Abstract
Iodine-131 (131I) is an essential and widely used radioisotope in thyroid diseases and animal experiments. Planar imaging has been considered the most popular method for 131I thyroid uptake radioactive activity quantification. The ROI defining section is essential and can affect the accuracy of quantitative results. However, a consistent method has not been proposed. In this study, a UC-ROI defining method based on ULWL setting and colour display grade was applied. Three steps were performed: image acquisition of five standard activity models and obtaining the exact value that the counts per radioactive activity contributes to the ROI; image acquisition of 20 rat thyroids and obtaining the counts of the ROI (thyroid); and calculating the rat thyroid radioactive activity and comparing these values with the true values. The accuracy of quantification activity of 131I in rat thyroid reached 2.62% ± 0.41%. The mean quantification within 5% could be achieved in 19 of 20 rat thyroids. No significant difference existed between calculated thyroid 131I activity and true values with a paired matched-test (t = −0.384, P = 0.706 > 0.05). The results indicated that with the UC-ROI defining method, more accurate thyroid uptake 131I radioactive activity quantification by SPECT planar imaging can be achieved in vivo rat study.
Similar content being viewed by others
Introduction
With the increasing use of radiopharmaceuticals both in clinics and animal experiments, precise estimation of the absorbed dose in patients is crucially required to evaluate the prognosis and side effects1. Planar images of a gamma camera were used for activity estimated several decades previously2,3; however, the deviation between estimated activity and true activity could be up to 10% or more for certain organs4,5,6, which was mainly caused by region of interest (ROI) definition, background, scatter and attenuation1,4,7. In recent decades, by the development of fused imaging technology of single photon emission computed tomography and computed tomography (SPECT/CT), attenuation correction, scatter correction and partial volume correction, absolute organ activity quantification has reached a new level. Using the method of quantitative SPECT/CT, the accuracy has increased and in a manner similar to positron emission computed tomography (PET), even the standard uptake value (SUV) used for diagnosis can be obtained8,9. However, because of the complexity to be achieved, quantitative SPECT/CT is rarely applied in clinical practice. For clinical practice and animal researches, a simple, accurate, easy-to-reach quantitative method of planar imaging is desired.
Iodine-131 (131I) is an essential and widely used radioisotope in thyroid diseases, and planar imaging has been considered the most popular method for 131I thyroid uptake radioactive activity quantification5,10,11,12. The sensitivity known as count rate per unit activity or calibration factor must be known in order to convert the acquired count rate to activity5. To obtain sensitivity data, pamphlet 16 of the Medical Internal Radiation Dosimetry committee (MIRD) recommend an independent measurement of a standard, with known activity, within a standardized geometry in air relative to the scintillation camera, using designed gamma camera acquisition parameter settings. The 131I radioactive activity of thyroid uptake is calculated using the equation ((thyroid counts − background counts)/(sensitivity × time))5,12,13. The thyroid counts and background counts are obtained by drawing an ROI of the target tissue and an ROI of the background region. For quantification of planar imaging, defining an ROI is an essential procedure that can affect the accuracy of the quantitative results. However, studies of thyroid radioactive activity quantification did not provide a consistent method for defining the ROI. Most researchers defined the ROI according to visual characteristics and drew a region containing the thyroid10,12, but the visual character is a fuzzy concept. Additionally, several researchers adopted other methods for defining the ROI of the target tissue, such as choosing an 85% isocount contour of the thyroid gland13 or considering system spatial resolution (FWHM)5. In a word, the criterion for defining the ROI of the target tissue is unified, and studies are rarely designed to use a normative, accurate and simple ROI defining method.
Recently, with the development of image data processing workstations, a more accurate ROI and background regions can be defined, which has made the exact quantification of planar imaging possible. In this study, we adopted a new method for defining the ROI border based on the upper limit window level (ULWL) and a ring-like shape of the background to obtain corrected ROI counts and achieve quantification of thyroid uptake 131I activity. An alternative conventional ROI defining method by visual characteristics was applied as a reference. Models with measured 131I activity and rats injected with 131I were used to verify accuracy. Three steps were performed to derive absolute values for radioactive activity in Bq (1 µCi = 3.7 × 104 Bq) from planar images and verify accuracy5,13: image acquisition of standard activity models and obtaining the exact value that the counts per radioactive activity contribute to the ROI; image acquisition of rat thyroids and obtaining the counts of the ROI (thyroid); and calculating the rat thyroid radioactive activity and comparing these values with the true values. We tried to verify the accuracy of the quantification in rats experiments.
Materials and Methods
Image Acquisition of Standard Activity Models
A plastic disc (diameter, 60 mm; height, 15 mm) covered with a layer of tri-distilled water (depth, 5-mm approximately) was used as the model. A 5-mm depth water is used for simulating the thyroid underneath the skin. The total dose of 131I activity was measured with a radioactivity meter (RM905a, Beijing Hicheer Sci-Tec Co., Ltd.) before being introduced into the model with a syringe. After repeated washing with the tri-distilled water in the model, the syringe needed another measurement for residual radioactive activity. The difference between the two measurements was regarded as the true value of radioactive activity remaining in the model.
A high energy general purpose (HGCP) collimator from a SPECT/CT was used to acquire planar images (GE Discovery, NM/CT 670). One collimator was proposed since only the anterior image was required. The scan was undertaken for 200 seconds with the energy window at 10%, centred on 364 KeV. The matrix was settled as 256 × 256, and no zoom factor was used. The standard models were measured at a distance of 10-cm from the HGCP collimator. A 10-cm distance was chosen because it was widely used in clinical practice and suitable for all patients, especially women with large breasts12.
We designed five kinds of radioactive activity models, 3.7 × 106 Bq (100 µCi) for Model 1 (M1), 7.4 × 106 Bq (200 µCi) for Model 2 (M2), 1.48 × 107 Bq (400 µCi) for Model 3 (M3), 2.96 × 107 Bq (800 µCi) for Model 4 (M4), and 5.92 × 107 Bq (1600 µCi) for Model 5 (M5). Images for each activity model were collected five times to avoid statistical fluctuation and measuring error.
Additionally, each image acquisition lasted for 200 seconds, which was much shorter compared with the radioactive half-time of 131I (approximately 8 days); therefore, we felt that the activity remained unchanged during acquisition.
Image Acquisition of Rat Thyroids
Twenty rats (weight, 290–310 g; Sprague Dawley; male) were studied for estimating thyroid 131I activity and evaluating accuracy. For each rat, a random amount of 131I activity, ranging from 1.11 × 107 Bq to 1.11 × 108 Bq (300 µCi to 3 mCi), was introduced through intraperitoneal injection. The images were collected approximately 6–8 hours after injection.
The imaging system and parameters of rat thyroids should be maintained with standard radioactive activity models; otherwise, the data of rat thyroid images and standard model images would not be comparable.
After imaging, the rats were sacrificed by isoflurane overdose immediately. The thyroids were dissected and measured in the radioactivity meter (RM905a, Beijing Hicheer Sci-Tec Co., Ltd.) to obtain the true activity values as the gold standard.
Additionally, for each rat, it took nearly 20 minutes from the beginning of imaging to the accomplishment of extracted thyroid activity measurement. The time was much shorter compared with the 131I half-time. For this reason, we regarded the value of rat thyroid uptake activity as a constant.
Method of ROI Selection Control
Traditionally, the ROI is drawn according to the visual characteristics, and many studies have been based on this method10,12. However, what has been ignored is that the visual characteristics are affected by the upper limit of the window level (ULWL) setting, which can affect the visual characteristics.
The relation between visual characteristics and ULWL is illustrated in Fig. 1 (copied from GE Healthcare, Direction 5336813-1ZH-CN Revision 2, Processing and Review System, User Guide). The horizontal axis represents the counts per pixel and the vertical axis represents the grade of colour display. The value of counts per pixel varies from 0 to the maximum according to the planar image. The grade of colour display improves from 0 to 255 (from the darkest to the brightest) in correspondence with the value of counts per pixel changing from the lower limit window level (LLWL) to the ULWL. The pixels with the same counts are shown in the same grade of colour display, similar to a contour line (Fig. 2 shows several colour display patterns and, in this research, the GE colour pattern was used). For the pixels that had higher counts than the stated ULWL, the colour display was the brightest.
To achieve an accurate ROI region, a proper ULWL should be adopted. For each planar image, both the standard models and rat thyroids, the value of ULWL was set according to the following equations:
t represents total counts of the whole planar imagep represents the amount of pixelsμ is a constant (equals to 50 in this study).
The sensitivity (count rate per unit activity) for a particular gamma camera (SPECT collimator) is stable, which means radioactive sources with larger radioactive activity will lead to a larger total counts in the planar image over the same period of time; in other words, t is affected by the amount of radioactive activity. The p is settled as a constant for an exact planar imaging depending on the matrix area. In this research, a matrix of 256 × 256 was applied for all the planar image acquisitions, so the p was equal to 65536.
The coefficient μ is a constant that can be set from 0 to 100 or more. The planar images with different μ values of the same radioactive source are shown in Fig. 3. The default system value for ULWL is equal to the maximum count per pixel. We can determine that it is not suitable if a small μ is applied because the border of the ROI region is not clear. In this research, we chose μ as 50 because the planar images had a better visual characteristic at this condition (additionally, other values such as 40 and 60 will also be fine). Once the μ is decided, each planar image should take the one and only coefficient for setting the ULWL and obtaining the ROI.
After the ULWL is set, a counter line of colour grade should be chosen for drawing the ROI border. The specific colour grade used for defining the ROI border should be maintained in all of the planar image processing. In our study, a colour display pattern named GE colour was used and a specific colour grade was adopted for drawing an ROI border in the manner of ROI-1 in Fig. 4b.
Since this method is based on ULWL setting and colour display grade, we will use the UC-ROI drawing method for short in the following. After drawing ROI, the total counts of the ROI and average counts per pixel of the ROI can be obtained from the imaging workstation.
Method of Background ROI Selection
Traditionally, the background ROI is drawn near the thyroid and is always defined over an area caudal and lateral in relation to that of the target tissue ROI5,10. However, as is illustrated in Fig. 4a, there are differences among the average background per pixel for different ROIs. Accurate quantification will not be achieved if the differences in background ROI selection remain. Therefore, we carried out a new method for background selection, i.e., a ring-like background ROI drawing method. As is shown in Fig. 4b, a narrow annular background region is drawn next to the target tissue ROI. The differences in counts per pixel around the target tissue ROI are averaged and eliminated by this method. Moreover, this method is easy to follow, and standardizing the background ROI selection will benefit the repeatability and reliability of results.
After the background ROI is selected, the average counts per pixel can be read from the imaging workstation.
Acquisition of calculated radioactive activity for rat thyroid
Once the total counts of ROI, the average counts per pixel of the ROI for the target tissue and the average counts per pixel for the background region are obtained, the corrected total counts for the target tissue ROI (δ) can be calculated as follows:
S represents total counts of the ROI for the target tissueV1 represents the average counts per pixel of the ROI for the target tissueV2 represents the average counts per pixel for the background region.
For standard models, δ1, δ2, δ3, δ4 and δ5 represent the corrected total counts for M1, M2, M3, M4 and M5, respectively. A1, A2, A3, A4 and A5 represent the exact radioactive activity measured by the radioactive meter for each model. The standard counts per radioactive activity (τ) contributing to δ can be acquired by the following equation:
τn is the calculated standard counts per radioactive activity contributing to each model (M1 ~ M5), and τ′ is the average of τn.
The radioactive activities of rat thyroids are calculated as follows:
A′ represents the estimated radioactive activity of the rat thyroid, unit of 3.7 × 104 Bq (1 µCi). δ′ represents the corrected total counts for the ROI of the rat thyroid.
After every A′ collected, the accuracy can be verified by a comparison of A′ and true radioactive activity of the rat thyroid measured by a radioactivity meter.
Conventional quantitative method by visual characteristics
ROI defining method by visual characteristics was applied as a reference. The calculating procedure (equations 3–5) kept consistent with UC-ROI defining method. But the methods for ROI selection and background ROI selection were different. For the reference, ROI was selected directly on the planar imaging of default ULWL by visual characteristics, without any adjustment of the ULWL value. And background ROI was selected according to the method shown in Fig. 4a.
Data processing and statistical analysis
All the image data were processed at the Xeleris Functional Imaging Workstation (GE, Xeleris version 3.1). SPSS software (Version No. SPSS Statistics 22.0 version) was used for statistical processing. The numerical data were reported as means ± standard error (S.E.). A P value less than 0.05 was considered significantly different.
Ethical approval and consent to participate
Approved by West China Hospital of Sichuan University Biomedical Research Ethics Committee (Number: IACUC-2016029a). All experiments were performed in accordance with relevant guidelines and regulations. This article does not contain any studies with human participants performed by any of the authors.
Availability of data and material
Please contact author for data requests (including images).
Data availability
All the data including the Excel database and the DICOM data are available.
Results
Results of the Standard Activity Model
All the images of five standard activity models were processed by drawing a UC-ROI and a circle background ROI (Fig. 5a). Because of manual operation and the syringe remaining, the actual activities of the five models were 3.589 × 106 Bq (97 µCi), 6.808 × 106 Bq (184 µCi), 1.3875 × 107 Bq (375 µCi), 2.9896 × 107 Bq (808 µCi) and 5.621 × 107 Bq (1533 µCi), with small differences compared with the designed values of standard activities. However, this would not affect the consequences.
The values of τn for each model were 159.6, 161.8, 161.1, 159.8 and 165.7, and the average τ′ was 161.6. There was good consistency among τn for different models with various radioactive activities (standard deviation, 2.467). The exact counts per radioactive activity (per 3.7 × 104 Bq, 1 µCi) contributing to the target tissue was ascertained as 161.6, with a fixed 200 s, 10-cm planar SPECT imaging acquisition, a UC-ROI drawing method and a ring-like background region selection.
Activity Quantification of 131I in Rat Thyroid
Twenty planar images of rats were acquired. According to the UC-ROI drawing method, the value of ULWL for each image was settled and the ROI of the target tissue was defined based on the same visual characteristics. All the values of S and V1 for each rat thyroid 131I image were collected. Then, a circle background region was drawn next to the UC-ROI border to obtain the value of V2 (Fig. 5b).
The results of rat thyroid 131I activity quantification are summarized in Table 1, including ULWL, δ′, calculated thyroid 131I activity (A′), true thyroid 131I activity and the deviation. The average for deviations of the calculated activity from the gold standard values was 0.69% ± 0.71%. The average for absolute deviations of calculated activity from the gold standard values was 2.62% ± 0.41%. The mean quantification within 5% could be achieved in 19 of 20 rat thyroids. No significant difference existed between calculated thyroid 131I activity and gold standard values with a paired matched-test (t = −0.384, P = 0.706 > 0.05).
Results of conventional quantitative method
The accuracy of rat thyroid 131I activity quantification by conventional method is shown in Table 2. Results of deviation for UC-ROI defining method is also listed as a contrast. For the conventional quantitative method, the average for deviations of calculated activity from the gold standard values was 13.51% ± 1.21%. Significant difference of the accuracy can be observed between UC-ROI defining method and conventional method with a paired matched-test (t = 8.461, P < 0.001).
Results of accuracy with different µ values
The results above are obtained on the condition that µ equals to 50. We also tested different µ values (µ = 20, 30, 40, 60, 70, and 80), and the averages for deviations of calculated activity from the gold standard values are shown in Fig. 6. The averages of deviations for each µ value are 3.77% ± 0.57%, 2.46% ± 0.36%, 2.38% ± 0.38%, 3.50% ± 0.48%,4.14% ± 0.61% and 4.58% ± 0.60%, respectively.
Discussion
Thyroid 131I radioactive activity quantification is widely applied in clinical practice and research. Radioactive iodine uptake (RAIU) is used to assess the thyroid function13,14 and is essential for providing clinical information before 131I-therapy in some kinds of thyroid disorders, such as hyperthyroidism15 and thyroid carcinoma16. After administration of 131I, detecting the amount of thyroid uptake radioactive activity will benefit the evaluation of therapeutic process and treatment outcomes. Many animal experiments are based on Iodine-131 so that a promoted method of radioactivity quantification for vivo animal study will be beneficial.
Traditionally, the 131I radioactive activity of thyroid uptake is calculated using the equation ((thyroid counts − background counts)/(sensitivity × time))5,12,13, where sensitivity is the count rate per unit activity. In our study, we simplified this equation and adopted counts per unit activity (τ) rather than count rate per unit activity. The time factor was neglected because the planar image acquisition for both standard models and rat thyroids was exactly the same, including the parameters such as scanning time (200 s), distance (10 cm), matrix (256 × 256), energy window (364 KeV ± 10%) and others. Considering the scanning time was settled, we used counts per unit activity (τ) directly.
In addition to the factor of scan time, τ can also be influenced by the distance from the radioactive source to the collimator. The value of τ will decrease as the distance grows5. Therefore, in order to obtain a quantitative result, a proper distance must be fixed for all the standard models and rat thyroids. In this study, a 10-cm distance was chosen because it was widely used in clinical practice and suitable for all patients12. The criterion distance for rat study was not found, so we took clinical practice standard directly.
After the control of systematic scan parameters, the accuracy of quantification is still affected by the target tissue ROI definition and background region selection. The ROI is mainly defined by visual characteristics. However, for the same image, with different ULWL settings, the visual characteristics will change and the ROI selection is influenced. To solve this, our study used a new and easy method for ROI definitions, a UC-ROI defining method based on ULWL and a colour display grade. We proposed a method for setting the ULWL based on the average counts per pixel (α), multiplied by coefficient (µ), as in equations (1) and (2). In this study, we adopt 50 as the µ value. As is shown in Fig. 6, other values will also be reliable. We tested different µ values, and though subtle differences existing, the accuracy of radioactivity quantification remained stable and acceptable generally. The key point is that the value of μ should stay the same for all data processing and ROIs defining. It is more accurate when µ equals to 30, 40, or 50, compared with other conditions.
A promotion for background selection was also taken in this study. The counts for the ROI of the target tissue needed to be corrected by the background to obtain a more reliable result1,2,10. However, as explained above, there are random errors with the conventional background correction. Drawing a circular background region is slightly more complex compared with the conventional method, but it is more stable and easy-to-follow.
The five standard models were designed with five different levels of radioactive activity in order to obtain a more exact value of τ. The standard models were covered with 5-mm depth water to imitate the soft tissue overlap of the thyroid of the rat. Note that though the diameter of the plastic model is 60 mm, the ROI selection should be based on visual characteristics rather than the actual model diameter. The values of τn for each model were 159.6, 161.8, 161.1, 159.8 and 165.7, with very slight differences existing. This was indicated by the UC-ROI defining method and ring-like background selection; the τ is a stable value for the standard models despite the changes in radioactive activity.
After using the τ obtained from standard models to estimate the radioactive activity of the 20 rat thyroids, the results met expectations perfectly. The mean quantification within 5% could be achieved in 19 of 20 rat thyroids, and no significant difference existed between the calculated thyroid 131I activity and gold standard values. Our results showed a much better accuracy (0.69% of deviation, and 2.62% of absolute deviation) compared with conventional quantitative method. Both of the two methods are based on visual characteristics. But for the conventional method, ROI was selected on the planar imaging of default ULWL directly and background ROI was selected in an unstable way. Thus in our study, for the conventional method, the average for deviations of calculated activity from the gold standard values was 13.51%, much higher than UC-ROI defining method (2.62%).
In addition to the accuracy, the simplicity and convenience are also the good features of our method. After setting the imaging parameter and image acquisitions of the standard models and target tissues, to quantify the radioactive activity, all we need to do is take another step before selecting the ROI—setting the ULWL. Unlike other studies, the scatter correction7 and partial volume effect correction8 are not necessary with our method. The ROI borders are chosen based on visual characteristics according to the UC-ROI defining method, so the CT component is not needed for obtaining the boundary. The attenuation can be simulated by the depth of water in the standard models. We used a 5-mm depth water for simulating the rat thyroid attenuation.
The UC-ROI method can be widely used in animal experiments. With a τ obtained from standard models, it is convenient to calculate the radioactive activity of thyroid uptake for animals after 131I treatment. In addition to the thyroid, 131I uptake of other organs (the liver, the bladder, stomach and etc.) can also be quantified by the UC-ROI method in the vivo rat study if a ROI border can be defined clearly. For instance, if a sort of liver targeted radiolabeled agent was applied in rat experiments, the UC-ROI defining method can be adopt for planar imaging quantification theoretically.
Limitations
This promoted method is one that will work best in animal studies, especially for these with minimum background. However, while the background increasing, the border of the target tissue will be mixed with the background and become undefined. In other words, this UC-ROI defining method will be inaccuracy in the conditions with high background.
Conclusion
We report an accurate and simple method for thyroid uptake 131I radioactive activity quantification by gamma-camera planar imaging in vivo rat study with a high-energy collimator of a commercial SPECT and image processing workstation by a UC-ROI defining method. The long-term goal is to verify the accuracy in other organs (like the liver, the bladder, stomach and etc.) and investigate more applications of this ROI and background defining method in animal experiments.
References
Norrgren, K., Svegborn, S. L., Areberg, J. & Mattsson, S. Accuracy of the quantification of organ activity from planar gamma camera images. Cancer Biother Radiopharm 18, 125–131 (2003).
Buijs, W., Siegel, J. A., Boerman, O. C. & Corstens, F. Absolute organ activity estimated by five different methods of background correction. Journal of Nuclear Medicine 39, 2167–2172 (1998).
Thomas, S. R., Maxon, H. R. & Kereiakes, J. G. In vivo quantitation of lesion radioactivity using external counting methods. Med Phys 03, 253–255 (1976).
Fleming, J. S. Technique for the Absolute Measurement of Activity Using a Gamma-camera and Computer. Phys Med Biol 24, 176–180 (1979).
Barquero, R. et al. I-131 activity quantification of gamma camera planar images. Phys Med Biol 62, 909–926 (2017).
Goris, M. L., Knox, S. A., Nielsen, K. R. & Bouillant, O. Organ Modeling in the Quantitation of Planar Images for Distribution Studies. Ccancer-AM Cancer Soc 73S, 919–922 (1994).
Green, A. J., Dewhurst, S. E., Begent, R., Bagshawe, K. D. & Riggs, S. J. Accurate Quantification of I-131 Distribution by Gamma-camera Imaging. Eur J Nucl Med 16, 361–365 (1990).
Zeintl, J., Vija, A. H., Yahil, A., Hornegger, J. & Kuwert, T. Quantitative Accuracy of Clinical Tc-99 m SPECT/CT Using Ordered-Subset Expectation Maximization with 3-Dimensional Resolution Recovery, Attenuation, and Scatter Correction. Journal of Nuclear Medcine 51, 921–928 (2010).
Ritt, P., Vija, H., Hornegger, J. & Kuwert, T. Absolute quantification in SPECT. Eur J Nucl Med Mol I 38(Suppl 1), S69–77 (2011).
Menon, B. K., Rao, R. D., Abhyankar, A., Rajan, M. G. R. & Basu, S. Comparative evaluation of 24-hour thyroid 131I uptake between gamma camera-based method using medium-energy collimator and standard uptake probe-based method. J Nucl Med T 42, 194–197 (2014).
Becker, D. et al. Procedure guideline for thyroid uptake measurement: 1.0. Journal of Nuclear Medcine 37, 1266–1268 (1996).
Sukumar Benjamin, R., Amro, A. & El-Desouki, M. I. Measurement of Iodine-123 thyroid uptake using a gamma camera with LEAP collimator. J Nucl Med T 27, 215–219 (1999).
Ke, C. et al. Quantitative Measurement of the Thyroid Uptake Function of Mouse by Cerenkov Luminescence Imaging. Scientific Reports 7, 5717 (2017).
Braverman, L. E. Evaluation of thyroid status in patients with thyrotoxicosis. Clin Chem 42, 174–178 (1996).
Andrade, V. A., Gross, J. L. & Maia, A. L. The effect of methimazole pretreatment on the efficacy of radioactive iodine therapy in Graves’ hyperthyroidism: One-year follow-up of a prospective, randomized study. J Clin Endocr Metab 86, 3488–3493 (2001).
Klein Hesselink, E. N. et al. Effects of Radioiodine Treatment on Salivary Gland Function in Patients with Differentiated Thyroid Carcinoma: A Prospective Study. Journal of nuclear medicine: official publication. Soci of Nucl Med 57, 1685–1691 (2016).
Acknowledgements
The authors are grateful to all the technologists in our department for their support for this publication.
Author information
Authors and Affiliations
Contributions
P.D. performed the statistical analysis. F.P. participated in the design of the study. G.S. participated in the design of the study and carried out SPECT planar imaging studies and analysis. H.C. participated in the design of the study and helped to draft the manuscript. Y.L. conceived the study, participated in its design and coordination, and wrote the manuscript. L.L. participated in the design and coordination of the study. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Li, Y., Cai, H., Shen, G. et al. Quantification of radioactivity by planar gamma-camera images, a promoted method of absorbed dose in the thyroid after iodine-131 treatment. Sci Rep 8, 10167 (2018). https://doi.org/10.1038/s41598-018-28571-y
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-018-28571-y
This article is cited by
-
In-situ x-ray fluorescence imaging of the endogenous iodine distribution in murine thyroids
Scientific Reports (2022)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.