Abstract
Numerous biomechanical studies comparing titanium (Ti) and cobalt-chrome-alloy (CCM) rods are described in the literature. However, there is a dearth of literature comparing the two rod materials in adolescent idiopathic scoliosis (AIS). Therefore, the purpose of this study is to compare the correction rates of Ti and CCM rods in the treatment of AIS with double major curves. We enrolled 45 patients with AIS who underwent surgery between 2009 and 2012. We divided patients into two groups, Group A (n = 29) treated with six-millimeter Ti rods and Group B (n = 16) treated with six-millimeter CCM rods. The rod-derotation maneuver was used for correction. We measured pre- and postoperative indices of coronal alignment (Cobb’s angle, coronal balance, T1-tilt, clavicle angle) and sagittal alignment (sagittal vertical axis, thoracic kyphosis, lumbar lordosis). In our study, there were no significant differences between the two groups with respect to demographics or curve characteristics (P > 0.05). In Group A, thoracic and lumbar curvature correction rates were 71.2% and 66.8% respectively, and in Group B they were 71.2% and 73.3%, respectively (P = 0.664 and 0.09). There were no significant differences between the two groups in coronal or sagittal factors (P > 0.05) except for greater postoperative lumbar lordosis in the CCM group (P < 0.001). In conclusion, Ti and CCM rods showed similar correction rates in the sagittal and coronal planes for the surgical correction of AIS with double major curves. Biomechanical studies of Ti and CCM rods in vitro is different in biological condition.
Similar content being viewed by others
Introduction
The goals of surgery for adolescent idiopathic scoliosis (AIS) are to correct the deformity and achieve sagittal and coronal balance with fusion of the least number of vertebral segments1. These goals are affected by factors including patient characteristics (e.g., curve pattern and flexibility) and surgical variables such as the type of device used (hooks or pedicle screws), rod type, and technique used to achieve curve correction (translation, rod-derotation, or direct vertebral rotation)2,3,4,5,6. Of these two considerations, surgical variables can be controlled and thus are modifiable. The medical community has been striving continuously to improve available instrumentation to yield better, safer, and stronger implants.
The pedicle screw-rod system has proven its efficiency due to a strong pull-out force and three-column fixation, and it has become the state-of-art technique for posterior spinal fixation7,8. Pedicle screw-rod constructs using titanium-alloy (Ti) have become popular in recent years. Ti rods have distinct advantages over their stainless steel predecessors in aspects of more flexibility, a higher inert nature, biocompatibility, corrosion resistance, and an apparent ability to integrate with the surrounding bone with lesser radiologic artifacts9. Despite their widespread use, there are concerns about Ti rod implant failure. Ti rods are subjected to weakening of forces during intra-operative rod contouring and corrective procedures in deformity correction surgeries.
As the pursuit of the ideal biomaterial continues, use of cobalt-chrome-alloy (CCM) rods has been on the rise. CCM rods have the merits of both Ti and stainless steel rods, but CCM has higher rigidity than Ti making it possible to use a lower-profile rod with similar strength. Other proposed advantages include relatively low resistance to infection and radiological artifact not significantly different from titanium9,10,11. However, the benefits of CCM over Ti have only been studied in vitro biomechanical studies so far, with rare clinical studies comparing their correction rates in vivo. With this background, our study was designed to compare and analyze the efficacy of CCM and Ti rod systems with regard to the correction rates of scoliotic curvature in AIS patients with double major curves requiring fusion surgery.
Material and methods
This research was designed as prospective, consecutive, non-randomized study and patients who underwent surgery for scoliosis between 2009 and 2012 were included. Inclusion criteria were: (1) 11 to 17 years old, (2) Risser grade 1 to 4, (3) scoliosis deformity with a double curve defined as Lenke type 6 which has thoracic and lumbar structure curve more than 30 degrees which also satisfying Suk’s guidelines for fusion as a double major curve (4) a minimum follow-up period of 5 years6,12. Patients undergoing revision spine surgery were excluded. This study was performed after obtaining approval of the institutional review board. All methods were carried out in accordance with relevant guidelines and regulations of our licensing committee. Informed consent for surgery and publication of identifying images were obtained and approved from both all patients and parents and/or legal guardians.
To avoid bias of surgical technique cases operated by a single surgeon were enrolled. Posterior spinal instrumentation and fusion were performed using mono-directional pedicle screws. The proximal level of fusion was neutral vertebrae and distal extent of fusion was determined according to Suk’s guidelines6,12. In Group A, 6-mm Ti rods (Titanium Alloy Ti 6AI-4V, Carpenter, PA, USA) was used and in Group B, 6-millimeter CCM rods (BioDur® Carpenter CCM® Alloy, Carpenter, PA, USA) was used. Correction was obtained with a simultaneous double rod-derotation maneuver. Thoracoplasty was performed in patients who had a significant rib hump height difference of more than 3 cm. Radiographs were taken at regularly with interval of 6 months, and final radiographs taken at last follow-up visit were compared with the preoperative radiographs.
Coronal Cobb’s angles, coronal balance (CB), first thoracic vertebral tilt (T1 angle), and clavicle angle (CA) were assessed to evaluate coronal balance13. Sagittal vertical axis (SVA), thoracic kyphosis (TK) and lumbar lordosis (LL) were assessed to evaluate sagittal balance14. Normal values based on the Korean sagittal profile as elaborated by Lee and colleagues were used as a ref. 15. For verifying inter and intra-observer reliability, all factors were measured twice by experienced orthopaedic surgeons and radiologists. After the first measurement, a second measurement was performed with a three-week of time-interval.
Clinical outcomes related with complication were recorded as form of intra-operative, immediate post-operative and last follow-up. As intra-operative complications, significant change of neuromonitoring signal and screw pull out during correction procedure were evaluated. As immediate complications, infection, pulmonary complication (hemothorax or pneumonia etc.), wound dehiscence, neurological deficit, abdominal discomfort were evaluated. And as late complications, delayed infection, reduction loss which was defined as change of angle more than 5 degrees, proximal and distal decompensation, halo sign or screw breakage on last follow-up radiography, kidney and/or liver complication related with metal debris were evaluated.
Statistical analysis was performed using SPSS Version 17.0 (IBM Corporation, Armonk, NY, USA). Pearson’s chi-square test or Fisher’s exact test was used to compare ordinal values between the two groups. The Mann–Whitney U test was used to compare numeric values between the two groups. If the P value is under 0.05, this value was considered to be statistically significant. However, because the enrolled patients for statistical analysis are relatively small in size, the power of statistically significant result was verified with G-power program Ver 3.19 (Universität Düsseldorf, Germany). Intra and inter-observer reliability were analyzed with intraclass correlation coefficient (ICC) test.
Results
Group A included 29 patients who underwent correction with Ti rods and Group B consisted of 16 patients who underwent correction with CCM rods for a total study population of 45 patients. Before comparison of two groups, intra and inter reliability of all measured factors were verified and all values were more than 0.9 (Table 1).
Both groups were matched for possible confounding factors that could affect the correction rate1,14. There was no significant difference in age, sex, height, weight, body mass index, Risser stage, preoperative Cobb’s angle, flexibility of major curve, Lenke curve type, fusion length, upper and lower fusion level or thoracoplasty between the two groups (Table 2). Pedicle screws were inserted in all fusion level, therefore the density of pedicle screws between Ti and CCM groups could be considered to be same. Postoperative complications occurred in 3 and 2 cases in Ti and CCM group, respectively. There was no significant difference in postoperative complications between the two groups (P = 0.991) and it was described detail in Table 2. Follow-up period was calculated 7.6 ± 1.8 and 6.0 ± 0.9 years in Group A and B. Late post-operative complications developed 3 cases in Ti group and 0 cases in CCM group (P = 0.267).
The change in Cobb’s angle of thoracic curvewas 41.5 ± 13.0 degrees in the group A and 43 ± 12.2 degrees in the Group B. The thoracic curve correction rate was determined using the formula “preoperative major curve (°) - postoperative major curve (°)/preoperative major curve (°) × 100%”. It was 71.2 ± 6.8% in the group A and 71.2 ± 10.0% in the Group B. Similarly, the lumbar curve correction rate was 73.3 ± 13.6% in the Group A and 66.8 ± 13.1% in the Group B. There was no statistical difference in the thoracic or lumbar curve correction rates between the Ti and CCM groups (P = 0.664, 0.09).
At last follow-up radiography, maintenance of corrected Cobb’s angle for thoracic and lumbar curve was evaluated between Group A and B and there was no statistical difference between immediate and last follow-up corrected Cobb’s angle for thoracic and lumbar curve in Group A and B (P = 0.711, 0.722 in Group A, P = 0.271, 0.662 in Group B).
There were also no significant differences in pre- and postoperative T1 and CA angles between the two groups (Table 3). The CB correction rate after surgery was 1.5 ± 13.5 mm in the Group A and 4.7 ± 12.8 mm in the Group B and the difference between groups was not statistically significant (P = 0.441) (Fig. 1).
Sagittal parameters between the groups were also similar, with no significant differences in SVA or TK (p = 0.485 for SVA, P = 0.086 for TK) (Table 4) (Fig. 2). However, the power for statistical results of TK is relatively weak, therefore, patients with CCM rods had greater postoperative lumbar lordosis (59.0 ± 3.6°) than patients with Ti rods (47.3 ± 9.1°) (P < 0.001). Postoperative TK in CCM rod group is larger than Ti rod group.
At last follow-up radiography, maintenance of corrected TK and LL were evaluated between Group A and B and there were no statistical difference between immediate and last follow-up corrected TK and LL in Group A and B (TK: P = 0.691 in Group A, P = 0.852 in Group B, LL: P = 0.351 in Group A, P = 0.652 in Group B).
Discussion
The rod-derotation maneuver after pedicle-screw insertion is a commonly used technique for curve correction in patients with AIS1,2,4. Mechanism of correction of rod-derotation is reducing the deformity by delivering the corrective force generated by derotation of rod anchored on screws to vertebrae. Rod-derotation maneuver is consisted of as follows; rod is pre-contoured fitting to the shape of scoliotic curve then pre-contoured rods are assembled to screws in concave and convex side and both rods are derotated 90 degree in counter clock wise direction using 3~4 vice grip.
For effective correction shape of pre-contoured rod should be maintained without deformation of flattening during derotation maneuver. Therefore, stiffer the rod is advantageous in maintain the pre-contoured rod shape. For these reasons cobalt-chrome-alloy which is approximately five times stiffer than titanium alloy is preferred by some surgeons. But not always cobalt-chrome-alloy bears higher correction rates due to pull out of screws from vertebra during rod derotation maneuver owing to too much stiffness of cobalt-chrome-alloy. On the other hand, even though titanium alloy is less stiff than CCM, Ti alloy has lower incidence of screw pull out due to its relative flexibility thus fulfilling its goal of correction without complications. Many studies compared effect of Cobalt-Chrome-alloy and Ti on correction rates in AIS surgery16,17,18,19,20,21,22,23,24. Etemadifar et al.20 reported that cobalt chromium-titanium (Co-Ti) rods were more effective than Ti only rods. However, their study used polyaxial pedicle screws which is less effective in transferring correction force to vertebral body due to freedom of head to body, and there was mixed use of relatively small-sized (5.5 mm) cobalt chromium and titanium on concave and convex side each. Considering these limitations, strong scientific validity is weak. Lamartina et al.5 reported better correction rates with a hybrid system using thicker rods (6 mm vs. 5.55 mm). However, this research did not exclusively studied pedicle screw constructs and also was not supported by an appropriate statistical analysis. Serhan et al.23 reported rod made of CCM and stainless steel materials are more corrective than that of titanium in-vitro study real scoliosis surgical conditions.
To overcome these limitations, this study was designed as prospective, consecutive, non-randomized method even though it was retrospectively analyzed. Double curve (Lenke type 6) only requiring fusion was enrolled and other variable mixed type of curves were excluded to avoid bias of patient selection. And for determination of distal fusion level (distal instrumented vertebra), Suk’s guidelines was used6,12. To the best of our knowledge, there are no papers that studied effect of material properties of the rod on correction rate with homogeneity of groups (e.g. single surgeon operation, single type of implant fixation, single pattern of curve).
With this background, study was carried out under the hypothesis that CCM (rigidity: 13.4 × 103 ksi) rod is stiffer than Ti (rigidity: 5.90 × 103 ksi) rod and thus can achieve better correction of scoliotic curvature. However, in contrary to initial assumption results showed no statistical difference in coronal or sagittal spinal parameters except for lumbar lordosis (Tables 3 and 4). Also this results contradicted the findings of a previously published study by Lamerain et al.19 that resulted as CCM rods produced higher correction rates in the frontal plane when compared to stainless steel rods of the same diameter (53.3% and 51.4%). But in that study patients operated during learning curve periods were not excluded thus showing higher correction rates among CCM rods group of patients operated after learning periods. Also, the study population was heterogeneous with inclusion of all types of curves in Lenke classification.
Our study is not the first study to show discrepancy between correction rates and rod stiffness of rods. The similar correction rates between stiffer and less stiff rods in our study can be explained by understanding that correctional force of rod-derotation is delivered to deformed vertebra via screws inserted in vertebra through pedicle. For the force of rod-derotation to be transferred to vertebra without loss of force, purchase of screws should be strong enough to resist to pulling out force of screws generated by derotation of rod. But in biologic conditions rod-derotation mechanical force exceeds purchasing power of screws resulting in pulling out or loosening of screws. Pulling out screws from vertebra fail to deliver the correction force effectively and results in loss of reduction sometimes. Ti rod although is less stiff than CCM rod, Ti rod is far stronger than the purchasing power of screws and thus Ti rod can achieve the effective correction.
In this study, sagittal parameter of lumbar spine (LL) definitively was affected by material property of rod (P < 0.001). And sagittal parameter of thoracic spine (thoracic kyphosis) also was affected by material property of rod in CCM-rod used group, even though it was not statistically proved (P = 0.086). Statistical results of TK had relatively weak value of power analysis (0.30) by small enrolled patients and /or absence of more definitive difference in measured thoracic kyphosis between two groups.
Abul-Kasim et al.21 also reported that larger rod diameter (stiffer rods) had a positive impact on the deformity correction in the sagittal plane with pre-operative hypokyphotic thoracic spine and mainly Lenke type 1 curve. In this study, sagittal profile in Lenke type 6 curve is also more affected by more rigid rod. These results could be explained as follows; firstly, various surgical techniques making rigid curve flexible mobile, multiple facetectomies to enhance efficient load transfer of mono-axial pedicle screw into vertebrae. Through these techniques, more powerful correction force of stiffer rod (CCM) can be transferred to vertebral body without pulling out of screws; Secondly, sagittal profile in lumbar and thoracic are compensatory working in interactively influencing with each other. The lumbar spine is relatively more flexible than thoracic spine. Hence when pre-contoured rods are maneuvered for deformity correction, the stiffer CCM rods would have lesser deformation in the thoracolumbar or lumbar region compared to Ti rods. This may be the cause of the greater postoperative lumbar lordosis observed with CCM rods and this large change of lumbar curvature also have effect on postoperative TK sequentially. In this study, postoperative TK in CCM-rod used group is larger than Ti-rod used group. However, the degree of change of postoperative TK is less than lumbar curvature. And the possible causes could be inferred to interactive action of thorax and lumbar curve each other and relative stiffness of thoracic spine by holding effect of rib cage.
In the mid-term follow-up for more than 5 years, it was confirmed that the coronal and sagittal balance was maintained regardless of the material properties of the rods. There were three delayed complications occurred in the Ti rod group and revision surgery was performed in infected case only. There was no statistical difference between the two groups (P = 0.267).
Our study was not without limitations. The study was not randomized and the number of enrolled patients was relatively small. Although this study was designed as a retrospective type, the data included in the study were collected and managed in a prospectively, consecutive, non-randomized method. Furthermore, to decrease the selection bias, titanium and cobalt-chrome rod were used to patient with time sequential pattern using all or none way. And to improve the quality of results in this study, we tried to apply to proper statistical analysis and verified power of statistical results, although achieving homogeneity of the enrolled patients in terms of a single curve was not possible. A large randomized study is required to provide more reliable information about the effect of rod stiffness on correction rates in patients with AIS. Various factors affect the outcome of such surgeries, such as the screw-bone interface, screw-rod interface, rod contouring and surgeon familiarity with the instruments. In spite of our efforts, these factors could not be controlled for in our study. However, we feel that our study is relevant given the fact that optimal rod stiffness to correct scoliotic curves has not yet been determined. In addition, the authors did not evaluate the clinical results before and after surgery. Clinical outcomes are not simply correlated to the radiological corrections but can be affected by changes in coronal and sagittal profile after operation. If the clinical outcomes were evaluated with health-related and functional outcomes questionnaire like SRS-22, the effect of rod material on the patients could be evaluated more carefully in terms of physical, mental and surgical satisfaction.
Conclusions
Ti and CCM rods showed similar correction rates in the sagittal and coronal planes for the surgical correction of AIS with double major curves. Biomechanical studies of Ti and CCM rods in vitro is different in biological condition.
References
Weinstein, S. L., Dolan, L. A., Cheng, J. C., Danielsson, A. & Morcuende, J. A. Adolescent idiopathic scoliosis. Lancet. 371, 1527–1537 (2008).
Anekstein, Y. et al. Reversing the concept: correction of adolescent idiopathic scoliosis using the convex rod de-rotation maneuver. Eur Spine J 21, 1942–1949 (2012).
Blondel, B. et al. Influence of screw type on initial coronal and sagittal radiological correction with hybrid constructs in adolescent idiopathic scoliosis. Correction priorities. Orthop Traumatol Surg Res 98, 873–878 (2012).
Lafon, Y., Lafage, V., Dubousset, J. & Skalli, W. Intraoperative three-dimensional correction during rod rotation technique. Spine (Phila Pa 1976) 34, 512–519 (2009).
Lamartina, C., Petruzzi, M., Macchia, M., Stradiotti, P. & Zerbi, A. Role of rod diameter in comparison between only screws versus hooks and screws in posterior instrumentation of thoracic curve in idiopathic scoliosis. Eur Spine J 20, S85–S89 (2011).
Suk, S. I. Pedicle screw instrumentation for adolescent idiopathic scoliosis: the insertion technique, the fusion levels and direct vertebral rotation. Clin Orthop Surg 3, 89–100 (2011).
Crawford, A. H., Lykissas, M. G., Gao, X., Eismann, E. & Anadio, J. All-pedicle screw versus hybrid instrumentation in adolescent idiopathic scoliosis surgery: a comparative radiographical study with a minimum 2-Year follow-up. Spine (Phila Pa 1976) 38, 1199–1208 (2013).
Jaquith, B. P. et al. Screws versus hooks: implant cost and deformity correction in adolescent idiopathic scoliosis. J Child Orthop 6, 137–143 (2012).
Binyamin, G., Shafi, B. M. & Mery, C. M. Biomaterials: a primer for surgeons. Semin Pediatr Surg 15, 276–283 (2006).
Shida, T. et al. Adherence ability of Staphylococcus epidermidis on prosthetic biomaterials: an in vitro study. Int J Nanomedicine 8, 3955–3961 (2013).
Yasuyuki, M. et al. Antibacterial properties of nine pure metals: a laboratory study using Staphylococcus aureus and Escherichia coli. Biofouling. 26, 851–858 (2010).
Suk, S. I., Kim, J. H., Kim, S. S. & Lim, D. J. Pedicle screw instrumentation in adolescent idiopathic scoliosis (AIS). Eur Spine J 21, 13–22 (2012).
Malfair, D. et al. Radiographic evaluation of scoliosis: review. AJR Am J Roentgenol 194, S8–S22 (2010).
Prince, D. E. et al. The effect of rod diameter on correction of adolescent idiopathic scoliosis at two years follow-up. J Pediatr Orthop 34, 22–28 (2014).
Lee, C. S., Chung, S. S., Kang, K. C., Park, S. J. & Shin, S. K. Normal patterns of sagittal alignment of the spine in young adults radiological analysis in a Korean population. Spine (Phila Pa 1976). 36, E1648–E1654 (2011).
Noshchenko, A. et al. Thermomechanical effects of spine surgery rods composed of different metals and alloys. Spine (Phila Pa 1976). 36, 870–878 (2011).
Hosseinpour-Feizi, H., Soleimanpour, J., Sales, J. G. & Arzroumchilar, A. Lenke and King classification systems for adolescent idiopathic scoliosis: interobserver agreement and postoperative results. Int J Gen Med 4, 821–825 (2011).
Richards, B. S., Sucato, D. J., Konigsberg, D. E., & Ouellet, J. A. Comparison of reliability between the Lenke and King classification systems for adolescent idiopathic scoliosis using radiographs that were not premeasured. Spine (Phila Pa 1976), 28, 1148–1156, discussion 1156–1157 (2003).
Lamerain, M. et al. CoCr rods provide better frontal correction of adolescent idiopathic scoliosis treated by all-pedicle screw fixation. Eur Spine J 23, 1190–1196 (2014).
Etemadifar, M. R., Andalib, A., Rahimian, A. & Nodushan, S. M. H. T. Cobalt chromium-Titanium rods versus Titanium-Titanium rods for treatment of adolescent idiopathic scoliosis; which type of rod has better postoperative outcomes? Rev Assoc Med Bras (1992) 64, 1085–1090 (2018).
Abul-Kasim, K., Karlsson, M. K. & Ohlin, A. Increased rod stiffness improves the degree of deformity correction by segmental pedicle screw fixation in adolescent idiopathic scoliosis. Scoliosis. 6, 13 (2011).
Johnston, C. E. II., Ashman, R. B., Baird, A. M. & Allard, R. N. Effect of spinal construct stiffness on early fusion mass incorporation. Experimental study. Spine (Phila Pa 1976) 15, 908–912 (1990).
Serhan, H., Mhatre, D., Newton, P., Giorgio, P. & Sturm, P. Would CoCr rods provide better correctional forces than stainless steel or titanium for rigid scoliosis curves? J Spinal Disord Tech 26, E70–E74 (2013).
Sudo, H. et al. Correlation analysis between change in thoracic kyphosis and multilevel facetectomy and screw density in main thoracic adolescent idiopathic scoliosis surgery. Spine J. 16, 1049–1054 (2016).
Acknowledgements
No funds were received in support of this work. No benefits in any form have been or will be received from a commercial party related directly or indirectly to the subject of this manuscript.
Author information
Authors and Affiliations
Contributions
Jae Hyuk Yang wrote the manuscript and prepared the figures. Seung Woo Suh contributed the collection of data, statistical analysis, and supervision. Dong-Gune Chang contributed the supervision, critical revision of the manuscript for important intellectual content.
Corresponding authors
Ethics declarations
Competing interests
We wish to confirm that there are no known conflicts of interest associated with this publication and there has been no significant financial support for this work that could have influenced its outcome. We confirm that the manuscript has been read and approved by all named authors and that there are no other persons who satisfied the criteria for authorship but are not listed. We further confirm that the order of authors listed in the manuscript has been approved by all of us. We confirm that we have given due consideration to the protection of intellectual property associated with this work and that there are no impediments to publication, including the timing of publication, with respect to intellectual property. In so doing we confirm that we have followed the regulations of our institutions concerning intellectual property. We understand that the Corresponding Author is the sole contact for the Editorial process (including Editorial Manager and direct communications with the office). He is responsible for communicating with the other authors about progress, submissions of revisions and final approval of proofs. We confirm that we have provided a current, correct email address which is accessible by the Corresponding Author and which has been configured to accept email from dgchangmd@gmail.com.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Yang, J.H., Suh, S.W. & Chang, DG. Comparison of surgical correction rates between titanium and cobalt-chrome-alloy as rod materials in adolescent idiopathic scoliosis. Sci Rep 10, 10053 (2020). https://doi.org/10.1038/s41598-020-66975-x