Introduction

Hemolysis (the rupture of red blood cells) can occur in medical procedures where blood is removed from the body1,2. For example, passing blood through a faulty dialysis machine can potentially risk the life of the patient3,4, and even drawing blood too quickly through a needle can lead to defective laboratory samples5. Preventing hemolysis is therefore an important design constraint for medical pumps, prosthetic organs, hypodermic needles, and blood extraction procedures6,7,8,9.+

The cytoplasm of red blood cells is rich in hemoglobin, an iron-containing biomolecule that can bind oxygen and is responsible for the red color of the cells. As red blood cells rupture, they release their hemoglobin into the plasma— which is mostly water— changing the plasma from being relatively colorless to having a red tint. The degree of hemolysis can be measured by separating the plasma from the red blood cells and analyzing the amount of cell-free hemoglobin10 using a spectrophotometer, which measures how much light of a given wavelength is absorbed by the sample. Spectrophotometry is considered the most accurate method for measuring hemolysis, but there are a number of other possible methods that do not require extracting red blood cells or using chemical analysis. For example, Tarasev11 suggested a blood hemolysis analyzer that can measure the amount of cell-free hemoglobin by using two or more wavelengths of light and comparing the different levels of light absorption. Karlsson12 and Lee13 both describe devices to identify hemolysis by the naked eye through the change in plasma color. None of these methods, however, allow clinicians to monitor hemolysis during a procedure and provide immediate information on the level of blood damage.

Recently, Zhou et al.14 proposed a complex method that could actively measure hemolysis by combining nanofilters, which actively filter the plasma from the red blood cells, and optofluidic sensors for evanescent absorption detection. The process is similar to spectrophotometry, in that it is analyzing light absorption and relating it to hemolysis level, with the added benefit of eliminating the need for specialty sample preparation.

We propose a much simpler technique that can detect hemolysis continuously in real-time by measuring the electrical resistance of the blood. Blood is naturally conductive, but the outer lipid bilayer of the red blood cell is insulating and so healthy blood cells do not contribute much to the overall conductivity18,19. As red blood cells rupture, however, they release their hemoglobin and raise the conductivity of the entire fluid (as illustrated in Fig. 1). Hence, the change in blood conductivity can be related directly to the level of hemolysis. Our proposed technique does not require external sources of light, the separation of the blood cells from the plasma, or specific chemical detection, yet it can determine the progression of hemolysis in real time.

Figure 1
figure 1

Simplified illustration of blood flow to show how hemolysis leads to a higher blood conductivity. The fluid becomes more conductive left-to-right as red blood cells rupture and release their hemoglobin.

Results and discussion

The method uses a test cell that consists of a small converging/diverging channel equipped with top and bottom electrodes. The cell was tested by inserting it in a laminar flow loop driven by a peristaltic pump, as shown in Fig. 2. The conductivity was measured using a high-quality inductance-capacitance-resistance (LCR) meter for continuous sampling, and a conventional conductivity probe for periodic sampling.

Figure 2
figure 2

Experiment schematic. Flow path: (1) magnetic stirrer with open reservoir; (2) peristaltic pump; (3) 3D printed channel with a conductive floor and ceiling in the test section.

The system was first tested using KCl saline solutions to achieve mass concentrations of 0.5%, 1%, 2%, and 4% in deionized water. This range of salinity was chosen to give changes in conductivity similar to that expected under moderate levels of hemolysis. Figure 3 shows the variation in time of the saline resistivity, \(\rho =Rlw\)/\(h\), where \(R\) is the resistance in ohms, \(l\) and \(w\) are the length and width of the plate, respectively, and \(h\) is the channel height (note that conductivity is defined as 1/\(\rho \)). As salt is added to the reservoir, there is an initial step change in the resistivity followed by more gradual asymptotic behavior as the salt dissolves. The conductivity probe measurements were made just before each stage of salt addition, and we see that they agree well with the LCR meter readings. The measurements also match well with the known conductivity of KCl solutions20.

Figure 3
figure 3

Saline resistivity, as measured continuously with the LCR meter and periodically with the conductivity probe, compared with the corresponding theoretical values.

The system was then tested using porcine blood. We are only concerned with the change in blood conductivity with damaged blood cells, thus porcine blood is a good model for human blood because it has similar blood cell size - porcine: 5.8 \(\mu \)m, human: 7.3 \(\mu \)m15; hematocrit - porcine: 32–50%, human: 36–51%16,17; and hemaglobin concentration - porcine: 10–16 g/dL, human: 12–17 g/dL16,17. Measuring the absolute resistivity of blood can be challenging, in that it can act as a dielectric18, and also the flow shear can align the orientation of the red blood cells and make the electrical properties of blood anisotropic19. Since we are only interested in measuring relative changes in blood resistivity for a given flow condition, these effects can be ignored. To control the level of damage, the blood was split into two separate 500 mL samples. One sample was left pristine while the other sample was mechanically damaged using an immersion blender (77% hemolyzed). The undamaged sample was used as the starting fluid in the flow loop, and then every 10 minutes 50 mL of the damaged blood sample was added into the stirring reservoir to increase the hemolysis levels in gradual steps. Samples of 1.5 mL were extracted twice at each blood damage level for direct hemolysis measurements using the spectrophotometer as described in the methods section.

Figure 4
figure 4

(a) Blood resistivity (left axis) and hemolysis percentage (right axis) as a function of time. (b) Change in blood conductivity as a function of hemolysis percentage; blue circles are the main trial, and all other symbols represent repeated experiments with different blood samples. Inset figure in (b) shows more moderate hemolysis levels.

Figure 4a shows the blood resistivity measured by the LCR meter (left axis) and hemolysis percentage measured using the spectrophotometer (right axis) over the 90 minute test period. Each addition of damaged blood causes a step change in the blood resistance followed by a slower asymptotic behavior as the mixture homogenizes. Figure 4b shows a direct correlation between the change in blood conductivity level (1/\(\rho \)) and the total hemolysis percentage. The trend is almost linear—where the correlation coefficients and \(p\)-values are shown in Table 1—and additional trials with different porcine blood samples show this trend is repeatable and is reliable down to 2–3% hemolysis, at least. However, the change of the slope between repeated trials emphasizes that the blood conductivity is sensitive to other factors that may change between samples.

Table 1 Correlation coefficients and \(p\)-values for the cases presented in Fig. 4b.

We see that a simple conductivity cell can be used to immediately detect qualitative changes in hemolysis. The measurement is continuous, in real time, and easy to implement in clinical practice. In its current state, the technology is limited to measuring relative changes in the blood for a given patient and cannot be used to measure quantitative blood damage because the electrical properties of blood samples may vary from one patient to another. (Despite this limitation, there are clinical situations where the technique might find use, e.g., it could act to detect dialysis machine malfunction that damages otherwise non-hemolyzed blood by comparing blood conductivity on the machine input and output.) Many factors can change blood conductivity, including patient gender, age, and health condition (e.g. comorbidities like hyperlipidemia or anemia), to name a few. Future studies are needed to refine the concept and clarify the impact of possible patient variables (hemoglobin/hematocrit concentration, blood conditions that may impact conductivity) and flow variables (laminar/turbulent, flow rate, temperature).

Material and methods

Measurements were made in the recirculating flow facility shown in Fig. 2. Fluid was driven by a peristaltic pump (Cobe 043600-000) at 41.67 cm3/s through a test channel and an open reservoir equipped with a magnetic stirrer (Sargent-Welch). The channel was custom designed and 3-D printed from a waterproof photopolymer (Watershed 11122XC) at the W.M. Keck Center for 3D Innovation at the University of Texas El Paso. The channel test section had length \(l=66\) mm, height \(h=7.6\) mm, and width \(w=25.4\) mm and was equipped with an electrically conductive brass floor and ceiling for flow resistance measurements. The test cell was designed to avoid any flow separation and keep the flow laminar so that flow stresses through the channel do not further damage blood cells. The open reservoir station allowed for direct conductivity measurements of the recirculating fluid, sample extraction, and mixing in other materials.

During the tests, the resistance of the fluid was continuously measured with an inductance-capacitance-resistance meter, or LCR meter (Keysight Technologies E4980AL) which can continuously read resistance with 0.1% accuracy. Measurements were recorded via LabView at 2 Hz, and tests were conducted for up to 90 minutes. Direct conductivity measurements were also made with a more conventional conductivity probe (Hach HQ14D), accurate to 0.5%, to validate the LCR meter readings.

Two working fluids were used for this study. First, a potassium chloride (KCl) based saline solution was used to validate our resistance measurements, which was made by mixing a known mass of KCl (EMD PX1405-1) measured via a precision scale (VWR 1002E) into room-temperature deionized water. Second, we used room-temperature porcine blood for testing hemolysis. The blood was purchased fresh through Lampire Biological Laboratories where it was obtained from healthy adult animals an added anticoagulant heparin. Each test trial was a unique Hamp/York female donor. The hematocrit ranges of the donors ranged from 39–50%, though some samples did not have the hematocrit measured by the source lab (specifically, cases numbers 3–5). Hemolysis levels were determined by measuring the relative levels of free hemoglobin in the blood plasma using a spectrophotometer (Beckman Coulter DU730).

The procedure to obtain hemolysis percentage measurements of blood samples using a spectrophotometer is as follows.

  • Prepare a Drabkin’s Solution by combining Drabkin’s Reagent (Sigma Aldrich D5941) to 1 L of deionized water.

  • To lyse the red blood cells, prepare a separate solution containing 100 mL of the Drabkin’s solution from Step 1 and 0.05 mL of 30% Brij 35 Solution (Sigma Aldrich B4184).

  • Set aside two 1.5 mL samples of undamaged blood into centrifuge tubes (samples measured via Eppendorf 5 mL adjustable volume pipette). Then, acquire and similarly store blood samples during the experiment.

  • Centrifuge the blood samples for 3 minutes at 6000 RPM to separate the red blood cells from the plasma. Leave one of the two 1.5 mL undamaged blood samples uncentrifuged.

  • In spectrophotometer cuvettes, mix 2 mL of the Drabkin’s Solution with 8 \(\mu \)L of plasma from the centrifuged blood samples (extracted with Eppendorf 10 \(\mu \)L adjustable volume pipette).

  • Mix 8 \(\mu \)L of the undamaged/uncentrifuged blood sample from Step 4 with 2 mL of the Drabkin’s + Brij 35 Solution from Step 2 into a spectrophotometer cuvette.

  • Ensure that all cuvette samples are well mixed and allow to rest for 15 minutes.

  • Zero the spectrophotometer using only the Drabkin’s Solution (2 mL) in a cuvette, this serves as the "blank” sample.

  • Using the spectrophotometer, measure and record the baseline reference case of the cuvette with the original undamaged blood sample from Step 5. Denote as \({A}_{0}\), where \(A\) is the absorbance at a wavelength of \(540\) nm.

  • Similarly, measure and record the fully damaged reference case from the lysed blood sample made in Step 6. Denote as \({A}_{\infty }\).

  • Lastly, measure and record the spectrophotometer readings from all of the experimental samples from Step 5.

  • The relative hemolysis of a given sample is given by \((A-{A}_{0})\)/\(({A}_{\infty }-{A}_{0})\).