentiation of arteries from veins in this fast flow situation.
For optimizing the temporal resolution, we set up a very fast TWIST_12 study protocol depicting data sets every 1.44 s and examined AVMs of the hand19. Thus, we wanted to study the method´s reliability and robustness in preferably small vessels. Furthermore, we investigated the additional value of different iterative reconstructions using four different regularization factors. Comparison between reconstructed sequences with the product TWIST_12 reconstruction was performed.
To the best of our knowledge, this is the first study to examine the effects, advantages, and disadvantages of rapid sequences with iterative reconstructions with different regularization factors on AVMs of the hand.
As we had hypothesized, the major advantage of the study protocol was the high temporal resolution enabling to differentiate between arteries and veins in all patients.
On TWIST_12 data sets, diagnostically relevant artefacts were absent. Non-involved interdigital arteries were uniformly definable in 45% of patients in comparison to 27% of patients on TWIST_3 standard images. Disadvantages of the study protocol were the low SNR, the lack of uniform delineation of non-involved hand arteries, and the detectability of arterial feeders in only 73% of patients.
Whereas the regularization factors 0.001 and 0.002 improved the SNR, the former one maximally, diagnostically relevant artefacts were significantly more frequent in comparison to TWIST_12 (bleeding: p = 0.016 and p = 0.031, ghosting: p = 0.008 and p = 0.008 respectively).
The regularization factor of 0.004 led to a less distinct improvement of SNR compared to iterative reconstructions with the factor 0.001. The uniform delineation of non-involved hand arteries improved significantly (p = 0.001) in comparison to the unmodified TWIST_12, arterial feeders were detectable in all patients, and diagnostically relevant artefacts decreased in comparison to iterative reconstruction with lower regularization factors.
The regularization factor of 0.008 improved the SNR in comparison to the study protocol, but to a lesser degree than all other regularization factors. Non-involved hand arteries (p = 0.002) and arterial feeders were detectable at a high rate. Furthermore, diagnostically relevant artefacts were extinguished.
Contrary to the results of other study groups, where the delineation of very small vessels improved after iterative reconstructions15, the delineation of non-involved interdigital arteries worsened with higher regularization factors resulting in a lack of its delineation in 70% of patients after reconstructing with a factor of 0.008.
Considering the uniform and nearly uniform delineation of non-involved interdigital arteries as diagnostically accurate, the best result was TWIST_3 in 82% of patients succeeded by TWIST_12 with regularization factor 0.002 (55%), and TWIST_12 with product reconstruction (45%).
Thus, we sub-classified all assessed findings into diagnostically accurate and into non-diagnostic categories and summarized results. Regarding the delineation of non-involved interdigital and hand arteries, the uniform and the nearly uniform depiction was regarded as diagnostically accurate. Furthermore, the identification of arterial feeders and the absence of relevant artefacts were deemed as diagnostically accurate. Results of diagnostically accurate results are displayed in ESM Diagram 2.
The best overall results were achieved with the regularization factors 0.004 and 0.008. Whereas the factor 0.008 extinguished all artefacts completely, it had the disadvantage of a lesser SNR and a not accurate delineation of non-involved interdigital arteries. Iterative reconstructions with regularization factor 0.004 outperformed the factor 0.008 in both categories on account of a higher rate of artefacts.
The results in this study were achieved with an average dose of 39.8% of a single dose of contrast medium only. Thus, iterative reconstructions might help to reduce the required contrast medium amount. Taking into account the detection of gadolinium deposition in the brain after repetitive gadolinium applications for MRI, a low dose of gadolinium has to be achieved, especially in this young patient group facing—due to their chronic disease—further contrast-medium injections for MRI20.
Despite the retrospective character of this study, results are promising for accurate imaging and delineation of AVM of the hand and might also be transferred for imaging of AVM of the feet. However, further investigations in a larger study population is necessary.
In conclusion, a high temporal and spatial resolution is mandatory for delineation and understanding of hemodynamic features in AVMs. Temporal resolution is sufficient using an extraordinarily high acceleration factor of 12, but is associated with decrease in SNR. The postprocessing with iterative reconstructions improves the SNR and results in diagnostically reliable data sets, and reduces volume of applied contrast medium.
Data availability
All data is included in the manuscript in terms of Table 1.
References
Mulliken, J. B. & Glowacki, J. Hemangiomas and vascular malformations in infants and children: A classification based on endothelial characteristics. Plast. Reconstr. Surg. 69, 412–422 (1982).
Redondo, P., Aguado, L. & Martinez-Cuesta, A. Diagnosis and management of extensive vascular malformations of the lower limb: Part I. Clinical diagnosis. J. Am. Acad. Dermatol. 65, 893–906 (2011).
ISSVA Classification of Vascular Anomalies © 2018 International Society for the Study of Vascular Anomalies. issva.org/classification. Accessed 5 Aug 2020.
Greene, A. K., Liu, A. S., Mulliken, J. B., Chalache, K. & Fishman, S. J. Vascular anomalies in 5,621 patients: Guidelines for referral. J. Pediatr. Surg. 46, 1784–1789 (2011).
Mulliken, J. B., Fishman, S. J. & Burrows, P. E. Vascular anomalies. Curr. Probl. Surg. 37, 517–584 (2000).
Webb, J. B., O’Brien, M., John, P. R. & Nishikawa, H. Early presentation of an extremity arteriovenous malformation. Br. J. Plast. Surg. 57, 785–788 (2004).
Yakes, W. F., Pevsner, P., Reed, M., Donohue, H. J. & Ghaed, N. Serial embolizations of an extremity arteriovenous malformation with alcohol via direct percutaneous puncture. AJR Am. J. Roentgenol. 146, 1038–1040 (1986).
Wohlgemuth, W. A. et al. The retrograde transvenous push-through method: a novel treatment of peripheral arteriovenous malformations with dominant venous outflow. Cardiovasc. Intervent. Radiol. 38, 623–631 (2015).
Wohlgemuth, W. A., Wolfle, K., Schuster, T., Schlimok, G. & Bohndorf, K. Hereditary vascular malformations: Classification, symptoms, diagnostics and prognosis. Zentralbl. Chir 137, 440–445 (2012).
Lim, R. P. et al. 3D time-resolved MR angiography (MRA) of the carotid arteries with time-resolved imaging with stochastic trajectories: comparison with 3D contrast-enhanced bolus-chase MRA and 3D time-of-flight MRA. AJNR Am. J. Neuroradiol.29 (2008).
Lee, B. B. et al. Consensus Document of the International Union of Angiology (IUA)-2013. Current concept on the management of arterio-venous management. Int. Angiol.32, 9–36 (2013).
Herborn, C. U. et al. Comprehensive time-resolved MRI of peripheral vascular malformations. AJR Am. J. Roentgenol. 181, 729–735 (2003).
Dunham, G. M., Ingraham, C. R., Maki, J. H. & Vaidya, S. S. Finding the nidus: Detection and workup of non-central nervous system arteriovenous malformations. Radiographics 36, 891–903 (2016).
Hammer, S. et al. Time-resolved magnetic resonance angiography (MRA) at 3.0 Tesla for evaluation of hemodynamic characteristics of vascular malformations: description of distinct subgroups. Eur. Radiol.27, 296–305 (2017).
Wetzl, J. et al. High-resolution dynamic CE-MRA of the thorax enabled by iterative TWIST reconstruction. Magn. Reson. Med. 77, 833–840 (2017).
Gratz, M., Schlamann, M., Goericke, S., Maderwald, S. & Quick, H. H. Evaluation of fast highly undersampled contrast-enhanced MR angiography (sparse CE-MRA) in intracranial applications - Initial study. Eur. Radiol. 27, 1004–1011 (2017).
Stalder, A. F. et al. Highly undersampled contrast-enhanced MRA with iterative reconstruction: Integration in a clinical setting. Magn. Reson. Med. 74, 1652–1660 (2015).
Steeden, J. A., Pandya, B., Tann, O. & Muthurangu, V. Free breathing contrast-enhanced time-resolved magnetic resonance angiography in pediatric and adult congenital heart disease. J. Cardiovasc. Magn. Reson.17, 38, https://doi.org/10.1186/s12968-015-0138-9 (2015).
Fellner, C. et al. Does TWIST with Iterative Reconstruction Improve Diagnostics of AVM of the Hand? (Program 2793, ISMRM 25th Annual Meeting, Honolulu, 2017).
Miller, J. H., Hu, H. H., Pokorney, A., Cornejo, P. & Towbin, R. MRI brain signal intensity changes of a child during the course of 35 gadolinium contrast examinations. Pediatrics 136, 1637–1640 (2015).
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Contributions
V.I.H.: Investigation, Data analysis, Methodology, Validation, Writing original draft. C. F.: Conceptualization, Optimization of the study protocol, Investigation, Data analysis, Methodology, Project administration, Supervision, Validation, Visualization, Writing review and editing. W.A.W.: Conceptualization, Methodology, Project administration, Writing review and editing. C.S.: Supervision, Writing review and editing. M.S.: Optimization of the study protocol, Software, Writing review and editing. C.F.: Software development, Validation, Writing review and editing. J.W.: Software development, Validation, Writing review and editing. W.U.: Investigation, Methodology, Supervision, Validation, Data analysis, Writing review and editing.
Corresponding author
Ethics declarations
Competing interests
Michaela Schmidt, Christoph Forman, and Jens Wetzl are employees for Siemens Healthcare, Erlangen, Germany. The study patients were examined with a Siemens MR scanner; Siemens provided the prototype for iterative reconstruction. Veronika I. Huf, Claudia Fellner, Walter A. Wohlgemuth, Christian Stroszczynski, and Wibke Uller declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Huf, V.I., Fellner, C., Wohlgemuth, W.A. et al. Fast TWIST with iterative reconstruction improves diagnostic accuracy of AVM of the hand. Sci Rep 10, 16355 (2020). https://doi.org/10.1038/s41598-020-73331-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-020-73331-6
This article is cited by
-
Low-dose contrast-enhanced time-resolved angiography with stochastic trajectories with iterative reconstruction (IT-TWIST-MRA) in brain arteriovenous shunt
European Radiology (2022)
-
Assessment of superior cavo-pulmonary anastomoses (Glenn shunt) by cardiac magnetic resonance imaging in comparison with multi-slice computed tomography
Egyptian Journal of Radiology and Nuclear Medicine (2021)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.