Introduction

Three-dimensional(3D) reconstruction and modeling of hand morphology has been increasingly gaining importance for many purposes, including manufacturing of custom-made glove1, analysis of hand surface and volume2,3,4, and designing of hand surgery and humanoid robot hand5,6,7. Various methods of 3D imaging have been extensively applied in oral and maxillofacial surgery as well as plastic surgery8,9,10,11, because they collect surface data rapidly, non-invasively, accurately, reliably, and cost-effectively. These methods allow for the development of equally accurate 3D models by means of direct anthropometry, which is considered as the “gold standard”12,13.

Most of the currently available 3D scanning systems have been designed for the face or chest with 180–190 capture, i.e., 3dMDface (3dMD, USA)14 and DI3D (Dimensional Imaging, Scotland)9. In our previous study, we have successfully obtained 3D images of chest from 40 patients with pectus excavatum by using a Prime Sense 3D sensor15. However, the human hand is an annular stereoscopic structure with five fingers, which make it markedly different from arc-shaped structures. A full 360 scanning is essential for 3D reconstruction of hand. Moreover, it is difficult for an individual to keep hand still, even for a few seconds. Kau and Richmond stated that soft tissues was inherently difficult to study16, because they are inevitably affected by movement. Therefore, a support to hand is commonly used to keep still, i.e. the template and glass support7,17. However, the possibility of refraction errors of the glass support cannot be ruled out18,19. The plaster model instead of actual hand is another approach to avoid hand tremble during scanning. Yu et al. fixed the plaster hand on a rotatable disc1, and rotated 36° after each image capture. However, it was time-consuming to obtain ten images captured from different angles and combine them into an entire 3D image.

Therefore, to reconstruct the 3D hand image, it requires an accurate and convenient setup with a full 360° and rapid capture. Gemini structured light-scanning system was specially designed for hand based on the current structured light-scanning equipment. In this system, two modular units were installed on opposite sides of the spout with a PC-controller desktop to collect the images of both dorsal and palmar aspect of hand simultaneously. The projection grating and data collection could be completed within one second to minimize the error caused by movement of hand.

This study aims to identify an efficient approach for 3D photographs of hand. We obtained 3D photographs of actual hand by Gemini structured-light scanning system, and compared accuracy of two indirect hand measurement techniques(Gemini structured light scanning and CT scanning) with direct measurement integument hand features. To minimize the potential discrepancies caused by soft tissue deformation and movement associated with actual hand, we adopted the plaster hand models for measurement.

Subjects and methods

Subjects

The subjects included in this study were ten adults [6 males, 4 females; mean age 29.6 years, range 26–35 years, body mass index (BMI): 20.98–23.99 kg/m2], without any preexistent hand deformities or scars on their left hands. 10 plaster hand models of left hand were prepared in advance. Each participant was asked to hold fingers out straight with the fingers spread far apart to take a negative impression. Then, plaster was added to create the hand model. The study protocol was approved by the ethics committee of our institution. All participants provided written the informed consent and they did not receive any remuneration for participating in this study.

Methods

3D photographs of Gemini structured light-scanning

Gemini structured-light scanning system was designed by Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences. Both modular units are equipped with a machine vision camera and industrial-grade system (with resolution of < 0.1 mm), which are synchronized in two captures. The system has a single-amplitude measurement range of 176 × 276 mm and single amplitude measurement points of 1.9 million. The projection grating and data collection could be completed within one second. The 3D photographs were taken with the subject flexion in elbow and extension in wrist joint, and the hand was kept with the fingers in a fully abducted and extended position (Fig. 1). The subject was instructed to keep the left hand absolutely still and the images of both the palmar and dorsal aspect of hand were taken. The data generated by scan was saved in the original point cloud. Wrap software (Geomagic, USA) was used to splice, encapsulate, fill, and smoothen the surface of 3D hand photograph (Figs. 2, 3), and the data was generated and saved in .STL format ultimately.

Figure 1
figure 1

Scanning of actual hand by Gemini structured-light scanning system.

Figure 2, 3
figure 2

Encapsulation and filling of 3D hand model data by Wrap software.

Measurement of 3D photographs

Measurements of each plaster hand model were obtained by three methods respectively: (1) direct measurement (DM); (2) 3D CT scanning (CT scanning); and (3) Gemini 3D structured-light scanning (SL scanning). 24 identifiable landmarks were chosen based on Hoevenaren et al.’ study of 3D hand photograph7 (Table 1), which has proven that intra- and inter-observer reproducibility of the most of these landmarks were high. A series of linear measurements based on the landmarks were chosen for comparative measurements (Table 2), according the standard used by Yu et al.1 and Hoevenaren et al.20.

Table 1 Landmarks of 3D photographs.
Table 2 Description of 19 measurements made on the hand.

In DM, a measuring tape (with a minimum scale of 1 mm and thickness of 0.4 mm) was used for measuring the circumference dimensions and a sliding caliper (with an accuracy of 0.01 mm) was used for measuring the length dimensions. The CT scanning was constructed in the palm-up position of hand model with Neuviz 128 (Neusoft, China), and the slice thickness was 0.4 mm. The spatial resolution of X and Y axis is 0.29 mm and the Z axis is 0.32 mm. The acquired DICOM datasets were imported into Mimics Research software 19.0 (Materialise, Belgium). The 3D photographs of hand were generated and saved in the format of .STL. In SL scanning, the plaster hand model was placed in the upright position in the middle of two modular units. After processed by Wrap software, the 3D photographs were also saved in .STL format.

Figure 4, 5
figure 3

Process of circumference measurement.

In DM, the measurements were taken following ISO 7250-1:2008 standard. In SL and CT scanning methods, the 3-matic Research software (Materialise, Belgium) was used open .STL format, and measure the 3D hand model. The length dimension was the distance between two points. For circumference measurement, a horizontal surface was made crossing the middle of joint (Fig. 4), and removed the distal area with trimming method from entity (Fig. 5). The perimeter of cross section was the circumference of joint. Two weeks later, the measurements of three methods were obtained again in the same manner.

Figure 6, 7
figure 4

3D photographs of actual hand by SL scanning.

Figure 8, 9
figure 5

3D photographs of plaster hand model by SL scanning.

Statistics analysis

All statistical analyses were performed with IBM SPSS Statistics, version 22 (IBM, USA). The means and standard deviations were calculated for each measuring technique. As a measure of reliability, the intraclass correlation coefficient (ICC) for absolute agreement was calculated on the basis of a two-way random effects analysis of variance (ANOVA). ICCs of the repeated measurements were used to test the reliability of three measuring techniques. The measurements obtained from CT and SL scanning were then compared with those obtained from DM by using paired t tests. The probability level of P < 0.01 was considered to represent statistical significance. The differences in the measurement between any of the methods greater than 0.5 mm were considered clinically significant.

The accuracy of the measurements obtained using the two 3D methods was expressed in terms of the absolute error (AE) and absolute percentage errors (APE). AE was defined as the measurement value subtracted by the reference value, which was calculated as the mean of the DM, whereas the two 3D values measurements were determined by the mean of the measurements made by the two independent examiners. APE was calculated using the following equation: APE = 100 × (AE/reference value).

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the ethics committee of Xiangya Hospital.

Informed consent

Informed consent was obtained from all individual participants included in the study.

Results

It reveals that Gemini structured light-scanning system can successfully establish a complete and smooth 3D hand photograph of actual hand (Figs. 6, 7) and plaster hand model (Figs. 8, 9). The average time of SL scanning and reconstruction of a hand model was 0.9 s and 2.4 min. In CT scanning of plaster hand model (Figs. 10, 11), the average time was 8.9 s for scanning and 6.5 min for reconstruction.

Figure 10, 11
figure 6

3D photographs of plaster hand model by CT scanning.

The results of DM between two observers were found to be very reliable (ICC; 0.925–0.995)(Table 3). The intra-observer reliability for CT scanning (ICC; 0.946–1.000) and SL scanning (ICC; 0.984–1.000) was also very high. Both CT and SL scanning showed higher reproducibility (P < 0.01, both) than DM. However, the two 3D scanning methods did not show any significant difference (P = 0.462).

Table 3 Intraclass correlation coefficients (ICC): intra-observer reliability of the three techniques.

For 6 of 19 measurements (31.58%), the significant differences were noted (P < 0.01) (Table 4). Significant differences were observed more often observed in the case of circumference dimensions (5/9, 55.56%) than in the length dimensions (1/10, 10%). In addition, CT scanning values were significantly greater for 3 of 4 (75.0%) measurements that differed; all values obtained with SL scanning were all significantly smaller than the DM.

Table 4 Mean and standard deviations (SD) of three anthropometric measuring techniques (all measurements in mm).

The mean and standard deviation of AE and the mean of APE were determined and the accuracy of the measurements was confirmed (Table 5). The mean AE for 10 subjects was very low both for CT scanning (0.29 ± 0.10 mm) and for SL scanning (0.30 ± 0.11 mm). The maximum AE and APE was 0.93 mm, 4.69 ± 2.33% and 0.94 mm, 4.88 ± 2.22% for CT scanning and SL scanning respectively.

Table 5 Absolute error (AE) and Absolute percentage error (APE) of the measuring techniques.

Discussion

To capture accurate 3D hand image rapidly with a full 360, we improved the design of structured light-scanning equipment. The Gemini structured-light scanning system is equipped with two modular units and allows for almost simultaneous capturing images of dorsal and palmar sides of hand. The design of double modular units is aimed to reduce experimental errors caused by the instability of the object. The system not only successfully established a complete and smooth 3D hand photograph, but also shortened the scanning and reconstruction time. The skin texture of the actual hand by SL scanning was also very delicate.

In this study, we investigated the reproducibility and accuracy of two 3D imaging techniques. To the best of our knowledge, this is the first study to compare SL scanning and CT scanning for their efficacy in 3D imaging of hand. Gemini structured-light scanning system is a new method. CT scanning is a routine technique used for diagnosis, reconstruction of soft tissue and planning of operation. Tan et al. introduced computed tomography angiography (CTA) for preoperative planning in complex toe-to-hand reconstruction21. In our team, CTA is also commonly used to map donor site and recipient vasculature22,23,24, and CT data were used to generate 3D models of complex soft tissue defect for individual design for perforator flap.

Three methods all showed high levels of intra-observer reliability: ICC; 0.925–0.995 for DM, ICC; 0.946–1.000 for CT scanning, and ICC; 0.984–1.000 for SL scanning. The plaster hand models instead of actual hand is one of the major reasons of high reliability. As the plaster hand models were rigid and fixed in shape, the influence of hand posture and movement on the results of hand measurement could be neglected. It may also prevent errors due to skin deformity associated with the use of the actual hand in DM. The reliability of two 3D measurements was greater than that achieved with DM. The tightness of the tape and the direction of the sliding caliper were subjective factors that might influence the values obtained in DM.

In this study, we evaluated the accuracy of Gemini structured-light scanning system in comparison with direct measurement and used the standard mentioned by Soghyia et al., i.e. that the measurement differences of more than 0.5 mm between any of the techniques were considered clinically significant25. Significant differences were observed in 3 of 19(15.8%) measurements. The mean AE and APE were 0.30 ± 0.11 mm and 4.88 ± 2.22% respectively in SL scanning. In 2003, Enciso and co-workers showed that a single-camera system had errors in the range of 0.48–1.55 mm26. Yu et al. showed the differences in the dimensions of interphalangeal finger joint circumference obtained by 3D image combined from ten captures and direct measurement yielded root mean square error (RMSE) of 1.97–5.07 mm1. The maximum AE obtained with Gemini structured-light scanning was 0.94 mm. This finding is comparable with the results of other studies on the accuracy of structured-light scanning applications27, and other imaging applications used for obtaining measurements of the hand1,28. Although the Gemini structured-light scanning system has not been employed for the clinical measurement of hand, the results of our study are encouraging.

The AE for the PIP circumference of the 4th finger and 1st finger length on the palmar side was greater than the value set for clinically significant difference between direct measurement and CT scanning, i.e., 0.5 mm. The mean AE of ten subjects was very low for the CT scanning(0.29 ± 0.10 mm) and for SL scanning (0.30 ± 0.11 mm). However, the mean APE of the five measurements (26.32%) obtained by CT scan and seven measurements (26.84%) obtained by the structured-light scan were greater than 5%. Fourie et al. evaluated the accuracy of standard anthropometric linear measurements of the head with three different 3D scanning systems and reported a mean AE of 0.76–0.89 mm and mean APE of 1.21–1.64% for the 3D scanning systems9. Although it appeared that the AE was lower in our study, the APE was much lower in the study by Fourie et al., considering that the dimensions of the head are greater than those of the hand. The 3D scanning system used for hand measurement is technically more precise than that required for head24.

The SL scanning offers several advantages, including rapid capture, photorealistic appearance, non-invasive nature, varying resolution quality, and high sensitivity of the technique. These advantages allow more rapid and non-invasive evaluation of the soft tissues of hand than conventional anthropometric techniques. However, one of the drawbacks of this method is the image defects caused by light irradiation and occlusion. This problem may be overcome by adding scanning heads29, or modifying the precise encoding and decoding algorithm of structured light30. However, due to inherent defects in the technology and distortion of light, none of the 3D imaging systems currently in use are accurate across the full field of view. Continuous improvement in technology and software has made it possible for researchers and clinicians to achieve realistic 3D imaging, although it may never be fully attainable. Furthermore, Lee et al. reported that subjective satisfaction achieved with the 3D semiautomatic measurement protocol (3D-SAMP) with respect to ease of hand measurement was significantly greater than that achieved for the direct measurement protocol28. This made the 3D-SAMP more preferable to direct measurement because of the shorter interaction time with the subject.

This study has a few limitations. First, the investigations were made in a small sample population. Second, we used plaster hand models instead of actual hands to minimize experimental errors. Yu et al.1 also explained that actual hands and plaster hands could differ in dimensions because of posture changes, muscular movements, and tissue compression in actual hands. Our study revealed that 3D images acquired by Gemini Structured light-scanning system and CT scanning were highly accurate. CT scanning is commonly used to harvest the static data to diagnose, reconstruct soft tissue. A real-time measurement of hand model and hand tracking is the tendency, and we intend to obtain hand measurements from live models to evaluate the hand contour and function, assess the hand/finger defect, and guide hand operation by Gemini Structured light-scanning system.

Conclusions

The Gemini structured-light scanning system not only successfully established a complete and smooth 3D hand photograph, but also shortened the scanning and reconstruction time. We compared two 3D measurement techniques with DM, and the results indicate that in terms of accuracy, the 3D imaging measurements obtained using the two indirect hand measurement techniques did not differ from those obtained by DM.